Supplemental Information: Particle Emission Characteristics of a Gas Turbine with a Double Annular Combustor Adam M. Boies, Marc E. J. Stettler, Jacob J. Swanson, Tyler J. Johnson, Jason S. Olfert, Mark Johnson, Max L. Eggersdorfer, Theo Rindlisbacher, Jing Wang, Kevin Thomson, Greg Smallwood, Yura Sevcenco, David Walters, Paul Williams, Joel Corbin, Amewu A. Mensah, Ramin Dastanpour and Steven N. Rogak Emission Indices Black carbon emission indices, EI(BC), were compared for three different laser induced incandescence (LII) instruments on three different lines. The results of measurements on the FOCA, Annex and SAMPLE III lines are shown in Figure S1, and demonstrate that the differences in measured EI(BC) among instruments and sample lines is less than the different EI(BC) at each given thrust setting. For each thrust setting the variability of the CFM56-5B42P EI(BC), indicated by the grey error bars, as measured by each LII overlaps demonstrating that no distinction can be made by the measurements from individual instruments or lines at a given thrust. These results establish that reported EI(BC) values using one LII on one line are not statistically different than using other LII results on other lines. Figure S1: Black carbon emission index, EI(BC), for CFM56-5B4-2P at various thrust settings as measured by LII on three separate lines shown in Figure 1 for 28th April. The inset is a rescaled version of EI(BC) for DAC operation. The variability for particle number based emission indices (EIn) was low and was omitted from the main article to allow for better view of symbols. SI Figure 2 below is a reproduction 1 of Figure 4 within the main article with the addition of 5 and 95% variability. The variability within the system was the greatest source of uncertainty within the system. Figure S2: Black carbon number emission index, EIn, for CFM56-5B4-2P as measured by CPCs with D50 cut points of 10 nm (open symbols) and 23 nm (cross and plus sign) on two separate lines (FOCA and SAMPLE III) for various thrust settings. Error bars represent the 90% variability interval within a given thrust setting. Air-to-Fuel Ratio Global air-to-fuel ratio was as determined from undiluted CO2 measurements according to the method defined by the Office of Aviation Research (2006). The air-to-fuel ratios relative to the stoichiometric air to fuel ratio (SAFR = 22.72) are shown in Figure S2 for varying thrust settings. Single pilot combustion occurred during low thrust settings (< ~25% full thrust) and resulted in a global lean combustion. However, the local stoichiometry within the pilot combustor is rich (not measured) and only mixes with excess air downstream of the pilot combustion zone. The rich combustion ensures stability of the flame. During dual annular combustion the global air-to-fuel ratio drops, and the second combustor operates in a locally lean-burn condition. Local stoichiometry data was not measured for either pilot combustor or DAC thrust settings. 2 Figure S3: Global air-to-Fuel (AFR) ratio for varying thrust settings over the range of thrust settings. 3 Effective Density The particle effective density was determined as a function of particle mobility in the manner described by Johnson et al. (In Press 2014) where π = 11.92 and π·fm = 2.76. The effective density of the mean mobility diameter is plotted in Figure S3 as a function of thrust setting. The effective density is seen to decrease from ~900 kg/m3 to ~720 kg/m3 with increased thrust setting during pilot combustion mode. When comparing these results to Figure 5 where it is shown that particle mobility diameter increases over the same thrust range, it is apparent that the effective density of particles decreases as the particle diameter increases. The change in effective density is also known to impact the signal decay from LII measurements, resulting in longer decay times for particles of greater effective densities. For DAC thrust settings (> ~25% full thrust) more than 90% of the effective densities are determined to be between 800 to 900 kg/m. In almost all cases the effective densities, as determined by using mobility measurements from the DMS500 (FOCA, SAMPLE and Gantry) are greater than the effective densities determined by using the nanoSMPS (SAMPLE) and longSMPS (SAMPLE/FOCA). Figure S4: Number mean aggregate effective density versus thrust as determined by massmobility relations at a range of engine thrusts. The effective density is derived from Eq. 2 and the mobility measurements shown in Figure 5 from DMS500 (FOCA – asterisk, SAMPLE – triangle, Gantry - Circle), nanoSMPS (SAMPLE – diamond), longSMPS (SAMPLE/FOCA – square). Aggregate Mobility Diameter and Mass The mean aggregate mobility diameter was measured for multiple CPMA-selected aggregate masses at different thrust settings (circle – 9%, triangle – 17%, square – 21%, star – 24% and 4 diamond – 31%). The results are shown in Figure S4 for both undenduded (white fill) or denuded (black fill), where there is a similar power-law relationship observed for all samples. These results are equivalent to those shown in Johnson et al. (In Press 2014). Figure S5: Measured mean aggregate mobility for a CPMA-selected aggregate mass. Thrust settings are represented by symbols (circle – 9%, triangle – 17%, square – 21%, star – 24% and diamond – 31%), face color indicates whether the samples were undenduded (white) or denuded (black) and line type indicates whether the samples were collected from the FOCA (solid) or Annex III line (dashed). 