Fasteners & Metals Testing Program Research Summary Report # 101 – 1st Q 2013 Analysis: 147 – Preliminary Test Analysis Name: Grain Size Materials: 316 Stainless Steel, Titanium (CP Ti Gr4) Overview Determination of the Average Grain Size of Stainless Steel and Titanium was conducted by CTS in the 1st quarter of 2013. For this test, all participants received two samples, Y09 (Stainless Steel) and Y10 (Titanium). Labs were required to perform mounting and surface preparation using polishing and etching techniques as described in ASTM E3 and ASTM E407. 28 participants submitted results to CTS for at least one sample. ASTM E112 and/or ASTM E1382 were the referenced standards for determining the average grain size. Below is a breakdown of the different methods used: Breakdown of Methods Used 1. 2. 3. 4. 5. 6. 7. Comparison (21) Automatic Image Analysis (3) General Intercept (2) Heyn Linear Intercept (1) Abrams Three-Circle (0) Planimetric (Jeffries) (0) Not Specified (1) Comparison 25 Automatic Image Analysis 20 General Intercept 15 Heyn Linear Intercept 10 5 0 Figure 1: This chart shows the different methods used to determine the average grain size. Analysis The results were analyzed using a two sample statistical model. The following tables and graphs represent the results of this preliminary test. The Data Flags assigned do not warrant corrective actions. This report is for study only and should not be used as proof of participation in a proficiency testing program for Grain Size Measurement. Due to the success of this test we have decided to offer two new grain size tests in 2014. We will be offering: 147-2014-1 Grain Size: Stainless Steel 148-2014-1 Grain Size: Inconel You may click on the links to visit our website for more information. Sign up today! ︵ ︶ y t i l a i t n e d i f n o c b a l e r u s n e o t d e s u y r a r o p m e t r e b m u n n o i t a c i f i t n e d i y r o t a r o b a l d e n g i s s A Key for Fasteners & Metals Program Web Summary Report - Lab Mean - The average of the test results obtained by the participant. Grand Mean - The average of the LAB MEANS for all included participants. Laboratories flagged with an X or an M (see DATA FLAG column) are excluded from the GRAND MEAN. Between-Lab Standard Deviation - An indication of the precision of measurement between the laboratories. The greater the spread of the LAB MEANS about the GRAND MEAN, the larger the BETWEEN-LAB STANDARD DEVIATION (and vice versa). Comparative Performance Value (CPV) An indication of how well a laboratory's results agree with the other participants. The CPV is a ratio indicating the number of standard deviations from the GRAND MEAN. CPV = (LAB MEAN - GRAND MEAN)/ BETWEEN-LAB STANDARD DEVIATION. The closer a laboratory's COMPARATIVE PERFORMANCE VALUE is to zero, the more consistent its results are with the other participants' data (and vice versa). Instr Code - A code indicating the manufacturer of the instrument used to perform the test (see separate INSTRUMENT CODE LIST for each test section). Data Flag - DATA FLAGS are assigned based on the simultaneous analysis of both samples tested. Refer to the following chart for an explanation of each symbol: Data Flag . e t i s b e w S T C e h t n o d e h