533570046 Preparation and characterization of highly lead-loaded red plastic scintillators under low energy X-rays. 2 Matthieu Hamel,1, Grégory Turk2, Adrien Rousseau2, Stéphane Darbon2, Charles Reverdin2 and Stéphane Normand1 1. CEA, LIST, Laboratoire Capteurs et Architectures Électroniques, F-91191 Gif-sur-Yvette Cedex, France 2. CEA, DIF, Bruyères-le-Châtel, F-91297 Arpajon, France 4 6 8 10 12 14 16 18 20 Abstract To the aim of development of a spatially resolved x-ray imaging system intended for Inertial Confinement Fusion (ICF) at the Laser Mégajoule (LMJ) facility, new plastic scintillators have been designed. The main characteristics are the following: fast decay time, red emission and good X-rays sensitivity in the range 10-40 keV. These scintillators are prepared by copolymerisation of different monomers with an organometallic compound. In this matrix are embedded two fluorescent compounds, allowing to shift the energy from the UV to the near IR spectrum. Different parameters were studied: fluorophores concentration, nature of the secondary fluorophore and lead concentration. An outstanding effective atomic number of 53 has been reached, for a loading of lead corresponding to 29%w. Thus, small cylinders were prepared and their performance under X-ray beam was studied and compared with inorganic Cerium-doped Yttrium Aluminium Garnet scintillator (Y3Al5O12:Ce3+). Eventually, such new scintillators or their next generation could replace expensive and brittle inorganic scintillators, inducing a strong industrial potential. 22 Keywords: Plastic scintillator; X-rays; red fluorescence; fast decay time; lead loading. 24 26 28 30 32 34 36 38 40 42 44 46 48 1. Introduction The development of resolved x-ray imaging system within 10-40 keV has to take into account hard radiative environment induced by ICF in the LMJ experiment chamber. Indeed, the image acquisition is difficult due to highly energetic particle and beaming emission resulting directly or indirectly from deuterium-tritium fusion reaction, which can destroy equipments close to the experiment chamber [1]. Hence any X-ray imaging system in those conditions has to be as less vulnerable as possible. To this aim, the perfect scintillator (whatever organic or inorganic) should gather the following requirements: a fast characteristic decay time (below 50 ns); a scintillation wavelength shifted as far as possible to the red wavelengths so as to eliminate Čerenkov blue light by means of optical filtering and a good photoelectric absorption of 10-40 keV X-rays. So far, the best choice should be YAG:Ce single crystal which is known for its excellent properties towards X-rays imaging properties but suffer from a response dependent with temperature [2], a peak scintillation wavelength located at 550 nm and a decay time too long (70 ns if single crystal and 130 ns if polycrystalline) for our applications. Lutetium Sulfide scintillator Lu2S3:Ce seemed to gather all the researched scintillating properties (28.000 photons per MeV, maximum scintillation peak at 592 nm and decay time of 32 ns) but the authors were limited with the small volume of the single crystal which was approximately 1 mm3 [3]. Also of interest, common organic scintillators (such as BC400) are not enough absorbent in the range 10-40 keV X-ray energies due to their low photoelectric cross section in this range. The photoelectric effect probability Ppe is approximately proportional to a power of effective atomic number Z, roughly between Z4 and Z5. For non-relativistic X-ray photon (E < 511 keV), we can use the following approximation that can be expressed as in equation (1): Corresponding author. Tel: +33 1 69 08 33 25; fax: +33 1 69 08 60 30. E-mail address: matthieu.hamel@cea.fr. -1- 533570046 Z 4.35 (1) Ex3 Common plastic scintillators have a low effective atomic number (ca. 5) compared to inorganic scintillators (ca. 30 to 40). Maximizing the photoelectric cross section, by choosing high effective atomic number components, within 10-40 keV is of prime importance, since we want to avoid X-ray absorption by Compton effect. Indeed Compton