Error description:

advertisement
Error messages after turning power on
Bios Test passed ROM Test passed RAM Test passed
– Audible Signal The red status indicator is flashing.
– Cause: Dual flashing rate: EPROM error
–
Triple flashing rate: RAM error.
–
Quadruple flashing rate: NOVRAM error.
FXX Configuration (only with AutoPRIME™)
– Incorrect machine configuration
EPROM ERROR
– System error.
– Check the seat of the plugs and the EPROM. Replace the EPROM, if necessary.
BRAM_#_XXXX_XXXX_XXXX
– System error.
– Switch the machine off and on again.
– Check the seat of the plugs and the BRAM. Replace the BRAM, if necessary. Then recalibrate.
RAM ERROR
– System error.
– Switch the machine off and on again.
– Check the seat of the plugs and the RAM. Replace the RAM, if necessary.
Keyboard Error
– Short-circuit on the keyboard.
– Switch the machine off and on again.
– Check the seat of the plugs. Possible short-circuit on the keys. Replace the front panel, if necessary.
Watchdog Error
– This error message can only be displayed shortly after switch-on.
– Switch the machine off and on again.
– Check the WD relay and components.
– Check CPU2/CPU1.
– Check the plug connectors on the monitor.
XX (not calibrated)
– NOVRAM error upon test request.
– Switch the machine off and on again.
– Recalibrate the function indicated. Replace the NOVRAM, if necessary.
NTC109 switched off
– No valid value has been filed during start in the NOVRAM. The differ-ence in temperature between
– NTC 109 and NTC 103 is too large.
– Switch off NTC 109 in the setup menu, or recalibrate the temperature.
1
Error description: T1 test
F 01 Bypass
– The heater relay is switched off.
– Acknowledgement (H_REL_W, X639/A12) → X632/A10, 0 V are missing.
F 02 Bypass
– The heater relay cannot be switched off by CPU2.
– Acknowledgement (H_REL_W, X639/Y12) → X632/A10, 12 V are missing.
– Control line (EM_H_OFF, X632/A9) → X639/A17, 12 V are missing.
F 03 Bypass
– The temperature measurement range is set to hot rinse.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are missing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are missing.
F 04 Bypass
– The extended bypass cannot be correctly switched by CPU2 (V24 = off, V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are missing.
F 05 Bypass
The extended bypass cannot be correctly switched off by CPU2 (V24 = on, V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.
F06 Bypass
– CPU1 fails to set the temperature control to hot rinse.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 12 V are missing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 12 V are missing.
F 07 Bypass
– The extended bypass cannot be correctly switched by CPU1 (V24 = off, V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are missing.
F08 Bypass
– CPU1 fails to reset the temperature control to dialysis.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are missing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are missing.
F09 Bypass
– The extended bypass cannot be correctly switched off by CPU1 (V24 = on, V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.
F95 Bypass
– System error
2
F01 opt. Detector
– CPU1 interprets the optical detector in a different way than does CPU2.
– Acknowledgement (OD_OUT, X633L/C7) → X632/A30 and the digital input of P.C.B. LP 633
– measure different levels.
F02 opt. Detector
– CPU2 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) → X632/A30, 0 V are missing.
– Detuning (ODSA, X632/C15) → X351/7 not 12 V
F03 opt. Detector
– CPU1 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) → digital input on P.C.B. LP 633.
– Detuning (ODSA, X632/C15) → X351/7, not 12 V.
F04 opt. Detector
– CPU recognizes that the optical detector senses opaque fluid ( required because of the test
– in the cleaning program)
– Acknowledgement X632/A30 not 12V
– AD28 defective
F96 opt. Detector
– System error.
3
F09 Bloodsystem
– Acknowledgement that CPU2 recognizes that the arterial blood pump is inactive (BP not running).
– Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 12 V are missing.
– Control line (BPSST_ART, X634L/B14) → X348a/1, 12 V are missing or
– ( BPST_ART, X634L/A14) → X348a/3, 12 V are missing.
F10 Bloodsystem
– Acknowledgement that CPU1 recognizes that the arterial blood pump is inactive (BP not running).
– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 12 V are missing.
– Control line (BPSTT_ART, X634L/B14) → X348a/1, 12 V are missing or
– (BPST_ART, X634L/A14) → X348a/3, 12 V are missing.
F11 Bloodsystem
– The arterial blood pump cannot be stopped by CPU1. CPU2 recognizes that the arterial blood pump
– remains active.
– Control line (BPSST_A, X634L/B14) → X348a/1, 0 V are missing, as well as
– (BPST_ART, X634L/A14) → X348a/3, 0 V are missing.
– Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 0 V are missing.
– The level is raised during the T1 test, or the up/down key on the air detector is blocked and the
– Level is constantly raised.
F12 Bloodsystem
– The arterial blood pump cannot be stopped by CPU1. CPU1 recognizes that the arterial blood pump
– remains active.
– Control line (BPSST_A, X634L/B14) → X348a/1, 0 V are missing, as well as
– (BPST_ART, X634L/A14) → X348a/3, 0 V are missing.
– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 0 V are missing.
F13 Bloodsystem
– Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is
– connected – (= ADKS active)
– Acknowledgement that CPU2 detects that the pump is inactive (pump is not running).
– Acknowledgement (BPSB_VEN, X348V/6) → X632/ B11, 12V missing
– Control line (BPSST_VEN, X634L/B15) → X348V/1,12V missing or (BPST.VEN, X634L/A15)
– → X348V/3, 12V missing
– Transistor T9 on P.C.B. LP754 defective
– IC5 on P.C.B. LP632 defective
– In 4008 HDF an HDF treatment was performed, followed by a cleaning program with the
– substituate pump running, then the T1 test has been re-started.The substituate pump must
– be switched off because otherwise the test step will fail to be passed (problem was
– corrected with SW 3.20 in 4008 H/S Systems: the substituate pump will be switched off
– automatically on starting a cleaning program).
F14 Bloodsystem
– Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected
– (= ADKS active) Acknowledgement that CPU1 detects that the pump is inactive (pump is not
– running).
– Acknowledgement (BPSB_VEN,X348V/6) → X633L/A13, 12V missing
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 12V or
– (BPST_VEN, X634L/A15) → X348V/3 not 12V
– lC16 on P.C.B. LP633 defective
– P.C.B. LP633 recognizes Single-Needle pump although it is not connected.
F15 Bloodsystem
– Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected
– (= ADKS active) CPU1 fails to stop the corresponding blood pump. CPU2 detects that the
– pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0V missing as well as (BPST_VEN,
– X634L/A15) → X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11, 0V missing
– Transistor T9 on P.C.B. LP754 defective
– IC5 on P.C.B. LP632 defective
– During the test the lines are inserted on the corresponding pump using the Start/Stop key.
– P.C.B. LP633 recognizes Single-Needle pump although it is not connected.
4
F16 Bloodsystem
– Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected
– (= ADKS active) CPU1 fails to stop the corresponding blood pump. CPU1 detects that the
– pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 0V as well as (BPST_VEN,
– X634L/A15) → X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 0V
– IC16 on P.C.B. LP633 defective
– P.C.B. LP633 recognizes Single-Needle pump although it is not connected.
F17 Bloodsystem
– Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is
– connected (= ADKS active)
– Although the recognition of the venous blood pump (ADKS) is not acknowledged,
– the 24-V supply voltage of the pump can be switched off.
– Acknowledgement line (ADKS, X348V/7) → X633L/A10 not 12V
– Acknowledgement (BPSB.VEN, X348V/6) → X633L/A13 not 12V
– Acknowledgement (BPSB.VEN, X348V/6) → X632/B11 not 12V
– Online-HDF has already been switched on during the T1 test.
– lC16on P.C.B. LP633 defective.
F18 Bloodsystem
– Applicable for SW 5.00/4.10 and higher, check of the BPUS signal (CPU, P.C.B. LP 632)
– At the beginning of the test step a maximum of 40s may pass until rotation has stopped.
– lf the blood pump is being activated, the rotation stop alarm must have been cleared.
– Acknowledgement line (BPUS, X348A/8) → X632/A13 not 0V
– Acknowledgement line (BPUS, X348A/8) → X632/A13 not 12V
– Blood pump speed is set to "0": preset speed during the T1 test.
F19 Bloodsystem
– Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
– (CPU, P.C.B. LP 631 via LP 633)
– At the beginning of the test step a maximum of 40s may pass until rotation has stopped.
– If the blood pump is being activated, the rotation stop alarm must have been cleared.
– 1. Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 0V
– 2. Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 12V
F95 Bloodsystem
– System error.
5
F01 Venous
– CPU1 (input board) shows a venous zero point deviation of more than ±12 mmHg (60 s).
– Control (VENT_Valve, X634R/C18) → X351/1 of the vent valve in the LD is defective.
– Acknowledgement (P_VEN, X351/4) → X633L/B5 that the voltage value is outside the zero point
– tolerance.
– P-venous has not been calibrated.
F02 Venous
– CPU2 shows a venous zero point deviation of more than ±12 mmHg (60 s).
– Control (VENT_Valve, X634R/C18) → X351/1 of the vent valve in the LD is defective.
– Acknowledgement (P_VEN, X351/4) → X632/C17, the voltage value is outside the zero point
– tolerance.
– P-venous has not been calibrated.
F03 Venous
– With detuning in positive direction, the achieved change in the venous display is less than 100
– mmHg (7 s).
– The test detuning is defective (PV_DET, X632/C18) → X351/2.
– Acknowledgement (P_VEN, X351/4) → X633L/B5, the change in voltage is too low.
– P-venous has not been calibrated.
F04 Venous
– The deviation in the measured value between CPU1 and CPU2 is higher than ±12 mmHg (if Pven >
– 100 mmHg).
– Acknowledgement (P_VEN, X351/4) → X633L/B5 and X632/C17 measure different voltage values.
– P-venous has not been calibrated.
F95 Venous
– System error.
6
F01 Airdetector
– CPU1 interprets the air detector signal in a different way than does CPU2.
– Acknowledgements (LDA1, X351/14) → X632/C13 and X633L/C10 recognize different signal levels.
F02 Airdetector
– The air detector alarm is not recognized by CPU2.
– Acknowledgement (LDA1, X351/14) → X632/C13, 0 V are missing.
– Transmission weakening (LDSA, X632/C16) → X351/10, 12 V are missing.
F03 Airdetector
– Air detector clamps acknowledgement (CPU2) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F04 Airdetector
– Air detector clamps acknowledgement (CPU1) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F05 Airdetector
– The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are missing.
– If the HDF option is used, this signal is not tested (special function).
F06 Airdetector
– Closing of the air detector clamp via the CPU2 control line was not possible.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.
F07 Airdetector
– Opening of the air detector clamp via the CPU2 control line was not possible.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.
F08 Airdetector
– Closing of the air detector clamp via the CPU1 control line was not possible, or CPU2
– Acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.
F09 Airdetector
– Closing of the air detector clamp via the CPU1 control line was not possible, or CPU1
– Acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X633L/C13, 0 V are missing.
F10 Airdetector
– The blood alarm message is missing.
– Acknowledgement (BL_AL, X634R/C15) → X632/C21, 0 V are missing.
– If the HDF option is used, this signal is not tested (special function).
F11 Airdetector
– Air detector clamps acknowledgement (CPU2) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F12 Airdetector
– Air detector clamps acknowledgement (CPU1) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F13 Airdetector
– The blood alarm signal has not been cleared (indicates alarm).
– Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are missing.
– If the HDF option is used, this signal is not tested (special function).
F14 Airdetector
– Raise level key on the air detector is constantly active.
– Acknowledgement (LEVEL_UP, X351/3) → X632/C11 not 0V.
7
F15 Airdetector
– Acknowledgement of the supply voltage for the ultrasonic output stage not between 6.5 and
– 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7) not 12V.
– Relay on AD28 failed to drop.
F16 Airdetector
– Acknowledgement of the supply voltage for the ultrasound output stage not >14.5V arter
– 3 seconds.
– Adapter board AD28 not connected.
– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7) not 16V/24V.
– Relay on AD28 is not controlled.
– No 10-Hz Signal at ALARM.REST (X351/12)
F17 Airdetector
– Acknowledgement of the supply voltage for the ultrasound Output stage not between 6.5 and
– 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 → X633L/25A Jumper to X633L/B7) not 12V
– Relay on AD28 failed to drop
F95 Airdetector
– System error.
8
F01 Display
– CPU1 failed to start the display test within 5 sec.
– The “test started” information transmitted via the serial interface is missing.
F02 Display
– CPU1 failed to complete the display test within 120 sec.
– The “test completed” information transmitted via the serial interface is missing.
F95 Display
– System error.
9
F01 Arterial
– With detuning in negative direction, the change achieved on the arterial display is less than 100
– mmHg (2 sec).
– Acknowledgement (P_ART, X348A/7) → X633L/B12, insufficient voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
F02 Arterial
– With detuning in positive direction, the change achieved on the arterial display is less than
– 100 mmHg (2 sec).
– Acknowledgement (P_ART, X348A/7) → X633L/B12, insufficient voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
F95 Arterial
– System error.
10
F01 Accumulator
– CPU1 failed to complete the battery test within 5 sec.
– The “test completed” information transmitted via the serial interface is missing.
F02 Accumulator
– The battery charge is insufficient for 15 min emergency operation (maybe no battery connected).
– The battery voltage (U_ACCU, ... ) → X633L/B21 dropped below 17.6 V.
– Acknowledgement (U_ACCU, ... ) → X 633L/B21 of the battery voltage is defective.
F03 Accumulator
– The test circuit on P.C.B. LP 639 is defective.
– The test level is incorrect (TESTBATT, X634R/C23) → X639/A10, the 12-V pulse
– is missing (100 ms).
F95 Accumulator
– System error.
11
F01 Bloodleak
– The blood leak detector and dimness not in the alarm-free state during the T1 test.
– Dimness channel contaminated (calcium precipitate, etc)
– Acknowledgement (BLL, X637/A18) → X633L/B10, voltage value within alarm tolerances (< 3V)
– Acknowledgement (BLL_DIM, X637A/21) → X633L/B11 voltage value within the alarm
– tolerances (<1.5V/>8V).
– DAC_BLL or DAC_DIM not within the tolerances (check calibration).
F02 Bloodleak
– A blood leak alarm / dimness alarm is not recognized during test detuning.
– Acknowledgement (BLL, X637/A18) → X633L/B10, voltage value not within alarm tolerances.
– Acknowledgement (BLL_DIM, X637A/21) → X633LB11 voltage value not within
– alarm tolerances (1.5V)
– Test detuning (BLL_DET, X632/A25) → X633L/B27, not 5 V
– Calibration of DAC_BLL or DAC_DIM is too high.
– Detuning (DAC_DIM, X634R/A11) → X633L/C3 impossible.
– Dimness calibration is set to potenitometer calibration (BR6 from pos. 1/2 to 2/3.
F03 Bloodleak
– After test detuning, the blood leak channel and dimness fails to enter the alarm-free state.
– Dimness channel contaminated (calcium precipitate, etc)
– Acknowledgement (BLL, X637/A18) → X633L/B10, voltage value within alarm tolerances.
– Test detuning (BLL_DET, X632/A25) → X633L/B27, not 0 V.
– Acknowledgement (BLL_DIM, X637A/21) → X633L/B11 voltage value within the alarm
– tolerances (<1.5V/>8V).
– DAC_BLL or DAC_DIM not within the tolerances (check calibration).
F95 Bloodleak
– System error.
12
F01 Temperature
– The temperature measuring range is not set to hemodialysis.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are missing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are missing.
F02 Temperature
– The actual temperature is less than 35.0 ºC (test running time > 15 minutes).
– Calibrate the temperature.
– The heater rod failed.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, voltage got stuck.
F03 Temperature
– The actual temperature is higher than 39.0 ºC (test running time > 15 minutes).
– Calibrate the temperature.
– The regulating sensor (NTC-2) is defective.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, voltage got stuck.
F04 Temperature
– The temperature failed to stabilize within 15 minutes.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24 is steadily changing
– (change > 0.3 ºC/15 sec).
F05 Temperature
– Detuning in positive direction not higher than 3 ºC (10 sec).
