Tutorial 6: Chromatography
Pre-read: chromatographic separation principles and how column operation determines
purity, yield, and process scalability.
1. (a) Protein X (distribution coefficient 0.1) is being separated by chromatography from
protein Y (distribution coefficient 1) using a 50 cm long/10 mm diameter column. The void
fraction is 0.25 and the characteristic peak width of protein X is 0.52 min. Determine the
selectivity, resolution and theoretical plate height for the mobile phase flow rate of 10 ml
min-1.
(b) Draw a Gaussian chromatographic peak. Label the peak with 4σ, 2.35σ and inflection
point and mention their significance.
(c) Also show 6σ resolution of two peaks. Comment on its significance.
(d) What is the relationship between number of theoretical plates (N) and shape of analyte
peak (broader and narrower)? Show it with a diagram of two peaks (broad and narrow).
(e) Describe peak asymmetry factor (T) with a tailing chromatographic peak.
(All the plots should be labelled properly)
2. Gel chromatography is used for commercial-scale purification of a proteinaceous
diphtheria toxoid from Corynebacterium diphtheriae supernatant. In the laboratory, a small
column of 1.5 cm inner diameter and height 0.4 m is packed with 10 g dry Sephadex gel;
the void volume is measured as 23 ml. A sample containing the toxoid and impurities is
injected into the column. At a liquid flow rate of 14 ml min-1, the elution volume for the
toxoid is 29 ml and the elution volume for the principal impurity is 45 ml. A column of
height 0.6 m and diameter 0.5 m is available for large-scale gel chromatography. The same
type of packing is used; the void fraction and the ratio of pore volume to total bed volume
remain the same as in the laboratory-scale column. The liquid flow rate in the large column
is scaled up in proportion to the column cross-sectional area; the flow patterns in both
columns can be assumed identical. The water regain value for the packing is given by the
manufacturer as 0.0035 m3 kg-1 dry gel.
(a) Determine the partition coefficients for the toxoid and impurity.
(b) Estimate the elution volumes in the commercial-scale column.
(c) What is the volumetric flow rate in the large column?
(d) Estimate the retention time of toxoid in the large column.
3. In a chromatographic separation column used for the adsorption of solute A onto an
adsorbent solid B, the adsorption isotherm is given by
πΆπΆπ π = ππ1 πΆπΆπΏπΏ3 = ππ(πΆπΆπΏπΏ )
where CS is mg solute adsorbed/mg adsorbent, CL is the solute concentration in liquid medium
(mg solute/ml liquid), and k1 is a constant.
K1 = 0.2 (mg solute adsorbed/mg adsorbent)/(mg solute/ml liquid)3. The porosity (void
fraction) of the packed column is ε = 0.35. The cross-sectional area of the column is 10 cm2
and M is 5 g adsorbent per 100 ml column volume. The volume of the liquid added is ΔV= 250
ml.
(a) Determine the position of the solute band in the column when the solute concentration
in the liquid phase at equilibrium is CL = 5×10-2 mg ml-1.
(b) Find the ratio of the travel distance of solute A (LA) to that of solvent B in the column
when CL = 5 × 10-2 mg ml-1.
4. Consider an ion-exchange chromatography column used to purify 20 g of a particular
protein. At a superficial velocity of 20 cm h-1 the peak exits the column with ymax at 80 min.
The standard deviation of the peak is 10 min. Estimate:
(a) How long must the column be run to achieve 90% yield?
(b) If we increase the flow to 40 cm h-1, how long must we run for this same yield if the
process is controlled by external mass transfer?
(use the table attached at the end for calculating the error functions)
5. Human insulin A and B from recombinant Escherichia coli are separated using pilot-scale
affinity chromatography. Laboratory studies show that the capacity factors for the proteins
are 0.85 for the A-chain and 1.05 for the B-chain. The dependence of HETP on liquid
velocity satisfies the following type of equation:
π΄π΄
π»π» = + π΅π΅π΅π΅ + πΆπΆ
π’π’
where u is the linear liquid velocity and A, B, and C are constants. Values of A, B, and C
for the insulin system are 2×10-9 m2s-1, 1.5 s, and 5.7×10-5 m, respectively.
(a) What is the minimum HETP? At what liquid velocity is the minimum HETP obtained?
(b) If a column with 1 m length/25 cm inner diameter is used with a liquid flow rate of 0.31
litres min-1, will the two insulin chains be separated completely?
Table: Error function values