Radiochemistry Practical Training for chemists Freie Universität Berlin Institute for chemistry and biochemistry 1 Contents Rules for working in radioactive laboratories Comparison of various detectors and nuclides Decontamination of radioactively contaminated surfaces Calculation of the γ-radiation dose rate γ-spectrometry; recording of γ-spectra β-spectroscopy (liquid-scintillation counter) Backscattering of β--radiation Self-absorption of β--radiation The Thorium Generator – Determination of the half-lives of different members of the natural thorium decay series Responsible for content Prof. Dr. U. Abram (abram@chemie.fu-berlin.de) Dr. Adelheid Hagenbach (hagenb@zedat.fu-berlin.de) 2 Rules for working in radiochemical laboratories Only the bare necessities (writing material, calculator) are allowed into a radioactive laboratory. Jackets as well as book-bags are to be stored outside of the radioactive laboratory in available lockers. Individuals, handling non-confined radioactive materials are obliged to wear suitable protective clothing (laboratory jacket, overshoes, disposable gloves) and equipment (dosimeter). When handling radioactive substances, eating, drinking, smoking, and applying make-up are prohibited in the laboratory. (§70-Handling of non-confined radioactive substances – StrlSchV). Special attention should be paid to open wounds. When working with non-confined radioactive substances, procedures which avoid the incorporation of radioactive substances and the contamination of participants are used. The occurrence of contamination immediately must be reported to an assistant. Operation of equipment with the mouth is never allowed (application of piston-driven pipettes or peleus balls). Special attention is required when handling dust-developing, radioactive solids (danger due to inhalation). According to §4 of StrlSchG, anyone who carries out a task must adhere to the following obligations. Any needless radioactive exposure or contamination of people, property or the environment must be avoided. Taking into account the state of science and technology and bearing in mind all circumstances of a particular case, any radioactive exposure or contamination of people, property or the environment must be kept as low as possible and below the limit defined in §8principles in the protection against radiation – StrlSchG. When handling non-confined radioactive substances, disposable gloves must be worn at all times. Disposable gloves are always regarded as being contaminated. All work with non-confined radioactive substances is performed in designated plastic trays (according to the “principle of double containment”). When transporting radioactive substances, sealable transport containers must always be used (“principle of double containment”). Non-confined substances are never to be touched by the hands (contamination danger). Tweezers must always be used. Storage containers are to be properly labelled and may only be used for the storage of radioactive substances (§91labelling obligation-StrlSchV). Solid radioactive wastes are stored in available, specially designated containers. Combustible and non-combustible wastes are collected separately. Liquid radioactive wastes are poured into available, specially designated containers – never down the drain (§99-Protection of air, water, and soil –StrlSchV). Prior to and after completion of the - Use the gloves-monitor experiment: - Wash your hands 3 4 Chapter 1 Comparison of Various Detectors and Nuclides Task Investigate which detectors are best suited for the detection of α-, β-, and γ-radiation! The nuclides 241Am, 14C, 204Tl and 60Co are used. Fundamentals 1. Production of nuclear radiation Nuclear radiation is produced by the decay of unstable atomic nuclei. These nuclei are stabilized by emitting particles (α- and β-radiation) or electromagnetic waves (γ-radiation). The emission of alpha and beta radiation is always accompanied by a nuclear transformation, meaning that a different chemical element is formed. The emission of γ-radiation itself does not lead to the formation of a different element; the same chemical nucleus simply goes from a higher energy to a lower energy state (which may be the ground state). However, γ-radiation often accompanies α- or β-decay of atomic nuclei. Other types of nuclear radiation are: electron capture reactions (“capture” of an electron near the nucleus by a proton), proton radiation, neutron radiation, or spontaneous fission of the nucleus. 1.1. α-Radiation During α-decay, helium nuclei are emitted. The mass number of the nucleus is reduced by four units and its atomic number by two. X Y A → X −4 Y −2 B + 24He Typical elements experiencing α-decay are the heavier elements toward the bottom of the periodic table of elements. α-Decay becomes possible when the emission of the (extremely stable) αparticle is energetically favorable for the entire system: A → B + 24 He + ΔE ΔE = (MA - MB - MHe) · c2 Fig. 1: Formation of α-radiation (241Am → 237Np + He2+) The emitted α-particles possess a defined energy, which depends on the particular nuclear decay. α-Spectra are line spectra, which can be used to characterize nuclides. Besides αspectra with just one line, there are also examples of α-spectra with multiple energetic transitions. This can be explained by the favorable formation of decay products in different excited states. 