NSF/ANSI 42 – 2007e © 2007 NSF This document is part of the NSF International standard development process. This document is subject to change and may be a draft and/or non-final version. Committee members may reproduce, quote from, and/or circulate this document to persons or entities outside of their organization after first providing NSF International with written notice of to whom and for what purpose this document is to be shared. NSF/ANSI Standard for Drinking Water Treatment Units – Drinking water treatment units – Aesthetic effects 3 Definitions to be added to NSF/ANSI 330 3.1 Reporting Limit: “The “Reporting Limit” (RL) for an analyte is the maximum level, for an undiluted sample, to which a laboratory may report a particular analyte as “Not Detected”.” The laboratory must have validated the particular analytical method to this level following the procedures as established in the referenced method. The laboratory shall have evaluated its MDL (method detection limit) and LOQ (limit of Quantitation) in reference to the “RL”. In all cases, the MDL must be less than the RL, and preferably, the LOQ should also be below the RL. Further, to ensure accurate quantitation, when preparing its calibration standards, the lowest calibration point should be at or less than the RL. For extracted techniques (i.e., 625), regarding the concentration of the lowest calibration point, the laboratory should apply the concentration factor due to sample preparation. For example, if for a typical sample one liter extracted, and the extract concentrated to 1.0 milliliter, for a factor of 1000, if the RL is set to 0.2 ug/L, then the lowest calibration point should be at or below 0.2 mg/L.” 4 Materials 4.1 Materials in contact with drinking water POE drinking water treatment units shall conform to the protocol and criteria in NSF/ANSI 61. POU drinking water treatment units shall conform to the protocol and criteria in this section. Materials in contact with drinking water shall not impart levels of extractable contaminants that exceed the Total Allowable Concentration (TAC) MCC or MDWL values specified in tables 1, 2 and 3 when evaluated and tested in accordance with 4.2. At a minimum, metallic products shall be evaluated to the TAC levels in Table 1 and products made from non-metallic components shall be evaluated to the TAC levels in Tables 2 and 3 when evaluated and tested in accordance with 4.2. NOTE – The concentration of active agents or additives used in the drinking water treatment process shall be evaluated in the product water as specified in 6.10. The concentration of active agents or additives used in the drinking water treatment process shall not be evaluated during extraction testing. 4.1.1 Complete formulation information on any material not certified as specifically compliant with the sections of the U. S. Code of Federal Regulations, Title 21, listed in table 3, shall be reviewed to determine whether the material presents a health effects concern in contact with drinking water and to assess the material's potential for contributing contaminants to the drinking water. 1 NSF/ANSI 42 – 2007e © 2007 NSF NOTE – As a minimum for those materials requiring submission of formulation information, the complete chemical identity or proportion by weight (in some cases approximate weights or proportions may suffice), ingredient sources of supply, documentation regarding the health effects concern of each ingredient in the material, and documentation regarding the suitability of each ingredient for use in potable-water-contact material shall be provided. 4.1.2 The product shall be tested in accordance with 4.2.3. If the product does not impart a concentration of an extractable contaminant at a level that exceeds the TAC either the MCC, MDWL, or advisory concentrations in tables 1, 2, or 3 4, the product shall be deemed to have met the requirements of 4. If the product does impart a concentration of an extractable contaminant at a level that exceeds the advisory concentration, but not the MCC or MDWL, the product shall be deemed to have met the requirements of 4, but the manufacturer shall be notified of the concentration of the extractable contaminant, and a new product sample(s) shall be immediately retested in accordance with 4.2.3.6. For the parameters in table 4, the required follow-up analyses shall also be performed after the product has been exposed according to 4.2.3.6, if they were not performed as part of the initial exposure under 4.2.3.2. 