ECE3424 Electronics Laboratory Experiment #0 GETTING STARTED: Workstation instrumentation and Prototyping environment vers 2.2 Intent: Make an assessment of the environment. Run a snake check on cables and instrumentation and a preflight status check of the workstation. Check out the features and capabilities of the instrumentation cluster. Your workstation is centered around a cluster of the following instruments: 1. Digital Multimeter: Serves as (a) DC or AC voltmeter (b) Ohmmeter, (c) DC or AC ammeter. 2. Power Supply: Three outputs: (a) 5V fixed, (b) 0-20VDC variable (c) 0-20 VDC variable. 3. Function Generator: Provides an electrical signal that is periodic in form of three basic types: (a) square (b) triangular (c) sinusoidal (d) special applications, and can accomplish modulation and sweep of these waveforms. 4. Oscilloscope: Multipurpose instrument for viewing the voltage V(t) of any node. It has two input channels which are under the control of a time sweep channel. It has lots of knobs and buttons and menu options and is smarter than the average engineer. 5. Prototyping platform (MFJ Box): This box is the operational platform for the experiments. It includes (a) an internal power supply (b) an internal function generator (c) a set of 8 switches arranged as a byte (d) a set of 8 lights arranged as a byte (e) several interface connectors and switches (f) two potentiometers. The working space of the prototyping platform is a site where a prototyping whiteboard may be emplaced using velcro strips. The workstation will look something like that shown by figure 1. It is a 6' lab workbench. Take particular note of the fact that under the counter there are two drawers that contain cables and wires necessary for hookup of the instrument cluster to the experiment. 1 The instrument cluster is the operational environment for execution of tests and experiments. As shown by Figure 2 it is made up of the 5 basic instruments identified earlier. You might take note that different instruments will have different connector terminals specific to their usage. Instrument I/O connectors are always female connectors. The ones that you will encounter are shown by figure 3b. The corresponding male connectors are identified by figure 3a. If there are incompatibilities in connector pairs it may take a few patches to accomplish a simple connection. The simplest way to do so is to use the "alligator clip" (see figure 3a). The preferred way to connect up a circuit is via the prototyping board (protoboard) of figure 4 and this is the platform that we populate for each experiment. 2 3 THE MEASUREMENT INSTRUMENTS: Most of its features are self-explanatory. *The one that will give you the most trouble is (4). You will forget to push it or un-push it. +Caution: The ammeter input (2) has zero internal resistance. Make sure that you do not apply it to a voltage source or you may burn up both the source and the meter. The Volt/Ohm input (1) is the setting that you will use most often. This input essentially has ~infinite internal resistance. It is designed to be used with voltage probe cables (figure 5b). The DMM inputs accept either banana plug cables or the voltage probes (as shown). 4 Most of its features are self explanatory. You should take note that there are three power supplies in the box. (1) is usually used to support digital circuits, whereas (2) and (3) are usually used for analog (amplifier and device evaluation) applications Pushbuttons (4) defines which of the power supplies (1), (2) or (3) will be displayed on the meter (5). Output controls (7) are associated with power supplies (1) (2) and (3), respectively. Pay particular attention to switch (6). This switch will enable one power supply to be “slaved” to the other, i.e. one supply will track the other, same voltage one-to-one. This feature is particularly important when we desire to have power rails that are ‘bipolar’, i.e. (+,-) V, with a GND (common node) between and we desire to vary these voltage supply rails concurrently. The only feature that will give you any trouble is (8). This is the internal GND. It must occasionally be connected to one polarity or another for the use of the three power supplies concurrently. Notice that the inputs only accept banana plug cables. Wires can be clamped into the terminals but this is usually more trouble than it’s worth. 5 SIGNAL GENERATOR (AWG) Agilent 33521A The Agilent 33521A is a menu-driven instrument with the local screen display toggled by magnitude levels selected by the function keys (6) consistent with the type waveform as selected by the button keys (7). Specific values are set using the numeric keypad (10) and second-level screen display options. Most of the operational features are denoted by text and icons and are reasonably self-explanatory. The main output is taken from (16) which is toggled on/off by the channel button just above (16). Since you have a single-channel output (the Agilent 33521A) you will only have one output channel and it will be located at site (16). Waveform type and the parameter selects are determined by the pushbuttons (7). At the beginning all you should do is figure out how to call up a sine wave and set its amplitude and frequency, and then toggle it along to the oscilloscope. Much of what the AWG (arbitrary waveform generator) can do must be ascertained by discovery. Although you will be taken through the basics, the rest of the story will call for a walk down any and all of the paths that this instrument offers. The AWG can realize a good many signal and sweep offerings, and your task, should you choose to accept it, is to investigate any and all of its capabilities and have them ready for service in the set of circuit and device tests called for by the experiments. 