Biomedical Instrumentation Notes
1
Pulse Oximeter
Reference: R.S. Khandpur, Handbook of Biomedical Instrumentation, pp. 289–
300
Definition
A pulse oximeter is a non-invasive instrument used to measure the oxygen saturation of
arterial blood (SpO2 ) and pulse rate using optical absorption techniques.
Principle (7 Marks)
• Pulse oximetry is based on the different light absorption properties of oxygenated and
deoxygenated haemoglobin.
• Oxygenated haemoglobin (HbO2 ) and reduced haemoglobin (Hb) absorb light differently
at specific wavelengths.
• Two wavelengths are used: red light (about 660 nm) and infrared light (about 940 nm).
• Oxygenated haemoglobin absorbs more infrared light and allows more red light to pass.
• Deoxygenated haemoglobin absorbs more red light and allows more infrared light to pass.
• Light transmitted through tissue varies according to the oxygen saturation level of haemoglobin.
• The ratio of absorption of red and infrared light is used to determine oxygen saturation.
• A microprocessor uses stored calibration curves to convert this ratio into SpO2 value.
Block Schematic Diagram
Red LED and Infrared LED
↓
Patient Tissue
↓
Photodetector
↓
1
Amplifier
↓
Signal Processing Unit
↓
Microprocessor
↓
Display
Working
1. Two LEDs emit red and infrared light through the patient’s finger or ear.
2. Oxygenated and deoxygenated haemoglobin absorb light differently.
3. A photodetector receives the transmitted light signal.
4. The signal is amplified and processed to separate the pulsatile component.
5. The microprocessor calculates the ratio of the two wavelengths.
6. Oxygen saturation (SpO2 ) and pulse rate are displayed.
Applications (7 Marks)
• Continuous monitoring of oxygen saturation in intensive care units.
• Monitoring patients during surgical operations.
• Monitoring respiratory diseases such as asthma and COPD.
• Monitoring premature infants in neonatal care units.
• Emergency monitoring in ambulances.
• Detection of hypoxemia in critical patients.
• Monitoring effectiveness of oxygen therapy.
2
Blood Cell Counter
Reference: R.S. Khandpur, pp. 480–498
Definition
A blood cell counter is an automated electronic instrument used to count and analyze different types of blood cells such as red blood cells, white blood cells and platelets.
2
Principle (7 Marks)
• Most electronic blood cell counters operate on the Coulter principle.
• Blood cells are suspended in an electrically conductive electrolyte solution.
• The diluted blood sample flows through a small aperture between two electrodes.
• A constant electric current flows through the aperture.
• When a cell passes through the aperture it displaces electrolyte.
• This displacement causes a temporary increase in electrical resistance.
• The resistance change produces a voltage pulse.
• The number of pulses indicates the number of cells present.
• The amplitude of each pulse corresponds to the volume of the cell.
Block Schematic Diagram
Blood Sample
↓
Dilution System
↓
Aperture Tube
↓
Electrodes
↓
Amplifier
↓
Pulse Analyzer
↓
Counter and Display
Working (7 Marks)
• The blood sample is first diluted with electrolyte solution.
• The diluted sample is aspirated through a narrow aperture tube.
• Electrodes are placed on either side of the aperture.
• A constant current flows between the electrodes.
• When a blood cell passes through the aperture it changes the resistance.
• This resistance change generates a voltage pulse.
3
• The pulse is amplified by an electronic amplifier.
• Pulse height analysis determines the size of the cell.
• The number of pulses counted corresponds to the number of cells.
Applications (7 Marks)
• Complete blood count testing in hospitals.
• Diagnosis of anemia.
• Detection of infections using white blood cell count.
• Diagnosis of leukemia and blood disorders.
• Routine medical health check-ups.
• Hematology laboratory analysis.
• Monitoring treatment such as chemotherapy.
3
Defibrillator
Reference: R.S. Khandpur, pp. 730–747
Definition
A defibrillator is a medical device used to deliver a controlled electrical shock to the heart
to terminate ventricular fibrillation and restore normal cardiac rhythm.
Principle (7 Marks)
• Ventricular fibrillation causes chaotic electrical activity in the heart.
• During fibrillation the heart cannot pump blood effectively.
• Defibrillation applies a strong electrical shock across the heart.
• The shock depolarizes all cardiac muscle fibers simultaneously.
• This interrupts the abnormal electrical activity.
• The heart cells enter a refractory period after the shock.
• The natural pacemaker (SA node) regains control of the rhythm.
• Normal cardiac rhythm is restored.
4
Block Schematic Diagram
AC Power Supply
↓
Step-up Transformer
↓
Rectifier
↓
Energy Storage Capacitor
↓
Control Circuit
↓
Electrodes
Working (7 Marks)
• AC supply is converted into high voltage using a step-up transformer.
• The voltage is rectified and used to charge a capacitor.
• The capacitor stores electrical energy.
