Date of submission of tenders has been extended up to 30.01.2016 at 14.00 hrs. General requirements for all equipment: All vendors should quote only their high end equipment. Providing misinformation lead to disqualification without stating any reasons. All claims should be supported with authentic published documents. Laboratory manuals are not acceptable. Published papers for claims should be submitted. Documents pertaining to installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ) for analytical equipment qualification should be submitted. All equipment should be quoted along with suitable branded equipment table/stand, dedicated for the equipment concerned. Amendments to the Tender No. * GC-MS/MS SL.NO. 1 Gas Chromatograph: Head space analyser of 100 vial capacity should be quoted as optional. MS/MS triple quadrupole system: Mass range: 10-1050 amu or better Vacuum system: A robust high efficiency vacuum system with minimum maintenance and utility with low noise level and system with high capacity oil free turbomolecular pump. EI-MRM mode IDL: WITH 1 µL in column injection and 100 msec or less dwell time, 30 m column; IDL of the machine for OFN at the same conditions should be ≤0.5 fg. XIC of IDL concentration should be provided. Performance requirements: 1. Sensitivity of the equipment – EI-MRM transition 272-222 at unit resolution, 1 µL on column injection and 100 msec or less dwell time, in 30 m column, the IDL at this condition should be ≤ 0.5 fg. XIC at IDL concentration should be provided. The % RSD for five sequential injections should not exceed <5%. Related details such as chromatogram, details of GC and MS parameters, injection volume, date and time of analysis using GFR Part II compliant software. Documentary proofs on IDL pertaining to the quoted machine should be submitted with the quote. The IDL claim should be demonstrable in the new machine. This is included in the IQ OQ requirement of the machine. 2. Documentary evidence to be provided. Limit of quantification for δ-methrin, quinalphosmalathion and chlorpyriphosin fish matrix with full details as above. Quantified concentration in the provided spiked matrix should be reported. Analysis report should be from the same machine quoted. 3. Analysis of pesticide residues using QuEChERs method. LOD for major pesticides in fish matrix with full details as above. 50 numbers QuEChERS tube and cleanup kit should be quoted for demonstration purpose. Accessories 1. Gas cylinders, minimum two nos. with necessary license and pressure gauge Calibrated digital flow meter with certification from an accredited lab Point No. 3.GC-MS/MS vials for autosampler: 5000 nos. Headspace vials 100 nos. and two crimpers should be quoted as optional. Point No. 7. Analytical standard for dioxins (7 congeners), furans (10 congeners), dioxin like PCBs (12 congeners) should be quoted and purchase of the same for CIFT should be facilitated. Pesticide multiresidue mix (at least 200 compounds) from Restek or equivalent should be provided. The standards should be QC compliant and necessary certificates should be provided. * SPECIFICATION FOR BENCH TOP QUADRUPOLE ICPMS COUPLED WITH HPLC/ION CHROMATOGRAPH Simple and easy to use ICP MS for multi-elemental analysis with high-throughput capability; capable of coupling with HPLC/IC for speciation of elements such as Cadmium, Chromium, Arsenic, Mercury andLead. The instrument should have 21 CFR II compliant standard software. Adequate literature and published articles in peer reviewed journals should be attached to support the claims. All vendors should quote under FOB basis in Foreign Currency. CIFT has his own freight forwarder for transporting and clearing the consignment. * Sl. No. 1: Sample Introduction, point 5: All the gases for Nebulizer, auxiliary, plasma, CCT gases, Reaction should be controlled by Mass Flow Controller or better system. * Sl. No. 1: Ion Optics system, point 1: The ICPMS must have transfer lens system with 90 degree ion optics system or better to provide high ion transmission across the entire mass range. * Sl.No.1: Quadrupole assembly, point 6: Scanning speed >3,000 amu / sec or higher for entire mass range * Sl.No.1: Performance specification, point 5: Sensitivity: Be/Li > 30 x106 cps/ppm Mg/Co >15.0 x106 cps/ppm In or Y > 40.0 x106 cps/ppm U or TI >40.0 x106 cps/ppm * Sl.No.1: Accessories and standards (ICP), point 3 and 4: Gas cylinders for required operation to be supplied (4 for carrier gas, others 2 each) Suitable gas for removing polyatomic interference in reactive mode * Point 6: Sensitive weighing balance (200g / 0.01 mg) * Point 7: On-Line 20 KVA UPS along with branded Amaron/Exide 150 AH batteries for a minimum backup time of 2 hr. * Point 9: Also quote for