Crossing the Threshold ( FDA Regulatory Requirements for Medical Device Manufacturers) DESIGN CONTROLS FDA 1 Medical Device Design Controls Introduction to the FDA Definitions Classes of devices Design control overview Risk assessment Verification and Validation testing Software Quality Assurance Labeling Post design transfer issues 2 Regulations CBER Biologics •21 CFR 600/601/610 Blood CDRH Devices CDER Drugs •21 CFR 56 (IRB’s) • 21 CFR 56 (IRB’s) •21 CFR 58 (GLP) • 21 CFR 58 (GLP) •21CFR 606 •21CFR 11 (Electronic records) 21CFR 1270, 1271 (tissue) •21 CFRR 800-1050 (devices) 21 CFR 58 (GLP) •21 CFR 807 (510(k)) 21CFR 11 (electronic records) •21 CFR 812 (IDE) •21 CFR 814 (PMA) • 21CFR 11 (Electronic records) •21 CFR 210, 211 (Drug GMP’s) • 21 CFR 312 (IND) • 21 CFR 314 (NDA) •21 CFR 21 CFR 820 QSR (GMP) 3 What is a Medical Device? Type of Product: An instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar related article… Intended use: …for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment or prevention of disease . . . or intended to affect the structure or any function of the body… Mode of action: … and which does not achieve any of its primary intended purposes through chemical action within or on the body or by being metabolized. FD&C Act, §201(h) 4 FDA Oversight in a Medical Device Life Cycle Research FDA review Design and Development Good Clinical Practice Clinical Trial Controls Good Laboratory Practice Investigational Devices Exemptions (IDE’s) Design Controls Good Lab Practices Document Controls Electronic Records Manufacture and Service Quality Systems Requirements Establishment Registration Labeling Controls Design controls Recalls Complaints Medical Device Reporting 510(k) Clearance PMA Document Controls Obsolescence Record Retention 5 12 CFR 820.30 Requirements All Class II and Class III devices, and some Class I devices require design controls. Written procedures required. Procedures are controlled via document control. Information about the design must be readily available to FDA – Design History Files. Design controls can continue through the manufacturing and service phase. 6 Intended use Class I-Simple, Low risk. General controls needed (registration, labeling, GMP) Class II- More complex, Medium risk. Product Class Need 510(k) approval (some exemptions) Class III- Complex, High risk. Generally life support, life sustaining, preventing impairment to human health or unreasonable risk to human life. Premarket Approval (PMA) needed prior to market. 7 Examples Class I Stethoscopes Tung Depressors Reagents used in Clinical Labs Powered Tooth Brushes Dental Chair Class II Catheters Dental Implants Biopsy Needles Ultrasound Imaging System Powered Wheelchair Class III Automatic Defibrillators Artificial Hip Joints Heart Valves Extended Wear Contact Lenses Left Ventricular Assist Devices 8 Quality System A Medical Device Quality System is designed to assure that products are Safe and Effective for their Intended Use and Consistently meet the specifications as defined by results of clinical and/or detailed technical design and validation 9 Design Control Elements 21CFR 820.30 Design Planning Design Input (Requirements) Design Output (Specifications) Design Reviews (Technical) Design Verification (Meets Specifications) Design Validation (Meets clinical needs) Design Transfer (Moves from Design to Manufacturing) Design Changes (Formal Process) Design History File (DHF) 10 Stage-Gate Method Defines phases of project. Uses design reviews and approvals as gates between phases. No .no No No Feasibility and planning OK Design Review OK Design and development OK Design Review OK Verification and Validation OK Design Review OK Transfer to manufacturing Design history file 11 Design Controls General Stage-Gate Process Revise Revise Phase I Feasibility and Planning Revise or Drop Project Drop Phase I Design Review approved Yes Document in Design History File Drop Revise or Drop Project Phase II design review approval Phase II, Design and Development Yes Feasibility/ Technical development Yes System integration Document approval and place deliverables in DHF Hardware development Design Inputcustomer input documents Feasibility assessment Design input specifications Regulatory/Clinical strategy Software development Initial Risk Assessment Preliminary Project Plan Initial Quality Plan Project planning Redo testing Phase III, Verification and Validation System Verification testing System Validation testing Labeling verification/ validation Design Review documentation Update or drop Phase III Design Review Document rationale in DHF Yes Document deliverables in the DHF Final Design Specifications Design calculations, Summary of non-validation testing conducted, System Risk Assessment Final software code as intended to be released for distribution Quality Plan Regulatory strategy Verification and validation test plan Labeling and user manuals Trace matrix between design