INFN-LNS DOSIMETRY COMMISSIONING OF THE LNS-INFN PROTON THERAPY FACILITY THE DOSIMETRIC CHARACTERISTICS OF NARROW PROTON BEAMS USED IN EYE THERAPY HAVE BEEN DEFINED BY USING DIFFERENT RADIATION DETECTORS IONIZATION CHAMBERS (parallel-plate,minithimble chambers) SILICON DIODES RADIOCHROMIC AND RADIOGHRAPHIC FILMS THERMOLUMINESCENCE DETECTORS 1) CENTRAL AXIS DEPTH-DOSE DISTRIBUTIONS 2) LATERAL (TRANSVERSE) PROFILES 3) OUTPUT FACTORS (FSDF) TO PROVIDE INPUT TO THE TPS AND FOR M.U. CALCULATIONS. QUALITY CONTROL PROCEDURES DOSIMETRIC COMMISSIONING OF 60 MeV PROTON BEAMS PRODUCED AT LNS FOR EYE THERAPY SHAPED WITH 5 - 30 mm CIRCULAR BRASS COLLIMATORS. INFN-LNS PROTON DEPTH-DOSE DISTRIBUTIONS THE PTW MARKUS PLANE-PARALLEL ION CHAMBER WAS USED IN THE COMMISSIONING AS THE REFERENCE DETECTOR FOR DEPTH-DOSE MEASUREMENT IN PROTON BEAMS (ICRU 59, IAEA 398) THE DESIGN OF THE MARKUS CHAMBER 3 1) Active volume=0.05 cm , 2) Electrode separation=2 mm, 3) Collector diameter=5.4 mm 4) tentrance window=2.3 mg/cm2 5) Exact Location of peff. PROVIDES DEPTH-DOSE DISTRIBUTIONS WITH HIGH SPATIAL RESOLUTION AND HIGH PRECISION • ) PMMA CUP IS PROVIDED TO SEAL THE CHAMBER FOR USE IN WATER • ) A SPACER IS PROVIDED FOR USE IN A SOLID PHANTOM, WHICH ALSO SERVES AS A HOLDER FOR A RADIOACTIVE SOURCE FOR STABILITY CHECKS. INFN-LNS PTW MARKUS PLANE-PARALLER ION CHAMBER IN WATER PHANTOM INFN-LNS 110 PTW- MARKUS ION CHAMBER BRAGG PEAK 100 90 Relative dose (%) 80 70 60 50 40 30 PLATEAU REGION 20 10 INFN - LNS 0 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 Depth in water (mm) CENTRE MAXIMUM RANGE EQUIVALENT ENERGY (mm) (MeV) PEAK PLATEAU RATIO F.W.H.M. (mm) Distal-dose falloff d90%-10% (mm) CATANA 30.60 60.2 4.68 3.29 0.81 CCO^ 31.0 60.4 4.85 3.65 0.80 PSI^ 30.0 60.0 4.47 ^ taken from BJR Supplement 25 (1996) 1.10 INFN-LNS THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beam no.DZA192 1001) INFN-LNS THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beam no.DZA192 1001) 100 Markus LNS 90 Scanditronix LNS Diode Relative Dose (%) 80 70 60 50 40 30 20 10 0 0 5 10 15 20 25 30 35 Depth in water (mm) CYLINDRICAL MINIDIODE SPECIFICATIONS Detector Material: Hi-pSi, high doped p-type silicon (preirradiated for use in proton beams). Detector diameter: 0.6 mm (t = 60 m) Distal-dose falloff (1) Distal-dose falloff (2) d90%-10% (mm) d80%-20% (mm 3.19 0.71 0.50 31.15 3.07 0.80 0.60 31.06 PEAK DEPTH (mm) PEAKPLATEAU RATIO