absorption scatters impinging X-rays resulting in increased absorption thickness, which degrades spatial resolution of the whole imaging system. Therefore a loading of plastic scintillators with heavy elements could increase the photoelectric cross section in our range of interest and could decrease the necessary thickness of scintillators to absorb sufficient amount of x-ray energy to acquire a resolved x-ray image of ICF target. Since the pioneering work of Pichat et al. [4] dealing with the addition of heavy metals in plastic scintillators, most of the research in this field has been produced mainly in the 50's and the 60's [5]. Since then, to the best of our knowledge we were not able to detect any important work. Commercial lead-loaded plastic scintillators cannot respect all the LMJ requirements already explained before: EJ-256 from Eljen Technology displays a blue wavelength and a loading ranging from 1 to 5%. It seems that loading up to 10% should be possible, but the manufacturer does not recommend it. Bicron BC-452 is available either with 2, 5 or 10%w Pb-loading but its emission peak is also located around 420 nm. Amcrys-H is able to reach 12%w of lead but do not precise the emission wavelength. It is noteworthy that all these plastic scintillators suffer from a dramatic decrease of their light output, due to the loading of lead. All these discrepancies prompted us to develop our home-made plastic scintillators which would embrace all the requirements. A decade ago an efficient method for the production of optical resins doped with lead was described [6], consisting of a matrix prepared from styrene, methacrylic acid and lead dimethacrylate. We decided to extend this method for the preparation of plastic scintillators and the results will be presented herein. Ppe 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94 2. Experimental Bis-N-(2,5-di-t-butylphenyl)-3,4,9,10-perylenetetracarbodiimide and Nile Red were used as received from Sigma Aldrich. Lead dimethacrylate was purchased from Fox Chemicals and used as received. Vinyltoluene, 2-hydroxyethyl methacrylate and methacrylic acid were distilled from calcium hydride. The synthesis of N-(2’,5’-di-t-butylphenyl)-4-butylamino-1,8naphtalimide was already described by some of us [7]. Absorption spectra were recorded with a Jenway 6715 spectrophotometer. Fluorescence spectra and quantum yields of fluorescence were obtained with a Horiba Jobin Yvon Fluoromax-4 spectrofluorimeter. Absolute quantum yields were obtained with an integration sphere. Decay times were observed under UV excitation of the scintillator. Effective atomic numbers were estimated with XµDAT software [8]. Typical scintillators developed in this experiment are ca. 2 inch diameter and a few mm thick. 4 different concentrations of lead have been studied: 5, 10, 20 and 27%w. The primary fluorophore was N-(2’,5’-di-t-butylphenyl)-4-butylamino-1,8-naphtalimide whereas the wavelength shifter was either bis-N-(2,5-di-t-butylphenyl)-3,4,9,10-perylenetetracarbodiimide or Nile Red. These two fluorophores were solvated in the appropriate amounts of vinyltoluene, methacrylic acid and lead methacrylate. The preparation of these scintillators was patented [9]. 96 3. Results and discussion -2- 533570046 98 100 102 104 106 108 110 112 114 116 3.1. Choice of the fluorophores The combination of radiation hardness, nanosecond fluorescence lifetime, high Stokes shift, quantum yield and good physical properties for a red fluorescent molecule is still a challenge for chemists [10]. To the best of our knowledge, only a single publication explained what composes a fast and red plastic scintillator [11]. This system used a long series of fluorophores, i.e. butyl-PBD, dimethyl-POPOP, perylene and rubrene. The initial 35 ns decay time was then reduced to 5 ns by exposing the polymerized sample to large irradiation doses, leading to an important decrease of the light output. Some other recipes exist for liquid scintillation [12] or optical fibre applications but compounds such as rhodamine B or ammonium salts are too polar to be conveniently dissolved into solid solutions of polymers derived from polystyrene. We decided therefore to develop our own fluorophores, and two systems were considered after several tests [13]. They are drawn in Figure 1. Figure 1: N-(2’,5’-di-t-butylphenyl)-4-butylamino-1,8-naphtalimide 3,4,9,10-perylenetetracarbodiimide 2 and Nile Red 3. 