Experiment 3. Gravimetric Determination of Calcium
Gravimetric Analysis
Gravimetric analysis will be used to determine how much calcium is in an unknown. This
lab is over two lab periods. Day 1 consists of determining the accurate masses of 3
crucibles. Day 2 consist of a precipitation reaction to convert the calcium solution into a
solid, that can be weighed.
Drying
Heating to drive off surface moisture is done by microwave (MW) as opposed to
conventional heating methods, like a hot plate. MW heating saves a lot of time because
conventional heating heats up the hot plate, then the crucible, then the precipitate. The
crucible then has to cool down because weighing hot will yield false readings on the
analytical balance. Heating should be done with sample in a weighing bottle (or
crucible), cap off, placed in a beaker. Be sure you can identify your beaker if multiple
students load into the MW at the same time. After heating, the sample must cool in a
desiccator so that moisture from the air does not adhere to the crucible. The weighing
bottle (or crucible) should be capped/sealed while in the desiccator. A 20-minute
cooling time is needed before weighing.
Precipitation
Precipitation occurs in two steps: nucleation and crystal growth. Nucleation is the
formation of very small, invisible, crystal nuclei, while crystal growth is further growth of
these original nuclei. The fewer original nuclei are formed, the larger will be the final
crystals. This will make filtration much easier. If supersaturation of the solution is high,
many nuclei will form; if it is low and controlled, fewer will form. We normally add only
one or two drops of precipitating agent at first to allow a relatively small number of
nuclei to form. After waiting a minute or so, more precipitating agent is added, slowly,
again fostering crystal growth rather than formation of new nuclei. Slow growth of old
crystals instead of rapid growth of new ones also reduces coprecipitation and other
contamination by foreign ions. Complete precipitation is then tested for in any
gravimetric procedure.
Filtering
This can be accomplished with sintered glass or porcelain crucibles. These have a
porous bottom through which solvent is pulled by vacuum. We will use porous porcelain
crucibles. The crucibles must be previously cleaned and dried to a constant weight.
Thorough cleaning generally involves drawing an acid through, then complete rinsing
with deionized water as a final step.
Drying and weighing to a constant weight (+0.2 – 0.3 mg) follow. (This means that two
consecutive heating-cooling-weighting cycles should produce the same weight, say to +
0.0003 g or so.) You must be able to distinguish one crucible from another. The filtering
step actually consists of decanting the “mother liquor”, washing the precipitate in the
original beaker, transferring it to the filtering crucible, and further washing in the crucible.
Read carefully the sections in the text concerning these operations. Washing in the
beaker results in a more thoroughly decontaminated precipitate. Once the precipitate is
packed into the crucible, washing becomes more difficult. It is a provable fact that
washing with several portions of wash liquid is much more effective than one wash with
the same total volume. See the text for this discussion. So as to prevent “creeping” of
the precipitate over the rim; never fill the crucible more than about ¾ full. Before
halting the washing procedure, a test is normally made to be sure that all excess
precipitating agent has been removed. To do this, a small quantity of the wash liquid is
caught and tested for any remaining precipitating agent.
Most “errors” are made in the precipitate transfer process. You must “quantitatively
transfer” the precipitate to the crucible from the beaker. Any loss or precipitate left
behind will result in low results. Transfer should be aided by a stirring rod. A separate
stirring rod used for each sample to prevent sample from crucible 2 into crucible 1.
Never lay the stirring rod down on the lab bench. Not only can it become contaminated,
but if any particles of precipitate were on it, they may be lost. You can lay these across
in a beaker.
After complete transfer and washing, the precipitate is ready to be dried to a constant
weight. Some precipitates are “ignited” to a known chemical species, normally an
oxide, if the chemical form of the original precipitate is unknown or of variable
composition. Note that once the crucibles are cleaned, they should not be handled
again with the fingers until the final constant weight is obtained. Again + 0.2 – 0.3 mg
agreement is desired on two successive heating-cooling-weighing cycles. The final step
is calculation of the amount or percentage of the desired species in the original sample.
Grading Scale:
Grade (%)
error (ppt)
100
<3
95
<3.7
90
<4.6
85
<5.8
80
<7.3
75
<9.1
70
<11.4
65
<14.3
60
<17.8
55
<22.3
50
>22.3
Gravimetric Determination of Calcium as CaC2O4• H2O
Calcium ion can be analyzed by precipitation with oxalate in basic solution to form
CaC2O4•H2O. The precipitate is soluble in acidic solution because the oxalate anion is a
weak base. Large, easily filtered, relatively pure crystals of product will be obtained if
the precipitation is carried out slowly. This can be done by dissolving Ca2+ and C2O4 2in acidic solution and gradually raising the pH by thermal decomposition of urea
(Reaction 27-2 in the textbook).
