MLAB 2360 Clinical I Microscopy Training Activity 1: Parts of the Compound Microscope Aperture 1 MLAB 2360 Clinical I Microscopy Training 2 Parts of the Microscope Defined: 1. Main Switch: On/Off 2. Rheostat: Light intensity knob 3. Coarse adjustment knob: This knob is used to raise and lower the stage of the microscope. The coarse adjustment is used to first bring the object into APPROXIMATE focus starting first with the stage as close to the objective lens without touching. Then move the coarse adjustment so that the stage moves away from the lens until the object is in relative focus. **Note: The coarse objective will ONLY be used when focusing with the 4X or 10 X objectives to avoid damage to the other objectives. 4. Fine adjustment knob: This knob is used to focus the image into a sharp, critical focus after achieving relative focus with the coarse adjustment knob. **Note: These microscopes are PARFOCAL microscopes, meaning that if you are properly focused on a lower objective (10X), you will be able to switch to a higher objective and maintain relative focus. Do NOT use the coarse adjustment when on 40X, 50X or 100X. 5. Field Iris Diaphragm: Using the field iris diaphragm ring, the diameter of the field iris can be adjusted to circumscribe (encircle) the field of view to exclude extraneous light and improve image contrast. We will use this “field diaphragm” when setting the microscope for Kohler illumination. 6. Condenser Lens: Below the stage is the condenser lens. This focuses light onto the object and is NOT involved in the magnification. The focusing adjustment is a rack and pinion movement to permit vertical movement of the condenser. Clear images are obtained only when the condenser lens is in proper focus (Kohler illumination). **Note: In Urinalysis and when using hemacytometers, it is often necessary to LOWER the condenser to achieve greater contrast. This reduces resolution, but it is helpful when searching for transparent elements in the specimen. Aperture Iris diaphragm: The recommended setting of this aperture iris diaphragm is between 75-85% of the NA of the objective in use. When using the 100X objective, the aperture iris diaphragm should be set to 0.9 on the scale. Condenser centering screws: These screws are used during the Kohler illumination procedure to ensure that the condenser is centered in the Field Iris diaphragm. 7. X-axis : Moves the stage left to right 8. Y-axis: Moves the stage forward and back 9. Objectives: These are lenses mounted on the revolving nosepiece. Each lens is marked with its Magnification/numerical aperture (NA) and Focal Distance. Most of our microscopes are equipped with 10x and 40X DRY objectives as well as 50X and 100X OIL MLAB 2360 Clinical I Microscopy Training 3 objectives. The DRY objectives use only air between the lens and the slide, while the OIL objectives utilize a SMALL drop of oil between the lens and the slide. The objective lenses must be kept clean. Use lens tissue or KIMWIPES to clean the outer, exposed surfaces of the lenses. Although the OIL immersion lens is designed to work in oil, the oil must be removed after used and before you put the microscope away at the end of each class period. **NOTE: Once you have used the 50x or 100x oil immersion objective to examine a slide, DO NOT “go back” to the 40x objective. Doing so will cause oil contamination of the 40x objective, rendering it useless until it can be cleaned. The oil will cause the gaskets in the objective to erode and cause permanent and EXPENSIVE damage to the objective. Oil is ONLY used on the 50X and I00x oil objectives. 10. Oculars: The ocular lenses usually magnify 10X. Thus the total magnification observed is the multiplication of the power magnification of the ocular times the objective. The oculars can be moved back and forth to adjust to the interpupillary distance of the student. When first using the microscope, adjust the ocular lenses back and forth until a circular field is viewed with both eyes open. The index dot indicates the interpupillary distance on the Interpupillary Scale. Adjust the Diopter: Looking through the right eyepiece with your right eye, rotate the coarse and fine adjustment knobs to bring the specimen into focus. Then look through the left eyepiece with your left eye (cover the right eye) and focus the specimen using the adjustment ring on the left ocular. This allows for both eyes to be focused individually. Using the Eye shades: o When using the microscope with eyeglasses, the eye shades should be in the folded back position. o When using the microscope WITHOUT eyeglasses, the eye shades should be extended toward you to prevent extraneous light from entering between the eyepiece and the eye. Reticles: (Eyepiece micrometers) They are clear circular glass inserts with a scale inscribed on them. The reticle or eyepiece micrometer sits at the focal plane inside the eyepiece lens of the microscope and allows the investigator to make accurate measurements of specimens. In binocular scopes like the ones you are using, there will be a reticle in ONE of the lenses. Reticles come in many varieties and in different diameters. MLAB 2360 Clinical I Microscopy Training 4 The "actual" distance between any two marks on the reticle are a function of the objective lenses only. The best way to calibrate your reticle is to use what is called a stage micrometer. This is a slide that has tiny marks of a known dimensions inscribed on it. By making a comparison of the marks on the stage micrometer to the marks on the reticle, one can establish the actual value for each mark on the reticle. When you look into your eyepiece lens, the markings will always be the same but the size of the image superimposed under them will get larger with more magnification. So, as you change to a higher power objective lens, the represented value between marks will change proportionately. MLAB 2360 Clinical I Microscopy Training Let's say each division of the metric stage micrometer above is 0.01mm or 10 µm or 10µ (10 micron). The eyepiece reticle is divided into 100 units. **Note: 1 micron = 1 micrometer so 1 µ=1 µm First determine how many divisions of the eyepiece reticle correspond to a certain distance on the stage micrometer by finding the FIRST line of the eyepiece reticle that perfectly lines up with the stage micrometer. Calculate the length which corresponds to one division of the eyepiece reticle. In this example, 30 divisions of the eyepiece reticle corresponds to 20 divisions