Visual field screening and the development of a new screening program David B. Henson, Ph.D. Abstract: The diagnosis of chronic open-angle glaucoma requires diligent and careful screening. Intraocular pressure and disc evaluation, when used in isolation, produce an unacceptably high number of false positives and false negatives, while full visual field analysis, as conducted with kinetic or full threshold static perimetry, takes too long for routine use. Screening of the visual field may offer a viable I alternative. The effect of : changing certain parameters of m visual field screening test are -evaluated. These include the (location of the stimuli, the number of stimuli and their suprathreshold increment. The results of an optimized visual field screening test are then evaluated with respect to the theoretical predictions. The results indicate that high levels of sensitivity and specificity can be achieved with visual field screening tests which are enough to be included as of the routine optometric examination. words: visual fields, ^Screening, glaucoma The increasing demands that have been placed upon the optometric profession to detect and diagnose disease have encouraged practitioners to perform more and more diagnostic procedures as part of their routine eye examination. One such procedure is a visual field examina- tion. While a very good case can be made for a full field examination to be performed on every patient, the time taken to perform this type of test is so great that most practitioners view this option as unrealistic. One alternative to performing a full visual field examination is to screen part of the visual field with a test that only takes a few minutes to conduct, and then to perform a full examination on those patients who fail the screening test. A second alternative is to apply preconditions such as raised intraocular pressure (IOP), abnormal disc appearance or a family history of glaucoma before ordering a full field examination. At present it is impossible to say which is the best of these two approaches to adopt, as we lack the precise data bases necessary to calculate the relative merits of different screening programs vs. different sets of preconditions. Within the literature, however, there are several reports discussing the parameters, such as number and position of stimuli, which affect the sensitivity and specificity of visual field screening programs. This paper reviews the data from a number of these studies and then briefly describes a new piece of visual field equipment with a screening program based on the results of these research programs. Nearly all the presented data relates to chronic open-angle glaucoma. The reasons for this are twofold. First, glaucoma exists in a fairly large percentage of patients (approximately 0.5% of the population), and it is therefore feasible to build up sufficiently large data bases to enable certain hypotheses to be tested. To build up data bases for other, more rare conditions, would require enormous research pro- grams that, at present, have not been conducted. The second reason is that chronic open-angle glaucoma is often asymptomatic. Its detection requires diligent and careful screening. Many of the other conditions which give rise to visual field loss also give rise to symptoms which, in themselves, would indicate the need for special diagnostic procedures and further evaluation. Sensitivity and specificity of IOP and optic disc evaluation Optic disc evaluation, IOP measurement, and visual field assessment still currently form the basis of glaucoma diagnosis. Before describing how sensitive and specific different visual field screening strategies are, this review will briefly discuss IOP and optic disc evaluation. With this information at hand, the clinician will be able to make a personal judgment as to the relative merits of the three types of test. There exists a considerable amount of data on the sensitivity and specificity of IOP measurement. Most of the epidemiological studies reveal1'4 that approximately 50% of the patients with glaucoma (usually defined as a repeatable visual field defect allied with a cupped disc) will, at the time of initial examination, have an IOP of less than 21 mmHg. Figure 1 gives the results from four of these studies. It is important to emphasize that the results given in Figure 1 are from unselected population groups. Studies that look at selected populations, such as those attending a glaucoma clinic, are often biased in favor of IOP measurements since many practitioners use an elevated IOP as a criterion for referral. Volume 60, Number 12,12/89 893 Bengtsson, 1981 12 c a s e s Cockburn, 1985 27 cas es Framingham, 1980 71 cases Armaly, 1980 98 cases Figure 1: Results from four epidemiologies! studies giving the number of glaucoma cases in each study and the proportion of glaucoma cases which, at the time of examination, had low IOP.1"343 The specificity of IOP measurements is also very low. The Framingham Eye Study3 found that 7.6% of the population had pressures >22 mmHg, and that the vast majority of these patients were nonglaucomatous. Put another way, if a practitioner relied