Optimisation of Computed Radiography chest imaging utilising a digitally reconstructed radiograph simulation technique Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust www.hullrad.org.uk Introduction - Literature • Lots of publications have shown that patient anatomy is the limiting factor in the reading normal structures, and detection of lesions (lung nodules) in chest images – – – – – – – Bochud et al 1999 Samei et al 1999, 2000 Burgess et al 2001 Huda et al 2004 Keelan et al 2004 Sund et al 2004 Tingberg et al 2004 • European wide RADIUS chest trial (2005) Introduction - Literature • Chest radiography is now generally considered to be limited by the projected anatomy • Patient anatomy = anatomical noise So if we want to optimize digital system for chest imaging, vital that anatomical noise is present in the images!!! Introduction • However, the radiation dose/image quality relationship must not be ignored • Doses must be kelp ALARP – ICRP 2007 – IR(ME)R2000 – (required legally in the UK) • We would therefore want system (quantum) noise present in an image for dose reduction studies Digitally Reconstructed Radiograph (DRR) • Hypothesis: – Can use CT data of humans to provide realistic anatomy (anatomical noise) • Clinically realistic computerized ‘phantom’ – Simulate the transport of x-rays through the ‘phantom’ and produce a digitally reconstructed radiograph (DRR – a simulation of a conventional 2D x-ray image created from CT data) – Add frequency dependent system noise post DRR calculation – Add radiation scatter post DRR calculation – Validate – Use for optimization studies DRR Algorithm: Virtual Patient • • • Virtual patient derived from chest portion of real CT datasets Voxel resolution = 0.34 x 0.34 x 0.8 mm CT number converted to linear attenuation coefficient (LAC) using tissue equivalent inserts – Measure mean CT No. in each insert – We know elemental composition of each so can derive LAC – Can derive relationship between CT No. and LAC CT dataset reorientated in the ‘PA’ direction Final CT axial batch (i.e. slices 681 to 700) Energy absorbed in CR phosphor of X-rays Intensity (E) A E I ( E ){1 exp x }dE exiting is calculated E max en 0 0 CT axial batch 2 (i.e. CT slices 21 to 40) PA slice N CT axial batch 1 (i.e. CT slices 1 to 20) PA slice 1 X-ray spectra derived from IPEM 78 X-ray attenuated exponentially through CT dataset using a ray casting method of DRR calculation Scatter Addition • Must add to DRR generated image – DRR algorithm does not calculate scatter • Measured scatter in CR chest images using lead pellet array • Use chest portion of RANDO phantom Based on slightly modified work by Bath et al, 2005 Noise Addition lung spine diaphragm Uniform noise image Corrected noise image Lung Nodule Simulation • Added artificial nodules to the CT data prior to Xray projection • Baed on work by Li et al 2009 Human DRR v Human CR DRR CR DRRs: 50 kVp v 150 kVp Scatter Rejection • The DRR algorithm can produce images with scatter rejection DRR No Rejection DRR Grid CR DRR DRR (a) CR Grid (b) (c) Obese Patients • DRR algorithm can also produce images of large/obese patients DRR CR (a) (b) Validation • Decided to validate with RANDO and real patient images: – Histogram of pixel values – Signal to noise ratios (SNR) • Important because signal and noise affects the visualisation of pathology – Tissue to rib ratios (TRR) • Pixel value ratio of soft tissue to that of rib • Important as rib can distract the Radiologist from detecting pathology Phantom Histograms CR - 60 kVp 10 mAs CR - 150 kVp 0.5 mAs Frequency 20000 10000 0 20000 10000 Pixel Value 00 35 00 00 33 00 31 29 DRR - 150 kVp 0.5 mAs 40000 Frequency 30000 20000 10000 0 30000 20000 10000 Pixel Value b Pixel Value d 00 35 00 33 00 31 00 29 00 27 00 25 00 23 00 21 00 19 00 17 00 15 00 35 00 33 00 31 00 29 00 00 27 25 00 23 00 21 00 19 00 0 17 00 00 c DRR - 60 kVp 10 mAs 15 27 Pixel Value a Frequency 00 25 00 00 23 21 00 19 15 00 00 00 00 35 33 00 31 00 29 00 00 27 00 25 23 00 21 00 19 00 00 0 17 15 30000 17 Frequency 40000 115 5000 117 0 7000 119 0 9000 221 0 1000 0 223 3000 0 225 5000 0 227 7000 0 229 9000 0 331 100 00 333 300 00 335 500 0 Frequency Frequency 1155 0000 1177 0000 1 1990 