Bangladesh BEST Programme Calibration of Balance Nihal Gunasekara Sri Lanka 1 Bangladesh BEST Programme Types of Balances • • • • • Electromagnetic force balance (Top loading) Single-pan, two-knife-edge balance Two-pan, three-knife-edge balance Triple beam balance Weighing scale 2 Bangladesh BEST Programme Analytical Balance Balance with an enclosed weighing chamber and a resolution 0.01 or 0.1 mg Micro Balance Balance with a resolution of 1 microgram 3 Bangladesh BEST Programme Micro Balance 4 Bangladesh BEST Programme Semi Micro Balance Balance with a resolution of 10 microgram Balance with a resolution of 0.1 microgram 5 Bangladesh BEST Programme Resolution or Discrimination The smallest change in mass which can be detect by the balance Sensitivity This is a measure of the ability of the balance to detect changes in the load applied to it 6 Bangladesh BEST Programme Response time Time taken for the weighing instrument to indicate the value of a change in load Stability A measure of the time for which the reading on the balance remains unchanged 7 Bangladesh BEST Programme Correction Correction = True Value - Reading Error Error = Reading - True Value 8 Bangladesh BEST Programme Departure From Nominal Value The amount by which the reading on an instrument depart from its nominal value Hysteresis The indications of a measuring instrument when the same value of the quantity measured is reached by increasing or decreasing that quantity 9 Bangladesh BEST Programme Standby Operating state of an electronic instrument where one part is switched off for energy saving reasons Accuracy class. (mass) Legal metrology term. Classification of weights for calibration purposes by OIML R111-1 weights of Accuracy class E1, E2, F1, F2, M1, M2, M3 from 1 mg to 50 kg 10 Bangladesh BEST Programme Different accuracy classes of mass 11 Bangladesh BEST Programme Repeatability Closeness of the agreement between the results of successive measurements of the same measurement carried out under the same conditions of measurement. Repeatability condition include : • the same observer • the same measuring instrument, used under the same conditions • the same location • repetition over a short period of time 12 Bangladesh BEST Programme Reproducibility Closeness of the agreement between the results of measurements of the same measurement carried out under changed conditions of measurement. A valid statement of reproducibility requires specification of the conditions change The changed conditions may include – principle of measurement – method of measurement – observer – measuring instrument – reference standard – location – conditions of use – Time 13 Bangladesh BEST Programme Influence Factors Air movements and heat sources Support surfaces and vibration Electrical and electromagnetic interference Magnetic effects Buoyancy effects 14 Bangladesh BEST Programme Influence factor Influence Parameter Source Effect Test for effect Precautions Vibration Various, Eg. Machinery, roads, railways, aircraft operator Changes forces on the weighing machine due to acceleration Primary Use of a high-mass support, set on a solid surface, for the weighing machine, independent of the adjacent flooring, Eg fixed to bedrock factor Environmental Vibration causes damages Stability of reading under constant load Secondary Accelerometer measurements Note : low frequency, vibration is likely to be more serious in its effects than high (acoustic) frequencies Rubber pads used to damp high frequencies may pass low frequency vibration, and may cause tilting of the weighing machine 15 Bangladesh BEST Programme Influence factor Influence Parameter Source Effect Test for effect Precautions Temperature: air, weight, weighing machine Sunlight, heating systems, operator’s body, greenhouse heating in the weighing chamber Temperature changes and differentials can change the weighing machine response, cause convection effects and change buoyancy Temperature monitoring instruments Avoid direct sunlight and other connections to the external environment; avoid changes in heat and lighting input. Leave balance cases ajar when not in use. Avoid excessive handling of equipment during weighing. Allow weighted object to acclimatize factor Environmental 16 Bangladesh BEST Programme Influence factor Influence Parameter Source Effect Test for effect Precautions Humidity Air supply ,air conditioning, changes in activity levels Buoyancy, absorbed moisture, condensation Rh monitoring instruments Avoid liquid water in the working