ADInstruments Flow Injection Potentiometric Analysis with PowerLab (Chart & Application Note Peaks) Physical Science AN101A Fluoride, 0.1 – 50 mM (2 – 1000 ppm), was determined potentiometrically using an aluminium wire working electrode and a gold wire reference electrode. A multiple cell design can be used with the cells connected in series for greater sensitivity. J. Catherine Ngila. School of Chemistry, University of New South Wales, PO Box 1, Kensington NSW 2033, Australia. Introduction Potentiometric analysis involves the determination of the voltage of an electrochemical cell which typically consists of two electrodes which are immersed in the solution to be measured. The system is designed so that the potential difference between the two electrodes is related to the concentration of analyte (usually a cation or anion) under study. The sensitivity of systems which generate small potentials can be increased by using a manifold in which a number of identical cells are connected in series, Figure 1. Provided the cell is constructed so that the electrical resistance between the cells of the manifold is much greater (×10) than the resistance between the electrodes in each cell then the overall potential is the sum of the potentials of the individual cells.1 For aluminium and fluoride determinations the electrodes can be simply constructed from short lengths (4.0 cm) 0.5 mm diameter aluminium (99.5%) and gold (99.0%) wire. The relevant half equations are: Al + mH 2O → [Al(OH)m]+3-m + 3e– + mH+ ...[1] Al + nF- → [AlFn]+3-n + 3e– ...[2] 2H+ + e– → H2 ...[3] The aluminium electrode is slowly sacrificed during the reaction (though it will remain useful for many runs). The presence of excess fluoride encourages the dissolution of the electrode via reaction 2, while the presence of aluminium ion suppresses reactions 1 and 2. The corresponding reaction occurring at the gold counter electrode is the reduction of hydrogen ion, reaction 3. For reproducible results the solution must be buffered to between 4 and 5. At higher pH’s precipitates of Al(OH)3 can form while at lower pH the formation of HF is a competing reaction. Flow injection analysis is based on the injection of a liquid sample into a continuous moving carrier solution (much like the injection of a sample for 1 gas or liquid chromatography). The injected sample forms a zone which, during its transport towards a detector, partially disperses within the carrier solution. The detector registers a signal which is related to the concentration of the sample in the carrier solution. In this case the detector is an electrochemical cell. The peak height of the signal, or the area under the signal peak, are the two parameters that are used as a measure of the total amount of sample in the injection. Typically a series of samples of known concentration are injected and the resulting peak areas plotted against sample concentration to produce a calibration curve. Subsequent samples of unknown concentration can now be injected and their concentration determined by reference of their peak area to the calibration curve. When potentiometry is combined with flow injection analysis continuous monitoring of the concentration of analyte is required. The combination of a PowerLab interface driven with Chart software provides a fast, accurate, and simple way to record the data. The program Peaks can then be used for post acquisition data analysis, extracting peak heights and peak areas. Method All solutions are made up in an acetate buffer, pH 5, made by the addition of 1 M acetic acid to 50 mM sodium acetate solution. The carrier solution is pushed through the system with a multi-roller peristaltic pump while sample solutions are injected into the stream at regular intervals. Flow tubing is made from PTFE, fitted with Omnifit® connectors. The electrode outputs are connected directly to the CH 1+ port of a PowerLab/200 via a standard BNC connector, using Chart to monitor the cell potential. The impedance of a PowerLab unit is 1 MΩ which can be used for low resistance cells. For high resistance cells, for example with electrodes coated with a polymeric sensing material, the electrodes should February 1998 About the author Ms Catherine Ngila is undertaking her Ph.D. in Analytical Chemistry at the School of Chemistry, University of New South Wales. She already holds a B. Ed. (Sc.) and M. Sc. (Environ. Chem.) from Kenyatta University. Her current research interests include developing multisensor detection systems for ionic substrates for use in potentiometric flow injection analysis systems. be connected to a pH meter or high impedance voltmeter with recorder output, which is in turn connected to the PowerLab unit (ADInstruments ML165 pH Amp is suitable). export to a third party spreadsheet or graphing program, Figures 3 and 4. Annotation When Chart is opened channel 1 is resized to fill the screen. Recording is commenced by simply clicking the Start button on the screen. During the recording session you can type annotations with the Comments feature – for example the concentrations of the standard solutions. The Notebook facility can also be used After recording has begun the carrier solution is run until a steady baseline signal is obtained (about to writeup overall experimental procedures. In this way all observations and comments are recorded 2-5 minutes). The