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, change in voltage insufficient.
– Detuning (T_DETADJ, X632/A23) → X633R/C21 insufficient.
F06 Temperature
– The monitor sensor indicates a constant value.
– NTC-3 defective.
F07 Temperature
– The test release is missing (max. test running time is 10 minutes).
– Run-time problem (software).
F08 Temperature
– CPU1 failed to transmit a Bibag status message within 3 sec.
– Run-time problem (software).
F09 Temperature
– Bibag NTC_BIB detuning not higher than 1 ºC.
– Acknowledgement (NTC_BIB, X633R/C15) → ADW on P.C.B. LP 633,
– change in voltage insufficient.
– Detuning (BIBAG_TE, X634R/A13) → X633R/A20 insufficient.
F10 Temperature
– Bibag temperature display outside of measuring range (15 to 45 ºC).
– Acknowledgement (NTC_BIB, X633R/C15) → ADW on P.C.B. LP 633.
F95 Temperature
– System error.
13
F01 neg. Pressure
– During the start phase a negative pressure of more than 450 mmHg has developed (max. test
– running time 120 sec),
– the hydraulic system is contaminated,
– the air separation pump started running.
F02 neg. Pressure
– Setting the dialysate pressure to the test pressure (–300 mmHg to –450 mmHg) was not possible
– (max. test running time 120 sec).
– Upon repetition of measurement, the range was extended from –260 mmHg to 490 mmHg.
– Leakage in the hydraulic system.
– The UF pump is defective.
– If the HDF filter test was skipped: Clamp the HDF filter.
F03 neg. Pressure
– The working point (116 digits) of the differential amplifier cannot be set correctly (max. test running
– time 120 sec).
– Pressure variations are too large.
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective.
– The acknowledgement (P_DIAL, X633L/B6) → X632/A29 is defective.
– The CI signal is missing (LP 633 → X632/B22).
F04 neg. Pressure
– Completion of pressure measurement was not possible (max. test running time 120 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective.
– The acknowledgement (P_DIAL, X633L/B6) → X632/A29 is defetive.
F05 neg. Pressure
– The air separation pump started running during the measurement phase.
– Acknowledgement (ACKN_ASP, X634L/B10) → X632/A19, 0 V are missing.
– ASP has been interrupted electrically.
F06 neg. Pressure
– The negative pressure holding test failed to be passed. The dialysate pressure loss exceeds ± 40
– mmHg (related to ten balancing chamber changeovers).
– Leakage in the hydraulic system.
F07 neg. Pressure
– Current increasing pulses were not recognized (min. 2x).
– 5-V balancing chamber pulses are missing (CI. X634R/A23) → X632/B22.
F95 neg. Pressure
– System error.
– In machines with HDF option, the negative pressure holding test is performed internally only; i.e.
– V24, V24B are closed and V26 is open.
14
F01 pos. Pressure
– The CPU1 mandatory priming program was not completed (10 sec).
– The solenoid valve V43 is not closed.
F02 pos. Pressure
– The loading pressure cannot be measured via the solenoid valve V26 in the hydraulic system
– (P-Dial. < 600 mmHg, 15 sec).
– Solenoid valve V26 mechanically not open.
– Solenoid valve V43 mechanically not closed.
– The balancing chamber is switched to passage during this test sequence. V24, V24B and V43 are
– closed; V26 is open.
F03 pos. Pressure
– The hydraulic system cannot be deaerated via the solenoid valve V43; the zero point of –125 to 55
– mmHg has not been reached (15 sec).
– Solenoid valve V26 mechanically not closed.
– Solenoid valve V43 mechanically not open.
– Zero point outside the –125 to 55 mmHg range.
– The balancing chamber is switched to passage during this test sequence. V24, V24B and V26 are
– closed; V43 is open.
F04 pos. Pressure
– The first working point (220 digits) of the differential amplifier cannot be set.
– Pressure variations are too large.
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective.
– The acknowledgement (P_DIAL, X633L/B6) → X632/A29 is defective.
F05 pos. Pressure
– Test detuning results in a change in the measuring range of more than 95 mmHg (60 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– The balancing chamber valve V36 or V38 (old water valve) is leaky..
F06 pos. Pressure
– Test detuning results in a change in the measuring range of less than 85 mmHg (60 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in voltage insufficient.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– V26 is leaky.
F07 pos. Pressure
– After detuning in the test there is a difference (P.diff > ± 9 mmHg) between the display and the
– differential amplifier.
– The voltage divider R23/R9 or the operational amplifier IC2 is defective.
– The operational amplifier IC1/IC3 is defective.
– The balancing chamber valve V36 or V38 (old water valve) is leaky.
F08 pos. Pressure
– Test detuning results in a change in the measuring range of more than 400 mmHg (20 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
F09 pos. Pressure
– Test detuning results in a change in the measuring range of less than 350 mmHg (20 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in voltage insufficient.
– Detuning defective (DIAL_DET_ADJ, X632/C20) → X633R/C22.
F10 pos. Pressure
– The second working point (116 digits) of the operational amplifier cannot be set correctly.
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective.
15
F11 pos. Pressure
– Change in the dialysate pressure after closing of the solenoid valve V43 (zero point change of ± 20
– mmHg within 15 sec)
– The solenoid valve V24B is not closed.
– The balancing chamber valve V36 or V38 (old water valve) is leaky.
– The balancing chamber is switched to passage during this test sequence. V43, V24B and V26 are
– closed; V24 is open.
F12 pos. Pressure
– The loading pressure cannot be measured via the solenoid valves V24 and V24B in the hydraulic
– system (P-Dial. < 600 mmHg, 15 sec).
– Solenoid valve V24 or V24B mechanically not open.
– The balancing chamber is switched to passage during this test sequence. V43 and V26 are closed;
– V24 and V24B are open.
F13 pos. Pressure
– The hydraulic system cannot be deaerated via the solenoid valve V43 (P-Dial. not equal to –125 to
– 55 mmHg, 20 sec).
– The solenoid valve V24 is not closed.
– V43 neither opens electrically nor mechanically.
– The balancing chamber is switched to passage during this test sequence. V24 and V26 are closed;
– V24B and V43 are open..
F14 pos. Pressure
– Zero point change after closing of solenoid valve V43 (20 sec). Standard: P-Dial. not equal to –125
– to 55 mmHg. HDF option: P-Dial. not equal to –125 to 60 mmHg.
– The solenoid valve V24 is not closed.
– The balancing chamber is switched to passage during this test sequence. V24, V26 and V43 are
– closed; V24B is open.
F15 pos. Pressure
– The loading pressure is below 780 mmHg ± 30 mmHg (10 sec).
– The loading pressure is too low.
F16 pos. Pressure
– During the start phase, the pressure dropped below 620 mmHg (meas-uring tolerance: ± 30 mmHg,
– max. test running time 120 sec).
– Large-size leakage in the hydraulic system.
– The UF pump spring is defective.
– The loading pressure is too low.
– The air spiration pump fails to occlude.
– Relief valve is leaking.
F17 pos. Pressure
– During the start phase, it was not possible to reduce the dialysate pressure to a value below 760
– mmHg (measuring tolerance: ± 30 mmHg, test running time 120 sec).
– The loading pressure is too high.
– The UF pump is defective.
F18 pos. Pressure
– The working point (116 digits) of the differential amplifier cannot be set correctly (test running time
– 120 sec).
– The pressure variations in the system are too large.
F19 pos. Pressure
– Completion of the pressure measurement was not possible (max. test running time 120 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The acknowledgement (P_DIAL, X633L/B6) → X632/A29 is defective.
16
F20 pos. Pressure
– The positive pressure holding test failed to be passed. While the flow was off, a pressure loss of
– more than ± 80 mmHg/min was detected in the hydraulic system.
– Leakage in the hydraulic system.
– The UF pump spring is defective.
– ASP fails to occlude.
– Relief valve leaking.
– V84 leaking.
F21 pos. Pressure
– The dialysate pressure cannot be set to a value between 460 and 760 mmHg ± 30 mmHg (10 sec).
– The heat exchanger is defective.
– Problem in the hydraulic system.
F22 pos. Pressure
– The air purging pump is not running during the test phase (2 sec).
– Control line (AIR_SEP+, X634L/A22) → ASP/..., 24 V are missing.
– Control line (AIR_SEP–, X634L/C22) → ASP/..., 0 V are missing.
– Acknowledgement (ACKN_ASP, X634L/B10) → X632/A19, 12 V are missing..
F23 pos. Pressure
– Pressure drop in the hydraulic system during the measurement phase (8 sec). Change more than +4
– digits or more than –8 digits.
– Leakage in the pump segment of the air purging pump.
– Leakage in the heat exchanger.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in voltage too large.
F24 pos. Pressure
– V24 valve error.
– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.
F25 pos. Pressure
– No pressure increase above 150 mmHg (change in pressure) after valve switching.
– Control signals of V24 and V24B mistaken for each other.
– Leakage in the external system (shunt interlock, dialysate lines, etc.).
– If the HDF filter test was skipped: Clamp the HDF filter.
F26 pos. Pressure
– No pressure compensation after opening of V43 (–125 mmHg to 55 mmHg).
– V24 got stuck (mechanically open).
– V43 not open.
– V26 leaking.
F27 pos. Pressure
– No pressure compensation after opening of V43 (-125 mmHg to 55 mmHg).
– V24 got stuck (mechanically open).
– V43 not open.
– V189 (retentate valve) leaking.
F28 pos. Pressure
– ASP functional test (running and delivery test)
– APS line segment is occluded
– APS line segment has been incorrectly inserted ( check direction of delivery)
– APS is not running ( electrical or mechanically)
– V87 electrically or mechanically closed.
F95 pos. Pressure
– System error.
17
F01 UF-Function
– The pause between the strokes of the UF pump 1 was shorter than 220 ms. Correct volume delivery
– is not ensured due to too short a return.
– CPU1 issued too high a pump rate.
F02 UF-Function
– The pulse time for the UF pump 1 is shorter than 180 ms. Correct volume delivery is not ensured
– due to too short an emission time.
– The monoflop on P.C.B. LP 634 is defective (IC42/R82/C47).
F03 UF-Function
– The pulse time for the UF pump 1 is longer than 500 ms. A maximum rate of 5000 ml/h is not
– possible.
– The monoflop on P.C.B. LP 634 is defective (IC42/R82/C47).
F04 UF-Function
– No activity of the UF pump 1 during the test (5 sec).
– Acknowledgement (UF_P1, X637/B23) → X632/A7, no LOW pulses.
– Control line (UF_P1, X634L/ABC23) → X637/B23, no LOW pulses.
F05 UF-Function
– The UF pump 1 cannot be stopped by CPU2.
– Control line (UF_P_EN, X632/C28) → X634R/A22, 5 V are missing.
– The reset input at IC42/pin 3 on P.C.B. LP 634 is defective.
F06 UF-Function
– The UF pump acknowledgement of CPU1 is defective.
– Acknowledgement (UF_P1, X637/B23) → X622L/C14, no LOW pulses.
F07 UF-Function
– The change in pressure after a stroke is less than 20 mmHg.
– The UF pump 1 is mechanically defective.
– Control line (UF_P1_CTL, X632/C27) → X634R/A24, no LOW pulse.
F09 UF-Function
– Dialysate pressure is outside the measuring range (15s).
– UF pressure tansducer defective
– D/A converter (IC 11) on P.C.B. LP 632 defective
– Operational amplifier (IC1/IC3) on P.C.B. LP 632 defective
F11 UF-Function
– The pause between the strokes of the UF pump 2 was shorter than 220 ms. Correct volume delivery
– is not ensured due to too short a return.
– CPU1 issued too high a pump rate.
F12 UF-Function
– The pulse time for the UF pump 2 is shorter than 180 ms. Correct volume delivery is not ensured
– due to too short an emission time.
– The monoflop on P.C.B. LP 634 is defective (IC42/R65/C45).
F13 UF-Function
– The pulse time for the UF pump 2 is longer than 500 ms. A maximum rate of 5000 ml/h is not
– possible.
– The monoflop on P.C.B. LP 634 is defective (IC42/R65/C45).
F14 UF-Function
– No activity of the UF pump 2 during the test (4 sec).
– Acknowledgement (UF_P2, X637/B26) → X632/C7, no LOW pulses.
– Control line (UF_P2, X634L/ABC24) → X637/B26, no LOW pulses
F15 UF-Function
– The UF pump 2 cannot be stopped by CPU2.
– Control line (UF_P_EN, X632/C28) → X634R/A22, 5 V are missing.
– The reset input at IC42/pin 13 on P.C.B. LP 634 is defective.
F16 UF-Function
– The UF pump acknowledgement of CPU1 is defective.
– Acknowledgement (UF_P2, X637/B26) → X633L/C23, no LOW pulses.
F17 UF-Function
– The change in pressure after a stroke of the UF pump 2 is less than 20 mmHg.
– The UF pump 2 is mechanically defective.
– Control line (UF_P2_CTL, X632/B24) → X634R/C11, no HIGH pulse.
18
F20 UF-Function
– The difference in volume between UF pump 1 and UF pump 2 is higher than 25% (range of
– tolerance 15% to 35%)
– The stroke volume of UF pump 1 or UF pump 2 has been misadjusted.
F95 UF-Function
– System error.
19
F01 Conductivity
– The conductivity failed to be within the scale limits or to be stable within 10 minutes
–
– Concentrate is not connected.
– Acknowledgement (COND_SIG, X633L/B8) → X632/A22, voltage outside the measuring range or
– unstable.
F02 Conductivity
– Detuning in positive direction not more than 0.5 mS (10 sec).
– Acknowledgement (COND_SIG, X633L/B8) → X632/A22 insufficient.
– Detuning (COND_DET, X632/A21) → X633L/B31 insufficient.
F03 Conductivity
– Detuning in negative direction not more than 0.5 mS (10 sec).
– Acknowledgement (COND_SIG, X633L/B8) → X632/A22 insufficient.
– Detuning (COND_DET, X632/A21) → X633L/B31 insufficient.
F04 Conductivity
– The conductivity cell indicates a constant value.
– The CD cell is defective.
F05 Conductivity
– CPU1 failed to transmit a Bibag status message within 3 sec.
– Run-time problem (software).
F06 Conductivity
– The Bibag CD detuning is not more than 1 mS/cm.
– Acknowledgement (COND_SIGNAL3, X633R/A12) → MP TP3 on P.C.B. LP 633, change in voltage
– insufficient.
– Detuning (COND_DET, X632/A21) → X633L/B31 insufficient.
F07 Conductivity
– The Bibag CD display is outside of the measuring range (46 to 84 mS/cm).
– Acknowledgement (COND_SIGNAL3, X633R/A12) → MP TP3 on P.C.B. LP 633.
– Conductivity outside the expected detuning range caused by wrong concentrate on the bicarbonate
– port or temperature too low.
F08 Conductivity
– CPU 1 fails to increase the working point (when the conductivity is <40mS/cm uncompensated)
– for the bibag conductivity by >5 digits
– Detuning (HOT_RINSE, X634R/C24 → X633R/A16) not 12V
– P.C.B LP 633 T2 or IC26 Defective
F95 Conductivity
– System error.
20
F02 Diasafe
– The balancing chamber was not stopped by CPU1 (24 sec).
– The message via the serial interface from CPU1 to CPU2 is missing.
– The current rise pulse is missing (CI, X634R/A23) → X633L/C31, no 5-V pulse.
F04 Diasafe
– CPU1 failed to complete a balancing chamber changeover within 20 sec (30 ml fluid not
– removed?).
– The message via the serial interface from CPU1 to CPU2 is missing.
– The current rise pulse is missing (CI, 634R/A23) → X633L/C31, no 5-V pulse.
F06 Diasafe
– During the pressure built-up phase, a vacuum pressure of less than –450 mmHg has developed
– (24 sec).
– Diasafe valve not open, control line (V_DSAFE, X632/B5) → X637/C16, 0 V are missing.
F07 Diasafe
– After the maximum fluid volume of 145 ml + 30 ml has been removed, the expected negative
– pressure of –300 mmHg to –450 mm Hg failed to build up.
