Model 6200 Total Sulfur and Total Nitrogen

advertisement
Antek
®
6200
Meet Sulfur Regulations and
Control Your Process with Continuous Measurements
Product Presentation
PAC Refinery Solutions
Your Process.
2
PAC Refinery Solutions
S
S
S
S
S/N
S
S/N
S
S
S
S
S
S
S
Your Process.
Optimized.
Ensure regulatory compliance
Receive actionable total sulfur data
3
Product Description
Antek 6200 Process On-Line Analyzer
Meet strict regulatory standards for sulfur
content in your high-value products
• Total sulfur and total nitrogen analyzer for
gas, LPG, and Liquid samples
• Excellent repeatability and reproducibility
over a wide range even during process
swings
• In tight correlation with primary test
methods
4
Product Description
Antek 6200
Dual-stream and dual-range
analysis
6200
Series
Sulfur UV
6200
Series
Nitrogen
High
Speed
Sulfur
(HSS)
5
Key Advantages
Meet strict regulatory
standards for sulfur
content in high-value
products
Produce lab accuracy
with process robustness
Optimize the process
by keeping close to
the specification limits
Control operational and
maintenance costs
6
Key Advantages
Meet strict regulatory standards for sulfur content
in high-value products
•
•
•
•
Measure Total Sulfur, Total Nitrogen, or both in a single analyzer
Precise measurement of gas, liquid, and LPG products
Direct injection system capable of measuring heavy gas/oil products
Sensitivity from 250 ppb to % levels
7
Key Advantages
Produce lab accuracy with process robustness
• Utilizing Proven Antek technologies with instruments running 24/7 in
process and laboratory globally
• Flexible combustion system performs under various process
conditions
• Customized sample conditioning
systems available for optimal
results for your specific application
• Correlates to ASTM D4629 and
D5453
8
Key Advantages
Optimize the process by keeping as close to the specification
limits as possible
• Quick analysis in 2-5 minutes per stream
• Fast response time to process changes – no carry over from
products
• Immediate and precise results for fast and reliable decisionmaking
• Enables process control by
easy integration with plant DCS
2000
1750
1500
mg/kg
1250
1000
750
500
250
0
9
Key Advantages
Control operational and maintenance costs
• Can use air or oxygen as the combustion gas, depending on the
application
• New pyrotube design extends preventative maintenance efforts
• Reduced cycling intervals extends the life of the injection valve
• Built to withstand process disruptions, and recovers quickly for
minimal downtime
10
Antek 6200 Unique Analysis
Unsurpassed Leadership
•
•
•
•
High speed analysis
Measures all products
Direct injection
Antek’s proven leadership
11
6200 Technology
Analytical Process Flow Path
SAMPLE CONDITIONING
SSR
SYSTEM
Process
SCS
SCS
Temperature
Controller
1000
1000
Supply
Pyro Gas
Furnace
Inlet Gas
Return
Pyro tube
Sample Injection
Valve
Combusted
sample
Flow
Control
Pyro
gas
Inlet
gas
Carrier
Membrane Dryer
Dryer
Dryer Vent
Scrubber
System Vent
Sulfur
Detector
Pressure
Switches
Combustion
Gas
Carrier
Gas
Power
Supplies
Interface &
Control
UV &
HV
13
Theory of Operation
14
6200 Specifications
Specifications
Method Compliance
Certifications/Area Classifications
Performance
Communication Outputs
Sampling
Utility Requirements
ASTM D5453, D6667, D4629, D5176, ISO 20846
ATEX Ex px ia IIC T4 Gb
• Analytical Range: 200 ppb for Sulfur and 250 ppb for Nitrogen (lower
detection limit) to % levels
• Precision: within <1% of full scale, for most applications
• Linearity: <1% of full scale at >1ppm concentrations
• Analysis Time: 2 minutes; analyses over 1000 ppm may take longer
• Cycle Time: 2 to 5 minutes, programmable
• 4–20 mA isolated at 750 Ω (standard)
• Discreet digital inputs and outputs for remote control and status indication
• RS-232, RS-422, RS-485 for Modbus RTU (standard)
• Modbus TCP/IP and Ethernet (optional)
•
•
Injection valve, 2 or 5 µl sample volume;
Sampling system available as recommended option
•
•
•
•
•
Carrier (Argon, Nitrogen*, Air): 99.975%, 50 psig, 5–15 cc/min
Instrument or Plant Air/Oxygen for combustion (application dependent)
Combustion gas (Oxygen, Air): 99.75%, 50 psig, 400 cc/min
Plant Air (purge): clean, dry, particle-free; 80–100 psig
Instrument Air: 80–100 psig, for injection valve operation and for
pneumatic output to sample conditioning panel
* Nitrogen cannot be used as carrier on a 6200 N Nitrogen system.
