Medical Image Formation Image Sciences Institute Computed Tomography (CT) Literature: Farr & Allisy-Roberts, chapter 4 Image Sciences institute / University Medical Center Utrecht Medical Image Formation X-Ray Tomography • Limits of radiography / fluoroscopy – 3D structures are collapsed into 2D image (obscuring of details, loss of one dimension) – Low soft-tissue contrast – Not quantitative • Features of X-ray CT – X-ray imaging modality (same principles of generation, interaction, detection) – Generation of a sliced view of body interior (“T”, Tomography from Greek tomos = slice) – Computational intensive image reconstruction (“C”) Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT Images Image Sciences institute / University Medical Center Utrecht 1 Medical Image Formation Principle of X-Ray CT Linear scan Source • In one plane, obtain set of line integrals for multiple view angles • Reconstruct crosssectional views Angular scan Object Detector Image Sciences institute / University Medical Center Utrecht Medical Image Formation Scanner design • Purpose: – Obtaining axial (transverse) cross-sectional images (“tomos”) of the human body from many x-ray projection images – Use computer to calculate tomographic images from projection data Image Sciences institute / University Medical Center Utrecht Medical Image Formation Realization of X-Ray CT • Mathematical basis for computed tomography by Radon (1917) • Idea popularized by Allan Cormack at Tufts Univ. (1963) • First practical x-ray CT scanner introduced by Godfrey Hounsfield of EMI Ltd., England (1972) Image Sciences institute / University Medical Center Utrecht 2 Medical Image Formation Example of an early design Image Sciences institute / University Medical Center Utrecht Medical Image Formation First Generation • EMI Mark I (Hounsfield), “pencil beam” or parallel-beam scanner (highly collimated source) ⇒ excellent scatter rejection, now outdated • 180° - 240° rotation angle in steps of ~1 ° • Used for the head • 5-min scan time, 20-min reconstruction • Original resolution: 80 × 80 pixels (ea. 3 × 3 mm2), 13-mm slice Image Sciences institute / University Medical Center Utrecht Medical Image Formation Second Generation • Hybrid system: Fan beam, linear detector array (~30 detectors) • Translation and rotation, however • Reduced number of view angles ⇒ scan time ~30 s • Slightly more complicated reconstruction algorithms because of fan-beam projection Image Sciences institute / University Medical Center Utrecht 3 Medical Image Formation Third Generation • Wide fan beam covers entire object • 500-700 detectors (ionization chamber or scintillation detector) • No translation required ⇒ scan time ~seconds (reduced dose, motion artifacts) • Reconstruction time ~seconds • Pulsed source (reduces heat load, radiation dose) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Fourth Generation • • • • Stationary detector ring (600 – 4800 scintillation detectors) Rotating x-ray tube (inside or outside detector ring) Scan time, reconstruction time ~seconds Source either inside detector ring or outside (rocking, nutating detectors) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Fifth generation • Multiple X-ray tubes or funnel shaped X-ray tube in which an electron beam scans rapidly around a semicircular target • Multiple stationary detector rings allow multiple slices to be scanned simultaneously • Owing to fast acquisition potential for cardiac CT Image Sciences institute / University Medical Center Utrecht 4 Medical Image Formation Electron Beam CT: Scanners for CV Imaging • • • • No moving parts Electromagnetically swept electron beam 50 ms scan time → imaging of beating heart Developed 1979 at UCSF (Boyd et al.), licensed to Imatron, Inc. • Multi slice capability Image Sciences institute / University Medical Center Utrecht Medical Image Formation Multi-Slice CT • Reduction of scanning time / increase in zresolution • Inferior scatter rejection compared to single-slice CT • Use of widened (in z-direction!) x-ray beam and detector array to acquire multiple (typ. 4, up to 16 slices simultaneously. Image Sciences institute / University Medical Center Utrecht Medical Image Formation Spiral CT • Continuous linear motion of patient table during multiple scans • Increased coverage volume / rotation • Pitch: Number of slice thicknesses the table moves during one rotation (typically ~1-2) Image Sciences institute / University