5 Primary Particle Diameter The volume area equivalent primary particle diameter was determined as described in §2.2, Eq. 4, which relates the volume area equivalent primary particle diameter within an aggregate to the measured aggregate mass and mobility. As described in the text accompanying Figure 6, volume area equivalent primary particle diameter increases with aggregate mobility, 0.8 whereby a power-law relationship of πva = 0.79 πm ± 25% encapsulates all but one of the measured data points. Individual power-law fits are shown in Figure S5, which demonstrate that higher R2 values (R2 > 0.96 for all but one thrust setting) can be achieved when fitting individual thrust settings. The resulting π·va values vary from 0.65 to 0.98 for the individual fits with πva values that vary from 1.39 to 0.44, where larger π·va correspond to smaller πva values. Figure S6: Volume area equivalent primary particle diameter as a function of aggregate mobility diameter as measured by mass and mobility analysis. Primary particle size is determined according to Eq. 4. The blue lines correspond to the empirical fit with the power law form of Eq. 5 for each thrust setting. 6 Mass-Mobility Exponent The mass-mobility exponent, π·fm , was compared for a method by which it was determined by effective density data to Eq. 2, and from a method of fitting the data to the volume area equivalent primary particle data. The ordinate represents the mass mobility exponent, π·fm , determined by taking the best fit to effective density data as conducted by Johnson et al. (In Press 2014). The abscissa represents the mass mobility exponent as determined by finding π·va from a least-squares fit to primary particle data as shown in Figure 6 and solving the relation, π·fm = 2π·α − π·va (2π·α − 3) (Eq. 5 in main article) with a π·πΌ =1.069. The difference in the Dfm is small and arises from the differences of conducting a least squares fit to the same data weighted in different manners. These results confirm that using the Eggersdorfer constant, π·α =1.069, allows for conversion between π·fm and π·va . Figure S7: Mass-mobility exponent, π·fm , determined by taking the best fit to effective density versus mass-mobility exponent as determined by a fit to primary particle data shown in Figure S5 for π·va and calculated using Eq. 5. Dynamic Shape Factor Dynamic shape factor may be determined by comparing the measured mobility and aerodynamic diameters. AMS-measured vacuum aerodynamic diameter was measured in parallel to mobility measurements, see Figure 1. The mean mobility diameter and mean vacuum aerodynamic diameter were plotted and compared to lines of constant dynamic shape factor, π, as defined by DeCarlo et al. (2004). The dynamic shape factors of the measured aggregates are equal to or less than 1, which is a non-physical result. It was concluded that 7 insensitivity of the AMS to vacuum aerodynamic diameters less than 50 nm biased the measurements, resulting in mean measured vacuum aerodynamic diameters that were skewed to larger diameters. Further work is needed to improve the sensitivity of AMS for aggregates with diameters less than 50 nm in order to make definitive measurements of dynamic shape factors for gas turbine BC particles. Figure S8: Vacuum aerodynamic diameter and mobility diameter of aggregates measured at seven different thrust settings (black diamonds). Lines of constant dynamic shape factor, χ, as defined by DeCarlo et al. (2004), are shown along with general relations for π for aggregates of less than and greater than 60 primary particles. TEM Image Processing A semi-automatic image processing program developed in MATLAB was used for the analysis of the TEM micrographs (Dastanpour et al., 2014). Grayscale TEM image (Fig. 8-a) are binarized by setting threshold level for the brightness of the image (Fig. 8-b). Aggregate projected area, maximum length and width, 2-D gyration radius, and projected area equivalent diameter of the aggregates are measured from the binary image. Primary particles are sized manually. To enhance the accuracy, large aggregates are cropped into several sections. Since primary particles are not perfect spheres, the reported diameter is determined by the mean of the diameter measured in two different directions (Fig. 8-c). 