WRS ─1B MICROPROCESSOR MELTING POINT APPARATUS OPERATING INSTRUCTION Please read through these operating instruction before using CONTENTS I. APPLICATION AND FEATURE………………………1 II. SPECIFICATIONS ………………………………….…1 III. WORKING PRINCIPLE …………………………… 2 IV. CONSTRUCTION ………………………………….….3 V. OPERATION …………………………………………...5 VI. MATTERS NEED ATTENTION……………………….8 VII. MAINTENANCE AND INSPECTION………………. .9 VIII. COMPLETE SET OF THE INSTRUMENT……………10 IX. COMMON BREAKDOWN AND HANDING…………11 I. APPLICATION AND FEATURE In according with the definition of physical chemistry, the melting point of substance means the temperature at which the substance changes from solid state to liquid state. In the field of organic chemistry, the measurement of melting point is a basic means of organic chemistry, the measurement of melting point is a basic means of recognizing the nature of substance, and is also one of the important place in the chemical industry and medicine researches.Also, the melting-point meter is a kind of necessary instrument in the production of medicines, perfumes, dyestuff and other organic crystal substances. Microcomputer melting point meter uses the technology of photoelectric detection and temperature display by liquid crystal. Type WRS─2A could measure three sample at the same time. It has the functions of automatic displaying of initial and final melting points, automatic recording of the melting curves and calculating average value. The temperature system uses the linearly corrected platinum resistor as the detecting element, and additionally, it uses the integrated circuits to realize the speedy setting of eight-step linearly heating rate for being selected .Since the initial and final melting-point readings can be stored up automatically, there is no 2A need for this function to be monitored manually. Type WRS—2 melting point meter uses the capillary stipulated by the pharmacopoeia as the sample tube. II. SPECIFICATIONS Measuring range of the melting-point: room temperature ~300℃ Setting rate of the “starting temperature”:50℃~300℃ no more than 3 min 300℃~50℃ no more than 5 min Setting accuracy of the “starting temperature”:±0.8℃ Minimum reading of the digital temperature display: 0.1℃ Rate of the linear heating:0.2℃/min, 0.5℃/min, 1℃/min, 1.5℃/min, 2℃/min, 3℃/min, 4℃/min, 5℃/min Deviation of linear heating rate: no more than 10% (setting value) Accuracy of measuring the melting-point: <200℃±0.5℃ 200℃~300℃±0.8℃ Repeatability: 0.3℃(rate is 1℃/min), 0.2℃(rate is 0.2℃/min) Size of the capillary: φ1.4mm(outside diameter) φ1.0mm(inside diameter) Loading height of the sample: 3mm Power source: 220V±22V, 100W,50Hz Dimensions: 398mm×278mm×210mm Weight(net): 12.5kg RS232 interface: band rate 9600,1 bit stop, 8bits data 1 III. WORKING PRINCIPLE The working principle of the instrument is based on the following fact: when a substance is in its crystal state, it reflects the light, when it is melted state, it transmits light. Therefore, in the process of melting, a transition of the transmissivity will occur with the rise of the temperature. Transmissivityδ B A Ta Tb Temperature T Fig.1 Fig.1 shows a typical melting curve (the temperature-transmissivity curve). In Fig. 1, the temperature ta(to which point A is corresponding) is called initial melting point; and the temperature tb (to which point B is corresponding) is called final melting point (or full melting point). The length of ta- tb is called melting range (or melting interval). Type microcomputer melting point meter uses photoelectric method to detect the change of the melting curve automatically. When the temperature reaches the initial or final melting point, the data will be displayed and stored until next measurement. Fig.2 shows the principle of the instrument. After going through the optical fibre, the light, given out, by the incandescent lamp through the transmissive holes of the electric furnace and the capillary tube-socket, and focuses on the inside of the capillary. The light penetrating the melting sample is received by the silicon photoelectric cell. When the temperature rises, sample is melting, the light flux increase, then display the temperature of initial, final melting point and curve after microcomputer processed the data. Temperature test adopts the platinum resistance as detector