Chem 5486, Instrumental Analysis Spring, 2005 • Text: Principles of Instrumental Analysis, 5th Ed., Skoog, Holler, Nieman, Harcourt Brace, 1998 • Lecture MW, 1:00 - 1:50 pm • Lab MW, 2:00 pm - 4:50 • Office hours by appointment • J. O’Brien, 321A Fondren Science • jobrien@smu.edu Lab Schedule • Jan 24 Spectrophotometric Analysis of a Mixture of Absorbing Substances • Feb 23 High Throughput Screening Via Microplate Absorbance Spectrophotometry (Dr. Hua) • Feb 28 Caffeine Analysis by High Performance Liquid Chromatography (A. Humason) • March 21 Gas Chromatography/Mass Spectroscopy Applications (A. Humason) • March 28 Gel Permeation Chromatography (Dr. Wisian-Neilson) • April 4&6 Thermogravimetric Analyzer TGA and Differential Scanning Calorimetry (Dr. Son) • April 20 Characterization of Semiconductors and Tour of Engineering Cleanroom, Junkins Hall Grading • • • • • • Four lab reports (100 points each) 400 pts Three exams (100 points each) 300 pts Final (200 points) 200 pts Oral Presentation 100 pts Total 1000 pts Lab reports are due within two weeks of completion of lab Important Dates to Know • • • • Jan 17 University Holiday Mar 12-20 Spring Break Mar 25 University Holiday Apr 4 Last day to drop a course Course Objectives • To learn about the theory, instrumentation, and applications of instrumental analysis methods • To gain experience acquiring, treating, and interpreting data • To gain experience reading and writing scientific documents and presenting orally • To gain exposure to a wide range of instrumental techniques and the fields associated with them Scope and Relevancy of Analytical Chemistry/Instrumental Analysis • Approximately 66% of all products and services delivered in the US rely on chemical analyses of one sort or another • Approximately 250,000,000 chemical determinations are performed in the US each day – NIST, 1991, from Managing the Modern Laboratory, 1(1), 1995, 1-9. Classification of Analytical Methods • Classical – Also called wet-chemical methods – Separation of component of interest (analyte) from the sample by precipitation, extraction, or distillation – Followed by gravimetric or titrimetric measurement for quantitative analysis • Instrumental – Use of new methods for quantitative analysis Instrumental Methods • Involve interactions of analyte with EMR – Radiant energy is either produced by the analyte (eg., Auger) or changes in EMR are brought about by its interaction with the sample (eg., NMR) • Other methods include measurement of electrical properties (eg., potentiometry) Instruments • Converts information stored in the physical or chemical characteristics of the analyte into useful information • Require a source of energy to stimulate measurable response from analyte • Data domains – Methods of encoding information electrically – Nonelectrical domains – Electrical domains • Analog, Time, Digital • Detector – Device that indicates a change in one variable in its environment (eg., pressure, temp, particles) – Can be mechanical, electrical, or chemical • Sensor – Analytical device capable of monitoring specific chemical species continuously and reversibly • Transducer – Devices that convert information in nonelectrical domains to electrical domains and the converse Selecting an Analytical Method • What accuracy is required • How much sample is available • What is the concentration range of the analyte • What components of the sample will cause interference • What are the physical and chemical properties of the sample matrix • How many samples are to be analyzed Accuracy vs. Precision • Accuracy – Describes the correctness of an experimental result – Absolute error – Relative error • Precision – – – – Describes the reproducibility of results Standard deviation Variance CV Figures of Merit • Precision – Degree of mutual agreement among data that have been obtained in the same way – A measure of the random, or indeterminate error of an analysis – FOM • • • • Absolute standard deviation Relative standard deviation Coefficient of variation Variance Bias • A measure of the systematic, or determinate, error of an analytical method • Bias = µ - xt • In developing