entirely upon IOP measurement as a criterion for referral, and referred all patients with an IOP >22 mmHg, he or she would refer 14 normals for every patient with glaucoma and would miss 50% of the glaucoma cases. Data from optic disc evaluation trials again reveal that the sensitivity and specificity of disc evaluation is less than perfect.5 While studies have shown that disc changes often precede field defects, the types of change that occur are usually only detected when serial photographs, taken over a period of several years, are analyzed.6-7 When Lichter8 presented highly skilled practitioners with a set of photographs and asked 894 them to differentiate glaucomatous eyes from normal eyes, he found that 20% of them were misclassified. Optimized visual field screening tests When designing an optimized visual field screening test there are several things that need to be considered. Type of visual field examination Static, suprathreshold techniques have repeatedly been advocated for visual field screening.9'12 They combine a high sensitivity with a short examination time. Kinetic tests, such as those conducted with the Goldmann perimeter, while being valuable for monitoring the extent of a defect, have a low sensitivity to defects.13 The quantitative threshold tests, popularized with the Octopus and Humphrey instruments, while furnishing more information Journal of the American Optometric Association concerning the depth of any defect, take longer to perform and do not appear to have any greater sensitivity.9 Some of the early suprathreshold static instruments used stimuli of constant intensity across the whole visual field.14 It is now realized that a reduction in errors will occur if the intensity increases toward the periphery at a rate which mirrors the retina's loss in sensitivity. In this situation the stimuli will be at a constant suprathreshold value at all retinal locations. Multiple stimulus vs. single stimulus In the multiple stimulus technique, which was pioneered by Harrington and Flocks when they developed the Harrington-Flocks Screener," patients are presented with patterns of two, three or four stimuli. They respond verbally, stating the number of stimuli that they see in each pres- entation. This technique has several advantages over the single stimulus technique: it speeds up the examination time, it maintains the patient's attention somewhat better, and by being semi-automated,16 it promotes a dialogue between the patient and the perimetrist which has certan psychological advantages.17 It has also been reported as giving a lower threshold than that of a single stimulus technique with a smaller standard deviation of results.18 negatives — would be at a minimum. As can be seen from Figure 4, this would occur with a test which had approximately 20 stimuli. Thus, the ideal number of stimuli to have in a screening test, given that the clinician equally weighs false positives and false negatives, would be 20. Many clinicians would argue, and rightfully so, that false positives (a normal person failing the test) and false negatives (a glaucomatous person passing the test) are not equal. They would be willing to have a higher number of false positives, providing there were fewer false negatives. This situation can be accommodated, providing the clinician specifies the relative weight given to false positives and 25 Location of stimuli To decide where to optimally place the stimuli in a visual field screening test for glaucoma, it is essential to know where glaucoma defects normally occur and whether there is any relationship between the size of the defects and retinal location. This information was provided in an earlier paper19 which examined the results from 104 eyes with early glaucomatous visual field loss. Some of the results from that paper are presented in Figure 2, which gives the frequency of defects within the central 30° of the visual field. This data can be used to derive the relationship between the number of stimuli in the visual field screening test and the sensitivity of that test. Each distribution of stimuli is optimized to the location of glaucomatous defects.20 The results of such an evaluation are given in Figure 3, where it can be seen that the sensitivity of the test is logarithmically related to the number of stimuli. Figure 2: Frequency distribution of early glaucomatous defects in 109 eyes. The highest peak, in the superior temporal region, represents 29 eyes. Data collected with Friedmann Mark II Visual Field Analyzer.19 Number of stimuli The data presented in Figure 3 can be used, along with data giving the specificity of a visual field test,21 to derive the receiver-operator characteristics of a visual field screening test. The receiver-operator characteristics are shown in Figure 4 for a test in which missing one nonblind spot stimulus constitutes a failure. A line drawn at 45° to the axes and tangentially to the curve touches the curve at the point where the number of errors — false positives and false Figure 3: Relationship between the sensitivity of a visual field screening test and the number of test stimuli. The failing criteria was set at one or more missed