ANNEX 1 to GN-21: Change Notification Checklist This checklist is provided to assist the Registrant in determining the required documents for types of changes proposed. Submission of this checklist in MEDICS is mandatory. NOTE: All the required documents must be submitted for the relevant sections of the CSDT to support the proposed changes to the device. The documentary requirements in this checklist are meant to cover the broadest aspect for each category of change. If any required documents as defined in the respective category in this checklist are not available or applicable for the change proposed, please provide a clarification or justification as appropriate. Please also refer to GN-15 for the relevant templates/documents which may be required for the registered devices. Provided? Type of Change Documentary Requirements Yes All category of changes Completed Annex 1 to GN-21: Change Notification Checklist Completed Annex 2 to GN-21: Summary Table of Change Notification (As per Annex 2 to this guidance document) If no, please provide justification Provided? Type of Change Documentary Requirements Yes Completed Annex 3 to GN-21: Medical Device Safety and Performance Declaration (As per Annex 3 to this guidance document)1 1A Quality Management System certificate(s) All changes to manufacturing and/or sterilisation facilities with no changes to the manufacturing and/or sterilisation processes. Device labelling with changes redlined and annotated beside each amended section (if applicable) Declaration letter from Product Owner on company letterhead stating that there is no change to existing manufacturing sterilisation process Sterilization validation report (if applicable) 1 Except changes implemented as a result of reportable AE/ FSCA and If no, please provide justification Provided? Type of Change Documentary Requirements Yes If no, please provide justification For changes in manufacturing/ sterilization site of medical devices containing medicinal products in an ancillary role, please contact the Medical Device Branch, Health Sciences Authority for further advice. 1B Quality Management System certificate(s) All changes to manufacturing processes (including changes made to outsourced processes) that result in a change in medical device specifications of a registered Device labelling with changes redlined and annotated beside each amended section (if applicable) Summary of new manufacturing process Validation studies covering new processes Pre-clinical studies Software validation report (For software only) Clinical studies (For operating principles and design characteristics change) (if applicable) Risk analysis Provided? Type of Change Documentary Requirements Yes If no, please provide justification For changes in manufacturing/ sterilization site of medical devices containing medicinal products in an ancillary role, please contact the Medical Device Branch, Health Sciences Authority for further advice. 1C Sterilisation technique All changes to sterilisation processes (including changes made to outsourced processes) Device labelling with changes redlined and annotated beside each amended section (if applicable) Sterilization validation report (including the Sterilisation Protocol, Sterilisation Standards applied. Sterility Assurance Level, Sterilisation Revalidation report) Quality Management System certificate(s) Post Sterilisation functional test report 1D Updated All changes to Quality Management System (QMS) Certificate(s) Quality Management System Provided? Type of Change Documentary Requirements Yes certificate(s) for manufacturing and sterilisation Device labelling with changes redlined and facilities annotated beside each amended section (if applicable) Reason for cancellation of QMS certificate (if applicable) 2A Pre-clinical studies All changes to the control mechanisms, operating principles and/or design registered medical device characteristics of a Risk analysis Clinical studies (if applicable) Device labelling with changes redlined and annotated beside each amended section Software validation report (For software only) Detailed summary of software changes (For software only) 2B Pre-clinical studies If no, please provide justification Provided? Type of Change Documentary Requirements Yes All changes in specifications of a registered Clinical studies medical device (if applicable) Risk analysis Device labelling with changes redlined and annotated beside each amended section Software validation report (For software only) Detailed summary of software changes (For software only) 3A Pre-clinical studies All changes to biological materials that involve a change in type, source, processing and/or supplier of cells, tissues, and/or derivatives of animal, human, microbial or recombinant origin without a Clinical studies (if applicable) Biological safety data change in the intended purpose of the biological material Information of sources/donors If no, please provide justification Provided? Type of Change Documentary Requirements Yes 3B List of materials making direct/ indirect contact All changes to materials or