F.W.H.M (mm) MARKUS PTW 30.14 4.68 DIODE SCANDITRON IX 30.07 4.89 DETECTOR PRACTICAL RANGE (d10%, ICRU 59) INFN-LNS MODULATED PROTON DEPTH DOSES FOR EYE THERAPY 110 100 20 mm SOBP RS 1.8 mm 90 Relative Dose (%) 80 14.5 mm SOBP RS 14 mm 70 60 50 11 mm SOBP RS 10 mm 40 30 20 10 0 0 5 10 15 20 25 30 35 Depth in water (mm) MODULATOR RANGE SHIFTER MODULATION (SOBP) DISTAL-DOSE FALLOFF (1) DISTAL-DOSE FALLOFF (2) d90%-10% d80%-20% (mm eye-tissue) (mm eye tissue) (mm eye tissue) MAXIMUM DOSE IN SOBP % BEAM RANGE (90% DISTAL) (mm eye-tissue) MOD: 000/00 RS: 14 mm 14.5 0.91 0.65 101.6 14.38 MOD: 010/02 RS: 10 mm 10.5 0.90 0.60 104 18.27 MOD: 009/02 RS: 1.8 mm 20.30 0.80 0.55 103.9 27.04 CCO 17 0.90 0.75 < 102 INFN-LNS MODULATED PROTON DEPTH DOSES FOR EYE THERAPY 15 MM SOBP vs RANGE SHIFTER 110 100 Relative Dose (%) 90 RS = 10 mm 80 RS = 2 mm RS = 14 mm 70 60 50 40 30 20 10 0 0 5 10 15 20 Depth in water (mm) 25 30 35 INFN-LNS BEFORE EACH TREATMENT PERIOD TREATMENT DEPTH DOSE PROFILES HAVE TO BE VERIFIED. TO VERIFY RANGE MODULATION AND MAXIMUM RANGE OF THE BEAM DEPTH-DOSE MEASUREMENT WHEEL (CCO DIODE SCANNER) WHEEL IS COMPUTER CONTROLLED, AND RELATIVE DOSE MEASURED AS RATIO BETWEEN WHEEL’S DETECTOR AND REFERENCE DETECTOR TOLERANCES 1) [(MEASURED RANGE) / (REQUIRED RANGE)]: 0.2 mm 2) [(MEASURED MODULATION) / (REQUIRED MODULATION)]: [ -0.1 mm // +0.9 mm ] INFN-LNS LATERAL OFF-AXIS BEAM PROFILES 1) LATERAL PENUMBRA: H 2) Field ratio: d80%-20% 90% fieldsize 50% fieldsize 3) L95% 4) SIMMETRY (AREA RATIO): 5) FLATNESS: RT % Sr ABS (a .b) 200% ab Pmax Pmin 100% Pmax Pmin 110 100 90 80 N.O.D.% 70 60 50 40 30 20 10 -20 1.3 -10 mm 0 0 10 20 X (mm) MD-55-2 RADIOCHROMIC FILMS WERE USED IN COMMISSIONING tissue equivalence – dose rate indipendence Linear dose response – high spatial resolution ONLY ONE CALIBRATION FILE IS NEEDED TO EVALUATE FILMS EXPOSED AT DIFFERENT DEPTHS Energy indipendence PMMA Phantom He-Ne Scanning Laser Densitomiter INFN - LNS PROFILE MEASUREMENTS MD - 55 RADIOCHROMIC FILM 120 100 = 25mm Relative dose 80 60 40 20 0 -20 -15 -10 -5 0 5 10 15 20 Distance from central axis (m m ) EXPERIMENTAL SET-UP (=25 mm) SOBP (WIDTH) H L95% (mm) Sr% (simmetry) Rt% (flatness) Lateral penumbra (mm) d80%-20% MOD: 000/00 RS: 14 mm 14 mm zirr.=7 mm 0.92 23 3% 3% 1.40 MOD: 010/02 RS: 10 mm 10 mm zirr.= 12 mm 0.92 23 2% 3% 1.30 MOD: 009/02 RS:1.8 mm 20 mm zirr.= 14 mm 0.93 23 3% 3% 1.15 18 zirr.