Three-Bolometer Devices from CSO Run 1 : Time Constants Matt Gardner (Caltech) Version: 9 December 2000 Acknowledgements Christine Allen (NASA/GSFC) provided the bolometers. Jeff Groseth (Caltech) glued and wirebonded the dies. Matt Gardner and C. Darren Dowell (Caltech) performed all measurements. Thanks also to C. Darren Dowell for feedback and guidance. Apparatus Cryostat: Caltech Barney Dewar, which is IR Labs HDL-10 with helium shield + Chase Research single-stage 3He fridge. RFI filtered connectors with 1-4 nF capacitance. Detector holder: Up to 6 each 10 mm × 8 mm dies attached with GE varnish to fiberglass PC board with gold-plated copper traces. PC board screwed to Invar base and covered with lid. Black paint on inside of lid. Invar base attached with screws and N grease to copper cradle, which is attached to fridge coldhead with screws and N grease. Eight each gold wirebonds (0.001”) provide electrical interface to each die. Calibrated Lakeshore G.R.T. attached to PC board with screw and N grease. Methods: Load resistors and JFETs inside dewar are used during measurements. 1) “Direct” -- Bolometer is biased with a square wave between (Vbias - V/2) and (Vbias + V/2). Output response is amplified by a Stanford Research 560 preamp and displayed on an oscilloscope for measurement. Measurement Period: All measurements took place between 2000 Dec. 1 and 2000 Dec. 7. All device models used herein are taken from “Test Report -- Three-Bolometer Devices from CSO Run 1”, version 5, written by C. Darren Dowell. Direct Measurement For a given base detector temperature, device models can predict the bias voltage, Vopt, which minimizes bolometer NEP at 0.03 Hz. Bolometer responses are measured for Vbias/Vopt = .5, 1, 2, and 4 (the last value was omitted on Dec. 1) at several base temperatures. The bias frequency is selected so that a full thermal settling of the bolometer is visibly apparent before the next bias edge. Sample bolometer responses from December 1 are shown in the figures below; indicated in red on Figure 1 are the voltages measured for every observed response, while measured times are shown in blue. settle VOS Vrise VOS / e OS Figure 1: Bias voltage (bottom) and response (top) of bare 5273.7200 T2 device at Tbase = 502 mK, Vbias = Vopt, RL = 21.48 M; both signals are amplified by a factor of 100. Measured settle is 13.2 msec. Figure 2: Same as Figure 1, except Tbase = 282 mK. Measured settle is 22.8 msec. Table 1: Direct Measurement of Thermal Time Constants Thermistor Type Bol. Loc. Date Tbase (mK) Vbias (mV) V (mV)T (mK) G (nW/K) e/ settle (ms) 5273.7200 T2 bare bolometer mid Dec. 1 282 15.7 3.92 310 0.118 1.066 33.2 5273.7200 T2 bare bolometer mid Dec. 1 282 29.4 3.92 359 0.158 0.848 22.8 5273.7200 T2 bare bolometer mid Dec. 1 282 58.8 7.84 441 0.240 0.610 12.8 Die 5273.7200 T2 bare bolometer 5273.7200 T2 bare bolometer 5273.7200 T2 bare bolometer mid mid mid Dec. 1 Dec. 1 Dec. 1 397 398 398 21.6 45.1 88.2 3.92 3.92 7.84 417 459 528 0.215 0.260 0.345 0.903 0.758 0.627 29.2 14.9 10.1 5273.7200 T2 bare bolometer 5273.7200 T2 bare bolometer 5273.7200 T2 bare bolometer mid mid mid Dec. 1 Dec. 1 Dec. 1 501 502 502 37.3 74.5 147.1 3.92 7.84 15.69 521 561 635 0.336 0.389 0.500 0.896 0.769 0.649 37.2 13.2 12.2 5273.7200 T4 5273.7200 T4 5273.7200 T4 5273.7200 T4 bare bolometer bare bolometer bare bolometer bare bolometer mid mid mid mid Dec. 7 Dec. 7 Dec. 7 Dec. 7 283 283 283 283 16.5 30.2 61.3 119.5 1.03 1.03 1.03 4.06 301 349 456 567 0.144 0.187 0.300 0.441 1.212 1.214 0.657 0.515 N/A 27.0 8.0 6.1 5273.7200 T4 5273.7200 T4 5273.7200 T4 5273.7200 T4 bare bolometer bare bolometer bare bolometer bare bolometer mid mid mid mid Dec. 7 Dec. 7 Dec. 7 Dec. 7 384 384 384 384 20.4 39.7 80.6 160.0 1.94 1.94 1.94 2.88 405 446 528 638 0.243 0.289 0.390 0.545 0.961 0.820 0.625 0.531 17.2 14.8 7.6 4.6 5273.7200 T4 5273.7200 T4 5273.7200 T4 5273.7200 T4 bare bolometer bare bolometer bare bolometer bare bolometer mid mid mid mid Dec. 7 Dec. 7 Dec. 7 Dec. 7 505 506 506 506 33.9 68.8 136.8 272.6 3.91 3.91 3.91 5.78 526 569 650 773 0.387 0.445 0.562 0.765 0.901 0.764 0.630 0.552 NOTE: T, G, and e/ are derived from device models; cases where Vbias = Vopt are shown in bold. These same measurements were performed on several devices from wafer 5347.7500, as provided in 32-element test arrays by NASA/GSFC, in April 2000 (“PopUp Detector Test Report”, version 2, by C. Darren Dowell); of those devices, type HGUNIT.2B are equivalent in design to the bolometers considered here. Comparing these devices at Tbase ~ 300 mK and Vbias = Vopt, we findsettle ~ 4 and 25 msec for wafers 5347 and 5273, respectively. We take settle to be a measurement of the bolometer thermal time constant, e. Another appropriate time constant for device description is = C/G, where C is the bolometer heat capacity. Device models predict that for nearly all measurements here, e/ < 1 (see Table 1); we assume that this ratio should be the same for both wafers at similar Vbias and Tbase, and will therefore ignore its effects in further comparison. Again considering Tbase ~ 300 mK and Vbias = Vopt, models predict G ~ .3 and .2 nW/K for wafers 5347 and 5273, respectively. If this were the only difference between the two wafers, then wafer 5273 would have e larger by a factor of only 1.5. In order to account for the observed time constant increase, devices from the current dies must have 4 times the heat capacity of those tested in April -- specifically, C ~ 1 and 4 pJ/K for wafers 5347 and 5273, respectively. 10.0 11.3 8.5 4.6