000 22110 000 22330 000 22550 000 2 2770 00 29 0 29 0 00 31 0 31 0 00 33 0 33 00 0 35 0 35 00 00 Frequency Frequency Phantom Histograms CR CR-- 90 90kVp kVp14mAs mAs 30000 20000 10000 10000 00 Pixel Value Pixel Value DRR- -90 90kVp kVp41mAs mAs DRR 30000 30000 20000 20000 10000 10000 00 PixelValue Value Pixel PATIENT - Histograms DRR His togram Frequency 60000 Typical histogram of average patient DRR 30000 0 1744 1995 2246 2497 2748 Pixe l Value Patie nt CR His togram Frequency 60000 Typical histogram of average patient CR image 30000 0 1733 2012 2291 Pixe l Value 2570 2849 SNRs • Good agreement in lung, spine and diaphragm areas of chest • Maximum deviation approx 15% – Mean deviation = 7% • Addition of frequency dependent noise not perfect: – CR system noise is ergodic (changes with time) – Noise added here is a snapshot (and so not ergodic) – However, quantum noise dominates over ‘ergodic noise’ so not such an issue – DQE is NOT constant with dose variation in image Validation - TRRs • Good agreement – Within 2% • As tube potential increases TRR decreases – Due to rib attenuating higher percentage in incident photons at lower potentials than soft tissue, thus forcing up TRR Validation - Radiologists • Have told me DRR images contain sufficient clinical data to allow diagnosis and subsequent optimisation • They have scored the images out of 10 – ‘are the images sufficiently like real CR images? – Average score of 7.8 Conclusions – DRR Algorithm • DRR computer program has been produced that adequately simulates chest radiographs of average and obese patients – Anatomical noise simulated by real human CT data – System noise and scatter successfully added post DRR generation that provides: • SNRs • TRRs • Histograms – in good agreement with those measured in real CR images • Provides us with a tool that can be used by Radiologists to grade image quality with images derived with different x-ray system parameters • WITHOUT THE NEED TO PERFORM REPEAT EXPOSURE ON PATIENTS • ACCEPTED FOR PUBLICATION IN THE BJR – AUGUST 2011 Optimisation of CR Chest Radiography using DRR Generated Images • In Hull chest exposure factors were not standardised (historical reasons!!!!) • Three main hospital sites: – 60 kVp & 10 mAs – 70 kVp & 5 mAs – 80 kVp & 5 mAs • In the last 6 months, four expert image evaluators have scored DRR reconstructed images – Two Consultant Radiologists – Two Reporting Radiographers • Scoring criteria based on European guidelines (CEC) What to Optimise? • Optimum tube potential for Average Patients (70 kg ± 10 kg) – Without scatter rejection (as per Hull protocol) – With scatter rejection (grid and air gap) • Optimum tube potential for Obese patients – Without scatter rejection – With scatter rejection • Is Scatter rejection indicated? • Dose reduction? Scoring Criteria • Images scored on a dual PACS monitor system • Image on right hand screen held at a constant kVp • Images on left hand screen displayed from 50 to 150 kVp in steps of 10kVp (approx) – Image 1 = 50 kV – Image 2 = 60 kV – Image 10 = 150 kV • Test images scored against reference image • All images matched effective dose STUCTURE – NORMAL ANATOMY GRADING IMAGE 1 IMAGE 2 IMAGE 3 Vessels seen approx approx 3 cm from the pleural GRADING VISIBILITY OF STRUCTURE margin -3 Definitely inferior to the reference image Thoracic vertebrae behind the heart -2 Reasonably inferior to the reference image Retrocardiac vessels Pleural margin -1 Slightly inferior to the reference image 0 Equal to the reference image +1 Slightly better than the reference image +2 Reasonably better than the reference image +3 Vessels seen en face in the central area Definitely better than the reference image Hilar region STUCTURE – LUNG NODULES GRADING Nodule in lateral pulmonary region Nodule in hilar region Are the ribs a distraction? Y/N [1] European guidelines on quality criteria for diagnostic radiographic images. CEC European Commission EUR 16260 EN (Luxembourg 1996) IMAGE 4 Tube Potential Optimisation Results – Average Sized Patients, no scatter rejection 0.6 VGAS 0.5 0.4 0.3 0.2 ` 0.1 0 40 60 80 100 120 Tube Potential kVp VGAS = average of Radiologists results 140 160 Results – Average Patients with scatter rejection - grid 0.3 0.2 0.1 0 VGAS 0 20 40 60 80 -0.1 -0.2 -0.3 -0.4 -0.5 Tube Potential (kVp) 100 120 140 160 Results – Average Patients with scatter rejection – air gap 0.2 0.1 0 40 60 80 100 VGAS -0.1 -0.2 -0.3 -0.4 -0.5 Tube Potential (kVp) 120 140 160 Results – Obese Patients without scatter rejection 0.2 0.15 0.1 0.05 0 0 -0.05 -0.1 -0.15 -0.2 -0.25 20 40 60 80 100 120 140 160 Results – Obese Patients without scatter rejection • Very weak trend for better image quality with higher kVp • Lower kVps probably ‘worse’ due to combination of: – Lack of penetration through obese patient – Increased scatter from obese patient (scatter to cassette changes very little with kVp) • This is not so with average patients – Less tissue (fat) so more radiation penetration – Less scatter from fat • It is likely that poorer radiation penetration and increased scatter from obese patient outweighs the inherent benefit of photoelectric contrast obtained from lower kVps Results – Obese Patients with scatter rejection - grids 1 0.8 0.6 VGAS 0.4 0.2 0 40 60 80 100 -0.2 -0.4 -0.6 Tube Potential (kVp) 120 140 160 Results – Obese Patients with scatter rejection – air gap 0.4 0.2 0 40 -0.2 -0.4 -0.6 -0.8 -1 -1.2 -1.4 -1.6 -1.8 60 80 100 120 140 160 Scatter Rejection Scatter Rejection vs no scatter rejection – average patients 1.4 1.2 VGAS 1 0.8 0.6 0.4 0.2 0 grid air gap Scatter Rejection Technique Scatter Rejection vs no scatter rejection – average patients • Superior image quality with scatter rejection technique • Grids performed much better than air gap • Statistically significant differences • BUT: – Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality – Answer in 100% of cases was NO – SCATTER REJECTION FOR AVERAGE PATIENTS IS NOT INDICATED Exposure time for average patients without scatter rejection and low tube potentials? • Scatter rejection not indicated • Therefore low kVps should be used (remember the graph) • European guidance recommends exposure times < 20 ms • Can we achieve this with low kVps?? • Modern Philips X-ray generator: – For lgM = 2 (Agfa CR specific Dose Indicator) – With 630 mA, all kVps are possible – Max exp time = 16 ms • At the expense of increased tube loading Scatter Rejection vs no scatter rejection – obese patients 1.6 1.4 1.2 VGAS 1 0.8 0.6 0.4 0.2 0 grid air gap Scatter Rejection Technique Scatter Rejection vs no scatter rejection – obese patients • Superior image quality with scatter rejection technique • Grids performed much better than air gap • Statistically significant differences • BUT: – Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality – Answer in 100% of cases was YES – ANTI SCATTER GRID USE FOR OBESE PATIENTS IS INDICATED Exposure times for obese patients with an anti-scatter grid • Remember that low tube potentials were superior with an anti-scatter grid!!!! • Need exposure times < 20ms • Is this possible with low kVps and scatter grid for lgM = 2?? • Scatter grid is a focused grid so have to use it in a fixed range of FDD – Nominal distance = 140 cm FDD – Range allowed = 115cm – 180cm • At 140 cm FDD lowest exp time = 17.6ms @ 109 kVp • At 115cm FDD lowest exp time = 20 ms @ 90 kVp • So are limited to 90 kVp Dose Reduction Dose Reduction? • Images were also presented at different doses Dose reduction? • Results suggest doses can be reduced by around 50% before image quality begins to suffer and become unacceptable • Therefore could half exposure mAs • Conclusions – Use of DRR algorithm to optimise CR chest imaging Average patients: – No scatter rejection is indicated – Therefore, low kVps (< 102 kVp) are indicated – Can have exposure times < 20 ms for all kVps • Obese patients – Anti-scatter grid is indicated – So low kVps (<102kVp) should be used – For exposure times < 20ms, are limited to 90 kVp • Doses – As low as 50% reduction possible • ACCEPTED FOR PUBLICATION IN BJR Optimisation & Standardisation in Hull? • Agreed with Consultant Radiologists: – 60 kVp & 10 mAs • After a ‘settling in’ period: – 60 kVp & 8 mAs – Want to go lower than 8mAs eventually!!! • Implication on patient dose? – – – – Using PCXMC effective dose calculation software: 80 kVp/5 mAs = 0.011 mSv 60 kVp/8 mAs = 0.006 mSv Approx 45% drop in effective dose