“RP-HPLC METHOD FOR THE SIMULTANEOUS ESTIMATION OF MOXONIDINE AND HYDROCHLOROTHIAZIDE IN BULK DRUG AND TABLET FORMULATION” SHILPA KORTI*, CHANNABASAVARAJ K P, SOMASHEKAR P L Department of Pharmaceutical Chemistry, Government College of Pharmacy, Bengaluru, Karnataka, India. Address for correspondence * Contact no- 09980676409 Email ID- Ksuhaani@gmail.com ABSTRACT A new, simple, sensitive, selective, precise, rapid, economic and isocratic reverse phase high performance liquid chromatography (RP-HPLC) method was developed and validated for simultaneous estimation of Moxonidine (MOX) and Hydrochlorothiazide (HCTZ) in bulk drug and tablet dosage forms. The separation was achieved by using C8 Phenomenex Luna (250 x 4.6mm, 5µm) column with a mobile phase consisting of acetonitrile and formic acid solution (0.2%v/v) in the ratio 50:50. The mobile phase was delivered at a flow rate of 0.8 mL/min. Analysis was performed at ambient temperature with detection wavelength at 245 nm. The retention times of MOX and HCTZ were found to be 3.0 and 4.8 min and the calibration curves were linear (r2=0.999) over a concentration range from 1-35µg/mL for MOX and HCTZ respectively. Limit of detection (LOD) and Limit of quantitation (LOQ) were 0.08µg/mL and 0.1µg/mL for MOX and 0.2µg/mL and 0.4µg/mL for HCTZ respectively. The developed method was validated for parameters like system suitability, specificity, linearity, accuracy, precision, ruggedness and robustness as per ICH guidelines and the results were found to be within the limits. So it can be used for the routine quality control of MOX and HCTZ in bulk sample and tablet dosage forms. KEY WORDS: Moxonidine, Hydrochlorothiazide, RP-HPLC. INTRODUCTION Moxonidine, chemically 4-Chloro-N-(imidazolidin-2-ylidene)-6-methoxy-2 methylpyrimidin-5-amine. It has empirical formula C9H12ClO5 shown in Fig 1. Moxonidine is a new generation centrally acting antihypertensive drug used for the treatment of mild to moderate essential hypertension. It is a selective agonist at the imidazoline receptor subtype 1 (I1). This receptor subtype is found in both the rostral ventro-lateral pressor and ventromedial depressor areas of the medulla oblongata. Moxonidine causes a decrease in sympathetic nervous system activity and therefore a decrease in blood pressure. [1, 2] Hydrochlorothiazide, chemically 6-chloro3,4-dihydro-2H-1,2,4-benzothiadiazine-7-sulphonamide1,1-dioxide. It has molecular formula C7H8ClN3O4S2 shown in Fig 2. Hydrochlorothiazide belongs to the thiazide class of diuretics. It reduces blood volume by acting on the kidneys to reduce sodium (Na+) reabsorption in the distal convoluted tubule.[3,4] The combination of Moxonidine (2mg) and Hydrochlorothiazide (12.5mg) is widely used to treat the patients suffering from essential hypertention. This combination also permits effective antihypertensive treatment over a longer period. On literature survey it was found that several analytical methods like Ultra Violet Spectrometer (UV), [5,6,7,8,9,10,11] High Performance Thin Liquid Chromatography (HPTLC), [12,13] High Performance Liquid Chromatography (HPLC), [14] Reverse Phase High performance Liquid Chromatography (RP-HPLC), [15,16,17,18,19,20] Ultra Performace Liquid Chromatography (UPLC), [21,22,23] Liquid Chromatography- ElectroSpray Ionisation - Mass Spectroscopy (LC-ESI-MS) [24,25,26,27,28] have been reported for the determination and estimation of Moxonidine and Hydrochlorothiazide individually. The extensive literature review revealed that no analytical