6 Tektronix TDS 202C OSCILLOSCOPE 7 The oscilloscope is a viewport that is usually used to view signals tapped out from a node of interest. For a high-quality instrument such as this one the probes are non-intrusive and will not induce any measurement artifacts. For more sensitive measurement situations higher quality low capacitance probes are employed, but since the lab does not usually need them they are left in the background while alligatorclip probes are clipped to the circuit or device under test. 8 You might take note of the fact that the MFJ box has several of its own internal instruments. (1) corresponds to a power supply, fixed voltages of (+,-) 12 and 5VDC. (2) is a signal generator. It may have some mild distortions in the sinusoidal output. (3) is a clock pulse output, with three fixedfrequency settings. (4a) and (4b) are some internal potentiometers, handy for some types of experiments. (5) is a single-pulse output. For logic experiments there is a `byte' of input switches (6) and a `byte' of output indicators (7). These are usually used in conjunction with (3) and (5). (9) is a push-button switch and (10) is a means to make connections off-panel to other instruments. 9 Incoming status check: (before each experiment) If your predecessor has left the workstation in good order and in the standard conclusion mode settings listed below you should be able to do a status check in a glance. (1) Check out the power supply via the DMM and confirm (1) power supply output and (2) tracking of the +20 and -20 voltage supply rails. (2) Check out the function generator via the Oscope. Set it for a sinusoidal output of 1 kHz with zero and amplitude at 1.0 Vp-p on Ch1. If necessary initialize the O-scope by the ‘autoset’ button. Then make a quick check to ensure that the offsets, amplitude controls, and (Oscope) amplifiers are functioning correctly. (3) Check out cables, and make sure that they do pass signals with no weakness or intermittent loss. Cables are a common villain in the loss of workstation integrity. Conclusion settings: at the end of every lab session Restore the instrument cluster to ready-mode settings (listed below). Leave instruments ON and waiting for the next user even if the next lab section meeting is next day or next week. (a) Power supply: Voltage: 10V. Display switch: 20V Tracking: Independent (b) DMM: Option: Volts DC Range: 20 (c) Function Generator: Frequency: 1 kHz (range = 103). Type: sinusoidal Offset: null position Amplitude: 1.0 Vp-p (d) Oscilloscope: Offsets: Settings: Ch1: 1cm above axis, Ch2: 1 cm below axis Ch1: 500 mV Ch 2: 500 mV, M: 500 us trig'd M Pos: 0.000 CH1: coupling DC BW limit: OFF Volts/div: Coarse Probe 1x Trigger: Edge slope: rising Mode: Auto Coupling DC (2) Return (a) cables to right drawer and (b) wires to right drawer for use by the next person. It is necessary and essential that we following this regimen in order that you and your brethren and sisteren can start up each lab experiment with maximum confidence in the instrument cluster, minimum incoming operational problems and minimum start-up overhead. 10 GETTING ACQUAINTED WITH THE INSTRUMENT CLUSTER (PROCEDURE): I. Power Supply and Digital Multimeter A. Set up the power-supply and DMM as shown by figure I-A-1. Set the tracking switch to independent. Set 20V voltage supply to 8V as defined by the meter on the power supply. Measure this value using the DMM (is it the same?). Repeat for 12V, 16V and 20V. Enter data in a short table. B. Repeat for the -20V power supply. Note that you will have to change the ‘display’ switch on the power supply. You have just performed a calibration of the two power supply variable sources. Hopefully both are linear. C. Set the tracking switch to parallel and the display switch to 20V. Adjust the 20V power supply to 8V, 12V, 16V and 20V, respectively. Using the DMM alternately between (+20V, -20V) and toggling the display switch, record in a short table, V+ for display output, V- for display output, V+ for DMM, V- for DMM. These should correlate. This process is the calibration of the tracking mode. Figure I-A-1. Instrumentation check 11 II. Function Generator (AWG) and Oscilloscope A. Set up the function generator and Oscilloscope as shown by figure II-A-1. B. Use the function generator menu and screen to set the function to sinusoidal, with frequency = 1.0 kHz, offset 0.0, and amplitude setting 1.0Vp-p. Push the ‘autoset’ button on the Oscope, which should give a clear clean display of this waverform. The the autoset has not done so adjust the offset for CH1 of the Oscope so that the marker falls 1 cm above the median. Measure peak-peak voltage of the sine wave by means of the V/cm scale and the ‘measure’ button. Is it anywhere close to the amplitude setting of the function generator? (why not?). While you are in this setting check out the effect of the ‘offset’ of the function generator. *The AWG is designed for an output load of 50standard. So it is necessary to include a (BNC) ‘tee’ with a 50 load on one of the tee outputs as shown by the figure below. If these output accessories are available (the tee and the 50 load) then attach as shown repeat the measurements of part B. Check your trigger menu. It should be set for: Type: edge, Source: CH1, Slope: rising , Mode: Auto, Coupling: DC. If it is not at these settings, then reset them