= ½ SOBP 3% 1.10 CCO CAL 18 zirr.= ½ SOBP 1.50 INFN-LNS KODAK XV FILMS AND SCANDITRONIX DIODE WERE USED IN COMMISSIONING IF CALIBRATION FILES PRODUCED FOR KODAK XV FILMS AT DEPTH OF SOBP ARE USED FOR TRANSVERSE BEAM MEASUREMENTS GAF MD55-2 = KODAK XV FILM MOD 010/02 RS:10 MM Lateral penumbra (mm) H L95% (mm) Sr% (simmetry) zirr.=12 mm d80%-20% KODAK XV 1.30 0.92 22.8 0.50 MD-55-2 1.27 0.92 22.7 0.90 120 Diodo MD55 100 X-OMAT V N.O.D. % 80 60 40 20 0 -25 -20 -15 -10 -5 0 5 X Axis (Water mm) 10 15 20 25 INFN-LNS BEFORE EACH TREATMENT PROTON BEAM PROFILES AT ISOCENTRE ARE TO BE CHECKED IN AIR X-Y DIODE SCANNING COMPUTER CONTROLLED DEVICE PLANNED TOLERANCES 1) LATERAL PENUMBRA (d80%-20%) 1.50 mm 2) BEAM SIMMETRY (Sr ) 3% 3) BEAM FLATNESS: Rt%3% 4) FIELD RATIO: H0.90 EXPERIMENTAL AND INTERPOLATED TRASVERSAL DOSE DISTRIBUTIONS (X Axis) 160 Collimator diameter [mm]: 25 Penombra sinistra [mm]: 1.2 Interpolation step [mm]: 0.1 Penombra destra [mm]: 1.3 R90/50: 0.89922 Omogeneità95% [mm]: 21.7 Simmetria [%]: 0.54297 Rt [%]: 2.7259 140 120 Path: Z:\Misure\Trattamenti Marzo2002\20mar2002\preliminari\profilo11.prf Dose [a.u.] 100 80 60 40 20 Experimental data Interpolated data 0 -30 -20 -10 0 Distance from axes [mm] 10 20 30 INFN-LNS DOSE MONITORING SYSTEM (IN-BEAM DOSE MONITORS) THE PROTON DOSE IS MONITORED BY TWO INDEPENDENT TRANSMISSION UNSEALED ION CHAMBERS, PLACED IMMEDIATELY UPSTREAM OF THE PROTON NOZZLE. TRANSMISSION CHAMBERS HAVE SEPARATE CABLING, BIAS SUPPLIES (800 V) AND CURRENT INTEGRATORS AND ARE ARRANGED AS A REDUNDANT COMBINATION TO TERMINATE THE BEAM. TRANSMISSION ION CHAMBERS ARE CALIBRATED DAILY AGAINST A REFERENCE PARALLEL-PLATE MARKUS ION CHAMBER, LOCATED AT THE ISOCENTRE. COMMISSIONING TESTS 1) PRECISION 1a) SHORT TERM PRECISION: (CV) OF THE RATIO R OF DOSE MONITOR UNITS TO DOSIMETER SCALE READING FOR n=5 CONSECUTIVE IRRADIATIONS OF 15 Gy. MEASURED CV = 0.1% 1b) LONG TERM PRECISION (WEEKLY STABILITY): MEASURED [(cGy/U.M.)WEEK] : 1.5%. 2) LINEARITY THE RATIO V BETWEEN THE MEASURED PROTON DOSE AND DOSE MONITOR UNITS SHALL BE WITHIN 1% AT ALL MONITOR SCALE READINGS, i.e. IN THE CLINICAL RANGE UP TO 15–20 Gy. MEASURED RATIO Vmax 1.005 Vmin INFN - LNS MEASURED DOSE, cGy ( Markus chamber) MONITOR CHAMBER LINEARITY DOSE MONITORING SISTEM TRANSMISSION ION CHAMBER 2400 2200 2000 1800 1600 1400 1200 1000 800 600 400 200 0 R2 = 1 (Dose rate = 51.10 Gy/min.) 0 50000 100000 150000 U.M. SENT 200000 250000 