1, bis-N-(2,5-di-t-butylphenyl)- 120 Actually the difference between the two combinations concerns the second fluorophore which is in the first system bis-N-(2,5-di-t-butylphenyl)-3,4,9,10-perylenetetracarbodiimide and in the second system Nile Red. Both wavelength shifters' absorption spectra fit well with the emission of N-(2’,5’-di-t-butylphenyl)-4-butylamino-1,8-naphtalimide 1. The photophysical characteristics of the three compounds are resumed in Table 1. 122 Table 1: Spectroscopic data of compounds 1-3. 124 (nm) Compound a max (L mol-1 cm-1) A 417 15,100 1 528 109,200 2 524 33,400 3 a Spectra recorded in spectroscopic toluene at the concentration 10-5 M. b Stokes shift ( 1 / max in cm-1). 1/ max A F 118 c (nm) max F 485 540 570 (cm-1) b 3,362 421 1,540 F (%) c 96 72 60 Absolute quantum yield of fluorescence. 126 128 130 1,8-Naphthalimide molecule 1 displayed excellent properties as its role of first fluorophore with a nearly quantitative quantum yield of fluorescence and a good Stokes shift. The choice of the second fluorophore was more tedious and balanced in favour to perylenediimide 2. Despite a very low Stokes shift, the quantum yield was good (72%) and the molar extinction coefficient very high, allowing thus to dope the scintillator with very low concentrations. -3- 533570046 132 Nevertheless, scintillation prepared with the second system allowed us reaching scintillation wavelengths above 600 nm, which is of interest for avoiding Čerenkov Effect (see below). 134 136 138 140 142 144 146 148 150 152 154 3.2. Loading of the scintillators Common plastic scintillators are known to have inefficient X-ray absorption over 5-10 keV energies, owing to their low effective atomic number Zeff (5.7 according to a polystyrenebased plastic scintillator doped with 1.5%w p-terphenyl and 0.05%w POPOP). As a matter, doping plastic scintillators with different lead concentrations should increase the effective atomic number and as a consequence the photoelectric cross-section in the 10-40 keV X-rays range. Figure 2 is a calculation of X-ray mass attenuation coefficient against photons energies for 3 scintillators: YAG:Ce, an undoped plastic scintillator and a plastic scintillator doped with 12%w Pb (Zeff ≈ 40; undoped and doped scintillator were simulated with the same composition feed except for lead). Results indicate that absorption at 20 keV was 20 times higher for loaded plastic scintillators compared with their unloaded cousins, and nearly the same absorption was obtained for some energies of interest between loaded plastic scintillator and YAG:Ce. We therefore focused our efforts on the loading of scintillators with ca. 10%w lead. Fewer and higher compositions were also tested for comparison. Different loadings with tin were also investigated but will not be discussed in this paper. Relevant information can be found in the published works of Cho et al [14] in the 70’s. Figure 2 (color online): simulation of absorption for three different scintillators (standard plastic scintillator: green curve; 12%w Pb-loaded plastic scintillator: red curve; YAG:Ce: black curve) obtained with XµDAT. The area of interest 10-40 keV is highlighted in blue. 156 158 3.3. Preparation of the scintillators Based on the different possibilities of fluorophores mixtures, fluorophores concentrations and lead percentage, various scintillators have been prepared. The matrix was a ternary system -4- 533570046 160 162 164 166 168 170 composed of vinyltoluene, methacrylic acid and lead dimethacrylate. In the case of the heaviest scintillator a binary system 2-hydroxyethyl methacrylate / lead dimethacrylate was used. Vinyltoluene was chosen since its polymer displays an excellent refractive index [15] of 1.61 at 580 nm which can balance the low refractive index of methacrylic acid [6] (1.50) Relevant properties of these samples are resumed in Table 2. Another important criterion is the X-rays absorption efficiency which is proportional to the value × Z4 [16], which allows to estimate by rule of thumb X-ray photoelectric absorption. As can be seen in Table 2, whereas Zeff increases rapidly, the density does not change dramatically. As a result, typical values of × Z4 are located around 3 – 3.2 × 106 for 10%w loaded plastic scintillator. For comparison, YAG:Ce displays a higher but close value of 4.77 × 106. 