11/19/2014 EORTC meeting for Radiation Therapy Technologists: RTT’s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled by Monte Carlo and Collapsed Cone algorithm for VMAT treatment planning Ilenia Iarrusso University of Torino, Italy Equipe: • 4 Physicists + 4 Dosimetrists • Management of DICOM images; • Contouring of some OAR (Organ At Risk) in CT simulation scan; • Planning of 3D treatment plans; • Patient Specific Quality Assurance (the focal point of this study); • LINAC QC. 1 11/19/2014 Our Department: structure and organization I will speak about Our study: purpose, methods and discussion Patient specific Quality Assurance control Our Department: structure and organization I will speak about Our study: purpose, methods and discussion Patient specific Quality Assurance control 2 11/19/2014 ELEKTA MOSAIQ Vers. 2.4.1 LINAC 1 Elekta SL75 X 6; Align RT Philips Brilliance Big Bore CT-Simulation, Tumor L.O.C., 4D-CT; Clarity LINAC 2 Elekta Synergy Velocity X 4-6 , MLCi2, cone beam CT, Clarity, Intuity Elekta ABAS; Anatom-e LINAC 3 Elekta Precise Atlas based autosegmentation X 6-10-25; e, MLCi2, EPID, Clarity LINAC 4 Elekta Axesse Elekta Monaco 3.3 Monte-Carlo Systems Film-and paperless since 2010 X 6-10-15; e, Beam Modulator, cone beam CT, Integrity 1.1, Symmetry, Intuity, 4D-Autoscan Gammknife Perfexion 2014 IORT Linac Elekta/Nucletron Masterplan vers. 4.3 Connection to unit 1 Radioterapia SGAS (3 Linacs) Distance 3 km Novac-7 NRT Connection to unit 2 Radioterapia S. Anna (1 Linac) Distance 500 m LINAC Elekta Synergy 2 Linac Siemens Primus LINAC Elekta Synergy X 6; e, MLC, EPID X 6-18; e, MLC, EPID HDR Brachytherapy The Department of Radiation Oncology of University of Torino is directed by Prof. Umberto Ricardi and the staff is composed of: 9 Radiation Oncologists 13 Residents 4 Medical Physicist 4 Dosimetrists 15 RTT Dedicated ward to assist critical patients and a Day Hospital for chemotherapy and to carry out a supportive care. UNIVERSITA’ DEGLI STUDI DI TORINO 3 11/19/2014 UNIVERSITA’ DEGLI STUDI DI TORINO UNIVERSITA’ DEGLI STUDI DI TORINO 4 11/19/2014 Our Department: structure and organization I will speak about Our study: purpose, methods and discussion Patient specific Quality Assurance control UNIVERSITA’ DEGLI STUDI DI TORINO This is the phantom used by Dosimetrist for Patient Specific QA UNIVERSITA’ DEGLI STUDI DI TORINO 5 11/19/2014 Scandidos Delta4 Detector type Diode Resolution(mm) 5/10 # of Detectors 1069 Max field size 20 Weight kg (detector/phantom) 24 UNIVERSITA’ DEGLI STUDI DI TORINO - VMAT plan is approved. - Plan parameters are copied to a phantom (Verification Plan) UNIVERITA’ DEGLI STUDI DI TORINO 6 11/19/2014 - Delta 4 phantom CT Phantom density override and couch absorption For no-coplanar arcs, beams are arranged to remove couch angle UNIVERSITA’ DEGLI STUDI DI TORINO - Positioning and centering of Delta 4 phantom on the couch UNIVERSITA’ DEGLI STUDI DI TORINO 7 11/19/2014 The planar measured dose is compared to that calculated by TPS UNIVERSITA’ DEGLI STUDI DI TORINO - Dosimetrist and Medical Physicist analyze the results UNIVERSITA’ DEGLI STUDI DI TORINO 8 11/19/2014 - The agreement is acceptable if the percentage of points with value of Gamma Index < 1 is ≥ 90%; When this value is < 90%, we perform a specific procedure: Yes Percentage of point ≥ 90% VMAT – QA is approved No Verification of workflow Measure is repeated VMAT – QA is approved Re-export Step and Shoot VMAT – QA is approved UNIVERSITA’ DEGLI STUDI DI TORINO Our Department: structure and organization I will speak about Our study: purpose, methods and discussion Patient specific Quality Assurance control UNIVERSITA’ DEGLI STUDI DI TORINO 9 11/19/2014 The purpose of this study was to verify the influence of the transmission of the Elekta Stereotactic Body Frame (ESBF) in VMAT treatment planning: UNIVERSITA’ DEGLI STUDI DI TORINO - Selection of one patient with Liver tumor for Stereotactic Body Radiation Therapy (SBRT) and planning of a treatment of 60 Gy in three fractions, using beams with energy of 6 MeV. UNIVERSITA’ DEGLI STUDI DI TORINO 10 11/19/2014 - Acquisition of CT images of the Delta 4 phantom, positioning it on the Elekta Stereotactic Body Frame (ESBF) UNIVERSITA’ DEGLI STUDI DI TORINO - Calculation of VMAT QA with and without ESBF, using Monte Carlo and Collapsed Cone algorithm. Dose calculated using Collapsed Cone algorithm with and without ESBF Dose calculated using Monte Carlo algorithm with and without ESBF UNIVERSITA’ DEGLI STUDI DI TORINO 11 11/19/2014 - Verification of the plan, positioning the phantom with and without the ESBF on the couch and acquisition of the dose. UNIVERSITA’ DEGLI STUDI DI TORINO - Analysis of results and comparison between the two TPS UNIVERSITA’ DEGLI STUDI DI TORINO 12 11/19/2014 Gamma Index passing rates for the two considered algorithms 110 100 90 80 70 60 50 Gamma Index 2mm / 2% 40 Gamma Index 3mm / 3% 30 20 10 0 Monaco Masterplan Without ESBF absorption Monaco Masterplan With ESBF absorption UNIVERSITA’ DEGLI STUDI DI TORINO Evaluation of ESBF absorption for VMAT treatment by means of patient specific QA procedure If absorption is considered, we can find lower agreement between measured and calculated doses Metal artifacts? Considering ESBF absorption, Monaco TPS better than Masterplan TPS AAPM TG 119 (90%). UNIVERSITA’ DEGLI STUDI DI TORINO 13 11/19/2014 The use of ESBF introduces an increased value of uncertainty in treatment planning for liver SBRT Monte Carlo algorithm may overcome this uncertainties UNIVERSITA’ DEGLI STUDI DI TORINO 14