inserting capillary tube-socket bottom directly, the signal pass through linearity correction, voltage amplification, set to A/D converter, keyboard into heating rate. D/A converter digital (setting temperature) analog voltage (Vg), adds another analog voltage (Vto), its deviation signal through a regulator drive heating execution. When the actual temperature signal of the furnace is higher then the analog temperature, the cooling fan will be turned on, so that the temperature will begin to decrease. When the actual temperature is lower than the analog one or does not reach the desired starting temperature, the heating current increase. With the system, the furnaces’s following function can be realized. 2 Capillary Photoelectric Cell Photoelectric Cell Photoelectric Amplifer Heating exection Platinum Resistance Light Source Cooling Fan High Accuracy Constant Current Melting Curve Meltin Point detection + Vt0 Σ - D/A A/D A/D Vg Microcomputer Temperature Display Linearity Correction Set and heat Fig.2 Remark: WRS—2 only one sample IV. CONSTRUCTION 2 Insert Capillary 1 LCD 3 Reset reset 4 Keyboard Fig .3 Fig.3 shows the front view of the instrument. 3 operating presentation Start_T 75 Now_T 75 ℃/min Up_Rate 1.0 ℃ δ final melting point temperature initial melting point temperature T Smelt 79.8 Fmelt 80.6 n=3 n=3 Fig.4 Smelt: average of initial M.P. Fmelt: average of final M.P. Operation presentation: 1) “E123”—not insert capillary press “heating” key or starting temperature too high 2) ↑— heating 3) ↓— cooling 4) press “ CAL” (calibration) key: i “OK C”—if the measurement is valid (three final M.P. of WRS—2A are intensive or curve of WRS—2 is continuous), press “OK”; if the measurement is invalid(final M.P. is disperse or curve is not continuous),press “C”. ii “A”—calibrating 5) “123C”—press “CLR”(clear)key,123 mean number of the curve Fig.4 shows the LCD of the instrument 7 8 9 4 5 6 1 2 3 0 HEAT reset CLR CAL Fig.5 Fig.5 shows the operation of keyboard 4 Power switch Fuse 3.15A Fuse 1A RS232 Interface Electric outlet Fuse socket(1) 4 Fuse socket(2) Fuse(BGXP-1A-20) Fuse(BGXP-3.15A-20) Fig.6 Fig.6 shows the back view of the instrument RS232 Interface 1 connect RS232 RX point 2 connect RS232 TX point 3 the earth shielded teminal Fig.7 Fig.7 shows the RS232 interface V. OPERATION A. Routine measuring of the melting point i. Turn on the power switch (see l in Fig.6), screen display the starting temperature and rate of the last time. After stabilized for 20 minutes, the cursor stop at “Start_T”( starting temperature see left first line in Fig.4),keyboard into or modify the “Start_T”, then press “ ”key. After enter the temperature, the cursor moves to “Up_Rate”(heating rate see right first line in Fig.4). ii. Keyboard into rate, then press “ ”key. After that, the cursor returns to “Start_T”. iii. Using “ ”key can move cursor to the position you want to modify. iv. When the actual temperature of furnace arrived setting temperature, 5 stabilized, insert the sample capillary (WRS—2A one, two or three capillary, WRS—2 only one capillary). v. Press “HEAT”(heating) key, operating presentation “↑”(see right second line in Fig.4), the apparatus begin to measure the sample. (If you forget to insert the sample capillary, press “heating” key, screen will display En, “n” shows the number of furnace, not display “↑”.) vi. When arriving the temperature of initial or final melting point, screen will display the value and curve. (The meaning of data please see Fig.4) vii. If not cut off power source, the data will be stored. Remark: Loading quality of the sample will affect the accuracy of measuring reading. The curve appearing wave hollow or discontinuous long distance is result from bad loading. viii. If the user want to measure other sample, input “Start_T”, press “ ” key, this moment the instrument will clear the curve automatically and begin to measure the new one. ix. Use “CLR” key: after finishing measurement every time, the apparatus display three curves. For some factors, you don’t satisfy the result. You could clear the curves and measure again. Operation as following: ① Press “CLR”key, operating presentation “1 2 3 C”. ② Input the number of curve you want to clear. Press “CLR” key again, finish the operation of “clear” Remark: from left to right, the number is 1, 2, 3; clear one curve, the average value will change. ③ After inputting the working parameters, insert the sample capillary ( need not insert the capillary into furnace you don’t measure this time), then press “HEAT” key. ④ WRS—2A display “En”, “n” shows the number of furnace you not measure. x. Use “CAL” key: long time use or season changing could result in larger error of measurement. The user should calibrate the instrument in order to ensure the precision. Operation as following: ① Press “CAL” key, display as Fig.8. 