– Large-size leakage in the Diasafe filter membrane and/or filter housing.
– Large-size leakage in the O-rings on filter holder/couplings.
– V26 electrically or mechanically not closed.
F08 Diasafe
– The test negative pressure of more than –300 mmHg has developed before the minimum fluid
– removal of 145 ml –30 ml has been achieved.
– The Diasafe filter is contaminated.
– The Diasafe filter was not correctly deaerated upon start of the test.
– V112 electrically or mechanically not open.
F09 Diasafe
– The zero point for pressure measurement cannot be set. The maximum test time has been
– exceeded (max. test time 5 min).
– Leakage in the Diasafe filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder/couplings.
– P.C.B. LP 632, IC3/pin 12 not fit in base or IC defective (differential amplifier).
F10 Diasafe
– The vacuum pressure to be achieved in the test failed to stabilize within the maximum test time of 5
– minutes (change > ± 16.7 mmHg/min).
– Leakage in the Diasafe filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder/couplings.
– Leakage in the hydraulic system.
– V 26 electrically or mechanically not closed.
F20 Diasafe
– It was not possible to prime (deaerate) the dialysate filter within 2 minutes.
– Flow problems.
– The priming program is permanently active (level sensor, osmosis water, or P.C.B. LP 633, IC36
– defective).
F95 Diasafe
– System error..
21
F01 HDF-Filter
– The Diasafe option has not been set although ON-LINE HDF has been selected.
– The DIP switch 2.1 on the CPU2 board is not set to ON.
F02 HDF-Filter
– CPU1 failed to stop the balancing chamber (24 sec).
– The message via the serial interface from CPU1 to CPU2 is missing.
– The current increasing pulse is missing (CI, X634/A23) → X633L/C31, no 5-V pulse.
F04 HDF-Filter
– CPU1 failed to comlete a balancing chamber changeover within 20 sec (30 ml fluid not removed?).
– The message via the serial interface from CPU1 to CPU2 is missing.
– Verify the current increasing pulse.
F06 HDF-Filter
– During the pressure-buildup phase, a negative pressure of less than –370 mmHg has developed
– (24 sec).
– The HDF filter is clamped/clogged.
– The Diasafe valve is not open, control line (V_DSAFE, X632/B5) → X637/C16, 0 V are missing.
F07 HDF-Filter
– After the maximum fluid volume of 255 ml + 60 ml has been removed, the expected negative
– pressure of –220 mmHg up to 370 mmHg failed to build up.
– Major leakage in the Diasafe/HDF filter membrane and/or filter housing.
– Major leakage in the O-rings on filter holder/couplings.
– V26 electrically or mechanically not closed.
F08 HDF-Filter
– The test negative pressure of less than –220 mmHg has developed, before the minimum fluid
– removal of 255 ml –60 ml has been achieved.
– The Diasafe/HDF filters are contaminated.
– The Diasafe/HDF filters were not correctly deaerated upon start of the test.
– V112 electrically or mechanically not open.
F09 HDF-Filter
– The zero point for pressure measurement cannot be set. The max. test time has been exceeded (10
– min).
– Leakage in the Diasafe/HDF filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder/couplings.
F10 HDF-Filter
– The negative pressure to be achieved in the test failed to stabilize within the maximum test time of
– 10minutes (change > ± 13.3 mmHg/min).
– Leakage in the Diasafe/HDF filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder/couplings.
– Leakage in the hydraulic system.
– V26 electrically or mechanically not closed..
F20 HDF-Filter
– It was not possible to prime (deaerate) the Diasafe filter within 2 minutes.
– Flow problems.
– The priming program is permanently active (level sensor, osmosis water, or P.C.B. LP 633, IC36
– defective).
F21 HDF-Filter
– It was not possible to correctly rinse/prime the HDF filter within 5 minutes (before the test).
– Flow problems.
– No conductivity.
– Conductivity at the upper or lower end of the scale range.
– The HDF pump is not running (e.g. because of open door).
– The delivery rate of the HDF pump is less than 380 ml/min.
– Line diameter not set to 8 mm.
– NTC6 permanently fails to detect fluid.
– Sieve on V43 clogged
22
F22 HDF-Filter
– It was not possible to correctly rinse/prime the HDF filter within 5 minutes (after the test).
– Flow problems.
– No conductivity.
– Conductivity at the upper or lower end of the scale range.
– Especially with biBag machines: check filter on V43.
– The HDF pump is not running (e.g. because of open door).
– The delivery rate of the HDF pump is less than 380 ml/min.
– Line diameter not set to 8 mm.
– NTC6 permanently fails to detect fluid.
– Sieve on V43 clogged
F95 HDF-Filter
– System error..
23
F01 ONLINE plus
F01 DIASAFE plus
F01 HPU
– Present options and DIP switch settings do not match.CPU1 machine status (MST),hydraulics
– processor (HPU)status and DIP switch /array 2 changed while test was in progress.
– ONLINE plus:
– CPU2:Array 2,DIP switch2 not OFF,DIP switch 3 not ON.
– DIASAFE plus:
– CPU2:Array 2,DIP switch 2 not OFF,DIP switch 3 not OFF.
– Machine status transmitted by CPU1 does not matching with the DIP switch settings of array 2.
– DIP switch/array 2 changed while test was in progress.
– Hydraulic processor has logged off.
F02 ONLINE plus
F02 DIASAFE plus
F02 HPU
– Dialysate outlet pressure (DA1)outside the acceptable range (10 s).
– DA1 test range:–125 mmHg ≤ P_dial ≤ 55 mmHg.
– Acknowledgement DA1 (P_DIAL, X633L/B6) → X632/A29.
– Acknowledgement DA2 (see hydraulics processor diagram).
F03 ONLINE plus
F03 DIASAFE plus
– Cross comparator of both pressure transducers (DA1/DA2) outside the acceptable tolerance (10 s).
–
– Acknowledgement DA1 (P_DIAL, X633L/B6) → X632/A29.
– Acknowledgement DA2 (see hydraulics processor diagram).
F04 ONLINE plus
F04 DIASAFE plus
F04 HPU
– Air pump running,although valve V43 is closed.
– Hydraulics processor,output stage etc.defective.
F05 ONLINE plus
– Door on Online-Sys module is open during rate test.
– Close module door.
F06 ONLINE plus
– Port 1 open during first pressure build-up phase.
– Close port 1 (Substituate port).
F07 ONLINE plus Port 2 open during first pressure build-up phase.
– Close port 2 (rinse port).
F08 ONLINE plus
Failure to reach the test pressure of PDIAL2 >795 mmHg within 12 s.
– Calibrate dialysate pressure.
– Replace DA2 (recalibration required).
– Air pump (185)or test valve (V183)defective.
– Hydraulic system or valve ONL3 (191)leaking.
– Air pump control (185)based on V43 status defective.
– Hydraulics processor defective.
F09 ONLINE plus
– The ONLINE system pump failed to comply with the first test rate of 100 ml/min ±
– ONLINE system pump control defective.
F10 ONLINE plus
– Monitoring unit (Hall sensor)for the pump rotor detects incorrect rotation of the rotor
– (setpoint rate 300 ml/min ± 25 %).
– ONLINE system pump control defective (outside the tolerance of ± 25 %).
– Hall sensor /electronics defective.
F11 ONLINE plus
–
–ONLINE system pump control defective.
24
F12 ONLINE plus
– After the ONLINE system pump was switched off in the test, the monitoring unit (Hall sensor)detects
– that the rotor failed to stop properly.
– Pump stop (output stage)defective.
– Hall sensor /electronics defective.
F13 ONLINE plus
– After the ONLINE system pump was switched off in the test,the actual rate of the module
– is > 0 ml/min.
– Pump stop (output stage)defective.
– Synchro-transmitter electronics defective.
F14 ONLINE plus
– After activation of the substituate pump,the monitoring unit (Hall sensor)for the pump rotor detects
– incorrect rotation of the rotor (setpoint rate:300 ml/min).
–
– Hall sensor /electronics defective.
F15 ONLINE plus
– After activation of the substituate pump,the system pump failed to comply with the test rate of
–
– Pump control defective.
F16 ONLINE plus
– Port 1 open during ONL valve test sequence.
– Close port 1 (substituate port).
F17 ONLINE plus
– Port 2 open during ONL valve test sequence.
– Close port 2 (rinse port).
F18 ONLINE plus
– Acknowledgement of ONL1 (V193)differs from the state the valve should have.
– Valve selection in the Online system module defective.
– Valve acknowledgement in the Online system module defective.
F19 ONLINE plus
– Acknowledgement of ONL2 (V192)differs from the state the valve should have.
– Valve selection in the Online system module defective.
– Valve acknowledgement in the Online system module defective.
F20 ONLINE plus
– Acknowledgement of ONL3 (V191)differs from the state the valve should have.
– Valve selection in the Online system module defective.
– Valve acknowledgement in the Online system module defective.
F21 ONLINE plus
– Leakage test ONL3 (V191)failed to be passed.The permitted pressure drop of ∆P < -10 mmHg has
– been exceeded or the test pressure is P ≤ 710 mmHg.
– Valve ONL3 (V191)in the Online system module leaking.
– Leaking system /tubing connections.
– Port 1 or 2 in the Online system module leaking.
F22 ONLINE plus
– Leakage test ONL2 (V192)failed to be passed.The permitted pressure drop of ∆P < -10 mmHg has
– been exceeded or the test pressure is P ≤ 710 mmHg.
– Valve ONL2 (V192)in the Online system module leaking.
– Leaking system /tubing connections.
– Port 1 in the Online system module leaking.
F23 ONLINE plus
– Leakage test ONL1 (V193)failed to be passed.The permitted pressure drop of ∆P < -10 mmHg has
– been exceeded or the test pressure is P ≤ 710 mmHg.
– Valve ONL1 (V193)in the Online system module leaking.
– Leaking system /tubing connections.
F24 ONLINE plus
– After the valves ONL 1 to 3 opened,the pressure drop in the system was insufficient
– (∆P < -100 mmHg).
– Valve ONL 1/ONL 2/ONL 3 electrically or mechanically not open.
– Kinked tubing.
– HDF filter strongly contaminated.
25
F25 ONLINE plus
– No pressure change of ∆P >40 mmHg within 15s.
– HDF filter membrane leakage (large leak).
– No HDF filter installed.
F26 ONLINE plus
F26 DIASAFE plus
– Insufficient test pressure (P < 750 mmHg) in the system.
– HDF filter membrane leakage (large leak).
– No HDF filter installed.
– Hydraulic system leakage.
F27 ONLINE plus
F27 DIASAFE plus
– After valve V189 had opened, the pressure drop in the system was insufficient (∆P < -70 mmHg).
– Valve V189 electrically or mechanically not open.
– Diasafe filter strongly contaminated.
– Filter upstream/downstream of V43 clogged.
F28 ONLINE plus
F28 DIASAFE plus
– Pressure in system fails to rise above P > 760 mmHg
– Diasafe filter membrane leakage (large leak).
– No Diasafe filter installed.
F29 ONLINE plus
F29 DIASAFE plus
– Pressure holding test failed to be passed. Pressure drop within a measurement time of 30 s
– too high (∆P > -10 mmHg).
– Diasafe and/or HDF filter membrane leakage.
F30 ONLINE plus
F30 DIASAFE plus
– During the pressure holding test valve(s)ONL1,ONL2 or ONL3 and/or V189 was/were closed
– (according to electronic acknowledgement signal).
– Valve control failed.
F31 ONLINE plus
F31 DIASAFE plus
F31 HPU
– Fill phase stopped
– Valve(s)V26 open and/or V24,V24b closed (according to electronic acknowledgement signal),or
– failure of the balancing chamber to switch 25 or 15 times within 120s.
– Valve control failed.
– Balancing chamber does not switch (e.g.only „Eigentakt “).
F32 ONLINE plus
– Valve(s)ONL1, ONL2 or ONL3 closed and/or V24 open or ports 1 or 2 open during rinse
– phase (according to electronic acknowledgement signal).
– Valve selection failed.
– User opened ports too early.
F33 ONLINE plus
– Rinse phase stopped. Valve V189 open (according to electronic acknowledgment signal)or failure
– of the balancing chamber to switch 34 times within 240 s.
– Valve selection failed.
– Failure to detect a current rise pulse.
– Conductivity not within the scale range. Possible cause: Concentrate and/or bicarbonate level
– sensor do not recognize CD, although present.
F34 ONLINE plus
F34 DIASAFE plus
– Failure to pass the pressure holding test. Maximum treatment time exceeded?
– Diasafe and HDF filter membrane leaking/used up.
26
F41 ONLINE plus
F41 DIASAFE plus
F41 HPU
– Test valve V183 is leaking.Pressure increase in system ∆P(DA2) > 30 mmHg within 4 s.
– V183 open, clogged, mechanically defective.
– Hydraulics processor, output stage etc.defective.
F42 ONLINE plus
F42 DIASAFE plus
F42 HPU
– Within 4s no pressure increase of ∆P(DA2) > 200 mmHg after the test valve V183 in the system
– has opened.
– V183 does not open, mechnical defect.
–- Air pump defective, does not run.
– Hydraulics processor, V183 and/or air pump output stage etc. defective.
F43 ONLINE plus
F43 DIASAFE plus
F43 HPU
– Failure to reach the lower pressure test range of ∆P(DA2) > 300 mmHg within 1 s after test valve
– V183 was closed.
– Hydraulics processor, output stage, etc. defective.
– ONLINE filter leaking.
F44 ONLINE plus
F44 DIASAFE plus
F44 HPU
– Upper pressure test range of ∆P(DA2) 750 mmHg exceeded within 4 s after test valve V183
– was closed.
– Hydraulics processor, output stage, etc. defective.
27
Error messages during dialysis
Voltage Failure
– The 24-V/12-V supply voltages are drifting.
– The machine enters the safe state and must be switched off/on.
– The 12-V or 24-V operating voltage is outside of the permissible range:
– 24 V: > 26 V / < 22.5 V.
– 12 V: > 13.5 to 15 V / < 10.5 V.
– Check the power supply unit.
– Power supply unit okay: Check the voltages applied to P.C.B. LP 633:
– +12 V: X633R/A, C31.
– +24 V: 24V_EM: X633L/B20.
24 V Switched Off
– The 24-V supply voltage has fallen below 5 V.
– The machine enters the safe state and must be switched off/on.
– Check the power supply unit.
– Power supply unit okay: Check the voltages applied to P.C.B. LP 633: +24V_EM: X633L/B20.
– Pull off all plug-in modules. As soon as the machine is running: replug each plug-in module
– individually with the machine switched off; determine the defective module and repair it.
– Completely loosen the hydraulic compartment connections.
– Caution: J1 must now be fitted on P.C.B. LP 630 since, without it, the machine would not be able to
– perform the watchdog test. Be absolutely sure to remove the jumper again for hemodialysis
– operation.
– With the machine running, check the short circuit in the hydraulic compartment for 24-V supply and
– the valves and pumps for short circuit.
CPU-II failed
– CPU2 fails to communicate via the serial interface.
– The machine enters the safe state and must be switched off/on.
– AD 10 (adapter board from software version 3.0) has contact problems (up to and incl. layout „C“).
– The software versions of CPU1 and CPU2 are mismatching.
– Hardware defect on CPU2.
Profile time diff.
– Deviation in time between CPU1 and CPU2.
– The error message is emitted 60 seconds after the start of the profile.
– The clock module on CPU1 (IC14) is defective; or calibrate the time in case of layout < D.
Cyclical PHT F01
– Balancing error.
– System leaky.
– Applicable to Diasafe machines: On CPU II, the DIP switch array 2, switch 1, is not set to “ON”.
Cyclical PHT F02
– Balancing error.
– System leaky.
Applicable to Diasafe machines: On CPU II, the DIP switch array 2, switch 1, is not set to ON
Cyclical PHT F03
– IC1 or IC3 on P.C.B. LP 632 is defective, or the system is leaky.
Cyclical PHT F04
– It was not possible to complete the test within a specific time interval.
28
V84 faultiness!
– Conductivity is recognized on the V84 electrodes.
– This error message is emitted for the first time at the end of the T1 test.
