Diagnostic X-ray Imaging 1. Historical 8 Nov 1895 - x-rays discovered by Röntgen. (Awarded 1901 Nobel prize for physics). 13 Jan 1896 - first clinical use. X-ray photograph of a woman's hand produced in Birmingham. During 1896 - used in high doses for the treatment of Breast Cancer. Adverse affects to patient and doctor were observed. 1913 - The first mammogram taken. Technique essentially ignored for decades. 1917 - Radon presented the concept of Computed Tomography (CT). 1972 - First clinical CT (or CAT) scanner produced by Hounsfield. (Awarded 1979 Nobel Prize for Medicine.) 1 2. Production of x-rays – Simple X-ray generator Electrons accelerated through large potential before hitting the anode target. Electron K.E. partially converted into E.M. radiation. X-rays emitted in all directions greatest intensity at right angles to the electron beam. Conversion process is very inefficient: 99% conversion to heat. Anode is usually rotated at high speed to help heat dissipation. 2 3. X-ray spectra Firstly, the electron deceleration which occurs close to nuclei in the target produces a wide continuous spectrum of x-rays (Bremsstrahlung or 'white x-rays') The variation in the intensity of the emitted x-ray photons as a function of photon energy can be explained as follows. First, imagine a very thin anode, and consider the production of X-rays, not the X-rays that finally emerge. Consider the intensity of X-rays produced in a small energy range E to E+dE, this will be equal to the number of photons/m2/sec multiplied by the photon energy E. Fewer high energy photons are produced but their energy is higher and the product is constant. Thus for a thin anode we would have (a) overleaf. 3 A thick anode may now be thought of as composed of a large number of thin layers. Each will produce a similar distribution to that shown in (a), but the maximum photon energy will gradually be reduced because the incident electrons lose energy as they penetrate the anode material. Thus the composite picture for X-ray production might be as shown in (b). However, before the X-rays emerge, the intensity distribution will be modified in two ways. First, X-rays produced deep in the anode will be attenuated in reaching the surface of the anode and secondly X-rays will be attenuated in penetrating the window of the X-ray tube. Both processes reduce the intensity of the low energy radiation more than that of the higher energies so the result is the solid curve shown in (c). In the absence of further filtration (see later) the X-ray energy corresponding to maximum intensity will be about one third of the highest energy X-ray photons. 4 Superimposed on the continuous spectrum are sharp characteristic lines. Caused by ejection of K and L shell electrons followed by outer shell electrons filling the vacancy which was created. 5 3.1 Effect of tube voltage on x-ray spectrum Maximum x-ray energy corresponds to full deceleration of the electron: Emax = eVmax = hfmax = hc / λmin so, λmin = hc / (eVmax) E(keV)=1.24/ λ(nm) (E=13keV, λ≈0.1nm) As tube voltage V is increased: spread of wavelengths increases intensity increases (total intensity V2) peak in intensity shifts to higher energy Effect of tube voltage on x-ray spectrum 6 3.2 Effect of tube current on x-ray spectrum As the tube current i is increased: rate of production of electrons at the cathode is increased intensity increases (total intensity i) maximum energy remains unchanged intensity profile remains the same Effect of tube current on x-ray spectrum 7 3.3 Effect of target material on x-ray spectrum Changing the target material changes the atomic number, Z: x-ray intensity changes - the probability of a collision and so intensity Z changes the characteristic lines Tungsten (Z = 74) is almost always used as a target material: reasonably high Z high melting point (3650 K) - vital because of heat production 8 4. Attenuation of x-rays As an x-ray beam propagates, photons are scattered out of the beam: hence the beam is attenuated. Number of photons scattered per unit beam area in distance Δx is ΔΦ = ΦσNΔx N is no. of nuclei per unit volume, σ is the "scattering cross-section" - fractional beam area that interacts Φ is the fluence (no. of photons per unit area). Intensity of x-ray beam I is rate of energy incident per unit area, so is proportional to Φ. Change in intensity over distance Δx is ΔI = -lσNΔx. 