4.1.3 Whole-system extraction testing may be waived if components, when separately tested, meet the requirements of this Standard and are assembled in a manner that does not introduce any new components, increase the surface area-to-volume ratio of previously evaluated components, or present potential concern based on cumulative factors. 4.2 Materials evaluation 4.2.1 Analytical methods All analyses shall be conducted in accordance with the applicable method(s) referenced in 2. 4.2.2 Exposure water Systems and components shall be exposed to locally available tap water that has been adjusted to contain 50 ± 5 mg/L total dissolved solids and 0.5 ± 0.05 mg/L free available chlorine, and to have a pH of 6.75 ± 0.25. Exposure water used to evaluate systems or components shall be 23 ± 2 C (73 ± 3 F). Any existing concentrations of extraction testing parameters listed in tables 1, 2, and 3 4 found to be present in the exposure water shall be subtracted from the values obtained in the analysis of the extractant water. 4.2.3 Exposure 4.2.3.1 The system or component(s) of a system shall be installed, flushed, and conditioned in accordance with the manufacturer's instructions using the exposure water specified in 4.2.2 at an initial inlet static pressure of 340 kPa (50 psig). 4.2.3.2 The system or component(s) shall be refilled with the exposure water specified in 4.2.2 and maintained for 24 h at a temperature of 23 ± 2 C (73 ± 3 F). A 2-L water sample shall then be collected in accordance with 4.2.3.3. The system or component(s) shall be flushed according to the manufacturer’s instructions, refilled, and maintained for another 24 h at a temperature of 23 ± 2 C (73 ± 3F). A second 2-L water sample shall be collected in accordance with 4.2.3.3. The system or component(s) shall again be flushed according to the manufacturer’s instructions, refilled, and maintained for a third period of 24 h at a temperature of 23 ± 2 C (73 ± 3 F). A third 2-L water sample shall be collected in accordance with 4.2.3.3. 4.2.3.3 A minimum sample volume of 2 L shall be collected at each sample point. If the water-holding volume of the product is greater than 2 L, the entire volume shall be collected in a suitable collection vessel, and a 2-L subsample obtained from this volume. If the water-holding volume of the product is less than 2 L, sufficient samples shall be exposed to provide the required 2-L volume of extractant water. 2 NSF/ANSI 42 – 2007e © 2007 NSF 4.2.3.4 All samples collected shall be composited and analyzed in accordance with 4.2.1. 4.2.3.5 Systems with adsorptive or absorptive media shall be tested with and without the media. Testing without media shall include removal of the media from the system, as well as removal of any non-media materials or ingredients that cannot be disassociated from the media or materials that would be released into the effluent of the system in the absence of the physical barrier provided by the media. An example is the binder used to produce carbon blocks. Normalization or adjustment for changes in wetted surface area from the normal configuration shall be taken into account. For instance, carbon block end caps will have more wetted surface area exposed without the carbon block attached; therefore, an appropriate adjustment in the end caps included in the exposure shall be made. 4.2.3.6 If the level of an extractable contaminant exceeds an advisory concentration in table 1, 2, or 4, the 72-h test exposure sequence in 4.2.3.2 shall be repeated three times using a new product sample(s). The extractant water from the third 24-h exposure of the third 72-h exposure sequence shall be analyzed to determine whether the concentration of the extractable contaminant has been reduced to a concentration less than or equal to the advisory concentration. 