– The error can be acknowledged for the duration of one hemodialysis procedure by pressing the
– Dialysis Start key. It is, however, not possible to switch off the flow (Bergström-/ISO-UF operating
– mode). Should the error occur during Flow OFF, the flow is switched on automatically.
– First of all, it must be verified whether a Rinse Failure F01, F02 or F03 occurred during the previous
– disinfection procedure (see listing of cleaning program errors). Should this be the case, a disinfection
– program I to IV (not Dis V) must be completed correctly. A further possibility of elimination is to
– perform the calibration program under the NOVRAM menu item (Reset V84).
– Should this not be possible, the error memory of the machine can be read out.
– Should this neither be possible, the test described below can be performed:
– Remove the disinfectant.
– Switch the machine off and on again.
– Perform or skip the T1 test.
– Should the error message be displayed again at the end of the test, it was generated by a machine
– error F01, F02 or F03 and can be cleared only by taking the measures described above.
– Should the message not be displayed again, a second test can be performed:
– Reconnect the disinfectant.
– Set the UF rate and switch on the UF unit.
– Should the error occur at this moment, there is a leakage on V84 (see listing of cleaning program
– errors).
Shunt Cover open
– P.C.B. LP 633 C24 (100n) temporarily short-circuited. (temporarily)
– Shunt interlock defective (check switches).
Voltage Failure
– P.C.B. LP 633 C84 (100n) temporarily short-circuited. (temporarily)
UF1 volume - Error
UF2 volume - Error
– Failure to pass the test for a UF pump. The fill volume for the secondary
– air separator is outside the 100 ml ±4 ml tolerance.
– Possible cause:
– The UF pump fails to deliver correctly (not calibrated or mechanical defect)
– If the test result is >104 ml, the problem can also be caused by air coming from a poorly deaerated
– dialyzer.
F327 UF-failure
– Pause between two UF1 pump strokes less than 220 ms.
– Possible cause:
– CPU-1 defective.
F328 UF-failure
– Pulse time of one UF1 pump stroke less than 180 ms.
– Possible cause:
– Controlling monoflop on LP 634 defective.
F329 UF-failure
– Pulse time of one UF1 pump stroke exceeds 500 ms
– Possible cause:
– Controlling monoflop on LP 634 defective.
F330 UF-failure
– Pick-up time of the UF1 pump exceeds 10 sec.
– Possible cause:
– Controlling output stage on LP 634 defective.
F331 UF-failure
– Theoretical/actual rate of the UF1 pump deviates by more than ± 10 %.
– Possible cause:
– Communication problem or CPU-1/CPU-2 software problem.
F332 UF-failure
– UF1 pump stopped for more than the maximum time period.
– Possible cause:
– Controlling output stage on LP 634 defective.
– UF pump interruption.
29
– Communication problem or CPU-1/CPU-2 software problem.
F333 UF-failure
– UF1 volume change more than 10 ml although UF is switched off.
– Possible cause:
– Communication problem or CPU-1/CPU-2 software problem.
F334 UF-failure
– Pause betweeen two UF2 pump strokes exceeds 220 ms.
– Possible cause:
– CPU-1 defective.
F335 UF-failure
– Pulse time of one UF2 pump stroke less than 180 ms.
– Possible cause:
– Controlling monoflop on LP 634 defective.
F336 UF-failure
– Pulse time of one UF2 pump stroke exceeds 500 ms.
– Possible cause:
– Controlling monoflop on LP 634 defective.
F337 UF-failure
– Starting time of the UF2 pump exceeds 10 sec.
– Possible cause:
– Controlling output stage on LP 634 defective.
F338 UF-failure
– Theoretical/actual rate of the UF2 pump deviates by more than ± 10 %.
– Possible cause:
– Communication problem or CPU-1/CPU-2 software problem.
F339 UF-failure
– UF2 pump stopped for more than the maximum time period.
– Possible cause:
– Controlling output stage on LP 634 defective.
– UF pump interruption.
– Communication problem or CPU-1/CPU-2 software problem.
F340 UF-failure
– UF2 volume change more than 10 ml although UF is switched off.
– Possible cause:
– Communication problem or CPU-1/CPU-2 software problem.
F341 UF-failure
– Mechanical UF1 pump failure.
– Possible cause:
– Broken spring.
– Contaminated filter.
F342 UF-failure
– Mechanical UF2 pump failure.
– Possible cause:
– Broken spring.
– Contaminated filter.
F343 UF-failure
– UF1/UF2 pump volume difference
– Possible cause:
– Delivery volume altered..
30
Error Messages on 4008 Display
HPU Error The hydraulics processor no longer answers CPU1.
– Problem on LP 926 board.
– Problem on Can distributor board.
– Problem on LP 763 board.
– Problem on motherboard.
HPU Error F00
– The HPU logs off with index STATUS_ER; no bit is set in the error bit field.
– Problem on P.C.B. LP 941.
– Problem on CAN distributor board.
– Problem on P.C.B. LP 763.
– Problem on motherboard.
HPU Error F01
– The cyclic communication has failed for more than 2 seconds.
– System error.
HPU Error F02
– The response to an event violated the time-out.
– System error.
HPU Error F03
– An error occurred in the program sequence.
– System error.
HPU Error F04
– Voltage drop (24V_SW) during HPU operation.
– 24V voltage supply on P.C.B. LP 941 failed (watchdog dropped).
HPU Error F05
– Watchdog test failed to be passed.
– Watchdog circuit on P.C.B. LP 941.
HPU Error F06.
– Reference voltage monitoring detected an error.
– Reference voltage circuit on P.C.B. LP941 is detective.
HPU Error F07
– The HPU was logged off by the monitor. Wil! not be displayed since CPU1 has already
– stopped the communication.
– System error.
HPU Error F08
– General valve malfunction: may occur in HPU SW 2.01 or 3.00.
– (Software versions before evaluation of the HPU errors).
– System error.
HPU Error F09
– Malfunction of the compressor (185).
– MV43 defective or activated.
– Compressor 185 defective or activated.
– Error on P.C.B. LP 941.
HPU Error F10
– Malfunction of valve MV39.
– MV39 detective or activated.
– Error on P.C.B. LP 941.
HPU Error F11
– Maltunction of test valve (183).
– MV43 detective or activated.
– MV183 defective or activated.
– Error on P.C.B. LP 941.
HPU Error F12
– Maiïunction of evacuation valve (188).
– MV188 detective or activated.
– Error on P.C.B. LP 941.
31
HPU Error F13
– Malfunction of retentate valve (189).
– MV189 detective or activated.
– Error on P.C.B. LP 941.
HPU Error F14
– Defective component on P.C.B. LP 941.
– Error on P.C.B. LP 941.
HPU Error F15
– Error in the HPU software. Valves are activated incorrectly.
– System error.
HPU Error F98
– Proceeding to the T1 test is not allowed after restart.
– System error.
HPU Error F99
– HPU fails without logging off.
– Damaged cable or similar problem.
– HPU Iogged off by CPUI.
– CRC error in the transfer HPU -> CPU1.
– BVM is connected via CAN and software < 3.20 is installed in the BVM.
– The VDE test was performed directly after turning the system on.
– Turn the system on at least 2 minutes before the test.
32
F01 Configuration
– The HPU DIP switch (CPU2,array 2,DIP 4)has been set (erroneously), the HPU does not answer.
– Correct DIP switch settings.
F02 Configuration
– The ONLINE plus ™ DIP switch (CPU2,array 2,DIP 3)has been set (erroneously), the ONLINE plus
– ™ module does not answer.
– Correct DIP switch settings.
F11 Configuration
– The HPU answered via the CAN bus,although the HPU DIP switch has not been set.
– Correct DIP switch settings.
F12 Configuration
– The ONLINE plus ™ module answered via the CAN bus although the ONLINE plus ™ DIP switch
– has not been set.
– Correct DIP switch settings.
F21 Configuration
– The HPU DIP switch has been set,the DIASAFE ® plus DIP switch (CPU2,array 2,DIP 1) has not
– been set.
– Correct DIP switch settings.
F22 Configuration
– The HPU DIP switch and the HDF DIP switch (CPU2,array 2,DIP 2)have been set.
– Correct DIP switch settings.
F23 Configuration
– The HPU DIP switch has been set and the old online pump answered (signal HDF_ON).
– Correct DIP switch settings.
– HDF_ON signal on LP 633R/A22 in ON position.
F24 Configuration
– The ONLINE plus ™ DIP switch has been set,the HPU DIP switch has not been set.
– Correct DIP switch settings.
F25 Configuration
– The HDF DIP switch has been set,the DIASAFE ® plus DIP switch has not been set.
– Correct DIP switch settings.
33
Warnings / Information
24 V Switched Off
– The watchdog relay has been switched off.
– Press the On/Off key to turn the machine off and on again. Call the service.
Air in system
– Air in the dialysate system.
– Switch flow on.
ASP failure !
– The air separating pump is active with out interruption for more than 60 seconds.
– Acknowledge by pressing the Start/Reset key.
BLD dimness warning
– The blood leak detector is contaminated, or large air quantities are present in the dialysate system.
– Acknowledge by pressing the Start/Reset key. Perform a cleaning program. See also message
– fillprogram.
Connect Cone. Line
– No concentrate connected.
– Check concentrate side.
CPU-II failed
– System error.
– Press the On/Off key to turn the machine off and on again. Call the service.
Cyclic PHT FXX
– Leakage in the balancing system during dialysis. Bypass operation. UF off.
– Acknowledge by pressing the Start/Reset key, considering a possible balancing error.
– Repetition after approx 13 min.
Cyclic PHT F04
– Blood alarm present for more than 3 minutes or failure to perform the
– cyclic pressure holding test.
– System error
– Press the Start/Reset key to acknowledge. Or acknowledge by pressing the Start/Reset key,
– considering a possible balancing error.
– Repetition after approx 15 min.
Dialines not conn
– The dialysate couplings are still in theshunt.
– Connect the couplings to the dialyzer.
– First warning approx 9 min.
Dialines not conn (only with AutoPRIME™).
– The dialysate couplings are still in the interlock shunt.
– Connect the couplings to the dialyzer.
Dialyzer failure (only with AutoPRIME™).
– Dialyzer integrity test failed.
– Press the Confirmkey. Discard the line set. Replace the dialyzer. Restart AutoPRIME™ or fill
– with saline.
Fillprogram
– Air is present in the dialysate system. The TMP limits are flashing and are in active.
– Acknowledge automatic airpurging (Fillprogram end)by pressing theStart/Reset key.
– If the message displays repeatedly, discontinue the treatment and call the service.
– Failure to do so may lead to a balancing error (incorrect UFvolume).
Fill-time reached
– Preset time for filling elapsed.
– Check the blood pump and the venous line set. Press the Confirm key.
– Then select Stop Priming? Or after 10 seconds Prime Online? with the Confirm key.
Flow off warning
– The dialysate flow has been switched off.
– Press the Start/ Reset key (unless flow is desired at the moment).
– or switch the dialysate flow on.
Fund. not allowed
– Will be displayed any time a function is not available.
– Observe correct operating sequence.
34
FXXX UF-failure
– UF pump or control function defective.
– Press the Start/ Reset key. Call the service.
Heparin pump alarm
– The heparin syringe has reached its end position. Heparin pump failure.
– Press the Start/ Reset key (no heparinization), or install a new syringe.
– (See Chapter 4.3.7 Heparin Pump Alarm.)
Heparin pump not on
– The heparin pump module is not on.
– Press the Start/ Reset key (no heparinization), or switch the heparin pump on.
Increase blood flow (only with AutoPRIME™)
– Blood pump rate too low.
– Use the A. key on the blood pump module to increase the blood pump rate.
Maximum filling time reached
– Refer to message Fill-time reached.
MODULE error (only with AutoPRIME™)
– Hydraulic processor malfunction.
NO-UF.
– Ultrafiltration is off.
– Press the Start/ Reset key (unless UF is desired at the moment),
– or switch on Ultrafiltration by pressing the UF I/O key.
Profiles paused.
– Profiles have been stopped.
– Restart profiles, or continue to stop them by pressing the Start/Reset key.
Profile time diff.
– Difference between UF time and profile time.
– Press the Start/ Reset key, or stop the profiles.
Relation BPR/UFR ?
– As compared with the delivery rate of the blood pump, the Ultrafiltration rate is too high.
– Press the Start/ Reset key. Reduce the Ultrafiltration rate, or increase the blood pump rate.
Set UF rate.
– It is impossible to reach the UF goal with the remaining UFtime and with the
– programmed UF rate.
– Turn the UF unit off. Adjust the UF parameters (e.g. the UF rate).
Shunt Cover open.
– The shunt cover has been opened.
– Close the shunt cover.
Timer 'Alarm'.
– The timer time has elapsed.
– Press the Start/ Reset key.
Timer stops UF-Rate.
– The UF time has elapsed. The UF pump is not running.
– Press the Start/ Reset key.
UF-Goal reached.
– The UF time has elapsed, and the UF goal has been reached. The UF pump is not running.
– Press the Start/ Reset key.
V84 faultiness !
– Malfunction of valve V84.
– Press the Start/ Reset key. Start disinfection (cleaning programs 1 to 4).
Ven con. -test failed (only with AutoPRIME™)
– Connection test failed.
– Check dialyzer connections, venous line clamp (line) and venous pressure measuring line.
Voltage Failure.
– One of the voltages is outside the limits (+5V, +12V, +24V).
– Press the On/Off key to turn the machine off and on again.
Wrong cone. Supply.
– Wrong concentrate composition.
– Check the concentrate side.
35
Description of machine errors during the cleaning programs .V84 monitoring
Rinse Failure F01
– End of the rinse-free program in Dis I to V (CDS: Dis I to IV).
– Conductivity has been recognized via V84, although the valve is still closed.
– This error message can be acknowledged by pressing the Rinse key.
Rinse Failure F02
– Disinfectant suction phase in Dis I to IV (CDS: Dis I to IV).
– Conductivity has not been recognized via V84, and the “Disinfectant empty ?”
– message has been acknowledged twice.
– This error message cannot be acknowledged.
– Switch the machine off and on again.
– Program DisV (only on machines with advanced hydrolics) .
– No conductivity detectedvia concentrate level sensor, and “desinfectant empty?” message
– Acknowledged twice.
– Error message cannot be acknowledged.
– Turn the machine off and on again.
Rinse Failure F03
– End of the suction phase in Dis I to IV (CDS: Dis I to IV).
– Conductivity has been recognized via V84, although the valve is already closed.
– This error message can be acknowledged by pressing the Disinfection key.
Rinse Failure F04
– End of the suction phase in Dis I to IV (CDS: Dis I to IV).
– The float switch does not recognize any fluid after the disinfectant has been sucked in.
– Aeration of the disinfectant container!
– This error message cannot be acknowledged.
– Switch the machine off and on again.
Rinse Failure F21
– Disinfectant solution phase in Dis I – IV
– Maximum permissible UF pump strokes (160) during the suction phase exceeded.
– Error message cannot be acknowledged.
– Switch the machine off and on again.
Rinse Failure F01, F02 and F03 cause the V84 monitoring flag to be set. I.e. after one of these error
messages has been released, Bergström or ISO-UF dialysis is no longer possible, since it is not
possible to switch the flow off. The V84 malfunction can be eliminated by correctly performing Dis I to
IV (CDS: Dis I to IV); Dis V cannot be used. Another possibility of elimination is to perform the
calibration program (by a service technician only) in the NOVRAM menu item (Reset V84).
36
PSW (pressure switch) monitoring
– The pressure switches are designed as make contacts. Technical data: 700 mbar ±20 mbar
– switching pressure.
– Delta pressure switch:
Switching point 700 mbar ± 20 mbar
– Alcatel-SEL-pressure switch: Switching range 675 – 805 mbar
– Envec pressure switch:
Switching point 700 mbar ± 20 mbar
Rinse Failure F05
– Rinse-free program with following Dis or HDIS or mandatory rinse as individual program in Dis I to V.
– It was impossible to open the pressure switch for PSW_104 (S124) (bicarbonate).
– Observe the pressure exerted on the ring line (max. 500 mbar)!
– Pressure peaks on distribution piping: Frequently occurs in distribution pipings with user pointsif e.g.
