Human Papillomavirus (HPV) DNA Detection with Genotyping on Roche cobas-4800, SurePath Specimens Performance Characteristics Legacy Laboratory Services performed validation studies of SurePath liquid based cytology specimens for HPV Genotype testing with the Roche cobas® HPV Test. The Roche method is FDA approved (IVD) on the cobas-4800 for testing pre- and post-aliquot ThinPrep (TP) specimens. SurePath (SP) samples were evaluated using an aliquot removed from the original SP vial after processing for Pap testing. Method Validation Study Results Patient Correlation Study Seventy (70) post-aliquot SurePath patient samples were tested for HPV using both Roche’s Cobas-4800 HPV Test and Qiagen’s Hybrid Capture-2 High Risk HPV DNA Test (Digene HC2). Of the 70 samples tested, 40 were Digene negative and 30 were Digene positive. All of the negative samples correlated. In contrast, 7 of the 30 Digene positive samples were negative on the Cobas-4800. Details are shown in tables 1, 2, and 3. The overall concordance for all samples was 90% (63/70), and concordance of positive samples was 76.7% (23/30). See tables 1-3. Table 1. HPV testing results, Cobas-4800 versus Digene. Cobas-4800 Digene HC2 Pos Neg Pos 23 0 Neg 7 40 Page 1 of 10 Table 2. Concordance of HPV testing by Cobas-4800 versus Digene method Samples included Number Cobas-4800 Concordance with Digene Total Samples 70 90% Digene HPV Negative 40 100% Digene HPV Positive 30 76.7% Table 3. Digene Positive Samples, Results on Cobas-4800 Number Percent Other HR+ HPV16+ HPV18+ 17 56.7% 1 3.3% 2 6.7% HPV-16 & Other+ 2 6.7% HPV-18 & Other+ 1 3.3% Total Positive Negative 23 76.7% 7 23.3% All 7 samples that had discrepant results were submitted to the Roche Molecular System Laboratory, in Pleasanton, CA for adjudication testing using a Linear Array HPV Genotyping test plus repeat testing on the Cobas-4800. Table 4 gives a list of the HPV subtypes that are detected by these different HPV methods. The Digene and Cobas-4800 methods detect the same subtypes except for HPV-66 which is not detected by Digene. The Linear Array method detects all the subtypes detected by Digene and Cobas-4800, plus an additional 23 subtypes, which are currently considered to be low risk HPV subtypes. Table 4. Comparison of Digene, Cobas-4800 & Linear Array HPV Genotyping Method Digene Cobas-4800 # Subtypes Detected 13 14 Linear Array 37 HPV Subtypes detected 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 68 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, 68 6, 11, 16, 18,26, 31, 33, 35, 39, 40, 41, 45, 51, 52, 53, 54, 55, 56, 58, 59,61, 62, 64, 66, 68, 69, 70, 71, 72, 73, 81, 82, 83, 84, IS39, CP6108 The adjudication testing confirmed 6 out of 7 of Legacy’s Cobas-4800 results. One sample (#6) gave an “invalid” result by both Linear Array and repeat Cobas-4800 testing, thus it could not be included in statistical analysis. Of the 6 