accordingly. Adjust the horizontal positioning knob (tie scale) and you will see the reference marker (on the upper axis) translate accordingly. Set it to 0.000. C. Measure the period of the sine wave using (a) ttime scale (500 s/cm, or 50ms full screen) and the cursors*, and (b) the ‘measure’ button. Take the inverse to determine f. It should be close to 1.0 kHz if not exactly, since these are digital instruments. Reset the function generator to f = 2kHz and 5kHz respectively, and repeat the O-scope to measure period and frequency and record in a short (Excel) table. Change time scale of the Oscope as needed. Reset the AWG function generator to 10kHz, 100kHz, 1MHz, respectively. This procedure is a typical calibration sequence for the time scale. *The cursor menu is invoked by the ‘cursor’ button (second row of buttons from the top). If you push the button adjacent to the screen sidebar menu labeled as ‘Type’ you will invoke the cursors and be able to select either the y-axis scale (Amplitude) or the x-axis scale (Time). The knob at the upper center of the panel controls the cursor position and the sidebar menu will the display the position coordinate value of the cursor(s). D. Repeat part C for a signal taken off of the ‘pulse’ output of the function generator (it is under the ‘waveform’ button). For the 1kHz frequency select a pulse width (under the ‘parameters’ button) of 250s. Continuing with other frequencies select a pulse width of 0.25 x T (= period) for each frequency. Frequency assessment will be nearly the same, but more accurate since the pulse interval is usually an easier read than the sinusoidal period. Check and confirm pulse width as well as the frequency. 12 There should not be any great surprises unless your instrument has drifted out of calibration. Figure II-A-1. Instrumentation check E. Reset function generator to 1.0 kHz and 1.0Vp-p amplitude. Connect CH2 to the output from the internal signal generator of the MFJ box, set to sinusoidal. You may have to do some patching to make these connections. Adjust amplitude of CH2 display so that it is nearly the same as that for CH1. Find the X-Y setting for the Oscope (look under the display menu) and invoke this setting. An X-Y pattern is generated is of the form of a "Lissajous" figure. The Lissajous figure is used to compare two signals. If they are different frequencies you will see a “rotating” pattern. Tune the MFJ frequency so that the pattern stops "rotating". When the pattern becomes stationary, the frequencies are synchronized. Adjust the amplitudes (and frequency) so that the figure is a perfect "O". If any distortion is present then it will show up as an imperfect "O" (or an imperfect ellipse). F. Change the frequency of the MFJ box to 4kHz and synchronize to that from the Agilent 33521 function generator. Make a sketch of the pattern. Now change the frequency of the function generator to 4 kHz synchronizing it to the MFJ box. Find the display setting, change to YT, triggered by CH1, and measure the period. It should be almost exactly 0.25 ms, because you have used synchronization to reset your signal frequency from 1kHz to 2 kHz. This process can be done a good bit more efficiently with electronics using a technique called "mode locking" and is the basis for most of the uplinks to satellite communications and GPS systems. It is also the technique used to make audio synthesizers (which require precise harmonics). G. Using the DMM and its `probe' cables, probe the sinusoidal output of the function generator with the DMM set to ACV. Repeat for outputs from the function generator of Vout = 2V, 5V respectively, and record results in a short table. In the table, compare 0.5*Vp-p of the Oscope display value with the ACV reading on the DMM. The DMM measurement is the RMS (root-mean-square) value of the periodic function. For sinusoidal output VRMS is approx 0.707 x V(peak). 13 ADVANCED USAGE of the AWG There is not much entertainment with the Oscope since it is mostly a viewport. However the AWG will generate as many exotic waveforms as there are chords on a guitar. Otherwise it may be regarded as a source of waveforms and waveform options that serve to exact a complete test and evaluation of circuits and devices, which is the charter of the set of lab experiments. The feature of principal interest is the frequency sweep function. It requires a little more setup but is a real timesaver in the measurement of diode and transistor parasitic capacitances. With the waveform set to a sinusoidal f = 1kHz and amplitude 1.0 Vp-p (as before), push the ‘sweep’ button and bring up its menu bar. Toggle the necessary buttons below the menu bar to accomplish the following settings: Type: Start freq: Stop freq: Sweep time: linear 1 kHz 10kHz 10 ms On your O-scope set the time scale to 100s (per cm). The trace you see will display the signature pattern for a sweep frequency on a linear time scale, and is used as a confirmation of frequency sweep. Later in life you will be expected to choose an AWG frequency sweep on the order of 10kHz to 1MHz for an Oscope sweep time of 1ms/cm with the full screen window of the O-scope 10 x 1ms. Get check-off from your instructor on all of these tests and data as you march through them. This is called an "informal report". This checkoff report has value of approx 50% of experiments that include a written laboratory report. **At the conclusion of the experiment please attend to the proper close-out procedure (page 10). Your brethren and sisteren will appreciate your attention to this courtesy. 14