CAP Accreditation Requirements for Validation of Laboratory Tests AMT 75th Educational Program and National Meeting Stephen J. Sarewitz, MD, FCAP July 9, 2013 www.cap.org v. 1.0 Biography Stephen J. Sarewitz, MD, FCAP • Board of Governors, College of American Pathologists • Member, Council on Accreditation, CAP • Former Chair, Checklist Committee, CAP Laboratory Accreditation Program • Former Chair, Area Committee on General Laboratory Practices, CLSI © 2013 College of American Pathologists. All rights reserved. 3 Objectives • Perform appropriate validation/verification studies in accordance with CAP requirements prior to implementing new tests, including FDA cleared/approved tests and laboratory-developed tests. • Understand validation requirements for FDA cleared/approved tests modified by the laboratory. • Understand the current status of validation requirements for laboratory-developed tests. © 2013 College of American Pathologists. All rights reserved. 4 Topics What this discussion covers: • Validation/verification of the test method itself What it doesn’t cover: other aspects of implementing new tests, such as: •Calibration or QC validation •Reagent handling •Documentation of test procedure •Training of personnel and competency assessment •Proficiency testing or alternative assessment •Sample handling © 2013 College of American Pathologists. All rights reserved. 5 Definitions Used in CAP Laboratory Accreditation Program • Validation: provision of objective evidence through a defined process that a test performs as intended [CLSI] • Verification: an abbreviated process to demonstrate that a test performs in substantial compliance to previously established claims • Sometimes the term “validation” is used to cover both of the above—can be source of confusion © 2013 College of American Pathologists. All rights reserved. 6 Types of Validation • Analytic validation: “…the process of assessing the assay and its performance characteristics and determining the optimal conditions that will generate a reliable, reproducible, and accurate…assay for the intended application.” --Required by CLIA and CAP Accreditation Checklists • Clinical validation: determining “the ability of a test to diagnose or predict risk for a particular health condition, measured by clinical (or diagnostic) sensitivity…, specificity, and predictive values.” --Not addressed by CLIA; in CAP only in Molecular Pathology and Microbiology (in Molecular section) Checklists (for nonwaived tests). • Clinical utility: “determining the net balance of health-related risks and benefits associated with the test’s use in practice [to] patient, family, healthcare organizations, and society.” --Subjective, complex, involves areas outside of laboratory diagnosis. Not in CLIA. Mentioned in CAP (MOL); no requirements. © 2013 College of American Pathologists. All rights reserved. 7 Analytic Validation Analytic Validation • Required by CLIA and CAP for non-waived tests (No CLIA reqs. for waived tests; CAP has very limited req. for waived tests) • CLIA and CAP requirements similar • Distinction between FDA-cleared/approved tests, and others: o For tests cleared or approved by FDA—verification required: ie, laboratory must show that it can obtain performance specifications comparable to those established by the manufacturer o Requirements the same for both moderate and high complexity tests approved/cleared by FDA o For tests not approved/cleared by FDA: laboratory must establish the performance specifications (= validating the test) © 2013 College of American Pathologists. All rights reserved. 9 CAP Definition of LDT A laboratory-developed test (LDT) is defined as follows - A test used in patient management that has all of the following characteristics: • The test is performed by the clinical laboratory in which the test was developed • The test is neither FDA-cleared nor FDA-approved, or is an FDA-cleared/approved test modified by the laboratory (sample types or the use of collection devices not listed in manufacturer instructions constitute modifications, by this definition) • The test was first used for clinical testing after April 23, 2003 © 2013 College of American Pathologists. All rights reserved. 10 Analytic Validation, continued • For FDA cleared/approved tests that have been modified by the laboratory, lab must validate the modification(s) or show that they don’t degrade test performance o Sample types and collection devices not listed in mfg labeling are considered modifications of the test (ex. body fluids) o Using a test in a patient population excluded in mfg labeling is a modification (ex. glucometer in critically ill patients) o Adjusting the reference range is not considered a modification o By definition, all modified FDA cleared/approved tests, and LDT’s*, are high complexity *Unless specifically categorized by FDA/CLIA as waived or PPM © 2013 College of American Pathologists. All rights reserved. 