Template for Electronic Submission to ACS Journals

advertisement
1
Supporting Information for
2
A calibrated Graphene-based chemi-sensor for sub-ppm NO2 detection operating at room
3
temperature
4
Filiberto Ricciardella,* Ettore Massera, Tiziana Polichetti, Maria Lucia Miglietta and
5
Girolamo Di Francia
6
ENEA Research Center, Piazzale E. Fermi, 1, Portici (Napoli), I-80055, Italy
7
Corresponding author: filiberto.ricciardella@enea.it
8
9
Optimization of device conductance value
10
The chemiresistor conductance value has been optimized in order to have a conductance
11
around 10-5 S by varying the volume of the dispensed suspension until the achievement of the
12
proper film conductivity: This particular conductivity range allowed to exploit at the best the
13
electronics of our experimental setup.
14
After preparing different graphene dispersions by chemical exfoliation, few microliters of
15
suspension were deposited by drop-casting onto quartz substrate and the sheet resistance was
16
measured through four-probes instrument (NNPSON RESISTAGE RG8).
17
The flake average thickness in the deposited film was estimated equal to 200 nm through
18
profilometric analysis (KLA TENCOR P-10) so that a conductivity value ranging between
19
104÷105 S/m corresponds to a sheet resistance of drop-casted films in the range of kΩ/sq.
20
1
21
22
Seebeck effect measurement
23
Exploiting the thermoelectric effect, this qualitative method allows to identify the type of the
24
carrier in semiconductors. In our case, this analysis appears to be particularly useful because the
25
interaction between graphene and analyte is strongly influenced by the majority carriers.1
26
On alumina strip (3 x 1 cm2), two pads at a distance of about 5 mm were created by spreading
27
drops of silver paste. Few microliters of graphene dispersion were deposited between the
28
metallic contacts. Once one pad was heated at 450°C, a positive voltage equal to 3mV was
29
measured between the electrodes. This behavior, typical of the p-type doped material, has been
30
inspected for each graphene dispersion, confirming the type of the material doping.
31
32
33
34
35
2
36
Alumina transducer details
37
The sketch of the transducer used as sensor substrate is depicted in Figure S1.
38
39
Figure S1. Sketch of transducer used as substrate for sensor
40
41
All the finger dimensions have been designed by ourselves and customized by 3M taking into
42
account our need and the structure of the test chamber. The graphene film covers the gold
43
interdigitated electrodes (IDE) for an area of about 49 mm2. Volt-amperometric measurements
44
were performed by applying the two probes on the larger pads of the transducer. The probes have
45
a diameter of about 500 m.
46
47
48
3
49
Gas Sensor Characterization System and test protocols towards NO2
50
51
Figure S2. Picture of Gas Sensor Characterization System (left) and corresponding circuital
52
scheme (right).
53
The device is located in a stainless steel test chamber (see fig. S2) placed in a thermostatic box.
54
The chamber is provided of an electrical grounded connector for bias and conductance measure
55
as shownon the right in fig. S2 .
56
A constant flow (500sccm) of the gas carrier, i.e., synthetic air crosses the test chamber. The
57
carrier can be properly humidified through a water bubbler placed in a thermostatic bath. In this
58
environment, characterized by controlled temperature and humidity, the conductance value of the
59
device in its equilibrium state is firstly measured (baseline); after that, an intentional disruption
60
of the equilibrium state is produced by introducing Nitrogen dioxide in a controlled amount and
61
by mixing it with the gas carrier via pneumatic valves and through programmable Mass Flow
62
Controllers. In the field of environmental monitoring, the NO2 concentrations needed to be
63
detected are lower than few parts-per-million (ppm), this poses a fundamental issue in
64
reproducing as much as possible these conditions in lab. In addition, the reproducibility of these
65
levels of concentration in the lab equipment appears no immediate since the pipes and the core of
4
66
test chamber is made of electro-polished stainless steel. This means that the NO2 molecules can
67
spread over and adhere also to the chamber walls, reducing the effective analyte concentration
68
interacting with the sensor. In order to minimize the effects of these limitations, within 30
69
minutes before the measurements, the maximum available concentration of NO2 is fluxed into
70
the chamber for 15 minutes through a protocol controlled via software. In order to check the
71
degree of agreement between the gas concentrations set in the protocols and those present into
72
the chamber, tests have been performed through an FTIR (THERMO ANTARIS IGSS) equipped
73
with a cryogenic detector and a cell having an optical path equal to 10 m. All these operations
74
guarantee the reproducibility of the condition of the test chamber. Hardware and software
75
implemented on a work station allow to control and record environmental parameters, device
76
bias and output signal, making possible to perform customizable automated tests on devices
77
(Protocols).