6 Start_T 75 Now_T 75 Up_Rate 1.0 Adjust δ ℃ 0-------Default 1-------Reset T Smelt Fmelt Fig.8 ② Press “0” key, recover the factory default value, press “1” key, set again. Display as Fig.9. Start_T 75 Now_T 75 Up_Rate 1.0 Adjust δ ℃ Temp1 Temp2 Temp3 T Smelt Fmelt Fig.9 a) The user could set the final melting point of standard sample according to own demand (one point, two points or three points calibration all could), press “ ”key, the apparatus will measure as above method. (If the temperature is less than 100 ℃ , you should input “0” at hundred digit, for example “80.6”, input “080.6”.) b) When finishing measure one sample, operating presentation “OK C”. Three final M.P. of WRS — 2A are intensive or curve of WRS—2 is continuous, press “ ” key; if the final M.P. is disperse or curve is not continuous, press “clear” key and measure again. c) Operating presentation “A”, enter calibrating state. “A” disappear, finish calibrating. ③ Press “CLR” key again, exit calibration state. 7 Remark: when calibration, setting temperature should 5℃ lower than standard sample’s final M.P; the rate of heating is 1℃/min. The accessory standard samples have sealed up. Take off the seal when using. xi. Use “RESET” key: press this key on top right corner when the instrument is off work or refresh the operating state. B. Connect with computer i. Use RS232 cable connect melting point meter and computer. ii. Insert the disk with the meter into computer. iii. When computer enter win9X, executing WRS—2A program(detail in file Readme). VI. MATTERS NEED ATTENTION i. ii. iii. iv. v. The sample must be baked in accordance with the requirements, and pulverized in a dry and clean crushing bowl. Use the“free-falling” method to knock the capillary, so as to make a firm loading of the sample. On the other hand, loading height of the sample must be identical. After being turned on, the temperature of the furnace will be rise to the starting temperature of last time. If the temperature is higher or lower than the desirable temperature, keyboard can be used to set speedily. Don’t set up the starting temperature exceeding the application range of the instrument. If not, the instrument will be destroyed. Different starting temperature of some samples will affect the result of measurement. So certain operational standards should be fixed.If the linear heating rate of “1℃/min” is selected, the starting temperature should be lower than the melting point by 3-5℃; and if the linear heating rate of “3℃/min” is chosen, the starting temperature should be lower than the melting point by 9-15 ℃ . Usually, the optimum measuring condition should be decided in accordance with the experiments. Since different linear heating rates will lead to inconsistency results of the measurements. It is necessary to set up certain standards. The higher the rate is, the greater the reading value will be. Thus, the melting point reading value of each step of rate can be unified by use of the experimental correction value. Before precisely measuring those samples (the melting point values of which are not known), the speedy heating or high rate may be used first to obtain the approximate melting point value. 