11 CAP Performance Specifications for Analytic Validation • Analytic accuracy (= bias) : systematic error • Precision : random error • Analytic sensitivity (limit of detection, LOD)* • Analytic specificity (interferences) • Reportable range • Reference range • In COM checklist (moved from GEN for 2011 edition) *Literally means ability of assay to detect a change in concentration of the analyte; assays with steeper calibration curves are more sensitive in this sense. However this is not the meaning of the term for purposes of validation/verification—LOD is what is meant. © 2013 College of American Pathologists. All rights reserved. 12 CAP Requirements…in COM Checklist (in GEN before 2011) Accuracy & Precision (COM.40300, ph II) Analytic sensitivity (LOD) (COM.40400, ph II) Analytic specificity (interferences) (COM.40500, ph II) Reportable range (AMR)(COM.40600, ph II) Reference range (COM.50000, ph II) FDA approved/cleared LDTs & modified FDA tests Verify mfger’s results Establish (= validation) Verify: manufacturer or literature documentation OK Reference literature or manufacturer documentation Verify* Establish Verify ** Establish** Establish; studies by manufacturer or in literature OK Establish* *Reportable range ( = AMR, generally) is the range of values that the method can directly measure without dilution or concentration, while meeting specifications for accuracy & precision --Details on establishing & validating AMR are in other checklists (ex. CHM, HEM, MOL) **In some cases labs may use manufacturer or literature data when verification/establishment of a reference range is not practical: ex. pediatric blood cell count / index parameters; therapeutic drug levels. © 2013 College of American Pathologists. All rights reserved. © 2013 College of American Pathologists. All rights reserved. 13 Analytic Validation – Some New Provisions in July 2013 Edition of COM Checklist • Validation studies must be performed in location where patient testing will be done o If instrument is moved, lab must verify that performance specifications (eg, accuracy, precision, etc.) were not affected by the move or new environment (change in ambient temp, humidity). (Intro.) • Multiple instruments of same make & model: each must be validated separately (Intro.) o Interpretation: − accuracy could be verified for 2nd instrument by comparison study with 1st instrument (15-20 samples) − No separate reference range study needed for 2nd instr., assuming comparison study showed absence of significant bias © 2013 College of American Pathologists. All rights reserved. 14 Analytic Validation – New Provisions in COM, continued • Reports for LDT’s contain a description of the method, statement that the test was developed by the laboratory, and appropriate performance characteristics (COM.40630, ph I) (similar item was in MOL & MIC; added to COM 2013) o Not a regulatory requirement, unlike FDA-mandated statement on reports of tests using ASR’s (see MOL, ANP, CYG, MIC) • Mfg instructions must be followed for FDA cleared/approved tests, OR lab must document validation of any modifications (COM.10600, ph II; moved from Procedure Manual section of COM) • For analytic interferences (COM.40500, II), lab has an appropriate plan of action when they are present © 2013 College of American Pathologists. All rights reserved. 15 CLIA --- Differences from CAP – Analytic Validation • For FDA approved/cleared tests: o No provisions on analytic sensitivity or analytic specificity in CLIA • No requirements for waived tests • No CLIA requirement for establishing or verifying performance specifications for any test system used prior to April 24, 2003. o CAP does NOT have any grandfather date; all non-waived tests subject to CAP analytic validation / verification requirements © 2013 College of American Pathologists. All rights reserved. 16 Waived Tests • No CLIA requirements • CAP: Verify reference range if practical POC.04525 (ph II): Reference intervals…are established or verified…. NOTE: If a…reference study is not…practical….then the POCT site should carefully evaluate the use of published data for its own reference ranges… If waived method is modified, then all CAP/CLIA validation requirements apply © 2013 College of American Pathologists. All rights reserved. 17 Validation/Verification Specifics • Neither CLIA nor CAP list specifics of validation/verification—numbers of samples, limits of acceptability, number of days required for studies, statistical analysis o Variability across laboratories in terms of environment, patient populations, use of tests – specifics thus need to be determined by laboratory director o Validation = more extensive study than verification • Exceptions: o Number of samples—HER2, estrogen receptor, progesterone receptor tests in breast carcinoma tissue o Concordance acceptability threshold – ER/PgR receptor in breast carcinoma o Number of data points for AMR (3 required) - No requirement for linearity studies per se….but AMR is “linearity lite” © 2013 College of American Pathologists. All rights reserved. 