Procedure Day 1
1. You should have 3 crucibles and a desiccator in your cabinet. Desiccant (clay like
absorbent material) has likely absorbed all the water it can. So “dried” desiccant
is needed.
2. Empty the desiccant that is in your desiccator into the plastic container. Do not
throw away.
3. Put fresh desiccant (in oven) in desiccator. It was oven heated overnight to
remove absorbed water.
4. Crucibles should be free of solid material from previous semesters. If not, clean
out with spatula.
5. Setup vacuum filtration with a side arm flask, rubber fitting, and a vacuum trap.
This will be used to wash your porous crucibles with acid. The crucibles will fit
inside the rubber fittings which are shown in blue in the image below. The
vacuum trap is to ensure that no liquid gets pulled into the house vacuum
system. If you have glass crucible (pictured far left), the fitting is a little different,
ask TA.
6. Clean 3 porous porcelain crucibles with 1:4 HCl:H2O by placing the crucible into
the fitting and running ~20mL HCl through each crucible using house vacuum,
followed by DI water.
7. Using a microwave, dry crucibles to constant mass. See bullet points below. Use
a Kimwipe or tongs, not your fingers, to handle the crucibles.
Constant Mass of Crucibles
1. It is wise to use the same analytical balance for all weights. If there is fluctuation
in the mass, then cover the balance with the plastic cover. This will keep any
airflow from affecting the mass.
2. Dry in MW by heating 4min, cool 2min on benchtop, heat 4min, cool 15min in
desiccator
3. Weigh on analytical balance (record all 4 dec places). Do not weigh hot.
4. MW again, 4min. Cool in desi, 15 min
5. Weigh crucible (use same analytical balance). Do not weigh hot.
6. Compare weights, are the masses close? < 0.3mg?
7. Repeat procedure with 4-min heating periods until successive masses agree
within 0.3 mg.
Procedure Day 2
8. Immediately turn on your hot plate to 100°C if using the large 10”x10” topped hot
plates and a setting of 5 for the older smaller hot plates.
9. Make an ice water bath: half ice, half water.
10. Collect new desiccant in your desiccator.
11. Collect about 50 ml of unknow solution.
12. Use a few small portions of unknown to rinse a 25-mL transfer pipet and discard
the washings.
13. Transfer 25 mL of unknown to each of three 250mL beakers. and dilute each with
~75 mL of 0.1 M HCl. Add 5 drops of bromocresol green indicator to each beaker
(yellow < pH 3.5, blue > pH 5.4).
14. Add a stir bar to your beaker.
15. Add ~25 mL of ammonium oxalate to each beaker while stirring.
16. Add ~15 g of urea to each sample, cover with a watch glass.
17. Boil until indicator turns blue (approx. 35min). Hot plate values of 230°C are
needed when using the large 10” x 10” topped hot plates.
18. Filter each hot solution through a weighed crucible, using suction. Add ~3 mL of
ice-cold water to the beaker, use a rubber policeman to help transfer remaining
solid to the funnel.
19. Repeat this procedure with small portions of ice-cold water until all precipitate
has been transferred.
20. Use two 10-mL portions of ice-cold water to rinse each beaker. Pour washings
over precipitate.
21. Dry precipitate, 1 min with vacuum suction, then 4min in microwave, followed by
several 2-min periods, with cooling for 15 min before weighing. The water of
crystallization is not lost. Bring each filter to constant mass. The product is
somewhat hygroscopic, so remove only one crucible from the desiccator at a
time and keep the desiccator lid closed.
22. Calculate the average molarity of Ca2+ in the unknown solution. Report standard
deviation.
Name ______________________________
_____
Date ________________ Section
Gravimetric Analysis of Calcium Report Sheet
Weight of empty crucible (g)
___________
Crucible
Crucible
Crucible
1
2
3
___________
Aliquot vol. of unknown Ca2+(mL) ___________
___________
Weight of crucible with precip. (g)
___________
Weight of precip. (g)
___________
___________
___________
___________
___________
___________
___________
___________
___________
(CaC2O4·H2O)
Moles of precip.
___________
Moles of Ca in precip.
___________
___________
___________
Molarity of Ca2+ in unknown solution (M) ___________
___________
___________
Average molarity of unknown Ca2+ solution (M)
___________
Std. dev. ____________
Show a sample calculation for the molarity of Ca2+ in the unknown solution below, and
on the reverse side, if necessary. Please indicate the formula weights used.