Conny Andersson
IEC 60118-4
Conny.Andersson@edin.se
www.univox.eu
Receiver = T-coil in Hearing Instrument (HI)
No current consumption
Transmitter = Loop Amplifier and Loop
World Wide Standard
www.univox.eu
Acoustic
Output
Input
Gain +60dB
Speech level=70dB
@1m distance
Loop Level:
Fixed @ 400mA/m
for all signals
Hearing
Instrument
M
T
switch
130dB
Coil Sensitivity
Same output level for M and T-position
www.univox.eu
130dB
Switching between M and T will give
approximately the same output level
It could be stated differently:
0dB difference between M and T-position
Related to acoustic level: The acoustic
output is the same for T and M-position
with a 70dB acoustical input signal
www.univox.eu
Hearing Instrument (HI)
Many parameters are fitted for the
individual. Can not be standardized
Hearing Loop system
International
Electrotechnical
Commission
www.univox.eu
All parameters are standardized
American National Standard Institute has
many abbreviations for T-coil functions
3.4 Terms relating to measurements with inductive inputs
3.4.1 high frequency average SPL for an inductive telephone simulator (HFA-SPLITS).
The high frequency
average of the SPLITS levels (see Clause 3.4.5). The abbreviation for this term is HFASPLITS.
3.4.2 high frequency average SPL in a vertical magnetic field (HFA-SPLIV). The highfrequency
average of the SPLIV levels (see Clause 3.4.6). The abbreviation for this term is HFA-SPLIV.
3.4.3 relative simulated equivalent telephone sensitivity (RSETS). Difference in decibels
obtained
by subtracting the RTG + 60 dB SPL from the HFA-SPLITS. The abbreviation for this term is
RSETS.
3.4.4 relative test loop sensitivity (RTLS). Difference in decibels obtained by subtracting
the RTG +
60 dB SPL from the HFA-SPLIV. The abbreviation for this term is RTLS.
3.4.5 sound pressure level for an inductive telephone simulator (SPLITS). Sound
pressure level
developed in the coupler by a hearing aid with the gain control at the RTS when the input is the
magnetic field generated by a Telephone Magnetic Field Simulator (refer to Clause 3.2.5). The
abbreviation for this term is SPLITS.
3.4.6 sound pressure level in a vertical magnetic field (SPLIV). Sound pressure level
developed in
the coupler by a hearing aid with the gain control at the RTS when the input is a 31.6 mA/m
sinusoidal
alternating magnetic field parallel to the vertical reference and the input selector of the hearing
aid set
to T-position. The abbreviation for this term is SPLIV.
Good News, ANSI S3.22 committee
has decided to make T-coil
specification mandatory for quality
test inspection
www.univox.eu
Magnetic Field Strength, 400mA/m
Coverage area/space (sitting and standing),
Should have minimum variations (+-3dB)
www.univox.eu
Variation less
than +-3dB
www.univox.eu
Magnetic field strength from electrical
appliances
Received directly by the T-coil
NOT by the hearing loop system
Less problematic due to stronger
regulations of electrical environment
www.univox.eu
Hearing instrument, microphone/telecoil response
Mic
T-coil
www.univox.eu
Hearing instrument, microphone/telecoil response
Reduced low
frequency
sensitivity
Mic
25dB
Tele-coil
www.univox.eu
Hearing
Hearinginstrument,
intrument, microphone/telecoil response
Mic
Tele-coil
www.univox.eu
Hearing instrument, microphone/telecoil response
Within
+-3dB
+-3dB
Mic
Tele-coil
www.univox.eu
It is therefore impossible to judge
the full quality of a loop system by
using a Hearing Instrument
Measurements are mandatory
www.univox.eu
The specifications for most of the
parameters for loops are for vertical
directions
This corresponds to HI
manufacturers placing the T-coil in
vertical position
www.univox.eu
Vertical field
0dB=full level
-3dB (@45°)
Silence!
www.univox.eu
-3dB (@45°)
Silence!
Vertical AND Horizontal Field
0dB=full level
-3dB (@45°)
full level
www.univox.eu
-3dB (@45°)
full level
Increased low frequency response
Hum harmonics remover
Default sensitivity should follow the standard
of 400mA/m (125ms, RMS)
If the sensitivity is set higher in the Hearing
Instrument there is a strong risk for high
distortion, clipping/saturation when entering
a correct installed Loop System.
A weak loop must not be compensated by
increasing the T-coil sensitivity with a
decreased Signal to Noise ratio.
www.univox.eu
Several channels and modulated/coded
loop system.
One “standard loop system for world
wide use in theatres etc
Other channels for school and domestic
use. Decreased Signal to Noise ratio.
www.univox.eu
www.univox.eu
www.univox.eu
Acoustic
Output
Input
Gain +60dB
Speech level=70dB
@1m distance
Loop Level:
Fixed @ 400mA/m
for all signals
Hearing
Instrument
M
T
switch
130dB
Coil Sensitivity
Same output level for M and T-position
www.univox.eu
130dB
Acoustic
Output
Input
Long-term signals:
Speech level=70dB
@1m distance
Loop Level:
Fixed @ 100mA/m
for all signals
Gain +60dB
Hearing
Instrument
M
T
switch
130dB
Coil Sensitivity
Same output level for M and T-position
www.univox.eu
130dB
Acoustic
Output
Input
Long-term signals:
Speech level=60dB
@1m distance
Loop Level:
Fixed @ 31.6 mA/m
for all signals
Gain +60dB
Hearing
Instrument
M
T
switch
120dB
Coil Sensitivity
Same output level for M and T-position
www.univox.eu
120dB