an analytical method, sources of bias should be identified and eliminated or corrected for with use of blanks or instrument calibration Standard Reference Materials (SRM) • Provided by National Institute of Standards and Technology (NIST) • Specifically prepared for validation of analytical methods • Concentration of constituents has been determined by – A previously validated reference method – 2 or more independent, reliable measurement methods – Analyses from a network of cooperating labs Sensitivity • Of an instrument or method is its ability to discriminate between small differences in analyte concentration • 2 factors limit sensitivity – Slope of calibration curve – Precision of measuring device Detection Limit • The minimum concentration or mass of analyte that can be detected at a known confidence level • Sm = Sbl + ksbl Dynamic Range • Extends from the lowest concentration at which quantitative measurements can be made (LOQ), to the concentration at which the calibration curve departs from linearity (LOL) • An analytical method should have a dynamic range of at least 2 orders of magnitude Selectivity • Of an analytical method refers to the degree to which the method is free from interference by other species contained in the sample matrix • No method is totally free from interference from other species Calibration of Instrumental Methods • Analytical methods require calibration • Process that relates the measured analytical signal to the concentration of analyte • 3 common methods – Calibration curve – Standard addition method – Internal standard method Calibration Curve • Standards containing known concentrations of the analyte are introduced into the instrument • Response is recorded • Response is corrected for instrument output obtained with a blank – Blank contains all of the components of the original sample except for the analyte • Resulting data are then plotted to give a graph of corrected instrument response vs. analyte concentration • An equation is developed for the calibration curve by a least-squares technique so that sample concentrations can be computed directly Standard Addition Method • Usually involves adding one or more increments of a standard solution to sample aliquots of the same size (spiking) Lab 1: Spectrophotometric Analysis of a Mixture of Absorbing Substances • Purpose is to determine the individual concentrations of a mixture of absorbing substances • Gain experience working with a UV-Vis Spectrophotometer • Practice several analytical techniques • Understand absorbance and application of the Beer-Lambert Law Background: Absorption of Radiation • Absorption – A process in which electromagnetic energy is transferred to the atoms, ions, or molecules composing a sample – Promotes particles from their normal room temperature state (ground state) to one or more higher-energy states. • Atoms, molecules or ions have a limited number of discrete energy levels • For absorption to occur, the energy of the exciting photon must exactly match the energy difference between the ground state and an excited state of the absorbing species Absorption Methods • Absorbance A of a medium is defined A = -log10T = log10P0/P • Beer-Lambert Law is defined A = Єbc b P0 P Absorbing solution of concentration, c Lab Report Write-up • • Introduction to spectroscopy, instrument basics, absorption principles and BeerLambert Law Experimental section – Specific instrumention (www.oceanoptics.com) – Experimental procedures • Results – Abs vs. wavelength spectra – Plots of concentration vs. absorbance, including equations of lines and R2 • Red at λ1 and at λ2 • Yellow at λ1 and at λ2 – Tables • • • • Dilutions Red absorbance by concentration Yellow absorbance by concentration Є values – Equations and unknown concentrations • • Conclusions References Spectrophotometric Analysis of a Mixture of Absorbing Substances Absorbance 1 0.9 0.8 0.7 0.6 0.5 Red 6 ppm 0.4 0.3 0.2 0.1 0 300 500 700 900 Wavelength, nm 1100 Badger Red 0.6 0.5 Abs 0.4 abs at l1 0.3 abs at l2 0.2 0.1 0 0 2 4 6 Conc, ppm 8 10 12 Signals and Noise • Analytical measurements consist of 2 components – Signal – Noise • Signal