Elemental Analysis of Unburned Matches by SEM/EDS for Forensic Comparison 1 Allen , 2 Neff , 1 Rushton Lance David Catherine G. 1Forensic Science Program and 2Chemistry Department, Marshall University, Huntington, West Virginia Introduction Matches can be found at a variety of crime scenes: arsons, bombings, clandestine labs, and so forth. A technique capable of comparing a match found at a crime scene to a book of matches found in a suspect’s possession would be valuable in associating the suspect to the crime scene. Andrasko1 used SEM/EDS to identify burned matches, and Dixon2 used the fibers in book matches to identify a match originating from a specific book. SEM/EDS analysis has also been used to analyze unburned matches, but not to distinguish one producer from another. Materials and methods Preparation Method 2 A razor blade was used to remove the entire match head from a match. The match head was ground with a mortar and pestle to homogenize the sample. This sample was placed between two glass microtome blades and pressed flat with a minimum pressure of 50 psi. Samples from five different matches of the same book were prepared for two of the match producers. Results A spectrum was collected from five matches from each book for all seven match producer (ex. A1a ─ Ale). The major elements in all the matches were the same (silicon, sulfur, chlorine, and potassium), But spectra were visibly different based on relative amounts of major and trace elements. SEM/EDS Parameters A JEOL 5310 LV scanning electron microscope with an Oxford Pentafet electron dispersive spectrometer detector with thin beryllium window and ISIS analysis package was used. The settings used were: 300 mtorr low vacuum, 25 keV acceleration voltage, 24 mm working distance, and 109 mAmps condenser lens current (spot size). To ensure the spectra reflected the real composition of the matches and not just a few particles in the match, a magnification of 350x was used. A live time of 300 seconds was used as the collection time of each sample. Figure 5. Dendrogram of the PCA analysis of samples prepared by method 2. Figure 2. Dendrogram of the PCA analysis. Shows the amount of dissimilarity between the samples. A large amount of variation can be seen in each set of match samples, thus not allowing for differentiation between one producer and another. Sample preparation method 2 was devised to improve uniformity of the sample, and reduce variation. Sample preparation method 2 was used to prepare more samples for U1 and L1. U1 was chosen because there was a large variability in the samples. L1 was chosen because the sample were similar to U1. It was hoped using preparation method 2 the U1 samples would have less variation and be distinct from the L1 samples. Sample Preparation Matches produced by seven different match companies were analyzed. A list of the companies along with the samples corresponding to those match producers can be seen in Table 1. The samples were prepared by two methods. The second method produced a more homogenous sample. Preparation Method 1 A razor blade was used to cut a flake of the match head from the match. This flake was then placed between two microtome blades and pressed flat with a minimum pressure of 50 psi. Samples from five different matches of the same book were prepared for all seven match producers. Figure 3. Lion Match Company sample (L1b) prepared by a flake of the match head cut off and pressed flat. Although the variation in each of the two sets (U1 and L1) decreased, the U1 samples still were not distinct from the L1 samples. This suggests reanalysis of all samples after being prepared by method 2 would reduce the variability in each set. Conclusions The spectra of some matches showed a visible difference in the amount and types of elements. These differences ranged from large variability in the amount of major elements to the presence or absence of trace elements. Even with the visible differences in many spectra, the variation in each match producer set made comparisons impossible. With the second method of sample preparation the particle size was significantly smaller. Thus, the homogeneity was improved and variations within a set reduced. While it has been shown that even with the second preparation method some matches are not distinct from others, groups of certain match producers may be distinct from other groups of match producers. Future work would include preparation of sample for the other five match producers using the second preparation method. These could then be analyzed by the SEM/EDS. The PCA analysis would then show if groups of match producers differ from other groups. Literature cited Andrasko, J. Identification of Burnt Matches by Scanning Electron Microscopy. J. Forensic Sci. 1978, 23, 637-642. Table 1. Match Producers Samples Atlas Match A1a ─ A1e Atlantis Match AT1a ─ AT1e DDBean & Sons DDB1a ─ DDB1e Diamond Match Company D1a ─ D1e Lion Match Corporation L1a ─ L1j Ohio Match Company O1a ─ O1e Universal Match Corporation U1a ─ U1j Dixon, K.C. Positive Identification of Torn Burned Matches with Emphasis on Crosscut and Torn Fiber Comparisons. J. Forensic Sci. 1983, 28, 351-359. Figure 4. Lion Match Company sample (Lig) prepared by entire match head being ground and pressed flat. Figure 1. Spectra of A1b, D1d, and L1e. Elements present are similar, but relative abundance differs. Energy windows were used to calculate the integral values of eleven different elements. Relative abundance ratios were calculated from the integrals of each element in relation to potassium. These ratios were compared by Principle Component Analysis (PCA). Acknowledgments The particle size is much smaller in the sample ground with mortar and pestle (method 2), figure 4, than in the sample prepared by method 1, figure3. The ratios for the samples prepare by method 2 were used for more PCA analysis. The authors would like to thank Dr Michael Norton for providing accessible and well equipped microscopy facilities, supported in part by NFS COBRE grant (period 2004-2009). We also appreciate the help of Dr. Graham Rankin with the PCA analysis and Dennis Ward with guidance on sample preparation.