8 Figure S9: TEM image processing steps: a) Original TEM image; b) Binarized TEM image; c) Manual primary particle sizing Although manual measurement of the primary particle size is time-intensive, unlike automatic methods (Bescond et al., 2014; Grishin et al., 2012), it does not require prior information on the size distribution of the monomers in aggregates. Grishin’s method is also sensitive to model parameters and has to be calibrated for different TEM images frequently. Line Loss Correction Factor As shown in Figure S9, the ratio of line loss corrected to measured particle number and particle mass concentrations were determined for each thrust setting. The particle number line loss correction factors (πΆπΉPN ) varied from 4.4 to 1.7 with the highest correction factor at low thrust setting, and the lowest at the highest thrust setting of pilot combustion mode. Similarly, the particle mass correction factors (πΆπΉPM ) were highest (2.2) at the lowest thrust setting and lowest (1.5) at the highest thrust setting of the pilot combustion mode. The line loss correction factors for mass (mean of 1.9) were less than the particle number (mean of 2.9) correction because of the inherent weighting of mass mobility distributions to larger particle sizes, thus lessening the impact of small particle losses. The πΆπΉPN and πΆπΉPM were determined using the downstream measured geometric mean and a fixed geometric mean (ππ =1.7). The results indicate that over the range of distributions measured, the impact of varying geometric mean had little effect (<10% relative error) on the overall line loss correction factors. 9 Figure S10: UTRC line loss penetration of 25 m line length and line temperature of 160°C. The line loss correction factors were determined for the particle number (PN) and particle mass (PM) for each thrust setting, shown in Figure S11. The line loss correction routine was applied to the measured size distributions, which were in general well characterized by single-mode log-normal distributions with geometric standard deviations πg = 1.73 (90% CI [1.65 1.80]). Correction factors are greatest for thrust settings with smaller particle diameters. In all cases, particle number correction factors were larger than particle mass correction factors, as particle number distributions are dominated by small particles whereas particle mass is proportional to the third moment of the particle size distribution. As shown there is little difference between correction factors determined by using the measured geometric standard deviation as opposed to a constant geometric standard deviation πg = 1.7. 10 Figure S11: Line loss correction factors for particle number and particle mass at various thrust settings. Line loss correction factors indicated by solid symbols were determined using the respective measured geometric standard deviation, while the open symbols represent correction factors that were determined with a constant geometric standard deviation (πg = 1.7). As shown in Figure S10, the corrected geometric mean mobility diameter is less than the measured geometric mobility diameter for all distributions due to the greater propensity for smaller particles to be preferentially lost during transport. The relationship between line loss corrected and measured geometric mean mobility diameter is well described by a linear relationship, πpc = 1.165 ππ − 8.8249 nm, where the norm of the residuals is 1.36 for the corrected versus measured geometric mobility diameter. The linear relationship is used to correct the reported geometric mobility diameter within the results section of the article. 11 Figure S12: Line loss corrected particle mobility diameter shown versus measured particle mobility diameter. 12 Engine Conditions Figure S13: Measured combustor inlet temperature (T3) and engine speed for CFM56-5B42P at various thrust settings. As shown in Figure S14 the individually measured datapoints for EI(BC), EIn, and geometric mean mobility diameter demonstrate the repeatability of the measured metrics at a set thrust point (~17%). In general measurements that were taken on the 28th and 30th April, 2015 agreed well, with deviations within an individual run that were less than the deviation between measurements taken on different days. Measurements that were taken on the 29th April, 2015 tended to be more highly variable due to deviations in the dilution system, affecting the concentration of CO2 measured and calculated dilution factor. During all measurements, despite the variability in thrust setpoint, the measured quantity remained relatively steady. The geometric mean particle diameter was relatively insensitive to thrust setting, but does appear to be affected by the line in which the measurement was taken on. In particular the SAMPLE system appeared to have higher small particle loss (<20 nm) which had the effect of increasing the geometric mean mobility diameter of particles measured on this line. 13 Variability of Particle Measurements Figure S14: (a) Black carbon mass emission index, EI(BC), (b) number emission index, EIn, and (c) geometric mean diameter as measured by LII, CPCs and DMS/SMPS systems, respectively. Data represents individual measurements (~1 Hz) from each system in order to demonstrate the repeatability of the measurements. 