Supplimentary Material for Manuscript (MS #L14-09947) “A Comparison between Bipolar Transistor and Nanowire FET Biosensors” by Sufi Zafar et al. Fig. s1: Measured collector current (IC) and base current (IB) as a function of applied voltage. (A) Standard HBT device with no solution, reference electrode and TiN sensing surface. (B) HBT sensor measured in pH 6 solution of 10 mM and 100 mM concentrations. (C) HBT sensor measured in 100 mM and pH 4 solution. (D) HBT sensor measured in 100 mM and pH 8 solution. As shown in the figures, IB 2x10-9 A over the entire sensing range. Also from (C), IC curves are observed to be independent of solution concentrations, thereby indicating (IBR) is negligible. 1 IC (A) 1.5 pH 8, 100 mM pH 4, 100 mM pH 6, 10 mM pH 6, 100 mM HBT (no sol.) 1.0 0.5 10 -10 IB (A) 10 -9 Fig. s2: (A) Measured dependence of IC on the base current IB for the standard HBT device (solid line) and HBT sensors with solution of different concentrations and pH values (symbols). The observed dependence for sensors with different solutions is indistinguishable from that for the standard device (with no solution). From equation 1, this implies that (IB ∙ R) is negligible. 2 Discussion on the sensitivity definition: As discussed in the manuscript, a sensor has two components: device and sensing surface. The relative change in the sensing signal (I /I) can be written as: I /I = (gm/I) whereand (gm/I) measure sensing surface and device contributions, respectively. The goal of the present study is to demonstrate the enhancement in the sensor performance metrics due to the device component of a sensor: i.e. bipolar transistor versus nanowire FET. In Fig. 2 (C), the sensitivity of bipolar transistor sensors are compared with that for nanowire FET sensors. These sensors have different sensing surfaces, and therefore it is important that the sensing surface contributions are excluded so as to make a fair comparison between sensors with different device components. Since the definition of I /I per volt measures only the contribution of the device component, use of this sensitivity definition enables us to unambiguously demonstrate the superior sensing characteristics of bipolar transistor based sensors in comparison to nanowire FET based sensors. Furthermore, use of this definition in Fig. 2 (C) enables us to verify the prediction of equation 3 which predicts that I /I per volt = q/kT, irrespective of bipolar transistor device fabrication details. An alternate sensitivity definition is I /I per pH. This definition includes the contributions of both sensing surface and device components and is therefore not appropriate for the present study. However, it is of interest to re-plot the Fig 2. (C) in accordance with the definition of I /I per pH. The re-plotted data is shown in Fig. s3. 3 I / I ( pH -1 ) Bipolar Transistor Sensors 2 Nanowire FET sensors [6] [19] [4] 0 -10 -8 10 10 -6 10 Sensing Current (A) Fig. s3: A comparison between bipolar transistor and nanowire FET sensors using alternate sensitivity definition of I /I per pH. 4 -7 VBE (V) 10 -10 2 SI (A /Hz) 10 0.7 0.8 0.9 1.0 pH 6,100 mM -13 10 -16 IC (A) -10 1x10 -9 4x10 -7 2x10 -6 7x10 10 -19 10 -22 10 1/f -25 10 -28 10 0.1 1 f (Hz) Fig. s4: Typical collector current noise power spectra (SI) for the extended gate SiGe HBT sensor with 100 mM phosphate buffer solution at pH 6 at different IC values corresponding to different VBE values. SI is observed to have a power law dependence on frequency (f) with exponent of ~1 at lower VBE = 0.7 V and ~1.8 at higher VBE = 1.0 V. This implies that that 1/f noise dominates at lower VBE whereas random telegraphic noise starts to dominate at higher VBE values. Similar SI dependence is also observed for a HBT sensor with 10 mM buffer solution and the HBT device only (i.e. without the extended base consisting of TiN sensing surface, solution and reference electrode). 5 1080 pH 7.01 IC (pA) 1077 1038 pH 7.03 pH 7.03 1036 0 30 60 time (s) 90 Fig. s5: pH sensing measurements to investigate the limit of detection for HBT sensor; measurements show detection of pH = 0.02 with a measured signal to noise ratio of ~ 80. Discussion on Fig. s5 We further investigate limit of detection (LOD) using pH sensing measurements. The measured SNR of 2x105 V-1 for the HBT biosensor (Fig. 3(c)) translates to a minimum detectable voltage of ~15 V (for a typical sensing measurement: sampling rate is ~ 10 Hz and the total measurement time of ~ 30 minutes which gives BW =3). Using the measured pH sensitivity of 55 mV/pH and voltage ~15 V, the LOD is ~ 0.0003 pH. Fig. s4 shows the detection measurements for pH change of ~ 0.02 with signal to noise ratio of 80 which corresponds to measured LOD of ~ 0.00025 pH, consistent with the deduced LOD from SNR measurements. This measured LOD ~ 0.00025 pH is significantly less than the reported value of 0.01 pH for nanowire FET sensors.16 6 10 -9 (a) 70 (b) o Tthermocouple ( C) SS(mV/dec) 10 60 59.0 61.2 63.1 66.1 68.2 o 25 35 45 60 70 THBT ( C) I c (A) 10 -10 -11 THBT= 0.6 + Tthermocouple 50 40 100 mM HBT only 30 10 100 mM, pH 6 -12 0.5 20 20 0.6 VBE (V) 30 40 50 Tthermocouple 60 70 o ( C) Fig. s6: (a) Dependence of IC on VBE measured at different temperatures for extended base SiGe HBT sensor with 100 mM buffer solution at pH 6. Symbols are measurements and solid lines are fits in accordance with eq. (1b); Tthermocouple is the temperature measured using thermocouple and SS is the sub-threshold swing estimated from the fits. (b) Comparison between THBT and Tthermocouple; THBT is temperature estimated from the measured sub-threshold swing (SS) of the HBT sensor. 7