– several patients are disconnected simultaneously and disinfection is started
– Switching point of pressure switch too low: Desired value = 700 mbar ± 20 mbar.
– Check acknowledgement of pressure switch on P.C.B. LP 633: bicarbonate : X633L/A19
Rinse Failure F06
– Rinse-free program with following Dis or HDIS or mandatory rinse as individual program in Dis I to V.
– It was impossible to open the pressure switch for PSW_102 (S123) (concentrate).
– Observe the pressure exerted on the ring line (max. 500 mbar)!
– Pressure peaks on distribution piping: Frequently occurs in distribution pipings with user pointsif e.g.
– several patients are disconnected simultaneously and disinfection is started
– Switching point of pressure switch too low: Desired value = 700 mbar ± 20 mbar.
– Check acknowledgement of pressure switch on P.C.B. LP 633: bicarbonate : X633L/A20.
Rinse Failure F07
– Rinse-free program, Dis, HDIS, or mandatory rinse in Dis I to V.
– Pressure drop during the monitoring phase on PSW_104 (S124) (bicar-bonate) or pressure build-up
– impossible.
– Check switching point of pressure switch.
– Check loading pressure (possible splinter or contamination in orifice 151, remove tube and purge
– tube from both ends.)
– Check negative pressure and test orfice 89
– For this purpose, remove tube and purge tube from both ends.
– Check valve 118 and filter 120
– Check CDS valve 104
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633 bicarbonate: X633L/A19.
– Cartridge filter upsream of degassing pump clogged or wrong filter ( filter for disinfectant container)
– Installed. Filters can be distinguished by different adapters.
Rinse Failure F08
– Rinse-free program, Dis, HDIS, or mandatory rinse in Dis I to V.
– Pressure drop during the monitoring phase on PSW_102 (S123) (concentrate) or pressure build-up
– impossible.
– Check switching point of pressure switch.
– Check loading pressure (possible splinter or contamination in orifice 151, remove tube and purge
– tube from both ends.)
– Check negative pressure and test orfice 89
– For this purpose, remove tube and purge tube from both ends.
– Check valve 117 and filter 119
– Check CDS valve 102
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633 concentrate: X633L/A20.
– Cartridge filter upsream of degassing pump clogged or wrong filter ( filter for disinfectant container)
– Installed. Filters can be distinguished by different adapters.
Rinse Failure F09
– Five minutes before the end of the mandatory rinse in Dis I to V.
– Pressure switch PSW_104 (S124) (bicarbonate) or PSW_102 (S123) (concentrate) did not open
– after pressure reduction.
– See Rinse Failure F12
37
Rinse Failure F12
– Rinse-free program with following Dis or HDIS or mandatory rinse as individual program in Dis I to V.
– The pressure switches for PSW_104 (S124) (bicarbonate) and for PSW_102 (S123) (concentrate)
– could not be opened.
– Membrane pumps fails to run.
– V 102 or V 104 Fails to open
– Observe the pressure exerted on the ring line (max. 500 mbar)!
– Pressure peaks on distribution piping: Frequently occurs in distribution pipings with user pointsif e.g.
– several patients are disconnected simultaneously and disinfection is started
– Switching point of pressure switch too low: Desired value = 700 mbar ± 20 mbar.
– Check acknowledgement of pressure switch on P.C.B. LP 633: bicarbonate : X633L/A20.
Rinse Failure F13
– Rinse-free program with following Dis or HDIS or mandatory rinse as individual program in Dis I to V.
– Pressure drop during the monitoring phase on PSW_104 (S124) (bicarbonate) or PSW_102
– (S123) (concentrate) or pressure build-up impossible.
– Check switching point of pressure switch.
– Check loading pressure (possible splinter or contamination in orifice 151, remove tube and purge
– tube from both ends.)
– Check negative pressure and test orfice 89
– For this purpose, remove tube and purge tube from both ends.
– Check valve 117/118 and filter 119/120
– Check CDS valve 102
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633 concentrate: X633L/A20.
– Bicarbonate : X633L/A19
– Cartridge filter upsream of degassing pump clogged or wrong filter ( filter for disinfectant container)
– Installed. Filters can be distinguished by different adapters.
– In case of F07, F08 and F13, the “DO NOT SWITCH OFF !!” message can, in addition, be
– alternately displayed.
– However, this message is displayed only if a mandatory rinse program is requested, since the
– concentrate and bicarbonate lines are still discharged before the machine is switched off..
38
Hydraulics test (check of V91, V99, V100) in machines with central delivery system
Rinse Failure F11
– Three minutes before the end of the mandatory rinse in Dis I to V .
– The pressure switch PSW_102 (S123) (concentrate) did not open after pressure reduction.
– Observe the pressure exerted on the ring line (max. 500 mbar)!
– Pressure peaks on distribution piping: Frequently occurs in distribution pipings with user pointsif e.g.
– several patients are disconnected simultaneously and disinfection is started
– Switching point of pressure switch too low: Desired value = 700 mbar ± 20 mbar.
– Membrane pumps fail to run
– V 102 fails to open electricallly or mechanically
– Check acknowledgement of pressure switch on P.C.B. LP 633: bicarbonate : X633L/A20.
V91/V100 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V V91 or V100 cannot be opened.
– V 91 or V 100 fail to opened P.C.B. LP 634: V91 = X634L/A12; V100 = X634L/c13
– V 91 or V 100 mechanically not open: check sieve (148) upstream of V 100, or valves clogged
– V 99 constantly open (electrically P.C.B. LP 634: X634L/B12 or mechanically
– V 102 not open
– Pressure switch for PSW_102 (S123) fails to switch.
V99 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V V99 cannot be opened.
– V99 fails to open electrically: P.C.B LP 634: X634L/B12.
– V99 fails to open mechanically: check sieve (149) before V99, or V99 clogged.
– Pressure switch for PSW_102 (S 123) fails to switch.
V130 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V
– V130 cannot be opened (applicable to machines with BIBAG only).
– V130 electrically defective: P.C.B. LP 634: X634L/A4.
– V130 mechanically defective or clogged
– Pressure switch for PSW_102 (S 123) fails to switch.
– Check tubbing for bicarbonate suction line and bibag block.
V188 Failure
– V188 electrically defective.
– V188 mechanically defective or clogged
– Pressure switch for PSW_102 (S 123) fails to open.
CDS Failure V104
– Three minutes before the end of the mandatory rinse in CDS: Dis I to IV.
– The pressure on PSW_104 (S124) cannot be built up.
CDS Failure V102
– Three minutes before the end of the mandatory rinse in CDS: Dis I to IV.
– The pressure on PSW_102 (S123) cannot be built up.
F14
– Shortly before the end of the mandatory rinse in Dis I to V (CDS: Dis I to IV).
– The hydraulics test has not been completed correctly, possibly caused by flow problems.
39
Hydraulics test (check of V91 and valve 98) in machines without central delivery system
F14
– Three minutes before the end of the mandatory rinse in Dis I to V.
– It was not possible to readjust the flow to 750 ml/min ± 50 ml/min. V91 defective.
V91 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V.
– After V91 has opened, a flow > 950 ml/min failed to develop. V91 or valve V98 defective.
F14
– Shortly before the end of the mandatory rinse in Dis I to V.
– The hydraulics test has not been completed correctly, possibly caused by flow problems.
40
Hydraulics test (check of V91, V99, V100, V130) in machines with BIBAG and without central
delivery system
Rinse Failure F15
– Three minutes before the end of the mandatory rinse in Dis I to V.
– DS (BIBAG pressure switch 134) could not be opened at the beginning of the test.
– Check pressure switch: Switching point: rated value: 100 mbar + 30
– Suction error of bicarbonate pump.
– V91 constantly open electrically or mechanically.
– V99/100 constantly closed electrically or mechanically.
V91 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V.
– It is impossible to build up pressure on DS (BIBAG pressure switch) via V91.
– Pressure switch fails to close mechanically: check switching point.
– V91 fails to open electrically: P.C.B LP 634: X634L/A12.
– V91 fails to open mechanically (possibly clogged)
– V130 electrically not closed: P.C.B. LP 634: X634L/A4.
– V130 fails to close mechanically (possibly clogged)
– Bibag connector leaking (check O rings)
– Sealing on the bicarbonate suction tube leaking.
– Check acknowledgement of the pressure switch on P.C.B. LP 633: X633L/A8.
– V99 constantly open electrically or mechanically.
V100 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V.
– It is impossible to build up pressure on DS (BIBAG pressure switch 134) via V100.
– V100 fails to open electrically: P.C.B LP 634: X634L/C13.
– V100 fails to open mechanically (possibly clogged)
– V91 constantly open electrically or mechanically.
– Concentrate pump fails to pump.
– Filter (148) clogged.
– Pressure switch fails to open.
Rinse Failure F16
– Three minutes before the end of the mandatory rinse in Dis I to V.
– DS (BIBAG pressure switch) cannot be closed. V99 or V130 is leaky, or V91 does not open.
– V91 fails to open electrically or mechanically.
– V99 constantly open electrically or mechanically.
– V130 constantly open electrically or mechanically.
– Sealing on the concentrate suction tube leaking.
– Pressure switch fails to open.
V99 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V.
– DS (BIBAG pressure switch 134) cannot be opened. V99 does not open.
– V99 fails to open electrically or mechanically.
– V100 fails to open electrically or mechanically.
– Pressure switch fails to open.
– V91 electrically or mechanically open.
– Filter (149) upstream of V99 clogged.
Rinse Failure F17
– Three minutes before the end of the mandatory rinse in Dis I to V.
– DS (BIBAG pressure switch 134) cannot be closed.
– V91 fails to open electrically or mechanically.
– V130 electrically or mechanically open.
– V100 electrically or mechanically open.
– Pressure switch fails to close.
41
V130 Failure
– Three minutes before the end of the mandatory rinse in Dis I to V.
– DS (BIBAG pressure switch) cannot be opened.
– V130 fails to open electrically or mechanically.
– Pressure switch fails to open.
– Check tubing for bicarbonate suction line and bibag block.
– Bicarbonate line squeezed at strain relief.
– Narrowing in the reducer on the bibag connector.
Rinse failure F20
– Impossible to close the pressure switch (134) via V91/100.
– V91 fails to open electrically or mechanically.
– V130/188 electrically or mechanically open.
– Pressure switch fails to close.
V188 Failure
– The pressure on pressure switch (134) cannot be reduced via V188
– V188 fails to open electrically or mechanically.
– Pressure switch fails to open.
– Check tubing for bicarbonate suction line and air separator block.
– Concentrate line squeezed at strain relief.
Rinse Failure F14
– Shortly before the end of the mandatory rinse in Dis I to V.
– The hydraulics test has not been completed correctly, possibly caused by flow problems.
42
V 39 test
V39 Failure
– On opening V39 a difference in pressure (averaged value V39 open – averaged value V39 closed)
– Is detected on the dialysate pressure transducer (182).
– Standard system: < 50 mmHg / bibag system: < 20 mmHg
– V39 fails to open / close electrically or mechanically ( possibly hydraulic processing unit defective).
– It is impossible to re-adjust the degassing pump (P.C.B. LP 634).
– V91, V99, V100 fail to open electrically or mechanically.
– Dialysate pressure transducer (182) defective or not calibrated ( possibly hydraulic processing unit
– defective).
– Filter 210 (upstream of degassing pump) clogged.
43
Further messages which may be displayed before or during a cleaning program
Accumulator empty!
– Battery voltage <17.2 V ±2.5 %
– Only in the event of a power failure in the cleaning programs.
– If the voltage drops below 17 V, the machine will switch off.
Blood Sensed by OD
– Start of a cleaning program in RI I to II, HR I to III, Dis I to V.
– The optical detector in the air detector module recognizes blood.
BIBAG connect. open
– Upon start of a cleaning program in RI I to II, HR I to III, Dis I to V.
– The BIBAG connector is not closed (cap not attached).
Bic line not conn
– Start of cleaning program in RI I to II, HR I to III, Dis I to V, or end of the disinfectant suction
– phase in Dis V.
– The bicarbonate plug is not connected to the rinse chamber.
– Reconnect the bicarbonate plug to the rinse chamber.
Bypass Valve failed
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– V26 is closed although it should be open.
– The program has stopped.
– The error message can be acknowledged by pressing the respective program key.
CDS connectors open
– CDS program active in CDS: Dis I to IV, HR I to II, RI I to II.
– The CDS connectors (concentrate or bicarbonate CDS connector) have been pulled off during an
– activated CDS program.
CDS failure V145/V147
– All cleaning programs, if DIP switches 4 and 8 on CPU1/array II are set to ON.
– 24 V have been switched off.
– The electric state of V145 or V147 does not correspond to the desired state.
– The error cannot be acknowledged.
Conc line not conn
– Start of a cleaning program in RI I to II, HR I to III, Dis I to V, or end of the disinfectant
– suction phase in Dis V.
– The concentrate plug is not connected to the rinse chamber.
– Reconnect the concentrate plug to the rinse chamber.
Connect CDS lines
– Upon start of a CDS program in CDS: Dis I to IV, HR I to III, RI I to II.
– The CDS connection lines are not locked in the CDS shunt box.
Connect Disinfectant
– Disinfectant suction phase in Dis V.
– Request to connect the disinfectant.
CPU-II failed
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The watchdog relay has dropped.
– Communication (RxD or TxD) may be disturbed.
Dialines not conn
– Start of a cleaning program in RI I to II, HR I to III, Dis I to V.
– The dialysate couplings are not connected to the shunt interlock.
Dial. Valve failed
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– V24 or V24B is closed although it should be open.
– The program has stopped.
– The error message can be acknowledged by pressing the respective program key.
Disinfectant empty ?
– Disinfectant suction phase in Dis I to V.
– Dis V: After the disinfectant has been sucked in, the float switch does not recognize any fluid.
– Dis I to IV, Dis VI: The V84 monitoring unit does not recognize any conductivity.
44
Disinf-Temp. too high
– Transition to disinfection in Dis I to V.
– Temperature at the end of the rinse-free procedure > 40 °C.
– Again and again, the rinse-free procedure is prolonged by 1 minute.
– A warning tone is emitted after 4 minutes.
– The message is automatically cleared, and it cannot be acknowledged.
Float-Switch Failure
– During a disinfectant program in the suction phase in Dis I to V (CDS:Dis I to IV).
– The lower switching point of the float switch is not reached within 20 sec.
– The program has stopped.
Flow alarm
– During a cleaning program in RI I to II, HR I to III, Dis I to V, all CDS.
– A current increasing pulse is not recognized for more than 12 seconds.
– The machine continues to run at Eigentakt (10 seconds).
– Upon error elimination, the message is automatically cleared.
HDF-Pump failure
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The HDF pump has stopped, or the speed deviates (rated value: 400 ml/min, actual value: ≤ 300
– ml/min; rated value: 150 ml/min, actual value: ≤ 100 ml/min).
– The error message can be acknowledged for one complete cleaning program run by pressing the
– respective program key.
– The machine emits the following inquiry: “Are you sure ?”.
Heater error
– During the CDS rinsing phase at the end of a hot rinsing program or a hot disinfection program in
– CDS: HR I to III, Dis II to IV.
– The heater signal (P.C.B. LP 633: X633R/A26) is not changing for > 40 sec.
High temperature
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– Temperature > 41 °C; > 90 °C during HR; > 91 °C during IHR.
– The machine continues to run.
– The alarm tone can be acknowledged.
– Upon error elimination, the message is automatically cleared.
Low temperature
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– Temperature < 33 °C; < 78.5 °C during HR.
– The machine continues to run.
– The alarm tone can be acknowledged.
– Upon error elimination, the message is automatically cleared.
No LD alarm
– Priming of the blood line system in RI I to II, HR I to III, Dis I to V.
– The drip chamber in the air detector module does not recognize any alarm.
Please Wait
– Disinfectant suction phase in Dis V.
– Disinfectant is sucked in via the concentrate pump.
Power Failure
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– Line voltage failed.
Press CONFIRM key
– Disinfectant suction phase in Dis V.
– After the disinfectant has been connected, the Confirm key on the menu panel must be pressed to
– start the suction procedure.
