NMR Training The goal of initial NMR training is to have a list (at least a mental list) of things you need or want to learn. Learning NMR never ends, in the sense there will always be more that NMR can do for you in your work. Think of the console training as going to office hours with questions before an assignment is due, rather than a test to be passed. NMR is “learn by doing”. It is a practical skill more than a theoretical one, although some understanding of “what’s under the hood” of a modern (black box) spectrometer is essential. What has worked best in training is to cycle repeatedly through the basic rote procedure of acquiring a simple 1D spectrum, filling in more details as the procedure becomes familiar, the questions arise. This primer addresses: Safety and local etiquette, necessary to get started, then Gaps often encountered in intermediate users’ knowledge, and “urban legends”. (Absolute beginners should ask for the tutorials of the chem4639 handouts). When the defaults will not be what you want. A menu of intermediate and advanced skills you can pick from, for your next goals. Things to know before you begin operating an NMR NMRs are expensive. They are delicate. Negligent use can be disastrous. Most damage is avoidable, some is not. If you understand how the instrument works, you will also develop some feeling for why things should be done a specific way. If, on the other hand, you do things your own way and disregard local procedures, you run the risk of damaging the instrument. Repairs are often thousands of dollars, but the greater cost is if research and classes cannot use the NMR for several weeks. If you are a beginner, relax: problems are rarely caused by attentive beginners. If you already know NMR, pay special attention: every NMR is different. The software can be set up in many different ways for a different user base or for different hardware capabilities. The shortcuts or procedures you learned at another site may not work here (may actually be dangerous, if not set up properly). Each instrument has idiosyncracies to work around -- the calibrations are certainly different. Ignoring the local etiquettes of how to leave the instrument ready, could effectively shut down the NMR. Your experience hopefully puts you further along in understanding what underlies the steps of locking, shimming, acquiring and processing, independently of the specific Varian or Bruker commands to do them. And so you may get up to speed more quickly. But if you want access to this instrument, you must be willing to work within our user community. [ insert homer simpson picture ] Basic Magnet Safety: NMR magnets are always live and always at field. They cannot be turned off like a light switch. Therefore, nothing ferromagnetic is allowed near them (within 3 ft of the 400). At the top of the list are tools and other heavy objects (wrenches, screwdrivers, metal chairs, and floor buffers). These can dent or puncture the magnet dewar, or cause a quench: the magnet is drawn to the object as strongly as the object is drawn to the magnet, if the magnet hits the side of the dewar it can trigger a quench. Also important are small object that can fly into the bore: paper clips, staples, bobby pins, metal barrettes, costume jewelry, or wallet chains. Items such as iPods, cell phones, and media storage devices can also be damaged by the magnetic field. ATM and credit cards may become unusable. Don’t get sloppy because the AS in AS400 means “actively shielded”, so it has 1/5’th the fringe field of a common (unshielded) magnet, because the next magnet you see won’t be. Please practice good magnetic hygiene and leave all of these items in your lab or next to the instrument’s computer when you approach the magnet. Leave the door open when you are in the room. This is a good habit in general around cryogens, but especially an NMR. The magnet contains a tremendous amount of energy that could be released as heat, boiling off a lot of liquid helilum and displacing a swimming pool volume of air in a few seconds. Such a quench happens rarely (about once per century per magnet), but if the magnet starts to blow like a steam whistle, immediately follow your instinct to run away out the open door. In low oxygen you black out without warning (you don’t get short of breath unless there’s high CO2). Don’t waste time learning software: learn what the commands do. It is possible to take spectra by memorizing commands, like cheatcodes to a video game, but that knowledge will be quickly obsoleted by a new version of software. Put a little thought into what the data would look like as a spreadsheet, and what each command does to the data. That knowledge is forever, and portable. What