EE09 L05 BIO- MEDICAL ENGINEERING MODULE-III Part-II Mohammed Anvar PK Al-Ameen Engineering College AP/ECE Electroencephalography (EEG) • Electroencephalography (EEG) is the recording of electrical activity of brain. • EEG measures voltage fluctuations resulting from ionic current flows within the neurons of the brain. • In clinical contexts, EEG refers to the recording of the brain's spontaneous electrical activity over a short period of time, usually 20–40 minutes, as recorded from multiple electrodes placed on the scalp • EEG measurement are obtained from electrodes placed on the surface of the scalp • EEG potential represent a summation of the action potential of the neuron in the brain • The patterns obtained from scalp are actually the result of the graded potentials on the dendrites of neuron in the cerebral cortex and other parts of the brain • EEG potentials have random-appearing waveforms with peak to peak amplitude ranging less than 10µV to over 100µV and BW 1Hz to 100Hz • Surface or subdermal needle electrode are used • The ground reference electrode is often a metal clip on the earlobe • Suitable electrolyte paste or jelly is used in conjunction with electrodes to enhance the coupling of the ionic potentials to the input of the measuring device • Placement of electrode on the scalp is commonly by the requirements of the measurement to be made • In addition to the electrodes the measurement of EEG requires a readout or recording device and sufficient amplification for the readout devices • Most EEG provide the capability of simultaneously recording EEG signals from several regions of the brain for each signal a complete channel of instrumentation is required • Thus EEG having 16 channels are available • Because of low-level input signals EEG must have high quality differential amplifiers with good common mode rejection • The differential preamplifiers followed by a power amplifier to drive the pen mechanism for each channel • To reduce the effect of electrode resistance changes, the input impedance of the EEG amplifier should be as high as possible-modern EEG input impedance is greater than 10Mohm • The readout in a clinical EEG is a multichannel pen recorder with a pen for each channel • Standard chart speed 30mm/sec but most EEG also provide chart speed of 60mm/sec for improve the detail of higher frequency signal Wave group of normal cortex • Alpha wave- 8 to 13Hz – Recorded mainly at vision region – Disappear when subject is sleep, change when subject change focus • Beta wave- 14 to 30 Hz – During mental activity • Theta wave- 4 to 7Hz – During emotional stress such as disappointment • Delta wave- below 3.5Hz – Occur in deep sleep and premature babies • Gamma wave- 36 to 44Hz – During sudden sensory stimuli EMG measurements • Electromyography (EMG) is a technique for evaluating and recording the electrical activity produced by skeletal muscles • Potential are measured at surface of body near muscle or directly from by penetrating skin with needle electrodes • Surface or needle electrodes pickup the potentials by contracting muscle fibers • The action potential from individual muscle fibers can be recorded under special conditions • The signal is a summation of all the action potential within the range of electrodes. • The overall strength of muscular contraction depends on the number of fibers energized and the time of contraction • The amplifier for EMG measurement must have high gain, high input impedance and a differential input with good common mode rejection • EMG has an oscilloscope read out instead of graphic pen recorder. The reason is that high frequency response is required • Most emg include an audio amplifier and loud speaker in addition to oscilloscope to permit the operator to hear the crackling sound of the EMG • This audio presentation is especially helpful in the placement of needle or wire electrodes into a muscle Nerve conduction velocity • Nerve conduction velocity is an important aspect of nerve conduction studies. It is the speed at which an electrochemical impulse propagates down a neural pathway • The conduction velocity in a peripheral nerve is measured by stimulating a motor nerve at two points a known distance apart • Subtraction of shorter latency(duration) form the longer latency gives the conduction time along the segment of nerve between the stimulating electrode • By measuring EMG down stream , a latency can be determined from the time displayed on the oscilloscope • Thus knowing the separation distance we can determine conduction velocity of the nerve • Conduction velocity, u=(D/(L1-L2)) Respiratory system • Respiration is the exchange of gases in any biological process • Entire process of inhaling from atmosphere, transporting O2 to cells, removing Co2 from cells and exhausting the waste products into atmosphere is called respiration • Circulating blood is the medium by which oxygen is brought to internal environment and by same mechanism Co2 is carried out • Exchange of gases between blood and external environment take place in lungs and external expiration Lung volume and capacities • Pulmonary test are designed for determination of lung volume of capacities 1. 2. 3. 4. 5. 6. Tidal volume(TV)-Volume of air inhaled and exhaled in a single breath Inspired reserve volume(IRV)-the amount air that can be inhaled beyond the tidal volume Expiratory reserve volume(ERV)-amount of air that can be forcibly exhaled beyond the tidal volume Residual volume(RV)-amount of air remaining in lungs even after a forceful maximal expiration Vital capacity- maximal volume that can be exhaled after maximal inhalation, vital capacity = RV+TV Total lung capacity(TLC)-is amount of gas contained in lugs at end of maximal inspiration, TLC=TV+RV+ERV+IRV 7. 