18 Some Guidelines Validation/Verification Guidelines FDA cleared/approved LDT Accuracy (bias)** At least 40 samples across AMR; could be > 100 Run study for 20 days Reportable range (= linear range) (=AMR)* 20-40 samples across AMR 2-3 samples at clinical decision points run daily for 5 days 3 points near low end, midpoint, and high end Reference range 20 samples 40-60 samples; 120 or more ideal Precision (random error) Same [See CLSI Guidelines (references at end of presentation)] *CAP checklist requirement [ex. CHM.13600 (II0]. AMR must be verified at least every 6 mos. **Recommend duplicate measurements of each sample (not in consecutive order) to detect errors, sample mix-ups, etc. © 2013 College of American Pathologists. All rights reserved. 20 Validation Guidelines---Using Statistics • Westgard: “Statistics are just tools for combining many experimental results…and summarizing all that data in just a few numbers…the statistics are used to make reliable estimates of the errors from the data….” • Precision (= random error) – actually, we measure imprecision o If determined within a run = repeatability; across multiple runs across multiple days = reproducibility. The latter is most reflective of actual lab practice. o Calculate standard deviation (SD) / coefficient of variation (CV) o Use F test to see if variance (=SD^2) of test method is statistically different from old method, or claim of manufacturer © 2013 College of American Pathologists. All rights reserved. 21 Validation Guidelines, continued • Accuracy / Bias (= systematic error): o Run comparison of methods study (test method, vs. reference method / lab’s previous method / manufacturer’s results, etc) − Line of best fit* (visually or using statistics program) gives linear regression equation Y = a + bX − Calculate correlation coefficient “r” – If r is high (>=.99), use regression line to find bias at analyte concentrations corresponding to critical decision points (ex. glucose: 126 mg/dL) *Line that minimizes the square of the distances of the data points to the line © 2013 College of American Pathologists. All rights reserved. 22 Validation Guidelines, cont. – If r < .975, regression equation not reliable; use paired t-test to determine if a bias is present at the mean of the data − Analytes with wide range (cholesterol, glucose, enzymes, etc) tend to have high r in comparison studies; analytes with narrow range (electrolytes) tend to have low r − r should not be used to determine the acceptability of a new method © 2013 College of American Pathologists. All rights reserved. 23 250 Y: new method mg/dL 200 Y = -7 + 1.18 * X 150 r = .99 100 Comp. method 50 50 0 0 50 100 150 200 250 New method Bias 52 +2 100 111 +11 150 170 +20 X: comparative method mg/dL High r……but significant proportional bias © 2013 College of American Pathologists. All rights reserved. 24 Validation Guidelines, continued • For AMR, 5 points is closer to ideal but 3 is sufficient (CHM.13600, ph II) o Separate linearity study not required: AMR is “linearity lite” • Reference range classically = central 95% of values for “normal” population. If 20 samples are run to verify mfg range, if <= 2 lie outside the mfg limits, then the mfg reference range is considered verified • Decision limits For some tests, reference range = a cut-off point or decision limit: ex., troponin, cholesterol. Studies for these tests would not be samples from a reference population, but rather studies of samples with levels on either side of the decision point, using clinical data or a comparative method. • Transference by calculation The lab may be able to”transfer” reference range from a previous method or a reference (or mfger) method (Westgard; CLSI guideline C28-A3). o If new method = Y and old method = X, Y(upper) = a + bX(upper), etc. o Limit transference to one change of methods, to prevent potentially amplifying errors by multiple transfer procedures (Westgard). o Perform 20-sample verification study if there is doubt re the reliability of the transfer © 2013 College of American Pathologists. All rights reserved. 25 Defining Allowable Total Error… Accuracy & Precision • FDA cleared/approved tests: = manufacturer specifications • For LDT’s: o Decision by Laboratory Director o Ideally, laboratory defines allowable total error in advance for the intended application of the test • Possible ways to establish allowable total error include: o CLIA acceptability limits for variability in proficiency testing results (403.909 – 403.959) − Ex. AST = +/- 20%, glucose +/- 6 mg/dL or 10% (greater), Na +/- 4 mmol/L, HDL-chol +/- 30% o Range of actual PT results for the analyte/method o Clinical decision points—literature, input from clinicians © 2013 College of American Pathologists. All rights reserved. 26 Allowable Error vs. Observed Error: Using Statistics to Help Analyze Accuracy & Precision (Systematic & Random Error) Observed total error should be less than allowable total error Observed (actual) total error can be defined as: bias + [3 x (SD of method)] (Westgard) © 2013 College of American Pathologists. All rights reserved. 