– The program has stopped.
Remove CDS lines
– Upon start of a CDS program in CDS: Dis I to IV, HR I to II, RI I to II.
– The CDS connection lines have been locked in the CDS shunt box
– already before the program has been started. They must be removed
– and reconnected, in order to ensure safe function of the reed contacts.
– Actuation of the Test key after a CDS cleaning program.
– The CDS connection lines are still locked in the CDS shunt box.
– Remove the CDS lines or connect a container.
45
Rinse after Disinf.
– Selection of a cleaning program, although a mandatory rinse has been requested in HR.
– A disinfection program has been stopped and subsequently a rinsing or hot rinsing program started.
Rinse required !
– During stored mandatory rinse in Dis I to V.
– The mandatory rinse has been interrupted (e.g. the machine has been switched off).
Shunt Cover open
– Start of a cleaning program or during a cleaning program in RI I to II, HR I to III, Dis I to V.
– The shunt interlock is not closed.
UF-Pump failed
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The UF pump has stopped or the rate deviates (2800 ml/h < UFR < 6000 ml/h).
– The program has stopped.
– The error can be acknowledged by pressing the respective cleaning program key.
UF-Pump 2 failed
– During a cleaning program in RI I to II, HR I to III, Dis I to VI.
– The UF2 pump has stopped (applicable only to machines with 4008 HDF).
– The error can be acknowledged by pressing the respective cleaning program key.
Upper Flow Alarm
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The cleaning flow increases to > 1000 ml/min. The program has stopped.
– The error can be acknowledged by pressing the respective cleaning program key.
V102 Failure
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– V102 has been opened electrically.
– 24 V are switched off.
– The error cannot be acknowledged.
V104 Failure
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– V104 has been opened electrically.
– 24 V are switched off.
– The error cannot be acknowledged.
Voltage Failure
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The 24-V/12-V supply voltages are drifting.
– This error can be acknowledged for 8 sec by pressing the respective program key.
Water alarm
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– The float switch transmits the “no water available” message for more than 10 seconds.
– The balancing chamber has stopped; V41 is permanently open.
– Upon error elimination, the message is automatically cleared.
Water alarm
– During a cleaning program in RI I to II, HR I to III, Dis I to V.
– For more than 30 seconds, the float switch fails to signal that water is required (not applicable to
– recirculation programs).
– The machine continues to run.
– Upon error elimination, the message is automatically cleared.
46
BLOEDPOMP FOUTMELDINGEN/STROKE versie 1.30
E.01
– Pompsegment diameter buiten het bereik.
– Novram defect of gewist
E.02
– De Hex schakelaar staat op een ongedefinieerde positie.
– Verkeerd ingestelde hex.schakelaar.
– 0= Art. bloedpomp
– 1= SN bloedpomp
– B= Keuze stilstand tijd(15-of 30") D= Bedrijfsuren (xlOO)
– E= Service mode (MTS)
– F= Cal. drukopnemer
E.03
– Drukopnemer niet gecalibreerd.
– Novram defect. Drukopnemer calibreren.
E.04
– Looptijd overschrijding in SN gebruik.
– De bloedpompen hebben het 4-voudige van het stroke-volume overschreden, zonder omschakeling.
– Art.BP max.stroke (50)
– SN.BP ingestelde waarde
– Filter verstopt?
– Drukopnemer defect?
E.05
– Stroke volume buiten ingesteld bereik.
– Novram defect of gewist
E.06
– Het omschakelpunt in SN ligt buiten het AD-bereik.
– Drukopnemer niet gecal. Drukopnemer defect.
E.09
– Fout in AD conversie
– Microprocessor defect
E.12
– Fout in draairichting
– Geen detectie van draaiing.
– Hall sensor defect.
– Geen magneet in rotor
– Rotor geblokkeerd
E. 13
– Foutieve stroom detectie tijdens bedrijf
– Stappenmotor sturing defect.
E.14
– Foutieve stroom detectie tijdens stilstand
– Stappenmotor sturing defect.
E.15
– Fout in toeren detectie circuit.
– Het werkelijke toerental komt niet overeen met de ingestelde waarde. -IC4 defect.
Strookvolume afhankelijke omschakelpunten: Onderste punt 75 mmHg.
Strokevolume
10
15
20
25
30
35
bovendruk
110
130
150
175
195
220
40
245
45
270
50
300
ml
mmHg
47
Error Messages ONLINE plus Module
ONL Error F00
– Online module error.
ONL Error F01 Watchdog
– The watchdog is triggered every 500 ms,and the +24V_WD voltage is not switched on.
– Component error in watchdogcircuit (also LP 785 circuit description).
ONL Error F02 Watchdog
– The watchdog is triggered every 250 ms,and the +24V_WD voltage is not switched off.
– Component error in watchdog circuit (also LP 785 circuit description).
ONL Error F03 Watchdog
– The watchdog is triggered every 750 ms,and the +24V_WD voltage is not switched off.
– Component error in watchdog circuit (also LP 785 circuit description).
ONL Error F04 Program
– Error in program sequence. (Time slice not processed correctly.) ESD event.
ONL Error F05 Program
– The +24V_WD voltage dropped to <17 V or was switched off.
– Component error in watchdog circuit (also LP 785 circuit description);
– 24 V supply from dialysis machine <17 V.
ONL Error F06 Communication
– Timeout of communication watchdog exceeded.
– Error in CAN transmission (line break,CAN driver and others).
ONL Error F07 Communication
– An event transmission from module was not confirmed by dialysis machine.
– Error in CAN transmission (line break,CAN driver and others).
ONL Error F08 Valve
– Deviation of valve positions ONL1,ONL2,ONL3 from nominal position.
– Defective valve output stage transistors.
ONL Error F09 T1-test
– T-test was completely skipped.
– T1-test not performed.
ONL Error F10 T1-test
– T1-testfor ONLINEplus faled and was therefore completely skipped.
– T1-test not performed.
ONL Error F11 Ref. voltage
– The reference voltage for the ADC of the processor is outside the tolerance (also initial tests).
– Deviation of one of the two reference voltages.ONLINE plus ™.
ONL Error F12 CRC
– Calculated CRC different from CRC stored in the EEPROM. Defective EEPROM.
ONL Error F13 EEPROM
– Normal and inversely stored value in EEPROM not identical.
– At module start: Defective EEPROM or EEPROM is new and must be initialised (start dialysis
– machine in service mode: EEPROM is autom.initialised).
– In operation:Writing or reading from EEPROM failed.
ONL Error F14 From monitor
– Dialysis machine disabled ONLINE plus ™ module.
– CAN bus or WTRE failed.
ONL Error F16
– Valve error ONL1.
ONL Error F17
– Valve error ONL1.
ONL Error F18
– Valve error ONL1.
48
Error Messages During the Functional Test of the BVM
Software error: cont. ntc test
– Faulty program execution: Reset BVM 4008.
– Timing problems: Reset BVM 4008.
Software error: program error
– Faulty program execution: Reset BVM 4008.
– Timing problems: Reset BVM 4008.
System error: adu defect
– AD converter AX132 supplies incorrect reference voltage (R115).
– MAX132 wiring defective / MAX 132 defective / IC31 (DG408) defective / IC33 (74HC4052) .
– defective / R115 faulty/defective / D18 faulty/defective / IC35 (ICL7650) faulty/ defective.
– Transistor T7 faulty / defective.
– Contact problems / Sensors not connected / Sensors defective.
System error: block temperature.
– Block temperature above maximum (40.5 °C).
– Block above 40.5 °C - Block temperature sensor defective Allow block to cool .
System error: calib P current.
– Transient response of Peltier element (control) defective.
– Voltage divider R83/R84 defective / R113 faulty / Contact problems / NTC defective / PT1000
– block sensor defective.
System error: crc error
– Flash module defective.
– Faulty programming / Contact problems.
System error: eeprom crc 1
– Faulty data in memory area 1 of EEPROM.
– Loss of data caused by current surge / Contact problems / EEPROM defective.
System error: eeprom crc 2
– Faulty data in memory area 2 of EEPROM.
– Contact problems / EEPROM defective.
– Resolder contacts - Replace EEPROM.
System error: eeprom crc 3
– Faulty data in memory area 3 of EEPROM.
System error: eeprom crc 4
– Faulty data in memory area 4 of EEPROM.
– Loss of data caused by current surge / Contact problems / EEPROM defective.
System error: ext. ram error
– External RAM defective.
– Contact problem / RAM defective.
System error: heat off by port
– Software control of Peltier element (on/ off) does not function (it is possible that the supply voltage
– is no longer applied to the Pettier element).
– Contact problems / Wiring T4^T5 defective / Transistors T4/T5 defective / Wiring of switched
– converter LM2576T (1C3) defective / Voltage divider R81/R82 defective / Hardware
– of over-temperature switch defective / Diode D14 defective.
System error: int. ram error
– Internal RAM in processor defective.
– Processor defective.
System error: no power supply
– 24 Volt failure (voltage fluctuations) without failure of 5 Volt.
– Voltage fluctuations in power supply unit Failure of 1C 38B (LM339A) 4008 has switched to
– battery operation.
System error: ntc control
– Over-temperature switch does not function.
– Contact problems / Over-temperature switch / wiring defective / IC38 (LM339A) defective /
– T10/T13 defective.
– Diodes D14/D15 defective / NTC faulty/defective / VCC (5V) faulty.
49
System error: ntc Ctrl delay
– Jumper wiring of over-temperature switch does not function properly.
– Wiring IC17(4060BT) defective / IC17 defective / Contact problems / CAN controller defective.
System error: Pelt. OVERTEMP
– OVERTEMP signal (connection over-temperature switch to switch for Peltier current) has
– inadmissible level.
– Connection interrupted CAN controller defective.
System error: US current
– Ultrasonic current not within allowed range:
– US module off: I =0 mA.
– US module on: I = [2.5/3.5 mA].
– Hybrid module not connected Wiring for current measurement defective.
– Contact problems Hybrid module defective.
System error: USonic switch
– Software control of hybrid module (on/off) does not function.
– Hybrid module not connected Wiring for current measurement defective (D17/T12/...).
– Contact problems Hybrid module defective / CAN controller defective (IC27).
System error: vcc error
– VCC out of range of 4.75/5.25 V.
– Voltage fluctuation of 5 Volt (VCC) / Transistor T10 defective / Linear controller IC1 (LT1129)
– defective.
– ADD channel for VCC-regulation defective.
– Reset BVM 4008.
System error: V Pelt. too high
– Peltier supply voltage above 8.8 V.
– Voltage divider R81/R82 faulty Switched converter LP2576T (IC3) defective.
– Short-circuit in control circuit of Peltier element control.
System error: V Pelt. too low
– Peltier supply voltage above 7.2 V.
– Voltage divider R81/R82 faulty Switched converter LP2576T (IC3) defective.
– Short-circuit in control circuit of Peltier element.
System error: watchdog error
– Hardware watchdog does not function or has triggered RESET.
– Software has not retriggered watchdog.
– Contact problems / Components of watchdog assembly defective / Wrong processor
– Quartz frequency.
– BVM 4008 reset (press A and V keys at the same time) (the set time basis may be wrong).
System error: watchdog test
– Hardware watchdog functions, but mono-flop for watch-dog status is defective (1C 118).
– Contact problems Components of watchdog assembly defective.
System error: 5 voltage error
– 5 Volt failure without failure of 24 Volt.
– Voltage fluctuation of 5 Volt (VCC).
– Reset BVM 4008.
Treatment error: door opened!
– Door was opened during treatment.
– Door was opened by user or patient.
50
Error Messages and Warnings Displayed by the BVM 4008 During Preparation and Treatment
BV-Control off by 4008!
– Blood volume control was aborted by the dialysis machine.
– This message is displayed briefly within one minute at certain intervals.
– After the Start/Stop is pressed, the warning will either be displayed again (control cannot be
– restarted) or blood volume control is restarted.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The determination of the relative blood volume (RBV) is not affected by the missing data
– connection.
BV-Control off instability
– Blood volume control was aborted by the BVM 4008.
– This warning is shown on the BVM display when the BVM measuring system is unable to
– determine current RBV measuring values for a certain time.
– This message is displayed briefly within one minute at certain intervals.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The relative blood volume (RBV) is displayed again as soon as the BVM 4008 can resume the
– measurement and the control function can be restarted with the Start/Stop key.
BV-Control off middle UP rate
– Blood volume control was aborted by the BVM 4008.
– This warning is shown on the BVM display when the remaining ultrafiltration rate (UF-rate) is
– higher than the quotient of initial ultrafiltration volume and total ultrafiltration time.
– Restarting of blood volume control (via the Start/Stop key) is possible in this case.
– This message is displayed briefly within one minute at certain intervals.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The determination of the relative blood volume (RBV) is not affected by the aborted blood volume
– control.
BV-Control off no start value
– The blood volume control could not be started because the RBV start value was not found.
– Note: The operating steps in -» Chapter 2.3.1 must be strictly followed.
– This message is displayed briefly within one minute at certain intervals.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The BVM 4008 continues its attempts to find an RBV start value and the control function can be
– restarted with the Start/Stop key.
BV-Control OFF RBV drift
– Blood volume control was aborted by the BVM 4008.
– This warning is shown on the BVM display when successive RBV measured values deviate too
– much from each other within a certain time window (excessive drift).
– This message is displayed briefly within one minute at certain intervals.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The determination of the relative blood volume (RBV) is not affected by the aborted blood volume
– control.
– If the measured values are within the tolerance, the control function can be restarted with the
– Start/Stop key.
BV-Control off UF rate limit
– Blood volume control was aborted by the BVM 4008.
– This warning is shown on the BVM display when the currently calculated ultrafiltration rate
– (UF-rate) is higher than a maximum allowable limit value.
– Restarting of blood volume control (via the Start/Stop key) is not possible.
– This message is displayed briefly within one minute at certain intervals.
– The message can be re-activated by pressing the Start/Stop key.
– Corresponding information for the user is also shown on the display of the dialysis machine.
– The determination of the relative blood volume (RBV) is not affected by the abortion of the
– blood volume control.
51
BVM not calibrated!
– This message is displayed when the BVM 4008 is switched on but the blood volume monitor has
– not been correctly calibrated (machine calibration) or when data stored in the machine are faulty.
– Machine calibration must then be performed again by the Fresenius Service.
– As a result of this warning, the BVM 4008 stops operation after further system tests have
– been performed.
– The error message: "system error: eeprom crc1" is shown on the display.
Cuvette unsuitable
– During cuvette calibration, a check is made on whether the cuvette meets the minimum
– requirements. If this is not the case, this message is displayed. Use of the cuvette to measure
– the blood volume is possible, but is not recommended because of potentially higher measuring
– errors. If this message is displayed repeatedly with various cuvettes, there might be a malfunction
– of the measuring head. The device must then be checked by the Fresenius service department.
– This message must be confirmed by pressing the Enter key before blood volume measurement
– can be performed (possibly with considerably reduced measuring accuracy).
Door Open
– The door of the measuring head has not been closed correctly. Therefore, measurement
– cannot be started.
No Blood
– The BVM 4008 expects blood in the line set. However, the measurement shows that no blood
– is present.
No Calibration
– At the start of dialysis, the BVM 4008 indicates that the cuvette used has not been calibrated.
– This message has to be confirmed by pressing the Enter key before blood volume measurement
– can be performed (possibly with reduced measuring accuracy).
No communication!
– After the Start/Stop is pressed, the BVM 4008 informs the user that no data connection
– (communication) exists between the BVM 4008 and the dialysis machine.
– This message is displayed briefly within one minute at certain intervals.
– The message can be subsequently re-activated by pressing the Start / Stop key.
– The activation of blood volume control is not possible without data connection.
– The determination of the relative blood volume (RBV) is not affected by the missing data
– connection.
No Saline Solution
– For a function to be executed (e.g. cuvette calibration), the BVM 4008 requires physiological
– saline or dialysate as a measurement medium, but the measurement shows the presence of
– another measurement medium.
– This message has to be confirmed by pressing the Enter key.
Service Mode
– The BVM 4008 has been set to the service mode (intended only for use by the Fresenius service
– department!). To return to the normal mode, the configuration switch No. 1 has to be set to OFF.