8. Inspiration capacity- maximum amount of gas that can be inspired after reaching the end expiratory level, IC=TV+IRV Functional residual volume capacity- volume of gases remaining in lungs at end of expiration level, • • FRC=RV+ERV FRC=TLC-IC Measurements in respiratory system Spirometry • The changes in lung volume has been measured in two – To measure changes in the volume of gas space within the body during breathing by using plethysmographic techniques – Spirometry-involves the measurement of gas passing through the air way opening • For the purpose of testing pulmonary functions frequently implemented directly by electronically integrated output of a flow meter placed at patients mouth( with nose blocked) • Continuously gas passing through the airway opening and to compute the volume it occupied with in the lungs this done by the device –spirometer • Spirometry- measurement of the changes in volume of lungs for testing of pulmonary function • The lung volume and capacities that can be obtained by measuring the amount of gas inspired or expired under a given set of conditions or during a specific interval can be obtained by spirometer • Spirometer composes of a movable bell inverted over a chamber of water • Inside the bell above the water lime is the gas that is to be breathed • the bell is balanced by a weight to maintain gas inside at atmospheric pressure so that its height above water is proportional to amount of gas in the bell • Breathing tube connects the mouth of the patient with the gas under the bell, the nose of patient is blocked • Thus as patient breathes into the tube , the bell moves up and down with each inspiration and expiration is proportion to amount air breathed in and out • A pen attached to balanced weight mechanism and writes on paper attached to drum recorder called kymograph • As the kymograph rotates the pen traces the breathing pattern of patient • Sometimes a rotational displacement sensor is fed to an OP amp which can be connected electronic strip chart recorder • Generally respiratory test are repeated two or three times and maximum values are used to ensure that the patient performed the test to the best of his ability Pneumotachograph • • • • Also known as “differential pressure device” Tube with fixed resistance Contains a bundle of capillary tubes or fine meshes This device utilizes the principle that air flowing through an orifice produces a pressure difference across the orifice that is a function of the velocity of air • The orifice consist of a set of capillaries or a metal screen • Since the cross section of the orifice is fixed, the pressure difference can be calibrated to represent the flow • Two pressure transducers or a differential pressure transducer can be used to measure the pressure difference Gas exchange and distribution • Once air is in the lungs oxygen and Co2 must be exchanged between the air and the blood in the lungs and between the blood and the cells in the body tissue alveoli Measurement of gaseous exchange and diffusion 1. Chemical analysis method– – – – – in this a gas sample of approximately 0.5ml is introduced into a reaction chamber by use of transfer pipet, at the upper end of the reaction chamber capillary an indicator droplet in this capillary allows the sample to be balanced against a trapped volume of air in the thermobarometer Absorbing fluid for Co2 and O2 can be transferred without causing any change in the total volume of the system The micrometer is adjusted so as to put mercury into the system in place of gas being absorbed The volume of absorbed gas is read from the micrometer calibration 2. Diffusion capacity using CO infrared analyzer– – – – – To determine the efficiency of perfusion of the lung by blood and the diffusion of the gases The most important tests are measure O2,Co2,pH and bicarbonate in arterial blood In this trying to measure the diffusion rate of oxygen from alveoli into the blood. Assuming that alveoli have equal concentration of oxygen actually this condition does not exist bcz of unequal distribution of ventilation in the lung Hence we use diffusion capacity or transfer factor used rather than diffusion Diffusion capacity in normal adult - 20 to 38 ml/min/mmHg it varies with depth of inspiration, increases during excise and decreases with low hemoglobin • In this method of measuring diffusion capacity involves the inhalation of low concentrations of CO. ml CO taken up/min TF or diffusion capacity • • • • • • • = Pco in alveoli (mm/Hg) For this measurement as well as for all methods requiring CO determination, a CO analyzer or a gas chromatograph is used The commonly used carbon monoxide analyzer utilizes an infrared energy source , a beam chopper, sample and reference cells, plus a detector and amplifier A milliammeter or a digital meter may be used for display Two infrared beams are generated one directed through the sample and the other through the reference The CO gas mixture flowing through the sample cell absorbs more infrared energy than does the reference gas The two infrared beams are measured by a differential infrared detector The out put is proportional to the amount of monitored gas in the sample cell, the signal is amplified and presented to the output display meter or recorder Measurement of gas distribution • The distribution of oxygen from the lungs to the