14 Three representative particle size distributions are shown in Figure S15 for representative thrust settings. Two settings were during pilot combustion (6% and 17%) and one during double annular combustion (31%). The measured size distributions are all well characterized by single-mode log-normal distributions, where the individual scans do not deviate greatly from the mean measured distribution over the entire setpoint duration. The particle size distributions corrected for line loss are a factor of 1.6 - 2.1 greater in total concentration than the measured distributions, and the geometric mean particle size is 9-18% smaller for corrected versus measured geometric mean particle diameter. Figure S15: Particle mobility distribution as measured by nano-SMPS on the SAMPLE line for (a) 6%, (b) 17% and (c) 31% thrust settings for 30th April. Colored lines represent individual scans, solid black line represents average of individual scans and the dashed line represents the line-loss corrected particle size distribution. 15 Measurement Uncertainty The measurement uncertainties for the various instruments are given in SI Table 1 below. A further discussion of the nature of the uncertainty and verification method is given thereafter. SI Table 1: Measurement uncertainties for various instruments and physical parameters in this study. Derived uncertainty determined in accordance with Abernethy et al. (1985). 1 Correspondence with NRC Canada, as determined in APRIDE-4 and APRIDE-5 campaigns. (Symonds, 2010) 3 (Johnson et al., 2013) 4 (Johnson et al., 2013) 5 As shown in SI Table 3 below. 6 TSI and Grimm CPC product manual. 7 ISO 15900:2009 8 (Eggersdorfer et al., 2012) 9 References given in main text. 2 16 Condensation Particle Counter Measurement of particle size and concentration by differential mobility analysis (DMA) and condensation particle counting (CPC), respectively, are well-established, NIST-traceable methods. SI Table 1 shows the counting accuracy for CPCs in single counting and photometric modes for the model CPCs used in this study. Both manufacturer’s (TSI and Grimm) estimate accuracy is better than 10% when used in the single-counting mode, as was always the condition in this work. SI Table 2: Counting accuracy for CPCs in single-counting and photometric modes. The concentration range (in cm-3) provided by the manufacturer for single-counting mode is shown in parenthesis. 3775 3772 3010 5435 (5430?) single – count photometric 10% (0 - 50,000) 20% 10% (0 - 10,000) n/a 10% (0 - 10,000) n/a 10% (0 - 23,000) n/a Detailed studies of CPC calibration and accuracy have revealed that CPCs can be calibrated to levels much more precise than manufacturer’s estimates (Fletcher et al., 2009; Giechaskiel et al., 2010; Owen et al., 2012). For example, Owen et al. (2012) reported the relative expanded uncertainty of a CPC (coverage factor of k = 2) is 2.8% over the range of about 1 particle·cm−3 to 104 particles·cm−3. However, because CPC counting accuracy during our test campaign was conducted on a “spot-check” basis and no instrument-specific calibration curve was used, a more conservative standard uncertainty of 10% is most appropriate to apply to the concentration data herein. Differential Mobility Analyzer Standard sizing accuracy of the DMAs used was not explicitly provided by the manufacturer, although the 3081 and 3085 DMAs are ISO 15900:2009 compliant and it is stated that “rigorous peer reviewed uncertainty analyses have been performed indicating TSI’s DMA has a sizing uncertainty of approximately <2% (TSI, 2015)”. Vasiliou (2005) found that particle sizing of 3085 and 3081 DMAs was within the NIST-stated uncertainty range of the measured PSL particles. Mulholland et al. (2006) conducted a detailed uncertainty analysis for 100 nm and 60 nm standard reference PSL particles. They found relative expanded uncertainty of a 3081 DMA (coverage factor of k = 2) for a particle diameter of 101.8 nm to be 1.1 nm and 0.63 nm for a 63.9 nm particles. For smaller particles like the gas turbine soot aggregates measured here, the uncertainty will be slightly higher due to the effects of Brownian motion on the DMA transfer function. Kim et al. (2005) estimated the uncertainty for a 20 nm particle in a 3085 DMA to be ~ 0.7%. As the case with the CPCs, the manufacturer’s estimate of uncertainty (i.e. ~2%) appears to best represent the uncertainty of the particle sizes measured for this work. Laser Induced Incandescence As determined by the LII mass calibration method, the uncertainty of the measurement ties to the uncertainty of the calibration and then any uncertainties associated with differences between the particulate measured during calibration and that measured in engine tests. This latter uncertainty is an ongoing topic in the SAE-E31 community (VARIANT and MANTRA 17 campaigns, where