SHOULDER ULTRASOUND COMPARISON Paige Fabre 13654584 Shoulder Portfolio: Images in the left column taken March 2014. Images in the center column taken May 2014. Comparison of Ultrasound Images The images below were taken in March and May respectively. In March I had only just started in musculoskeletal sonography and had not completed many examinations. March Examination This examination was conducted with heavy supervision. At this point I would consider myself a novice. As I was very new to the world of MSK my supervising sonographer on this day had to assist me in many areas. I was assisted with probe orientation, image annotation, patient positioning, anatomy orientation and my own ergonomic positioning. Although the images taken were diagnostic, there were several factors that could have been improved. Places for improvement: Confidence – As I was very unsure of myself I often hesitated before taking an image. At times I would scan through a diagnostic image as I was unsure that it was correct. Probe orientation and anatomy knowledge – I had struggled with this throughout the scan as I was not used to scanning rounded anatomy. I was also very confused about my hand position versus the orientation of the tendons. Timing - The scan took almost 30 minutes to complete. This meant that the patient was relatively uncomfortable for an extending period of time. The patient was given several breaks from position holding during the scan. . Image quality – this will be reviewed further in the image comparison. The main areas include depth control, focus, the demonstration of fibres, overcoming anisotropy. May Examination This examination was also conducted with supervision however it was much less than was needed for the March examination. Overall, I believe that my technique has improved such that I would now consider myself as a beginner in this area. As I am still yet to complete and be confident in undertaking the examination on my own I do not believe that I have reached the point of competency. Areas of improvement: Confidence – I have become more confident in my ability to conduct the scan. I believe that this is due to my increasing exposure to this type of ultrasound. At our practice this is a very common scan so there are many people to practice on. Probe orientation and anatomy – With my increased exposure to these scans I have become more aware of the skill required to produce a diagnostic image. I am being trained by a variety of sonographers who have exposed me to their ways of obtaining or remembering how to obtain the necessary images. Timing – This scan took me 15 minutes to complete. I believe that this due to my improved knowledge and skill when conducting the scan. I do however believe that as my training progresses this will change as when I am scanning on my own I will need to make decisions on my own rather than rely on my tutors to agree that an image is acceptable. Image quality – This will again be discussed further below however I believe that many of the above issues have been addressed. Areas for Improvement Pathology recognition – For the most part during the scan I am still relying on my trainers to confirm if my assumptions about the condition of the rotator cuff are correct. This again suggests that I am still at the beginner level and not yet competent. Personal ergonomics – As I have conducted more of these scans I have less awkward with the way in which I conduct the scan. I still however find myself occasionally slumping or over extending/flexing my wrist while concentrating on the scan. Recognition of patients comfort – Although I have decreased amount of time that the patient is in the static position, at times during the scan when I was focusing on the image I would fail to recognise the patient’s fatigue. I think that this is another indication that I am still in the beginning phase and not yet competent in the examination. Paige Fabre 13654584 Shoulder Ultrasound Image Comparison An improvement of factors can be noted in these images. The second image is more magnified with a decreased depth and focal zone more in-line with the biceps tendon resulting in better visualisation. Biceps Transverse Prox Similar to the previous image and improvement in factors have allowed for better visualisation of the biceps tendon. Moreover the bony surface is also more clearly seen in the second image. Biceps Mid In these images of the biceps in long the fibrillar pattern can be seen clearly however the images are taken at slightly different angles. Though the later image has better imaging factors the earlier images has more of the fibres parallel to the beam. This may be due to the difference in patient habitus however it is most likely user error. This is an area I may need to improve on in the future. Biceps Long Prox These images are very similar in comparison. The second image however is more detailed. This is most likely due to the increased attention paid to improving the image and overall visualisation. Biceps Long Dist Paige Fabre 13654584 The images of the subscapularis tendon are where I see the most improvement so far. The technical factors have been improved to allow for better visualisation of the tendons. The transducer position has also been improved in the second image centring on the tendon attachment allowing for better visualisation and confirmation of the attaching fibres. The use of TGC has also improved and is seen as there is fewer artefacts noted the bone. Subscapularis Long The images here share a similar comparison to the above. Artefacts in the image have been reduced and the tendon is viewed more clearly. The fascicles in the tendon are more readily seen in the second image. Subscapularis Trans Paige Fabre 13654584 Supraspinatus Trans Prox Factors have again been improved in these images. These images are a good example in the difference in annotation during the examinations. Different trainers have different expectations of annotation required on the images. This however makes a limited difference in the diagnostic quality of the image itself. These images of the supraspinatus again demonstrate the improvement in the diagnostic quality. Supraspinatus Trans Mid Paige Fabre 13654584 This comparison is a reminder to me that constant machine manipulation is needed to produce images of high diagnostic quality. Supraspinatus Trans Distal The following supraspinatus images demonstrate the improvements made to image quality. As seen in the images of the subscapularis, there has been an improvement in probe position allowing for centring and overall better visualisation of the fibres at the insertion. Supraspinatus Long Ant Paige Fabre 13654584 On review of these images labelling of the mid portion should be included as to avoid radiologist confusion Supraspinatus Long Mid Supraspinatus Long Post Paige Fabre 13654584 These comparative images are a good example of the improvement of imaging improving diagnostic quality. Both images are of symptomatically tender AC Joints however the later image shows the bony surface of the joint more clearly allowing for better visualisation of the joint space. AC Joint Imaging factors have improved the second image diagnostically. The differences in the infraspinatus tendon of the second patent are most likely due to the 25 year age difference between them and the first patient. Infraspinatus Long Insertion Paige Fabre 13654584 Further improvements in technique are seen here however the image of the second patient may have been improved with the use of a lower frequency transducer. Glenoid Labrum SG notch imaging has again seen improvement with the change in imaging factors and technique. The bony element of the notch is markedly clearer in the second image. Spinoglenoid Notch Paige Fabre 13654584 Though factors on both images have been utilised improvement could have been made to both. The second image is slightly overgained. The first image has only one of three focal zones in the region of the bursa. Issues with these images I think are the result of the challenges faced scanning this region dynamically. This is something I will have to work on in the future. SAB in Abduction Paige Fabre 13654584