Medical Center Utrecht 5 Medical Image Formation Helical Reconstruction • Projections for one slice do not lie in one plane • Interpolation from data outside the slice plane necessary 1st 2nd 3rd Interpolation: 4th Rotation 1st 2nd 3rd 0° 0° 180° 180° 360° 360° -1 0 1 360 Degree Linear 4th Rotation -0.5 0.5 Standard (180 Degree Linear) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Multi Slice Spiral Scanning • Interweaving multiple helices ⇒ increased data density • Allows higher pitch (faster scan speed) pitch = 4 x single slice pitch Image Sciences institute / University Medical Center Utrecht Medical Image Formation X-ray Sources Image Sciences institute / University Medical Center Utrecht 6 Medical Image Formation X-Ray Tubes • Bremsstrahlung X-ray tubes – Fixed anode: oil-cooled – Rotating Anode • Collimator assembly used to control beam (slice) width (~1.0 - 10 mm) • Power: ~120 keV @ 200-500 mA → spectrum ~30 – 120 keV • Rotating geometry requires slip rings Image Sciences institute / University Medical Center Utrecht Medical Image Formation Rotary Gantry Image Sciences institute / University Medical Center Utrecht Medical Image Formation X-ray Detectors Image Sciences institute / University Medical Center Utrecht 7 Medical Image Formation Detector Performance Desired: • High detection efficiency • Fast response • Large dynamic range (ratio of largest to smallest detectable signal) • Linearity (proportional to X-ray intensity) • Insensitive to temperature-voltage variations • Low cost • Small in size • Reliable Image Sciences institute / University Medical Center Utrecht Medical Image Formation Gas Ionization Chambers • Measurement of conductivity induced in a gas volume by the ionizing effect of X-rays. – X-rays ionize gas molecules – Ions are drawn to electrodes by electric field • Number of ion pairs N produced ∝ X-ray intensity Collimator Anode + - - + + + + - Ammeter Cathode Image Sciences institute / University Medical Center Utrecht Medical Image Formation Gas Ionization Chambers Characteristics • To optimize efficiency usually filled with Xenon (high Z) under pressure (up to 30 atm) & Cheap & Excellent stability & Large dynamic range & High spatial resolution ' Low efficiency Image Sciences institute / University Medical Center Utrecht 8 Medical Image Formation Scintillation Detectors • Scintillation material converts X-ray energy into flashes of visible light • Light is measured using photomultiplier tube (PMT) or photo diode (PD) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Principles of CT • Rotating anode emits beam • Beam passes collimator after travelling through the patient • Detectors (+/- 1000) detect signal (logarithm of radiation intensity) • During 360 degrees rotation +/- 300 projection images of 2/3 ms (1 second per transverse image) Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT reconstructie • Example: 300x1000 measurements to reconstruct an image of 256x256 pixels • Method: “Back-projection” Image Sciences institute / University Medical Center Utrecht 9 Medical Image Formation CT: combining multiple views into one image Result: axial (=transverse) slices Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT acquisition source X-ray source object detectors detectors Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT reconstruction backprojection backprojection filtered backprojection Image Sciences institute / University Medical Center Utrecht 10 Medical Image Formation Phases in the backprojection process profiles Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT Scanner Performance Image Sciences institute / University Medical Center Utrecht Medical Image Formation Contrast Mechanism CT similar to X-Ray • Projection data given by raysum: I i = I 0,i exp −∑ µij ∆x i, j • Attenuation in diagnostic range due to photoelectric effect (pe) and Compton scattering (cs) µij = µ pe ,ij ( ρ , Z , ne , E ) + µcs ,ij ( ρ , Z , ne , E ) µ pe ∝ ρ ne Z −3.4 E 3.1 µcs ∝ ρ ne f ( E ) Image Sciences institute / University Medical Center Utrecht 11 Medical Image Formation Beam Hardening • Lower-energy X-rays more heavily attenuated than higher-energy X-rays • Effect accentuated for long paths / high-Z material • Compensation: – Equalization of path lengths (water bag) – Pre-hardening (filtering) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Hounsfield Units • Assign calibrated values to gray scale of CT images • Based on measurements with the original EMI scanner invented by Hounsfield HU original = 500 µ − µW µW → HU = 2 HU original • Relates the linear attenuation coefficient of a local region µ to the linear attenuation coefficient of water • -1000 for air, 0 for water, approx. 