Course: Chem 271 Column Chromatography and Redox Biochemistry (Lab 3) Laura LeClerc Lab partner: Monica Treder Student I.D.: 9764763 Experiment performed: March 1st, 2011 Date Submitted: March 1st , 2011 Results Table 1: LDH activity results Fraction # Colour 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Pink/brown Pink/brown Pink/brown Pink/brown Yellow Yellow Yellow Yellow - Before Ascorbate After addition of 25 uL of 1M Ascorbate LDH activity %LDH activity (ug/min) 0% 0% 0% 0% -0.018 0.5% -3.168 89.8% -3.529 100% -0.377 10.7% -0.026 0.7% -0.0123 0.3% -0.015 0.4% - Absorbance of Cytochrome c (at = 550 nm ) 0.370 0.755 % change: 51.0 % Enzyme activity assay stopped at Fraction 10 due to time constraints Table 2: Ascorbate reduction Absorbance of FMN (at = 450 nm ) 0.682 0.665 % change: 2.5 % Sample Calculations Table 1: % LDH activity Fraction 6: 0.377*100%/3.529 = 10.7% Table 2: % change FMN: 100 – [(100*0.665)/0.682] = 2.5% Questions Question 1: One would expect molecules to elute from a gel filtration chromatography column in order of size, largest first to smallest last. This would be because in gel filtration, the beads are made with indentations and inner cavities that trap and slow the passage of smaller molecules, while larger molecules pass through unaffected by the beads. Cytochrome C has a molecular weight of 13000 g/mol (1), LDH from rabbit muscle has a molecular weight of 140000 g/mol but in a buffer pH 5 dissociates into two 70 000g/mol dimers (2), and FMN is 456.344 g/mol (3). Seeing as all three of them are proteins one can assume that these vast differences in molecular weight can translate at least vaguely to their relative sizes, so the expected order of elution would be LDH first, then Cytochrome C, then FMN. Question 2: The order in the fractions observed was LDH activity first (sample 5,) cytochrome c’s reddish-brown colors (samples 7-10) and then FMN’s yellow colors (samples 11-14). This is exactly as predicted. Question 3: Without assaying LDH activity there would have been no way to detect when the LDH eluted (which allows us to find out the order in which the proteins eluted). It was also done to detect specifically in which samples LDH was present and where it was most concentrated. Question 4: The wavelength chosen for measuring cytochrome C reduction was = 550 nm, and for FMN, = 450 nm. For FMN, = 450 nm is obviously the highest peak of absorbance with the largest difference between the oxidized and reduced molecule’s absorbance, making it optimal as a choice. For cytochrome C, = 550 nm and = 375 nm present approximately the same difference, however = 550 nm is a far better choice: this because there is also a very large peak for FMN’s difference between the oxidized and reduced molecule at = 375 nm, and this could come to interfere with the discreeter results that cytochrome c’s analysis would yield seeing as its peaks (and difference) are much smaller. This is why = 550 nm, lying outside the range of absorbance for FMN altogether, was chosen. While one could argue that there is a possibility of affecting the results with impurities from cytochrome C when measuring FMN’s change of absorbance because cytochrome C also has an absorbance difference between oxidized and reduced states at = 450 nm, this is a much smaller interference. Question 5: At = 550 nm, if Cytochrome c is reduced the absorbance should increase. It did, and by a significant margin (increased by 51.0 %) so one could safely claim that it has been reduced. At = 450 nm, if FMN is reduced the absorbance should drop. It did, but only by a small margin (2.5%) but the preparation of an ascorbate-added sample and its measurement was repeated to verify the results, and exactly the same result was yielded with unerring accuracy. One could attribute the change to a difference in the material, use of equipment, impurities or a very small reduction having been achieved. Question 6: At 400 nm, Cytochrome C’s measured absorbance would be much higher than at 550 nm (perhaps between 1-2). However there should be little to no percieved difference between the oxidized and reduced samples. At 400 nm, FMN still has a reasonable difference between the oxidized and reduced forms, however seeing as at a higher difference there was only a 2.5% drop in absorbance, the difference of absorbances percieved before and after the ascorbate was added should have been even smaller. Bibliography (1) J. Powlowski, M.J. Kornblatt, P. Joyce, J. Turnbull & Mihai Ciortea. (2010.) “Laboratory & Tutorial Manual CHEM 271”. Laboratory 4, p.87. (2) Worthington Biochemical Corporation. (2011.) <http://www.worthingtonbiochem.com/LDH/default.html>. Visited 15/03/2011. (3) European Bioinformatics Institute. (2011.) <http://www.ebi.ac.uk/msdsrv/chempdb/cgi-bin/cgi.pl?FUNCTION=getByCode&CODE=FMN>. Visited 15/03/2011.