8 vi. vii. viii. ix. x. VII. If there is a standard sample, it can be measured first. According to the requirements, with the same starting temperature and heating rate to perform comparison measurement. The initial and final melting point readings of standard sample can be used as the basis for check. If you have a melting point standard powder, the standard powder can be used as the temperature transmission standard; and the results can be corrected in accordance with the nearby readings of the standard powder. The sample to be measured had better to be loaded in five capillaries at a time. After being measured respectively, the mean value of the three intermediate readings is considered as the result of the measurement, so that the accidental error induced by the preparation of sample and loading of capillaries can be eliminated. After measuring the higher melting point sample, measure the lower, you could input its “starting temperature” directly and press “cooling” key. For a few colored sample with poor melting characteristics or the completed carbonized samples, the plastic cut -slice and micro samples can be measured by type WRX-1S thermal values analyzer with microscope. The high molecular products such as polymer can be measured by type WQD-1A dropping point and softening point apparatus. Before inserting the capillary into the instrument, the stains and dirt on the outface should be wiped off with a piece of soft cloth, otherwise, the dirt will be accumulated on the bottom of the socked after several times, and this will lead to the failure of the test. MAINTENANCE AND INSPECTION i. ii. iii. The instrument should be used in a dry and ventilated room. The instrument should be prevented from touching any water by all means, and it should be kept from being moistened. It uses a three cores power supply plug. The ground terminal should be connected to the earth, which can not be replaced by neutral line. The capillaries used in the instrument must be the products provided by our factory. Never use the capillaries drawn manually. They are easy to break. The capillaries used in precision measurement should be picked and chosen. When the capillary is broken in the tube socket, the power source must be turned off first. After the furnace is cooled down, insert a piece of 1mm drill (an accessory) into the broken capillary. Then lift 9 the drill slowly, so as to take out the capillary. If there are some broken glasses in the tube socket, you can pull tube socket out, glasses can be poured out or knocked out. If also not clear completely, you can use the drill to break glasses, then pour out them. Put the tube socket to the former position in accordance with the original direction (the notch of the furnace). Finally, the instrument should be checked. If having some problem, please connect with our repair department. iv. when measuring, display “En”, maybe the following reason: i) Starting temperature too high ii) Forget inserting the capillary with sample iii) Bad loading v. Accuracy test: the accuracy testing should be carry out with three samples to be tested with heating rate 1.0℃/min, such as naphthalene (final M.P. 81.00℃), adipic acid (final M.P. 152.89℃), anthraquinone (final M.P. 285.96℃) Are applied. The measurement should be carried out for five times respectively, with starting temperature’s being set less 5℃ than final M.P.. The testing result can be found from the mean value of the remainder three intermediate reading which has been deleted the minimum and maximum. The deviation would be coorrespondent stipulated accuracy. vi. Repeatability test: under the specified heating rate 0.2℃/min and 1.0 ℃/min, the same conditions of the sample preparation, the measurement of the melting point standard powder (sulfanilamide) is performed for five times. If the standard deviation (℃)is calculated in accordance with the following formula, the deviation will be less than the specified tolerance: x i x n S= 2 i 1 n 1 Where: x i ─ the measured value (℃) of each time; x ─ the mean value (℃); n ─ times of the measurement(=5) vii. Calibration: see step “V/x” Beg to inform user: i) Check final M.P. value of standard periodically, if no obvious change occur, need not indicating value calibration ii) Dry samples strictly, fill it in capillary firmly, otherwise, the calibration value will be beyond the scope of standard. 10 VIII. COMPLETE SET (REEER TO PACKING LIST) IX. COMMON BREAKDOWN AND HANDING Breakdown Appearance ANALYSES FOR REASON Turn on the instrument, No power source. no display. The fuse is broken. Others. Value not accurate. Parts are broken. The change of environment Is large Computer no display The lines connected with RS232 are not correct. The program is wrong Handing method Connect with power source. Change fuse. Send to our repair department. Calibrate again. Check the lines and weld well Send to our repair department. 11