78
79
In Figure S2 a picture of the stainless steel testing chamber (on the left) and the circuital
scheme for the electrical characterizations (on the right) are reported.
80
In the field of environmental monitoring, the NO2 concentrations needed to be detected are
81
lower than few parts-per-million (ppm), this poses a fundamental issue in reproducing as much
82
as possible these conditions in lab. In addition, the reproducibility of these levels of
83
concentration in the lab equipment appears no immediate since the pipes and the core of test
84
chamber is made of electro-polished stainless steel. This means that the NO2 molecules can
85
spread over and adhere also to the chamber walls, reducing the effective analyte concentration
86
interacting with the sensor. In order to minimize the effects of these limitations, within 30
87
minutes before the measurements, the maximum available concentration of NO2 is fluxed into
88
the chamber for 15 minutes through a protocol controlled via software. In order to check the
5
89
degree of agreement between the gas concentrations set in the protocols and those present into
90
the chamber, tests have been performed through an FTIR (THERMO ANTARIS IGSS) equipped
91
with a cryogenic detector and a cell having an optical path equal to 10 m. All these operations
92
guarantee the reproducibility of the condition of the test chamber.
93
94
An example of standard protocol for sensing measurement at fixed NO2 concentration is set as
follows:
95
•
1200 seconds in carrier gas at constant relative humidity (RH) and temperature
96
•
15 seconds exposure to 4300 parts-per-billion (ppb) of NO2
97
•
225 seconds exposure to 350 parts-per-billion (ppb) of NO2
98
•
180 seconds at zero gas flow
99
•
1200 seconds of flushing with carrier gas.
100
The first step, also called baseline, is required to stabilize the conductance value and to
101
determine the starting point of the response at the gas inlet. Since the test chamber volume is
102
about 40 cl, a “double step” for the gas inlet is needed to more rapidly reach the steady state
103
conditions in terms of gas concentrations: in the first 15 sec the gas flow is set to very high
104
concentration value; in the subsequent 225 sec the gas flow is set to the target concentration. The
105
flushing step is used to recover the conductance to the initial condition.
106
107
The typical conductance dynamic response of the sensor upon the just described run is showed
in Fig. 4 in the paper; in that case, the carrier gas is N2 and the RH is set at 50%.
108
The standard protocol for the sensing measurements referred to Fig. 5 in the main text consists
109
of different pulses of NO2 followed by short recovery steps, operating in conditions of humidity
110
and carrier gas similar to those reported in the last runs. For all the concentrations, except for the
111
exposure to 50 ppb, the protocol is set as follows:
6
112

2 minutes exposure to NO2
113

4 minutes flushing with carrier gas
114
115
In the case of 50 ppb, the only change regards the exposure time window that requires 4
minutes to achieve an appreciable conductance variation.
116
Likewise, for the sensor calibration, the test protocol is constituted by sequential steps at
117
different analyte concentrations: after the initial acquisition of the baseline, the device was
118
exposed to NO2 for 2 minutes and left to recover in inert gas flow for 28 minutes assisting the
119
partial desorption of the gas and avoiding the poisoning of the sensing film.
120
121
7
122
Derivative Method and calibration protocol
123
The employed method for the sensor analysis and calibration is focused on the derivative of
124
the signal output. In particular, the derivative is calculated as the average of the incremental ratio
125
according to the following relation implemented by OriginPro® software:
126
𝒇′ (𝒙𝒊 ) =
𝟏 𝒚𝒊+𝟏 − 𝒚𝒊 𝒚𝒊 − 𝒚𝒊−𝟏
[
+
]
𝟐 𝒙𝒊+𝟏 − 𝒙𝒊 𝒙𝒊 − 𝒙𝒊−𝟏
127
where xi and yi are generic time instant and value conductance, respectively.
