ERRORS AND TREATMENT OF ANALYTICAL DATA Introduction: Every physical measurement is subject to a degree of uncertainty that can never be completely eliminated but, at best, can only be reduced to an acceptable level. The amount of uncertainty is often difficult to quantify and requires additional effort and good judgement on the part of the chemist. Nevertheless, it is a task that cannot be neglected because an analysis of totally unknown reliability is worthless. Thus the task of the analytical chemist goes beyond that of correctly performing the manipulations and taking the readings required in a procedure. To obtain meaningful results the following steps must also be carried out: 1. Results of each analysis must be properly recorded and calculated. 2. Since analyses are done in replicate (usually two to five times), the analyst must determine the best value to report. Although the best value is often the arithmetic mean, or average, of the individual results, there is often the question of whether to include a result that seems out of line with the others. 3. Finally, the analyst must evaluate the results obtained and establish the probable limits of error that can be placed on the final result. Unfortunately there exists no simple, generally applicable means by which quality of an experimental result can be assessed with absolute certainty. In this unit we will consider the types of errors encountered in analyses, methods for their recognition, and techniques for estimating and reporting their magnitude. Definition of Terms: Mean and Median: Individual results for a set of replicate measurements will seldom be identical and it is thus necessary to select a central, "best" value for the set. The central value of the set ought to be more reliable than any of the individual results and the variations among the results ought to provide some measure of reliability of the chosen "best" value. Either of two quantities, the mean or the median may be chosen as the central value of a set of measurements. The mean, arithmetic mean, and average ( X ) are synonymous terms for the numerical value obtained by dividing the sum of a set of replicate measurements by the number of individual results in the set: X BB/CH408 X1 X 2 X n n X i n 1 The mean of n results is n times as reliable as any one of the individual results. Thus there is a diminishing return for accumulating more and more replicate measurements. The mean of 4 results is twice as reliable as one result; the mean of 9 results is 3 times as reliable; the mean of 25 results is 5 times as reliable, etc. The median (Md) of an odd number of results is simply the middle value when the results are listed in increasing or decreasing order; for an even number of results, the median is the average of the two middle results. In general, the mean is a better measure of the central value, but in certain cases, such as with a wide spread of a small set of results, the median may give better results. Example 1: Calculate the mean and median for 10.06, 10.20, 10.08, and 10.10 Accuracy and Precision: Whereas accuracy refers to how close a measured result is to the true value, precision refers to agreement (repeatibility) amoung a group of experimental results; but says nothing about how close the results are to the true value. A group of measurements may be very precise (agree closely) but may still be inaccurate. Ideally, all measurements should be both precise and accurate. precise but not accurate accurate and precise not accurate or precise Accuracy is the nearness of a result ( Xi ) or the mean ( X ) of a set of results to the true value (). Accuracy is usually described in terms of absolute error, E, which is the difference between the measured and accepted value: E = ( Xi - ) or ( X - ) Example 2: Calculate the mean % Cl- and the absolute error of the mean for a sample, given that the true value is 24.36% and three measured results were: 24.39, 24.19, and 24.36%. BB/CH408 2 Often a more useful quantity than the absolute error is the relative error, which is expressed as a percentage (pph) or as parts per thousand (ppt) of the accepted value: pph ( X i ) 100 and ppt ( X i ) 1000 Calculate the relative error in pph and ppt for the mean % Cl- determined in Example 2. Calculate the relative error in pph and ppt and the absolute error for Example 1, given that the true value is 10.18 . Just as accuracy is reported in terms of its inverse, error, likewise, precision is reported in terms of its inverse, deviation (meaning scatter, spread, dispersion, variation, etc. of