tissues and Co2 from tissues to the lungs takes place in the blood • Oxygen is carried by hemoglobin of the RBC, Co2 is carried through chemical process in which Co2 and water combine to produce carbonic acid, which is dissolved in the blood • The amount of carbonic acid changes the pH of the blood • In assessing the performance of the partial pressure oxygen(Po2) and PCo2 in the blood , the percentage of oxygen in the hemoglobin and the pH of blood are useful Respiratory and therapy equipment • When a patient is incapable of adequate ventilation by natural process , mechanical assistance must be providedinstrumentation involved in providing mechanical assistance – respiratory therapy • Inhalators– the term generally indicates a device used to supply oxygen or some other therapeutic gas to a patient – Inhalators are used when a concentration of oxygen higher than the air required – The inhalator consists of a source of the therapeutic gas , equipment for reducing pressure and controlling the flow of gas and a device for administrating the gas Ventilators and respirators • Ventilators and respirators are used interchangeably to describe equipment that may be employed continuously or intermittently to improve the ventilation of the lungs and to supply humidity or aerosol medications to the pulmonary trees • The respirators and ventilators are classified as assistor – controllers and can be operated any three different modes 1. 2. 3. in assist mode inspiration is triggered by the patient. A pressure sensor respond to the slight negative pressure that occurs each time the patient attempt to inhale and triggers the apparatus to begin inflating the lungs. In the control mode breathing is controlled by a timer set provide the desired respiration rate. Controlled ventilation is required for patients who are unable to breathe on their own In the assist-control mode the apparatus is normally triggered by the patients attempt to breathe as in the assist mode . However the patient fails to breathe within a predetermined time , a timer automatically triggers the device and inflated the lungs • The various types of ventilators in clinical use two types • Pressure-cycled • Positive pressure assistor-controller Artificial heart valve • An artificial heart valve is a device implanted in the heart of a patient with valvular heart disease. When one of the four heart valves malfunctions, the medical choice may be to replace the natural valve with an artificial valve. This requires open-heart surgery. • Valves are integral to the normal physiological functioning of the human heart. • Natural heart valves are evolved to forms that perform the functional requirement of inducing unidirectional blood flow through the valve structure from one chamber of the heart to another. • Natural heart valves become dysfunctional for a variety of pathological causes. • Some pathologies may require complete surgical replacement of the natural heart valve with a heart valve prosthesis Heart lung machine • A medical equipment that provides Cardiopulmonary bypass, (temporary mechanical circulatory support) to the stationary heart and lungs • Heart and Lungs are made “functionless temporarily” , in order to perform surgeries • • • • CABG Valve repair Aneurysm Septal Defects Heart is Stopped Blood circulated systemically bypassing the heart and lungs Blood diverted through tubes and is pumped to maintain flow Temperature regulation of blood and gaseous exchange is done Parts • Five pump assemblies • Venous Cannula • Arterial Cannula - dual-stream aortic perfusion catheter / meshed cannula • Venous Reservoir • Oxygenators • Heat Exchangers • Cardiotomy Reservoir and Field Suction • Filters and Bubble Traps • Tubing and Connectors Centrifugal Pumps Roller • Centrifugal pumps consist of plastic cones, which when rotated rapidly, propel blood by centrifugal force. • Forward blood flow, varies with the speed of rotation and the after load of the arterial line. • Centrifugal blood pumps generate up to 900 mm Hg of forward pressure, but only 400 to 500 mm Hg of negative pressure. Hence, less gaseous micro emboli. • Centrifugal pumps produce pulse less blood flow • Roller pumps consist tubing, which is compressed by two rollers 180° apart. Forward flow is generated by roller compression and flow rate depends upon the diameter of the tubing, rate of rotation. Roller Pump Impeller Pump Centrifugal Pump Five pump assemblies : • A centrifugal or roller head pump can be used in the arterial position for outside body circulation of the blood. • Left ventricular blood return is accomplished by roller pump, drawing blood away from the heart. • Surgical suction created by the roller pump removes accumulated fluid from the general surgical field. • The cardioplegia delivery pump. • Emergency Backup of the arterial pump in case of mechanical failure. Venous Reservoirs • Reservoirs may be rigid (hard) plastic canisters ("open" types) or soft, collapsible plastic bags ("closed" types). • The venous reservoir serves as volume reservoir • Facilitates gravity drainage, • Venous bubble trap present, • Provides a convenient place to add drugs, fluids, or blood, and adds storage capacity for the perfusion system. Oxygenators Mebranous Bubble Membranous Oxygenators • Imitate the natural lung by interspersing a thin membrane of either micro porous polypropylene or silicone rubber between the gas and blood phases. • With micro porous membranes, plasma-filled pores prevent gas entering blood but facilitate transfer of both oxygen and CO2. • The most popular