– The BVM 4008 can be started only by switching the dialysis system on again.
52
Error Messages and Warnings Displayed by the Dialysis System During Preparation and
Treatment
BVM error
– Error was caused by the BVM 4008, see display of the BVM 4008.
– See error description of the BVM 4008.
BVM: bpr too low
– During BVM calibration, the effective blood pump rate is too low (< 290 ml/min).
– Set the effective BPR to greater than 300 ml/min, otherwise the circulating saline solution is too
– cold for calibration.
BVM: enter critRBV
– Attempt to start the control function without predefining the critical RBV.
– Enter the critical RBV, then press the Start/Stop key again.
BVM failure: CAN
– Failure was caused by the 4008:
– The CAN communication to the BVM 4008 was aborted.
BVM failure: DOOR
– The door of the BVM module was opend while treatment was in progress.
BVM failure: INT1
– Failure was caused by the 4008:
– Faulty assignment of UFR.
BVM failure: INT2
– Failure was caused by the 4008:
– Change to UFR that is not allowed.
BVM failure: INST
– Failure was caused by the 4008:
– The measured values of BVM 4008 were instable for too long.
BVM failure: INTO
– Failure was caused by the 4008:
– UFR of BVM was negative.
BVM failure: INTX
– Failure was caused by the 4008.
BVM failure: PWRF
– A power failure without battery was detected and the control in operation before the power failure
– was terminated.
– Continue UF with the displayed remaining average UFR (UF key). Control is no longer possible.
BVM failure: SEC2
– Failure was caused by the 4008:
– UF rate was too high.
BVM failure: SEC3
– Failure was caused by the 4008:
– Treatment goal cannot be achieved.
BVM failure: SEC4
– Failure was caused by the 4008:
– UF rate was too high.
BVM failure: SEC5
– Failure was caused by the 4008:
– UF rate was too high.
BVM failure: UFPE
– CPU2 stopped the UF pump because of an erroneous configuration.
BVM failure: 100%
– Failure was caused by the 4008:
– The BVM 4008 took too long to find the100%RBV.
BVM not calibrated
– The cuvette in the BVM 4008 is not calibrated and the Start/Stop key on the BVM 4008 was
– pressed.
– Calibrate cuvette, if still possible, then restart control. Control is not possible without a calibrated
– cuvette.
53
BVM not possible
– The dialysis machine was in the SN Click-Clack mode and the Start/Stop key on the BVM module
– was pressed.
– Select an appropriate treatment mode, then restart the control function.
BVM not ready!
– The status of the BVM 4008 is set to not ready to measure (e.g. no blood in cuvette) and the Start/
Stop key on the BVM 4008 was pressed.
– Restart control when the BVM 4008 is ready to measure.
BVM: only ISO-time
– Attempt to start the UF control function while standard ISO was in progress.
– Abort standard ISO. Enter only the ISO time. Restart ISO, then start the UF control function.
BVM: pArt too low
– During BVM calibration, the arterial pressure is too low (< - 100 mmHg).
– Make sure the cone on the saline solution bag is well separated, otherwise there is the risk of
– degassing.
BVM: UF-control?
– Control could be switched on but has not been for a period of 3 minutes.
– 1. Switch on control
– 2. Press the Dialysis Start key.
BVM: UF-goal ???
– Abort of UF control because the UF goal will not be achieved with controlling in the remaining
– dialysis time.
– Continue to ultrafiltrate (UF key) at the predefined average UF rate without controlling.
– Caution: High risk of drop in blood pressure!
– Increase the UF time or reduce the UF goal, then restart UF and control.
– Caution: The physician must change the prescription of the treatment!
BVM + ISO t=...
– Standard message displayed during dialysis treatment with ISO and UF control
Dialysis with BVM
– Standard message displayed during dialysis treatment.
First start UF !
– The UF unit on the dialysis machine was not switched on and the Start/Stop key on the
– BVM 4008 was pressed.
– Start UF unit, then restart control.
H(D)F with BVM
– Standard message displayed when using "old" HF or "old" HDF with UF control whether
– with or without SN.
HDF with BVM
– Standard message displayed during HDF with ONLINE^™ and UF control
HF with BVM
– Standard message displayed during HF with ONLINE/^™ and UF control
MODULE Error
– The CAN communication to the BVM 4008 was interrupted while not controlling.
No blood in system
– The optical detector of the 4008 has still not detected any blood and the Start/Stop key on
– the BVM 4008 was pressed.
– Wait until the optical detector detects blood, then restart control.
SN-HDF with BVM
– Standard message displayed during SN HDF with ONLINE^™ and UF control.
SN - HF with BVM
– Standard message displayed during SN HF with ONLINE/^™ and UF control.
SN with BVM
– Standard message displayed during SN dialysis with UF control
Stop UF-profile
– The dialysis machine was performing a UF profile and the Start/Stop key on the BVM 4008
– was pressed.
– Terminate UF profile, then restart control.
54
UF-control aborted
– The control was aborted before the UF goal was reached. This can be caused by the user
– (UF key or Start/Stop key) or by an error condition in the control.
– Confirm warning with the Dialysis Start key.
– When aborted by the user, the control can be switched on again.
– When one more of the following error messages is displayed after the Dialysis Start
– key has been pressed for10 seconds, the control cannot be switched on again.
UFR limit!
– Twice the current UFR exceeds the maximum UFR on the DIP switches or the UFRmax
– that was set on the BVM module.
– Warning. The control function is unable to take advantage of the entire UFR range.
Wrong UF-param.
– No UF goal was entered on the dialysis machine or a UF rate was entered that exceeds the
– max. permissible UF rate of the BVM, and the Start/Stop key on the BVM 4008 was pressed.
– Enter the UF goal or enter a UF rate that does not exceed the max. UF rate specified in the
– BVM menu, start UF, then start the control function.
4008 HDF with BVM
– Standard message displayed during 4008 HDF with UF control.
– SN with 4008 HDF impossible
4008 HF with BVM
– Standard message displayed during HF with UF control.
– SN with 4008 HF impossible
55
ERROR MESSAGES BPM
Group 0: Serious hardware errors
If an error of group 0 occurs, safe operation is no longer guaranteed. After pressing of the Stop / RESET
key, the Blood Pressure Monitor, therefore, performs a new initialization (system reset).
CAUTION!
Should an error of this group occur repeatedly, you may not perform any further measurements on the
patient. You must, under all circumstances, notify the Technical Customer Service.
E00 BPM RAM Check Error
– Writing on and reading out of the internal RAM delivered different results.
E01 BPM ROM Check Error
– CRC test run turned out to be negative.
E02 Module RAM Check Error
– Error in the RAM of the internal measuring module.
E03 Module ROM Check Error
– Error in the ROM of the internal measuring module.
E04 EEPROM Error
– CRC error during the check of the default values of the alarm limits.
E05 BPM CPU Check Error
– Functional check of CPU turned out to be negative
EO8 Program Run Error
– The program was not processed in the correct order.
E09 Module Single Failure Error
– Hardware error of the internal measuring module.
Group 1: Errors during blood pressure measurement
Should an error of group 1 or 2 occur, the Blood Pressure Module delivers the error codes listed in the
error name column instead of the measured results. The device automatically attempts up to three
times to repeat the measurement in order to achieve correct measurement results. If the STOP / Reset
key is pressed, normal operation is continued. These error codes are not stored in the EEPROM.
E 10 Measurement Error E03
– The inflation pressure failed to reach at least 10mmHg after 12 seconds.
– No cuff connected. Leakage in the line set.
E11 Measurement Error C11
– The inflation time exceeded 60 seconds.
– Leakage in the line set.
E 12 Measurement Error C 12
– The cuff pressure fell below 10 mmHg without completion of the measurement yet.
– Leakage in the line set, which occurred during the measurement.
E 13 Measurement Error C 13
– Movement artefact.
– The patient must be put in the appropriate resting position.
E 14 Measurement Error C 14
– Selected inflation pressure too low.
– The patient's systolic pressure exceeds the selected inflation pressure.
– The BPM 4008 automatically increases the inflation pressure by 50 mmHg and repeats the
– measurement.
E 15 Measurement Error C 15
– Detected variations in pressure are not caused by pulsations.
– External influence on the cuff. Take care that the upper arm including cuff is positioned freely.
E 16 Measurement Error C 16
– Cardiac arrhythmia.
– Allow adequate blood return; then restart the measurement; if necessary, check the patient's pulse
– quality.
E 17 Measurement Error C 17
– Total measurement time exceeded 180 seconds (including repeated attempts).
– Allow adequate blood return and restart the measurement.
56
E 18 Measurement Error C 18
– More than 100 pulsations during one measurement.
– Allow adequate blood return; restart the measurement; if necessary, check the patient's pulse
– quality.
E 19 Measurement Error C 19
– Cuff pressure exceeded 320 mmHg.
– In case this error occurs repeatedly, notify the Customer Service.
Group 2: Errors during blood pressure measurement
For description see group 1.
E20 Measurement Error C20 Pulse too weak.
– If necessary, select a different place for the cuff, or remove clothes from the upper arm.
E21 Measurement Error C21
– Inflation pressure reached too quickly.
– The cuff for neonatus was used in the adult mode.
– The pressure line to the cuff is occluded.
Group 3: BPM transmission errors
Errors of this group indicate failures in the communication between the blood pressure module and the
BPM 4008 or errors in the internal transmission of data. After the STOP / RESET key has been pressed,
a system reset is initiated.
CAUTION!
Should an error of this group occur repeatedly, you may not perform any further measurements on the
patient. You must, under all circumstances, notify the Technical Customer Service.
E30 Module Busy Timeout Error
– Occurs if the CTS signal of the internal measuring module is high for more than 30 sec.
E31 Host to Module Data Error
– If the internal measuring module receives an invalid sequence of commands, it reacts with the
– System Reset Response. In this case, the module must be re-initialized. This necessitates a
– system reset.
E32 Module to Host Data Error
– Indicates that the host received an invalid sequence of commands.
Group 4: DUART transmission errors
Errors of this group indicate failures in the communication between the BPM 4008 Blood Pressure
Monitor and the host system ( hemodialysis machine or service computer) or errors in the transmission
of data. Pressing the STOP/ RESET key continues normal operation.
CAUTION!
Should an error of this group occur repeatedly, you may not perform any further measurements on the
patient. You must, under all circumstances, notify the Technical Customer Service.
E40 Channel A Send Buffer Overflow
– Channel A was blocked too long, e.g. by transmission of the XOFF control combination without
– release of XON.
Group 5: PC bus errors
Errors of this group indicate failures in the connection between the touch panel and the EEPROM.
After the STOP / RESET key has been pressed, a system reset is performed. These error codes are not
stored in the EEPROM.
CAUTION!
Should an error of this group occur repeatedly, you may not perform any further measurements on the
patient. You must, under all circumstances, notify the Technical Customer Service.
E50 PC-Bus Timeout Error
– EEPROM or keyboard port does not respond within the time specified (SCL signal remains LOW).
E51 EEPROM Read Error
– Error during reading of the default values.
57
E52 EEPROM Write Error
– Error during writing of the default values
E53 Keyboard Read Error
– Error during reading of the keyboard port.
E54 Keyboard Write Error
– Error during writing on the keyboard port.
E55 Errorcode Reading Error
– CRC checksum error during reading of error codes.
Group 6: CAN bus errors
Error group 6 is reserved for errors of the CAN bus and will be defined only when the CAN bus is
implemented.
CAUTION!
Should an error of this group occur repeatedly, you may not perform any further measurements on the
patient. You must, under all circumstances, notify the Technical Customer Service.
E60 reserved
E61 reserved
58
Handling of Error Messages of the BTM
The errors are classified into 8 categories according to the most significant digit of the error number
E0xxx
The measured value is beyond the allowed range.
E1xxx
The BTM is technically faulty.
E2xxx
There is an operating error which can be corrected. As soon as the problem is eliminated,
the BTM continues its function.
E3xxx
A grave operating error has occured. The BTM is switched off.
E5xxx
A value of the internal parameter memory (see Chapter B 12.4), which is open to change
By + service technicians only, is beyond the allowed range or the parameter memory is
faulty.
E6xxx
Software error.
E7xxx
An incident occuring during recirculation measurement resulted in an unsuccessful
termination of the measurement.
E8xxx
An incident during an active control function resulted in the termination of the control
function.
Grave Errors
Errors of the category E0xxx, E1xxx, E3xxx, E5xxx und E6xxx indicate a grave problem. For
reasons of safety, the BTM is turned off. In such an event, control of dialysate temperature is again left
to the dialysis machine. The BTM terminates all functions and only displays the error message. There
is no longer any influence on the treatment. A grave error can be recognized by all of the 7 LED’s
being switched off. Press RESET-key combination to restart the BTM. It then behaves in the same
way as after power-on.
Before pressing RESET, make sure that there is no defect or any other failure relevant to
safety.
Warnings / Indications
Errors of the category E2xxx, E7xxx and E8xxx do not result in a turnoff of the BTM. They are to
understand as indications or warnings for the user. In most cases, a fast flashing LED of a key in
connection with the concerned function indicates the error cause. Partly these messages are
answered by pressing ENTER; partly, they extinguish automatically as soon as the cause is
eliminated.
Handling of Error Messages
Error messages occur due to operating errors, inadmissible conditions while a control function or a
recirculation measurement is active, and technical defects.
Error messages are displayed as four-digit error numbers with the letter E in front. In Chapter C2
possible causes and the nessesary measures to take are listed for all error messages. If a grave error,
e.g. caused by a technical problem, occurs, the BTM will terminate operation, which is indicated by all
LEDs being switched off. The most frequent errors are:
E2001
Arterial blood line has not been detected.
E2002
Venous blood line has not been detected.
E2003
Both arterial and venous blood line have not been detected.
E2203
Temperature of sensor heads is to high for tube detection, (also see B4.4)
E7601
Blood flow has been changed during recirculation test.
E7801
During recirculation measurement a blood alarm has occured or dialysis treatment has been
completed.
E8800
Dialysis treatment was stopped or interrupted for a longer period of time while a control
program has been active.
59
C2 List of Error Messages
E0xxx Measuring Results out of Allowed Range
E0001
E0002
E0003
E0004
E0010
E0020
E0030
E0040
E0100
E0200
E0300
E0400
E0500
E0600,
E0601
Arterial line temperature < 12°C.
Arterial line temperature > 42 °C.
Arterial sensor head temperature < 12°C.
Arterial sensor head temperature > 42°C.
Venous line temperature < 12°C.
Venous line temperature > 42°C.
Venous sensor head temperature < 12°C.
Venous sensor head temperature > 42°C.
Problem: The allowed temperature measuring range is 12°C to 42°C. This error may
occur
due to a technical defect of the BTM or the dialysis machine or in case of the BTM
being
used to measure fluids beyond the allowed temperature range.
Solution: First of all check, whether line temperature is really beyond the
range of 12-42°C
or adjacent to the limits. If this is not the case, do not continue using the BTM.
Call service.
The measured value of dialysate temperature is < 32 °C. Solution: Call service.
Dialysate temperature is > 41 °C and the blood pump is running (BFR > 10 ml/min).
Problem: As dialysate temperatures above 41 °C are critical, usually the BTM is
turned off, except the dialysis machine is placed in hot rinse mode. The BTM is not
able to determine for every type of dialysis machine whether it is in hot rinse mode or
not. Instead the blood flow rate is monitored and dialysate temperatures > 41 °C are
tolerated only if the blood pump stands still.
Solution: If the blood pump has been set in motion during hot rinsing or during the
phase of cooling down, wait until dialysate temperature has decreased below 41 °C.
Thereafter the BTM may be restarted with the RESET key combination. If the blood
pump has not been set in motion, the BTM or the dialysis machine is defect or
decalibrated, respectively.
The measured blood flow rate is not valid (negative, <-10 ml/min).
Solution: Check cable connection (20-pin cable to the monitor and to the blood
pump).
In case of recurrent errors call service.