methods were reported for simultaneous estimation of Moxonidine and Hydrochlorothiazide in pharmaceutical formulations. Hence new modified method was developed for simultaneous estimation of Moxonidine and Hydrochlorothiazide by RP-HPLC in pharmaceutical dosage forms. The proposed method was validated as per ICH guidelines. [29, 30] HN N HN H3C O Cl H N S S Cl N N CH3 Fig. 1. Structure of Moxonidine H2N O NH O O O Fig.2. Structure of Hydrochlorothiazide MATERIALS AND METHODS Chemicals and Reagents: HPLC grades, acetonitrile and all other analytical grade reagents were purchased from Rankem and Loba Chemie India. HPLC grade water was prepared using Milli-Q water purification system. Class A glassware is used throughout the experiment. Reference Standard MOX and HCTZ are gift samples were obtained from Micro labs, Bengaluru. MOXAVAS-H, Macleodes tablets were purchased from local market. Instrument & Chromatographic conditions: The chromatographic system consists of LC-20AT SHIMADZU -SPD-20A equipped with UV detector and Autosampler injector with 20 µl loop volume. Chromatographic separations were carried out using C8 Phenomenex Luna column (250x4.6mm, 5µ). The mobile phase consisting of acetonitrile and formic acid solution (0.2%v/v) in the ratio 50:50. The flow rate was 0.8mL/min and detection was carried out at 245nm using UV detector. Preparation of standard stock solutions of MOX and HCTZ: Accurately 10 mg of MOX and HCTZ were weighed separately and transferred into a clean and dry 10 ml volumetric flask, dissolved with sufficient volume of diluent consists of ACN and Water (50:50). The volume was made up to 10 ml with diluent to get a concentration of 1000 µg/ml (Stock I). 1 ml of resulting stock solution was further diluted to 10 ml volumetric flask with diluent to get a concentration of 100µg/ml for both MOX and HCTZ respectively (Stock II). From stock II, 1 ml was transferred into a 10 ml volumetric flask and final volume was then made up to 10 ml with diluent to get a concentration of 10µg/ml of MOX and HCTZ respectively. Preparation of sample stock solution: Twenty tablets of MOXAVAS-H each containing 0.2 mg of MOX and 12.5 mg of HCTZ were weighed and finely powered. Powder equivalent to 0.2 mg of MOX and 12.5 mg of HCTZ were taken together and transferred into a clean, dry 100 ml volumetric flask. The powder was first dissolved in diluent and sonicated for 10 mins. The resulting mixture was then filtered through whatmann filter no 0.45μ. The final volume of filtrate was made up to 100 ml with diluent. 1ml of the resulting solution was diluted to 10ml with diluent to get a concentration of 2µg/ml and 125µg/ml of MOX and HCTZ respectively. METHOD DEVELOPMENT A variety of mobile phases were investigated in the development of a RP-HPLC method for the analysis of MOX and HCTZ. A mixture of Acetonitrile and 0.2% v/v Formic acid solution (50:50v/v) was found to be the most suitable mobile phase for ideal separation of MOX and HCTZ. The solvent mixture was filtered through a 0.45 μ whatmann filter paper and sonicated before use. It was pumped through the column at a flow rate of 0.8 mL/min. The column was maintained at an ambient temperature. The detection of the drug was monitored at 245 nm. Under these optimized chromatographic conditions the retention time obtained for the drugs MOX and HCTZ was 3.0 min and 4.8 min respectively. A typical chromatogram showing the separation of the drugs is as shown in Fig: 3. Fig: 3 Chromatogram of Moxonidine and Hydrochlorothiazide