to noise ratio – S/N = x/s = mean / standard deviation Sources of Noise in Instrumental Analysis • Chemical Noise • Instrumental Noise – Thermal noise – Shot noise – Flicker noise – Environmental noise Signal to Noise Enhancement • Hardware • Software – Ensemble Averaging – Boxcar Averaging – Digital Averaging • Fourier transformation An Introduction to Spectrometric Methods • Spectroscopy – Interactions of various types of radiation with matter • Electromagnetic radiation (light, X-Rays) • Ions and electrons Properties of EMR • Described by means of sine wave – Wavelength, frequency, velocity, amplitude – Particle model of radiation is necessary – Represented as electric and magnetic fields that undergo sinusoidal oscillations at right angles to each other and the direction of propogation • vi = n li • Frequency is determined by source and remains invariant • Velocity depends on medium • Velocity (air or vacuum) = c = 3.00 x 108 m/s = l n Transmission of Radiation • Refractive index – A measure of the interaction of radiation with the medium it travels through hi = c/vi Scattering of Radiation • Small fraction of radiation is scattered as it passes through a medium – Rayleigh Scattering (elastic) • Scattering by molecules with wavelengths smaller than wavelength of radiation • Its intensity is proportional to 1/l4 – Raman Scattering (inelastic) Diffraction of Radiation • All types of EMR exhibit diffraction • Is a consequence of interference • A parallel beam of radiation is bent as it passes a barrier or slit • nl = BC sin q (Bragg Equation) The Photoelectric Effect • Experiments revealed that a spark jumped more readily between 2 charged spheres when their surfaces were illuminated with light • EMR is a form of energy that releases electrons from metallic surfaces • Below a certain frequency, no additional sparks (electrons) are observed • E = hn (Einstein) • eV0 = hn - w • E = hn = eV0 + w Emission of Radiation • EMR is produced when excited particles (atoms, ions, or molecules) relax to lower energy levels by giving up their excess energy as photons • Excitation can be brought about by – Bombardment with electrons – Irradiation with a beam of EMR • Radiation from an excited source is characterized by an emission spectrum – Plot of relative power of emitted radiation vs wavelength or frequency – Types of spectra • Line • Band • Continuum Absorption of Radiation • In absorption, EM energy is transferred to atoms, molecules comprising the sample • Absorption promotes these particles from RT state to a higher-energy excited state • For absorption to occur, the energy of the exciting photon must exactly match the energy difference between the ground state and one of the excited states of the absorbing species Atomic Absorption • Passage of radiation through a medium that consists of monoatomic particles results in absorption of a few frequencies • Simplicity is due to small number of possible energy states for the absorbing particles Molecular Absorption • More complex because the number of energy states is large compared to isolated atoms • The energy, E, associated with the molecular bands: E = Eelectronic + Evibrational + Erotational Components of Optical Instruments • Stable source of radiant energy • Transparent sample container • Device that isolates a restricted region of the spectrum • Radiation detector • Signal processor and readout Sources of Radiation • Source must generate a beam of radiation with sufficient power • Output must be stable for reasonable periods • Radiant power of a source varies exponentially with the voltage of its power supply – Continuum (tungsten) – Line (lasers) Laser Sources • • • • High intensity Narrow bandwidth Coherent nature of output Components – Lasing medium – External source – Mirrors Laser Mechanism • Pumping – Excitation of active species – Causes population inversion • Spontaneous Emission – Random process yielding incoherent monochromatic radiation • Stimulated Emission – Emitted photon has same energy and is in phase with photon that caused the emission – Therefore is coherent • Absorption – Competes with stimulated emission Wavelength