Precision Power Measurement Solutions from Bird Agenda • National Standards Traceability- Challenges & Bird’s Solution • RF Metrology Paths at Bird Electronic Corporation – – – – – High power RF Calorimetry Low power microwave attenuation Low power microwave power MCS (master calibration system) Test Setups & system considerations • 4020 Series Power Sensors and the 4421 Power Meter • Typical Field Power Measurement Systems National Standards TraceabilityChallenges & Bird’s Solution Generic Traceability Path National Reference Standard Measurement Reference Standard Working Standard Power Sensors NIST Bird Metrology Bird Manufacturing Facility Power Measurement Requirements of the Semiconductor Industry Various frequency & power combos 40 kW Power 13.56 Mhz Frequency Accuracy Capability of the Scientific Community Bird’s performance range & capability Power NIST, NPL etc. Frequency Accuracy Capability of the Scientific Community Bird’s performance range & capability Calorimetry Path Precision Attn & Power Path Power NIST, NPL etc. Frequency Bird’s Multi-Path Solution High Power RF Calorimetric Path Low Power Precision Attenuator RF & Microwave Path NIST AC & DC Standard NIST Attenation Standard Primary Lab Low Power RF & Microwave Power Path < 10 mw Primary Standard Thermistor CN Mount Working Standard Thermistor Mount Precision 60 Hz Power Analyzer Working Standard Measurement Ref. Standard MicroCalorimeter < 10 mw AC Voltage & Current Stds. NIST Fixed Attenuator Set Working Standard NIST Standard High Power Calorimeter VNA Coupler Verification Cal Factor Verification MCS Transfer Standard Couplers + Power Meter Test Setups 4027, 4028 4024, 4025 Model 43 Calibration Subtleties of the Bird System • • • • • +/- 1% calibration requirements dictate daily calibration +/- 3% is calibrated every 6 months +/- 5% is calibrated annually Cross correlations are on-going and constant Multiple paths are used to cross correlate high power & high frequency standards • It is capital intensive, time consuming, and demands high skill levels, but worth every effort in order to guarantee the high accuracy demands of the semiconductor industry RF Metrology Paths at Bird Electronic Corporation • • • • • High power RF Calorimetry Low power microwave attenuation Low power microwave power MCS (master calibration system) Test Setups & system considerations High Power RF Calorimetric Path NIST AC & DC Standard Primary Lab AC voltage & Current Stds. Working Standard Precision 60 Hz Power Analyzer Measurement Ref. Standard High Power Calorimeter • Calorimetry is the critical link between high power AC standards & high power RF standards Calorimetric Power Meters 6091 Power (kW) = .263 x flow rate (GPM) x T (0C) 8860 Calorimeter Block Diagram Bird Metrology Manufacturing Facility Characteristics of Calorimetric Power Meters • Highly Accurate, Especially When Using 60Hz Substitution Technique • Measures True Heating Power, Regardless of Harmonic Content or Modulation Characteristics of Signals • Requires Careful Setup and Maintenance, Due to Coolant Characteristics • Long Settling Time Specific Heat of Water AC Substitution Method 60 Hz AC Source Precision AC Power Meter RF Calorimeter RF Source • Measure 60 Hz power into calorimeter w/AC Power meter • Adjust calorimeter display to match AC power meter • Accuracy of AC standard has now been transferred to calorimeter • When RF is supplied to load, read calibrated watts from calorimeter display AC Substitution Technique • Use Low Distortion 60Hz Source • Calibrate Calorimeter Using Precision 60Hz Power Meter (Accuracy = <0.1%) • Apply Unknown RF Source to Calorimeter • Adjust Coolant Flow Rate to Maintain ΔT Across Load of > 2º C • Allow 1 hour For Stabilization Transfer of Accuracy from