area unless rh is too low . factor Environmental Air conditioning, humidity controller 17 Bangladesh BEST Programme Influence factor Influence Parameter Source Effect Test for effect Precautions Airflow Air conditioning, wind, convection, operator movement Varies force on weighing machine Primary Avoid temperature gradients. draught free working space, enclose weighing factor Environmental Stability of reading under constant load Secondary machines Enclose weighing machine and observe change of reading stability 18 Bangladesh BEST Programme Influence factor Influence Parameter Source Effect Test for effect Precautions Magnetic Constructional steelwork, ferromagnetic weight (s) electrical induction earth’s magnetic field, moving-coil meters Change of current in force compensation coil, direct attraction to/from surrounding structure Motiondependent restoring forces Primary Avoid weights with residual magnetism (de-gauss) Test with compass or gaussmeter Limit magnetic permeability of weights weighed check for magnetic fields with gaussmeter Screen with mumetal factor Environmental Use a magnet to check for induced reading changes Secondary Monitoring readings when weights are rotated and at varying heights above the weight receptor (place on/under non-magnetic mount) 19 Bangladesh BEST Programme Source of error Type of error Load cell hysteresis Repeatability Electrical Interference Repeatability Mains fluctuation Zero Incorrect horizontal support Sensitivity, Zero, Repeatability Mechanical errors Repeatability 20 Bangladesh BEST Programme ACCURACY CLASS (OIML R111-1 :2004) Nominal value Tolerance in mg Class E1 Class E2 Class F1 Class F2 Class M1 Class M2 Class M3 10 kg 5.0 16 50 160 500 1600 5000 5 kg 2.5 8.0 25 80 250 800 2500 2 kg 1.0 3.0 10 30 100 300 1000 1 kg 0.5 1.6 5.0 16 50 160 500 200 g 0.10 0.3 1. 0 3 10 30 100 100 g 0.05 0.16 0.5 1.6 5 16 50 50 g 0.030 0.10 0.3 1.0 3 10 30 10 g 0.020 0.06 0.20 0.6 2.0 6.0 20 5g 0.016 0.05 0.16 0.5 1.6 5.0 16 2g 0.012 0.04 0.12 0.4 1.2 4.0 12 1g 0.010 0.03 0.10 0.3 1.0 3.0 10 100 mg 0.005 0.016 0.05 0.16 0.5 1.6 50 mg 0.004 0.012 0.04 0.12 0.4 10 mg 0.003 0.008 0.025 0.08 0.25 5 mg 0.003 0.006 0.020 0.06 0.20 0.003 0.006 0.020 0.06 0.20 1 mg 21 Bangladesh BEST Programme Lowest class of weights required for different types of weighing machine Resolution Capacity 100g 10g 1g 10mg 1mg 0.1mg 0.01mg or less M1 M1 F2 F1 E2 M1 M1 F2 F1/E2 E2 E2 E2 E2 Up to 200g 200g to 1 kg 100mg 1 kg to 30kg M2 M2 M1 F2 E2 30kg to 100kg M2 M1 F2 F1 E2 Above 100 kg M2 M1/F2 F1 E2 22 Bangladesh BEST Programme Calibration of Top Loading and Analytical Balances 23 Bangladesh BEST Pregramme Before use a Balance………… • • Keep the Balance power on appropriate time to warm up. Clean the balance as best as no contamination on the balance pan. • • • • If it is applicable, level the balance before use. Don’t move the balance after the calibration. Lint or any other soft gloves should be worn by the user. A pair of forceps or a pair tweezers with thermal insulating ends should be used to lift and place weights. 24 Bangladesh BEST Programme Before use a Balance • Make sure with the soft placement of weights on the balance pan as level best of possible. • Readings should be taken after the balance is stabilized well. • The place should be free from vibration. • Any other relevant precaution practiced by the user. 25 Bangladesh BEST Programme Balance Calibration The calibration of the electromagnetic force compensation balance consists of the following tests: • • • • • Test 1 - Scale Value Test 2 - Repeatability of Reading Test 3 - Departure from Nominal Value Test 4 - Off centre Loading Test 5 – Hysteresis 26 Bangladesh BEST Programme Scale Value Internal Scale Adjustment –Internal Mass External Scale Adjustment – External Mass Procedure – Balance Operating Manual 27 Bangladesh BEST Programme Repeatability of Reading Objective of Measurement : The repeatability test is carried out to determine the ability of the balance to give under defined conditions of use closely similar responses for repeated measurements The balance Standard Deviation is the “repeatability of measurement” 28 Bangladesh BEST Programme Repeatability of Reading : • M 1 -Set the reading to zero without any mass on the pan • M 2 - Record the zero reading of the balance (z) • M 3 - Place the mass on the pan and record the reading (r) • M 4 - Remove the mass from the pan • M 5 – Repeat M2 to M4 ten times • during the process the balance should not be zeroed 29 Bangladesh