Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted AT 10.1 AT 10.1 AT 10.2 DE Introduction DE Contents last paragraph Table of p. 2 ed The links to the chemicals and apparatus are wrong. Corrected ed The last paragraph (The amount of….) should come before 10.1, so that it is valid for 10.1 and 10.2 changed ed The links to the chemicals and apparatus are wrong. corrected te To my opinion, the proposed method is not applicable to fertilized peat, because materials of this type can contain large amounts of nitrates, which are not detected by the method (see cl. 6) Please proof the validation in EN 13654-1 and eliminate this material in the table, if there are no facts which contradict the experience given. Fertilized peat is not in the scope Upps Recomm.: soil, treated biowaste and sludge in the title ed Referred numbers of pages are not correct always Re-numbering Corrected DE 1 Scope te Nitrogen present in the form of nitrate or nitrite in samples can not be determined by this method. This should be stated more clearly Replace sentence by: “Nitrogen present in the form of nitrate and nitrite can not be determined by this method. Some nitrogen forms will not be determined totally (see clause 6).” Corrected DE 1 1st sentence ed A word seems to be missing This standard is applicable to… Formal corrections are made DE 3 ed Singular correct? … definitions apply: See above DE 3.1 ed Definition can be given more concretely Mass of nitrogen… See above te/ed Better understanding It condenses in the cooling system and rinses into a conical flask, filled with boric acid solution. This solution is titrated with sulfuric acid, using an indicator to mark the endpoint. Changed to proposal ed Improve english writing Replace “Kjeldahltube” by “Kjeldahl digestion vessel (flask or tube)” Changed to proposal ed Better understanding Nitrogen, bound by N-N and N-O linkages… Changed to proposal DE 5 DE 5 DE 6 4th line 2nd sentence 2nd 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 1 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted DE 6 3rd sentence te Concretization Furthermore, of the inorganic fraction nitrates and nitrites are detected only partially Changed to proposal DE 6 4th sentence ed Fluctuations in water content? Variation of water content? “Fluctuations in water content” DE 6 4th sentence ed Fluctuations during titration (what does it mean?) ? deleted DE 6 last sentence ed Meet this expression what it is really meant? The amount of sulfuric acid needed, depends on the composition of the sample which has to be analyzed (see table 2). Changed DE 6 table ed Numbering and headline are missing Use the information under the table as headline and insert “Table 2:”. Changed DE 6 table, headline of column 2 ed Better understanding Consumption of sulfuric acid (36 mol/l) in ml/g Changed DE 7 Headline ed Headline is missing 7 Reagents corrected DE 7.2 ed Potassium sulfate is only one component of the catalyst mixture “Catalyst mixture” (only) as headline Changed DE 7.3 ed The information given is a concretization of 7.2, not a new reagent. Delete the number 7.3 and re-number all the following reagents; check if they are cited in the text correctly. Formal mistakes corrected DE 2nd 7 ed Incorrect numbering 8 corrected DE 2nd 7.1 ed Number of the headline and cited number of the apparatus are not correct 8.1…digestion stand (8.2) s.a. DE 2nd 7.1 te The comment contradicts the information given before (is the tube of 50 ml adequate for all analyses or not?). Specify the information before or delete the comment. corrected DE 8.2 ed …evaporate the fume? …conduct away the fume deleted 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 2 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted DE 9.1 DE 9.1 DE 9.2 te To my opinion, the development of a standard which covers sampling of all materials cited is nonsense. Either it will be very voluminous or not concrete enough. The same situation is given as with sample pretreatment (9.2). Delete the information and refer to the special sampling procedures. Reference cited 3rd line ed What means “closure” material? PTFE is a very expensive material. Its use is not necessary, if only total nitrogen has to be analysed. …containers of appropriate material, e.g. PE Changed to “PE” 2nd section ed The information “During …avoided” concerns on 10.1 Delete in 9.2 and insert in 10.1 Added in 10.1 Ed/te The terms “semimicro” and “macro version” are not explained explain Not changed as these are common expressions ed The information given concerns on 9.2 Delete it in 10 and insert it at the end of 9.2 correcetd DE DE DE 10.1 DE 10 DE 10.1 1st sentence ed Grinding is not referred before. Omit “ground” (instead of “grinded”) or specify grinding in cl. 9.2. Changed to proposal DE 10.1 1st sentence ed Better understanding …, or the corresponding dry mass of undried sample… Changed to proposal DE 10.1 ed The referring numbers are not correct. Check and correct them. Corrected DE 10.1 3 last subsections ed The information given seem to be declarative to the foregoing text Make three notes. DE 10.1 1st subsection ed …adopted? …adapted. Changed DE 10.1 3rd subs. ed What means “working instructions”? Has to be specified, e.g. “in the manufacturers manual.” Changed to proposal DE 10.2 ed The referring numbers are not correct Check and correct them. s.o. 