STABILITY OF CALIBRATION (long term ) 103 cGy/U.M. 102 DOSE MONITORING SISTEM TRANSMISSION ION CHAMBER 101 100 99 98 97 0 1 2 3 tim e (days) 4 5 6 INFN-LNS DOSE RATE MONITORING (BEAM INTENSITY MONITOR) THE PROTON BEAM RATE IS MONITORED BY THE VOLTAGE SIGNAL (VSF) PROVIDED BY THE FIRST SCATTERING FOIL, LOCATED IN VACUUM IMMEDIATELY UPSTREAM OF THE EXIT WINDOW. IBEAM (nA) = 0.0412 + 6.0835 VSF (R2=0.9999) (VSF) IS USED BY THE C.S. TO STOP THE BEAM IF THE RATE XCEEDS A PRESET LIMIT. NORMAL TREATMENT BEAM CURRENT 4 nA CLINICAL DOSE RATE =1220 Gy/min. COLLECTED CHARGE (Markus, nC) 1.05 1.04 U.M. =K 1.03 1.02 Alarm current 1.01 1 Zmarkus = 1 mm Full energy beam 0.99 Normal current 0.98 0.97 0 1 2 3 4 5 6 PROTON BEAM CURRENT (nA) 7 8 9 10 INFN-LNS ABSOLUTE DOSIMETRY (BEAM CALIBRATION, cGy/U.M.) FOR REFERENCE DOSIMETRY, AT LNS, A PLANE-PARALLEL MARKUS ION CHAMBER IS USED IN A WATER PHANTOM, BY EXTENDING TO PROTON BEAMS THE FORMALISM OF THE IAEA 381 CODE OF PRACTICE ON THE USE OF PLANE-PARALLEL ION CHAMBERS IN HIGH ENERGY ELECTRON DOSIMETRY. DwQ ( Peff ) M Q N Dpp,air,Q0 Wair Q Wair Q 0 S w,air Q pQ MQ M ( C ) kT ,P k pol ks (sw,air)q (nC) ( TAB. 7.1 ICRU 59) N pp D , air,Q0 M ref cGy ref N Dref,air Pcav Pcelref cGy/nC M pp (Wair/e)cp = 1.031 (ICRU 59) pQ = 1.00 (ICRU 59, IAEA TRS-398, MED.PHYS. 1995 :22) FOR DOSE CALIBRATIONS OF INDIVIDUAL PATIENTS, THE CALIBRATION IS MADE IN A UNIFORM DOSE REGION AT THE MIDDLE OF SOBP, AT ISOCENTRE, WITH THE 25 mm DIAMETER REFERENCE COLLIMATOR ON BEAM DELIVERY NOZZLE. FOR DAILY DOSE CHECKS, A PMMA PHANTOM BLOCK IS USED: z PMMA zwater 0.87 A RADIOACTIVE CHECK DEVICE (90Sr) IS USED FOR OPERATIONAL AND CONSTANCY CHECKS OF DOSIMETRY SYSTEM (1%, NCRP). INFN-LNS AN EXRADIN T1 THIMBLE ION CHAMBER [A-150 TE, 0.05 cm3, NK(60Co)] CAN BE USED AT LNS IN THE WATER PHANTOM, ACCORDING TO ICRU 59 Dw , p Gy M Q nC kT ,P k pol ks 0.6311 Gy 0.9534 1.031 S / water ,air nC BECAUSE OF THE Cavity length AND Outer diameter of the chamber, USE IS LIMITED TO ( ) 15 mm AND (SOBP WIDTH) 12 mm. WORK IN PROGRESS NOW AVAILABLE BY PTW ND,w,60Co FOR T1 EXRADIN AND MARKUS LNS CHAMBERS ADOPTION OF IAEA TRS-398 CODE OF PRACTICE Dw,Q = MQ ND,w,Q0 kQ,Q0 ESTIMATED UNCERTAINTY OF Dw,Q UNDER REFERENCE CONDITIONS (1 S.D.) ICRU 59 (Nk=1%) IAEA TRS-398 (ND,w,Q0 =1%) EXRADIN T1 2.6% 2.1% PTW MARKUS 3.1% 2.5% INFN-LNS PROTON DOSIMETRY INTERCOMPARISON (ECHED, ICRU 59) BEFORE STARTING LNS PROTON THERAPY FACILITY ) INTERCOMPARISON (1) AT CCO (PHYSICA MEDICA, VOL.XV, N.3) LNS (PTW MARKUS, T1-EXRADIN) CCO (A-150 FW-IC18, REFERENCE DOSE) (Dw) [(PTW MARKUS) / (FW-IC18)]: 1% (Dw) [(T1-exradin) / (FW-IC18)]: 1.4% ) INTERCOMPARISON (2) AT PSI (PHYSICA MEDICA VOL.XVII, S.3, PTCOG XXX 1999) 1) CCO (FW-IC18, MARKUS) 2) LNS (MARKUS, EXRADIN-T1, PTW PMMA FARMER) 3) PSI (PTW GRAPHITE FARMER, MARKUS) 4) TERA ( MARKUS, EXRADIN-T1) % = 1.2% max (%) = 3.5% INFN-LNS DOSIMETRY INTERCOMPARISON AT PSI (FULLY MODULATED BEAM) 6 Deviation from the mean (%) 4 2 0 -2 Standard deviation of the mean = 1.13% Difference between maximum and minimum dose = 3.3% -4 -6 -8 MARKUS EXRADIN FARMER MARKUS FWT 683 LNS T1 LNS LNS CCO CCO FWT 725 MARKUS EXRADIN MARKUS FARMER CCO TERA T1 TERA PSI PSI INFN-LNS OUTPUT FACTORS (FSDF) ) AT THE INITIAL CALIBRATION OF CYCLOTRON THE OUTPUT DOSE RATE HAS BEEN MEASURED FOR ALL COLLIMATORS AREA ENCOUNTERD IN THE CLINICAL PRACTICE. ) TO EVALUATE IF THERE ARE SIGNIFICANT DROPS OF DOSE PER MONITOR UNIT WHEN COLLIMATOR DIAMETER DECREASES TO A FEW MILLIMETERS. ) THE MOST RELIABLE RELATIVE OUTPUT VALUES ARE PROVIDED FROM RADIOCHROMIC DETECTORS, ESPECIALLY FOR NARROWEST BEAMS (VATNITSKY). NO SIGNIFICANT DECREASE OF BEAM OUTPUT (cGy/U.M.) FOR COLLIMATOR DIAMETER UP TO 5 mm 110 100 90 Relative output 80 GAF Markus Diode1 Diode2 TLD2 TLD1 70 60 50 40 30 MODULATED BEAM Z = 21 mm WED PMMA PHANTOM 20 10 0 0 2 4 6 8 10 12 14 16 Collimator diameter (mm) 18 20 22 24 26 28 INFN-LNS OUTPUT FACTORS (FSDF) ) FROM EXPERIMENTAL RESULTS WE CAN STATE THE LOWER BOUND OF COLLIMATOR DIAMETER FOR WHICH FRELATIVE OUTPUT IS MEASURED ACCURATELY WITH EACH DETECTOR 110 100 90 Relative output 80 GAF Markus Diode1 Diode2 TLD2 TLD1 70 60 50 40 30 MODULATED BEAM Z = 21 mm WED PMMA PHANTOM 20 10 0 0 2 4 6 8 10 12 14 16 18 20 22 24 26 Collimator diameter (mm) 1) MARKUS CHAMBER CAN BE USED IN PROTON BEAMS WITH 12 mm 2) TLDs AND SCANDITRONIX DIODE CAN BE USED UP TO =8 mm IN CLINICAL PRACTICE AT LNS GAF-DETECTOR IS USED FOR EVALUATION OF OUTPUT FACTOR FOR SHAPED NARROWEST BEAMS 28 INFN-LNS PATIENT DOSES (STRAY RADIATION) (PERSONAL MONITORING, TWO PATIENTS TESTED) A) TWO PERSONAL BADGE-DOSIMETERS WERE PLACED ON THE CHEST OF PATIENTS DURING THE WHOLE PROTON TREATMENT. 1) ENEA FILM BADGE PERSONAL DOSIMETER (, , x) 0.05 mSv 2) NRPB PADC NEUTRON PERSONAL DOSEMETER PATIENTS COMPLETING TREATMENT 0.20 mSv INDUCED DOSE RATE 15 Sv/h (10’) 1.5 Sv/h