material formulation (of with human body If no, please provide justification non-biological origin) including changes to device coating or surface modification techniques, that make direct/indirect contact with body tissues and Pre-clinical studies fluids, or is absorbed by the body 3C Information on radiation source All changes to materials that are used for shielding Information on materials for shielding of in medical devices emitting ionising radiation radiation 3E All changes to concentration or drug specifications (DS) of medicinal substances in medical devices that incorporate medicinal substances Contact the Medical Device Branch, Health Sciences Authority for further advice in an ancillary role 4A Pre-clinical performance evaluation data All changes to materials that affect performance Clinical performance evaluation data Provided? Type of Change Documentary Requirements Yes specification of a registered IVD medical device, including changes to biological materials that involve a change in type, source, processing and/or supplier of cells, tissues, and/or derivatives of Information on source of material animal, human, microbial or recombinant origin Biological materials (including supplier of materials) 5A Regulatory approval documents from the All changes to the labelling of medical devices that reference agencies for the change involve addition and/or revision of the approved indications for use Device Information Device labelling for new product(s) with changes redlined and annotated beside each amended section Declaration of conformity document Pre-clinical studies Clinical studies If no, please provide justification Provided? Type of Change Documentary Requirements Yes Risk analysis Software validation report (For software only) Changes that involve a reduction of indications for Reasons use not due to device safety, quality or efficacy indications for the reduction of approved considerations Device labelling for new product(s) with changes redlined and annotated beside each amended section 5B Reasons All changes to the labelling of medical devices that contraindications, warnings and precautions involve addition, removal and/or revision for or removal of of warnings, precautions and/or contraindications OR involves addition of contraindications, warnings Device labelling with changes redlined and annotated beside each amended section Pre-clinical studies and/or precautions not due to safety, quality or efficacy issues. addition Clinical studies Risk Analysis If no, please provide justification Provided? Type of Change Documentary Requirements Yes 5C Preclinical Studies Labelling changes that only - instructs the user to use the device in a different manner from that originally approved OR - involve a change from ‘professional use only’ to ‘home use’ Device labelling with changes redlined and annotated beside each amended section Clinical Studies Software validation report (For software only) Risk analysis 5D Labelling changes that involve clarification of Device labelling with changes redlined and existing instructions for use, not affecting safety, annotated beside each amended section quality or efficacy of the medical device If no, please provide justification Provided? Type of Change Documentary Requirements Yes 5E Labelling changes that only - involve changes in layout, colour, font sizes and design, without change in prominence of precautions, warning and contraindications OR - involve addition and/or removal of languages not Device labelling with changes redlined and annotated beside each amended section required by the Authority OR - involve the addition of reference agencies’ approvals (e.g. CE marking) 5F Device labelling with changes redlined and Other labelling changes annotated beside each amended section Details of changes and the reason for changes Documents supporting detailed above (if applicable) proposed changes If no, please provide justification Provided? Type of Change Documentary Requirements Yes 6A Justification for addition of product(s) to be Addition of new medical devices to a device listing grouped within the registered product based on GN-12: Guidance on Grouping of Medical Device for Product Registration Annex 2 for GN-17 and GN-18: List of Configurations of Medical Devices to be Registered Regulatory approval documents from the reference agencies Device Information Device labelling with changes redlined and annotated beside each amended section Declaration of conformity document Letter of Authorisation (GN-15) Pre-clinical studies If no, please provide justification Provided? Type of Change Documentary Requirements Yes Clinical studies (if applicable) Risk analysis Software validation report (For software only) Manufacturing information Unless the change Justification for addition of product(s) to be involves the addition of new devices of the same grouped within the registered product based design, within the existing range of sizes already on GN-12: Guidance on Grouping of Medical registered. Device for Product Registration OR involves a design change that does not affect the