1000 for bone CT # = 1000( µ t − µ w ) / µ w Image Sciences institute / University Medical Center Utrecht Medical Image Formation Applications • brain images inner ear Image Sciences institute / University Medical Center Utrecht 12 Medical Image Formation Thorax peripheral bronchial carcinoma emphysema Image Sciences institute / University Medical Center Utrecht Medical Image Formation Abdominal Imaging screening pancreas gastro-intestinal kidneys liver multi-phase study of the liver Image Sciences institute / University Medical Center Utrecht Medical Image Formation Spine Image Sciences institute / University Medical Center Utrecht 13 Medical Image Formation CT angiography Image Sciences institute / University Medical Center Utrecht Medical Image Formation CTA circle of Willis maximum intensity projection Image Sciences institute / University Medical Center Utrecht Medical Image Formation Volume Rendering Image Sciences institute / University Medical Center Utrecht 14 Medical Image Formation Dynamic cardiac CT Coronary arteries on the heart Catharijne 07-11-2002 Image Sciences institute Conferentie / University Medical Center Utrecht Medical Image Formation Tomoscan M/EG Image Sciences institute / University Medical Center Utrecht Mobile CT Medical Image Formation Philips Tomoscan M Image Sciences institute / University Medical Center Utrecht 15 Medical Image Formation Application areas of mobile CT Radiology Operating Room Intensive Care Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT Guided Navigation Image Sciences institute / University Medical Center Utrecht Medical Image Formation CT Guided Navigation Image Sciences institute / University Medical Center Utrecht 16 Medical Image Formation Surgical navigation Image Sciences institute / University Medical Center Utrecht Medical Image Formation Image Sciences institute / University Medical Center Utrecht Medical Image Formation Advantages Inherent to CT? Q: Why the immediate popularity of initial CT, although – low resolution (802 pixels @ 3 mm × 3mm × 13 mm), – limited application (head only), – slow (5 min. scan time, 20 min. reconstruction)? A: CT copes with the following undesirable features of radiography: 9 3D structures are collapsed into 2D image (obscuring of details, competing anatomical patterns, loss of one dimension) 9 Not quantitative Increased Contrast Resolution 9 Low soft-tissue contrast Decreased Structure Noise Image Sciences institute / University Medical Center Utrecht 17 Medical Image Formation Image Contrast in Radioscopy • Signal contrast: C= E0 EB − ES EB µB h µS D x d Cr = G ( µ S − µ B ) d E EB • Assume linear D(E) curve of gradient G and small difference in µ (as for soft tissue): ES EB ES Image Sciences institute / University Medical Center Utrecht Medical Image Formation • Luminance function of display (similar to E(D)-curve) results in contrast: CCT = µ B − µS µB Gray Scale Image Contrast in CT µ (HU) CCT = Gray Scale • Windowing: Values below lower window setting are all black µ B − µS µB − µL • ⇒ Enhanced display contrast (Choice of contrast / latitude) µL µB µ (HU) Image Sciences institute / University Medical Center Utrecht Medical Image Formation Comparison • Comparing contrasts: CCT = Cr / G ( µ B − µ L ) d Because (µB − µL) ~ 200 HU ⇒ Radiography contrast better for large d (~ cm) or strong contrast • Use of windowing possible because of large dynamic range of detection / display in CT • (Desire for digital detectors in radiography) Image Sciences institute / University Medical Center Utrecht 18 Medical Image Formation Something for nothing? C • Limiting factor is noise Blur Noise Blur + Noise d Image Sciences institute / University Medical Center Utrecht Medical Image Formation Noise-Contrast resolution • Detailed analysis shows: SNR r C /N := r r ∝ d −1 SNR ct CCT / N CT • CT superior for smaller structures @ low intrinsic contrast (soft tissue), but limited resolution (~0.5 mm) • Radiography better suited for larger structures and/or strong intrinsic contrast, able of higher resolution (Example: microcalcifications in mammography, d ~50 µm) Image Sciences institute / University Medical Center Utrecht 19