other samples, 4 were negative by Linear Array and 2 were positive for low risk HPV subtypes that are not detected by either Digene or Cobas-4800. These results confirmed Legacy’s Cobas-4800 results. Table 5 summarizes these findings. Based on these findings, all of Legacy’s Cobas-4800 “discrepant” test results were judged to be true negatives (excluding sample #6 that was “invalid”). Page 2 of 10 Table 5. Summary of adjudication testing Legacy Digene Legacy Cobas-4800 Roche Lab # HPV High Risk Other High Risk HPV-16 HPV-18 Roche Linear Array Roche Cobas 4800 1 Pos Neg Neg Neg Neg Neg 2 Pos Neg Neg Neg HPV 70 Pos Neg 3 4 5 6 Pos Pos Pos Pos Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Invalid Neg Neg Neg Invalid 7 Pos Neg Neg Neg HPV 84/53 Pos Neg Table 6. Sensitivity & Specificity of Cobas-4800 HPV Genotyping with SurePath samples True Pos False Pos True Neg False Neg Total Sensitivity Specificity Accuracy Cobas-4800 23 0 46 0 69 100% 100% 100% Digene 23 6 40 0 69 100% 87% 91% Sensitivity = TP/(TP+FN) *100% ; Specificity = TN/(TN+FP)*100% Accuracy = (TN+TP)/(TN+TP+FP+FN)*100% These studies suggest that the discrepancies between the Digene HC2 and Cobas-4800 methods are related to a combination of causes, including cross-reactivity on Digene with low risk HPV subtypes (table 5, samples 2 & 7). Other discrepancies might be related to differences in cut-off sensitivity settings for determining positivity. Page 3 of 10 Precision Study For each viral genotype (HPV 16, 18 and 68), a negative SurePath patient pool was spiked with each positive type and divided into 10 aliquots. Each aliquot was run in duplicate daily for 10 days. Legacy's results showed a positive precision of 98.3% (59/60) and a negative precision was 100% (20/20) which concurs with Roche’s results for precision studies performed with ThinPrep samples. Table 7. Positive Precision Study Results Run # Run Date 1 12/2/2013 HPV 16 Neg/Pos Ct HPV 18 Neg/Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos Neg Pos Pos Ct HPV 68 Neg/Pos Ct 37.4 Pos 36.9 Pos 36.1 36.8 Pos 37.9 Pos 36.5 2 12/3/2013 37.3 Pos 36.4 Pos 37.4 36.7 Pos 36.8 Pos 36.3 3 12/4/2013 38.0 Pos 37.0 Pos 37.1 38.0 Pos 36.7 Pos 36.9 4 12/5/2013 37.2 Pos 37.0 Pos 36.8 38.0 Pos 36.9 Pos 37.0 5 12/6/2013 37.6 Pos 36.9 Pos 37.1 37.6 Pos 36.7 Pos 36.4 6 12/9/2013 38.5 Pos 37.5 Pos 38.5 38.7 Pos 37.4 Pos 37.9 7 12/10/2013 37.1 Pos 36.9 Pos 37.2 38.0 Pos 37.3 Pos 37.1 8 12/11/2013 38.1 Pos 36.3 Pos 37.3 39.8 Pos 36.4 Pos 38.2 9 12/12/2013 40.2 Pos 36.7 Pos 37.4 na Pos 37.0 Pos 37.8 10 12/13/2013 38.6 Pos 37.0 Pos 38.4 38.9 Pos 37.8 Pos 38.5 Average Ct 38.0 37.0 37.3 SD 0.94 0.43 0.73 CV% 2.46 1.16 1.95 *Ct = Crossing Threshold; HPV16 Ct positive cutoff = 40.5; HPV18 Ct positive cutoff = 40.0; HPV68 (High Risk pool) Ct positive cutoff = 40 Page 4 of 10 Table 8. Negative Precision Study Results Run # 1 Run Date 12/2/2013 2 12/3/2013 3 12/4/2013 