172 174 176 178 180 For all scintillators the percentage of lead was expressed a priori from the starting feed of the preparation. Two elemental analyses were performed on samples #2 and #5 at the Laboratoire Central d’Analyses du CNRS, Solaize, France. The experimental values slightly differed from the experimental, but towards the upper values, with %Pb measured at 12.34 and 29.51%, instead of 11.0 and 27.4%, respectively. This gap was explained by a slight evaporation of volatile solvents while heating the scintillator. Table 2: General properties of the lead-loaded plastic scintillators. These data will be discussed along the paper. Ref. %w Zeff Decay time Determination Zeff4 1st fluorophore 2nd fluorophore max F 3 6 Pb (g/cm ) (/10 ) (%w) (%w) #1 10.9 1.20 40.26 3.15 1 (0.5) 2 (0.02) (nm) 586 #2 #3 #4 #5 #6 #7 #8 #9 11.0 5.4 21.9 27.4 21.9 10.9 10.9 10.9 1.18 1.12 1.54 1.55 1.38 1.12 1.16 1.12 40.31 32.93 49.60 53.11 49.60 40.25 40.19 40.19 3.11 1.32 9.32 12.33 8.35 3.01 3.03 2.92 1 (0.05) 1 (0.05) 1 (0.05) 1 (0.05) 1 (0.05) 1 (0.5) 1 (1) 1 (1) 2 (0.002) 2 (0.002) 2 (0.002) 2 (0.002) 2 (0.002) 3 (0.02) 2 (0.04) 3 (0.05) 579 578 580 591 578 632 544 630 (ns) coefficient τ1 12.0, I1 0,98 τ2 46.0, I2 0.02 13.26 12.78 11.28 9.22 12.09 n.d. n.d. n.d. 0.997 0.989 0.992 0.996 0.997 0.993 n.d. n.d. n.d. 182 184 186 188 190 3.4. Physical results 3.4.1. Scintillation wavelength Cherenkov light intensity in the sense of number of photons per material unit length and per wavelength width interval is known to decrease with 1/2 versus wavelength [17] To this aim, various compositions of scintillators were studied so as to shift the maximum of emission to longer wavelengths (Figure 3). Thus, we succeeded in preparing a plastic scintillator with the first fluorophore absorbing in the UV region and a wavelength shifter allowing fluorescence close to 590 nm. This has been possible by using molecules with good Stokes shift, such as 1,8-naphthalimides [18]. 192 -5- 533570046 194 196 198 200 202 204 206 208 210 Figure 3 (color online): normalized scintillation emission of various scintillators compared with Čerenkov effect (normalized to wavelength of minimal spectral sensitivity of considered CCD camera) and the relative CCD spectral sensitivity. Scintillation performances were evaluated under X-rays excitation. A tungsten-anticathod Philips RX FFL tube powering at 40 kV and 40 mA delivered an X-ray bremsstrahlung emission with a maximum of fluence close to 35 keV stopping at 40 keV. The beam was shuttered, leading to a square section of 1 cm² on the scintillator. Scintillation photons were collected with an optical fibre derived from the trajectory of the X-ray beam. Three different scintillators were tested, depending on their concentration of lead: scintillators #3, #1 and #6 with 5.4, 10.9 and 27.4 %w, respectively. Data are represented in Figure 4. It showed that according to Figure 2, X-rays were not well-absorbed when a small amount of lead was incorporated inside the matrix: almost no scintillation could indeed be detected for the lowest doped plastic scintillator #3. The two other scintillators displayed a blue-shifting emission of 6 and 11 nm for #1 and #6, respectively, compared with UV excitation (data not drawn but resumed in Table 2). Nevertheless, only the emission relative to the second fluorophore 2 was detectable, which means that the cascade of wavelength-shifting is efficient. Under the same conditions, a better intensity was observed for the highest loaded plastic scintillator. 