data). Both absolute deviation and relative deviation are used. Measures of absolute deviation include range and standard deviation. Relative deviation is reported as relative standard deviation. Range, (R), in a set of data is the absolute value of the numerical difference between the highest and lowest result. In Example 1, the range is: 10.20 - 10.06 = 0.14 Calculate the range for the data in Example 2: Standard deviation () is a more valid and useful measure of precision (absolute deviation). The standard deviation of an infinitely large set of experimental data is the square root of the average of the square of the individual deviations from the mean (): BB/CH408 X i N 2 , where N = # of results and approaches infinity. 3 This equation is valid for N > 30 samples, but has limited use for everyday analytical analyses where typically only 2 to 10 replicates are analysed. When the number of data is small, the standard deviation of the samples (s) rather than the standard deviation of an infinite population () is calculated using the following formula: s X i X 2 N 1 where the true population average () is replaced by the sample mean ( X ) and the denominator changes to N-1 rather than N. Example 3: A student calibrated a 10mL pipet by repeatedly weighing the amount of water delivered by the pipet and converting this to mL using a density at 20 ºC of 0.99820 g/mL. The following data were obtained: 9.9720, 9.9670, 9.9550, 9.9620, and 9.9640 grams. Calculate for the volume in mL: a) the median b) the mean c) the absolute error of the mean d) the relative error of the mean e) the range f) the standard deviation (use the stats mode of your calculator) The relative precision, called the relative standard deviation (rsd) or the coefficient of variation (cv) is calculated as the as follows: cv = s 100 X Calculate the cv for Example 3. BB/CH408 4 Example 4: The normality of a solution is determined by four separate titrations, the results being 0.2041, 0.2049, 0.2030, and 0.2043 . Assuming that the true normality is 0.2047, calculate the median, mean, absolute error of the mean, relative error of the mean in pph and ppt, range, standard deviation, and cv. SOURCES, TYPES, EFFECTS OF, DETECTION AND REDUCTION OF ERRORS AND DEVIATIONS: SOURCES 1. instruments 2. methods 3. operations DETERMINATE ERROR -constant or proportional -unidirectional (biased) -assignable -measurable, systematic REDUCED ACCURACY INCREASED ERROR INDETERMINATE ERROR -random -unbiased -not assignable -not measurable REDUCED PRECISION INCREASED VARIATION Types of Error: BB/CH408 5 Both errors and deviations, collectively called uncertainties, are classified into two broad categories, determinate errors and indeterminate errors. Determinate errors are those that, as the name implies, are determinable and that usually can be avoided or corrected. A determinate error is ordinarily unidirectional or biased, that is, it will cause all replicate analyses to be either high or low. They may be constant , as in the case of an out of calibration balance, or proportional, as in the case of and impure reagent-the more reagent used, the greater the amount of impurity present. The significance of a constant error generally decreases as the size of the sample increases. Compare the relative error (in ppt) of a constant balance error of -0.0010g on a 0.5000g sample versus a 15.0000g sample. Compare the relative error (in ppt) of a proportional error of +0.20% moisture in a supposedly dry reagent when 0.5000g sample and 15.0000g sample are weighed out. Indeterminate errors: If a measurement is sufficiently coarse, repetition will yield exactly the same result each time. For example, in weighing a 50g object to the nearest gram on a top loading electronic two-place balance, only by extreme negligence could a person obtain different values for a set of replicate weighings. On the other hand , most measurements can be refined to the point where it is mere coincidence that replicates agree to the last recorded digit. Eventually, the point is approached where unpredictable and imperceptible factors introduce what appear to be random fluctuations in the measured quantity. These are called indeterminate errors. Instruments that may be at or beyond their performance limits fluctuate due to noise and drift in an electronic circuit, temperature variations, and vibrations. Often, indeterminate errors arise from the inablility of the eye to detect slight differences in reading a buret, filling a pipet, or reading the dial of an analog instrument. Indeterminate