Selectors • Narrow bandwidth is required • Filters • Monochromators, consisting of – Entrance slit – Collimating lens (or mirror) – Grating (or prism, historical) – Focusing element – Exit slit Radiation Transducers • Convert radiant energy into an electrical signal • Photon transducers – Photomultiplier tube (PMT) • Contain a photoemissive surface • Emit a cascade of electrons when struck by electrons • Useful for measurement of low radiant power Component Configuration for Optical Absorption Spectroscopy Source Lamp Photoelectric Transducer Sample Holder Wavelength Selector Signal Processor and Readout Atomic Absorption Spectrometry • Most widely used method for determination of single elements in analytical chemistry • Quantification of energy absorbed from an incident radiation source from the promotion of elemental electrons from the ground state • Technique relies on a source of free elemental atoms electronically excited by monochromatic light Sample Introduction in AAS • Flame – Method of supplying atom source – Utilizes a nebulizer in conjunction with air/acetylene flame – Solvent evaporates – Metal salt vaporizes and is reduced to complete the atomization process • Radiation source is a hollow cathode lamp Graphite Furnace AAS • Samples are atomized by electrothermal atomization • Provide an increase in sensitivity and improved safety compared to Flame AAS instruments • Applications Mass Spectrometry • Relies on separating gaseous charged ions according to their mass-to-charge ratio (m/z) • Widely used in conjunction with other analytical techniques Operating Principles • • • • • Sample inlet Sample ionization Ion acceleration by an electric field Ion dispersion according to m/z Identification of ion mass Mass to Charge Ratio • Obtained by dividing the atomic or molecular mass of an ion, m, by the number of charges, z, of the ion • Most ions are singly charged Vacuum Systems • Reduce gas molecules in analysis chamber – Ion detectors need low pressure environment – Don’t want to detect gas in system – Want to detect analyte before it can collide or react with gas in system – Surface analysis: a need for analyzing an unoxidized surface Vacuum Pump Selection • Atmosphere to 10-2 Torr – Roughing pumps • 10-4 – 10-8 Torr – Diffusion pumps – Turbomolecular pumps – Cryogenic pumps • 10-9 – 10-12 Torr – Ion pumps Vacuum Regions • Rough Vacuum – 760-1 Torr • Mid Vacuum – 1 – 10-3 Torr • High Vacuum – 10-3 – 10-7 Torr • Ultrahigh Vacuum (UHV) – <10-7 Torr Vacuum Characteristics at Room Temperature Pressure, Fraction of Atoms Pressure, Torr Number Density Mean Free Path, cm Time for One Monolayer, s 1/1,000 0.76 2.5 x 1016 0.0065 3 x 10-6 1/10,000 7.6 x 10-2 2.5 x 1015 0.065 3 x 10-5 1/100,000 7.6 x 10-3 2.5 x 1014 0.65 3 x 10-4 1/1,000,000 7.6 x 10-4 2.5 x 1013 6.5 3 x 10-3 1/10,000,000 7.6 x 10-5 2.5 x 1012 65 3 x 10-2 1/100,000,000 7.6 x 10-6 2.5 x 1011 650 3 x 10-1 Molecular Absorption • Measurement of Transmission and Absorption • Limitations to Beer-Lambert Law – Concentration – Chemical deviations – Polychromatic Radiation Raman Spectroscopy, Chapter 18 March 2 • Review concepts from Chapter 6, Fig 6-20, and Section 6C-5 – Relaxation processes – Emission processes • Fluorescence • Phosphorescence • Theory and mechanisms of Raman Spectroscopy – 18A1, 18A2, 18B1, 18C – Homework: 18-2, 18-3 Fluorescence and Phosphorescence • Following absorption – Nonradiative relaxation • Loss of energy in a series of small steps • Energy of molecule is conserved – Fluorescence Emission • Excited State analyte molecule returns to the GS producing radiative emission (a photon is emitted) • ~10-5 s – Phosphorescence Emission • Similar to fluorescence but process is > 10-5 s • Due to relaxation from an excited triplet state Raman Scattering • In addition to fluorescent and phosphorescent emission, scattering of photons off analyte molecule occurs • Raman scattering – Inelastic – Scattered photons have a frequency that differs from incident photons • Rayleigh scattering – Elastic – Scattered photons have the same frequency as incident photons History • Effect discovered by Indian physicist C.V. Raman in 