For Internal Use Only PL-ELS 2100 & PL-ELS 2100 Ice Evaporative Light Scattering Detectors Introduction The PL-ELS 2100 and PL-ELS 2100 Ice (Integrated Cooled Evaporator) Evaporative Light Scattering Detectors are universal detectors for liquid chromatography. They are designed and built at Polymer Laboratories’ facility in Shropshire, UK, now a part of Varian, Inc. The unique design combines high sensitivity with low temperature operation. Figure 1: PL-ELS 2100/2100 Ice Evaporative Light Scattering Detector The PL-ELS 2100/2100 Ice is designed for analytical applications where compounds lack a UV chromophore or fluorophore and are thermally sensitive, particularly in the pharmaceutical sector. The PL-ELS 2100/2100 Ice is complementary to LC-MS as both techniques share similar chromatographic requirements. Principle of Operation The operation of the PL-ELS 2100/2100 Ice consists of three consecutive steps, namely Nebulization, Evaporation and Optical Detection (see Figure 2). During the nebulization stage, the mobile phase is passed through a fine needle and combined with a gas, such as nitrogen or air, to form a plume of uniform aerosol droplets. The size and shape of the droplets* determine the sensitivity and repeatability of the detector. The droplet size distribution is influenced by changes in the nebulization gas flow as well as other factors, such as mobile phase composition. Large droplets are removed by condensation or collision with the chamber walls, and subsequently directed to a waste outlet. The nebulization process can be heated independently of the evaporation tube, useful in minimizing changes in mobile phase viscosity during gradient elution. The aerosol plume formed during nebulization then passes through a heated coil or tube where the solvent is evaporated, leaving residual particles of the analyte. The * Please note, for purposes of clarity, droplet refers to a droplet of mobile phase liquid formed during nebulization. Particle refers to the solid form of the analyte during the detection process. Page 1 of 24 For Internal Use Only temperature of this heated tube is controlled by the user and is commonly set according to the mobile phase composition or compound volatility. It is important to use as low a temperature as possible for this step, to minimize sample volatilization. Typically, low temperature evaporation is required for low molecular weight pharmaceuticals. However, this type of compound is analyzed in highly aqueous eluents, making their detection at ambient temperature difficult with other ELS detectors. To overcome this problem, the PL-ELS 2100/2100 Ice uses additional dry gas during the evaporation step to facilitate desolvation without increasing temperature. This 'evaporation gas' is controllable by the user according to the type of mobile phase. For example, high boiling point eluents, such as water, require higher evaporator gas flows than low boiling point solvents, such as dichloromethane. This technology enables the PL-ELS 2100 Ice to evaporate water at sub-ambient temperature (ca. 15°C). Figure 2: General Schematic of an ELSD The analyte particle exits the drift tube and enters an optical chamber where the solute particles pass through a collimated light beam. The light source is typically a halogen lamp or a monochromatic laser-emitting diode. The light scattered by the particle is detected using a photomultiplier or photodiode, and the magnitude of the scattered light is dependent on the size of the particle formed. The size of the analyte particle is dependent on several factors, including sample concentration, physical properties of the mobile phase, gas flow and sample volatility. Typical detection limits lie in the range 1-50ng on-column, which can sometimes be 2-3 orders of magnitude lower than with UV detection. ELSD response is non-linear because its signal is related to the absolute quantity of the compound and independent of the analyte’s optical properties. Consequently, it does not obey Beer’s Law. A simple logarithmic manipulation of the data will produce a linear relationship. Some caution must, therefore, be exercised for applications that demand a very low limit of detection or very high accuracy. ELSD Theory There are four main processes by which the path of electromagnetic radiation or light can change direction when passing through a medium containing a suspended particulate phase. These are shown in Figure 3. Page 2 of 24 For Internal Use Only Figure 3: Scattering Mechanisms in an ELSD The importance of each of these processes depends on the radius of the particle (r) compared to the wavelength () of the incident light. Rayleigh scattering is -2 predominant when r/ is < 5x10 . When particle dimensions are greater than /20, they no longer behave as point sources, and Mie scattering becomes predominant. Once particle size approaches the wavelength of incident light, then reflection and refraction begin to prevail. In order to decide which mechanism is predominantly responsible for the “scattering” observed in the PL-ELS 2100/2100 Ice, an estimate of the size of the particles involved compared to the wavelength of the incident light can be made: D0 585 u where D0 na 597 1000Q Qa = = u = = = = Q = = Qa 0.45 1.5 na D 3 / na D 2 mean drop diameter number of drops in the size range with diameter D liquid surface tension liquid density liquid viscosity relative velocity between the gas stream and the liquid stream volumetric flow rate of liquid volumetric flow rate of gas The particulate size may be varied by altering the gas velocity, the eluent flow rate, the temperature of the nebulizer and the initial solute concentration. Changes in the solute concentration and variations in the atomizer gas pressure influence the solute particle size. This relationship gives the instrument a maximum sensitivity around r/ = 4. Detection declines rapidly when values for r/ are above 5 or below 2.5. When r/ < 2.5, the interference effects, typical of Mie scattering, cause the deflected light to be low in intensity at the measuring angles. As the Page 3 of 24 For Internal Use Only particles increase in size, reflection and refraction become dominant and sensitivity increases. A further increase in the particle size causes the ratio of surface area to volume to decrease, thus the sensitivity decreases. The distribution tails as diameter increases, the largest particle in a distribution generally reaching twice that of the mean. Consequently, although there is undoubtedly some Mie and Rayleigh scattering, the observed phenomena are predominantly due to reflection and refraction since the majority of the particles are larger than the incident wavelength. The relative importance of refraction and reflection can be understood by examining the effects of the incident light on a single spherical particle whose equilateral axis lies in the same plane as the photodetector and light source. With this configuration, refraction is of greater significance than reflection. The majority of organic compounds have refractive indices between 1.3 and 1.5. Changes in the refractive index within this range will not greatly affect the quantity of light reaching the detector. This accounts for similarities in the sensitivity of the instrument to various compounds. Therefore, the PL-ELS 2100/2100 Ice is useful as a universal detector, providing that the material under investigation is non-volatile under the operating conditions of the instrument. Instrument Design Nebulizer The design and function of the nebulizer in the PL-ELS 2100/2100 Ice directly influences all subsequent stages of operation, consequently the nebulizer design is key to the performance of the PL-ELS 2100/2100 Ice. Figure 4: PL-ELS 2100/2100 Ice Nebulizer Cross Section The nebulization stream can be categorized into three distinct aerosols, namely primary, secondary and tertiary. The primary aerosol describes the plume formed Page 4 of 24 For Internal Use Only straight after the nebulizer tip, whereas the tertiary aerosol is the portion of droplets/particles that reaches the detection cell (see Figure 4). Consequently, it is the properties of the tertiary aerosol that are important for ELSD performance, however, the properties of the tertiary aerosol are dependent on the characteristics of the primary aerosol. The nebuliser chamber has a glass lining to prevent droplets beading on the inner surface, which would otherwise cause excessive turbulence and poor reproducibility. Within the PL-ELS 2100/2100 Ice, the nebulizer gas flow is fixed at 0.4 SLM, as this is the optimum gas flow for the operation for the nebulizer. Changing nebulizer gas to reflect the mobile phase characteristics is not required with the PL-ELS 2100, as the nebulizer is highly efficient. Low viscosity solvents are nebulized as effectively as high viscosity solvents. The PL-ELS 2100/2100 Ice nebulizer can be heated to 90°C, which can help in reducing viscosity changes across a solvent gradient. Typically, the nebulizer temperature is set to the same value as the evaporator temperature, although nebulizer temperature has minimal effect on sensitivity. Evaporator Design PL-ELS 2100 The technique of ELSD relies on removing solvent by evaporation prior to detection. For non-volatile compounds, this approach is straightforward, as the evaporator temperature can be set at, or close to, the boiling point of the mobile phase. In order to analyze semi-volatile species, the ELSD needs to operate at 30°C, which makes the evaporation of solvents, such as water, problematic. As Figure 4 shows, to dry a 10µm water droplet at 30°C takes 2.4 times longer than at 50°C. Therefore, as you reduce the evaporation temperature, the evaporation tube needs to be longer, in order to remove the solvent. If the droplet size is doubled from 10µm to 20µm, the problem is exacerbated, because a 20µm droplet takes 4 times longer to evaporate than a 10µm droplet at 30°C. The PL-ELS 2100 has the shortest evaporation tube on the market, therefore, to remove solvents such as