Ministry of Interior of the Republic of Serbia Police Directorate Criminalistic Police Department National Forensic Center Regional Forensic Division in Novi Sad Novi Sad, 06.12.2013. CONTRACTING AUTHORITY'S CLARIFICATIONS Contract title: Procurement of laboratory equipment (GC/MS, GC-FID, HPLC) for project: Novi Sad - Szeged cross-border cooperation in development of harmonized methods and production of common database for the analysis of designer drugs Publication Ref: HUSRB/1203/214/025-TD3 No. 1. 2. Question Answer In the section "Technical specification and technical offer" for the item number 1.1 requested Ion source is "noncoated inert EI source". EI source is the part of GCMS which should be inert and that is the main characteristics of ion source. Noncoated is description which lead to one manufacturer and it should be removed. Please, provide us with technical specification without trademark of specific manufacturers. In the section "Technical specification and technical offer" for the item number 1.1 adjustable Ionization energy in the range 5-240eV has been requested. Ionization energy which is in common use for almost all application is in range of 60-l50eV. It is known that electrons with energy lower than 10eV have difficulties to ionizing. When it is needed CI (Chemical ionization) mode is used for this soft ionization. It is not needed to have this low range below 10eV. Please, provide us with technical specification with energy range above 10eV. 1 New product specifications are published on the website of the Ministry of Interior of the Republic of Serbia: http://www.mup.gov.rs/cms_cir/oglasi.nsf/ja vne_nabavke.h as part of CORRIGENDUM No.1. to the TENDER DOSSIER New product specifications are published on the website of the Ministry of Interior of the Republic of Serbia: http://www.mup.gov.rs/cms_cir/oglasi.nsf/ja vne_nabavke.h as part of CORRIGENDUM No.1. to the TENDER DOSSIER Ministry of Interior of the Republic of Serbia Police Directorate Criminalistic Police Department National Forensic Center Regional Forensic Division in Novi Sad 3. 4. 5. In the section "Technical specification and technical offer" for the item number 1.1 has been requested "Ionization current 0-3I5 uAmp". Ionization current is trademark of every manufacturer, it depends from model. This range is not applicable in work. Nobody will use 0uA for any kind of analysis it is not applicable. Also, in technical specification for every instrument it is shown what is the best current range for that model. That is the reason way you have technical specification for every model and that's why you have instrument qualification after installation with service engineers to show you how to use this model and how to make its best performance. Please, provide us with technical specification for ionization current that is in range of 5-250uA and more. In the section "Technical specification and technical offer" for the item number 1.1 requested Quadrupole temperature is up to 200ºC? As we describe above in previous question, every technical specification is different for every manufacturer and it is trademark for every model. That's why Agilent heat quadruple and Shimadzu does not. Both have the best performance but different technical solution. With this constraint of specification it is for you to decide with solution you like best but we would appreciate if you consider different technical construction of Agilent since sensitivity and versatility is the same even better. Please provide us to participate with different technical solution of requested since there is no leak of sensitivity and versatility (without heating quadruple). In the section "Technical specification and technical offer" for the item number 2.2, has been requested "full control of all CC and ALS parameters”. ALS is short name for Automatic Liquid Sampler and it is Agilent trademark and shortcut. Since it is not applicable for anyone else but Agilent, please provide us with technical specification without this particular short name. 2 New product specifications are published on the website of the Ministry of Interior of the Republic of Serbia: http://www.mup.gov.rs/cms_cir/oglasi.nsf/ja vne_nabavke.h as part of CORRIGENDUM No.1. to the TENDER DOSSIER GCMS analysis of complex matrix require heated ion path (from ion source to the detector) in order to reduce contaminations and instrument down time, to improve sensitivity and repeatability. Instrument with independently heated ion source and mass filter have possibility of thermal cleaning, which reduce instrument down time and increase instrument lifetime. New product specifications are published on the website of the Ministry of Interior of the Republic of Serbia: http://www.mup.gov.rs/cms_cir/oglasi.nsf/ja vne_nabavke.h as part of CORRIGENDUM No.1. to the TENDER DOSSIER "ALS" is abbreviation for Automatic Liquid Sampler. Software for instrument control must control parameters of GC and Automatic Liquid Sampler. It is not a trademark of any manufacturer. Every device capable of injecting liquid sample into gas chromatograph can be considered as ALS. Ministry of Interior of the Republic of Serbia Police Directorate Criminalistic Police Department National Forensic Center Regional Forensic Division in Novi Sad Same is valid for abbreviations GC (gas chromatograph), MS (mass spectrometer) or MSD (mass spectrometric detector) or SQ (single quadrupole). Full controll of all components of the instrument by software is requested, including "ALS". ANNEXES: -CORRIGENDUM No.1 to the TENDER DOSSIER -ANNEX II + III : TECHNICAL SPECIFICATIONS + TECHNICAL OFFER 3