METHOD VALIDATION The validation of developed method was done as per ICH guidelines [29, 30] which include Linearity, LOD and LOQ System suitability, Precision (system precision, method precision), Accuracy, Ruggedness, Robustness (flow rate, wavelength, and mobile phase ratio) and Assay. Linearity: Linearity is the ability of the method to produce results that is directly proportional to the concentration of the analyte in samples with given range. The various concentration of working standard solutions of MOX and HCTZ were made by pipetting 0.1-3.5 ml from stock (II) into a series of 10 ml volumetric flask and diluted to 10ml with diluent to get the final concentration of 1 to 35µg/ml solutions. 20 µl of each of these working standard solutions of MOX and HCTZ ranging from 1 to 35 µg/ml were injected into a chromatograph at flow rate of 0.8 mL/min. Retention time and peak area obtained were recorded and standard calibration curve was plotted for MOX and HCTZ and linearity equations were derived. The calibration curves were shown in Fig: 4 and Table 1 Sl. No 1 2 3 4 5 6 7 8 9 10 11 12 13 Table 1: Linearity data for MOX and HCTZ Volume of MOX Final Peak and HCTZ Concentration Volume area* for working standard (µg/ml) (ml) MOX solution(ml) 36215 0.1 10 1 0.2 10 2 65703 129582 0.4 10 4 194637 0.6 10 6 253032 0.8 10 8 320829 1.0 10 10 383460 1.2 10 12 1.4 1.5 2.0 2.5 3.0 3.5 10 10 10 10 10 10 Peak area* for HCTZ 12249 24925 46927 73248 94323 120017 146652 449716 172148 14 15 483285 184672 628706 245927 20 777387 293683 25 937368 358572 30 1072093 414494 35 *Average of five determinations 1200000 MOX y = 30719x + 10857 R² = 0.9996 1000000 HCTZ y = 11873x + 2076.7 R² = 0.9993 Area 800000 600000 400000 200000 0 0 5 10 15 20 25 30 35 40 Concentration µg/ml Fig: 4 Linearity plot for Moxonidine and Hydrochlorothiazide Limit of Detection and limit of Quantification: The limit of detection (LOD) is the smallest concentration that can be detected but not necessarily quantified as an exact value. The limit of quantitation (LOQ) is the lowest amount of analyte in the sample that can be quantitatively determined with precision and accuracy. For estimation of LOD, LOQ and visualization method was followed. In visualization method lower dilutions of the standard drugs of MOX and HCTZ were successively prepared, injected in to the chromatograph and response obtained was recorded and presented in Table 1. Table: 2 Result of LOD and LOQ Drug MOX HCTZ LOD (µg/ml) 0.08 0.2 LOQ (µg/ml) 0.1 0.4 System suitability parameter: A Standard solution of MOX and HCTZ working standard was prepared as per procedure and was injected six times into the HPLC system. The system suitability parameters were evaluated from standard Chromatograms obtained by calculating the % RSD of retention times, tailing factor, theoretical plates and peak areas from six replicate injections. From the system suitability studies it was observed that all the parameters were within limit. Hence it was concluded that the Instrument, Reagents and Column were suitable to perform the assay. The Fig: 3 and Table 3 shows the results. Table: 3 System Suitability Parameters System Suitability Factor MOX HCTZ Acceptance Criteria Theoretical plates 3044 6603 >2000 HETP (mm) Tailing factor 49 1.2 23 1.2 <2 Resolution 7.83 >2 Specificity: Specificity is a procedure to detect quantitatively the analyte in presence of component that may be expected to be present in the sample matrix. Commonly used excipients in tablet preparation were spiked in