water at ambient temperature, dry nitrogen gas is added during the evaporation step; this is referred to as “evaporation gas”. This evaporation gas reduces the relative humidity (vapor loading) of the surrounding gas in the evaporation tube, which in turn reduces the drying time. Figure 5 shows the effect that evaporation gas has on the drying time of water droplets at 30°C. By reducing the relative humidity from 70% to 50% for a 20µm droplet, the drying time is halved. Therefore, the addition of gas during the evaporation step allows the evaporation tube to be shorter, and facilitates low temperature operation at ambient temperature. Page 5 of 24 For Internal Use Only Figure 4: Drying Time for Water Droplets at different Evaporator Temperatures Figure 5: Drying Time for Water Droplets at 30°C under different Relative Humidities The evaporation gas is a user defined parameter and adjusted according to the mobile phase conditions. PL-ELS 2100 Ice The PL-ELS 2100 Ice operates in the same way as the PL-ELS 2100, the only difference is the PL-ELS 2100 Ice can operate at sub-ambient temperatures. . The PL-ELS 2100 Ice comprises a peltier cooled/heated evaporation tube as shown in figure 5. This allows the PL-ELS 2100/2100 Ice to operate at sub-ambient temperatures, such that, the evaporation tube can be heated to 80° and cooled down to 10°C, depending on the ambient temperature. The evaporator tube has a cool/heat rate of ca. 4°C/min. Page 6 of 24 For Internal Use Only Figure 7: PL-ELS 2100/2100 Ice peltier cooled/heated evaporator section Optical Design The optical bench of the PL-ELS 2100/2100 Ice has been specifically designed to maximize sensitivity, while keeping the background noise to a minimum. The light source is a short wavelength (480nm) Light Emitting Diode (LED), which provides maximum sensitivity for small particles. The shorter the wavelength of light, the more efficient the scattering process becomes because the smaller particles appear larger to the incident radiation, hence refraction and reflection predominate. The LED source is also a cool light source (unlike a halogen lamp), so internal heating of the instrument, which would otherwise prevent low temperature operation of the ELSD, does not occur. A feedback monitor maintains a constant output over the lifetime of the ELSD. The focal path length of the LED is such that the light beam covers the largest area of the particle stream, consequently, the optics are ‘bent’ to accommodate this (see Figure 8). To maximize detection of the scattered light, the photomultiplier (PMT) detector module is mounted as close as possible to the scattering region. The PMT is also arranged at the optimized angle of 120°C to the incoming radiation to ensure maximum sensitivity. The PMT and LED on the PL-ELS 2100/2100 Ice have been “matched” to the same wavelength (480m), because the PMT is at its optimum sensitivity at this wavelength. Page 7 of 24 For Internal Use Only Figure 8: Cross Section of Optical Bench in the PL-ELS 2100 To prevent condensation forming on the optical surfaces, the optical bench on the PL-ELS 2100/2100 Ice is kept at a constant temperature of 50°C and is “swept” by nitrogen gas at 0.4 SLM. Digital Signal Processing/Filtering The output of the PL-ELS 2100/2100 Ice is converted to a digital signal prior to the analog output. Consequently, the PL-ELS 2100/2100 Ice offers two digital features as outlined below. Gain Function (PMT) This parameter sets the factor by which the output signal is amplified by the detector. The gain setting does not change the sensitivity of the detector. It merely amplifies (or divides) the captured signal by the inputted factor (see figure 9). The gain can be adjusted from 1 to 10, in increments of 0.1. When setting a PMT (or Gain) value, both the signal and noise are simply amplified by the value set. The raw signal output displayed on the parameter screen when the values are changed will reflect this increase or decrease in signal amplification. However, the instrument output and the output displayed on the main operating screen are automatically zeroed to 10mV following a PMT change, and thus the recorded baseline position will remain unchanged. The magnitude of the baseline noise will be the obvious indication of the change in PMT. Page 8 of 24 For Internal Use Only Figure 9: Example of Changing the PMT from 1 to 8 Smoothing Function (smth) The smoothing function is defined as “the number of data points over which the data is averaged”, and can be regarded as a digital time constant. For most applications, the default value of 1 (0.1secs) is satisfactory, however for noisy or spiky data, then it may be beneficial to average the raw data by increasing this smoothing parameter. The maximum smoothing width that can be applied is 50 (5 secs). The value is changed by increments of 1. Caution must be exercised when using this function, as Figure 10 shows; the peak width can become wider if the smoothing function is set too high. Note: The data rate transmitted