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 3 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted 2 last subs. ed The information given seem to be declarative to the foregoing text Make two notes. Note 10.2 rewritten Upps. Recomm.: other methods of detection should be allowed DE 10.2 DE 11 ed Subdividing into two clauses seems to be appropriate. Headlines: 11 Calculation and 12 Expression of results corrected DE 11.1 ed In the headline “Method” is redundant Calculation corrected DE 11.1 Symbols in the equations ed mt means not a mass but a mass portion wmd corrected DE 11.1 Explanation of Symbols ed mt is … wmd is the dry residue (mass portion of oven dry matter) of the used material in %, determined according to the standard procedure, which is applicable to it corrected DE 11.1 Explanation of Symbols ed The referring numbers are not correct Check and correct them. corrected DE 13 and he whole document table ed The tables should be numbered Give the tables numbers yes DE 13 table Ed/Te Express s rel with two decimal places Express s rel with two decimal places Not changed- format error, during validation it should be done. Ge Pretreatment of samples should possibly be equal to all different parameters. This will save a lot of time. Pretreatment of samples should therefore be prescribed in cluster 1 Here dry and fresh samples can be used Upps. Recomm.: It should be clear that liquid samples can be used as well as dry and NL 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 4 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted moist samples NL 2 ed Replace the WI number 29292030 with WI 292030 and add the complete title. Changed to proposal NL 2 ed Add the sampling standard (as mentioned in chapter 9) in chapter 2. Refrence inserted by Horizontal NL 3 ed Draft terms and definitions according to CEN rules. yes NL 3.1 Ed Other analysis methods should be mentioned. Now this test could only be done with titration as it is defined NL 6 ed Add table title and draft table according to CEN rules. NL 7 Ed Chapter header is missing Insert the last word of chapter 6 Reagent as chapter header 7 corrected NL 7.3 Te Other digestion mixtures, as potassiumsulfate, coppersulfate and selenium or potassiumsulfate and selenium should also be possible Mention proposed mixtures in the text Selenium should be avoided due to its toxicity and due to the fact , that it does not increase digestion efficiency Upps. Recomm.: The TiO2 CuSO4/KSO4 catalyst should be the only catalyst allowed unless another expert group has stated, that other catalysts are as effective and safe table Spectrophotometric measurements, manual or automatic Modern Kjeldahl devices generally contain a distillation and titration unit, Upps. Recomm.:other analytical methods are allowed. Note is added corrected 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 5 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted NL 7.7 Te Sulfuric acid should replaced by hydrochloric acid Sulfuric acid should be replaced by hydrochloric acid HCl is now mentioned as alternative NL 7 Ed Chapter 7 Apparatus and 7.1 should be Chapter 8 en 8,1 Make changes to 8 and 8.1 Formal mistakes are changed NL 7.1 Ed Reference to 6.2 is wrong Reference should be to 8.2 Corrected NL 10 Ed Avoid hanging paragraphs. Add a subtitle to the first paragraph. NL 10 Ed Replace 0.5 with 0,5 according to CEN rules/ NL 10.1 Ed References in this chapter are wrong Correct the references NL 10.5 te Standard now makes analysis of each sample in duplicate obligatory. This is not something to be determined in a standard Remove this. Instead add a procedure of checking the quality of the analysis by analyzing a reference sample in each batch of samples. This may be a home made sample or a certified reference material. Results should be controlled with a control chart. NL 11.1 ed Draft formula and formula expression according to CEN rules. corrected NL 13 table ed Add table title and draft table according to CEN rules. s.o. NL Annex A table ed Add table title and draft table according to CEN rules. s.o. NL Bibliography ed Remove any duplicate ISO/EN standard already listed in chapter 2. s.o. Corrected Changed 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 6 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted GB General T GB Title T GB Table of applicability T GB Scope GB Principle This method is not fit for purpose because of its insistence on using dried samples, which