9 Integrating gives logel = -σNx + k. Let l0 be intensity when x = 0, so k = loge l0 So l = l0e-σNx Define the linear attenuation coefficient as μ =σN, so I = I0e-μx. N and μ, both related to the density ρ of the medium: N = NAρ/M , NA is Avogadro's number and M is the molecular or atomic mass, So μ = σNAρ/M So high density media yield high μ values. μ also depends on Z through σ and M. Mass attenuation coefficient, μ/ρ = σNA/M, depends on Z (through M) and photon energy (through σ ). σ also depends on Z 10 4.1 Attenuation mechanisms Attenuation = Absorption + Scatter Dependent on incident photon energy (E). Medical imaging requires best contrast and least damage. (i) Simple scattering photon energy << electron binding elastic collision μ / ρ Z2 / E When X-rays pass close to an atom, they can excite electron vibrations. The process is one of resonance, such that the electron vibrates at the same frequency as the incident X-ray photon. This is an unstable state and the electron quickly reradiates this energy in all directions at exactly the same frequency as the incident photons. The process is one of scatter and attenuation without absorption. The electrons that vibrate in this way must remain bound to their nuclei – thus the process involves bound electrons. Binding energy increases with Z. 11 (ii) Photoelectric effect photon energy > binding energy all photon energy given to an inner electron which is ejected. Characteristic x-ray emitted ejected electron ionizes atoms along its path until its kinetic energy is dissipated μ / ρ Z3 / E3 (crude approximation) 12 (iii) Compton Scattering photon energy >> binding energy photon energy transferred to an outer 'free' electron which is ejected ejected electron energy depends on angle through which incident photon scattered (size of arrow in figure below) photon continues with reduced energy electron dissipates energy by ionizing atoms along its path μ / ρ independent of Z and falls slowly with E 13 (iv) Pair production Very high incident photon energies (>1.02 MeV) Pair of anti-particles (electron+positron) formed in nuclear coulomb field μ / ρ Z2 and rises very slowly with E Thus overall attenuation is a combination of (i) to (iv) thus for a path length x through an object I = I0 e-μ1 x e-μ2 x e-μ3 x e-μ4 x I = I0 e -(μ1 +μ2 +μ3 +μ4) x 14 4.2 Relative importance for medical imaging Left – Relative attenuation coefficients between materials. Right – Attenuation mechanisms in water (similar to soft tissue) Contrast decreased as incident photon energy is increased Best contrast: photoelectric effect (Z3) Scattering causes image blurring High energy attenuation causes damage (through ionization and heating) 20 - 100 keV used for diagnostic radiology (photoelectric effect and Compton scattering) > 100 keV used when contrast between bone and surrounding structure not required (e.g. imaging lungs) High Z elements may be used to improve contrast e.g. injected NaI to investigate circulatory system or barium sulphate in gastro-intestinal system 15 16 17 18 4.3 Filtration of x-ray sources Want maximum intensity in useful energy range. high energy extreme controlled by varying tube voltage Low energy photons absorbed in or near skin: no effect on contrast just an increased dose Iow energy: photoelectric effect dominates AI commonly used as a filter: μ/ρ proportional to Z3/E3 so low energies preferentially absorbed. (Al 1-3mm thick) overall reduction of x-ray intensity; peak shifted to higher energy ('hardening' beam) 19 4.4 Dose and exposure X-rays cause damage to tissue through ionization, so x-ray dose must be kept small. Dose defined as: D = Energy absorbed in mass mass Units: Gray (Gy); 1 Gy = 1 J kg-1 Old unit: rad (radiation absorbed dose) 1 Gy = 100 rad Absorbed dose difficult to measure, so often use measured Exposure to determine dose. An Exposure of 1 C kg-1 in air produces 1/e electrons per kg of air (C Coulomb). Dose (Gy) f x Exposure (C kg-1) [f = 34 in air] In soft tissue f = 34 to 40, higher for bone. Depends on photon energy. Old unit: Röntgen (R); 1 C kg-1 = 3876 R. 20 5. X-ray photographs Advantages: Cheap Equipment easy to maintain and use Readily available Relatively safe 21 5.1 Blurring (i) Scattering Scattering leads to loss of spatial resolution. scattered x-rays eliminated using a grid of lead strips between patient and film. strips angled to receive only direct beam strips continually moved to avoid being imaged 22 (ii) Patient movement Involuntary movement of organs can lead to blurring overcome by using short exposure times. to maintain exposure must use high intensities (increased dose rate) Often image intensifier used instead. Fluorescent screens placed either side of the film (e.g. zinc cadmium sulphide doped with silver) Metal backing plate used to avoid backscatter and x-ray leakage 10-40 intensification factor possible Some definition lost by diffusion of fluorescent light 23 (iii) Focal spot size penumbra shadow formed if large effective focal spot used (a) angling anode gives smaller effective focus (penumbra decreased) (b) patient-film distance also minimised 24 5.2 Limitations Features behind e.g. bones, are difficult to Image photograph 2-D projection of a 3-D object cannot establish where the feature is in the path the contrast is a sum of attenuations in the path (so possibility of ghost objects) Can get around these problems by taking two projections 25 6. Linear Tomography Tomography is a method of viewing a slice through the body. (Greek Tomos - slice) source and film moved in opposite directions at the same speed any point in the image plane will appear in the same position on the film all other points image at different positions on film as source and film are moved and thus image blurred (e.g. point closer to source than image plane - dotted lines indicate path of X-ray photons) depth resolution of approx. 