212 214 Figure 4: emission spectra under 35 keV X-rays for scintillators #3, #1 and #6 (ca. 5, 10 and 27 %w-Pb) -6- 533570046 216 218 220 Except Čerenkov spectrum, the other interest of emission spectrum of scintillator is the spectral adaptation towards its detector, represented by the spectral matching factor SMF [19]. It is a number without dimension between 0 and 1 (1 meaning a perfect fit), which is calculated with the formula (2): SMF E ( ) S CCD ( )d (2) E ( )d 222 Where E() is the normalized emission spectrum of the scintillator and SCCD() is the spectral sensitivity of the detector, a CCD camera in our case. 224 Table 3: Spectral Matching Factor of various scintillators towards the Pixis CCD camera. Scintillator YAG:Ce #1 #2 #3 #4 #5 #6 226 SMP 0.993 0.966 0.999 0.940 n.d. 0.989 0.983 It is clear from Table 3 that all plastic scintillators display a good adaptation for their use with this CCD camera, owing to their principal emission wavelength higher than 550 nm. 228 230 232 234 236 238 240 242 244 246 248 250 252 254 3.4.2. Decay time For the purpose of the project, it was mandatory to find the fastest scintillators so as to meet the requirement of limited time of image acquisition of ICF target at LMJ facility [20].Decay times of our scintillators were measured on a single-photon counting chain. The excitation was produced by electrical discharges in a lamp containing gaseous hydrogen, which resulted in UV bursts impinging on the tested scintillator. The light emitted by the scintillator was filtered around the principal wavelength emission, so as to eliminate stray light, and then collected by an XP2020 photomultiplier tube, whose anode signal was redirected to an electronic chain of temporal discrimination, allowing analyzing the signal with precision of 100 picoseconds. The temporal signal of the excitation pulse was recorded and all the decay signals of scintillators were deconvoluted from excitation. Resulting signals are drawn in Figure 5. This Figure showed the decay time of the 10.9%w lead-loaded plastic scintillator #1, along with well-known standard scintillators measured in the same conditions. Among the nine lead-loaded plastic scintillators tested, only one (sample #1) was presented a doubleexponential fit. Determination coefficients for deconvoluted signals are excellent (Table 2), which mean that the fit is in perfect agreement with the measured curve. As can be seen, loaded plastic scintillator presented a slower decay (between 9.2 and 13.2 ns, depending on the feed composition, see Table 2) than unloaded blue scintillating plastic scintillator NE102 but remained fast enough for X-ray imaging with limited time during ICF experiments at LMJ facility. This decay time was 7 to 10 times faster than YAG:Ce which was estimated between and 70 [21] and 119 ns [22] (at the maximum emission wavelength of 530-550 nm), and close to the commercial red scintillating plastic scintillator BC430 (16.8 ns at 580 nm, data not shown in Figure 5). Different inorganic scintillators suitable for X-ray spectrometry such as BGO obviously did not display an appropriate decay time, according to Figure 5. Finally, no relevant afterglow was detected since after 100 ns only 0.1% of the initial light intensity was remaining. -7- 533570046 256 Figure 5 (color online): Normalized luminous intensity of various scintillators relative to time. 258 260 262 264 266 268 It is noteworthy that the presence of lead has a strong influence on the decay time of the scintillator. As one can see, the fastest scintillator is also the heaviest (Table 2, entry #5, 27.4%w) which is the only scintillator presenting a decay time below 10 ns. Quenchers are known to reduce not only the scintillation yield, but also the slow component of the signal, and thus the decay time. They are commonly used for the preparation of ultra-fast plastic scintillators [23]. Another point is the concentration of fluorophores, which is also a trick for reducing the decay time [24], despite again in low scintillation yield. In our case, we indeed observed a slight decrease of the decay time by comparing entries #1 and #2 which have exactly the same composition except the concentration of fluorophores which is 10 times higher for #1. As a result, the decay time was reduced by nearly 25%. 