1 The half lives of α-emitters vary widely. Besides very short-lived nuclei (for example 212Po, t1/2 = 3 x 10-7 s), there also exists very long-lived α-emitters (for example 232Th, t1/2 = 1.4 x 1010 a). 1.2. ß-Radiation β-Radiation can be described as the conversion of a neutron to a proton (β--radiation) or as the conversion of a proton to a neutron (β+radiation). The result of these conversions is an energetically more favorable situation in the nucleus, achieved through a “balanced” ratio of protons and neutrons. The number of neutrons that are needed to stabilize an atomic nucleus Fig. 2: Formation of β--radiation depends on the number of protons. The stable (14C → 14N + e-) nuclei of lighter elements have a proton/neutron ratio of about 1:1, while the stabilization of the heavier elements requires more neutrons than protons. A line drawn along the stable nuclei of the nuclide chart is called the “β-stability line.” Nuclides with an excess of neutrons (or a deficit of protons) approach this line through β--decay. Nuclei with a neutron deficit (or proton excess) sustain β+-decay. As a result of these nuclear transformations, negatively charged β-particles (electrons) or β+-particles (positrons) are emitted. At the same time, either an antineutrino (in the case of β--decay) or a neutrino (β+decay) is also emitted (conservation of momentum). The energy that is released by β-decay is distributed among the decay products in such a manner that only a statistical energy distribution with typical parameters Emax (maximum energy when all of the energy is imparted to the β-particle) and Fig. 3: Statistical energy distribution of β--particles Emean (average value of β-energy, approximately one-third of Emax) can be assigned (Fig. 3). During β--decay, the atomic mass number of unchanged. However, the atomic number is increased by one unit. X Y A → X Y +1 the atom remains B + −10ß − + ν During β+-decay, the atomic mass number of the atom also remains unchanged. However, the atomic number is reduced by one unit. X Y A → X Y −1 B + +10ß + + ν Positrons resulting from β+-decay react immediately with surrounding electrons to produce two γ-quants (annihilation of a positron). 2 1.3. γ-Radiation γ-Radiation is created by the transition of an atomic nucleus from an excited state to the ground state (or to another excited state of lower energy). The energy difference between the states is released in the form of an energy-rich electromagnetic radiation (γ-radiation). γRadiation is coupled to α- or β-decay, when these nuclear transformations do not result directly in the ground state of the product nucleus (Fig. 4). γ-Radiation without accompanying particle radiation is obtained from so-called meta-stable nuclear isomers. This refers to nuclides, in which the emission of γ-radiation occurs long enough after their creation that they can be separated from the “mother isotope.” Such metastable nuclear isomers are distinguished with the symbol “m”; for instance 99mTc (t1/2 = 6 h) or 137mBa (t1/2 = 3 min). γ-Radiation of one individual disintegration is monoenergetic and γ-spectra, which are line spectra, are characteristic of the individual de- Fig. 4: Formation of γ-radiation cay processes. The emitted energy corresponds (60Co → 60Ni ( + e- + ν ) + γ-quant) almost exactly (recoil effects) to the energy difference between the excited and ground state. 2. Interaction of radiation with matter The high energy radiation emitted during a nuclear transformation cannot be directly observed. Its existence can be seen only through its interaction with matter. In this process, the kinetic energy of the radiation is distributed to its surroundings. This energy transfer occurs predominately through scattering processes at the electron shells and the atomic nuclei of the irradiated material and is manifested in production of ions. α- radiation The detection and quantitative measurement of nuclear radiation are primarily realized through its ionizing β-- radiation character. The ionizing character of nuclear radiation depends on the energy and specific type of radiation (Fig. γ- radiation 5). This is mainly due to the different ionization mechanisms of α-, β- and γFig. 5: Ionizing effect of nuclear radiation in air radiation. 2.1. Absorption of α-radiation α-Particles emitted during a distinct nuclear decay are monoenergetic. They usually possess energies of 2-10 MeV. Primarily, inelastic collisions with the outer shell electrons are observed. In these processes, 0.1-1 keV of energy is transferred to the electrons. The average energy for the production of an ion in air is 32.5 eV. Collisions with atomic nuclei are relatively rare. Since the mass of α-particles are much larger than that of the particles with which they normally collide, the path of the α-particle undergoes only a slight change in direction. 3 This results in a short and nearly straight flight path, which depends on the energy of the α-particle and the nature of the absorber. α-Radiation is a strongly ionizing form of radiation, which in air forms (assuming an average energy) approx. 