errors are revealed by small differences in successive measurements made by the same analyst under virtually identical conditions, and they cannot be predicted or estimated. These accidental errors follow a random or normal distribution; therefore, mathematical laws of probability can be applied. The relative frequency and magnitude of indeterminate errors are described by the normal distribution or Gaussian curve or bell curve which is shown. It is apparent that BB/CH408 6 there should be very few large errors, many small errors, and, since the curve is symmetrical, there should be an equal number of positive and negative errors. f r e q u e n c y 68.3% of results 95.5% 99.74% -3 -2 - + +2 +3 quantity measured, X Sources of Errors: Both determinate errors and indeterminate errors (deviations) have three sources; instruments, methods, and operations. We will discuss determinate errors first 1. Sources of determinate instrument errors: All measuring devices are potential sources of determinate errors. For example, pipets, burets, and volumetric flasks frequently deliver or contain volumes slightly different from those indicated by their graduations. These differences arise from such sources as their use at temperatures that differ significantly from the calibration temperature, distortions in the container walls due to heating while drying, errors in the original calibration, or contaminants on the inner surfaces of the containers. Many determinate errors of this type are readily eliminated by calibration. Measuring devices powered by electricity are commonly subject to determinate errors. Examples include decreases in the voltage of battery operated power supplies with use, increased resistance in cricuits because of dirty electrical contacts, vibration, and currents induced from 110-V power lines. Again, these errors are usually detectable and correctabe; most are unidirectional. BB/CH408 7 2. Sources of determinate method errors: These errors are often introduced from nonideal chemical or physical behavior of the reagents and reactions upon which an analysis is based. Sources include slowness or incompleteness of some reactions, instability of some species, the nonspecificity of most reagents, and occurrence of side reactions which interfere with the desired reaction. For example, coprecipitation yields positive errors in gravimetry as does insufficient washing of a precipitate, while overwashing a precipitate yields negative errors. Visual indicators which change colour before or after equivalence point will cause determinate method errors. In many cases this can be corrected using a blank reagent. Determinate method errors are the most serious errors for the analyst. They are the most likely to remain undetected and require changes in the procedure to be corrected. 3. Sources of determinate operational errors: These include personal errors and can be reduced by experience and care of the analyst in the physical manipulations involved. Operations in which these errors may occur include transfer of solutions, effervescence and “bumping” during sample dissolution, incomplete drying of samples, parallax error in reading a buret, tilting a pipet while it drains, colour blindness or poor colour judgement in finding a titration end point. Other personal errors include gross mistakes in calculations, transposing numbers while recording data, reading a scale incorrectly, reversing a sign, and prejudice in estimating measurements. Effects of determinate errors: Determinate errors are often constant or proportional. In either case, they are unidirectional and bias the results, creating a negative or positive error and thus reducing the accuracy. Occasionally, determinate errors may be variable, as in the case of a buret which is out of calibration by different amounts throughout its length. In such infrequent cases, precision decreases and deviation increases. Correction of determinate errors: Instrumental errors are usually found and corrected by calibration procedures. Most personal errors can be minimized by care and self-discipline. Good chemists develop the habit of always rechecking instrument readings, notebook entries, and calculations. Method errors are particularly hard to detect. Identification and compensation for systematic errors of this type require one or more of the following : a) Analysis of standard samples either purchased or prepared in house b) Independent analysis by outside accredited laboratories BB/CH408 8 c) Blank determinations are run, in which all steps are performed