128
When the sensor is exposed to several sequential steps at different analyte concentrations (see
129
Figure 5 in the text) for the same exposure time, the effects of the algorithm on the signal is well-
130
rendered. The linear trend of the maxima with the concentration leaps out, clearly highlighting an
131
increase of one order of magnitude upon NO2 exposure passing from 100 ppb to 1000 ppb. This
132
biunivocal corrispondance has been exploited to the aim to calibrate the sensor according to a
133
above mentioned calibration protocol.
134
The error bars in the calibration plot (Figure 6 in the paper) have been calculated by
135
considering the difference between two subsequent points in the neighborhood of the maximum
136
of each derivative peak.
137
The employed analysis method allows to compare devices fabricated starting from different
138
batches and/or having a large different parameters such as initial conductance and response to the
139
analyte. At the same time, the method reveals so versatile that it can be extended to all devices
140
affected by slow and incomplete recovery.
141
142
8
143
State of the art for the sensors operating at Room Temperature
144
A brief comparison between our sensor results gas sensor devices operating at Room
145
Temperature and based on sensing materials other than graphene, is reported herein.
146
147
-
Sensitive film based on nanostructured metal oxides
148
Metal oxides are considered the best option for this kind of application and SnO2 is one of the
149
most investigated material. This oxide is a very good choice for sensors working at high
150
temperatures (>350°C), but when RT scenarios have to be considered, their performances
151
strongly get worse and different approaches (i.e. nanostructured morphologies) have to be
152
considered. For instance, Khuspe et al. show that a fairly high response (about 20%) can be
153
obtained although at much higher concentration (100 ppm).2 However, the device shows a very
154
short time of response (7s) but the operating temperature is still fairly high, around 200°C . In
155
comparison, our device is certainly slower but it is much more sensitive (10 times at least) and it
156
truly operates at RT.
157
ZnO is another widely investigated metal oxide for NO2 detection. Shishiyanu et al. report on
158
a doping process also finalized to decrease the device operating temperature. They find a
159
sensitivity around a few %/ppm at RT (therefore again our device seems more promising) with a
160
response time in the order of minutes (comparable to our device).3
161
162
-
Sensitive film based on an organic material.
163
Some conjugated polymers have been investigated as possible sensitive materials exploiting
164
the “doping” effects that some chemicals can exhibit during the molecule-polymer interaction. In
165
the recent paper, Ji et al., it is for instance shown that a high change of the relative response at
9
166
RT can be observed (about 20%) but at 5 ppm of NO2, with a time of response that strictly
167
follows the gas pulse (tens of minutes).4 Also in this case our device seems to be more
168
promising, at least in terms of response time.
169
170
-
Sensitive film based on Carbon Nanotubes (CNT).
171
CNTs based sensors show a very interesting response to NO2. The most widely cited papers, in
172
this respect, is the paper by L. Valentini et al. Their device shows a 6% response to 100 ppb and
173
are sensitive to the gas up to 10 ppb. However the device operates at a fairly high temperature
174
(165 °C) and, more importantly, the characterizations have been performed in dry air, that is, far
175
from normal operating conditions.5 Our device operates, on the contrary, at relative humidity
176
concentrations similar to those normally observed in urban environments.
177
178
179
180
10
181
182
183
184
185
186
REFERENCES
1. F. Schedin, A. K. Geim, S. V. Morozov, E. H. Hill, P. Blake, M. I. Katsnelson, and K. S.
Novoselov, Nature Mat. 6, 652 (2007).
2. G. D. Khuspe, R. D. Sakhare, S. T. Navale, M. A. Chougule, Y. D. Kolekar, R. N. Mulik, R.
C. Pawar, C. S. Lee, V. B. Patil, Ceram. Int. 39, 8673 (2013).
187
3. S. T. Shishiyanu, T. S. Shishiyanu, O. I. Lupan, Sens. Actuators B 107, 379 (2005).
188
4. S. Ji, H. Wang, T. Wang, D. Yan, Adv. Mat 25, 1755 (2013).
189
5. L. Valentini, I. Armentano, J. M. Kenny, C. Cantalini, L. Lozzi, S. Santucci, Appl. Phys.
190
Lett. 82, 961 (2003).
191
11
Download