specifications, risk assessment and verification and validation test plan. Production prototype units for verification and validation testing Bill of materials Quality documentation plan (inspection procedures, Device Master Record, etc.) Design review Yes Phase IV, Design documentation and transfer Final design release Yes Yes Document approval and place deliverables in the DHF Document approval and place deliverables in the DHF Regulatory approvals Design documentation Final failure analysis/Risk Management Labels/labeling Quality Systems Testing plan Design Systems verification test plan/Trace matrix documentation Design product brochures & literature Pilot build test protocol with approvals Reliability Testing Material requirements Manufacturing processes Packaging validation Initial manufacture Sterilization validation of commercial Shelf life testing units Design and manufacturing V&V testing report Prior design review minutes review Essential Requirements Checklist Design History file Technical File or Design Dossier (If required per the Regulatory strategy) Yes Commercial manufacture Completed design of the product as manufactured Approved vendors list Complete DMR Manufacturing process validations as required Final labeling/DFUs Final marketing literature DHF sign off Design review and minutes 12 Design Planning Feasibility Studies Risk Assessments Project Plan Defines Interfaces with Others Stage-Gate Methodology Constantly Changing 13 Design Input-Feasibility Where What Customers Technical Papers Medical experts Service people Intended Use Technical Requirements Safety Issues How Documented Approved Filed Formal Change Control System 14 Risk Assessment Clinical Risk Analysis A compilation of the possible causes of death, serious injury or harm to either patient or user, from a procedural or medical basis, without regard to the specific device used for the procedure, based on actual clinical reference, prior art, published data, or documented experiences of knowledgeable clinical practitioners. These risks are associated with use of the device as detailed in its intended use statement and when used per the device’s instructions for use. Manufacturing Risk Analysis A compilation of the possible causes of death, serious injury or harm to anyone involved in the supply, testing, packaging, shipment, or disposal of a given device, as referenced by prior art, published data, or documented experiences or opinions of person(s) knowledgeable with the device and the methods of supply, testing, manufacture, packaging, shipment, and disposal of the device. Also, the manufacturing risk analysis contains the compilation of possible causes of a device to fail to perform to specifications in its intended use environment for the design life of the product which are caused by limitations or risks in the manufacturing process. 15 Risk Assessment Feasibility Clinical Risk Summary Design Input Preliminary Design Specification Trace Matrix: (links between) Specification Risk Analysis Fault Table Test Plan Preliminary Risk Assessment: Mitigations: Final Risk Assessment: Risk Management Document, Approvals 16 ISO 14971 Risk Assessment Example Example of a Hypothetical Risk Assessment for a Electronic System to Monitor Patient Core Body Temperatures 17 Definitions Harm: Physical injury or damage to the health of people, or damage to property or the environment Hazard: Potential sources of harm Risk: Combination of the probability of Occurrence of Harm and the Severity of the harm Risk Analysis: Systematic use of available information to identify hazards and estimate the risk Residual Risk: Risk remaining after protective measures and mitigations are taken Severity: A measure of the possible consequences of the risk As Low As Reasonably Practicable (ALARP): The residual risk is reduced to a level which is as low as can be reasonable implemented without sacrificing patient safety or clinical utility. The risk/benefit ratio is determined to be acceptable in light of technical feasibility and economic feasibility of implementing additional controls. 18 Classification of Residual Risk Severity Occurrence Detection Risk Quadrants Current Controls Corrective Actions 1 –10 Scale 1- 10 Scale 1-3 Scale 1- Risk, None to Little None required None required 2- Risk, Minimal to Moderate Recommended Recommended Required Required if no existing controls. (ALARP) Redesign * Redesign * 3- Risk, Significant 4- Risk, Serious 19 Rating Effect SEVERITY TABLE Description of Rating 1 None No effect. Device operates as intended. 2 Very Minor Some customers notice defect. Device operates as intended. No effect on patient or clinician 3 Minor Device operates as intended. Slight effect on patient, clinician or user. 4 Very Low Patient comfort or convenience is slightly Reduced but with no patient, clinician or user injury. 5 Low Comfort or convenience is severely Reduced but with no patient, clinician or user injury. 