standard samples are injected. About one minute is required for the sample to pass along with the data rather using a separate notebook. through the system then a further two minutes is allowed to elapse while carrier solution is flushed through to eliminate the possibility of Data Analysis contamination between consecutive samples. Thus The two parameters of importance are the peak the total time for a single measurement is 3 height and peak area. The peak height is related to minutes. During this time the signal rises from its the concentration of the analyte in the sample while baseline value, reaches its maximum, then returns to its original level. After all of the standards have the peak area is a function of the total amount of analyte administered in the injection. Both been injected a series of samples of unknown concentration can be measured in a similar manner. quantities can be used for analytical purposes. Data Display You can view the data, expand the horizontal or vertical scale, use the Zoom Window to magnify a particular region. Make a hard copy of the data by printing the results with the Print command in the File menu. Finally the data can be saved as a compact Chart file, or as an ASCII text file for As an alternative to injection, the inlet tubing can be dipped into a standard solution, then placed back in the blank and this process repeated for all the standards and then the samples. In this procedure the peak heights will give a measure of concentration but since the “dipping” takes place over slightly different periods of time the peak area can no longer be related to total amount of analyte. Perspex electrochemical cell manifold Output PowerLab ADInstrume + CH1 CH2 + CH3 + To Waste CH4 + + Trigger Au Al Au Al Carrier Solution from peristaltic pump Au Al Injection Port Figure 1. Diagram showing connection of a cell manifold suitable for measuring aluminium and fluoride ion concentrations. There may be any number of Au/Al cells connected in series but one to six pairs usually gives good results. For more resistive cells the electrodes should be connected to a MLA165 pH Amp (or other pH meter with recorder output) which is in turn connected to the PowerLab. AN101A 2 February 1998 Use can be made of the in built Data Pad feature of Chart to determine peak heights, however, it is much easier to save the data as a Chart file then use Peaks to reopen this file. Set the baseline, threshold and noise level as required then click on each peak to give an instant readout of its height and area. This data can also be extracted from the Report Window by cutting and pasting into a spreadsheet or graphing program in readiness for the preparation of the calibration curve, Figures 3 and 4. with a commercial membrane ion selective electrode and a pH meter. More to do 3. D. B. Hibbert, P.W. Alexander, and P. Yatiman, Mikrochimica Acta, 108: 93 (1992). References 1. D.B. Hibbert, ‘Introduction to Electrochemistry’, Macmillan, London, 209–210 (1993). 2. D.B. Hibbert, P.W. Alexander, .S. Rachmawati, and S.A. Caruana, Analytical Chemistry, 62: 1015–19 (1990). The method can also used for the indirect 4. P.W. Alexander, P.R. Haddad, and M. Trojandetermination of aluminium ion in the range 0.5 – owicz, Analytica Chimica Acta, 171: 93 50 ppm, using a constant fluoride concentration of (1985). 1.0 mM. Nitrate can be similarly determined in the range 2 – 1000 ppm using a single flow–through cell fitted Figure 2. A trace collected by Chart from apparatus similar to that shown in Figure 1. Only a single pair of electrodes was used. AN101A 3 February 1998 Trademarks MacLab and PowerLab are registered trademarks, and Chart and Scope are trademarks, of ADInstruments Pty Ltd. Other trademarks are the properties of their respective owners. Addresses International ADInstruments Pty Ltd Unit 6, 4 Gladstone Road Castle Hill, NSW 2154 AUSTRALIA Phone:+61 (2) 9899 5455 Fax:+61 (2) 9899 5847 Email:enquiries@adi.com.au Web: http://www.adinstruments.com North America ADInstruments 1949 Landings Dr Mountain View CA 94043 U.S.A. Phone:+1 (650) 965 9292 Fax: +1 (650) 965 9293 Email: info@adinstruments.com Europe ADInstruments Ltd Grove House Grove Road, Hastings East Sussex, TN35 4JS UNITED KINGDOM Phone: +44 (1424) 424 342 Fax: +44 (1424) 460 303 Email:enquiries @adi-europe.com Japan ADInstruments Japan Inc. Okajima Bldg 2-10-1 Iwamoto-cho Chiyoda-ku, 101 Tokyo JAPAN Phone:+81 (3) 5820 7556 Fax:+81 (3) 3861 7022 Email:adijapan@po.iijnet.or.jp Figure 3. The results shown in Figure 2 have been imported into Peaks for display and analysis. The baseline has been flattened, 5 point moving average smoothing applied, the peaks have been automatically located and their relative areas and heights determined. The peak areas normalised to the largest peak have also been determined and can be seen in the Report Window. 4.0 Your local distributor: 3.5 pF 3.0 2.5 2.0 1.5 0.0 0.2 0.4 0.6 0.8 peak height (V) Figure 4. Data in the Report Window can be cut and pasted into spread sheets or exported to other graphing programs in order to construct a calibration graph. This calibration graph was prepared with Igor Pro™. If pF ( = -log10[F-] ) is plotted against signal then a straight line relationship becomes Copyright. All rights reserved. AN101A apparent, E = 1.280 - 0.342 pF, where E is the peak height potential in volts. 4 February 1998