happens if you learn on a Varian but they have a Bruker at your first job? Asking "Where’s the chloroform lock?' or "what shims are responsive on this instrument" makes a good impression; In contrast, "what button do I push", not so much. Focus on learning your immediate need, but recognize you can keep learning NMR forever. To be a competent chemist you need an appreciation (not necessarily fluency) of the things you don't know that support your professional expertise. (Or produce the artifacts in your spectrum you must explain or avoid.) Blaming your tools is unprofessional, merely reveals ignorance. Pay attention, watch other students, ask questions, compare notes. . There are also a lot of "urban legends" handed down in NMR. Like "lock power should be less than lock gain". There is a natural tendency to be shy about one’s lack of NMR knowledge, to avoid risking losing access to the NMR, or perhaps to save face. The opposite is true. Asking naïve questions will get you an answer tailored to your understanding and goals. Accounts only get cancelled if a user i) presents a problem or risk from not knowing something; and ii) persistently resists learning it after being told how and why it is important. Don’t be shy about having questions about NMR or software. I’ve been at this for twenty years, and I still see new commands or capabilities of vnmr software I’ve never known were there. About half of the problems that come to me require looking up or figuring out new things. Understanding what you are doing on the console can get into a lot of interesting subjects: synthetic chemistry, signal processing, computer programming, math, physics. “No day without a tip” This document outlines the steps of taking an NMR spectrum the way we do it here. it relies more on the command line than on automation. The basic steps are : Making the sample Inserting the sample -- Making samples. Use deuterated solvents. Use a solvent with TMS. Chemical shifts are (by definition) with reference to TMS. Referencing to solvent (e.g. chloroform at 7.26 ppm) is ok for a quickie, but it is not publishable. TMS evaporates readily (it will disappear from samples in a week or so, even from a capped tube. It is trivial to add TMS to a sample by pouring vapors from a bottle of TMS over an open NMR tube. Non-deuterated solvents can be suppressed, but this is an advanced technique – about 2 hr to learn and ½ hr each time you do it. The sample volume should be 700 uL -- at least 4 (preferably 5 cm) in the NMR tube. Making it smaller is hard to shim: you will get worse S/N and resolution. 3 cm samples only make sense when spending 4 hr or more acquiring a severely masslimited sample, for expert at shimming, with time to spend a half-hour at it. NMR tubes vary from $1 "disposable" tubes to $20 or more. Cheap tubes are fine for most routine synthetic work, as long as they are straight enough, so they don’t scratch the probe when they spin. Expensive tubes can be shimmed to better linewidth: they have more uniform wall thickness. Cheap tubes can't be shimmed as tight and may give spinning sidebands of a few percent. But they are fine for most work. It is important that the tube not be curved, or it can scratch the probe when it spins. Wilmad HIP tubes are fine. Even a tube that rolls true on a flat surface can be a problem. If in doubt, bring a tube to the NMR500 and I have a tester (a straight, tight sleeve: if the tube fits in the sleeve and spins smoothly, it is ok as far as the instrument is concerned.) Advice on washing NMR tubes is given in an appendix. -- Correct positioning of the NMR tube in the spinner. Clean the tube thoroughly after inserting it into the spinner. Scrub it vigorously with an acetone-soaked kimwipe, then dry off the dirty acetone with a dry section of the Kimwipe. There is a film of something on most new NMR tubes that shows up in spectra acquired for 5 min or longer. The O-ring in the spinner was put there by the devil to coat your sample with a mix of silicone oil, water bath fungus, skin flakes, sharpie ink, cellotape adhesive. (Is chemistry the only profession where you wash your hands *before* you go to the bathroom?). Even if you are taking a quick spectrum of a concentrated sample, keep that gunk out of the probe. Plean clean all samples, habitually. No exceptions. If the NMR tube slides too easily in the spinner, use the other spinner and email Roger that the O-ring needs replacing. When the spinner drops into the magnet, the tube can keep going and bottom out in the probe. Position the clean tube using the depth gauge. Center the sample on the CL mark, but not lower than the maximum depth (bottom of the guide). The sample must not only completely fill the