6 Moderate Product is inoperable with no patient or user injury. 7 High Possible transient (reversible) minor injury to patient or user. 8 Very High Transient minor injury to patient or user (possibility of further surgical procedures). 9 Hazardous - 10 Hazardous – Irrevers ible Possibly can contribute to death, severe injury, permanent significant disability or severe occupational illness in patient or user. Can cause irreversible patient or clinician harm. (including for example organ failure, limb loss or death) Example Negligible Marginal Critical Needle stick Exposure to blood borne pathogens Catastrophic 20 Occurrence Rating Probability of Failure Description of Rating DFMEA Failure Rate PFMEA 1 Improbable Failure is unlikely. Failure unlikely. No failures ever associated with almost identical processes. 1 in 1,500,000 (~ 0.000067%) 2 Remote Relatively few failures. Isolated failures associated with almost identical processes. 1 in 150,000 (~ 0.00067%) Isolated failures associated with similar processes. 1 in 15,000 (~ 0.0067%) Generally associated with processes similar to previous processes that have experienced occasional failures. 1 in 2000 (~ 0.05%) Generally associated with processes similar to previous processes that have often failed. 1 in 80 (~1.25%) Failure is almost inevitable. 1 in 8 (~ 12.5%) 3 4 Occasional Occasional failures. 5 6 Probable Repeated failures. 7 8 Frequent Failure is almost inevitable. 1 in 400 (~ 0.25%) 1 in 20 (~ 5%) 9 1 in 3 (~ 33 %) 10 1 in 2 ( 50%) 21 Detection Rating Probability of Detecti on Description of Rating DFMEA 1 PFMEA Almost Certain Design Control will almost certainly detect a potential Cause of Failure or subsequent Failure Mode. Current Controls almost certain to detect failure mode or Cause. Very High Very high chance Design Control will detect Cause of Failure or subsequent Failure Mode. Very high likelihood that Current Controls will detect failure mode or Cause. High High chance Design Control will detect Cause of Failure or subsequent Failure Mode. High likelihood that Current Controls will detect failure mode or Cause. 2 Moderate Moderate chance Design Control will detect Cause of Failure or subsequent Failure Mode. Moderate likelihood that Current Controls will detect failure mode or Cause. 3 Low Low chance Design Control will detect Cause of Failure or subsequent Failure Mode. Low likelihood that Current Controls will detect failure mode or Cause. Remote Remote chance Design Control will detect Cause of Failure or subsequent Failure Mode. Remote likelihood that Current Controls will detect failure mode or Cause. Absolute Uncertai nty Design Control will not detect a potential Cause of Failure or subsequent Failure Mode. No known Controls available to detect Failure Mode or Cause. 22 Quadrant Map Occur rence 10 9 Quad 4 8 7 Quad 3 6 5 Quad 2 4 3 2 Quad 1 1 1 2 3 4 5 6 7 8 9 10 Severity 23 Risk Assessment Table Clinical Risk Assessment Cause ID #Potential Clinical Risk CRA 01 CRA 02 CRA 03 CRA 04 CRA 05 CRA 06 CRA 07 Possible effects Potential causes Patient Core Temperature exceeds physiological limits Patient Core Temperature exceeds physiological limits Patient Core Temperature exceeds physiological limits Severe Hyperthermia (Seizure, Death, Brain Damage) Severe Hyperthermia (Seizure, Death, Brain Damage) Severe Hyperthermia (Seizure, Death, Brain Damage) Probe has a intermittent or “noisy” signal due to EMI in area Probe is loose or disconnected Patient Core Temperature exceeds physiological limits Patient Core temperature exceeds physiological limits Patient Core temperature lower than physiological limits Patient Core temperature lower than physiological limits Severe Hyperthermia (Seizure, Death, Brain Damage) Severe Hyperthermia (Seizure, Death, Brain Damage) Patient enters Hypothermia Patient enters Hypothermia Initial State S O Of effect Of cause or failure Controlling Action's)/ Design Mitigations Score Post Mitigation State . Score (Quad) 18 Q3 Spec Ref S O D Of effect Of cause or failure Of cause or failure Audible and Visual Check Probe alarm EMC testing to UL/IEC 60601-1-2 Requirements Audible and Visual Check Probe alarm 9 2 1 9 2 1 18 Q3 CDS-019 Factory calibration window set for 400 Series thermistor Software Check for probe range Audible and Visual Check Probe alarm Audible and Visual High Temp Alarm 9 2 1 18 Q3 CDS-002 9 2 1 18 Q3 CDS-012 9 6 54 CDS-015 9 6 54 Probe not in 9 calibration window Wrong Temp Probe used 7 63 Infection, Drug reaction, disease state 9 4 36 Patient not being appropriately monitored 9 5 45 Audible and Visual High Temp Alarm Labeling and Training 9 2 1 18 Q3 CDS-015 Probe has a intermittent or “noisy” signal due to EMI in area Probe is loose or disconnected 9 4 36 Audible and Visual Low Temp Alarm Design for EMI immunity 9 2 1 18 Q3 CDS-016 9 6 54 Audible and Visual Low Temp Alarm Audible and Visual Check Probe alarm Design fro interlocking connector 9 2 1 18 Q3 CDS-018 24 Design Output Final design specifications Quantitative Documented Approved Final specifications are contained in the design history file. Final risk assessments completed. Clinical testing may be needed. 