spares and consumables including standards required for all the operations for three years * Sl. No.2 : Specifications for Ion Chromatography-High Performance Liquid Chromatography Ion Chromatography System should have the capability to connect with ICP-MS for metal speciation studies Quaternary Gradient Pump with built in degassing system Flow rate: 0.001 - 10 ml/min Flow rate accuracy: <0.1% Pressure Range: 6000 psi Autosampler and optional eluent generator * 3 sections for Separation, Detection and Automation * Houses Electrochemical detectors, columns * Lower Section Temperature Range: 10 - 70⁰C in 0.1⁰C increments. * Capable of housing 2 columns of 1-9 mm: 250 cms Length ED Electrochemical Detector * Provides efficient detection for multiple analysis. * Detection Types: Single and dual electrochemical detection. Compatible with 0.4 mm, 2mm to 4 mm id columns * Interchangeable Working Electrode: Gold glassy carbon, platinum, silver * Reference Electrode: pH - Ag/AgCl combination * Auto ranging facility Software for system control, data acquisition, data processing and reporting Suitable PC & Printer and Nitrogen gas cylinder are to be quoted. UV-Visible detector with suitable PEEK Ion-Exchange columns for analyzing Transition Metals / Heavy Metals Cu, Co, Cd, Ni, Mn, Pb, Zn and speciation. (a) UV-Visible Detector with PEEK cell. Optical System : Dual beam optic design Monochromator Wavelength Range : 190-900 nm in 1 nm increments Wavelength Accuracy : ± 1 nm Bandwidth : 6 nm Noise : Typically <±2.5 mAU at 254 nm Drift : < 0.1 mAU/h Cell Path Length : 10 mm Cell Volume : 10 or more µL HPLC Chromatography b) Two sets of columns for anions/cations as well as reverse phase for analyzing trace elements, heavy metals and their speciation. System should have the capability to connect with ICP-MS for metal speciation studies The HPLC System shall include the following individual stackable self-contained HPLC modules Quaternary solvent delivery system Flow rate - 0.001 to 5 ml/min, with ± 0.5% or ± 1ul/min accuracy and precision of less than ± 0.075% RSD, Maximum pressure setting range should be 15000 psi or more Leak sensor should be included as safety feature and maintenance kit with reservoir tray (with 4 solvent bottles complete with fittings) should be included Degassing unit: On line membrane degassing unit with maximum operating flow rate up to 10ml/min or more per flow line Diode Array Detector: Photodiodes: 512 diodes or more Light source: Deuterium and Tungsten Wavelength range : 190 nm to 800 nmwith accuracy ± 1 nm and reproducibility ± 1 nm The flow cell must be temperature controlled from ambient +5 0 C to 40 0C Manual injector kit Manual injector kit with S-S manual injector valve and Rheodyne 7725 injector Auto sampler injector Sample injection system with variable injection volume between 0.1µL to 50µL and zero sample loss during injection Flow line rinse capability both before and after sampling with less than 0.005% carry over Should have safety features like automatic rack and vial recognition and leak sensor System software System software must be compatible with wireless networking and controllable from a web based interface via net work Expert function with automatic purging of mobile phase if pressure falls below specified value Should be able to store minimum of 20 analysis files Columns C 18 Column 150mm*2.1mm*3.1u- with guard column – 1 No Sonicator Sample & Solvent Filtration Kit with suitable Vacuum Pump Accelerated Solvent Extractor for Sample Preparation C8 Column 150mm*2.1mm*3.1u- with guard column – 1 No C 18 Column 250mm*4.6mm*5u- with guard column – 1 No C 8 Column 250mm*4.6mm*5u- with guard column – 1 No Data Management system Hard ware o Intel core i5 Duo processor with 2GB RAM or higher and 500GB hard disk or higher with pre-installed Windows 10 Software o Operation of the system should be easy and intuitive o Security features of software must be comply fully with FDA 21 CFR Part 11 o It should cover full digital instrument control with qualitative and quantitative processing, report creation, self-diagnosis and auto- tuning o Should have the ability for simultaneous monitoring and quantitation of 8 different wave lengths o Provision of online help function for users 5 L capacity with heating facility It should include oven, pump, fluid sensors, collection bottles and Display & key board SPECIFICATION FOR REAL TIME PCR Sl.No. Real Time PCR, point 11: LED Excitation Source with Photodiode / Photo Multiplier Tube as Detection Source or better Sl.No. 10: Power Supply for Vertical / horizontal and Blotting Apparatus, point 13 Warranty 3 years Sl. No. 11: Consumables for RT PCR Required Reagents/consumables for continuous operation of equipment: 2 sets