detection coil observed volume (dotted box in guage) but the ends of the sample should be at least 1 cm from the coil. The magnetic field bends in & out of the sample at the ends (the sample/air interface) these cannot be shimmed. If the interface is 1 cm away, you can shim quickly with Z1C and Z2C to under 1 Hz. If the interface is closer, you need to shim Z1 and Z2 ten times at different Z3 and Z4 settings to find the best combination. If centering a large sample would put it deeper than the max depth, then the ends will be far enough away the asymmetry doesn't matter. Please do not mess with the max depth adjustment of the depth gauge. It should be 69 mm for the 4nucleus probe; 65 mm for the broadband probe (backup probe). -- NEVER DROP A SAMPLE INTO THE MAGNET WITHOUT THE LIFT/EJECT AIR FLOWING. Get in the habit of feeling the sample float before letting go of it. Turn the lift air on with software. (The manual switch is only there if an sample must be removed when the computer is down; it requires higher air pressure and won't support a sample). - Spin the sample at 20 Hz. Spinning averages out poor X or Y shims, making it easier to get good Z shims on the first pass. It typically sharpens peaks from 1.3 Hz (nonspinning) to 1.0 Hz (spinning), which is hardly noticeable in most spectra. Spinning can introduce spinning sidebands around your peaks, and noise in 2D spectra. It is best turned off to see small peaks near big solvent peaks, and for all 2D spectra. - Locking Locking does three things. 1) You make a fine adjustment of the magnet field (Z0) to tune your sample to the "tuning fork" in the RF cabinet (spectrometer frequency), so our spectrum is in the spectrometer’s spectrum window (sw); 2) When the lock is turned on ("field frequency lock") the spectrometer will then auto-adjust Z0 to compensate for trains, elevators, or magnet drift that would otherwise move the lines while the spectrum is acquiring, and thereby blur your spectrum. 3) The lock level gives a signal to shim on. The lock display on a Varian shows the "difference frequency" between the sample and the tuning fork. You adjust Z0 to make this difference 0 (step function). Each hump in the display is about 10 Hz of frequency, and is about one click of the 16's button of Z0. You will see 2 humps if you are either 20 Hz too high (sharp) or 20 Hz too low (flat). A complication is that the intensity of the trace is random. Try it: scan small values of Z0 (e.g. the 4's button) you will see sometimes the humps are large, but often small or 0 (it depends on the relative phase of the tuning fork and the signal from the ping'ed sample). There is an urban legend that lock power should be lower than lock gain. What is true is the lock power must not be too high. There is a specific number for each solvent (written in a chart near the spectrometer). If the lock power is too high, it will saturate (suppress) the solvent peak instead of stimulate it. If this happens the lock level will oscillate wildly, and will not reflect how good the shims are. The gain can be any number that puts the lock level between 20 and 90%. Another urban legend is you should shim to 70% lock level. -- Shimming Shimming means adjusting the magnetic field to be exactly the same at the top, middle, and bottom of the tube - within one ppb (one part-per-billion). Note 1 ppm on your NMR spectrum would be one ppm of variation in the magnetic field. ) Shims Determine Linewidth The Z direction is the long axis of the sample; X and Y are transverse. Shimming is an art. Explore it on your own. Lesson 1 below should get you decent shims in a couple of minutes. The field must be the same at the top, middle, and bottom of the tube. The lineshape you get is the sum of all the volume elements 2 Hz marginal; 1 Hz good; 0.5 Hz limit B(z) ppm ppm type bestshim to load a good starting shim set. Start with the 4's button of Z1C. if it goes 4 clicks, check Z2, z and go 2 clicks. Good Homogeneity When you go through the maximum, step back to it using the 1's button. Get to a point where any Z1 change or z2 change reduces the lock level. Bad Homogeneity Reduce lock gain if the lock level goes above 100% (and you can't see changes). You can click about once per second for CDCl3 or D2O. With acetone or methanol, wait for the lock level to stabilize before clicking again – about 3 sec. The lock level is the height of the solvent deuterium line. The area is fixed, so the height depends on the linewidth. You should be able to get good (1 Hz) shims using the 4's button on Z1C and Z2C The 64's button of Z1 is about equal to the 1's button of Z1C. If you start shimming on Z1, it may hit its limit (of 2048) before you are shimmed. The vnmr software