Agilent 1260 Infinity Quaternary LC System

advertisement
Selected System Specifications - Agilent 1260 Infinity Isocratic LC System
Pump principle
Single channel
Flow rate range
0.05 – 5 mL/min at 600 bar
5 –10 mL/min at 200 bar
Maximum pressure
600 bar
Injection principle
Flow-through design, no overfill (drawn volume equals injected volume). Autosamplers are
available for vials only or for vials and micro-titer plates. Manual injector up to 600 bar
available.
Injection volumes
0.1 – 100 µL, variable in 0.1 µL steps
Column dimensions
Length: up to 300 mm, ID: 0.05 – 8 mm
Temperature range
10 degrees below ambient to 80 °C (or to 100 °C with 1290 Infinity Thermostatted Column
Compartment)
Temperature stability
< ± 0.15 °C (< ± 0.05 °C with 1290 Infinity Thermostatted Column Compartment)
Detection data rate
80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher
data rates available with 1290 Infinity detectors
UV noise level
As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)
Maintenance and system test
Front access to all maintainable parts.
Maintenance instructions through multimedia CD.
Time for full system test (OQ/PV) < 4 hours.
Standard GLP features
Early Maintenance Feedback (EMF) tracks lamp burn time, usage, number of injections with
limits and feedback messages
Extended GLP features
(for certain systems, refer to specifications of individual modules)
Data Recovery Card (DRC) for data security, preventing data loss in case of communication
breakdowns.
Radio Frequency Identification (RFID) for data traceability. RFID tags for flow cells and UV
lamp allow unambiguous traceability of measurement conditions.
System control
Through local workstation software, client/server network data system and/or local handheld
control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.
Agilent 1260 Infinity Quaternary LC System - Selected System Specifications
Pump principle
Low pressure quaternary mixing, integrated high performance degassing unit
Flow rate range
0.05 - 5 mL/min at 600 bar
5 - 10 mL/min at 200 bar
Maximum pressure
600 bar
Injection principle
Flow through design, no overfill (drawn = injected volume). Vial-only autosampler and vial
and micro-titer plate autosampler available.
Injection volumes
0.1 - 100 µL, variable in 0.1 µL steps, higher injection volumes optionally available
Column dimensions
Length: up to 300 mm, ID: 0.05 - 8 mm
Temperature range
10 °C below ambient - 80 °C
With optional 1290 Infinity Thermostatted Column Compartment up to 100 °C
Temperature stability
< ± 0.15 °C
With optional 1290 Infinity Thermostatted Column Compartment
< ± 0.05 °C
Detection data rate
80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher
data rates available with 1290 Infinity detectors
UV noise level
As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)
Maintenance and system test
Front access to all maintenance parts
Maintenance instructions through multimedia CD-ROM
Time for full system test (OQ/PV) < 4 hours
Standard GLP features
Early maintenance feedback-EMF
(tracks lamp burn time, usage, number of injections with limits and feedback messages)
Extended GLP features
(for certain systems, refer to specifications of individual modules)
Data Recovery Card (DRC) for data security (prevents data losses in case of communication
breakdowns – depends on actually chosen detector)
Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV
lamp allow unambiguous traceability of measurement conditions)
System control
Through local workstation software, client/server network data system and/or local handheld
control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.
Download