s i l b u p t r o p e r e h t n i a t a d s t i e t a c o l o t b a l a g n i t t i m r e p e l i h w WebCode Statistically Included/Excluded ACTION REQUIRED * INCLUDED CAUTION - review testing procedure and monitor future results. Results fall outside the drawn 95% ellipse but within a 99% ellipse that is calculated but not drawn. Labs flagged with an * do not typically receive a specific note regarding the flag. If this error is repeated in future rounds, however, a lab may need to stop and review its testing procedures. The initial data flag is not cause for alarm. X EXCLUDED STOP - immediate review of data and/or testing procedure is required (all tests exept Chemical Analyses). Results fall outside the 99% ellipse. See the specific note following the data for more information on why the data are excluded. For Chemical Analyses see an additional Memo. M EXCLUDED PROCEED - lab was unable to report data for at least one sample. However, a lab receiving two or more M flags for a test may need to stop and review its testing procedures. Graph - For each laboratory, the Lab Mean for the second sample (y-axis) is plotted against the Lab Mean for the first sample (x-axis) with each point representing a laboratory. The horizontal and vertical cross-hairs are the Grand Means for each sample. When 20 or more laboratories are included in the statistics, an ellipse is also drawn so that 95% of the time a randomly selected laboratory will be included inside the ellipse. Plotted data flags are explained above. Labs not receiving a data flag appear as points on the plot. Interlaboratory Testing Program for Metals Analysis 147 - Preliminary Test Results Conducted in 2013 Average Grain Size - ASTM Grain Size Number (G) ASTM E112, ASTM E1382 Sample Y09 - Stainless Steel Sample Y10 - Titanium Data Flag Lab Mean Diff. from Grand Mean CPV 072MLD M 7.20 0.25 0.40 No Data Reported Comparison 072MLD M 7.46 0.51 0.83 No Data Reported General Intercept 7.50 0.55 0.89 WebCode 0U5MXA Lab Mean Diff. from Grand Mean 6.40 -0.34 CPV Analysis Method Comparison -0.60 36GPW5 M 6.20 -0.75 -1.22 No Data Reported Comparison 4FDZPD X 4.90 -2.05 -3.34 5.70 -1.04 -1.86 Comparison 5TBE74 8.10 1.15 1.87 8.00 1.26 2.26 Comparison 6LCTD8 7.00 0.05 0.08 6.20 -0.54 -0.96 Comparison 6U5UO1 M 7.19 0.24 0.39 No Data Reported Heyn Linear Intercept 6U5UO1 M 7.21 0.26 0.42 No Data Reported Automatic Image Analysis 6ZX4H4 M 7.20 0.25 0.40 No Data Reported Comparison 7D92UM X 5.40 -1.55 -2.52 5.80 -0.94 -1.68 7.20 0.25 0.40 6.80 0.06 0.11 Comparison -1.50 Automatic Image Analysis 8C6U4B BYEY8W 5.90 -0.84 Comparison 5.97 -0.98 -1.60 CCDGZP M 7.80 0.85 1.38 No Data Reported Comparison CCXZ96 M 8.48 1.53 2.48 No Data Reported Automatic Image Analysis E6226X 7.40 0.45 0.73 6.70 -0.04 -0.07 Comparison F82621 7.30 0.35 0.57 6.70 -0.04 -0.07 Comparison H8D8DX 6.40 -0.55 -0.90 6.00 -0.74 -1.32 Not Specified HRCF22 7.00 0.05 0.08 7.20 0.46 0.83 Comparison HTQ63P 6.00 -0.95 -1.55 6.90 0.16 0.29 Comparison IBJ6P0 7.00 0.05 0.08 6.70 -0.04 -0.07 Comparison PVYBEW 7.30 0.35 0.57 6.30 -0.44 -0.78 Comparison QWY8EW M 7.00 0.05 0.08 No Data Reported Comparison R3U1WJ M 6.40 -0.55 -0.90 No Data Reported Comparison SCGMKX 7.56 0.61 0.99 7.60 0.86 1.54 General