1 SUPPLEMENTARY MATERIALS 2 Figure S1 can be found at DOI: xxx-xxxx.s1. 3 Figure S2 can be found at DOI: xxx-xxxx.s2. 4 Table S1 can be found at DOI: xxx-xxxx.ts1. 5 Table S2 can be found at DOI: xxx-xxxx.ts2. 6 7 Table S1. Total Irradiance of Light Source (250-900 nm) and the UV Component 8 (250-400 nm) and % of total measured in all three light intensities (1,000, 2,000 and 9 4,000 Lux) under all conditions (light, light with UV filter, light with borosilicate tube 10 and no light). Light Intensity (lux) 4000 4000 4000 2000 2000 2000 Condition light source light source with UV cut off filter light source with borosilicate tube light source light source with UV cut off filter light source with borosilicate tube Irradiance Irradiance 250 – 900 250 – 400 2 nm (μW/cm ) nm (μW/cm2) % of total light source output 8841.9067 9.4836 0.1073 % % of UV portion of light source output 100 % 7780.8779 1.1380 0.0129 % 12.000 % 8016.6621 8.5984 0.0972 % 90.6667 % 7780.8779 3.7783 0.0486 % 100 % 6994.9306 0.3817 0.0049 % 10.1010 % 7309.309545 3.5494 0.0456 % 93.9394 % IRBP protects retinol from photodegradation 1000 1000 1000 0 light source light source with UV cut off filter light source with borosilicate tube no light source 5894.604472 1.4517 0.0246 % 100 % 5501.630841 0.0968 0.0016 % 6.6667 % 5501.630841 1.3550 0.0230 % 93.3333 % 0.0747 0.0012 0.0008 % 0.0127 % 11 12 Table S2. Recorded Temperature (at the test samples inside the test tube; with and 13 without light source at 4,000 Lux or 8,842 W/cm2) at time 0, 10, 20, 30 min of 14 experiment. Time (min) 15 16 Sample 0 10 20 30 60 1 (4000 Lux) 22 22 22 22 22 2 (4000 Lux) 22 22 22 22 22 3 (4000 Lux) 22 22 22 22 22 4 (dark) 22 22 22 22 22 5 (dark) 22 22 22 22 22 6 (dark) 22 22 22 22 22 17 18 Figure S1. UV Emission Spectra of Light Source (A) and Transmission Profile of 19 FGL400 UV filter (B). Total irradiance of light source is 8,842 W/cm2 from 250-900 20 nm (light intensity: 4,000Lux). Emission spectra for the light source were first obtained 21 with a PTI QuantaMaster 500 spectrophotometer. Fig.S1A shows spectra taken under 22 conditions with the light bulb (blue line; 4,000 Lux or 8,842 μW/cm 2, 250-400 nm), with the 23 light bulb and an UV short pass filter (red ), and with the light bulb and borosilicate test 2 IRBP protects retinol from photodegradation 24 tubes (green), and with the light bulb turned off for a background spectrum (black ). 25 Based on the manufacturer’s specification, the UV shortpass filter transmits minimal light 26 with a wavelength shorter than 400 nm but transmits nearly 90% in the visible and IR 27 (Thorlabs FGL400, see transmittance profile in Fig. S1B). A borosilicate glass tube (used 28 for holding the samples in experiments in the present study) was used to evaluate how 29 much of the emitted light was absorbed by the borosilicate test tube. In both cases, the 30 UV filter or the test tube was placed directly in front of the entrance slit to the 31 monochromator at the same distance from the light source as the original samples. To 32 derive power densities, spectra were first background corrected (black line). Total power 33 density was measured using a SpectraPhysics power meter to obtain power in Watts and 34 dividing by the sensor area. Power densities were measured under the same conditions 35 (all four lines in Fig. S1). By integrating under the entire curve for condition (red line) and 36 correlating that to 100% of the power density, the corresponding power densities for 37 spectral regions can be obtained by a simple ratio of the area of interest (250 – 400 nm) 38 to the total area (250 – 900). In this way, the power densities for UV range (250-400 nm) 39 were calculated using the data for all conditions. This experiment was repeated for all 40 four light intensities used for our retinol degradation study. All power densities and % 41 light source are reported in Table S1. 42 Figure S2. Reversal of Retinol Photodegradation by UV Filter and UV/VIS Filter. 43 Percent change in retinol concentration at time 0 and 30 min by indicated 44 treatments are presented in (A). Transmission profile of UV-VIS filter FEL1000 is 45 shown in (B) [Note: See Fig. S1 B for transmission profile of UV filter ]. To examine 46 the effect of UV light on retinol degradation,2, experiments were performed using a short 3 IRBP protects retinol from photodegradation 47 pass UV filter (to minimize UV light; see Fig. S1 B for transmittance profile of FGL400 48 from Thorlabs) and a long pass UV/VIS filter (to minimize UV/VIS light; see Fig. S2 B for 49 transmittance profile of FEL1000 from Thorlabs). To accommodate these light filters, a 50 sample chamber was connected to the light chambers by a short circular cylinder (with a 51 2 inch diameter). 52 Retinol samples were exposed to light intensity of 4,000 Lux (or 8,842 μW/cm 2; measured 53 at the sample) for 30 min with and without filter. Changes in retinol level were measured 54 by a spectrophotometer (at 325 nm) and also by HPLC. Fig. S2 shows that light 55 exposure for 30 min significantly reduced retinol concentration (from 100% to 25%). The 56 removal of UV light (by the FGL400 filter) resulted in an almost complete reversal of this 57 degradation (i.e. returned to 81% from 25%) suggesting that the UV component of the 58 light source was mainly responsible for retinol degradation. Furthermore, the exclusion of 59 both UV and VIS source (by a UV/VIS filter FEL1000) did not significantly change this 60 level of reversal of retinol degradation, suggesting that light from the VIS range did not 61 contribute a significant role in retinol degradation (n=3, mean and standard errors are 62 show in the figure). 63 64 65 66 67 4 IRBP protects retinol from photodegradation 68 Figure S1 69 70 71 5 IRBP protects retinol from photodegradation 72 Figure S2 120 100 % ATOL 80 60 40 20 0 TO DARK 73 74 T30 DARK T30 LIGHT at 4,000 lux A B 6 T30 LIGHT at 4,000 lux + UV Filter T30 LIGHT at 4,000 lux + UV/VIS Filter