4 12/5/2013 5 12/6/2013 Neg/Pos Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Run # 6 Run Date 12/9/2013 7 12/10/2013 8 12/11/2013 9 12/12/2013 10 12/13/2013 Neg/Pos Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg Analytical Sensitivity Using a purchased 16, 18, and 68 high risk HPV, ThinPrep media and SurePath media were spiked with the same volume of virus suspension. Various dilutions of each were compared to see if the end-point (negative result) for SurePath was equal to or higher than the ThinPrep. This was done to determine if the SurePath specimens are at least as sensitive as the ThinPrep. The results are displayed below in tables 9-11. Samples prepared in SurePath media had the same (HPV-68) or better sensitivity (HPV-16 & 18) than samples prepared in ThinPrep media. Table 12 summarizes these findings. Table 9. Comparison of ThinPrep (TP) and SurePath (SP) for HPV Genotype 16 Dilution x16 x16 x16 x32 x32 x32 x64 x64 x64 x128 x128 x128 x256 x256 HPV 16 Crossing Threshold (Ct), Positive cutoff = 40.5 ThinPrep (TP) SurePath (SP) Pos/Neg Ct Pos/Neg Ct Mech Fail* NA Pos 33.7 Mech Fail NA Pos 33.8 Pos 34.9 Pos 35.1 Pos 35.4 Pos 34.7 Pos 35.0 Pos 34.9 Pos 35.9 Pos 36.0 Pos 36.4 Pos 35.8 Pos 36.7 Pos 35.9 Pos 38.1 Pos 36.6 Pos 37.7 Pos 36.9 Pos 37.4 Pos 36.7 Pos 38.3 Pos 38.0 Pos 39.5 Pos 37.1 Page 5 of 10 x256 Pos 39.1 Pos 37.1 x512 Pos 39.4 Pos 40.1 x512 Pos 39.3 Pos 39.0 x512 Pos 40.2 Pos 38.4 x1024 Neg NA Pos 40.1 x1024 Neg NA Pos 39.4 x1024 Pos 40 Pos 39.7 *Note: “Mechl Fail” - instrument failure; not a sample failure. Recovery for this error is to repeat testing of the sample affected. Table 10. Comparison of ThinPrep (TP) and SurePath (SP) for HPV Genotype 18 HPV 18 Crossing Threshold (Ct), Positive cutoff = 40.0 ThinPrep (TP) SurePath (SP) Dilution Pos/Neg Ct Pos/Neg Ct x16 Pos 37.1 Pos* 35.7 x16 Pos 37.7 Pos 35.9 x16 Pos 37.0 Pos 36.1 x32 Pos 38.5 Pos 37.0 x32 Pos 38.7 Pos** 36.8 HPV 18 continued Crossing Threshold (Ct), Positive cutoff = 40.0 ThinPrep (TP) SurePath (SP) Dilution Pos/Neg Ct Pos/Neg Ct x32 Pos 38.5 Pos*** 36.8 x64 Pos 38.9 Pos 37.6 x64 Pos 39.6 Pos 37.2 x64 Neg NA Pos 37.6 x128 Neg NA Pos 38.1 x128 Neg NA Pos 38.0 x128 Neg NA Pos 38.0 x256 Neg NA Pos 39.1 x256 Neg NA Pos 38.6 x256 Neg NA Pos 39.1 x512 Neg NA Neg NA x512 Neg NA Pos 39.8 x512 Neg NA Pos 39.5 x1024 Neg NA Neg NA x1024 Neg**** NA Neg NA x1024 Neg NA Pos 39.6 SurePath Pos*, Pos **, & Pos *** were also positive for HPV 16 with Ct's of 39.9, 40.3, and 39.9. ThinPrep Neg**** - This sample was negative for HPV 18, but positive for Other HR HPV Genotype (Ct-38.2). Page 6 of 10 Table 11. Comparison of ThinPrep (TP) and SurePath (SP) for HPV Genotype 68 X64 X64 X64 x128 x128 x128 x256 x256 x256 x512 x512 x512 x1024 x1024 x1024 HPV 68 Crossing Threshold (Ct), Positive cutoff = 40.0 ThinPrep (TP) SurePath (SP) Pos/Neg Ct Pos/Neg Ct Pos 34.2 Pos 34.8 Pos 34.4 Pos 35.4 