BEST Programme Repeatability of Reading: Performed at three points : Near zero Near half load Near full load Bangladesh BEST Programme DATA RECORDING Mass 1 : No Zero reading (z) Mass reading (r) Mass 2: Corr. reading (r-z) Zero reading (z) Mass reading (r) Mass 3: Corr. reading (r-z) Zero reading (z) Mass reading (r ) Corr. reading (r-z) 1 2 3 4 5 6 7 8 9 10 31 Bangladesh BEST Programme Repeatability of Reading Mass 1: Mean value Std. Dev. ( σ ) Max. Diff. Uncer. Type A Mass 2: Mass 3 : : : : : 32 Bangladesh BEST Programme Repeatability of Reading Repeat three different values of M ; near zero, half maximum and maximum load. Load on Pan Standard Deviation of Reading (mg) Maximum Difference Between Successive Readings (mg) 0.05 0.067 0.2 100 0.052 0.1 200 0.052 0.1 (g) Bangladesh BEST Programme Departure from Nominal Value: Objective of Measurement : “The departure from nominal value measurements are carried out to determine the deviation of the balance reading from the expected value” This is the Correction to Balance Reading 34 Bangladesh BEST Programme Departure from Nominal Value: 1 Divide the range into at least 10 equally spaced steps 2 The masses corresponding to M, 2M, 3M, 4M, 5M, 6M etc. until the capacity of balance is reached 3 In addition choose four more test points corresponding to 20 %, 40 %,60%and 80 % of the first step for single range. 35 Bangladesh BEST Programme Departure from Nominal Value: 1 - Read the zero z 2 - Place the weight on the pan &record reading r 3 – Remove the weight momentarily and place it back on 1 1 the pan and note the new reading r2 4 - Remove the weight and read the zero z 5 – Repeat 1 to 4 for weights of masses 2M,3M,4M,5M etc 2 until the capacity of balance is reached 36 Bangladesh BEST Programme Departure from Nominal Value: • Average zero reading z = (z + z )/ 2 • Average mass reading r = (r + r )/ 2 • Difference d = r – z i 1 i i i 1 2 2 i • Correction c = M – d i i i • M is the conventional value of the mass i 37 Bangladesh BEST Programme Correction to Balance Reading Pan Load Conv. Value Mi Reading M- 100 100.02 0.00 100.00 0.00 100.02 2M 3M 4M 5M 6M 7M 8M 9M 10M Mean 0.00 100.01 Difference ri - zi 100.01 Correction Mi - di 0.01 Bangladesh BEST Programme Departure from Nominal Value: Reading Correction(mg) Expanded Uncertainty k=2 (g) Before the Calibration After the Calibration +(mg) 4 -- 0.0 0.096 8 -- 0.0 0.097 12 -- -0.1 0.097 16 -- 0.1 0.098 20 0.2 -0.0 0.099 40 -- 0.0 0.098 60 -- 0.0 0.098 80 0.4 -0.1 0.099 100 -- -0.1 0.099 120 0.6 -0.1 0.099 140 -- -0.1 0.10 160 0.4 -0.2 0.10 180 -- -0.3 0.10 200 0.5 -0.3 0.11 220 -- -0.3 0.11 39 Bangladesh Best Programme Off Centre Loading: Objective of Measurement : “The off center loading is carried out to find the variation in the displayed weighted as the object is moved to various positions on the weighing pan” 40 Bangladesh BEST Programme Off Centre Loading This test to find the variation of the reading for several place of the balance pan. Test should be done for rectangular type balance pans and as well as for circular type balance pans. A single weight should be used 3 4 1 5 2 3 4 1 2 5 41 Bangladesh BEST Programme Off Centre Loading: Back (g) 100.0000-100.0000=0.0000 Center (g) Left (g) 100.0000-100.0000=0.0000 100.0001-100.0000=0.0001 Max Difference =0.0002 g Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time Right (g) 99.9999-100.0000= -0.0001 Front (g) 100.0001-100.0000=0.0001 42 Bangladesh BEST Programme Off Centre Loading: Mass of 100g was moved to various positions on the pan. The balance readings obtained relative to the centre reading Centre(g) Front(g) Back(g) Left(g) Right(g) Maximum Difference (g) 0.0000 0.0001 0.0000 0.0001 - 0.0001 0.0002 Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time 43 Bangladesh BEST Programme Hysteresis: Objective of Measurement : “The hysteresis measurements are performed to determine the difference in the indication at the same load , depending upon the direction of approach to that load ie increasing or decreasing the load” 44 Bangladesh BEST Programme Hysteresis Perform the test at one point about mid range • • • • Zero the balance and record the reading z Place a mass M on the plan and record the reading r Add extra mass M to bring the balance close to full range Remove the extra mass M and record the reading with the mass M still on the pan • Remove the mass M and record the reading r • Repeat above steps three times • Take the average of three readings 1 1 1 2 2 1 1 2 45 Bangladesh BEST