Center for Advanced Brain Imaging Georgia State University and Georgia Institute of Technology 831 Marietta St, Atlanta GA 30332, USA phone (404) 385-8619 :: fax (404) 385-8620 www.cabiatl.com Screening form decision tree. The screening form is designed to identify whether items may be a hazard or cause artifacts in a 3T scanner (this is referred to as being MRI safe and MRI compatible, respectively). Therefore, the list below outlines our standard operating procedure for interpreting items on the checklist. For more details, refer to www.mrisafety.com Metal fragments in your eyes -Exclusion Are you a metal worker or have you worked with metal lathes? -Check history Cardiac pacemaker -Exclusion Aneurysm clip -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Note that modern clips are typically MRI safe, but we wish to exercise extreme caution. Any type of internal electrode(s) Pacing wires, Cochlear Implant, etc... -Exclusion Implanted insulin pump -Exclusion Swan-Ganz catheter -Exclusion Halo vest or metallic cervical fixation device -Exclusion Hearing aid -Exclusion if not removable, if removable, technologist and scientist should ensure that residual hearing is sufficient for task, and whether the participant is comfortable without normal auditory communication during scan (e.g. they might not be able to hear technoligist between scans) Any type of intravascular coil, filter or stent -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Implanted drug injection device -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Any type of foreign body, shrapnel or bullet -Unknown metals will be exclusion Heart valve prosthesis -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Any type of ear Implant -Exclusion Penile prosthesis -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Any type of surgical clip or staple -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Vascular access port 1 -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Intraventricular shunt -Exclusion unless surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Artificial limb or joint -Exclusion unless removable or surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Dentures -Typically MRI safe but not MRI compatible: nonferrous, but can cause substantial image artefacts. Ask participants to not wear their dentures. Diaphragm (in place), IUD -Exclusion if in pace. Typically MRI safe, but not all varieties have been tested (http://www.mrisafety.com/safety_article.asp?subject=35). Latex allergies -Inform participant that panic button uses latex, wrap panic button with cloth. Neurostimulator -Exclusion Wire mesh -Exclusion unless removable or surgeons note is available specifying a make and model listed as 3T safe on mrisafety website. Any type of electronic,mechanical or magnetic implant -Exclusion Any type of implant held in place by a magnet -Exclusion Any implanted orthopedic items (e.g. pins,rods, screws, nails, clips, plates, wire, etc...) -MRI technician records detailed information about model and location of implant. Implanted cardiac defibrillator -Exclusion Tattoo or tattooed eyeliner -Tattoosmay cause artifacts and have very low risk of irritation and warming. Participants with large tattoos on the head will be excluded, people with tatttoos at other locations will be warned of potential heating... Amateur or prison tattoo -Exclusion Pregnancy or suspected pregnancy: -Exclusion Surgery with implant: -Collect verbal history and determine if any exclusion checkboxes described above were incorrectly marks Injured by foreign object -Collect verbal history – exclude if even remote chance of metal fragment Injury to eyes -Collect verbal history – exclude if even remote chance of metal fragment 2