a pre-weighed quantity of drugs and then area was measured and calculations done to determine the quantity of the drugs. Precision: The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. System precision: Successive six injections of standard solution (six replicates) were injected into a HPLC chromatograph, the peak area and chromatograms obtained were recorded. The % relative standard deviation was calculated for peak areas of replicates. Method precision: Intraday precision Intraday precision of test method is demonstrated by 5 injections of the same batch (same conc) of samples at 0, 1, 2, 3, 4, 5, 6 and 12hrs on same day. Inter day Precision Inter day precision of test method is demonstrated by 5 injections of the same batch (same conc) of samples on 3 successive days. The amounts of MOX and HCTZ in sample were calculated from their respective standard linearity equations. The results obtained were presented in Table 4-8 Table: 4 System Precision Data for MOX and HCTZ Replicates MOX Peak Area* HCTZ Peak Area* 1 311076 119728 2 310856 121910 3 320829 120017 4 310933 119188 5 316886 123317 6 321166 119470 315291 120605 Average 4983.11 1641.28 Standard Deviation 1.58 1.36 %RSD Table: 5 Method Precision Data for MOX and HCTZ Replicates MOX Concentration Peak Area* (µg/ml) HCTZ Concentration Peak Area* (µg/ml) 1 320886 10.09 121817 10.09 2 3 4 ;5 6 320163 316575 320876 319933 320651 10.07 9.95 10.09 10.06 10..08 10.06 0.05 0.53 120178 119470 119188 121820 120986 9.95 9.89 9.86 10.09 10.02 9.98 0.10 0.97 Mean Standard Deviation %RSD Table: 6 Intraday Precision Data of MOX and HCTZ Replicate 1 2 3 4 5 6 7 Time Interval MOX Peak Area* Conc. (μg/ml) 1hr 320642 10.08 2hr 310946 9.77 3hr 320184 10.07 4hr 320817 10.09 5hr 320676 10.09 6hr 319945 10.06 12hr 318966 10.03 10.02 Mean 0.12 Standard Deviation 1.24 % RSD HCTZ Peak Area * Conc. (μg/ml) 120953 10.01 119467 9.89 119152 9.86 119698 9.91 121900 10.09 121836 10.09 120477 9.97 9.97 0.10 1.02 Table: 7 Interday Precision Data of MOX and HCTZ Replicate 1 2 3 Date Interval Day 1 Day 2 Day 3 MOX Peak Area* Conc. (μg/ml) 320765 10.09 318958 10.03 319844 10.06 HCTZ Peak Area* Conc. (μg/ml) 121260 10.04 120810 10.00 119880 9.92 Mean 10.06 9.99 Standard Deviation % RSD 0.03 0.29 0.06 0.59 Table: 8 Results of Precision parameters Precision Parameters System Precision MOX % RSD 1.58 HCTZ % RSD 1.36 Acceptance Criteria < 2.0% Method Precision 0.53 0.97 < 2.0% Intra day Inter day 1.24 0.29 1.02 0.59 < 2.0% < 2.0% Ruggedness: Ruggedness was determined by the assay of sample sets by different analysts, the peak area and the chromatograms were recorded. The percentage assay of standard drug was calculated from the peak areas of replicates. The results obtained were mentioned in Table 9-11. Table: 9 Intermediate Precision Data of Analyst 1 MOX HCTZ Replicates Peak Area* %Assay Peak Area* %Assay 319666 100.53 120280 99.56 1 318998 100.31 120781 99.98 2 319886 100.60 119964 99.29 3 320895 100.93 119188 98.64 4 320157 100.69 121317 100.04 5 317813 99.92 119778 99.13 6 100.50 99.50 Mean Table: 10 Intermediate precision data Analyst 2 MOX HCTZ Replicates Peak Area* %Assay Peak Area* %Assay 1 2 3 4 5 6 316574 320854 310878 320719 319058 319847 99.52 100.91 97.67 100.87 100.33 100.59 99.98 120840 119823 121498 119994 119979 123116 100,03 99.17 100.58 99.32 99.30 101.95 100.06 Mean Table: 11 Results of Ruggedness Drug Analyst 1 Analyst 2 Acceptance Criteria MOX 100.50% 99.96% 90-110% HCTZ 99.50% 100.06% 90-110% Accuracy: Accuracy is expressed as