270 272 274 276 3.4.3. Scintillation efficiency One of the main factors describing a scintillator is the scintillation efficiency (expressed in visible photons emitted per absorbed MeV of X-ray energy, i.e. in ph.MeV-1). As can be seen of Figure 7, scintillation efficiency is represented versus characteristic decay time. The points corresponding to scintillators are graphically situated with respect to a merit curve expressed as in the relation (3): 100 t 278 e dt 0 e t 10 4 h .MeV 1 (3) dt 0 280 The curve of merit expresses that the scintillator emits as much as 104 hν.MeV-1 in less than 100 ns. The region of interest where ideal scintillators are located is over this curve. Being -8- 533570046 282 284 286 288 290 292 294 given that no known scintillating efficiency exceeds 105 hν.MeV-1, the region of interest is limited down by the merit curve and upper by the maximal known efficiency. The relative scintillation efficiencies of our materials were compared with YAG:Ce, known to deliver nearly 8,000 ph/MeV at peak wavelength of 550 nm under X-ray irradiation [25]. They were measured by analyzing the CCD image intensity of scintillators irradiated by an Xray source term peaking at 40 keV. The tested scintillators were placed at 40 cm of the X-ray source and imaged through a 4 meters-long optical relay system to a CCD camera. The X-ray source term being known by previous measurements with Amptek CdTe detector and knowing the spectral sensitivity of the CCD camera, and knowing the scintillating efficiency of YAG:Ce, we can deduce the scintillating efficiency of scintillators. The experimental setup is shown on Figure 6. Figure 6: Experimental setup of relative scintillating efficiency measurement 296 298 300 Our best lead loaded scintillator, namely #1, has its emission spectrum peaking at 580 nm under X-rays excitation with a scintillating efficiency of 200 ph/MeV. This has been considered as the main drawback by the Authors, compared to YAG:Ce and other known compositions, and will be improved in the future. -9- 533570046 302 304 306 308 310 312 314 316 318 320 322 324 Figure 7 (color online): scintillation efficiencies of different organic (round points) and inorganic (square points) scintillators. The color of the points indicate the scintillation wavelengths, i.e. 400-450 nm for violet-blue, 520-550 nm for green and 560-590 nm for yellow-orange. 3.4. Discussion The goal of this project was the preparation of plastic scintillators sensitive to 10-40 keV Xrays which presented a red scintillation, a fast decay time and a good scintillation efficiency. Among these four properties, three of them have been fulfilled. Only the scintillation efficiency has to be improved, a progression of one decade is expected for the next generation of scintillators. Two issues could be pointed out. First is probably a weak energy transfer from the matrix where interactions between radiation and matter occur to the primary fluorophore. It is admitted that the absorption spectrum of the primary fluorophore has to fit correctly with the emission spectrum of the matrix. Molecule 1 presents two absorption bands: 260-300 and 360-470 nm. It could be possible that the energetic cascade is not good enough. We will therefore consider the use of another fluorescent molecule able to cover the emission spectrum of the matrix and transfer the energy close to the absorption of compound 1. Second is the influence of lead which could act as a quencher of the scintillation process. Quenching of scintillation by tetraphenyl lead and other organolead molecules has already been discussed [26] The Authors explained that the quenching should occur by energy transport to the triplet level of the quencher, possibly by dipole-dipole interactions. Whatever happens, the best solution in our case seems to find the best compromise so as to be both sensitive to 10-40 keV X-rays and enough fluorescent to afford good scintillation efficiency. 