The measured blood flow rate is > 550 ml/min. However, a maximum flow rate of 500
ml/min has been selected by the setting of the DIP switch (S 1; LP 906). The measured
blood fow rate is > 650 ml/min, whilst a maximum flow of 600 ml/min has been selected by
the setting of the DIP switch 2 (Sl; LP 906)
Solution: Check DIP switch no. 2 (S 1; LP 906) to make sure the right type of blood
pump has been selected (s. Chapter D3). If this does not correct the problem,
call service.
The measured recirculation is invalid (>= 1).
Problem: Possibly arterial and venous blood lines have been exchanged or have not
been inserted properly into the sensor heads!
Solution: Check the whole extracorporeal system.
E1 xxx Technical Defect
If errors of this category occur (left figure =1), there is a technical problem of the BTM (except E 1103:
configuration may not be correct).
60
E1031
E1032
Arterial sensor head does not reach thermal equilibrium.
Venous sensor head does not reach thermal equilibrium.
Explanation: During normal operation, the sensor head of the BTM is heated to the
temperature of the blood in the tube. This condition, called "thermal equilibrium of the
sensor head", is a prerequisite for a precise blood temperature measurement. After a
change in blood temperature it takes up to 3 minutes until equilibrium is reached
again. The error message appears if the equilibrium is not reached after 10 minutes.
Solution: If a sensor head is unusually hot the service should be called. Check the
two fuses Fl and F2 on the BTM board (LP 906). If they are both intact or if one of
them blows repeatedly, call the service.
E1101
E1102
E1103
CRC error in message from dialysis machine.
Instruction cannot be executed by dialysis machine.
Timeout-error during data reception (microprocessor-controlled dialysis-machine).
Problem: All 3 messages occur due to data transfer problems between BTM and
dialysis machine. In case of El 103, the DIP switch no. 3 (S 1; LP 906) may be set
wrong.
Solution: Check position of DIP switch no. 3 (Sl; LP 906) for selection of the dialysis
machine type (in case of El 103). Check cable connection between BTM and dialysis
machine. If you are unable to remove the problem this way, call service.
E1104
E1105
E1106
E1107
E1108
E1109
Data from dialysis machine unusable.
Inconsistent answer received from dialysis machine.
Send buffer overflow.
No acknowledge received for byte sent to dialysis machine.
Timeout error: No answer to instruction received from dialysis machine.
Communication dialysis machine to communication coprocessor interrupted.
Problem: All messages occur due to data transfer problems between BTM, dialysis
machine and communication coprocessor (if present). In case of El 107, the DIP
switch no. 3 (Sl; LP 906) may be set wrong.
Solution: Check position of DIP switch no. 3 (S 1; LP 906) for selection of the dialysis
machine type (in case of El 107). Check cable connections between BTM, dialysis
machine and communication coprocessor (if present). If you are unable to solve the
problem this way, call service.
Hardware error on board LP906.
The parameter memory (EEprom) is defect; CRC-test was negative.
The program memory (Eprom) is defect; CRC-test was negative.
Solution: BTM is no longer operable; service has to be called.
E1200
E1300
E1301
E2xxx Operating Errors
Messages of this category indicate removable problems mostly due to operating errors. Error no.
Explanation
E2001
E2002
Arterial blood line has not been detected.
Problem: While the venous blood line has been detected at least 2 minutes ago, the
arterial line has not been detected. Therefore no control function or recirculation
measurement can be activated. This message also occurs if blood line temperature is
below 33.5°C.
Solution: Check arterial blood line to make sure it is properly inserted into the arterial
(upper) sensor head and that the sensor head is closed. After the line has been
inserted properly, it may take some time (up to about 1 minute) until the message
does no longer appear on selection of a control function or recirculation
measurement.
Venous blood line has not been detected.
Problem: While the arterial blood line has been detected at least 2 minutes ago, the
venous line has not been detected. Therefore no control function or recirculation
measurement can be activated. This message also occurs if blood line temperature is
below 33.5°C.
Solution: Check venous blood line to make sure it is properly inserted into the venous
(lower) sensor head and that the sensor head is closed. After the line has been
61
E2003
inserted properly it may take some time (up to about 1 minute) until the message does
no longer appear on selection of a control function or recirculation measurement.
Both arterial and venous blood lines have not been detected.
Problem: Neither of the two blood lines has been detected, while the blood pump has
been running for at least 2 minutes. Therefore no control function or recirculation
measurement can be activated. This message also occurs if the blood temperatures
of both lines are below 33.5 °C.
Solution: Check both blood lines to make sure they are inserted properly into the
sensor heads and that the sensor heads are closed. After the lines have been
inserted properly, it may take some time (up to about 1 minute) until the message
does no longer appear on selection of a control function or recirculation
measurement.
Messages E2001 - E2003 may occur especially if a control function or recirculation measurement has
already been selected during priming process (which does not make sense). In the event that one of
these 3 messages occurs, flashing LED's on the keys T-ART or T-VEN indicate which line (arterial,
venous or both) is to be checked (see also Chapter B4.4).
E2202
E2203
If this operating mode is selected, blood line temperatures (not fistula temperatures) are
displayed.
Explanation: Display of line temperature has been selected by DIP switch no. 6 (S 1;
LP 906). After starting of the BTM this message simply indicates the choosen
operating mode, avoiding that line temperature instead of fistula temperature will be
displayed accidentally for a longer period of time.
Solution: If display of the line temperatures is intended, the message can be erased
by pressing ENTER. On the other hand, if fistula temperatures should be displayed,
turn off the dialysis machine, correct position of DIP switch no. 6 (S 1; LP 906) and
than restart the dialysis machine.
Temperature of sensor heads is to high.
Explanation: At the beginning of a dialysis treatment and during the priming process,
respectively, the BTM checks whether the blood lines are properly inserted into the
sensor heads. For this purpose the temperature of the sensor heads has to be initially
below 32.5°C. If this is not the case, the display of this error message demands to
wait until sensor heads have cooled down adequately. Afterwards the message
extinguishes automatically. This message usually occurs if there is a longer period of
time with the blood pump not working following the priming procedure. Consequently
the BTM passes into the inactive mode and then restarts tube detection (see also
Chapter B4.3).
Solution: Wait until sensor heads have cooled down adequately and message
extinguishes automatically. However this message can be skipped by pressing
ENTER. In this case
detection of tubes is not reliable any longer! The user is responsible for
deactivating an essential safety feature of the BTM!
E3xxx Grave Operating errors
Messages of this category indicate operating errors which result in turnoff of the BTM.
E3004
E3101
E3201
62
Arterial and venous line exchanged and inserted this way into the sensor heads. This error
will be detected some time after starting a control program.
Solution: Insert lines properly; press RESET to restart BTM.
Maximum operating time of the BTM has been attained.
Solution: Press RESET to restart BTM.
Dialysis machine without microprocessor control: The connection between the BTM and
monitor of the dialysis machine (LP433) is not installed or interrupted.
Solution: Turn off dialysis machine, install connection properly, restart dialysis
machine.
E5xxx Inadmissible Data in Parameter Memory (EEprom)
Errors of the category E5xxx occur only after an inadmissible input in the parameter memory, e.g.
through the service technician. The parameter memory is loaded with calibration parameters as well
as a plurality of user specific parameters, which especially allow to set options for control functions
and for the recirculation measurement (see Chapter B12.4, D6). For each of the parameters upper and
lower limits are prescribed, which are checked every time the BTM has been turned on. If the stored
parameter is beyond these limits an error message is displayed and the BTM deactivated. The
following table lists the error number and the inadmissible parameter. The parameter in question has
to be corrected by a service technician. Errors also occur if parameters are inconsistent. In the event
of E5000 - E5021 occuring the last two figures to the right represent the parameter number in the
parameter memory. The BTM is not servicable until all parameters are within the limits prescribed.
Error no.
E5000
E5001
E5002
E5003
E5004
E5005
E5006
E5011
E5012
E5013
E5014
E5015
E5016
E5017
E5018
E5019
E5020
E5021
E5022
E5023
E5024
E5025
E5026
E5027
E5028
E5051
E5052
E5053
E5054
E5055
E5100
Inadmissible parameter
Room temperature
Actual size of the positive phase of the recirculation bolus (°C)
Actual size of the negative phase of the recirculation bolus (°C)
Actual duration of the recirculation bolus
Maximum size of the positive recirculation bolus to be selected manually (from the
keyboard)
Minimum size of the negative recirculation bolus to be selected manually
Maximum duration of the recirculation bolus to be selected manually
Preset variation of body temperature with time (for control function)
Minimum of variation of body temperature with time to be selected manually
Maximum of variation of body temperature with time to be selected manually
Preset thermal energy flow rate (for control function)
Minimum thermal energy flow rate to be selected manually
Maximum thermal energy flow rate to be selected manually
Maximum dialysate temperature permitted during a control function
Minimum dialysate temperature permitted during a control function
Maximum rate of change of dialysate temperature with time permitted during a control
function
Maximum dialysate temperature permitted during a recirculation measurement
Minimum dialysate temperature permitted during a recirculation measurement
Calibration parameter
Calibration parameter
Calibration parameter
Calibration parameter
Calibration parameter
Calibration parameter
Bolus type for recirculation measurement without parameter input
Parameters 17, 18, 20, 21 are inconsistent
Parameters 11, 12, 13 are inconsistent
Parameters 14, 15, 16 are inconsistent
Parameters 1, 2, 4, 5 are inconsistent
Parameters 3, 6 are inconsistent
Parameters in program memory are invalid
E6xxx Software Error
Errors of type E6xxx are software errors. In the event of an error of this type occuring, the error
number should absolutely be told to the service technician. If the error has no problematic
consequences, working with the BTM can be continued.
E7xxx Termination of Recirculation Measurement
Several conditions may result in an unsuccessful termination of recirculation measurement. Previous
to bolus production, arterial and venous temperatures as well as blood flow have to be adequately
constant. If this is not attained 10 minutes after selecting the recirculation test, recirculation
measurement will break off and an error message will appear on display. Each of the three figures on
the right of the error number separately indicates a reason for the breakoff, if the figure is not zero.
63
Error no.
E7xxl,
E7xx3
E7xx2,
E7xx3
E7xlx,
E7x3x
E7x2x,
E7x3x
E71xx,
E73xx
E72xx,
E73xx
Explanation
Arterial blood temperature increases or decreases too rapidly.
Venous blood temperature increases or decreases too rapidly.
Variation of arterial blood temperature in time is too high.
Variation of venous blood temperature in time is too high.
Blood flow rate is not sufficiently constant.
Blood flow rate is too low (<120 ml/min)
E.G. error no. E7203 indicates that blood flow rate is too low and arterial as well as
venous temperature decrease or increase too rapidly.
The following error messages indicate conditions which occurred during bolus production and
bolus analysis:
Error no.
Explanation
E7601
E7602
has
E7603
Blood flow rate has been changed by user during recirculation test
Unexpected reaction of venous temperature: Instead of an expected increase a decrease
been observed.
No positive temperature bolus has been recorded within 4 minutes after adjusting dialysis
temperature.
E8xxx Termination of the Control Program
After selection of a control program the BTM first waits for both blood temperatures and the the blood
flow rate to be sufficiently constant. If this is not attained within 15 minutes after selection of the
program, the control function will terminate and an error message will be displayed. Each of the three
figures on the right of the error message separately indicates a reason for the turnoff, if the figure is
not zero.
Error no.
Explanation
E8xx1,
E8xx3
Arterial blood temperature increases or decreases too rapidly.
E8xx2,
E8xx3
Venous blood temperature increases or decreases too rapidly.
E8xlx,
E8x3x
Variation of arterial blood temperature in time is to high.
E8x2x,
E8x3x
Variation of venous blood temperature in time is to high.
E81xx,
E83xx
Blood flow rate is not sufficiently constant.
E82xx,
E83xx
Blood flow rate is too low «120 ml/min).
E.g. the error message E8231 means that blood flow rate is too low, the temporal variation of the
arterial as well a.s the venous blood temperature is too high, and the arterial blood temperature
increases or decreases to rapidly.
The remaining messages are provoked by finishing the dialysis treatment or interruptions over a
longer period of time.
E8701
E8800
64
The normal operating mode of the BTM could not be attained within 15 minutes after
starting the control program.
Problem: E.g. repeated blood alarms.
Dialysis treatment was stopped or interrupted for a longer period of time while a control
program has been active.
C3 Common Questions
In this chapter questions shall be answered and aids be given to solve problems that might rise with
the daily use of the BTM. Each problem is annotated briefly and a reference to the chapters containing
detailed explainations is given. This way tedious looking through the operating instructions shall be
avoided.
The questions and problems are classified in 3 categories. In the event of a problem occuring look
over the headings of the different categories and try to find the one that comes closest to your
problem.
Problem: The Meaning of the Displayed Information of the BTM is Unknown
.36.52
A figure with a point in front is being displayed. The point indicates that the BTM is not in
normal operation mode at present. Either normal operating mode has not been reached at
the beginning of dialysis (Chapter B4.3 - B4.5) or a blood alarm has occured
(Chapter B 10.1).
E 2203
The letter 'E' followed by any four-digit figure (except '8888') is being displayed. This way
the BTM displays a warning, a remark or an error message. General information concerning
this is given in Chapter B10.2. For information on a particular message refer to the list of all
error messages in Chapter C2.
E8888 The opposite indication appears in connection with the self-testing of the BTM immediately
after starting it (Chapter B4.1). If the indication does not extinguish after a few seconds, the
DIP switch no. 1 (Sl; LP 906) is set wrong (Chapter D3).
≡ -100
The symbol on the left followed by a figure is being displayed. The symbol indicates that the
value shown refers to a control function. Either control parameters are being entered
(Chapter B6.1, B7.1) or displayed (Chapter B 12.1).
c –13.4 The symbol on the left followed by a figure is being displayed. The symbol indicates that the
value shown refers to the recirculation measurement. Either bolus parameters are being
entered (Chapter B 11.2) or the the result of the recirculation measurement is being
displayed.
-This indication shows that the BTM is inactive (Chapter B4.3).
- - This indication appears whenever the attempt is made to call up data not available at
present with the RESULT-key (Chapter B8.2, B5.5, B12.1, B12.2).
- - - - - It is indicated that the blood lines have not yet been detected so that the respective fistula
temperature or thermal energy flow rate cannot be displayed (Chapter B4.4, B5.2).
Problem; Unknown Meaning of a Steadily Shining or Flashing LED
At first it has to be distinguished between steadily shining and fast, slowly or intermittently flashing
LEDs (definition see Chapter B2).
LED-Status
One of the 4 green LEDs shows a steady light.
The LED of the T-ART or T-VEN key is slowly Hashing.
The LED of the 'E' key is flashing fast.
The LED of the T point SET or 'E point
SET key shows a steady light
The LED of the T point SET or 'E point
SET key is slowly Hashing.
Explanation
The value of the parameter indicated
on the key is being displayed
(ChapterB5.1).
The arterial or venous sensor head
has not yet reached thermal balance
(Chapter B5.3).
While selecting a control program, the
BTM allows resetting of the energy
balance being displayed (Chapter
B6.1.B7.1).
The respective control function is
being active (Chapter B6.1,B7.1).
Parameter input for the respective
control function is expected (Chapter
B6.1, B7.1). The function is not active
yet.
65
The LED of the T point SET or 'E point
SET key is flashing fast.
The LED of the REC key shows a constant light.
The LED of the REC key is slowly flashing.
The LED of the REC key is Hashing fast.
The LED of the REC, T point SET or 'E point
SET is flashing intermittently.
No LED is on and 'Exxxx' is being
displayed (xxxx= any figure).
66
On account of a problem the
respective control function was
terminated (Chapter B6.3, B7.3).
Recirculation measurement is being
active (Chapter B8.1)
Parameter input for recirculation
measurement (Chap. Bl 1.2) is
expected.The measurement is not
active yet. If an injection bolus has
been selected, the Hashing indicates
that injection is not allowed yet
(Chapter B11.3).
The recirculation measurement is
finished.
The measuring result or an error
message is
being displayed or can be called
(Chapter B8.2).
During the respective control function
or the bolus production a dialysate
temperature limit was reached
(Chapter B 10.3).
A grave error has occured (Chapter B
10.2).
The BTM has stopped operation.
Download
Study collections