percentage recovery by the assay of known added amounts of analyte carried out at three different levels (80%, 100%, and 120%). Excellent recovery and low relative standard deviation value showed that the method was suitably accurate for potency assay of Moxonidine and Hydrochlorothiazide simultaneously in the drug substances. The results were tabulated in the Table 12-13. Table: 12 Results of Accuracy of MOX Level 80% 100% 120% Stand Conc. (µg/ml) Sample Conc. (µg/ml) Peak Area* Total Conc. (µg/ml) I II 8 2 312891 9.83 Amt of standard recovered (µg/ml) 7.83 8 2 311678 9.79 7.79 97.41 III I 8 10 2 2 320072 378432 10.07 11.97 8.07 9.97 100.82 99.66 II III I II III 10 10 12 12 12 2 2 2 2 2 367854 377987 425767 430987 430871 11.62 11.95 13.51 13.68 13.67 9.62 9.95 11.51 11.68 11.67 96.21 99.51 95.89 97.30 97.27 Replicate % Recovery 97.90 Table 13: Recovery study data for HCTZ Replicate Level 80% 100% 120% Stand Sample Conc. Conc. (µg/ml) (µg/ml) I 8 II III I II III I II III 8 8 10 10 10 12 12 12 125 125 125 125 125 125 125 125 125 Peak Area* Total Conc. (µg/ml) 1580106 132.91 Amt of standard recovere d (µg/ml) 7.91 1580278 1579912 1600446 1601566 1602871 1626975 1627766 1625962 132.92 132.89 134.62 134.72 134.83 136.86 136.92 136.77 7.92 7.89 9.62 9.72 9.83 11.86 11.92 11.77 % Recover y 98.86 99.04 98.66 96.22 97.17 98.26 98.80 99.36 98.09 Robustness: The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. Robustness of the method was investigated under a variety of conditions like change in flow rate (± 0.1 ml/min), wavelength (± 2 nm) and mobile phase composition (± 5 from absolute value). % assay for all the robustness studies is as shown in Table 14. Table: 14 Results of Robustness Parameters Area Change in flow rate % Assay of MOX 104.51 Area 117993 % Assay of HCTZ 104.38 0.7ml 331888 0.8ml 320895 100.93 121900 100.92 0.9ml 309524 97.23 119059 96.05 Mobile phase ratio (ACN: Buffer) (55:45) (50:50) (45:55) 304913 320092 302891 95.72 100.67 95.07 114914 120900 115981 95.04 100.92 96.05 Wavelength in nm 243 314957 96.56 117993 97.63 245 247 320895 310524 100.93 97.55 121900 119059 100.92 98.53 Assay: To provide an exact result which allows an accurate statement on the content or potency of the analyte in a sample. The proposed method was also evaluated by the assay of MOX and HCTZ in their combined dosage formulation. The results were mentioned in the Table 15. Table: 15 Assay results of Tablet formulation Peak Peak area*for area*for MOX HCTZ 71962 1449741 Amount of drug recovered MOX (µg/ml) Amount of Drug recovered HCTZ (µg/ml) % Assay for MOX % Assay for HCTZ 1.99 121.93 99.46 97.54 RESULTS AND DISCUSSION Literature review reveals only few UV and HPLC methods have been reported for the estimation of MOX and HCTZ alone in bulk and tablet dosage formulation till so far. So, new simple, rapid and accurate RP-HPLC method has been developed and validated for the simultaneous estimation of MOX and HCTZ in bulk and tablet dosage formulation. The separation was carried out by using C8 Phenomenex Luna (250 x 4.6mm, 5µm) column with a mobile phase consisting of acetonitrile: buffer (0.2%v/v formic acid solution) in the ratio 50:50v/v. The mobile phase was delivered at a flow rate of 0.8 ml/min with detection at 245 nm. The retention times of MOX and HCTZ were found to be 3.0 and 4.8 min and the calibration curves were linear (R2=0.999) over a concentration range from 1-35µg/ml for MOX and HCTZ respectively. Limit of detection (LOD) and Limit of quantitation (LOQ) were 0.08µg/ml