4.3 Gas chromatography/mass spectroscopy (GC/MS) analysis 4.3.1 General requirements for GC/MS analysis When determined to be required by a toxicologist following a product-specific formulation review, Semivolatile and Volatile Mass Spectral scans shall be performed on non-metal products or components, and shall include full-range scans to monitor for non-target compounds. Testing for Semi-Volatiles (EPA Method 625 and Volatiles EPA Method 524.2) shall be conducted using the required Target Compounds in Tables 2 and 3 and the laboratory’s reporting limit shall be no greater than the reporting limits listed in Tables 2 and 3. Contaminants are considered either as “Target Compounds” (targets) or as “Tentatively Identified Compounds” (TICs). Target compounds are those analytes for which the analytical system has been specifically validated for, and for the samples in question specifically calibrated in accordance with the referenced analytical procedure. Contaminants that have been identified and quantified by comparison to authentic standards shall be compared to the acceptable exposure concentration as determined in accordance with NSF/ANSI 61 Annex A and NSF/ANSI 61 Annex B. TICs are identified by comparison of the spectrum of the unknown to the mass-spectral reference library utilizing “Probability Based Matching” (as available from instrument manufacturers) as well as interpretation by the GC/MS chemist. The laboratory shall report the TIC with the best match factor except in the following circumstances. 1. Due to the complex nature of GCMS interpretation and identification, when reviewing the list of possible matches for any particular TIC peak, the laboratory has the authority to assign the identification to a compound “hit” with a lower numeric match factor from the library search algorithm. 2. The laboratory may determine that none of the returned compounds by the automated search algorithm is a good match for the unknown peak. In this case the compound is reported as a “Unknown” 3. The laboratory may utilize manual spectral interpretation to identify the peak in question. The library used during the analysis shall be NIST version 2007 or more current version. For TICs, the concentration is estimated by comparison of its total ion area response to the total ion area response of the nearest internal standard. For TIC’s, a response factor of “1” (one) shall be utilized for the purposes of calculating the TICs estimated concentration. Note: To ensure clear communication of the potential range in concentration of tentatively identified compounds, as part of the laboratory report shall be the typical range of response factors from their 524.2 and 625 calibrations (based upon the required compound lists within this standard). Because reporting this range is to give clearer understanding of the potential range in concentration to the user 3 NSF/ANSI 42 – 2007e © 2007 NSF of the TIC concentrations, it is acceptable for the laboratory to determine this range on an annual basis. An example of an appropriate laboratory statement would be; when performing concentration estimates for tentatively identified compounds, the laboratory utilizes an estimated response factor of “1”. In performing methods 524.2 and 625 the typical response factors for calibrated compounds are 0.2 to 2.0 and 0.1 to 1.5 respectfully. Because these factors are applied as a reciprocal, this would mean an overall potential concentration range for any particular TIC from 0.5 the reported concentration to about 10 times. TICs that have been identified and quantified by comparison to a known mass spectrum, or by spectral interpretation by a qualified chemist, or both shall be compared to the acceptable exposure concentration as determined in accordance with Annex A and Annex B of NSF/ANSI 61. The concentration threshold for all TICs and unknowns shall be no greater than 3.0 ppb. Upon toxicological review, TICs may require verification of identification and quantitation as a target compound. This would necessitate performing the requisite validation studies utilizing an authentic standard of the compound in question. In addition, the product manufacturer shall assist the testing laboratory in the identification of an authentic standard for the compound and an appropriate analytical method, if applicable, so that confirmatory identification and quantification can be performed. Contaminants that are detected by GC/MS analysis but are not identified and quantified against a known mass spectrum or authentic standard shall be evaluated as follows: a) The laboratory shall