by the ELSD is a constant 10Hz, regardless of the smoothing setting. Page 9 of 24 For Internal Use Only ELS2100 DAC QC - SMOOTHING 300 Output (mV) 250 NO SMOOTHING 200 adca-00 5 SEC SMOOTHING 150 adca-00 100 50 0 0 20 40 60 80 100 Tim e (s) Figure 10: Example of Smoothing Data In addition to the digital features, the User is also permitted to change the power of the LED source of the PL-ELS 2100/2100 Ice. The LED power enables the brightness of the LED to be adjusted from 0% (off) to 100% (default). This feature is extremely useful if the sensitivity of the detector needs to be reduced, bringing large peaks back on scale (e.g. Preparative Chromatography). Connections The eluent from the chromatography system is connected to the front of the instrument via the low dead volume Valco bulkhead connector provided (see Figure 11). The liquid inlet port is connected directly to the nebulizer by a short length (190mm) of capillary tube, giving a delay volume from port to nebulizer tip of ~5µl. The instrument should be supplied with clean, dry nitrogen gas at a head pressure of 60psi. A 4mm push-in connector is provided at the rear of the instrument for a convenient connection to the gas source. To prevent unnecessary gas usage, an automatic, but controlled, gas shut off valve is integrated into the gas inlet manifold. This will only allow gas to pass into the instrument when the instrument is operating. Should the instrument default to a standby mode, the gas valve will close after 15 minutes. The PL-ELS 2100/2100 Ice is fitted with a standard RS232 (DTE-DCE) 3-wire serial interface. The serial RS232 connector provides a 24bit (10Hz) digital output for connection to a chromatographic acquisition device. This facilitates data acquisition via a PC running Galaxie™ Workstation The PL-ELS 2100/2100 Ice can be controlled from a PC using the RS232 interface and the PL-ELS 2100/2100 Ice graphical control software. If there are no available serial ports on the PC, a serial to USB adapter can be used via the USB port on the PC. Page 10 of 24 For Internal Use Only The I/O connector of the PL-ELS 2100/2100 Ice can be connected to auxiliary equipment to pause or stop the operation of a pump or autosampler if the PL-ELS 2100/2100 Ice reports an error condition. The PL-ELS 2100/2100 Ice is equipped with 2 contact closures, which are normally open, and the available configurations are shown in Table 1. Outputs Inputs I/O Description User contact closure – normally open Pump stop contact closure – normally open TTL active low TTL active high Ground (to case) Remote A/Z Pin No. 4 & 12 3 & 10 2 & ground 9 & ground 1, 5, 6,11 7 & ground [Firmware version 1.0.15. Units with serial # 004-161 and before will also require a wiring modification on the main PCB] Table 1: Control I/O Connections Detector Output (1V) Serial RS232 Connector –Digital Output Solvent Inlet Firmware Flash Upgrade Connector I/O Connector Gas Inlet Liquid Waste Outlet Figure 11: Front & Rear View of the PL-ELS 2100 Page 11 of 24 For Internal Use Only Detector Electronics The PL-ELS 2100/2100 Ice is controlled using a single PCB with surface mounted electronics. The main board controls all the functions of the detector. It contains Flash RAM, which allows the firmware to be upgraded very quickly via the firmware port on the rear of the detector. The main board also includes an internal vapor sensor, a DAC and the PL-ELS 2100 Ice has a TEC control device. The PL-ELS 2100/2100 Ice also include device identification, which allows multiple ELSDs to be connected to one PC, e.g. Using Galaxie CDS. The Device ID is set at 1 in the factory, but is programmable in the field. Please refer to the service manual for programming the Device ID. Front Panel Control The PL-ELS 2100/2100 Ice can be controlled through the graphical display on the front of the instrument. Using this interface, you can select method, evaporator temperature, nebulizer temperature, gas flow and digital features, using the interactive menu bar at the bottom of the display. Table 2 briefly describes the functions of each of the key commands. Function MODE METHOD EVAP NEB GAS or Description Options Access to the parameter settings menu Modes of instrument operation STANDBY RUN Displays and changes the instrument DEFAULT method 1-10 programmable methods* Displays and changes the evaporator OFF, 25-120°C (PL-ELS 2100) temperature OFF, 10-80°C (PL-ELS 2100 Ice) Displays and changes the nebulizer 0-90°C (in 1°C increments) temperature Displays and changes the gas flow 0.90-3.25 SLM Home (or lock) position displays actual Real time display can be values and actions any changes made accessed by pressing the up and in the menu options down arrow keys when in this The locked icon () will be displayed position. with the instrument is controlled by the PC based software Table 2: PL-ELS 2100/2100 Ice Operating Functions * The method editor software supplied with the instrument enables the methods to be edited and stored on a PC and downloaded to the instrument. The