25 Design Reviews Formal Process Required for Phase Approval Checklists Minutes Attendees- one not associated with items reviewed Areas covered Action items/open issues Open items closed for final release Formal design review prior to release for manufacture and distribution 26 Design Verification and Validation Demonstrates that all the risks have been mitigated. Demonstrates that specifications have been met. Uses a trace matrix between risk assessment, specs and V&V plans. Clinical trials may be needed to demonstrate safety and/or effectiveness. 27 Design Verification And Validation Verification - meets specification Validation - meets intended use Written procedure required. Testing must be documented, reviewed and approved. Software must be verified and validated. Manufacturing processes must be verified and validated. 28 System Verification and Validation Product Requirements Specification complete System Validation Test Plan generated Software Requirements Specification generated and approved No Plan approved ? Software Validation and Verification Plan drafted Yes Software Development Detailed design No Plan approved ? System integration completed Yes Plan executed Pilot Run completed Yes No Software V & V acceptable ? System Validation and Verification Test Plan executed Software Validation and Verification Test Report generated No System V & V testing OK ? Yes System Validation and Verification Test Report generated Final Design Review 29 Software Quality-Design Controls System Design Specs and System V&V Activities SRS SDS Unit level Risk and SRS trace Software Verification Testing Unit verification activities 30 Design Transfer Design moves from R&D to manufacturing Manufacturing and production specifications are documented Manufacturing risk assessment may be needed Manufacturing IQ, OQ, PQ IQ - Installation Qualification (Equipment) OQ - Operational Qualification( 1st ones meet specs) PQ - Performance Qualification (Consistently repeatable) 31 Design Changes All changes to the design after release must be formally controlled (Change Control). Re-validation may be needed Continues for the life of product. Documentation control system is necessary. 32 Labeling 21 CFR 801 Section 201(k) defines "label" as a: "display of written, printed, or graphic matter upon the immediate container of any article..." The term "immediate container" does not include package liners. Any word, statement, or other information appearing on the immediate container must also appear "on the outside container or wrapper, if any there be, of the retain package of such article, or is easily legible through the outside container of wrapper." Section 201(m) defines "labeling" as: "all labels and other written, printed, or graphic matter (1) upon any article or any of its containers or wrappers, or (2) accompanying such article" at any time while a device is held for sale after shipment or delivery for shipment in interstate commerce. 33 Rx Medical Device Labeling Intended Use Indications for Use Contraindications for Use Warnings, Cautions Description of the Device User Instructions Specifications Corrective Actions (Troubleshooting) 34 Labeling Controls Need for a label or Labeling identified Requirements determined Draft labeling created and reviewed Validation of labeling Translations needed? Select qualified Supplier Final labeling approval via QSP0-0002 ECO process Translations created File and control per QSP0-0001 Documentation Control Process 35 Labeling Verification Labeling must be verified prior to FDA review and product release. Users should also review labeling. Risk assessment “labeling” mitigations must appear as warnings or cautions. 36 Design History File Record of the Development Process Plans Specifications V&V Test Results Design Reviews Changes to the Design 37 Class Exercise-Design Controls Dr. Bright and Dr. Idea have found a novel way to produce a machine to determine if a heart attack patient has additional blockage in the coronary arteries that may be caused by the surgical bypass procedure (CABG). The machine noninvasively measures arterial flow by using Doppler sonar to determine if the arteries are blocked. It can be used in a patient’s home, by itself, on post heart attack patients who may be at risk for additional heart attacks. It transfers the data to a monitoring station at a EMS facility for 24/7 monitoring. They have formed a company (The Bright-Idea Company), built a prototype and tested it in the lab on sheep and pigs. It worked great. Now they want to begin marketing it for use on humans. 1. 2. 3. Is the machine a medical device? What steps should Dr. Bright and Dr. Idea take before they can begin marketing the machine? What documents do they need to have on file? 38 Questions 39