326 328 330 4. Conclusion Due to the harsh conditions encountered at the LMJ facility, YAG:Ce which is so far the best compromise for X-ray imaging in ICF conditions presents also many drawbacks. We propose therefore herein the preparation and some characterizations of highly lead-loaded red - 10 - 533570046 332 334 336 scintillating fast plastic scintillators. New developments permitted to load plastic scintillators with very high concentrations of lead without dramatic loss of optical properties. Calculations of the X-rays absorption performances showed that our matrix is comparable with YAG:Ce. A smart combination of fluorescent molecules allowed obtaining pretty fast plastic scintillators with monoexponential decay times close to 10 ns. From this first generation of plastic scintillators, only the scintillation efficiency had to be increased. The challenge is now to ameliorate this efficiency from 2.5% to almost 25%, relative to YAG:Ce. 338 340 342 Acknowledgements We greatly thanks Gilles Ledoux and Christophe Dujardin from LPCML, Université de Lyon 1, for their precious know-how and their measuring chains that allowed to extract decay times and spectral emissions of our scintillators. 344 References 1 J. L. Bourgade, R. Marmoret, S. Darbon, R. Rosch, P. Troussel, B. Villette, V. Glebov, W. T. Shmayda, J. C. Gomme, Y. L. Tonqueze, F. Aubard, J. Baggio, S. Bazzoli, F. Bonneau, J. Y. Boutin, T. Caillaud, C. Chollet, P. Combis, L. Disdier, J. Gazave, S. Girard, D. Gontier, P. Jaanimagi, H. P. Jacquet, J. P. Jadaud, O. Landoas, J. Legendre, J. L. Leray, R. Maroni, D. D. Meyerhofer, J. L. Miquel, F. J. Marshall, I. Masclet-Gobin, G. Pien, J. Raimbourg, C. Reverdin, A. Richard, D. Rubin de Cervens, C. T. Sangster, J. P. Seaux, G. Soullie, C. Stoeckl, I. Thfoin, L. Videau, C. Zuber, Rev. Sci. Instrum. 79 (2008) 10F301. 2 T. Yanagida, T. Itoh, H. Takahashi, S. Hirakuri, M. Kokubun, K. Makishima, M. Sato, T. Enoto, T. Yanagitani, H. Yagi, T. Shigetad and T. Ito, Nucl. Instr. Meth. A 579 (2007) 23. 3 J. C. van't Spijker, P. Dorenbos, C. P. Allier, C. W. E. van Eijk, A. R. H. F. Ettema, G. Huber, Nucl. Instr. and Meth. B 134 (1998) 303. 4 L. Pichat, P. Pesteil, J. Clément, J. Chim. Phys. 50 (1953) 26. 5 J. Dannin, S. R. Sandler, B. Baum, Intern. J. Appd. Radn. Isotopes 16 (1965) 589. 6 Q. Lin, B. Yang, J. Li, X. Meng, J. Shen, Polymer 41 (2000) 8305. 7 M. Hamel, A.-M. Frelin-Labalme, V. Simic, S. Normand, Nucl. Instr. and Meth. A 602 (2009) 425. 8 http://www-nds.iaea.org/reports/nds-195.htm 9 M. Hamel, S. Darbon, S. Normand, G. Turk, French Patent Application (2010/12/21). 10 P. A. Cahill, Radiat. Phys. Chem. 41 (1993) 351. 11 I. B. Berlman, Y. Ogdan, Nucl. Instr. and Meth. 178 (1980) 411. 12 J. M. Flournoy, C. B. Ashford, in Liquid Scintillation Counting and Organic Scintillators, 1989, p. 83. 13 M. Hamel, V. Simic, S. Normand, unpublished experimental results. 14 Z.H. Cho, C.M. Tsai, IEEE Trans. Nucl. Sci. 22 (1975) 72. 15 T. Depireux, F. Dumont, A. Watillon, J. Colloid Interface Sci. 118 (1987) 314. 16 A. Koch, C. Raven, P. Spanne, A. Snigirev, J. Opt. Soc. Am. A 15 (1998) 1940. 17 D. E. Groom, S. R. Klein, Eur. Phys. J. C15 (2000) 163. 18 M. Hamel, V. Simic, S. Normand, React. Funct. Polym. 68 (2008) 1671. 19 E. H. Eberhardt, Appl. Opt. 7 (1968) 2037. 20 G. Turk, C. Reverdin, D. Gontier, S. Darbon, C. Dujardin, G. Ledoux, M. Hamel, V. Simic, S. Normand, Rev. Sci. Instrum. 81 (2010) 10E509. 21 www.detectors.saint-gobain.com 22 E. Mihóková, M. Nikl, J. A. Mareš, A. Beitlerová, A. Vedda, K. Nejezchleb, K. Blažek, C. D’Ambrosio, J. Lumin. 126 (2007) 77. 23 P. B. Lyons, S. E. Caldwell, L. P. Hocker, D. G. Crandall, P. A. Zagarino, J. Cheng, G. Tirsell, C. R. Hurlbut, IEEE Trans. Nuc. Sci. 24 (1977) 177. 24 T. M. Undagoitia, F. von Feilitzsch, L. Oberauer, W. Potzel, A. Ulrich, J. Winter, M. Wurm, Rev. Sci. Instrum. 80 (2009) 043301. 25 G. Blasse, B.C. Grabmaier, in Luminescent Materials, ed. Springer (1994). 26 E. A. Andreeshchev, V. S. Viktorova, S. F. Kilin, K. A. Kovyrzina, Y. P. Kushakevich, I. M. Rozman, V. M. Shoniya, Opt. Spectrosc. 57 (1984) 624. - 11 -