and 0.1µg/ml for MOX and 0.2µg/ml and 0.4µg/ml for HCTZ respectively. The developed method was validated for parameters like system suitability, specificity, linearity, accuracy, precision, ruggedness and robustness as per ICH guidelines and the results were found to be within the limits. CONCLUSION The reverse phase high performance liquid chromatographic method for the determination of Moxonidine and Hydrochlorothiazide in bulk drug and tablet dosage form was found to be accurate and precise. The elution of the analytes by thus developed and validated isocratic method is shorter than the gradient method reported so far. Thus, the proposed RP-HPLC method can be successfully applied for the routine quality control analysis of Moxonidine and Hydrochlorothiazide in bulk and tablet dosage form. ACKNOWLEDGMENT: The authors are thankful to Micro Labs, Bengaluru, for providing gift samples of MOX and HCTZ for the research work. REFERENCES: 1. European Pharmacopoeia. European Pharmacopoeia Commission, 5th ed: 2064. 2. British Pharmacopoeia. Monographs Medicinal and Pharmaceutical Substances.2009; I & II: 4057. 3. The Merck Index, Merck and Co., Inc. 2006; 14th ed: 4785. 4. Indian Pharmacopoeia. The Indian Pharmacopoeia Commission, Ghaziabad 2007; 2:576. 5. Hapse SA, Wagh.VS, Kadaskar PT, Dokhe MD Shirsath AS. Spectrophotometric estimation and validation of Hydrochlorothiazide in tablet dosage forms by using different solvents. Der Pharma Chemica 2012; 4 (1):10-14. 6. Reema J, Pinak P. Development and validation of first order derivative method for simultaneous estimation of Quinapril Hcl and Hydrochlorothiazide in combined pharmaceutical formulation. Int J Pharm Res Sci 2013; 1(1): 16-25. 7. Mallikarjuna Rao N, Gowri Sankar D. Simultaneous estimation of s (-) Amlodipine and Hydrochlorothiazide in bulk and tablet dosage form by simultaneous equation method. Int J Pharm Pharmaceutical Sci Res 2011; 1: (1) 15. 8. Ilango K, Shiji K. Spectrophotometric methods for simultaneous determination of Telmisartan and Hydrochlorothiazide in tablet dosage form. Res J Pharm Bio Chem Sci 2012 Oct-Dec; 3(4):1438-45. 9. Deshpande MM, Mahajan MP, Sawant SD. Simultaneous estimation of Valsartan and Hydrochlorothiazide in fixed dose combination in UV spectrophotometry. Int J Pharm Sci Res 2012; 3(1): 236-40. 10. Rekha G, Sunil K, Paras S. Spectrophotometric simultaneous determination of Hydrochlorothiazide and Telmisartan in combined dosage form. J Applied Pharm Sci 01 (01); 2011: 46-49. 11. Meyyanathan SN, Arunadevi SB Bhojraj S. Simultaneous estimation of Nebivolol Hydrochloride and Valsartan and Nebivolol Hydrochloride and Hydrochlorothiazide in pharmaceutical formulations by UV spectrophotometric methods. Indian J Pharm Edu Res 2010 Apr-Jun; 44(2):156-9. 12. Rajendra K, Kamlesh G, Obaid Qureshi M. Stability indicating HPTLC Method for determination of Moxonidine in Pharmaceutical Preparations. Int J PharmTech Res 2012 Jan-Mar; 4(1):358-63. 13. Shah NJ, Suhagia BN, Shah RR, Patel NM. HPTLC method for the simultaneous estimation of Valsartan and Hydrochlorothiazide in Tablet Dosage Form. Indian J Pharm Sci 2009 Jan-Feb; 71(1): 72-4. 14. Balamuralikrishna K, Syamasundar B. Development and validation of high performance liquid chromatographic method for the simultaneous estimation of Candesartan cilexetil and Hydrochlorothiazide in combined tablet dosage form. Der Pharma Chemica 2010; 2(6):231-7. 