27 Example – New Method for Serum Aspartate Aminotransferase (AST) • Total allowable error: o Lab determines it should be <= CLIA proficiency testing limit for AST, which is +/- 20% of target value • Lab’s new AST method: o Bias is + 2 U/L compared to reference method o Imprecision = standard deviation = +/- 3 U/L • Consider target value of 50 U/L: o Total allowable error = +/- 20% x 50 = 10 U/L o Observed total error = bias + (3 x SD) = 2 U/L + (3 x 3 U/L) = 11 U/L • Observed total error of 11 exceeds total allowable error of 10 so test method is not acceptable (Note: in this calculation, the absolute value of bias and SD are used) © 2013 College of American Pathologists. All rights reserved. 28 AMR and Calibration Reqs. in Coagulation (HEM chklst): Apply Only to Directly Measured Analytes • Method must be calibrated (HEM.38008, ph II) • There must be criteria for calibration verification (HEM.38006, ph II): o Changes of reagent lots for chemically / physically active components, unless lab can show that lot changer does not affect patient test results o Unacceptable QC (not correctable by other means) o Major maintenance or change of critical instrument component o When recommended by manufacturer o At least every 6 mos. • System must be recalibrated if cal ver fails to meet lab’s criteria (HEM.38007, ph II) © 2013 College of American Pathologists. All rights reserved. 29 AMR and Calibration in Coagulation (HEM Chklst): Apply Only to Directly Measured Analytes, continued • AMR is validated at low, mid- and high range (HEM.38009, ph II) o If materials used for cal/cal ver include the above values, then cal/cal ver also covers AMR validation o Revalidate every 6 mos, or following change in major instr. components or lots of critical reagents unless lab can show that change of lots does not affect results o AMR validation not required for methods that measure an analyte quantitatively or semiquantitatively and report a qualitative value based on a threshold • For result greater or less than AMR, numeric result not reported unless sample is diluted, concentrated or processed by a mixing procedure to bring result into the AMR (HEM.38010, ph II) © 2013 College of American Pathologists. All rights reserved. 30 Coagulation: Directly Measured Analytes • Often immunoassays (EIA, immunoturbidity, chromogenic) o Ex., protein C ag, free & total protein S ag, von Willebrand factor ag, LWM heparin, quant. ddimer • May include tests reported in % if traceable to a standard • Whether or not test involves a reference/standard curve not relevant • Does not apply to analytes measured by functional assay with results convertible to concentration by a formula © 2013 College of American Pathologists. All rights reserved. 31 Other Reqs. in COM Other Requirements in COM/GEN 1. Lab director or designee qualified as director must approve validation studies prior to reporting patient results (COM.40000, ph I). 2. Lab must list all LDT’s implemented in past 2 years for review by inspector (COM.40200, ph I). 3. Lab’s test methods, including performance specifications and supporting validation/verification data, must be available to clients and inspection team (COM.40700, ph II). Lab may require clients to treat information as confidential. 4. If method changes such that results may be significantly altered, the change must be explained to clients (COM.40800, ph II). 5. Lab evaluates reference range and takes corrective action as appropriate (COM.50100, ph II). a. Change of method or patient population 6. GEN .20377 (II) Retain validation records while method is in use and at least 2 years afterwards © 2013 College of American Pathologists. All rights reserved. 33 Other Requirements in COM, continued 7. Intermittent testing (ex. seasonal testing for influenza): o A test is considered to be taken out of production if 1) patient testing not offered; and 2) PT / alternative assessment is suspended o When suspended test is put back on-line: 1. PT or alternative assessment must be performed within 30 days prior to re-starting patient testing 2. Performance specifications are verified as applicable within 30 days prior to restarting patient testing 3. Competency is assessed for analysts within 12 mos. Prior to restarting testing (COM.40100, ph II) (Note: for tests for which PT is required, if a PT challenge is not offered in the 30 days prior to restarting testing, lab may do alternative assessment) © 2013 College of American Pathologists. All rights reserved. 