Selected System Specifications - Agilent 1260 Infinity Isocratic LC System Pump principle Single channel Flow rate range 0.05 – 5 mL/min at 600 bar 5 –10 mL/min at 200 bar Maximum pressure 600 bar Injection principle Flow-through design, no overfill (drawn volume equals injected volume). Autosamplers are available for vials only or for vials and micro-titer plates. Manual injector up to 600 bar available. Injection volumes 0.1 – 100 µL, variable in 0.1 µL steps Column dimensions Length: up to 300 mm, ID: 0.05 – 8 mm Temperature range 10 degrees below ambient to 80 °C (or to 100 °C with 1290 Infinity Thermostatted Column Compartment) Temperature stability < ± 0.15 °C (< ± 0.05 °C with 1290 Infinity Thermostatted Column Compartment) Detection data rate 80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher data rates available with 1290 Infinity detectors UV noise level As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell) Maintenance and system test Front access to all maintainable parts. Maintenance instructions through multimedia CD. Time for full system test (OQ/PV) < 4 hours. Standard GLP features Early Maintenance Feedback (EMF) tracks lamp burn time, usage, number of injections with limits and feedback messages Extended GLP features (for certain systems, refer to specifications of individual modules) Data Recovery Card (DRC) for data security, preventing data loss in case of communication breakdowns. Radio Frequency Identification (RFID) for data traceability. RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions. System control Through local workstation software, client/server network data system and/or local handheld control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN. Agilent 1260 Infinity Quaternary LC System - Selected System Specifications Pump principle Low pressure quaternary mixing, integrated high performance degassing unit Flow rate range 0.05 - 5 mL/min at 600 bar 5 - 10 mL/min at 200 bar Maximum pressure 600 bar Injection principle Flow through design, no overfill (drawn = injected volume). Vial-only autosampler and vial and micro-titer plate autosampler available. Injection volumes 0.1 - 100 µL, variable in 0.1 µL steps, higher injection volumes optionally available Column dimensions Length: up to 300 mm, ID: 0.05 - 8 mm Temperature range 10 °C below ambient - 80 °C With optional 1290 Infinity Thermostatted Column Compartment up to 100 °C Temperature stability < ± 0.15 °C With optional 1290 Infinity Thermostatted Column Compartment < ± 0.05 °C Detection data rate 80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher data rates available with 1290 Infinity detectors UV noise level As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell) Maintenance and system test Front access to all maintenance parts Maintenance instructions through multimedia CD-ROM Time for full system test (OQ/PV) < 4 hours Standard GLP features Early maintenance feedback-EMF (tracks lamp burn time, usage, number of injections with limits and feedback messages) Extended GLP features (for certain systems, refer to specifications of individual modules) Data Recovery Card (DRC) for data security (prevents data losses in case of communication breakdowns – depends on actually chosen detector) Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions) System control Through local workstation software, client/server network data system and/or local handheld control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.