real-time output of the photomultiplier can also be displayed graphically via the graphical interface. This provides the user with a means of monitoring detector output without data acquisition. Access to this display is achieved by pressing the up/down arrow keys while in home (lock) mode. Further or key presses changes the display range. Page 12 of 24 For Internal Use Only Modes of Operation The PL-ELS 2100/2100 Ice can be operated in two modes, STANDBY and RUN, both of which are described below. STANDBY The STANDBY mode is the “ground state” of the PL-ELS 2100/2100 Ice, which is initiated automatically after power on. In this mode of operation, the heaters are off, and the gas manifold valve is closed. The STANDBY mode gives the User a control platform in which to setup the operational parameters (gas flow, nebulizer and evaporator temperatures) before switching the unit into RUN mode. The instrument will default to STANDBY mode should an error occur. When the instrument is switched from RUN mode to STANDBY mode following a command or error, then the gas management system is invoked and the gas set to minimum flow of 1.2 SLM for 15 minutes before the gas manifold valve is closed. This minimum “blanket” gas is enough to nebulizer and evacuate solvent should the instrument default to STANDBY mode with solvent still flowing [firmware 1.1.8 and greater]. RUN The RUN mode is the normal operational mode. The instrument is now being controlled at the set temperature and gas flow, and the system is fully operational. During heating or cooling the instrument will display ‘NOT READY’ to show that the system has not reached the set conditions. When the instrument has equilibrated, ‘READY’ will be displayed, and the instrument is ready for use. Ten custom methods can be stored onboard for specific applications (see Table 3 for the 10 default methods), but these cannot be edited via the keypad on the front of the instrument. Custom methods can be created using the PL-ELS 2100/2100 Ice method editor software, and subsequently downloaded onto the instrument. Method 1 2 3 4 5 6 7 8 9 10 Evap (°C) 0 30 50 50 70 90 90 120 120 120 PL-ELS 2100 PL-ELS 2100 Ice Neb (°C) Gas (SLM) Evap (°C) Neb (°C) Gas (SLM) 0 1.6 30 30 1.6 30 1.6 25 30 1.6 50 1.6 20 30 1.8 50 1.6 15 30 2.2 50 1.6 20 80 1.6 50 1.6 40 40 1.4 50 1.2 50 50 1.0 50 1.2 80 60 1.0 90 0.9 80 80 0.9 90 2.8 80 90 3.0 Table 3: PL-ELS 2100/2100 Ice Default Methods Error Conditions The PL-ELS 2100/2100 Ice is equipped with a number of sensors and error checking facilities to ensure safe operation. If an error is detected, the instrument gives an audible warning and a visible description of the error condition. In the event of an error condition, the unit defaults into the STANDBY mode in which the heaters and Page 13 of 24 For Internal Use Only gas are turned off. A complete list of instrument errors and remedial actions are given in the troubleshooting section of this manual. Once the source of the problem has been corrected, the error message will be cleared automatically on the instrument display. Computer Control The PL-ELS 2100/2100 Ice can be controlled via a PC running the ELSD control software (as supplied), or Galaxie workstation. The control software will automatically adjust depending on the model of the ELSD connected to it. Control Software This Windows-based control software offers a second and complementary level to total detector control. An intuitive single control panel provides control as well as a comprehensive monitoring system. Operational parameters can be easily manipulated, saved or loaded by using the flexible Methods Editor, enabling rapid set up and custom method archiving. Control of the PL-ELS 2100/2100 Ice is through a simple control window. Within the control window, the status of the detector is displayed, together with options to control the various instrument parameters. The control window is effectively divided into three main sections: 1. CONTROL View the current mode of operation Change the current mode of operation between Standby and Run Change the Current Method of operation between a stored custom method and the default method ‘XXX’ Change Parameters in the default method ‘XXX’ 2. PARAMETER STATUS Displays the actual values against the set parameters for: Evaporator Temperature Nebulizer Temperature Gas Flow 3. PHOTO DETECTOR Displays current detector output Displays current gain setting Remote Autozero Page 14 of 24 For Internal Use Only Figure 12: PL-ELS 2100/2100 Ice Control Software The PL-ELS 2100/2100 Ice control software offers all the features offered on the instrument control panel, and thus the instrument settings, including gain adjust, smoothing and LED intensity, can be adjusted via the Edit menu (EDIT / Instrument settings menu option). Changes made to these parameters need to be downloaded to the PL-ELS 2100/2100 Ice using the UPDATE button. The PL-ELS 2100/2100 Ice control software allows the User to set up an automated sequence of events during a 24-hour