34 Validation Items in Other Checklists – Examples (Full List at End of Presentation) • MOL.30900 (II) Validation studies include samples for each reportable genotype • NOTE: Assays for genetic disorders with a limited number of possible genotypes (e.g. hereditary hemochromatosis) should confirm the ability of the assay to detect these genotypes. Assays for genetic disorders with considerable allelic heterogeneity and/or significant numbers of private mutations (e.g. cystic fibrosis or hereditary nonpolyposis colorectal cancer) should confirm the accuracy of the methodology used to provide a high degree of assurance that the assay will detect targeted genotypes. Various sample types may affect…analytical performance....laboratories may need to establish sample-specific analytical and clinical performance characteristics. The number of samples depends on the intended use of the test. …laboratories are encouraged to review the cited references for guidance and provided confidence intervals to estimated performance characteristics. • MOL.31015 (II) Validation studies include expected specimen types (frozen, fresh, paraffin-embedded, blood, etc.) © 2013 College of American Pathologists. All rights reserved. 35 Validation Items in Other Checklists – Revision for 2013 In subsections “FDA cleared/approved non-amplification methods” and “FDA cleared/approved target & signal amplification methods & sequencing: • MIC.64770 and 64815 (II): If the laboratory tests sample types or uses collection devices other than those listed in the package insert, the laboratory performs validation studies to document adequate performance of the test. o NOTE: Results from tests performed on sample types not listed in the pkg. insert may be reported without complete validation only if…1) validation studies are ongoing but have not been completed; 2) the sample type is encountered rarely, precluding an adequate number for validation studies. Under these circumstances, the test report must include a disclaimer stating that the sample type has not been validated. © 2013 College of American Pathologists. All rights reserved. 36 Complications – Qualitative Tests • Performance specifications must be established/verified as applicable. Ex. reportable range may not be applicable. For tests with a cut-off, precision is considered to be variation in the test result near the cut-off point. • Can use kappa statistic to determine the degree to which agreement between 2 comparative methods is result of chance: kappa = [P(o) – P(e)]/[1-P(e)] where P(e) = agreement by chance and P(o) = observed agreement P(e) = [%test1(pos) x %test2(pos)] + [%test1(neg) x %test2(neg)] kappa values: 0 = chance agreement; 1 = perfect agreement .41 - .60 = moderate agreement .61 - .80 = substantial agreement See Viera AJ, Garrett JM. Fam Med 2005;37(5):360-363. © 2013 College of American Pathologists. All rights reserved. 37 Complications – “Gold Standard” Problem • New method is considered superior in analytic sensitivity to old established method—how to handle discordant results in method comparison study. o Troponin replacing CK-MB in dx of myocardial infarction o Molecular tests for microorganisms replacing culture or immunologic tests—Chlamydia, Trichomonas • Approaches: o Use clinical data (chart review) − Ex. clinical & EKG evidence of MI o Use reference method if available—must test all samples, not just discordant ones © 2013 College of American Pathologists. All rights reserved. 38 Complications -- Body Fluids -- New Checklist Item 2013 • COM.40620 (ph II) o Testing of body fluid specimens using methods intended for other specimen types (e.g. blood….) has been validated…for accuracy, precision, analytic sensitivity, analytic interferences, and reportable range. o NOTE: applies to B.F. testing offered as a routine, orderable test.: − Written procedure required − Lab director determines extent of validation studies − Blood performance specs. may be used if matrix effects can be excluded (by studies using mixtures of samples, spiking, dilution) − Reference range must be reported, unless result is reported as comparison to blood level (ex. pleural fluid protein, LD, cholesterol; ascitic amylase) – literature data OK − Alternative assessment or PT required © 2013 College of American Pathologists. All rights reserved. 39 New Body Fluid Validation Requirement, continued • NOTE: for clinically unique tests, validation may not be practical o Add comment: ex., “The reference range and other performance specifications have not been established for this body fluid. The test result must be integrated into the clinical context for interpretation.” © 2013 College of American Pathologists. All rights reserved. 40 Migration of Analytic Validation Requirements from GEN to COM Requirement GEN no. of citations COM no. of citations Approval of validation studies by director Accuracy/precision Analytic sensitivity Interferences Reportable range Data available to clients 94 18 9 9 16 1 176 65 29 24 33 23 Communicate significant method changes Establish/verify reference intervals Re-evaluate reference intervals as necessary Total 0 14 3 164 2 30 13 395 --First seven months of 2011: GEN contains requirements --First seven months of 2012: COM contains requirements --395/164 = 140% increase in citations at on-site inspections © 2013 College of American Pathologists. All rights reserved. 