AC to RF VSWR AC (60 Hz) RF frequency • Calorimetric load has virtually identical response at both AC & RF -0.15 9/ 11 9/ 12 9/ 13 9/ 14 9/ 15 9/ 16 9/ 17 9/ 18 9/ 19 9/ 20 9/ 21 9/ 22 9/ 23 9/ 24 9/ 25 9/ 26 9/ 27 9/ 28 9/ 29 9/ 30 10 /1 10 /2 9/ 10 Calorimetric Stability % Error 0.2 0.15 0.1 Error 0.05 0 -0.05 -0.1 Days Low Power Precision Attenuator RF & Microwave Path NIST Attenuation Standard < 10 mw • Provides the important link between low power, high frequency attenuation values & high frequency coupling values NIST Fixed Attenuator Set Working Standard VNA Coupler verification MCS Transfer Standard Working Standard VNA Precision Coupler • Transfers the accuracy of the VNA to the precision coupler when the coupling value is determined Attenuation Standards • Attenuation kit traceable to NIST VNA Attenuation Kit Low Power RF & Microwave Power Path NIST Standard MicroCalorimeter < 10 mw Primary Standard Thermistor CN Mount Working Standard Thermistor Mount MCS Transfer Standard Cal Factor verification Provides the link between high frequency low power standards and high frequency power meters Working Standard Thermal Power Meter CN Thermister Mount • Cal factor of power meter is verified with reference to Thermister mount MCS Transfer Standard MCS Transfer Standard Provides the combinational accuracy of calibrated high frequency power & coupling standards into a single calibrated device that can be used as a measurement standard in a high frequency, high power test setup Directional Coupler - Thermal Power Meter MCS Standard Characteristics of Directional CouplerThermal Power Meter Standards • Wide Dynamic Range • Useful Frequency Range Determined by Directional Coupler • Complicated Error Budget – Internal Reference Uncertainty – Mismatch Uncertainty – Calibration Factor Uncertainty • Fundamental Accuracy Limited by Knowledge of Directional Coupler Attenuation, as well as Power Meter Error Sources. • Mismatch Uncertainty is a Major Contributor to Total Uncertainty Precision Power Measurement Test Setups Test Setups 4027, 4028 4024, 4025 Model 43 4027A +/-1% Calibration System These two measurements must agree within +/- .2% Test Results 4027A Test Results 1.00 Difference from Calorimeter (%) 0.80 0.60 0.40 SN 11569 MCS-59 w ith 11569 0.20 SN 11596 MCS-59 w ith 11596 0.00 -0.20 SN 11597 5 10 15 20 25 30 35 40 45 50 55 60 MCS-59 w ith 11597 SN 11598 MCS-59 w ith 11598 -0.40 -0.60 -0.80 -1.00 Elapsed Time (minutes) A Typical Field Calibration Setup Bird 4020AM Power Sensor p2 p1 p2 RF Matching Network p1 p2 p1 Bird Oil load 5 kW RF Generator at 13.56 MHz p2 Plasma Etching Chamber Bird 4421 Power Meter Mismatches are present at each interconnection of system components Mismatch Uncertainty p 2p 1p 2 S p 1p 2 S p1 p2 S p 2S Total reflected signal p2 +/- p1p2p2 = p2’ Mismatch Uncertainty VSWR (apparent) = 1 + p2’ p2 +/- p1p2p2 = p2’ 1– p2’ VSWR (apparent) = 1 + ( p2 +/- p1p2p2 ) = 1 - ( p2 +/- p1p2p2 ) 1 + p2 +/- p1p2p2 1 - p2 -/+ p1p2p2 Recognize that this expression can be approximated as the product of Very small contribution 1 + p2 1 – p2 x 1 +/- p1p2p2 1 -/+ p1p2p2 = 1 +/- p1p2p2 + p2 +/- p1 p32 1 -/+ p1p2p2 + p2 +/- p1 p32 Then: VSWR (true) x 1 +/- p1p2p2 1 -/+ p1p2p2 ~ VSWR (apparent) Mismatch Uncertainty VSWR (true) x 1 +/- p1p2p2 1 -/+ p1p2p2 ~ VSWR (apparent) • The true VSWR is multiplied by an uncertainty factor which can only be controlled by carefully choosing the reflection coefficients (p1 and p2) at the source and test points Lower limit of multiplier factor = 1 - p1p2p2 1 + p1p2p2 Lower uncertainty limit of measured VSWR = F- Upper limit of multiplier factor = 1 + p1p2p2 1 - p1p2p2 Upper uncertainty limit of measured VSWR = F+ = F- 1 + p2 1 – p2 = F+ 1 + p2 1 – p2 Mismatch Uncertainty