Measurements in Fluid Mechanics 058:180:001 (ME:5180:0001) Time & Location: 2:30P - 3:20P MWF 218 MLH Office Hours: 4:00P – 5:00P MWF 223B-5 HL Instructor: Lichuan Gui lichuan-gui@uiowa.edu http://lcgui.net Lecture 21. Temperature measurement 2 Temperature measurement Temperature scales - three temperature scales in use today, Fahrenheit (F), Celsius (C) and Kelvin (K) Fahrenheit temperature scale - 32 for the freezing point of water - 212 for the boiling point of water - interval divided into 180 parts Celsius, or centigrade, scale - 0 for the freezing point of water - 100 for the boiling point of water - conversion formula: F = 9/5C + 32 Kelvin temperature scale - base unit in International System (SI) of measurement - zero point at absolute zero - difference between the freezing and boiling points of water is 100 degrees - conversion formula: K = C + 273 3 Temperature measurement Thermometers Thermal expansion thermometers - liquid-in-glass thermometers - bimetallic thermometers Thermocouples - based on the thermoelectric effect Resistance thermometers - based on the relationship between temperature and electric resistance - include metallic resistance sensors (RTDs), and semiconductor resistance sensors Coil elements 4 Temperature measurement Liquid-in-glass thermometers - a bulb, a reservoir in which the working liquid can expand or contract in volume - a stem, a glass tube containing a tiny capillary connected to the bulb and enlarged at the bottom into a bulb that is partially filled with a working liquid. The tube's bore is extremely small - less than 0.5 mm in diameter - a temperature scale is fixed or engraved on the stem supporting the capillary tube to indicate the range and the value of the temperature. The liquid-in-glass thermometers is usually calibrated against a standard thermometer and at the melting point of water - a reference point, a calibration point, the most common being the ice point - a working liquid, usually mercury or alcohol - an inert gas is used for mercury intended to high temperature. The thermometer is filled with an inert gas such as argon or nitrogen above the mercury to reduce its volatilization. - constriction may be used to measure maximal or minimal temperature 5 Temperature measurement 𝑻𝒏𝒐𝒏−𝒊𝒎𝒎 Liquid-in-glass thermometers - Typical resolution: 0.05-1 K 𝑵 - Immersion types: partial Immersion (inserted in fluid up to marked line) total immersion (inserted in fluid up to liquid column) complete immersion (entirely immersed in fluid) 𝑻𝒊𝒎𝒎 - total immersion required to avoid errors due to temperature difference between immersed and non-immersed sections - stem correction necessary with partial immersion thermometers e.g. for mercury-in-glass thermometers: ∆𝑇 = 0.00016𝑁 𝑇𝑖𝑚𝑚 − 𝑇𝑛𝑜𝑛−𝑖𝑚𝑚 𝑁- number of degrees (K) of the non-immersed portion 𝑇𝑖𝑚𝑚 - temperature of the immersed portion 𝑇𝑛𝑜𝑛−𝑖𝑚𝑚 - temperature of the non-immersed portion - disadvantage: poor spatial and temporal resolutions - advantage: excellent laboratory standards for calibration of other instruments 6 Temperature measurement Bimetallic thermometers - two thin plates of different materials - vastly different thermal expansion coefficient - bonded together tightly with one end fixed & another free - curvature of the assembly changed due to temperature variation - helically or spiral coiled assembly used to amplify motion resulting from temperature change Bimetallic thermometer (flat, spiral strip) - bimetallic assemblies also used in thermostatic controls - typical resolution about 1% of full scale, and maximal operation temperature around 500C 7 Temperature measurement Thermocouples Seebeck effect - Any electrical conductor will develop a potential difference (thermoelectric voltage) between two of its points that have a temperature difference. Thermocouple configuration - two dissimilar metallic wires (e.g. A and B) joined firmly at two junctions - one junction exposed to the temperature of interest (e.g. T1) - the other one (reference junction) kept at known constant temperature (e.g. T2) - reference junction conventionally immersed in an ice bath for reference temperature of 0C - constant reference temperature also provided with an electronically controlled heated block - common types of thermocouples and their properties - typical resolution in the order of 1C - highest temperature of 2930C by tungsten-rhenium type 8 Temperature measurement Thermocouples Sensor & measuring circuit 9 Temperature measurement Resistance thermometers Resistance temperature detectors (RTDs) - pure metal thermometers of platinum (most popular and accurate), nickel, or copper - typical resolution of 0.1 K, possible high resolution of 0.0001 K - non-linear response fitted with low-order polynomials e.g. resistance Rpt of platinum RTD in the range 0f 100-700C described by Callendar-Van Dusen equation Rpt0 – resistance at 0C Cold-wires - similar construction to HW - high frequency response in temperature measurement Thermistors - semiconductor elements whose resistance is a very strong function of temperature. - extremely high sensitivity to temperature - non-linear response 10 Discussion on homework assignment - Questions and Problems: 8 on page 286 11 Homework - Read textbook 12.1-12.2 on page 290 - 296 - Questions and Problems: 6 on page 305 - Due on 10/14 12