period. Using a time-triggered event approach (based on the PC clock), the User can select a method of choice to be loaded and/or the PL-ELS 2100/2100 Ice status changed. System Test The PL-ELS 2100/2100 Ice control software has an additional feature of System Test, where the performance of the detector can be checked to ensure it is operating to the required standard set at the factory. When the User selects this feature (TOOLS / Run system test menu option), the software will run the detector through a number of tests and analyse the information it gains, providing the User with an analysis of the results. Polymer Laboratories advises that this test is used occasionally to check the performance of the instrument, or when a problem is encountered. System Test Page 15 of 24 For Internal Use Only The PL-ELS 2100/2100 Ice can store 10 customs methods, which can be selected via the front of the instrument or via the control software. The Method Editor program allows the User to create, edit and download these custom methods for method development purposes or specific applications. The program can be launched from the control software via this icon Select method # Edit the control parameters Download to PL-ELS 2100 Add name and description Figure 13: Method Editor window Detector Performance Detector Noise Detector noise on the PL-ELS 2100/2100 Ice can originate from three sources: electrical, gas and solvent. The electrical noise should not exceed 0.1mV, while the gas noise should not exceed 0.2mV. If the solvent has been evaporated successfully, the baseline noise with eluent flow should not exceed 0.3mV. However, the presence of non-volatile additives or impurities in the solvent will dramatically increase the baseline noise. HPLC solvents should have a “residue after evaporation/ignition” value of <1ppm. The baseline noise can be smoothed using the smoothing function on the PL-ELS 2100/2100 Ice, as shown in Figure 14. Page 16 of 24 For Internal Use Only Output (mV) ELS2100 DAC QC - NOISE 41 40.5 40 39.5 39 38.5 38 37.5 37 36.5 36 400 Key: PL-ELS 2100 smoothing setting: 10adca-0011 (1 sec) 30 (3 sec) 40adca-0012 (4 sec) 50adca-0013 (5 sec) adca-0014 450 500 550 600 Tim e (s) Figure 14: Effect of Smoothing Function on the Baseline Noise of the PL-ELS 2100 Sensitivity The PL-ELS 2100/2100 Ice, as with other ELSDs, has a typical LOD range of 1-50ng on-column. For the PL-ELS 2100, the LOD of caffeine is 19ng or for glucose is 50ng oncolumn. However, caution must be exercised when comparing limits of detection between ELSDs. The sensitivity of any ELSD is compound specific depending on the type of mobile phase, flow rate and design of the ELSD. For non-volatile compounds, optimum sensitivity is achieved at high temperature, while for semi-volatile compounds, low temperature operation is required to minimize loss through volatilization. In addition, direct injections can be performed on ELSDs and sometimes an LOD may be quoted which refers to the direct injection measurement. The only true way of comparing ELSDs is by a head-to-head comparison, and even then, the operating temperatures of the ELSD must be correctly optimized to achieve the highest sensitivity. It is for this reason that most ELSDs are sold by demonstration. Linearity ELSD is a non-linear technique, due to the four main scattering processes: 1. 2. 3. 4. Rayleigh Scattering Mie Scattering Reflection Refraction Page 17 of 24 For Internal Use Only The dominance of each of these processes depends on the radius of the analyte particle (r) compared to the wavelength () of the incident light. However, the intensity of these four scattering mechanisms is not equal. Therefore, at high concentrations, particle size is large and refraction and reflection dominate, hence the response is linear in fashion. However, when the concentration of the analyte is lowered, the size of the analyte particle gets smaller, and weaker scattering mechanisms of Mie and Rayleigh dominate. The consequence of this disparity in scattering strength gives the evaporative light scattering detector a non-linear response, especially at low concentrations (see Figure 15). Figure 15: Scattering Mechanisms in ELSD and their Effect on Response The analyte concentration and peak area response for an ELSD are related by the following expression: A = amb Where, A is the measured peak area, m is the sample mass and a and b are coefficients depending on droplet size, concentration, the nature of the solute, gas and liquid flow, as well as vaporization temperature. By plotting concentration vs peak area for this relationship, a curved line is obtained. The chart in Figure 16 shows data plotted on a standard axis. A linear trend line does not fit the data well (R2=0.9902), but a quadratic expression gives R2 as 0.995. Page 18 of 24 For Internal Use Only Comparison of Trendlines fits for Customer data 600000 Linear (Series1) Power (Series1) 500000 y = 1785.5x1.4702 R2 = 0.9948 y = 12268x - 83991 R2 = 0.9902 peak area 400000 300000 200000 