report the molecular weight or, if no molecular ion is identifiable by the GC/MS chemist, a minimum value for the molecular weight (ifor example, if the highest mass ion for the TIC has an m/z of 143, then report MW >=143). b) The laboratory shall report chemical class information if this determination is possible. c) The laboratory shall report the presence of the common halogens chlorine and bromine utilizing their characteristic “M+2” patterns. d) The product material formulation(s) shall be reviewed for potential identification of the unknown contaminant(s) as an ingredient or byproduct; e) The manufacturer shall be notified and requested to provide supporting information that enables identification of the unknown contaminant(s); f) Structure activity relationships (SAR) shall be utilized when sufficient structural identification of the unknown contaminant(s) can be made; and g) Alternative methods of analysis that may identify the unknown contaminant(s) shall be considered, such as classifying the unknown into a chemical class. Contaminants that can be identified after performing one or more of the above steps shall be compared to the acceptable exposure concentration as determined in accordance with Annex A and Annex B of NSF/ANSI 61. In addition, the product manufacturer shall assist the testing laboratory in the identification of an authentic standard for the compound and an appropriate analytical method, if applicable, so that confirmatory identification and quantification can be performed when needed. Contaminants detected by GC/MS analysis for which no identification can be made after performing the above steps shall not be considered in the determination of product compliance to this Standard. When unknown contaminants are detected in the extractant water, the testing laboratory shall report the analytical results to the product manufacturer. Notes: The product manufacturers should assist the testing laboratory in a continuing effort to indentify the unknown contaminant(s) until specific identification is achieved. Items “B” and “C” above may be automated utilizing software available from NIST available with their mass-spectral database. The minimum instrument operation requirements for GC/MS analysis shall be in accordance with USEPA Method 625 with the addition of the following modifications: 4 NSF/ANSI 42 – 2007e © 2007 NSF – The average chromatographic peak area of each internal standard in the calibration curve shall be determined. The chromatographic peak area of each internal standard in the continuing calibration shall be greater than 50% and not more than 200% of that average; – While a continuing calibration check (CCC) is performed, concentrations of 10% of the target compounds for each analysis (e. g., base/neutral, base/neutral/acid, acid) shall be allowed to fall outside of the range of 70% to 130% (outlier) of the true value. None of the concentrations shall be allowed to fall below 50% or above 200% of the true value. If a positive sample analyte result is identified for any outlier, a second CCC shall be performed. If the second CCC determines the sample analyte result no longer to be an outlier, the sample shall be reanalyzed. However, if the second CCC also determines the analyte to be an outlier, a new calibration curve shall be determined and the sample shall be reanalyzed; and – If commercially available mass spectral libraries are utilized, a minimum size of 100,000 compounds shall be required. 4.4 Pass Fail Criteria Contaminants that extract from products when tested in accordance to section 4.2 shall be evaluated to the TAC levels in Tables 1, 2 and 3. Contaminants that extract from products when tested in accordance to section 4.2 that are not listed in Tables 1, 2 and 3 shall be evaluated in accordance to NSF/ANSI 61 Annex B (TAC) concentrations and/or Annex A if the contaminant is not listed in the Annex B tables. 