safety, quality or efficacy of the medical device (e.g. changes that improve device ergonomics, aesthetic modification of the device Annex 2 for GN-17 and GN-18: List of Configurations of Medical Devices to be Registered Regulatory approval documents from the reference agencies Device Information If no, please provide justification Provided? Type of Change Documentary Requirements Yes Device labelling with changes redlined and annotated beside each amended section Declaration of conformity document Letter of Authorisation (GN-15) Pre-clinical studies (where applicable) Software validation report (For software only, where applicable) Manufacturing information If no, please provide justification Provided? Type of Change Documentary Requirements Yes Unless the change Involves addition of Class A non-sterile medical Declaration on Registrant’s letterhead to state (1) The added models are class A non- device accessories that complement the registered medical device as a SYSTEM/ GROUP sterile (2) the name of the device affected (3) the device identifier (4) the name of Product owner for class A non-sterile device (5) name and address for the manufacturing site(s) for class A nonsterile device Letter of acknowledgement from the product owner of the Class A non-sterile device, that they have no objection that their said Class A non-sterile product is to be listed on SMDR as a part of the registered SYSTEM /GROUP. (applicable only where, the product owner of the Class A non-sterile device differs SYSTEM/GROUP) from that of the If no, please provide justification Provided? Type of Change Documentary Requirements Yes The quality management system certificate Unless the change Involves addition of Class A non-sterile medical device accessories that complement the registered medical device as a SYSTEM/ GROUP Declaration on Registrant’s letterhead to state (1) the added models are class A nonsterile (2) the name of the device affected (3) the device identifier (4) the name of Product owner for class A non-sterile device (5) name and address for the manufacturing site(s) for class A nonsterile device If no, please provide justification Provided? Type of Change Documentary Requirements Yes Letter of acknowledgement from the product owner of the Class A non-sterile device, that they have no objection that their said Class A non-sterile product is to be listed on SMDR as a part of the registered SYSTEM /GROUP. (applicable only where, the product owner of the Class A non-sterile device differs from that of the SYSTEM/GROUP) The quality management system certificate The quality management system certificate Annex 2 for GN-17 and GN-18: List of Configurations If no, please provide justification Provided? Type of Change Documentary Requirements Yes 6B/6C Justification All deletions of models from device listing product(s) to be grouped within the registered OR product Changes which increase or reduce the number of Propose: Justification for deletion of product(s) devices in a pack of a registered device or justification for addition of product(s) to be OR grouped with the registered product based on reduction in the number of models listed due to GN-12: Guidance on Grouping of Medical obsolescence and not due to safety, quality or Device for Product Registration. for addition or deletion of efficacy considerations Annex 2 for GN-17 and GN-18: List of Configurations of Configurations Device labelling for new product(s) (if applicable) 6D Declaration of Conformity document All changes to Product Name only (including identifier) Declaration letter from Product Owner on company letterhead to state that there is no change to device in all aspects, including intended use, technical specifications. If no, please provide justification Provided? Type of Change Documentary Requirements Yes Annex 2 for GN-17 and GN-18: List of Configurations of Configurations Device labelling with changes redlined and annotated beside each amended section Letter of Authorisation (GN-15) 6 E (i) Declaration of Conformity document All changes to Product Owner Declaration letter, from Product Owner on company letterhead, to state that they will undertake responsibility to provide post market support and assistance related to the medical devices <state device name> already supplied under the former product owner’s name (if applicable) Device labelling with changes redlined and annotated beside each amended section Letter of Authorisation (GN-15) If no, please provide justification Provided? Type of Change Documentary Requirements Yes 6 E (ii) Device labelling with changes redlined and A change in Product Owner Address annotated beside each amended section Letter of Authorisation (GN-15) 6F Declaration of Conformity document A change in regulatory status on rejection or withdrawal (in any reference agencies) for models included in the SMDR Listing. Existing regulatory approval certificate(s) Documents from relevant regulatory authorities citing reason for the change in regulatory status Reason for company to withdraw from regulatory authorities If no, please provide justification