for some materials will mean loss of ammoniacal-N and consequently underreporting the agronomic fertilizer replacement value, which will lead to excessive mineral fertilizer and risk of losses. In growing media it could risk quality problems. Sample homogenization and drying can improving laboratory precision but if some of the analyte is lost in the process the result could be precise but wrong. Method 6162 stresses the importance of keeping soils cool and not letting them dry out because of the risk of losing ammoniacal-N but this method ignores that risk. The title should reflect the Scope of HOR which is “soils, sludges and treated biowaste” [not necessarily in that order] it is not “biowaste” and it is not “sewage sludge” The use of fresh samples is mentioned as alternative. Upps. Recomm.: It should be clear that liquid samples can be used as well as dry and moist samples Determination of Kjeldahl Nitrogen in soil, treated biowaste and sludges Changed to proposal given in Uppsala What is “domestic sludge” and what is “industrial sludge” “fertilized peat” in EN 13654 was a growing medium “sediment” is outside the scope of HOR “bark humus” in EN 13654 was actually “composted bark” and it was a growing medium constituent or a soil improver, it was not a waste Improve the precision of this table – the current terminology is inadequate, misleading or wrong. Restrict the table to the Scope of HOR, i.e. soils, sludges and treated biowastes Table deleted T The wording in EN 13654 is better The Scope of HOR is Soils, sludges and treated biowastes” - “related waste” is outside the Scope of HOR The Kjeldahl method determines ammonium-N, nitrate-N, nitrite-N and organic-N … Nitrogen in NN linkages, N-O linkages and some heterocyclics (especially pyridine) is only partially determined. Nitate and nitrite-N is not determined in the proposed method. Other terms are used as proposed T If samples are dried there is a risk of losing ammoniacal-N, especially from digestates; it is preferable to kepp samples cool between collection and the time that they are analysed and then to extract the sample in its “as received” condition. Is titanium dioxide/copper sulphate [note dioxide spelling] really needed? EN 13654 uses potassium sulphate/copper sulfate Delete the restriction to “dried” samples Fresh and dry samples can be used A detailed survey on catalyst mixtures , which are as effective as the used mixture Should be subject of a 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 7 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted research work in future and exceeded the proposal of the desk study, which aimed at comparing the effectivity of A mixture containing selenium or TiO2 GB Interferences and sources of errors T GB 9.1 T GB 9.2 T 10 T Like the Scope this is also erroneous about the limitations of the method Is PTFE really necessary? It is expensive. Isn’t polyethylene or polypropylene good enough? This repeats the Scope error of not qualifying biowaste with “treated” and including “related waste” It also repeats the error of requiring dried samples Homogenising the sample improves laboratory precision but if the real material is heterogeneous the act of homogenizing might lose some N – the result could be a precise but wrong answer Rewrite digestion temperature and impurities as sources of error added Check necessity for PTFE Changed to “PE” Rewrite Changed to “ treated waste” GB Procedure FIN 7 tec Header is missing:7 Reagents All formal mistakes corrected FIN 8 tec 7 Apparatus should be 8 Apparatus See above FIN 7.1 tec Corrected FIN 10.1 7.1 should be 8.1 and in the same sentence …suitable for digestion stand (6.2) should be (8.2) In the first sentence …flask or tube (5.2) should be (8.1). Para1 tec A homogeneous test sample is ment, which means e.g. manual homogenising not milling, which will result in ammonium losses due to temperature increase Corrected 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 8 of 9 ISO electronic balloting commenting template/version 2001-10 Template for comments and secretariat observations Date: 2006-03-28 Document: 6-16-1 Nutrients Kjeldahl Horizontal 6-16-1 Nutrients Kjeldahl Nitrogen Author of comments: ON/Austrian Standards Institute 1 2 (3) 4 5 (6) (7) MB1 Clause No./ Subclause No./ Annex (e.g. 3.1) Paragraph/ Figure/Table/N ote (e.g. Table 1) Type of comment2 Comment (justification for change) by the MB Proposed change by the MB Project Horizontal observations on each comment submitted Para1 tec FIN 10.1 In the second sentence ..Add 10 ml sulfuric acid (7.1) should be (7.1) Corrected 1 MB = Member body (enter the ISO 3166 two-letter country code, e.g. CN for China; comments from the ISO/CS editing unit are identified by **) 2 Type of comment: ge = general te = technical ed = editorial NOTE Columns 1, 2, 4, 5 are compulsory. page 9 of 9 ISO electronic balloting commenting template/version 2001-10