100000 0 0 10 20 30 40 50 60 Concentration Figure 16: Concentration vs Peak Area for the PL-ELS 2100/2100 Ice response For ease of manipulation, all ELSD users employ the logarithm of equation 1 to give the following expression: log A = b log m + log a Log-Log Plot of Customer data 6 5.8 y = 1.4702x + 3.2518 R2 = 0.9948 5.6 log peak area 5.4 5.2 5 4.8 4.6 4.4 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 log Concentration Figure 17: ELSD data from figure 15 expressed with Log Axes Therefore, by plotting the log concentration vs the log peak area, the more familiar linear fit can be applied to the data (see Figure 17). The value of b has been shown to be highly dependent on the mobile phase eluents and temperature, with typical values ranging from 0.8-1.2. Calibration data is not forced through the origin, as the detector will give a small response at zero concentration due to fluctuations in electrical and gas noise. Page 19 of 24 For Internal Use Only Uniformity of Response The majority of pharmaceutical compounds possess a UV chromophore, hence UV detection is the first choice detection method. However, the use of a UV detector can limit the sensitivity and efficiency of a separation as the optimum detection wavelength may not be compatible with the mobile phase eluent. In addition, different responses are obtained for different species at the same concentration due to differences in their extinction coefficient. By contrast, since all analyte particles scatter light, the ELSD can detect all samples with high sensitivity and accuracy, irrespective of their chemistry or optical properties. Consequently, when developing chromatographic methods for active ingredients using ELSD, the analyst can select the appropriate solvent system to give the optimum separation without compromising the sensitivity or resolution. ELSD also provides a uniform response to all components, as highlighted in figure 18 for aspirin and phenacetin. A solution containing equal quantities of aspirin and phenacetin produces an equal response for both compounds using the PL-ELS 2100, whereas UV detection exhibits a lower response for aspirin. This is due to a difference in extinction coefficient between the two compounds. This example highlights the advantage of using ELSD when analyzing mixtures of pharmaceutical compounds. Figure 18: Comparison of the Uniformity of Response between ELSD and UV/Vis for Analysis of Aspirin and Phenacetin at Equal Concentrations Dispersion Evaluation of peak dispersion of the PL-ELS 2100/2100 Ice for caffeine was undertaken at 3 eluent flow rates. A PL-ELS 2100/2100 Ice and Sedex 75 ELSD were evaluated at flow rates of 0.5ml/min, 0.25ml/min & 0.1ml/min. Figure 19 shows the plate counts measured at each flow rate for the two detectors. The PL-ELS 2100/2100 Ice was evaluated with both a 0.007” and 0.010” ID nebulizer needle. The PL-ELS 2100/2100 Ice uses a 0.010” ID needle for nebulization as the 0.007” ID needle is prone to blocking. Page 20 of 24 For Internal Use Only Plate counts vs Flow rate 35000.00 30000.00 Unit #049 (10 thou) Unit #058 (7 thou) SEDEX 75 (HPLC) Plate counts N (m) 25000.00 20000.00 15000.00 10000.00 5000.00 0.00 0 0.1 0.2 0.3 0.4 0.5 0.6 Flow rate (ml/min) Figure 19: Dispersion of PL-ELS 2100/2100 Ice for different Nebulizer Needles PL-ELS 2100 Ice Heating/Cooling performance The PL-ELS 2100 ICE can be heated or cooled very rapidly in order to switch between ELSD methods. Figure 20 shows the cooling rate of the PL-ELS 2100 ICE PL-ELS 2100 Cooling curve from 50-30°C. 50-30°C 50 Temp (°C) 45 40 35 30 0.0 1.0 2.0 3.0 4.0 Time (min) Figure 20: PL-ELS 2100 Ice evaporator cooling curve The PL-ELS 2100 Ice has a heating/cooling rate of ca. 4°C/min, however the performance of the peltier unit is dependent on ambient temperature. If the ambient temperature is >25°C, then the PL-ELS 2100 Ice can only achieve a minimum temperature of 15°C.Whereas at an ambient temperature of <25°C, a minimum evaporator temperature of 10°C can be achieved. Page 21 of 24 5.0 For Internal Use Only Best Practice for the PL-ELS 2100/2100 Ice ELSD Optimization To achieve optimal performance from the instrument, certain operational conditions should be met. An understanding of the nebulization/evaporation process will aid in selecting the best operational parameters for that application. In theory, optimal performance is achieved by producing a specific solute droplet size with a narrow size distribution. The nebulized droplet size is affected by adjusting the eluent flow rate, evaporation temperature and gas flow rate. The effect of eluent flow rate on the instrument operation is dependent on the volume of eluent that must be volatilized and on the droplet particle size. A higher eluent volume may require a higher evaporation temperature, a higher gas flow rate, or both, to achieve optimal conditions. The PL-ELS 2100/2100 Ice is designed to receive eluent flow rates up to 5ml/min at ambient temperatures, even for 100% water. However, at sub-ambient