Intercept SN7SQY 6.70 -0.25 -0.41 7.00 0.26 0.47 Comparison UEYF1X 6.80 -0.15 -0.25 6.40 -0.34 -0.60 Comparison ZOG38A 6.00 -0.95 -1.55 7.00 0.26 0.47 Comparison Summary Statistics Sample Y09 - Stainless Steel Sample Y10 - Titanium Grand Means 6.95 ASTM G 6.74 ASTM G Btwn Labs Stan Dev 0.61 ASTM G 0.56 ASTM G Avg Within Lab Stan Dev 0.50 ASTM G 0.35 ASTM G Statistics based on 16 of 28 reporting participants Interlaboratory Testing Program for Metals Analysis 147 - Preliminary Test Results Conducted in 2013 Average Grain Size - ASTM Grain Size Number (G) ASTM E112, ASTM E1382 Sample Y09 - Stainless Steel Sample Y10 - Titanium 6.95 ASTM G 6.74 ASTM G Analyst Comment: An ellipse wasn't drawn because an insufficient number of labs reported the Average Grain Size for both speciemens. +2σ 8.0 ZOG38A 8.5 UEYF1X ZOG38A UEYF1X SN7SQY SCGMKX R3U1WJ QWY8EW PVYBEW IBJ6P0 HTQ63P HRCF22 H8D8DX F82621 E6226X CCXZ96 CCDGZP BYEY8W 8C6U4B 7D92UM 6ZX4H4 6U5UO1 6U5UO1 6LCTD8 5TBE74 4FDZPD 36GPW5 0U5MXA 072MLD 3.0 SN7SQY SCGMKX R3U1WJ QWY8EW PVYBEW IBJ6P0 HTQ63P HRCF22 H8D8DX F82621 E6226X CCXZ96 CCDGZP BYEY8W 8C6U4B 7D92UM 6ZX4H4 6U5UO1 6U5UO1 6LCTD8 5TBE74 4FDZPD 36GPW5 0U5MXA 072MLD ‐1.0 072MLD GM 072MLD CPVs Stainless 2.0 Titanium 1.0 0.0 ‐2.0 ‐3.0 ‐4.0 Figure 2: This chart shows the CPVs for each lab. Average Grain Size Stainless Titanium 7.5 7.0 6.5 ‐2σ 5.5 6.0 5.0 4.5 Figure 3: This chart shows the Average Grain Size values for each lab. The Grand Means and G at 2σ away from the Grand Means are shown for reference. WebCode: PVYBEW Sample Y09: 316 Stainless Steel Etchant: 60% Nitric Acid Mag: N/A Method: G= CPV = Comparison 7.3 0.57 Mag: N/A WebCode: PVYBEW Sample Y10: Titanium (CP Ti Gr4) Etchant: Krolls Method: G= CPV = Comparison 6.3 -0.78 Mag: 100x Mag: 100x WebCode: UEYF1X Sample Y09: 316 Stainless Steel Etchant: 10% Oxalic Method: G= CPV = Comparison 6.8 -0.25 WebCode: UEYF1X Sample Y10: Titanium (CP Ti Gr4) Etchant: Krolls Method: G= CPV = Comparison 6.4 -0.60 Mag: 100x WebCode: F82621 Sample Y09: 316 Stainless Steel Etchant: PWA - 17 Method: G= CPV = Comparison 7.3 0.57 Mag: 100x WebCode: F82621 Sample Y10: Titanium (CP Ti Gr4) Etchant: Krolls Method: G= CPV = Comparison 6.7 -0.07 Mag: 100x WebCode: HTQ63P Sample Y09: 316 Stainless Steel Etchant: 10% Oxalic Method: G= CPV = Comparison 6.0 -1.55 WebCode: HTQ63P Sample Y10: Titanium (CP Ti Gr4) Etchant: 186 - 10% HF 5% Nitric Acid Mag: 100x WebCode: SCGMKX Sample Y09: 316 Stainless Steel Etchant: 40% Nitric Acid Method: G= CPV = Comparison 7.56 0.99 Mag: 500x WebCode: 6U5UO1 Sample Y09: 316 Stainless Steel Etchant: 10% Oxalic Methods: G= CPV = Mag: 100x Heyn Linear Intercept, Automatic Image Analysis 7.19, 7.21 0.39, 0.42 Method: G= CPV = Comparison 6.9 0.29 Mag: 100x Mag: 100x WebCode: H8D8DX Sample Y09: 316 Stainless Steel Etchant: #149 HCL, H20, H202 Method: G= CPV = Not Specified 6.4 -0.90 WebCode: H8D8DX Sample Y10: Titanium (CP Ti Gr4) Etchant: 2% HF Mag: N/A WebCode: CCDGZP Sample Y09: 316 Stainless Steel Etchant: Not Specified Method: G= CPV = Comparison 7.8 1.38 Mag: 100x WebCode: 072MLD Sample Y09: 316 Stainless Steel Etchant: 9% Oxalic Method: G= CPV = Comparison, General Intercept 7.2, 7.46 0.4, 0.83 Method: G= CPV = Not Specified 6.0 -1.32