Mu (%) = 100 [(1 ± Pg Pl)2 – 1] Where: Pg = Reflection Coefficient of Source Pl = Reflection Coefficient of Load Pg and Pl are FREQUENCY DEPENDENT QUANTITIES! VSWR Mismatch Uncertainty 3 1.1 source VSWR VSWR Uncertainty 2 1.1 source VSWR 1.5 source VSWR 1 1.5 source VSWR 2.0 Source VSWR 0 1 2 3 -1 4 5 2.0 Source VSWR 2.5 Source VSWR 2.5 Source VSWR 3.0 Source VSWR 3.0 Source VSWR 3.5 Source VSWR 3.5 Source VSWR -2 -3 Load VSWR Transmission Uncertainty p 2p 1p 2 S p 1p 2 S p1 p2 S p 2S Total transmitted signal S(1 +/- p1p2) +/- dB (ripple) = 20 log | 1- p1p2 | Transmission Uncertainty High point flatness Ripple averaged out Low point • If data is taken at discrete points, then each individual reading carries an uncertainty of +/- x dB Measurement uncertainty Transmission Uncertainty 4.5 4 Uncertainty +/- dB 3.5 3.5 Source VSWR 1.1 Source VSWR 3 1.5 Source VSWR 2.5 2.0 Source VSWR 2 2.5 Source VSWR 3.0 Source VSWR 1.5 3.5 Source VSWR 1 1.1 Source VSWR 0.5 0 1 2 3 Load VSWR 4 5 Transmission Uncertainty 0.25 Uncertainty +/- dB 0.2 1.1 Source VSWR 1.15 Source VSWR 0.15 1.2 Source VSWR 1.25 Source VSWR 0.1 1.3 Source VSWR 1.35 Source VSWR 0.05 0 15 20 25 30 Load Return Loss (dB) 35 40 Example of Typical RF System Error Budget Prototype RF Delivery System Gain/Mismatch Analysis Power Sensor Matching Network Cable Cable Cable RF Generator No. Termination Device Return Loss (dB) Input 1 2 3 4 5 6 7 8 9 10 Generator cable Power Sensor cable Matching Network cable termination Total 25.00 30.00 30.00 30.00 30.00 37.00 Output 25.00 30.00 30.00 30.00 30.00 37.00 Gain/Loss (dB) Mag. -0.010 -0.025 -0.010 -0.050 -0.010 -0.11 (db) Ripple 0.03 0.02 0.02 0.02 0.01 +/- 0.05 Watt Budget Watts In Watts out dbm out 1700.0 1696.1 1686.4 1682.5 1663.2 1,700.0 1,696.1 1,686.4 1,682.5 1,663.2 1,659.4 62.30 62.29 62.27 62.26 62.21 62.20 Low Uncertainty 1,642.1 62.15 62.24 Hi Uncertainty 1,676.8 1659.4 stage loss Cumulative loss 3.9 9.7 3.9 19.3 3.8 0.0 3.9 13.6 17.5 36.8 40.6 40.6 Delta Watts 34.7 Prototype RF Delivery System Gain/Mismatch Analysis Powe r Se nsor Ma tching N e twork Ca ble Ca ble Ca ble RF Ge ne ra tor No. T e rmina tion Device Return Loss Input 1 2 3 4 5 6 7 Generator cable adapter adapter Power Sensor adapter adapter cable adapter Matching Network adapter adapter cable adapter adapter termination Total 25.00 26.44 26.44 30.00 26.44 26.44 26.44 26.44 30.00 26.44 26.44 26.44 26.44 26.44 37.00 (dB) Output 25.00 26.44 26.44 26.44 30.00 26.44 26.44 26.44 26.44 30.00 26.44 26.44 26.44 26.44 26.44 Gain/Loss Watt Budget (dB) Mag. (db) Ripple Watts In Watts out dbm out stage loss Cumulative loss -0.010 -0.005 -0.005 -0.025 -0.005 -0.005 -0.010 -0.005 -0.050 -0.005 -0.005 -0.010 -0.005 -0.005 0.05 0.04 0.04 0.02 0.04 0.04 0.04 0.04 0.02 0.04 0.04 0.04 0.04 0.04 1700.0 1696.1 1694.1 1692.2 1682.5 1680.5 1678.6 1674.7 1672.8 1653.7 1651.8 1649.9 1646.1 1644.2 1,700.0 1,696.1 1,694.1 1,692.2 1,682.5 1,680.5 1,678.6 1,674.7 1,672.8 1,653.7 1,651.8 1,649.9 1,646.1 1,644.2 1,642.3 62.30 62.29 62.29 62.28 62.26 62.25 62.25 62.24 62.23 62.18 62.18 62.17 62.16 62.16 62.15 3.9 2.0 1.9 9.7 1.9 1.9 3.9 1.9 19.1 1.9 1.9 3.8 1.9 1.9 0.0 3.9 5.9 7.8 17.5 19.5 21.4 25.3 27.2 46.3 48.2 50.1 53.9 55.8 57.7 Low Uncertainty 1,547.7 61.90 62.41 Hi Uncertainty 1,742.6 57.7 Delta watts 194.9 -0.15 +/- 0.26 1646.1 4027A +/-1% Calibration System These two measurements must agree within +/- .2% Effects of Harmonics on Power Measurement • 4027 Power Sensor Detector Scheme is Very Sensitive to Harmonics in the Signal. • 4027 is Calibrated with Signals Having Harmonics of Less than –60dBc. • Signals with Harmonic Content Greater Than –60dBc will Cause Offsets in Power Readings • Effects of Harmonics are Determined not Only by Diode Response, but Also by