41 Clinical Validation CAP Clinical Validation Requirements—Limited to MOL / MIC* New text for 2013 in MOL, Intro. To Validation section: •…clinical validity, which includes .…clinical sensitivity, clinical specificity, positive and negative predictive values in defined populations or likelihood ratios, and clinical utility should … be considered, although individual laboratories may not be able to assess these parameters within their own patient population, especially for rare diseases. However, patients without disease can typically be tested to assess clinical specificity. If clinical validity cannot be established within a laboratory, it is appropriate to cite scientific literature that established clinical sensitivity and specificity. (Clinical utility: determining the net balance of health-related risks and benefits associated with the test’s use in practice [to] patient, family, healthcare organizations, and society.) *except for clinical claims © 2013 College of American Pathologists. All rights reserved. 43 Clinical Validation Requirements in CAP, cont. • MOL.31590 (II) Clinical performance characteristics of each assay are documented o Diagnostic sensitivity & specificity, positive & negative predictive values, likelihood ratios*, clinical utility* o May depend on clinical setting, genotype/phenotype associations when these vary with particular mutations or polymorphisms, & genetic/environmental/epigenetic factors affecting clinical expression of a genetic alteration • For FDA cleared/approved tests, literature or mfg data can be used • For laboratory-developed tests (LDT’s)—lab should perform studies in-house o Use clinical, biopsy & radiologic findings, other lab results* o Literature may be used for very rare conditions, or very common conditions for which clinical validity is well-established in literature o Establishing clinical validity may require extended studies…that go beyond the purview…of the individual laboratory (*New text, 2013) © 2013 College of American Pathologists. All rights reserved. 44 CAP Clinical Validation Requirements, continued • COM.40640, for 2013 (was in MOL) (II): All clinical claims made by the laboratory about an LDT are validated. • NOTE: Clinical claims may include statements about a test's diagnostic sensitivity and specificity, ability to predict the risk of a disease or condition, clinical usefulness, or cost-effectiveness. Clinical claims may be found on the test report or in other information distributed by the laboratory (websites, test catalogues, newsletters, memoranda, advertisements, etc.). Laboratories are not required to make clinical claims about a test, but any claims made by the laboratory must be validated. In general, the laboratory should validate claims through a clinical study, but for rare conditions or well-accepted uses of a test, reference to published peer-reviewed literature is acceptable. © 2013 College of American Pathologists. All rights reserved. 45 CAP Clinical Validation Requirements in MIC • Introduction to subsection on LDTs in Molecular section of MIC: • Diagnostic sensitivity and specificity must be determined relative to some "gold standard" (eg, biopsy findings, clinical findings, etc.). The sensitivity of an assay equals [TP/(TP+FN)] X 100 and the specificity of an assay equals [TN/(TN+FP)] X 100. (TP=true positive, TN=true negative, FN=false negative, FP=false positive.) Determinations of sensitivity and specificity should be done in a "blinded" fashion (ie, without prior knowledge of the patient's disease status). For some infections, it may not be possible to identify large numbers of positives (ie, patients with the infection) to establish the diagnostic sensitivity of the assay. In such instances, the laboratory should procure as many positive cases as is reasonably possible for method validation and in addition cite any publications that have investigated the diagnostic sensitivity of the assay. © 2013 College of American Pathologists. All rights reserved. 46 Further Issues with Clinical Validation 1. Why restrict clinical validation requirements to MOL and MIC? a. The importance of clinical validation is mainly (but not exclusively) in molecular/genetic LDT’s. b. If applied to all lab areas, clinical validation requirements would capture LDT’s used for— i. Flow cytometry markers ii. Coagulation factor assays iii. Manual microbiology cultures iv. Immunohistochemistry v. Conventional pap smear 2. Future revisions in clinical validation requirements not clear --pending FDA decisions on regulation of LDT’s a. Requirements will most likely be risk-based b. CAP has proposed partnership with FDA © 2013 College of American Pathologists. All rights reserved. 47 Future of LDT Validation Potential Regulatory Change for LDT’s • 7/19/2010: FDA announces intent to regulate LDT’s • FDA has asserted that laboratory tests are medical devices, subject to its jurisdiction • Medical Device Amendments (1976) to Food, Drug & Cosmetic Act (1938) define a medical device as: © 2013 College of American Pathologists. All rights reserved. 