Pos 34.2 Pos 34.8 Pos 37.2 Pos 36.1 Pos 35.6 Pos 35.4 Pos 35.2 Pos 34.3 Pos 36.2 Pos 36.1 Pos 36.3 Pos 37.1 Pos 37.8 Pos 36.7 Pos 36.9 Pos 38.4 Pos 38.4 Pos 36.3 Pos 38.4 Pos 37.7 Pos 38.9 Pos 36.7 Pos 38.2 Pos 37.9 Mech Fail NA Pos 39.0 Table 12. Summary of Analytic Sensitivity Studies Media type Last positive dilution HPV 16 TP SP x512 x1024 HPV 18 TP SP x64 x512 HPV 68 TP x1024 SP x1024 Analytical Specificity SurePath material was spiked with a cocktail of mixed vaginal microbial flora to determine if the flora caused a false positive HPV result. The cocktail contained Lactobacillus acidophilus, Candida albicans, Streptococcus agalactiae (Group B strep), E. Coli, and Peptostreptococcus anaerobius. A negative SurePath patient pool was prepared by Cytology. This negative pool was run on the Cobas-4800 before and after it was spiked with the cocktail. All testing, including before and after spiking, was negative for HPV-16, HPV-18, and HPV-High Risk Other. No false positive reactions were observed (table 13). Table 13. Spiked sample results Panel ID 1. SP Neg Pool 2. SP Neg Pool Results Negative Negative Page 7 of 10 3. 200 µL Micro Cocktail + 800 µL SP Neg Pool Negative 4. 500 µL Micro Cocktail + 500 µL SP Neg Pool 5. SP Neg Pool 6. SP Neg Pool 7. SP Neg Pool 8. SP Neg Pool Negative Negative Negative Negative Negative Carryover Carryover testing on the Pap processor was checked by performing a run on the Cobas 4800 before and after the Pap processing. A 1:10 dilution using SurePath media and a HeLa cell line (positive for HPV 18) was used for the positive sample source. The results are listed in table 14. There was no observable carryover from the positive HPV 18 samples on either the pre or post Cobas 4800 results. Table 14. Carryover study results, HPV testing on Cobas-4800 before and after SurePath processing. Pre SurePath Aliquots - 4800 Results Sample ID Result Ct Neg Neg NA Neg Neg NA Neg Neg NA 1:10 HPV 18 Pos 18 30.1 1:10 HPV 18 Pos 18 30.2 Neg Neg NA 1:10 HPV 18 Pos 18 30.3 1:10 HPV 18 Pos 18 30.1 Neg Neg NA Neg Neg NA Neg Neg NA Neg Neg NA 1:10 HPV 18 Pos 18 29.5 1:10 HPV 18 Pos 18 29.8 Neg Neg NA 1:10 HPV 18 Pos 18 30 1:10 HPV 18 Pos 18 28.9 Neg Neg NA 1:10 HPV 18 Pos 18 29.9 1:10 HPV 18 Pos 18 29.6 Neg Neg NA Post SurePath Aliquots - 4800 Results Sample ID Result Ct Neg Neg NA Neg Neg NA Neg Neg NA 1:10 HPV 18 HPV18+ 30.0 1:10 HPV 18 HPV18+ 29.4 Neg Neg NA 1:10 HPV 18 HPV18+ 29.6 1:10 HPV 18 HPV18+ 30.0 Neg Neg NA Neg Neg NA Neg Neg NA Neg Neg NA 1:10 HPV 18 HPV18+ 29.5 1:10 HPV 18 HPV18+ 29.6 Neg Neg NA 1:10 HPV 18 HPV18+ 29.9 1:10 HPV 18 HPV18+ 29.6 Neg Neg NA 1:10 HPV 18 HPV18+ 29.0 1:10 HPV 18 HPV18+ 29.6 Neg Neg NA Page 8 of 10 Interference Study A SurePath positive pool was produced by combining positives of Other HR HPV, HPV 16 and HPV 18. The pool was tested by adding two volumes of the Acrometrix® Inhibition panel using the Low, Mid and