• The operator selects the required energy level.
• Electrodes are placed on the patient’s chest.
• When triggered the capacitor discharges through the electrodes.
• The electrical shock passes through the heart.
• The shock depolarizes the myocardium and stops fibrillation.
• The heart resumes normal rhythm.
Applications (7 Marks)
• Treatment of ventricular fibrillation.
• Treatment of ventricular tachycardia.
• Emergency cardiac resuscitation.
• Use in hospitals and intensive care units.
• Portable use in ambulances.
• Implantable cardioverter defibrillators for high-risk patients.
• Emergency life-saving device in cardiac arrest.
5
4
Pacemaker
Reference: R.S. Khandpur, Handbook of Biomedical Instrumentation, pp. 709–
729
Definition
A cardiac pacemaker is an electronic medical device used to regulate the heartbeat by delivering electrical impulses to the heart when the natural pacemaker fails to maintain a normal
rhythm.
Principle (7 Marks)
• The normal rhythm of the heart is controlled by the sinoatrial (SA) node.
• When the SA node fails or heart rate becomes abnormally slow, artificial pacing is required.
• A pacemaker generates periodic electrical pulses to stimulate cardiac muscle.
• These electrical pulses depolarize the myocardium and initiate heart contraction.
• The pulses are delivered through electrodes placed in contact with the heart.
• The pacemaker maintains a predetermined heart rate using a timing circuit.
• Modern pacemakers can sense intrinsic heart activity and adjust pacing accordingly.
• The device ensures that the heart maintains adequate cardiac output.
Block Schematic Diagram
Battery (Power Supply)
↓
Timing Circuit
↓
Pulse Generator
↓
Output Amplifier
↓
Electrodes / Leads
↓
Heart
6
Working (7 Marks)
• The pacemaker continuously monitors the electrical activity of the heart.
• If the heart rate falls below a preset level, the pacing circuit is activated.
• A timing circuit determines the interval between pacing pulses.
• The pulse generator produces electrical impulses of controlled amplitude and duration.
• These impulses are amplified by an output stage.
• The electrical pulses are delivered to the heart through implanted electrodes.
• The stimulus depolarizes the cardiac muscle causing heart contraction.
• The pacemaker continues to generate pulses at regular intervals to maintain normal heart
rhythm.
• Modern pacemakers can sense natural heartbeats and inhibit unnecessary pacing.
Applications (7 Marks)
• Treatment of bradycardia (abnormally slow heart rate).
• Treatment of heart block conditions.
• Control of cardiac arrhythmias.
• Post-operative cardiac rhythm stabilization.
• Implantable devices for long-term cardiac rhythm control.
• Temporary pacing during cardiac emergencies.
• Maintaining adequate cardiac output in patients with conduction disorders.
5
Ventilator
Reference: R.S. Khandpur, Handbook of Biomedical Instrumentation, pp. 859–
873
Definition
A ventilator is a medical device used to assist or replace spontaneous breathing by delivering
a controlled mixture of air and oxygen into the lungs when a patient is unable to breathe
adequately.
7
Principle (7 Marks)
• Ventilators operate on the principle of artificial ventilation.
• They deliver gas to the lungs using positive pressure.
• During inspiration, gas is pushed into the lungs causing lung expansion.
• During expiration, pressure is released allowing the patient to exhale.
• The ventilator controls parameters such as tidal volume, respiratory rate and airway pressure.
• Sensors monitor pressure, flow and volume of air delivered.
• A control system adjusts the ventilation cycle according to patient requirements.
• Alarm systems warn medical staff if abnormal conditions occur.
Block Schematic Diagram
Air and Oxygen Supply
↓
Gas Mixing System
↓
Pressure and Flow Regulator
↓
Control System (Microprocessor)
↓
Inspiratory Valve
↓
Patient Breathing Circuit
↓
Exhalation Valve
↓
Monitoring and Alarm System
Working (7 Marks)
• Air and oxygen are supplied from cylinders or hospital pipeline systems.
• The gases are mixed in a gas mixing unit in the required proportions.
• Pressure and flow regulators control the gas delivery parameters.
• A microprocessor-based control system regulates breathing cycles.
• During inspiration, the inspiratory valve opens allowing gas to enter the lungs.
8
• The lungs expand as gas is delivered under positive pressure.
• After the inspiratory phase, the exhalation valve opens.
• The patient exhales either passively or through controlled pressure release.
• Sensors monitor airway pressure, flow rate and volume.
• Alarm systems activate if abnormal pressure or flow conditions occur.
Applications (7 Marks)
• Respiratory support for critically ill patients in intensive care units.
• Artificial respiration during surgical procedures and anaesthesia.
• Treatment of respiratory failure.
• Support for patients with lung diseases such as pneumonia or COPD.
• Neonatal ventilation for premature infants.
• Emergency respiratory support in ambulances.
• Life-support systems in severe trauma or cardiac arrest patients.
9