with no sample present and the result is subtracted from each real analysis. d) Variation in sample size can detect constant errors as already discussed. Sources of indeterminate error: Indeterminate errors are generally chance or random. Examples include visual judgement in reading the etch mark of a pipet, the mercury level in a thermometer, and the position of an indicator on a scale. Other sources include variable pipet drainage time, variable temperatures of solutions and glassware, variable hot plate temperature, etc. All of these assume that no personal error is involved, but rather just the natural limitations of the chemist and his/her equipment, and methods. Effects of indeterminate error: Indeterminate errors cause random fluctuation or scatter in results which is another name for decreased precision and increased deviation. Increased scatter of results will usually also decrease accuracy as well. Correction of indeterminate errors: Indeterminate errors cannot be eliminated. Careful work and instrument servicing can only reduce the magnitude of indeterminate error. Replicate analyses followed by statistical analysis of data allows indeterminate error to estimated and reported so as to produce realistic results. CONFIDENCE INTERVAL AND CONFIDENCE LIMITS Frequently a chemist must make use of an new method to analyse a sample. Limitations of time prohibit numerous replicate analyses to determine a good value of the population mean () and the population standard deviation (). As indicated earlier, the sample mean ( X ) and the sample standard deviation (s) can and should be calculated in these cases however these parameters are, at best, only estimates. The use of confidence limits is probably the most meaningful way to report a result when uncertainty due to few replicates is present. In 1908, an English chemist, W. S. Gosset, writing under the pen name of “Student” (to avoid chastisement by his boss) developed a sound statistical method for estimating the true mean with only X and s based on limited data. BB/CH408 9 The numerical range (confidence interval) within which the true mean will be found is given by the following formula: Confidence limits of X t s n The quantity t (called Student’s t) is a statistical factor (obtained from a table) which is dependent upon the the number of degrees of freedom (the number of individual results less one, i.e., n-1 ) and the confidence level or probability level in %. Values for t at various probability levels in % # of # of degrees factor for confidence level observations (n) of freedom (n-1) 80% 90% 95% 99% 2 1 3.08 6.31 12.7 63.7 3 2 1.89 2.92 4.30 9.92 4 3 1.64 2.35 3.18 5.84 5 4 1.53 2.13 2.78 4.60 6 5 1.48 2.02 2.57 4.03 7 6 1.44 1.94 2.45 3.71 8 7 1.42 1.90 2.36 3.50 9 8 1.40 1.86 2.31 3.36 10 9 1.38 1.83 2.26 3.25 11 10 1.37 1.81 2.23 3.17 21 20 1.73 2.09 2.85 1.65 1.96 2.58 1.29 Example 5: A soda ash sample is analysed in the lab by titration with standard HCl. The analysis is performed in triplicate with the following results: 93.50, 93.58, and 93.43% Na2CO3. Within what range are you a) 90%, b) 95%, and c) 99% confident that the true value lies? Note that in order to obtain higher condidence, much larger confidence limits are obtained. BB/CH408 10 Example 6: A chemist determined the percentage of iron in an ore, obtaining the following results: X = 15.30, s = 0.10, n = 4 Calculate the 90% and 99% confidence interval of the mean. Example 7: Suppose 10 results have a mean of 56.06%, and the cv = 0.375. Calculate the 90% confidence limits for the true mean. s decrease, with the n result that the confidence interval is narrowed. So the more measurements you make, the more the range is narrowed within which the true value will probably lie. A 90% confidence limit means that the stated limits of X will contain the true value of the mean every 9 times out of 10. Note that as the number of measurements increases, both t and REJECTION OF A RESULT When a set of data contains an outlying result that appears to differ excessively from the mean, the decision must be made to accept it or reject it. If a valid outlier is rejected, the mean becomes biased; however by retaining a spurious result, the mean is biased in the other direction. In either case, accuracy is lost. Unfortunately, there is no sound statistical rule that always shows whether the outlier was caused by an error or by chance variation. The only reliable basis for rejecting a result is knowing that an error occurred during analysis, in which case that result is always rejected. The