stimuli at a suprathreshold intensity 0.5 log units above the patient's threshold. Data was collected with a Friedmann Mark II Visual Field Analyzer.20 Volume 60, Number 12,12/89 895 false negatives. For example, if the clinician weighted a false positive as being half as important as a false negative, then rather than draw the line at 45° to the x axis, it should be drawn at 30°, where the ratio of the angle between the axes equals the relative weight given to the different types of error. As can be seen in Figure 4, the ideal number of stimuli to incorporate in the screening test has now increased to 30. Suprathreshold increment The extent of the suprathreshold increment will affect the overall performance of a threshold-related, suprathreshold test, one in which an estimate of the threshold is obtained at the beginning of the test before the intensity is stepped up to its suprathreshold value. If a suprathreshold increment is chosen that is only marginally above the patient's threshold, then there would be a large number of false positives, as it is highly likely that the patient will miss, by chance, one or more of the stimuli. If the suprathreshold increment was made very large, then the number of false negatives would become high as patients with relative defects may, as a result of the high intensity stimuli, see all the stimuli and pass the screening test. The relationship between suprathreshold increment and the number of errors — false positives and false negatives — can be quantified with a mathematical technique called information theory. The results from such an analysis indicate that the total amount of information in a test increases with the suprathreshold increment up to 0.6 log units, and then decreases beyond this level. These results indicate that suprathreshold increments of approximately 0.6 log units would be the best to choose for an optimized screening strategy. Quantifying visual field defects An aspect of visual field testing that Figure 4: Effect upon sensitivity and specificity of varying the number of stimuli in a visual field test. The numbers placed along the solid curve give the number of stimuli in a visual field test, all optimally located to detect the largest number of defects. As the number of stimuli increases, the sensitivity of the test increases, while the specificity, probability of a normal patient missing one or more of the stimuli, decreases. A: The broken line, drawn at 45° to the abscissa, touches the curve at the point where the number of stimuli in the screening test would give a minimum number of errors, false positives and false negatives. B: The broken line, drawn at 30° to the abscissa, touches the curve at the point where the number of stimuli in the screening test would give twice as many false positives as false negatives.30 896 Journal of the American Optometric Association has received a great deal of interest in the past few years is that of quantifying the results of an examination. There are two basic types of quantification: the first gives you an indication of whether a defect exists, while the second gives you a measure of the extent of any defect. In visual field screening tests we are primarily concerned with the former, i.e., whether a defect exists. The latter form of quantification is useful once a defect has been established to ascertain whether it has progressed or not and whether any treatment has been successful. Quantification systems are, in effect, a form of data reduction. They take the large amount of data that is produced in a visual field examination and reduce it to either one or a series of different numbers. When a single number is produced the result is often called the visual field score. But regardless of how you score the visual field — be it simply a count of the number of missed stimuli or the result of a more complex calculation — you will always encounter a certain degree of variability. There is, therefore, no error-free cut-off value of a visual field score that can be used to classify patients as either defective or normal. In deciding whether a patient does or does not have a visual field defect most researchers have chosen to compare the patient's visual field score with that obtained from a normal population. They then provide the practitioner with an indication of the difference between a given patient's score and that of the normal population. Over the years there have been a large number of papers describing different scoring techniques. Some have been developed for kinetic results such as those obtained with the Goldmann bowl perimeter.22'31 Other techniques have been developed to deal with the results from static techniques in which the threshold is measured at each stimulus location.32"35 Both of these two groups of quantification systems have primarily been designed to give a measure of the extent of a defect for either litigation purposes or for monitoring the progress of any field defects. They cannot be applied to suprathreshold techniques, which, as pointed out earlier, are the preferred ones for visual field screening. Crick36 and Demailly and Papos37 both described scoring systems for the