Dispelling the Myths of Cleaning Validation Destin A. LeBlanc Cleaning Validation Technologies Capital Chapter PDA Gaithersburg, MD October 30, 2002 1 CV myths z “Things the FDA doesn’t allow” z Covers 8 myths z Will discuss issues relating to those myths z Rationale: not unnecessarily restrict scientifically justified options 2 Myth #1 z “Regulatory authorities don’t like rinse water sampling” z Fact: FDA and PIC/S guidance documents says rinse water sampling is one of two acceptable sampling methods 3 Issues z “Direct measure” of target residue z Relating rinse water concentration to potential contamination z Rinse recovery z Adequate coverage of rinse solution 4 Myth #2 z “You must correlate rinse sampling results with swab sampling results” z Fact: Rinse and swab measure two different things; don’t expect correlation 5 Issues z Swabs focus on small area z Rinses focus on larger area z Swab measures worst case z Rinse measures average 6 Issues z If both done correctly on same surfaces, may pass on rinse but fail swab z If both done correctly on same surface, if swabs pass, rinse should also pass 7 Myth #3 z “You can’t use non-specific analytical methods” z Fact: Non-specific methods such as TOC are widely used and are accepted by regulators 8 Issues z TOC limit set on dose based calculations, not PW/WFI specs z Calculate and express limit as active z Convert analytical TOC value to active z Compare measured value to limit 9 Issues z Assume worst case,all TOC due to target residue z Note: Correctly applied, TOC is more stringent than specific method for target residue 10 FDA support z Human Drug CGMP Notes -- “We think TOC or TC can be an acceptable method for monitoring residues routinely and for cleaning validation.” 11 Myth #4 z “If you use TOC, you must correlate it with HPLC” z Fact: As long as TOC is validated with appropriate standards, do not need to “correlate” with HPLC 12 Issues z What’s point of running both TOC and HPLC on same standard for correlation? z Method validation of TOC with target analyte is adequate and sufficient 13 Issues z In CV protocols, TOC will never correlate with HPLC results z TOC is subject to interferences z Can’t express exactly how much target residue present, but can assure is at or below measured amount z As long as interference increase TOC, will be worst case 14 Myth #5 z “Any residue is unacceptable” z Fact: With newer methods or with TOC, will always find some residue 15 Issues z Detection limits of analytical methods achieve lower levels z Issue is whether residue is medically safe and whether it affect product quality z But, any visible residue is generally unacceptable 16 FDA support z Human Drug CGMP Notes -“Should equipment be as clean as the best possible method of residue detection or quantification?” ANSWER: “No…” 17 Myth #6 z “Dose-based (MAC) limits calculations are unacceptable” z Fact: Are referenced in FDA and PIC/S guidance documents 18 Issues z Have been misused z Safeguards against unreasonably high limits z Consider cumulative residues from equipment train z Default limits (such as 10 ppm) z Visually clean criterion z Reasonable “safety” factors 19 Issues z Are defaults arbitrary? z Yes, but so what? z If medically safe limit is X ppm, and I set my limit is below that, from a regulatory perspective, should there be a concern? 20 Issues z Consider other medical or safety concerns unrelated to “dose” z Allergenic z Cytotoxicity z Reproductive z May hazards result in – z Limits = LOD of best method z Dedicated equipment 21 Myth #7 z “Recovery percentages at different spiked levels should be linear” z Fact: Recovery percentages are highly variable. It is not reasonable to expect linear response 22 Example Spike Recovery 1.0 µg/cm2 91% 2.0 µg/cm2 81% 3.0 µg/cm2 71% 23 Issues z Swabbing is a manual procedure (analogy to manual cleaning) z High variability in recoveries recovery for one individual z High variability in recoveries among individuals z As practical matter, will use lowest 24 Myth #8 z “You can’t validate manual cleaning” z Fact: You will validate manual cleaning processes 25 Issues z Manual cleaning more variable than automated processes z Consistency of manual cleaning depends on adequate detail in written procedure and adequate training of operators z Requires more attention to validation maintenance 26 Origin of myths z Probably misinterpretation or misapplication of 483’s z Example: “Your use of rinse water sampling is inappropriate to….” z Faulty conclusion: Can’t use rinse water sampling z Correct response: Use rinse sampling correctly 27 Suggestions Don’t chase latest 483 z Design a comprehensive, defendable cleaning validation program z Confirm (or disprove) “You can’t…” statements by regulatory documents (Human Drug CGMP Notes, Warning Letters, Guidance Documents, GMPs) z 28 z Q&A z Discussion 29