49 Medical Device Amendments (1976) • “instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article….intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment or prevention of disease in man…” • “Enforcement discretion”: FDA may choose not to enforce applicable regulations © 2013 College of American Pathologists. All rights reserved. 50 FDA’s Rationale for End of Enforcement Discretion • Increase in type & number of LDT’s • Many developed by commercial labs or biotech companies • More like commercially-developed tests rather than tests used in limited way by local laboratory • LDT viewed as “route to market” driving venture capital funding • Some LDT’s aggressively marketed to clinicians or directly to consumers • Public needs assurances that diagnostics are sound & reliable • Uneven playing field for IVD manufacturers because distributed “test kits” had to go through FDA review prior to marketing, while LDTs did not. • No post-market reporting or recall requirements apply to LDTs. © 2013 College of American Pathologists. All rights reserved. 51 Planned FDA LDT Regulation • 3 guidance documents anticipated; o Overall regulatory framework o Registry Requirements o Description of the Synergies Between CLIA Regulations and FDA Quality System Regulation • Initial timeline: o proposed regulations mid – end 2013 with period for comments o final regulations 2014 - 2015 • Current status: delayed for uncertain period © 2013 College of American Pathologists. All rights reserved. 52 CAP Approach • Meetings with FDA • CAP proposes public-private partnership with CMS (CLIA) & FDA with 3-tier risk based system of accreditation/regulation...initially proposed 2009 o Covers analytic and clinical validation o Applies to all LDT’s © 2013 College of American Pathologists. All rights reserved. 53 CAP Proposed Risk Classification Principles for LDT’s • Laboratory assigns risk; submits analytic/clinical validation & other info to CAP • LDT’s put on-line before 4/24/2003 are grandfathered. • CAP confirms risk level & informs lab o Appeal process offered by CAP © 2013 College of American Pathologists. All rights reserved. 54 Classification Principles for LDT’s, continued Classification Principles Oversight Low Test used with other clinical information; not used alone to determine Rx -Lab notifies CAP & CAP confirms risk class -Lab performs validation -Accreditor inspects Incorrect result unlikely to cause serious harm Lab makes no claims that test alone determines prognosis or Rx © 2013 College of American Pathologists. All rights reserved. 55 Classification Principles for LDT’s, continued Classification Principles Oversight Moderate Test often (but not necessarily) used to predict prognosis or determine if patient can receive a specific Rx -Lab performs validation -CAP performs desk review of validation materials & must approve before test goes on-line; FDA informed -Accreditor inspects Lab may make claims about clinical accuracy Incorrect result may cause serious harm Test method is well understood and independently verifiable © 2013 College of American Pathologists. All rights reserved. 56 Classification Principles for LDT’s, continued Classification Principles Oversight High Test predicts risk, progression or patient eligibility for a specific Rx AND uses proprietary algorithm; result cannot be tied to methods used or interlab comparisons cannot be performed -Lab performs validation & submits to FDA -FDA reviews validation before test goes on-line -Accreditor inspects Incorrect result may cause significant harm AND test method not well understood or not independently verifiable © 2013 College of American Pathologists. All rights reserved. 57 Elements of LDT Validation Defining the Disorder/Test/Clinical Scenario Analytical Validity • Encompasses the development stage of an LDT and serves to establish a testing procedure and identify the intended use of the test. • The test’s ability to accurately and reliably measure the analyte of interest in the clinical laboratory, and in specimens representative of the population of interest. • The ability of a test to diagnose or Clinical Validity © 2013 College of American Pathologists. All rights reserved. predict risk for a particular health condition, measured by clinical (or diagnostic) sensitivity, clinical (or diagnostic) specificity, and predictive values. 58 FDA Proposed Risk Classification • Low risk: Little potential for injury, adjunctive tests identifying one of many characteristics of a tissue or cell with little clinical impact Potential non-serious injury, relatively easy to detect false result—tests where multiple findings used to direct Rx; disease monitoring tests • High risk: Serious injury if incorrect, difficult to detect false result—companion diagnostics, cancer diagnosis, serious communicable diseases © 2013 College of American Pathologists. All rights reserved. 59 Questions? accred@cap.org