Q-test or “Rejection Quotient” is an easy method of testing for flyers in small data sets. The results are arranged in decreasing or increasing order. The difference between the suspect value and its nearest neighbour is divided by the range of all the results, giving a fraction. This is compared to tabulated values of Q for a 90% confidence limit. If it is equal to or greater than the tabulated value, the suspect value is rejected. range d BB/CH408 11 Q d range Example 8: # of results (n) Q 3 0.94 4 0.76 5 0.64 6 0.56 7 0.51 8 0.47 9 0.44 10 0.41 Use the Q test to test the seven results in this sample: 5.12, 6.82, 6.32, 6.22, 6.02, 6.32, and 6.12 . Arrange the results in increasing order and test both the highest and lowest results. The lowest value is tested first. If it is not rejected, then the largest value is tested. If the smallest value is rejected, the range is recalculated without the rejected value, and the largest value is then tested; and so forth. For the common situation or 3 results, the calculated Q (quotient) is compared with 0.94. To reject a questionable result, the value of 0.94 makes it necessary that the questionable value deviate quite widely from two values which agree quite closely. For example, intuition would reject 6.00% from a set of results such as 5.00%, 5.07%, and 6.00%. Yet the Q test would not permit it to be rejected. In cases like this it is recommended that 1 or 2 more results be obtained and the Q test be applied again. If the questionable result still cannot be rejected, then the median is reported rather than the mean, since the median is less influenced by an outlier. For 4 results, this case involves averaging the two middle results (the “interior average”). Example 9: Four results obtained for the normality of a solution are 0.1014, 0.1012, 0.1026, and 0.1015 . Calculate the best central value after applying the Q test. Despite the validity of the Q-test, it must be applied with good judgement as well. For example, consider three results obtained by titration 96.00, 95.01, and 95.00%. The calculated Q is 0.99, and so 96.00% would be rejected by the Q test. However, this BB/CH408 12 leaves only two results, which are virtually identical. The relative deviation of these values is given by: 95.01 95.00 1000 01 . ppt and this is unrealistically low for titrations. 95 It is clear that 95.01 and 95.00% are “lucky” results in that they are accidentally so close together. The average of these two results would not necessarily be close to the true value. In such a case it would be wise to run one or two additional analyses and reevaluate the data. Example 10: Analysis of a calcite sample yielded CaO percentages of 55.95, 56.00, 56.04, 56.08, and 56.23 . The last value appears anomalous. Calculate the best central value for the material. Example 11: A student obtained the following values for the normality of a solution: 0.0990, 0.0991, 0.0992, and 0.0998 . a) Can any result be rejected by the Q test? b) What should the normality be reported as? c) A fifth result was run and a value of 0.0991was obtained. Can 0.0998 now be discarded? Explain. d) Calculate the 90% confidence interval of the mean in c). BB/CH408 13 SIGNIFICANT FIGURES AND COMPUTATION RULES When a computation is made from experimental data, the error or uncertaintity in the final result is often calculated using computation rules for significant figures. The principal advantage of this method is that it is easy. The principal disadvantage is that only a rough estimate of uncertaintity is obtained. Despite its limitations, we will review this procedure along with a better method for determining how uncertainty of measurements affects results. Significant Figures: Most scientists define significant figures in a measurement as “all digits that are known for certain plus one estimated digit”. The following exercise will illustrate. The length of a desk top is measured with 3 different meter sticks (A, B, and C) as shown below. For each meter stick, record the length of the desk top in common notation, the number of sig figs, and the length in scientific notation. 