High Hemolytic Vials. Table 15 shows results of the interference study. No inhibition was noted with any of the positive HPV Genotypes up to a hemolysis concentration of 3.8 µg/dL. Table 15. Hemolysis interference study results run with SurePath samples on Cobas-4800 Panel ID Low Interference High Interference Add 200 µL of Heme Panel to 800 µL of SP Pool Add 500 µL of Heme Panel to 500 µL of SP Pool Results Hemolytic Low 1.0 µg/dL Pos/Neg Positive Pos/Neg Positive Hemolytic Mid 1.9 µg/dL Positive Positive Hemolytic High 3.8 µg/dL Positive Positive Stability Study Stability of SurePath media was tested with media spiked with genotypes HPV 16, HPV 18 and HPV 68. Dilutions were made that approached the lower end of the method sensitivity (two dilutions before the endpoint as observed in the Analytical Sensitivity study). These prepared samples were tested for baseline results (Day 0), then separate aliquots were stored at room temperature (21-23°C) and refrigerated (2-8°C). Duplicates from each storage group were tested at Day 0, Day 3, Day 5, Day 7, Day 10, Day 14, Day 21, and Day 28. The results are listed in tables 16 and 17. Results of these studies show that the SurePath specimens are stable for 14 days when refrigerated and for 3 days when stored at room temperature. Table 16. Stability of SurePath specimens - Refrigerated Day # Day 0 Day 3 Day 5 Day 7 HPV 16 HPV 18 HPV 68 Pos/Neg (Ct Value) Pos 36.9 Pos 38.2 Pos 39.6 Pos 39 Pos 36.4 Pos 37.8 (Ct Value) Pos Pos Pos Pos Pos Pos (Ct Value) 37.9 39 39.4 38.1 35.1 37.1 Pos/Neg Pos Pos Pos Pos Pos Pos (Ct Value) 38.3 37.2 38.4 38.1 37.1 36.8 Pos/Neg Pos Pos Pos Pos Pos Pos Page 9 of 10 Day 10 Day 14 Day 21 Day 28 (Ct Value) 38 38.4 39.2 38.7 37 37.9 Pos/Neg Pos Pos Pos Pos Pos Pos (Ct Value) 38.7 37.7 39.4 38.5 37.8 37.4 Pos/Neg Pos Pos Pos Pos Pos Pos (Ct Value) 37.1 38.1 38.2 38.2 37.6 38.7 Pos/Neg Neg Pos Pos Pos Pos Neg (Ct Value) n/a 39 39 39.6 38.1 n/a Pos/Neg Pos Pos Pos Pos Pos Pos (Ct Value) 40 39 38.3 38.7 39.6 39.5 Table 17. Stability of SurePath specimens - Room Temperature Day # Day 0 Day 3 Day 5 Day 7 Day 10 Day 14 Day 21 Day 28 HPV 16 Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos/Neg (Ct Value) Pos 36.9 Pos 38.8 Neg n/a Neg n/a Neg n/a Neg n/a Pos 40.2 Neg n/a Pos 38.2 Pos 38.2 Pos 39.2 Neg n/a Neg n/a Neg n/a Pos 40.2 Neg n/a HPV 18 Pos 39.6 Pos 38.4 Pos 39 Pos 38.9 Neg n/a Neg n/a Neg n/a Neg n/a Pos 39 Pos 38.7 Pos 39.8 Pos 38.7 Pos 39.2 Pos 39.5 Neg n/a Neg n/a HPV 68 Pos 36.4 Pos 37.4 Pos 39.3 Pos 38.8 Neg n/a Neg n/a Neg n/a Pos 39.9 Pos 37.8 Pos 38.3 Pos 37.7 Neg n/a Pos 40 Neg n/a Neg n/a Neg n/a Conclusion: SurePath, post-quot specimens are acceptable for patient testing to evaluate for HPV Genotyping with the Roche Cobas® HPV test on the Cobas-4800. Specimens must be run within 3 days after collection if kept at room temperature, or within 14 days if refrigerated. Page 10 of 10