supports a very wide range from fine to coarse, because it is the same software used for many NMRs -- from microNMR up to imaging magnets. Additionally, on some spectrometers the shim values range from -32K to +32K, others (like the 400) only from -2K to +2K. Z1 to Z5 make corrections as a polynomial series ( f(x) = c + c1x + c2x2 + c3x3 + c4x4 + ...) The bottom line in shimming is whether the lineshape of your H1 or C13 spectrum is good enough to see what you need to see. The TMS line is usually sharp, and good to check your shims on. TMS can be shimmed to 0.7 Hz (but getting 0.7 Hz shims on TMS may be a waste of spectrometer time if others are waiting). Your compound will have a natural linewidth from 1.2 to 1.5 Hz. After a quick shimming, start acquiring, and check the TMS line. If yourshims aren’t good enough, abort acquisition and shim some more. The most common reasons for shimming problems are: Bad starting shims (type bestshim and continue) The sample is too small (dilute it to 4 cm) The sample was not centered on the coil (reposition and re-insert) A bad tube (throw it out and use a new one) The sample precipitated, or didn’t dissolve. You added solvent to solid in the NMR tube, and there is a high conc at the bottom (shake it up). -- Tuning. We don’t tune the probe of the 400. Just be aware that on many spectrometers it is a good idea to tune the probe with your sample inserted. It is mentioned in passing here so you won't look stupid if they do it at your next job. -- Set Up Acquisition Typically click (MainMenu) > Setup > H1, CDCl3 (click other buttons as appropriate for different solvent, etc.) You can also load a previous parameter set, just retrieve a spectrum, then type su ('setup") to load them from the Sun Computer into the spectrometer (the RF cabinet). Generally you acquire your proton 1D into exp1. If you will be taking a 2D, you copy your 1D paramters into another experiment directory (exp2), join exp2, then setup the 2D. If you load a parameter set for CDCl3, and your sample is in DMSO, some of your peaks may be outside of the spectrum window. Either click setup again for the correct solvent, or type solvent='DMSO' Set the number of scans, e.g. nt=16. If your sample is dilute, increase the number of scans (transients). 10 times the number of scans takes 10 times longer but ives 3 times the S/N. Practical limits are: H1 spectrum of 10 ug overnight C31 spectrum of 10 mg overnight. There is a tradeoff between sensitivity and resolution. If your sample is dilute, you can get better sensitivity (S/N per hour) if you take small pulses quickly: d1=0, np=16k, pw(30) If you need accurate integrals increase time between scans: d1=3, np=32K pw(60) Type dps to see your current pulse sequence. -- Start the acquisition type ga. -- Annotate the spectrum. While waiting for acquisition, type gettext. A minimal text file includes, file name (sample name, if different), your name and date, and location where data is stored, e.g. LM-Bz-ascorbate-top in CDCl3, .05% TMS John Smith, 23 August 2011 NMR400:smith Any samples in your refrigerator or shelves without name and date are worthless and disposable, the same is true of NMR data. If you don't record where data can be found, it doesn't exist, you never completed the synthesis, and shouldn't get paid or recommended for a job. 3x the sample will acquire 10x faster; 10 times the sample will acquire 100x faster Set the reference as soon as there are enough scans to see it. It is possible to copy and paste the filename from the text file into the "save FID" dialog. It is a good habit to save the data as soon as it is done (in case it is accidentally deltered, e.g. by another ga), and then to save it again after processing (phase, set reference, set integrals, ) While data is acquiring, look at the FID (type df). This is the signal recorded by the spectrometer. It is essentially an oscilloscope trace of a bell ringing, and fading out over a second or so. A plot of amlitude vs. time. To see what frequencies are present, you need to do a Fourier transform. - Display and Plotting The most noticeable skill on the spectrometer is your ease to zoom in and out of regions of the spectrum, measure couplings and frequencies, and create plots that can make a point, sharply. This is something everyone manages to learn well. In this draft I prioritize the things people miss and can’t move on without. Things You Have Yet to Learn Bilge Rat Galley Slave Deck Ape Able-Bodied How to take a simple 1D proton How to lock on different solvents How to shim Z1 and Z2 together (second-order optimization) how to shim higher-order shims Reduce noise with line broadening (what is an FID?) Identify water and other solvents in spectra. Plots