0 A 1 0 .1 .2 .3 .4 .5 .6 .7 .8 .9 1.0 .1 .2 0 .1 .2 .3 .4 .5 .6 .7 .8 .9 1.0 .1 .2 common notation number of sig figs B C scientific notation A B C How does the sensitivity of a measuring instrument affect the number of significant figures obtainable in a measurement? BB/CH408 14 Note that by using ruler A we are able to exactly measure the object's length to the nearest meter and estimate the length to the nearest 0.1 meters (2 sig figs). Note that by using ruler B we are able to exactly measure the object's length in meters and in 0.1 meters and estimate the length to the nearest 0.01 meters (3 sig figs). Note that by using ruler C we are able to exactly measure the object's length in meters, 0.1 meters, and 0.01 meters and estimate the length to the nearest 0.001 meters (4 sig figs). For graduated devices in general (e.g., thermometers, Mohr pipets, burets, graduated cylinders, etc.) the smallest unit we can exactly measure is the smallest division on the measuring scale. We may then estimate 1 digit more. All of these (and only these) digits are significant digits (sig figs). For instruments with digital readouts, the final digit on the display is, in fact, an estimate as determined by the instrument. The uncertainty of the last digit, unless otherwise known, is taken to be 1. Likewise for data in literature, the last digit reported is assumed to have an absolute uncertainty of 1, unless otherwise stated or known. Exercise: Complete the following table of sig figs, tolerances, and absolute uncertainty for some common lab measuring devices. Calculate the relative uncertainties in ppt. Note that tolerance is the manufacturer’s guaranteed limit of uncertainty and will usually produce a constant bias (determinate error) of either high or low readings. Uncertainty in reading these devices is due to limitations of the analyst’s eye or limitations of the instrument in reading itself (in the case of electronic balances). These uncertainties produce indeterminate, random, unbiased error. device example analytical balance tolerance absolute uncertainty 100.0000g _ 0.0001 top loader balance 100.00g _ 0.01 class A volumetric pipet 10.00mL 0.02 _ 25.00mL 0.03 _ class A 50mL buret 41.00mL 0.1 0.02 × 2 class A volumetric flask 50.00mL 0.05 _ 250.00mL 0.12 _ 1000.00mL 0.30 _ 29.7mL 1 0.3 50mL graduated cylinder BB/CH408 # of sig figs relative uncertainty 15 Rules for Significant Digits in Numbers A mass of 4.56g weighed to 3 sig figs on a top loader balance can be reported as 4,560mg or 4,560,000µg, however, changing its units cannot affect its # of sig figs. Scientists must follow conventions for unambiguously reporting and interpreting numerical data. One method is to report all data in scientific notation which only list digits which are significant. When common notation is used, the following rules for sig figs are followed. Learn these rules for determining sig figs in common notation! 1. Nonzero digits: 1, 2, 3, 4, 5, 6, 7, 8, and 9 are always significant. 6.2 16.2 16.25 two significant digits three significant digits four significant digits 2. Leading zeros: zeros that appear at the start of a number, are never significant because they act only to fix the position of the decimal point in a number less than 1. ` 0.564 0.0564 three significant digits three significant digits 3. Confined zeros: that appear between nonzero numbers are always significant. 104 1004 three significant digits four significant digits 4. Trailing zeros: zeros at the end of a number are significant only if: a) the number contains a decimal point or b) the number contains an overbar 15400 1540.0 15.4000 three significant digits five significant digits six significant digits 5,600 5,600 four significant digits three significant digits Practice: Determine the number of significant digits in the following numbers: a) 345 _______ f) b) 32000 _______ g) 10700 122.0 _______ _______ c) 0.0078 _______ h) 10.04 _______ d) 9.0068 _______ i) 20 _______ e) 5.0200 _______ j) BB/CH408 0.604 _______ 16 Rounding off Numbers If the digit following the last significant digit is greater than 5, the number is rounded up to the next higher digit: 9.47 to 2 significant digits = 9.5 If the digit following the last significant digit is less than 5, the number is rounded to the present value of the last significant digit: 9.43 to 2 significant digits = 9.4 If the last digit is exactly 5, the number is rounded to the nearest even digit: 8.65 to 2 significant digits = 8.6 8.75 to 2 significant digits = 8.8 8.55 to 2 significant digits = 8.6 Practice: Round the following numbers to 2 sig figs. a) 2249 _____ d) 2250.001 ____ b) 2250 _____ e) 0.07750 ____ c) 2251 _____ f) 0.07749 ____ Uncertainty in Measurement In scientific work, we recognize two kinds of numbers: exact numbers - values known exactly - numbers which are defined or counted - defined: 12 donuts in a dozen 1000 grams in a kilogram -counted: 25 students in a classroom inexact numbers - values are uncertain - numbers which