temperatures, eluent flow rates should be limited to 1ml/min for optimum performance It is recommended that the instrument be operated at the lowest evaporator temperature required for complete evaporation of the eluent. Optimal temperature settings will be dependent on the volatility of the analyte and not the just the eluent. Only volatile eluents should be used. Non-volatile eluent components (such as nonvolatile salts) will not evaporate and may collect inside the instrument, requiring removal and cleaning of the evaporator tube. The gas flow rate is the most varied parameter when optimizing performance. The detector operates using clean, dry nitrogen at flow rates up to 3.25 SLM. Adjustments to the gas flow rate can have a marked effect on detector response. The maximum response is achieved when reflection and refraction increase as the particle size approaches the optimum diameter. For method development purposes, the nebulizer and evaporator temperatures of the PL-ELS 2100/2100 Ice should initially be set at 30°C/15°C, with a gas flow of 1.6SLM/2.4 SLM. An injection of the analyte should be made and the response recorded. A second injection should be made at slightly higher temperature (eg 40°C/20°C, gas 1.4/1.8 SLM), and a third injection at 80/50°C (gas 1.2 SLM). In doing this, the optimum conditions of the PL-ELS 2100/2100 Ice will be obtained when the highest response of the analyte is observed. The gas flow setting changes the amount of gas passing up the evaporation tube and is adjusted according to the volatility of the mobile phase. As the temperature is increased, the gas flow value can be lowered. Excess evaporation gas will reduce the sensitivity of the detector, therefore, the gas parameter should be operated as low as possible on the PL-ELS 2100/2100 Ice. It is recommended to run the PL-ELS 2100 Ice at low flow rates (eg <1ml/min) if possible, because sub-ambient evaporation is more efficient at these lower flow rates and sensitivity is improved. Chromatography Requirements The PL-ELS 2100/2100 Ice requires the use of volatile mobile phases and additives in order to function correctly. The PL-ELS 2100/2100 Ice is compatible with most Page 22 of 24 For Internal Use Only HPLC mobile phases, with the exception of DMF & DMSO. Sodium or potassium salts and mineral acids are not compatible with the ELSD. Those additives that are compatible with ELSD are shown in Table 4. pKa pKb Acids Trifluoroacetic acid Formic acid Acetic acid Carbonic acid 0.3 3.75 4.75 6.37 13.70 10.25 9.25 7.63 72.4°C 100.7°C 116.0°C - Bases Ammonia Methylamine Ethylamine Triethylamine 9.25 10.81 10.66 11.01 4.75 3.19 3.34 2.99 -33.35°C 16.6°C -6.3°C 89.3°C Buffers Ammonium formate Pyridinium formate Ammonium acetate Ammonium carbonate pH range BP 3.0-5.0 3.0-5.0 3.8-5.8 5.5-7.5 and 9.3-11.3 Ion Pair Reagents Pentafluoropropionic acid ~0.6 Heptafluorobutyric Acid ~0.6 Nonafluoropropionic acid ~0.6 Pentadecafluorooctanoic acid ~0.6 Tridecafluoroheptanoic acid ~0.6 Table 4: ELSD Compatible Mobile Phase Additives MP 120°C 111°C 97°C 120°C 140°C 189°C 175°C Maintenance There is a potential for impurities to accumulate in the evaporation tube and nebulizer due to the nature of the instrument. Consequently, it is highly recommended that the detector is cleaned on a regular basis to prevent this build up of contamination. The simplest way to prevent fouling of the instrument is to include a wash cycle in a sequence of injections, or to ensure that clean mobile phase is flushed through the system prior to switching the unit off. If the instrument does become contaminated, a “steam clean” can be performed whereby the instrument is heated to maximum temperature and the gas flow is set to 3.0 SLM. Water is then passed through the instrument for 16 hours (or overnight) to remove the contamination. Alternatively, the instrument can be flushed with a suitable solvent, if water isn’t applicable, for the same period of time. Page 23 of 24 For Internal Use Only Specifications Light Source Detector Temperature Range Gas requirements Blue LED 480nm Photomultiplier tube with additional digital signal processing Evaporator PL-ELS 2100 PL-ELS 2100 Ice Nebulizer Flow rate Pressure operating range Maximum Pressure Eluent Flow rate Analogue Output Digital Output Communication Outputs Input Instrument Operation Power Requirements Detector Status Size Weight Unpackaged Packaged Unpackaged Packaged ambient-120°C (1°C increments) 10-80°C (1°C increments) Ambient-90°C (1°C increments) Up to 3.25 SLM @ 60 psi @25°C With integrated automatic, controlled gas shut-off valve 60 – 100 psi (4-6.7 bar) 100 psi (6.7 bar) 0-5 ml/min 0-1V FSD 24bit digital data, 10Hz via serial port Serial I/O (RS232) 1 User Contact closure Pump stop: 1 contact closure 1 TTL +ve 1 TTL –ve Auto zero Graphical Vacuum Fluorescent display 5 button keypad 10 predefined methods PC based method utility program 90/120V AC or 220/250V AC 50/60 Hz 2A max Standby, Run 200x450x415 mm (wxdxh) 360x700x600mm (wxdxh) 11kg 16 kg Page 24 of 24