instrument response differences with source have been observed). The uncertainty of the calibration is principally defined by the uncertainty of the NIOSH method which is suggested to be 16.7% from the NIOSH 5040 standard document. NPL suggests a higher value – 25% (95% confidence limit). From engine test comparisons (A-PRIDE4, APRIDE5), differences less than +/- 25% have been observed, which is the uncertainty level assumed in this study. Centrifugal Particle Mass Analyzer The uncertainty in the CPMA measurements were explored in detail by Symonds et al. (2013) and they determined a 2.8% uncertainty in the CPMA mass-to-charge setpoint at a 95% confidence interval (CI). The uncertainty in the standard DMS500 measurements were determined by Symonds (2010) to be 10% in mobility size at a 95% CI. Differential Mobility Spectrometer The DMS500 used in the CPMA-DMS system was modified and therefore the standard DMS500 uncertainties do not apply. The uncertainty of the modified DMS was determined by comparing it against a DMA and knowing the uncertainty in the DMA measurements (Mulholland et al., 1999). The results of this comparison was indicated that the mDMS has a mobility diameter sizing uncertainty of 3.04%-3.35% (at a 95% CI) depending on the particle mobility size. Further details regarding this comparison and the uncertainty of the mDMS are shown in the supplemental information of Johnson et al. (2013). Johnson et al. (2013) also found the uncertainty in the effective particle density measured using the CPMA-mDMS system was 9.5%-10.4% (at a 95% CI) depending on the particle mobility size. Transmission Electron Microscopy Average standard error for TEM primary particle diameter (either dp or dva) is ~2 nm (1.6 nm). Measured standard deviations and standard errors are reported in SI Table 3. Imaging magnification varied from 100k to 300k for the whole range of test points, which typically corresponds to ~1 to 0.6 nm/pixel resolution. The image processing procedure takes into account the exact magnification. Although these changes in resolution may influence the results slightly, almost all images taken from ONE test point (RPM/load) are at the same magnification. 18 SI Table 3: Standard deviations and standard errors for TEM measured primary particle diameters. RPM 1000 1600 1845 2600 2630 2810 3025 3400 3700 3950 4500 STD 6.1 8.08 16.93 6.13 9.76 4.68 12.4 5.45 5.79 4.35 5.99 d p [nm] STE 0.95 1.2 4.52 2.5 1.45 0.71 2.77 0.82 1.14 0.87 1.11 AVG 7.79 1.64 d va [nm] STD STE 6.05 0.95 8.34 1.24 15.19 4.06 6.05 2.47 9.86 1.47 4.74 0.72 12.64 2.83 5.34 0.8 5.97 1.17 5.83 1.17 6.11 1.13 7.83 1.64 19 SI Bibliography Abernethy, R., Benedict, R., Dowdell, R. (1985). ASME measurement uncertainty. Journal of Fluids Engineering. 107:161-164. Bescond, A., Yon, J., Ouf, F., Ferry, D., Delhaye, D., Gaffié, D., Coppalle, A., Rozé, C. (2014). Automated determination of aggregate primary particle size distribution by TEM image analysis: application to soot. Aerosol Science and Technology. 48:831-841. Dastanpour, R., Rogak, S. N. (2014). Observations of a Correlation Between Primary Particle and Aggregate Size for Soot Particles. Aerosol Science and Technology. 48:1043-1049. DeCarlo, P. F., Slowik, J. G., Worsnop, D. R., Davidovits, P., Jimenez, J. L. (2004). Particle Morphology and Density Characterization by Combined Mobility and Aerodynamic Diameter Measurements. Part 1: Theory. Aerosol Science and Technology. 38:1185-1205. Eggersdorfer, M. L., Pratsinis, S. E. (2012). The Structure of Agglomerates Consisting of Polydisperse Particles. Aerosol Science and Technology. 46:347-353. Grishin, I., Thomson, K., Migliorini, F., Sloan, J. J. (2012). Application of the Hough transform for the automatic determination of soot aggregate morphology. Applied optics. 51:610-620. Johnson, T. J., Olfert, J. S., Symonds, J. P. R., Johnson, M., Rindlisbacher, T., Swanson, J. J., Boies, A. M., Thomson, K., Smallwood, G., Walters, D., Sevcenco, Y., Crayford, A., Durdina, L., Bahk, Y. K., Brem, B., Wang, J., Dastanpour, R., Rogak, S. N. (In Press 2014). Effective density and mass-mobility exponent of aircraft particulate matter. Journal of Propulsion & Power. Johnson, T. J., Symonds, J. P., Olfert, J. S. (2013). Mass–mobility measurements using a centrifugal particle mass analyzer and differential mobility spectrometer. Aerosol Science and Technology. 47:1215-1225. Mulholland, G. W., Bryner, N. P., Croarkin, C. (1999). Measurement of the 100 nm NIST SRM 1963 by differential mobility analysis. Aerosol Science and Technology. 31:39-55. Office of Aviation Research, (2006). Determination of Fuel/Air Mass Ratios for Jet Fuels at Their Flash Point Temperatures, in: Transportation, U.S.D.o. (Ed.). DOT/FAA/AR-02/96, http://www.tc.faa.gov/its/worldpac/techrpt/ar02-96.pdf. Symonds, J. (2010). Calibration of Fast Response Differential Mobility Spectrometers. National Physical Lab., Metrology of Airborne Nanoparticles, Standardisation and Applications, London. 20 21