with expansions or stacked plots Export plots to powerpoint or word processor. Use multiple exp’s (e.g. acquire in exp1, process in exp2) Acquire multiple spectra overnight Retrieve saved data Predict Spectrum of a chemical structure Back up your data (copy to another computer – ssh) Off-line data processing (ACD, Mestre, spinworks) Basic unix (list directory, change directory, copy files) Acquire 13C spectrum. Set parameters for sensitivity vs. accuracy Measure a 90-degree pulse (prelim for all 2D). Set gain. (describe signal path) Basic 2D spectrum (e.g. gCOSY) – setup, process, plot Ship’s Mate Measure C13 90o pulses Set decoupler power 2D proton-carbon spectra Create and use shaped pulses Use pulsed field gradients -- When the Defaults Are Not What You Want There will come a day when the defaults won't work for you. The default parameters are a compromise between sensitivity and accuracy. One day you may have barely enough material to get a spectrum overnight, and need to squeeze out another factor of 2 in sensitivity. Another day you may have a contaminant, and you need more accurate integrals to distinguish the two mixed compounds. For sensitivity set a shorter time between pulses, (and use smaller tip angles); For quantitative accuracy, set a longer time between pulses. This section very briefly describes the relevant of the parameters, and gives brief samples of the principal tools for understanding them: The pulse sequence, The vector model A 90 degree pulse calibration At right is how we sketch the basic 1D pulse and acquire sequence. The sample is pulsed to stimulate emission of the signal, which is acquired and recorded, until it dies out. pw pulse width d1 at relaxation delay acquisition time You can see your pulse sequence on the Varian by typing dps. time This is repeated and the signal added: Sample Free Induction Decay (FID) Recorded Relaxatio n Delay time Typical defaults would be: a 45 degree pulse, 1 second acquisition time 1 second relaxation delay What’s a 45-degree pulse? The angle refers to the vector model. 30o pulse At equilibrium Pulse Width 1 µsec 45o pulse 3 2 30o pulse 60o pulse 6 5 4 90o pulse 45o pulse 8 7 60o pulse 90o pulse Figure: At top, the magnetization of the sample is tipped more, by longer pulses, up to 90 degrees. The intensity of the spectrum (bottom) is proportional to the xy component of the tipped magnetization. For 1D NMR the pulse is not critical – very little difference is seen in the spectrum whether you used a 45 degree or 60 degree pulse. For 2D NMR, however, knowing the length of the 90-degree pulse is critical. The default calibration is an approximate guess. For 2D NMR the 90 degree pulse is calibrated by taking a series of spectra, with increasing pulse lengths, as shown below: 90o 0o 2 µsec 4 6 8 180o 10 13 16 360o 270o 19 21 24 27 30 A spectrum line when excited by pulses of indicated length. The 360-degree pulse above is 32 usec, so the 90-degree pulse is 8 usec. 33 FID (free induction decay, “time domain”) np at =1 sec =16K points another tremendous degree of optimization and control comes in the process of the Fourier Transform. At left is the data recorded ft Fourier transform Spectrum (“frequency domain”) np =16K points The FID can be thought of as a sound recording of a bell ringing, and dying out in a second or so. (Actually, a chorus of bells). The NMR spectrum is plot of the frequencies present in the chorus, obtained by a Fourier transform of the FID. There are some processing tricks in how you do this Fourier transform that can improve your S/N significantly, or can sharpen peaks if you need to resolve multiplets or measure coupling constants. The Fourier transform algorithm in the software requires that the number of points must be a power of two. (If the number of data points acquired is not a power of two, zeroes may be added to the end of the FID. Typically it is 16 Kb or 32 Kb. ) The number of points in the spectrum is the same as the number of points in the FID. A surprising fact is adding more points to the end of the FID (making it longer) increases the density of points in the spectrum, rather than making the spectrum wider. We can use this if you want to see narrow lines or 2 Hz splittings: zeroes may be added to the FID to get better digital resolution (more points on each peak) in the spectrum. The above is a cartoon version of the first principles, to help you understand what the commands are doing. You don’t need to understand them to get started – and probably can’t understand them until you get started. But you will need to become a competent NMR spectroscopist if you want to become a competent chemist -to justify the structures of your compounds in publications.