5 NSF/ANSI 42 – 2007e © 2007 NSF Table 1 – Extraction testing parameters (Metals) Parameter aluminum antimony arsenic barium beryllium cadmium chromium copper lead manganese mercury nickel selenium thallium Total Allowable Concentration (TAC) 0.5 0.006 0.010 2.0 0.004 0.005 0.1 1.3 0.015 0.3 0.002 0.1 0.05 0.002 Reporting Limit (RL) mg/L 0.100 0.001 0.001 0.100 0.001 0.001 0.010 0.100 0.001 0.010 0.001 0.010 0.001 0.001 6 Reference Method(s) 200.7, 200.8 200.8, 200.9 200.8, 200.9 200.7, 200.8 200.7, 200.8, 200.9 200.8, 200.9 200.7, 200.8, 200.9 200.7, 200.8 200.8, 200.9 200.7, 200.8 200.8, 245.1 200.7, 200.8 200.8, 200.9 200.8, 200.9 NSF/ANSI 42 – 2007e © 2007 NSF Table 2 – Extraction testing parameters (Semi-Volatiles) Analyte CAS Number 2,4,6-Trichlorophenol 2,4-Dichlorophenol 2,4-Dimethylphenol 2,4-Dinitrophenol 2,6-Di-tert-butyl-4methoxyphenol 2-Chlorophenol 2-Methylnaphthalene 2-Nitrophenol 2-Phenyl-2-propanol 3,3-Dichlorobenzidine 3-and 4-Methylphenol, m&pcresol 4,6-Dinitro-2-methylphenol 4-Chloro-3-methylphenol 4-Nitrophenol 4-tert-Butylphenol or p-tertButylphenol Acenaphthene Acenaphthylene Acetophenone Anthracene Benzo(a)anthracene Benzo(a)pyrene Benzo(b)fluoranthene Benzo(g,h,i)perylene Benzo(k)fluoranthene Benzothiazole Bis(2-ethylhexyl)adipate Bis(2-ethylhexyl)phthalate 88-06-2 120-83-2 105-67-9 51-28-5 489-01-0 Butyl benzyl phthalate 85-68-7 95-57-8 91-57-6 88-75-5 617-94-7 91-94-1 106-44-5 534-52-1 59-50-7 100-02-7 98-54-4 83-32-9 208-96-8 98-86-2 120-12-7 56-55-3 50-32-8 205-99-2 191-24-2 207-08-9 95-16-9 103-23-1 117-81-7 Total Allowable Concentration (TAC) 5.0 3.0 100.0 3.0 30.0 50.0 0.8 700.0 3.0 200.0 0.2 1.0 400.0 6.0 1000.0 7 Reporting Limit (RL) Reference Method(s) 1.0 1.0 2.0 1.0 5.0 625 625 625 625 1.0 0.4 1.0 0.6 1.0 1.0 625 625 625 625 625 1.0 1.0 1.0 2.0 625 625 625 0.4 0.4 0.6 0.2 0.4 0.2 0.2 0.6 0.2 3.0 0.5 1.0 1.0 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 NSF/ANSI 42 – 2007e © 2007 NSF Table 2 – Extraction testing parameters (Semi-Volatiles) Analyte Chrysene Dibenzo(a,h)anthracene Diethyl phthalate Dimethyl phthalate Di-n-butyl phthalate Di-n-octyl phthalate Fluoranthene Fluorene Indeno(1,2,3-cd)pyrene Isophorone Naphthalene N-Nitroso-di-n-butylamine N-Nitrosodi-n-propylamine N-Nitrosodiphenylamine o-Cresol or 2-methylphenol Pentachlorophenol Phenanthrene Phenol Phenyl sulfone Pyrene CAS Number 218-01-9 53-70-3 84-66-2 131-11-3 84-74-2 117-84-0 206-44-0 86-73-7 193-39-5 78-59-1 91-20-3 924-16-3 621-64-7 86-30-6 95-48-7 87-86-5 85-01-8 108-95-2 127-63-9 129-00-0 Total Allowable Concentration (TAC) 6000.0 3.0 700.0 3.0 400.0 100.0 0.1 0.1 70.0 1.0 3.0 2000.0 3.0 8 Reporting Limit (RL) Reference Method(s) 0.3 0.4 1.0 1.0 2.0 1.1 0.2 0.4 0.2 0.5 0.5 0.7 1.0 0.3 1.0 0.5 0.2 0.5 0.2 0.6 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 625 NSF/ANSI 42 – 2007e © 2007 NSF Table 3 – Extraction testing parameters (Volatiles) Analyte 1,1,1,2-Tetrachloroethane 1,1,1-Trichloroethane 1,1,2,2-Tetrachloroethane 1,1,2-Trichloro-1,2,2trifluoroethane 1,1,2-Trichloroethane 1,1-Dichloroethane 1,1-Dichloroethene 1,1-Dichloropropene 1,2,3-Trichlorobenzene 1,2,3-Trichloropropane 1,2,3-Trimethylbenzene 1,2,4-Trichlorobenzene 1,2,4-Trimethylbenzene 1,2-Dibromo-3chloropropane 1,2-Dibromoethane 1,2-Dichlorobenzene 1,2-Dichloroethane 1,2-Dichloropropane 1,3,5-Trimethylbenzene 1,3-Dichlorobenzene 1,3-Dichloropropane 1,4-Dichlorobenzene 2,2-Dichloropropane 2-Butanone 2-Chloro-1,3-butadiene 2-Chlorotoluene 2-ethyl-1-hexanol 2-Hexanone 4-Chlorotoluene 4-Isopropyltoluene 4-Methyl-2-pentanone CAS Number 630-20-6 71-55-6 79-34-5 76-13-1 79-00-5 75-34-3 75-35-4 563-58-6 87-61-6 96-18-4 526-73-8 120-82-1 95-63-6 Total Allowable Concentration (TAC) 10.0 200.0 2.0 Reporting Limit (RL) Reference Method(s) 0.5 0.5 0.5 524.2 524.2 524.2 0.5 3.0 524.2 3.0 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 3.0 40.0 70.0 96-12-8 0.2 0.2 524.2 106-93-4 95-50-1 107-06-2 78-87-5 108-67-8 541-73-1 142-28-9 106-46-7 590-20-7 78-93-3 126-99-8 95-49-8 104-76-7 591-78-6 106-43-4 99-87-6 108-10-1 0.2 600.0 5.0 5.0 0.2 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 5.0 5.0 0.5 5.0 5.0 0.5 0.5 2.0 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 600.0 75.0 4000.0 50.0 3.0 3.0 7000.0 9 NSF/ANSI 42 – 2007e © 2007 NSF Table 3 – Extraction testing parameters (Volatiles) Analyte Acetone Acetophenone Acrylonitrile Allyl chloride Benzene bis(2-Chloroethyl)ether or Di-(2-chloroethyl) ether Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane Carbon disulfide Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane cis-1,2-Dichloroethene cis-1,3-Dichloropropene Cyclohexanone Dibromochloromethane Dibromomethane Dichlorodifluoromethane