The Biuret Method for the Determination of Total Protein Using an

Application
Note: 51859
The Biuret Method for the Determination of
Total Protein Using an Evolution Array
8-Position Cell Changer
Nicole Krueziger Keppy, Michael W. Allen, Ph.D, Thermo Fisher Scientific, Madison, WI, USA
Introduction
Key Words
• Biuret Method
• 8-Position Cell
Changer
• Total Protein
• UV-Visible
Spectroscopy
One commonly used method for determining the total
protein in a sample is the Biuret method. The Biuret method
is based on the complexation of Cu2+ to functional groups
in the protein’s peptide bonds as shown in Figure 1.
The formation of a Cu2+-protein complex requires two
peptide bonds and produces a violet-colored chelate
product which is measured by absorption spectroscopy
at 540 nm. Over a given
NH 2
NH 2
concentration range,
the measured absorption
at 540 nm is linear
with respect to the
NH
NH
concentration of total
protein. This relationship
Cu 2+
allows a standard curve
to be created that is
NH 2
NH 2
used to calculate the
Figure 1: Biuret reagent reacts with an
concentration of an
alkaline solution of CuSO4 to form a violet
unknown sample.
chelate compound
Experiment and Results
The Biuret reagent was prepared by adding 3 g of
CuSO4• 5H2O and 9 g of sodium potassium citrate to
500 mL of 0.2 N NaOH solution, followed by the
addition of 5 g of KI. The resulting solution was then
brought to a total volume of 1 L with 0.2 N NaOH.
Alternatively, the Biuret reagent is available from a
variety of sources including Thermo Fisher Scientific.
Protein standards and the sample were prepared with
saline solution (8.5 g/L) according to Table 1. 3.0 mL of
Biuret reagent was added to each standard and sample,
the solution was mixed well and incubated at room
temperature for 30 minutes.
The standards and sample were analyzed using the
Biuret method included in Thermo Scientific VISIONcollect
software with biological tests. To begin the measurement,
select the Biuret Method from the provided method files
in Quantification Mode. The experimental method and
8-position cell changer set-up are shown in Figures 2
and 3 respectively.
Figure 2: Experimental Method Set-up
using VISIONcollect
Test Tube Number
Reagent
3.0 mg/mL
BSA (mL)
Protein
Sample (mL)
Saline
Solution (mL)
Final
Concentration
(µg/mL)
1
2
3
4
5
6
7
8
–
0.2
0.4
0.7
1.0
2.0
3.0
–
–
–
–
–
–
–
–
1.0
3.0
2.8
2.6
2.3
2.0
1.0
–
2.0
0
200
400
700
1000 2000 3000
Table 1: Preparation of Protein Standards and the Sample
–
Figure 3: Multi-Cell Method Set-up using VISIONcollect software
Standards 2 to 7 were measured at 540 nm using
standard 1 as the reference sample, or blank. A linear fit
was applied to the standard results in Table 2 to obtain
the standard curve shown in Figure 4. The resulting
calibration curve exhibits a linear relationship with a
correlation coefficient (R2) of 0.9996. The unknown
sample was measured in Quantification mode. Using
the calibration curve the concentration of protein in
the sample was calculated to be 1553 µg/mL, as shown
in Table 2.
Solution
Standard 2
Standard 3
Standard 4
Standard 5
Standard 6
Standard 7
Unknown Sample
Concentration (µg/mL)
Absorbance
200
400
700
1000
2000
3000
1553
0.0238
0.0541
0.0862
0.1304
0.2514
0.3817
0.1972
In addition to these
offices, Thermo Fisher
Scientific maintains
a network of representative organizations
throughout the world.
Table 2: Results of Protein Standards and the Sample using Biuret Method
Conclusion
Automated quantitative analysis of protein is performed
quickly and easily using the Thermo Scientific Evolution
Array UV-Visible spectrophotometer. The VISIONcollect™
software includes a pre-configured method for the Biuret
assay, allowing further customization to individual laboratory
protocols. Integration of the sample measurement and data
analysis into VISIONcollect software saves time and improves
laboratory throughput by eliminating post-measurement
data manipulation. The 8-position cell changer enables
measurements to be taken without exchanging the cells
between measurements further increasing the efficiency
of your methods.
Figure 4: Biuret-Protein Complex Spectrum and Calibration Curve
Africa-Other
+27 11 570 1840
Australia
+61 2 8844 9500
Austria
+43 1 333 50 34 0
Belgium
+32 2 482 30 30
Canada
+1 800 530 8447
China
+86 10 8419 3588
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland / Norway /
Sweden
+46 8 556 468 00
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9434
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 608 276 5659
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
South Africa
+27 11 570 1840
Spain
+34 914 845 965
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
www.thermo.com
©2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Electron Scientific
Instruments LLC, Madison, WI
USA is ISO Certified.
AN51859_E 12/09M
Part of Thermo Fisher Scientific