1928 • Nobel Prize in 1931 • Not widely used until 1960s when lasers became available – Raman lines are 0.001% of intensity of source Raman vs. Infrared Spectroscopy • Techniques are complementary • Raman spectra can be obtained from aqueous solutions – Main advantage over IR • More detail on Raman vs. IR next week Mechanism of Raman Scattering • Spectral acquisition – Powerful laser source • Visible or near-IR monochromatic radiation • Emitted radiation (Figure 18-2) – Stokes scattering: E = hn – DE – anti-Stokes scattering: E = hn + DE – Rayleigh scattering: E = hn • Wavelength is equal to that of excitation source Units • Wavenumbers (cm-1) are convention – Easy to convert between wavelength and frequency Infrared Spectroscopy, Chapter 16 March 10, 2005 • • • • Theory of Infrared Spectroscopy Components Read Sections 16A, 16B Homework: 16-2 IR Spectral Regions, Table 16-1 Region Wavelength Range, mm Wavenumber Range, cm-1 Frequency Range, Hz Near 0.78-2.5 12,800-4,000 3.8E14 – 1.2E14 Middle 2.5-50 4,000-200 1.2E14-6.0E12 Far 50-1,000 200-10 6.0E12-3.0E11 Most used 2.5-15 4,000-670 1.2E14-2.0E13 Dipole Changes During Molecular Vibrations • IR radiation is not energetic enough to cause electronic transitions • To absorb IR radiation, a molecule must undergo a net change in dipole moment due to its vibrational (or rotational) motion • If n of EMR matches a vibrational frequency of the molecule, a net transfer of energy occurs – Results in change in amplitude of vibration – Absorption of radiation occurs Types of Molecular Vibrations • Stretching – Continuous change in interatomic distance along axis of atomic bond • Bending – Characterized by a change in angle between 2 bonds • • • • Scissoring Rocking Wagging Twisting Simple Harmonic Oscillator • Model which approximates atomic stretching vibrations • Vibration of a single mass attached to a spring hung from an immovable object (Figure 16-3a) : F = -ky Vibrational Frequency 1 vm 2 k m 1 vm 2 k 1 m 2 1 v 2c k m m1m2 m m1 m2 (16-7) k (m1 m2 ) m1m2 5.3 x10 12 k m (16-9) (16-14) (16-8) Vibrational Modes • Linear molecules 3N-5 (number of possible molecular vibrations) • Polyatomic molecules 3N-6 (number of possible molecular vibrations) Infrared Sources • Inert solid electrically heated to 1500-2200K • Nernst Glower – Rare earth oxides formed into a cylinder – Formed into a resistive heating element, 1200-2200K • Globar Source – Silicon carbide rod, also electrically heated, 13001500K – Greater output than Nernst Glower below 5 mm • Tungsten Filament Lamp – Used in near-IR region of 4,000-12,800 cm-1 • Infrared lasers Chromatographic Separations, Chapter 26 March 30, 2005 • • • • 26 A, General Description 26 B, Migration Rates 26 C, Zone Broadening and Column Efficiency 26 D, Optimization of Column Performance – Through 26 D-5, General Elution Problem • Example 26-1, Homework 26-16 and 26-17 • Chapter 28 G through 28 G-3 General Description • In all chromatographic separations, the sample is transported in a mobile phase – Gas, liquid, or supercritical fluid – fundamental classification • Mobile phase is forced through an immiscible stationary phase – Column or solid surface • As a consequence of differences in mobility, sample separates into bands or zones Chromatograms • Plot of analyte concentration vs. time • Positions of peaks on time axis identify components of sample • Areas under peaks provide quantitative measure of amount of each component • Figure 26-4 Migration Rates of Solutes • Distribution constant Amobile ↔ Astationary cs K cm • Retention Factor tR tM k A tM Chromatographic Peak Shape • Similar to normal error or Gaussian curve • Attributed to additive combination of random motions of solute molecules in chromatographic zone • Peak represents behavior of average molecule • Breadth of band is directly related to residence time in column and inversely related to mobile phase velocity Column Efficiency • Plate height 2 LW H 2 16t R • Plate count, N = L/H • Maximum efficiency occurs at minimum H Column Resolution • Resolution, Rs, provides a quantitative measure of its ability to separate analytes 2[(t R ) B (t R ) A ] Rs WA WB