Directional Coupler Response Characteristics, as well as Phase Relationships of Harmonic. Effects of Harmonics on Power Measurement Worst Case Errors % Error % Error % Error Harmonic with One with Two with Three Level Harmonic Harmonics Harmonics -55 0.36% 0.70% 1.00% -50 0.63% 1.10% 1.80% -45 1.10% 2.10% 2.90% -40 1.90% 3.90% 5.80% Effects of Modulation on Power Measurement • Detector Scheme Used in 4027 is Sensitive to Amplitude Modulation of the Signal. • Magnitude of Change in Power reading is Related to Power Level and Instrument Range. • Approximate Error: – At 10% of Full Scale: 5% AM Results in 2% Error – At 90% of Full Scale: 5% AM Results in 8% Error Additional Tips for Making Accurate Power Measurements • Know the effects of the mismatches present in the system architecture on the power measurement uncertainty • Avoid the use of multiple adapters or non-compensated (high VSWR) adapters between cables and components • Perform a system error budget to quantify the effects of mismatches and component tolerances in the system • Avoid the use of long interconnecting cables, as the ripple period will be more frequent as the length is increased for a given frequency • Use coupler based measurement techniques when the load is unstable or poor in performance compared to the system line impedance • Averaging techniques over wider frequency bands can be effective in minimizing the effect of mismatch uncertainties 4020 Series Power Sensors and the 4421 Power Meter 4421/4020 Series Power Meters • Highly Accurate, Highly Repeatable Power Meter System • Long Product History, Introduced in 1988 • Has Become the Power Meter of Choice in Semiconductor Processing Applications • Extremely Wide Dynamic Range 4027A Precision Power Sensor Model Power Range Frequency VSWR Range Directivity Insertion Loss 4027A12M 300 mW to 1 kW 10-15 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A250K 3 W to 10 kW 250-400 kHz 1.0 to 2.0 28 dB <0.05 dB 4027A400K 3 W to 10 kW 400-550 kHz 1.0 to 2.0 28 dB <0.05 dB 4027A800K 3 W to 10 kW 800-950 kHz 1.0 to 2.0 28 dB <0.05 dB 4027A2M 3 W to 10 kW 1.5-2.5 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A4M 3 W to 10 kW 3-5 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A10M 3 W to 10 kW 10-15 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A25M 3 W to 10 kW 25-30 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A35M 3W to 10 kW 35-45 MHz 1.0 to 2.0 28 dB <0.05 dB 4027A60M 3W to 6kW 45-65 MHz 1.0 to 2.0 28dB <0.05 dB • Designed for Service in Semiconductor Processing Applications • ± 1% Accuracy at Calibration Points • Several Models to Address Specific Semiconductor Power Levels and Frequencies 4027 true average responding detector scheme VOUT = VIN 2 5.77 • First generation diode detectors operate over transition region of diode response curve limiting use in modulated communications systems. • The entire dynamic range of the 4027 series sensor is contained within the square law operating range of the detector • Sensor will behave similar to a thermal device, responding to the heating power of the signal being measured 4027A Power Sensor 4027F Power Sensor LP Filter LP Filter 4027A Power Sensor 1 2 3 4 5 Error Budget Error Source 4027A Limit R2 Calibration Standards Uncertainty ± 0.4% 0.16 Frequency Response Error 0.0% 0 Dynamic Linearity ± 0.5% 0.25 Temperature Effects ± 0.5% 0.25 Noise ± 0.5% 0.25 Worst Case Error ± 1.9% RSS (Probable) Error ± 0.91% Notes: Based upon a temperature range of +35 degrees celcius. 