15. Svetlana M, Biljana O, Mira Z, Ljiljana Z, Ana P. Development and validation of reversed phase high performance liquid chromatographic method for determination of Moxonidine in the presence of its impurities. J Pharm Biomed Anal 2012 Feb 5; 59:151-6. 16. Maitreyi Z, Bhavita D, Jemish P, Viral B, Amit K. Simultaneous estimation of Olmesartan medoxomoil and Hydrochlorothiazide by validated reversed phase high performance liquid chromatography. Int J Pharm Res Bio Sci 2012; 1(1). 17. Padmini LV, Firdose SK, Madhavi S, Javed Mohammed M and Buchi NN. Simultaneous estimation of Lisinopril and Hydrochlorothiazide in bulk, pharmaceutical dosage forms and in dissolution samples by RP-HPLC-PDA method. J Chem Pharm Res 2013; 5(8): 195-202. 18. Malavika L. Vishwakarma, Uttam D P, Satish G P, Kiran V. Mangaonkara. A rapid method for the simultaneous estimation of Amlodipine, Hydrochlorothiazide and Losartan by RP-HPLC. Global J Anal Chem 2010 Jun-Jul; 1-7. 19. Vania M, Lily P, Boyka T. RP-HPLC method for the simultaneous determination of Lisinipril and Hydrochlorothiazide in pharmaceutical formulation. Int J Pharm Sci Rev Res 2013 Sep-Oct; 22(1):253-6. 20. Naman D, Avani S, Patel CN. Validated reverse phase high performance liquid chromatographic method for simultaneous estimation of Olmesartan medoxomil, Hydrochlorothiazide and Amlodipine besylate in newly designed pharmaceutical dosage form. Int J Res Pharm 2012; 3(11):178-182. 21. Antil P, Kaushik D, Jain G, Srinivas KS, Indu Thakur. UPLC Method for Simultaneous Determination of Valsartan & Hydrochlorothiazide in Drug Products. J Chromat Seperation Techniq 2013; 4(5):1-5. 22. Vanitha C, Bindu P, Alagar RM, David B, Pradeep KP.Development and validation UPLC method for simultaneous determination of Hydrochlorothiazide and Eprosartan. J Adv Pharm Edu Res 2013 Jul-Sep; 3(3):171-175. 23. Bheru S, Malviya, Dr. Deepti J, Prince S, Pramod H, Vidhi J. Development and validation of UPLC method for simultaneous determination of Irbesartan, Hydrochlorothiazide, Amlodipine Besylate and Atorvastatin. Int J Drug Discovery Herbal Res 2012 Oct-Dec; 2(4):481-6. 24. Arjun S, Dr.Sivakumar T, Dr.Jain DA, Umesh M, Arun KY, Tarun S. High throughput liquid chromatography ionspray tandem mass spectrometry (LC-ESIMS/MS) method development and validation for quantification of Hydrochlorothiazide and Bisoprolol in human plasma. Int J Pharm Res Dev 2011 May; 3(3): 11 – 22. 25. Rudolph M, Janssen W, Strassner M. Determination of Moxonidine in plasma by gas chromatography-negative ion chemical ionization mass spectrometry. J Pharm Biomed Anal 1992 May; 10(5):323-8. 26. Lara FT, Maha FT, Mamoun IA, Manal HK, Adi IA. Simultaneous determination of Irbesartan and Hydrochlorothiazide in plasma using HPLC coupled with tandem mass spectrometry: Application to bioequivalence studies. J Pharm Bio Anal 2010 March 11; 51(4):985-90. 27. Luhua Z, Li D, Xin W. Determination of Moxonidine in human plasma by liquid chromatography-electrospray ionisation-mass spectrometry. J Pharm Biomed Anal 2006 Jan 2013; 40 (1):95-9. 28. Ariadni V, Michael K, Irene P. Development and validation of a liquid chromatographic/electrospray ionization mass spectrometric method for the determination of Benazepril, Benazeprilat and Hydrochlorothiazide in human plasma. Journal of Mass Spectrometry 2006 May; 41(5): 593–605. 29. ICH guidance, validation of analytical methods: definition and terminology. International Conference on Harmonization, Q2A: Geneva; Nov, 2005. 30. ICH guidance, validation of analytical procedures: methodology. International Conference on Harmonization, Q2B: Geneva; Nov, 2005.