are measured eg. - speed of an automobile - temperature of a cup of soup - volume of water in a beaker Exact numbers are assigned an infinite number of significant figures for calculations Inexact numbers have a number of significant figures equal to the number of digits known for certain plus one more. Significant Figures in Calculations - a rough method for estimating error/precision of results One additional digit (one more than the number of sig figs) is usually carried on all values throughout a calculation and the answer is rounded to the correct number of sig figs. In calculations involving multiplication, division, roots, and powers: round off the answer so that it has only as many significant figures as the value in the calculation with the fewest significant digits. e.g. 35.63 0.5481 0.0530 100% 88.5470578 % 88.5% 1.1689 The answer only contains 3 significant figures since this is the least number of sig figs amoung the values in the calculation. 100 % is an exact number having an infinite number of significant digits. In addition and subtraction: the answer is only as precise as the least precise measurement. e.g. Suppose we have 3 measurements of length to be added, i.e., 6.6 m 18.74 m 0.766 m 26.106 m 26.1 m Note that the least precise measurement, 6.1 m, was only reported to the nearest 0.1m, therefore the answer cannot be reported with greater precision. Note that the correct answer has more significant digits than one of the measurements. Remember that the answer of additions or subtractions cannot have more precision than the least precise measurement. For example, if approx. 1 L of Pepsi is divided equally between 3 people, each person will not get 0.333 L, but rather 0.3L. Sig Figs in Logarithms and Antilogarithms: The pH of a 2.0 × 10-3 M HCl solution is calculated as follows: pH = -log (2.0 × 10-3 ) = -( log 2.0 + log 10-3) = -(+0.30 - 3) = 2.70 Although the logarithm appears to have one more sig fig than the original number, it only has 2 sig figs as well. The number in front of the decimal place (the “characteristic”) only indicates the power of 10 in the original number. The log of 12.1 is 1.083 and the antilogarthm of 0.072 is the number 1.18 . All the digits in the “mantissa” of a log (digits after the decimal) are significant. 18 Calculating Uncertainty of Final Results (a better method) Estimate the uncertainty of each piece of data (measurement). All uncertainties, both determinate and indeterminate, are handled mathematically as determinate errors. For calculations involving addition and subtraction the absolute uncertainty of each measurement is added together. For calculations involving multiplication and division the relative uncertainty of each measurement is added together. Procedure for addition and subtraction: Retain as many decimal places in the result as in the least precise reading (same as before). To determine the limits of uncertainty in the result, add the absolute uncertainties of each value. Examples: For each of the following, add the numbers and report the answer to the correct number of decimal places and report the absolute uncertainty of the answer. a) 50.1 ( 0.1), 1.36 ( 0.02), 0.5182 ( 0.0001), and 6.453 ( 0.003) answer 58.4 ( 0.1) b) 14.23 + 8.145 - 3.6750 + 120.4 answer 139.1 ( 0.1) c) 0.50 ( 0.02) + 4.10 ( 0.03) - 2.63 ( 0.05) answer 1.97 ( 0.10) 19 Procedure for multiplication and division: Convert the absolute uncertainty of each value to its relative uncertainty (in ppt) and add all these together to obtain the total relative uncertainty of the final result. Convert the relative uncertainty of the result to absolute uncertainty. Keep only as many decimal places in the result as the first occupied decimal place in the absolute uncertainty. Examples: For each of the following, carry out the indicated operations and report the answer to the correct number of decimal places and report the absolute uncertainty of the answer. (10.00 0.02) (5.000 0.001) 2.50 0.01 a) answer 20.0 ( 0.1) b) The percentage Cr is calculated from a titration as follows: %Cr 40.64mL 01027 . mmol / mL 51996 . mg / mmol / 3 100 346.4mg answer 20.88 ( 0.05) c) Compare these with the method of only using sig figs: (0.98 0.01) (1.07 0.01) = 1.05 0.02 (1.02 0.01) (1.03 0.01) = 1.05 0.02 Since the uncertainties in these numbers are the same, both results should have the same number of decimal places. The sig fig method would not work here nor in the next example. d) 24 0.452 100.0 e) . ( 01 . ) 0.050( 0.001) 14.3(01. ) 116 820(10) 1030(5) 42.3(01. ) answer 0.108 ( 0.005) 20