4027A Typical Linearity, 13.56 MHz 0.30 0.20 4027A10M Serial # 0.10 S/N#11758 S/N#11759 0.00 0.50 1.00 1.50 2.00 2.50 3.00 S/N#11822 S/N#11823 S/N#11817 % Error -0.10 Power Levels (kW) S/N#11818 S/N#11820 S/N#11821 -0.20 S/N#11824 S/N#11825 S/N#11826 -0.30 -0.40 -0.50 4027A Typical Linearity, 12 MHz 4027A10M Serial # Power Levels in (kW) .5kw 1kw 1.5kw 2.0kw 2.5kw 3.0kw 11596 11597 11598 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.60 0.20 0.10 -0.07 -0.50 0.00 Short Term Drift at Elevated Power Level 4027A10M S/N#11820 100w 1000w 3000w Time (Min) 15 4027A10M 4027A10M 4027A10M Min Warmup MCS-59 (w) (w) MCS-59 (w) (w) MCS-59 (w) (w) 0 99.8 99.8 1009 1009 3.01 3.01 5 99.6 99.6 1008 1008 3.01 3.01 10 99.5 99.5 1009 1009 3.00 3.00 15 99.6 99.6 1010 1010 2.99 2.99 20 99.7 99.7 1012 1012 2.98 2.98 25 100 100 1013 1013 2.98 2.98 30 100 100 1011 1011 2.99 2.99 35 100 100 1009 1009 2.99 2.99 40 99.6 99.6 1011 1011 2.98 2.98 45 99.3 99.3 1013 1013 2.99 2.99 50 99.6 99.6 1015 1015 2.99 2.99 55 99.4 99.4 1013 1013 2.97 2.97 60 99.3 99.3 1009 1009 2.97 2.97 Higher Power 4028 Capability 4028A 1-5/8” or 3-1/8” Transmission Line • Similar Accuracy To Other 4027 Models • Uses Larger Transmission Line (1-5/8” or 3-1/8”) Flanged or Unflanged • Power Measurement Capability Up To 40Kw Typical Field Power Measurement Systems A Typical Power Measurement Setup Utilizing a Directional Coupler Wattmeter Forward watts Reverse watts Amplifier Source p2 p1 p1 p1 p2 Directional Coupler p2 High Power Termination Coupler Based Measurements Advantages: • Typically not limited by power- very little power dissipated • Typically have good thru line reflection coefficients • Forward Power readings are basically isolated from load stability issues • Allows in-line monitoring of signal with actual system load Disadvantages: • Must know the coupling value very accurately • Directivity limits reflected power reading • Frequency bandwidth limited A Typical Power Measurement Setup utilizing an Attenuator and Thermal Power Meter Wattmeter Thermal Sensor Amplifier Forward watts Source Attenuator • Knowledge of the attenuation factor and stability is crucial to making a precise power measurement Attenuators and Their Effect On Accuracy Forward Power attenuation +/- attn. tolerance Output power Low reading High reading +/- % fwd power change 1000 30 0.01 1 0.998 1.002 0.23 1000 30 0.1 1 0.977 1.023 2.28 1000 30 0.25 1 0.944 1.059 5.59 1000 30 0.5 1 0.891 1.122 10.87 • Assuming nominal attenuation value can lead to significant errors • Errors can be minimized by calibrating the attenuator at the specific frequency or band of frequencies Attenuator Based Measurements Advantages: • Wideband frequency response, DC coupled • Convenient to use, eliminates a termination Disadvantages: • Limited in power dissipation • Attenuation accuracy is often not precise, not as stable • Reflection coefficients are generally higher Attenuator Based Measurements • Uncertainties Associated With This System – Input Mismatch Uncertainty (Typically Small Due to Low Input VSWR) – Output Mismatch Uncertainty – Uncertainties Associated With Thermal Power Meter – Attenuation Factor Uncertainty – Stability of Attenuation Factor Over Temperature – Additional Thermal Errors Due To Excessive Load Temperatures Affecting Thermal Power Sensor When an attenuator is used, obtain the calibrated attenuation factor from the manufacturer (or make the measurement yourself) for best possible precision measurements. Summary • National Standards Traceability- Challenges & Bird’s Solution – Bird’s multi-path solution and test capabilities make it unique in the industry • RF Metrology Paths at Bird Electronic Corporation – High accuracy transfer of standards at every step of the way – Know the concepts behind the error sources in a test setup • 4020 Series Power Sensors and the 4421 Power Meter – +/- 1% power sensor ideal for semiconductor industry • Typical Field Power Measurement Systems – Know your system and the errors associated with it