John Siller Laboratory 2 GC/MS and GC/FID Introduction The purpose of this lab was to familiarize and learn the operations of the GC/MS and GC/FID. To do this we ran a number of hydrocarbon samples and reviewed the spectrum of peaks. The gas chromatograph separates volatile compounds into its molecular components. Mass spectrometry and flame ionization detectors are two methods of detecting the components that are separated by the GC. The GC is a very good instrument to use when attempting to separate and identify a number of analytes from a mixture. Gas chromatographs are found in a wide variety of laboratory’s including labs that focus on the analysis of gasoline and its components. There are a variety of detectors for gas chromatographs each with a different purpose. The detectors for this lab are the mass spectrometer and the flame ionization detector. The mass spectrometer is useful for a number of anyltes including volatile hydrocarbons. The flame ionization detector is also useful for a wide range of flammable products. In this experiment we ran octane, 1-octanol, and t-butyl methyl ether on both gas chromatographs with different detectors. In addition we prepared different mixtures of these hydrocarbons and ran them to see separation. Procedure Day 1 GC/MS Only 1 sample of 1-octanol was run due to the helium running out. 0.1 mL of octanol was diluted with 10 mL of chloroform. 1 μL of this solution was injected into GC with 1 µL of air. Peaks were observed at 1 minute and 5 minutes. Day 2 GC/MS Multiple samples of octane, 1-octanol, and t-butyl methyl ether were run including mixtures containing each analyte. Again 0.1 mL of analyte was diluted with 10 mL of chloroform and 1 μL was injected with 1 µL of air. Different peaks were observed for different anyltes. GC/FID Same anyltes as GC/MS were run including mixtures of each. Similarly 0.1 mL of anylte was diluted with 10 mL of chloroform and then 1 μL was injected with 1 µL of air. Peaks were observed at different retention times depending on the anylte(s). Results All spectra can be found in my lab book. Below is a list of anyltes and there corresponding retention times for each peak. Normally only one peak was observed per analyte and this peak is used to identify each peak in the mixture spectra. All analytes are diluted in chloroform solutions as previously stated. GC/MS Analyte 1-Octanol Retention Time 0.85 min 4.95 min Octane 0.65 min 1.25 min 1.70 min 3.05 min T-butyl methyl ether 3.05 min Mixture 0.75 min 0.95 min Table 1: GC/MS retention times for analytes completed. Octane’s spectrum was very messy so those are just the four largest peaks. GC/FID Analyte 1-Octanol Octane T-butyl methyl ether 1-Octanol and t-butyl methyl ether n-butanol, decane, and heptane Retention time 0.75 min 0.80 min 0.85 min 0.80 min† 0.55 min 0.58 min 0.65 min 5.35 min Table 2: GC/FID retention times for anylates completed. Analysis and Conclusion This was our first lab after our two weeks of expert work and as a result it did not go off as smoothly as we thought it would. Gas chromatographs are great for collecting detailed data but require a good amount of time per sample. On the first day of our lab, Monday, we only were able to run two samples of 1-octanol before the helium tank ran out and we were forced to quit for the day. This put us pretty far behind and we were not able to run as many separations as I would have wanted. We were also not able to sit down and figure out the best way to analyze our chromatograms as a result of this time crunch. If it were possible I would have liked to spend at least another week on the gas chromatographs to finish collecting our data and come up with spectra that we could properly analyze and interpret. Many of our spectra, especially on the GC/MS, were not as good looking as I hoped, but running multiple samples of each analyte were simply not possible due to time constraints. If I were to repeat this lab in the future I would run at least three samples of each analyte to confirm the accuracy of the spectra. Additionally we were not too sure on the method we selected for the temperature variance of the chromatogram, and if any future work would be done on them I would like to test to see the optimal method for separation. Because this was our first lab on a new instrument on our own however, I am not entirely disappointed with the results. My father used to work with a GC/MS every day in his lab at ExxonMobil so I know how useful they can be for the analysis and separation of different anyltes.