multiple stimulus suprathreshold technique of the Mark I Friedmann Visual Field Analyzer, while Batko et al38 and Henson and Dix39 have developed systems for the Mark II instrument. The Mark II Friedmann Visual Field Analyzer differs from the Mark I in that it presents more stimuli (98 rather than 46) and has a more convenient technique for recording the data. The technique devised by Henson and Dix39 for the Friedmann was later modified and incorporated into the Henson CFS2000/ CFA3000.40'41 A characteristic of early glaucomatous visual field loss is that the missed stimuli in a static test tend to be clustered together. A patient who misses four neighboring stimuli is more likely to have a scotoma than one who misses four stimuli that are unconnected and scattered throughout the visual field. It is important that any quantification system which is designed to detect early defects can differentiate between these two situations. The value of cluster analysis has been demonstrated by both Chauhan and Henson21 and Henson and Dix,39 and it is expected that the next generation of computerized visual field equipment will incorporate this type of analysis in its screening programs. The number and type of calculations necessary to perform a cluster analysis cannot realistically be performed without the aid of a computer. Clinical trials with a new visual field instrument Much of the data obtained from the research reviewed in this paper has been used to design an optimized visual field screening program which has been incorporated in a new visual field instrument, the Henson CFS2000/CFA3000. This instrument's screening program uses multiple stimulus patterns in a threshold-related suprathreshold strategy, which is eccentrically compensated to the normal gradient of sensitivity across the retina. This program is divided into three stages: a screening stage in which just 26 stimuli are presented in locations optimized for the detection of glaucoma; an extended stage which increases the number of stimuli to 66; and a quantification stage which increases it further to 132 stimuli. The CFA3000 also includes a program that uses a full threshold technique with allied quantification techniques. The screening stage is designed for routine use in patients with no suspicion of visual field loss. This stage normally takes less than 3 minutes to perform on both eyes. The extended stage is designed for use on patients with some suspicion, such as raised IOP or a family history of glaucoma, and for those patients who have missed one or more of the screening stimuli. The quantification stage, as the name implies, is for use on patients in whom a defect has been found. It is used to map the extent of the defect and to give a numerical score that represents the extent of visual field loss. Clinical trials with the CFS2000/CFA3000 has shown that the screening stage has a sensitivity to glaucomatous visual field defects of 90% and a specificity of 88%.4M2 The extended stage, which would be performed on all patients who failed the screening test, has a specificity of 100%. Conclusion The diagnosis of glaucoma is currently based upon three clinical measures: an evaluation of the optic disc, a measurement of the IOP, and an investigation of the visual field. While many optometrists routinely examine the optic disc and measure lOPs, relatively few routinely investigate the visual field. The data reviewed in this paper have shown that high sensitivity and specificity levels can be reached when visual field screening pro grams are optimized to the detec tion of glaucomatous damage. Op tometrists frequently state that rou tine investigation of the visual field is not time-efficient. This statement is often based on experiences gained with instruments and programs that were designed to quantify loss rather than to screen the visual field. It is hoped that the data presented in this paper will indicate that visual field testing has come a long way in the past few years and that routine vis ual field screening is now both fea sible and desirable in optometric practice. •• Submitted for publication 11/88 Accepted 9/89 Dept. of Optometry University of Wales PO Box 905 Cardiff CF1 3YJ Wales Acknowledgment The author has a proprietary interest in the Henson CFS2000 and Henson CFA3000 visual field analyzers. References 1. Bengtsson B. The prevalence of glaucoma. Br J Ophthalmol 1981; 65:46-9. 2. Armaly MF. Lessons to be learned from the collaborative glaucoma study. Surv Ophthalmol 1980; 25:139-44. 3. Leibowitz HM, Kreuger DE, Maunder LR, et al. The Framingham eye study monograph. Surv Ophthalmol 1980; 24(suppl):335-610. 4. Cockburn DM. Glaucoma enigma. Am J Optom Physiol Opt 1985; 62:913-23. 5. Gloster J. Quantitative relationship between cupping of the optic disc and visual field loss in chronic simple glaucoma. Br J Ophthalmol 1978; 62:665-9. 6. Pederson JE, Anderson DR. The mode of progressive disc cupping in ocular hypertension and glaucoma. Arch Ophthalmol 1980; 98:490-5. 7. Sommer A, Pollack I, Maumenee AE. Optic disc parameters and the onset of glaucomatous field loss. Arch Ophthalmol 1979; 97:1444-8. Volume 60, Number 12,12/89 897