of the stage micrometer. Each division of the stage micrometer equals 10µ so 20 divisions of the stage micrometer would equal 200 µ. To calculate the value of one division of the eyepiece reticle we would divide 200µ by 30 and the result would be 6.67µ per reticle division. The reticle value, in this case 6.67µ, would apply only to the objective for which the calibration was made. Each microscope objective must be calibrated independently. **Note: Care should be taken when removing dust covers from microscopes in order to avoid pulling an eyepiece off of the scope and having it drop to the floor or workbench. Dropping eyepieces can damage them beyond repair. **Note: Eyepieces should be kept clean. Fingerprints can blur images and mascara can “appear” in the field of view. A fresh piece of lens paper should be used to clean each eyepiece. Gently “fogging” the eyepiece with your breath before wiping can aid in removing fingerprints. Rubbing mascara when it is on a lens can etch the glass; therefore, it is important to “blow off” as much of the mascara as possible before wiping the lens. Activity 2: The Compound Microscope Video http://www.austincc.edu/biocr/1406/labv/microscope/index.html Activity 3: Review the parts of the Microscope http://www.austincc.edu/biocr/1406/laba/microscope/index.html Activity 4: Watch Virtual Compound Microscope Demonstration and Practice on Your Microscope http://www.udel.edu/biology/ketcham/microscope/scope.html 5 MLAB 2360 Clinical I Microscopy Training 6 1. Place microscope on the workbench as shown in the first video. 2. Select a specimen slide from the slides provided by the instructor. 3. As the instructor goes through the Virtual Microscope Demonstration, find the parts of the microscope and perform the tasks in the checklist. Activity 5: Kohler Illumination http://youtu.be/fNTNZ8nWPd4 Illumination of the specimen is the most important variable in achieving high-quality images in microscopy and critical photomicrography or digital imaging. Köhler illumination was first introduced in 1893 by August Köhler of the Carl Zeiss corporation as a method of providing the optimum specimen illumination. The object of the process is to align the condenser to ensure that the cone of light emitted from the illuminator (light source) is focused directly on the specimen and the apertures are set to maximize contrast. The primary advantage of Kohler illumination is even illumination free from glare; however, in urine samples and when using a hemacytometers, the condenser is often lowered to increase contrast then raised to increase resolution. Kohler illumination is not warranted in these circumstances, but IS recommended in the hematology, microbiology, and parasitology areas of the lab. Adjusting Contrast: 1. Place a specimen slide onto the microscope stage and secure between the specimen holders. 2. Set the interpupillary distance to ensure comfort. 3. Place the 10X objective lens into position and focus the specimen using the coarse adjustment knob. a. Close the right eye. Use the fine adjustment knob to sharpen the image. b. Close the left eye. Focus is obtained using the diopter ring shown below. The diopter ring is normally located on the right eyepiece and moves both in the counter clockwise and clockwise directions to achieve the necessary focus. MLAB 2360 Clinical I Microscopy Training 7 4. As stated earlier, the aperture iris diaphragm knob should be set on the aperture iris diaphragm scale to a setting between 75-85% of the NA of the objective. We will perform the Kohler illumination on the 10X objective which has a NA of 0.25 (as seen below); therefore, you are to set the aperture diaphragm knob to approximately 0.2 on the scale. Objective Magnification Where video begins NA a. This can be done visually as shown in the video. Remove the removable eyepiece and place it carefully on the table. b. Look through the ocular. c. When you move the aperture iris diaphragm knob from right to left you will see a hexagon or octagon depending on the model expanding then becoming a circle of light. You want to set the aperture iris at approximately 75-85% of the capacity (gray circle) by sliding the aperture iris knob. d. Replace the eyepiece carefully. Centralizing Field Diaphragm 5. With the 10X objective in the viewing position, focus the specimen using the coarse and fine adjustment knobs. This gets the stage in the appropriate position. 6. Once focused, close the diaphragm. This can be done by moving the field iris diaphragm knob to the far left. 7. While looking through the oculars adjust the condenser knob until the hexagon/octagon of light/specimen is at its sharpest. 8. Continue looking through the oculars and centralize the condenser by twisting the condenser centering screws. 9. Open the field iris diaphragm by moving the field iris diaphragm to the far right. 10. You have now achieved Kohler illumination. MLAB 2360 Clinical I Microscopy Training 8 Activity 6: Focus with an OIL Objective 1. Select a blood smear that the instructor supplies. 2. Place the specimen slide onto the microscope stage and secure between the specimen holders. 3. Set the interpupillary distance to ensure comfort. 4. Place the 10X objective lens into position and focus the specimen using the coarse adjustment knob. i. Close the right eye. Use the fine adjustment knob to sharpen the image. ii. Close the left eye. Focus is obtained using the Diopter ring. 5. When the image is focused with the 10x objective, carefully hold the nosepiece and move the 100X oil objective into viewing position. **Note: These slides have had oil used on them before, therefore, be careful to go from 10X to 100X WITHOUT passing the 40X over the slide. 6. Once the 100x oil objective is in place, apply a SMALL drop of oil to the space between the objective and the slide. 7. Use ONLY the FINE adjustment knob to focus on the specimen. **Note: These microscopes are PARFOCAL microscopes, meaning that if you are properly focused on a lower objective (10X), you will be able to switch to a higher objective and maintain relative focus. 8. Use the X and Y axis knobs to move through the slide. 9. When you have looked at a minimum of 5 fields (visible area of oculars), lower the stage using the coarse adjustment knob, remove the slide, and GENTLY remove the excess oil off the slide using a Kimwipe. 10. Using a clean Kimwipe and lens cleaner, clean the 100X oil objective. 11. Properly put away your microscope as shown in the first video.