1 mm possible 26 6.1 Selection of image plane Consider a rod connecting source (S) and film cassette (F), with pivot, P, at a variable point as in diagram. The slice through body in focus changes as pivot moves up and down. The thickness of the cut is controlled by the size of tomographic angle θ. The greater the value of θ the thinner the slice will be. This is because even structures very close to the pivot are blurred, since their image moves significantly on the film. If θ=0 it is a conventional radiograph. If the movement runs parallel to an elongated body structure (e.g. femur) there will be comparatively little blurring of the long edges even when not in the pivot plane. A circular instead of linear movement can help in such cases. 27 7. Computed Axial Tomography (CT or CAT) Radon showed that it is possible to build an image of an unknown object given an infinite number of projections through object CAT relies on obtaining many projections from different axial positions. Each ray in a linear projection provides an intensity data point that depends on the cumulated absorption coefficient. Distance between projections ~0.5 mm 28 7.1 Tomographic reconstruction Main type of tomographic reconstruction is filtered back projection. Integration of absorption coefficient (μ) over path, must be 'undone'. achieved by 'smearing' μ, over the path grid of pixels created, and value of μ from ray added to value of each pixel on line - repeated for each ray and each projection pixel values are divided by number of values of μ added, and converted to a grayscale for display each pixel actually a volume element (voxel), typically 1 to 10 mm deep 29 30 31 32 33 34 35 7.2 Scanner development Scanner design continually modified to improve image quality and increase scan speed. (i) First generation one ray source and one detector source and detector moved linearly to obtain line scan source and detector rotated in 1° intervals, and linear scan repeated. Repeated for 180°. Solid state scintillation, or argon ionization chamber detector used. Advantages: Excellent scatter rejection Disadvantages: Long scan time (up to 5 mins) 36 (ii) Second generation narrow fan source beam (10°) used. detector is a linear array of 30 detectors source and detector array are moved linearly source and detector are rotated in 10° intervals to cover the 180° arc. Advantages: Much shorter scan times (typically 20 sec) 3 times as many measurements, so improved image resolution. Disadvantages: Decrease in scatter rejection Each detector must be gain-balanced. 37 (iii) Third generation Larger fan angle to cover whole body. 750 detectors arranged along the circular arc source and detector array rotated to take a number of scans Advantages: No translation and many more detectors, so very fast scans (typically 5 sec). Disadvantages: Much increased cost 750 rotating detectors mean cable problems Detectors must be balanced 38 (iv) Fourth generation Complete ring of detectors (1000 detectors) Wide fan x-ray source rotated Advantages: Only source cabling needs to be rotated. Detectors are self-balancing: at some time each detector views a beam that does not pass through the body. Stable x-ray source required Increased scan speed (1 second) Disadvantages: Worse scatter rejection: each detector sees all beam directions. X-ray source must rotate on a ring closer to patient: decreased resolution (higher magnification). Can be overcome by bending detector ring out of the way (nutation). 39 40 (v) Fifth generation (cine-CT) Developed to achieve very fast scan times to monitor pulsating organs (e.g. heart or liver) No moving parts Complete ring of detectors (as in 4th generation) X-ray source: ring of tungsten around the patient used as target for an electron beam, which is directed by magnetic field. Advantages: 50 ms scan times. Disadvantages: Very low intensity x-rays, therefore poorer quality images. 41 42 43 44 Here the average attenuation along a particular direction is put in every cell along that direction. If there is already a value in the cell the new average for the new direction is added to the existing value. Thus the cells in the first column each start with the average vertical attenuation per cell along the first column (i.e. 20/3) followed by the average attenuations in the horizontal direction (20/3 for first row) and diagonal directions (25/3 for top left to bottom right cells and 5/1 for diagonal through just top left hand cell). The resulting attenuation coefficients for each cell calculated in this way are ringed. 45