design uses sheaves of hollow fibers connected to inlet and outlet manifolds within a hard-shell jacket. Bubble Oxygenators • Venous blood drains directly into a chamber into which oxygen is infused through a diffusion plate (sparger). • The sparger produces thousands of small (approximately 36 µm) oxygen bubbles within blood. • Gas exchange occurs across a thin film at the blood-gas interface around each bubble • Produce more particulate and gaseous microemboli are more reactive to blood elements. Heat Exchangers • Control body temperature by heating or cooling blood passing through the perfusion circuit • Temperature differences within the body and perfusion circuit are limited to 5°C to 10°C to prevent bubble emboli Filters and Bubble Traps • In the circuit, micro emboli are monitored by arterial line ultrasound or monitoring screen filtration pressure. • Depth filters consist of porous foam, have a large, wetted surface and remove micro emboli by impaction and absorption • Screen filters are usually made of woven polyester or nylon thread. Tubing • Medical grade Polyvinyl Chloride (PVC) tubing • It is flexible, compatible with blood, inert, nontoxic, smooth, nonwettable, tough, transparent, resistant to kinking and collapse, • Can be heat sterilized • The Duraflo II heparin coating ionically attaches heparin to a quaternary ammonium carrier (alkylbenzyl dimethyl - ammonium chloride), which binds to plastic surfaces. Perfusion Monitors and Sensors • A low-level sensor with alarms on the venous reservoir and a bubble detector on the arterial line are desirable safety devices. • Flow-through devices are available to continuously measure blood gases, hemoglobin/hematocrit , and some electrolytes • Temperatures of the water entering heat exchangers Sterilization : • Ethylene dioxide is commonly used • 4 hours of sterilization at 55°C or 18 hours at 22°C . • Disadvantages of ethylene dioxide , are the toxicity and explosive nature • Disposable tubing ,reservoirs and oxygenator • Steam sterilization as PVC can withstand heat Hemodialysis • In medicine, hemodialysis (also haemodialysis) is a method that is used to achieve the extracorporeal removal of waste products such as creatinine and urea and free water from the blood when the kidneys are in a state of renal failure. • Hemodialysis is one of three kidney failure therapies (the other two being kidney transplant and peritoneal dialysis). An alternative method for extracorporeal separation of blood components such as plasma or cells is aphaeresis. • Hemodialysis can be an outpatient or inpatient therapy. Routine hemodialysis is conducted in a dialysis outpatient facility, either a purpose built room in a hospital or a dedicated, stand alone clinic Cont… •Blood is removed from the body and pumped by a machine outside the body into a dialyzer (artificial kidney) •The dialyzer filters metabolic waste products from the blood and then returns the purified blood to the person •The total amount of fluid returned can be adjusted •A person typically undergoes hemodialysis at a dialysis center •Dialysate is the solution used by the dialyzer Cont… • Waste products (urea, creatinine,…ets) move from blood into the dialysate by passive diffusion along concentration gradient • Diffusion rate depends on; 1. The difference between solute concentrations in the blood and dialysate 2. Solute characteristics 3. Dialysis filter composition 4. Blood and dialysate flow rate Lithotripsy • Lithotripsy is a medical procedure involving the physical destruction of hardened masses like kidney stones, • Lithotripsy (also called ESWL- Extra-Corporeal Shock Wave Lithotripsy) is a method is breaking up kidney stones in the kidneys or tract ,using ultasound shock waves. – Electromagnetic lithotripsy – Electrohydraulic lithotripsy – Endoscopic lithotripsy – Extracorporeal shock wave lithotripsy – Laser lithotripsy Infants incubators • Premature infants are babies born prior to the normal 36 or 37 days, so they are unable to control there temperature with the new environment and also they could have some problems with their respiratory systems so as the heart diseases. • Heat is lost via evaporative, conductive, convective and radiative means as shown in figure (1). What is Infant Incubator • Infant incubator is a biomedical device which provides warmth, humidity, and oxygen all in a controlled environment as needed by the new born. • It can be consider as therapy device. Simple Block Diagram Incubator Types • Transport Incubator • Intensive Care Incubator • Radient Warmer Temperature degree regulator method in Incubator • Linear Method : automatic control (ON or Off) • Proportional control method Proportional control method Temp Control Signal (Turn ON & Off for Heater) Monitoring Respiration Rates Parameters Control in Infant Incubator • Temperature control. • • Air temperature mode Skin temperature mode • Humidity control. • Humidity is defined as the percent of water evaporated molecules in the air. • This is important for an infant baby, because if level of humidity was low the baby's skin will be dry and cause a lot of health problems to him. • Humidity probe: • humidity transducer. Humidity transducers principle is the capacitive changing types that mean any change in humidity will cause a change in capacitive and the change in capacitive will be translated to a change in voltage using bridge circuit. • Oxygen control. Alarms in Incubator • • • • • Air flow alarm (fan stop alarm) High temperature alarm Power failure alarm Probe alarm. Set temperature alarm