16
Theory of Operation- Total Sulfur 1
Pyro Oxidation of Sample
•
Sample is introduced into an oxygen rich environment and combusted @ >1000°C.
– Sulfur species converted to SO2;
– Hydrocarbons are converted to water vapor and CO2
CxHy-S-N + O2 (excess) = H2O+ CO2+O2+SO2+NO
Or
R- N + R-S + O2  CO2 + H2 O + NO + SO2
•
Water vapor removed with membrane dryer
– Reverse osmosis
17
Theory of Operation- Total Sulfur 2
Pyro UV fluorescence of Sulfur
•
Oxidized sample is exposure to UV light
– SO2 Is converted to SO2*
– On returning to stable state, energy is emitted as a flash of light which is
measured by a photomultiplier tube
– Amount of energy produces is directly proportional to the number of sulfur atoms
in the sample
SO2 + hv = SO2* = SO2 + hv
18
Theory of Operation – Total Nitrogen
Pyro Chemiluminescence's of Nitrogen
•
Oxidized sample is Mixed with Ozone
– NO Is converted to NO2*
– On returning to stable state, energy is emitted as flash of light which is measured
by a photomultiplier tube
– Amount of energy produces is directly proportional to the number of bound
nitrogen atoms in the sample
NO + O3  NO2*  NO2 + hv
19
6200 Results
Sulfur in diesel sample
Switching from 1800 to 400 mg/kg sulfur
2000
Concentration (mg/Kg S)
1800
1600
1400
1200
1000
800
600
400
200
0
1
3
5
7
9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41
Injections (5 minute analysis)
No sample carry over!
21
Sulfur Gasoline
20 ppm scale 7.5 ppm standard oxygen vs air
Repeatable Measurements!
22
Sulfur Testing Diesel
20 ppm scale 7 ppm standard oxygen vs air
Repeatable Measurements!
23
Measurement Stability
DIESEL SAMPLE WITH 11.8 mg/kg SULFUR INJECTED 3048 TIMES OVER A PERIOD OF 10
DAYS. AVERAGE VALUE OF 11.7 ± 0.16 mg/kg ( 1.4 %RSD).
14
50 ppm full scale
13.5
13
mg/kg
12.5
12
11.5
11
10.5
10
0.00
1.01
2.01
3.02
4.03
5.03
6.04
7.05
8.05
9.06
10.07
Days
24
Sample Conditioning System
A Sample Conditioning System (SCS) Removes Contaminants & Water, Regulates Sample
Pressure, & Controls Sample Flow To Ensure Samples Introduction Under Constant Conditions.
Sample Conditioning System
Cal
Standard
Selection
Cal 1
Cal 2
Stream
or
Calibration
Selection
Stream 1
Fast
Loop
Stream 2
Fast
Loop
Filtered
sample
Stream
Selection
Bypass
flow
Return to Process
Low
Pressure
Header
Flow
Control
Flow
Indication
Sample
To
Analyzer
Sample
from
Analyzer
25
Typical Sample System
26
Summary
• Total sulfur and total nitrogen analyzer for gas, LPG, and
liquid samples
– Analyzes a wide range of sulfur and nitrogen monitoring
applications
– Robust technology
– Agreement with primary test methods
• With its fast and accurate results, the 6200 helps refineries:
– Protect catalysts
– Meet regulations and optimization targets.