Diethyl ether Ethyl acrylate Ethyl methacrylate or Ethylmethacrylate Ethylbenzene Hexachlorobutadiene Isoprene or 2-Methyl-1,3butadiene Methyl acrylate Methyl methacrylate CAS Number Total Allowable Concentration (TAC) 6000.0 200.0 Reporting Limit (RL) Reference Method(s) 5.0 5.0 5.0 0.6 5.0 0.5 524.2 524.2 524.2 SIM 524.2 524.2 111-44-4 5.0 0.3 524.2 108-86-1 74-97-5 75-27-4 75-25-2 74-83-9 75-15-0 56-23-5 108-90-7 75-00-3 67-66-3 74-87-3 156-59-2 10061-01-5 108-94-1 124-48-1 74-95-3 75-71-8 60-29-7 140-88-5 3.0 90.0 80.0 80.0 10.0 700.0 5.0 100.0 0.5 0.5 0.5 0.5 0.5 5.0 0.5 0.5 0.5 0.5 0.5 0.5 0.5 5.0 0.5 0.5 0.5 2.0 1.0 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 524.2 1.0 524.2 0.5 0.5 524.2 524.2 5.0 524.2 1.0 1.0 524.2 524.2 67-64-1 98-86-2 107-13-1 107-05-1 71-43-2 80.0 70.0 4.0 30000.0 80.0 80.0 3.0 10.0 97-63-2 100-41-4 87-68-3 700.0 78-79-5 96-33-3 80-62-6 3.0 10000.0 10 NSF/ANSI 42 – 2007e © 2007 NSF Table 3 – Extraction testing parameters (Volatiles) Analyte CAS Number Methyl tert-butyl ether Methylene chloride Naphthalene n-Butyl acrylate n-Butylbenzene n-Propylbenzene 1634-04-4 75-09-2 91-20-3 141-32-2 104-51-8 103-65-1 o-Xylene 2 or 1,2-Xylene Pentachloroethane sec-Butylbenzene Styrene t-Butyl alcohol or t-butanol or tert-butanol tert-Butylbenzene Tetrachloroethene Tetrahydrofuran Toluene Total Xylenes trans-1,2-Dichloroethene trans-1,3-Dichloropropene Trichloroethene or Trichloroethylene Trichlorofluoromethane Vinyl chloride 95-47-6 76-01-7 135-98-8 100-42-5 75-65-0 98-06-6 127-18-4 109-99-9 108-88-3 156-60-5 10061-02-6 79-01-6 75-69-4 75-01-4 Total Allowable Concentration (TAC) Reporting Limit (RL) Reference Method(s) 5.0 100.0 10.0 0.5 0.5 0.5 1.0 0.5 524.2 524.2 524.2 524.2 524.2 0.5 524.2 0.5 2.0 0.5 0.5 524.2 524.2 524.2 524.2 10000.0 3.0 100.0 9000.0 524.2 5.0 50.0 1000.0 10000.0 100.0 4.0 0.5 0.5 5.0 0.5 0.5 0.5 0.5 524.2 524.2 524.2 524.2 524.2 524.2 524.2 5.0 0.5 524.2 2000.0 2.0 0.5 0.2 524.2 524.2 11 NSF/ANSI 42 – 2007e © 2007 NSF Table 3 4 – Materials listed in U. S. Code of Federal Regulations, Title 21, not requiring formulation review Sections 172.880 178.3700 172.888 178.3720 172.878 172.884 172.886 178.3710 173.25 173.65 178.3620 Part 184 solvents Material petrolatum synthetic petroleum wax white mineral oil odorless white petroleum hydrocarbons petroleum wax ion exchange resins – provided that the sub-section stating the composition of the resin is specified divinyl benzene copolymer mineral oil Direct food substances affirmed as generally recognized as safe – when used in accordance with any conditions of use specified for the substance. Solvents that may be considered for solvent bonding without review are limited to acetone, methyl ethyl ketone, cyclohexanone, and tetrahydrofuran. Mixtures such as solvent cements shall be evaluated against NSF/ANSI 61 or shall be subject to formulation review. NOTE – Solvent bonding is not recommended, as solvents soak into synthetic materials and leach back out into water at relatively high levels for long periods of time. In addition, solvents can contaminate the work area and can be adsorbed by carbon in the work area. Solvents, which have been reprocessed or recycled, shall not be used. 12 NSF/ANSI 42 – 2007e © 2007 NSF Table 4 – Non-specific extraction testing parameters Advisory concentration (mg/L) USEPA method(s) phenolics 0.05 420.4 total organic carbon (TOC) 1.0 415.2 Advisory concentration (mg/L) USEPA method(s) 50 160.1 No follow-up recommended 351.2 Analysis for specific compound(s) in material formulations, ammonium analysis, Base/Neutral scan by GC/MS2 (for nonpolar compounds), LC/MS3 (for target polar compounds) Required parameter Formulation dependent parameter total dissolved solids (TDS) total kjeldahl nitrogen (TKN) 1The 0.5 Additional testing1 Analysis for specific phenolic compound(s) in material formulations, Base/Neutral/Acid scan by GC/MS2 Analysis for specific compound(s) in material formulations, Base/Neutral/Acid scan by GC/MS2 (for non-polar compounds), LC/MS3 (for target polar compounds) Additional testing1 additional testing required may include one or more of the analyses. 2 Gas chromatography with mass spectroscopy (GC/MS). 3 Liquid chromatography with mass spectroscopy (LC/MS). 13 NSF/ANSI 42 – 2007e © 2007 NSF 14