Programme Hysteresis Mass M = g Pan Load Reading 1 Reading 2 Reading 3 Zero ( z1 ) M ( m1 ) M + M1 M ( m2 ) Zero ( z2 ) m1 - m2 z1 - z2 46 Bangladesh BEST Programme Hysteresis: Pan Load (g) Hysteresis Error(mg) 0 0.0 100 0.1 47 Bangladesh BEST Programme Hysteresis: Mass Value VS Display Value 100 Hysteresis error 80 90 Display Value 70 60 Increasing 50 Decreasing 40 30 20 10 0 0 20 40 60 80 100 Mass Value 48 Bangladesh BEST Programme Calibration of Single Pan two Knife Edge Balances 49 Bangladesh BEST Programme Balance Calibration The calibration of the single pan two knife edge balance consists of the following tests: • • • • • Test 1 - Repeatability of Reading Test 2 - Departure from Nominal Value Test 3 – Built-in-Mass Test 4 - Off centre Loading Test 5 – Hysterisis 50 Bangladesh BEST Programme Balance Calibration The calibration of the single pan two knife edge balance consists of the following tests: • Test 1 - Repeatability of Reading- same as electronic balance • Test 2 - Departure from Nominal Value- same as electronic balance • Test 3 – Built-in-Mass • Test 4 - Off centre Loading- same as electronic balance • Test 5 – Hysteresis- same as electronic balance 51 Bangladesh BEST Programme Balance Calibration • • • • • • The procedure to determine the Built-in- Mass With all dials set to the zero and no load on the pan, release the balance and record the reading z1 Set the "1" on the dial to be tested, place a calibrated mass, release the pan and record the reading m1a Remove the mass , replace it and record the reading m1b Return the dial to zero, remove the mass from the pan and record the zero z2 Repeat the above steps for all dial settings of the balance (i.e. 2 , 3, 10 , 20 ,…………..90 , 100 , 200 ) Follow the same procedure used for linearity test to 52 Bangladesh BEST Programme Calibration of Triple Beam Balances 53 Bangladesh BEST Programme Balance Calibration The calibration of triple beam balances consists of the following tests: • Test 1 - Repeatability of Reading - same as electronic balance • Test 2 - Departure from Nominal Value (Linearity Test) same as electronic balance 54 Bangladesh Best Programme Triple Beam Balance Before the calibration • Clean the weighing pan and other parts of the balance taking due care • Level the balance • Exercise the balance by making several weighing 55 Bangladesh BEST Programme Triple Beam Balance Repeatability of readings: • Set the reading to zero without any mass on the pan • Record the zero reading zi of the balance without any mass on the pan • Place the mass M on the pan and record the mass reading mi • Remove the mass M from the pan • Repeat the above steps ten times (The balance should not be zeroed during the series of readings) • Repeat for this for near zero, half load and near maximum 56 Bangladesh Best Programme Triple Beam Balance • Select a minimum of 10 equally spaced points ( M , 2M ,…..10M ) on the measuring range in such a way they cover the entire range or the part of the measuring range to be calibrated. In addition four test points corresponding to 20%,40%,60% and 80 % of the first step • Record the zero reading z1 • Set the slider scale to the first step on the measuring range (M) , place the calibrating mass(es) on the pan until the balance comes to equilibrium and record the mass reading mi • Remove the mass(es), place it back on the pan and record the mass reading mi2 • Remove the mass(es) and record the zero reading zi+1 • Repeat the above steps for the remaining points ( 2M , 3M ,….) 57 Bangladesh Best Programme Weighing Scale (digital or analogue) The calibration of weighing scales consists of the following tests: Test 1 - Repeatability of Reading - same as electronic balance Test 2 - Departure from Nominal Value (Linearity Test) same as electronic balance 58 Bangladesh BEST Programme Calibration Results of a Set of Weights –Class F1 Nominal Value Conventional Value Uncertainty k=2 Max. Perm. Error 1 mg 1 mg – 0.001 mg 0.006 mg ±0.02 mg 2 mg 2 mg + 0.011 mg 0.006 mg ±0.02 mg 5 mg 5 mg+0.002 mg 0.006 mg ±0.02 mg 20 mg 20 mg – 0.005 mg 0.01 mg ±0.03 mg 50 mg 50 mg + 0.009 mg 0.012 mg ±0.04 mg 100 mg 100 mg + 0.011 mg 0.015 mg ±0.05 mg 500 mg 500 mg + 0.033 mg 0.025 mg ±0.08 mg 1g 1 g – 0.028 mg 0.03 mg ±0.1 mg 2g 2 g + 0.022 mg 0.04 mg ±0.12 mg 20 g 20 g + 0.009 mg 0.08 mg ±0.25 mg 50 g 50 g – 0.12 mg 0.1 mg ±0.3 mg 100 g 100 g – 0.22 mg 0.15 mg ±0.5 mg 200 g 200 g + 0.04 mg 0.3 mg ±1 mg Issued by an accredited laboratory (DKD) 59 Bangladesh BEST Programme Thank you 60