10,000 ion pairs per cm traversed (Fig. 6). Fig. 6: Ionizing effect of α-radiation in air 2.2. Absorption of β-radiation The energy released by β-radiation can be distributed to the electron and the antineutrino in different ways. A continuous energy spectrum is obtained for β--radiation, with a characteristic maximum β--energy for each β--emitter. The energy of β--particles is in the range between a few keV and approximately 10 MeV. When β--particles collide with shell electrons, much larger changes of direction are expected than when α-particles collide with shell electrons since the masses of the two collisions partners (β--particles and shell electrons) are roughly equal. Because of these large changes of direction and since β--particles are not monoenergetic, no defined range is observed (see Chapter 7: Backscattering of β--radiation). The ionization density of β−-radiation is substantially smaller than that of αradiation. The formation of approx. 100 ion Fig. 7: Ionizing effect of β-radiation in air pairs per cm air can be assumed for βradiation of average energy (Fig. 7). 2.3. Absorption of γ-radiation γ-Radiation is electromagnetic radiation (photons), which possesses neither resting mass nor charge. Because of this, absorption is only possible through scattering processes at shell electrons and/or atomic nuclei. The most important interaction mechanisms are photoreactions, Compton reactions and pair formation processes. A more detailed description can be found in Chapter 4 “γ-Spectrometry, Recording of γ- Fig. 8: Ionizing effect of γ-radiation in air Spectra“. With the formation of around 10 ion pairs per cm traveled in air, γ-radiation has a relatively small ionization potential (Fig. 8). The relatively small interaction of γ-radiation with matter presents a problem in regard to radiation protection since the radiation has a much larger range than α- or β-radiation and can, therefore, have a direct effect on the body. Since the interactions are proportional to the atomic number of the matter, lead absorbers and lead-glass windows are used while working with γ-radiation. 2.4. Range of nuclear radiation Due to their different interaction mechanisms, α-, β-, and γ-radiation have very different ranges. Since emitted α- and β-particles engage in intense interactions with matter, their energy is completely transferred to their surroundings within a short distance. On one hand, this leads to a large number of ionization events, but, on the other hand, also leads to a small range for 4 these types of radiation. Because of this, they can be shielded quite effectively. However, there exists a large potential danger from the incorporation of α-emitters in particular since a cascade of ionization processes may take place directly in biological tissue, leading to the introduction of a large Radiation Air Aluminium Muscle (water) amount of energy. This causes direct damage to cell material, whereas β- and espe0.32 0.0002 0.0004 α cially γ-radiation are expected to have ß394 0.206 0.438 comparatively smaller interactions. Since γ-radiation concerns electromagnetic Table 1: Range (in cm) of α- and β−-radiation (1 MeV) radiation, specification according to “range” is senseless. Because of this, values of “half-value thickness” for numerous materials are given. These values correspond to the distance in which the intensity of the penetrating radiation is reduced by one-half. Examples of ranges of α- and β-radiation Water Concrete Iron Lead and half-value thicknesses of γ-radiation are arranged in the Tables 1 and 2. These 9.85 4.66 1.47 0.893 values and knowledge of the different ionizing mechanisms play a crucial role in Table 2: Half-value thicknesses (in cm) of γ-radiation (1 MeV) the selection of suitable shielding materials. 3. Detectors for the measurement of ionizing radiation For the measurement of radiation, the following reactions during the irradiation of matter are predominantly utilized: Ionization in gases (Ionization chamber, proportional flow counter, release counter) Scintillation in solids and liquids (α, β- and γ-scintillation counters) Ionization in solids (semiconductor-detectors) 3.1. Ionization of gases (gas-filled detectors) Gas-filled detectors all possess the same fundamental construction. They consist of a gasfilled chamber with a built-in, charged capacitor. When ionizing radiation enters this chamber, ions are produced, which lead to the (partial) discharge of the capacitor. Measurement of the so produced electric flow makes it possible to record the ionization processes (Fig. 9). The level of the detected signals depends on the ionization properties of the respective types of radiation. The actual design of the radiation Fig. 9: General measurement setup for a gas ionization detecmeasuring instruments can vary tor widely from this general description and, therefore, influences the measurement geometry and applications. 5 The strength of the registered signals can be greatly influenced by the applied voltage since only at relatively low applied voltage it is possible to exclusively measure the ionization processes generated by radiation. At higher voltages, secondary ionizations caused by the strongly accelerated charged particles play a larger role. On the one hand, this leads to a considerable increase in the strength of the signal, but, on the other hand, reduces the ability to distinguish between the individual types of radiation (consider the different ionization properties of α-, β-, and γ-radiation). For this reason, three different measurement systems based on gas-filled detectors can be used. These measurement systems are mainly distinguished by the applied voltage and the resulting number of ionization processes (Fig. 10). In an electric field, the charged Fig. 10: Relationship between applied voltage and signal particles, which are formed by strength in a gas-filled detector the influence of radiation, are accelerated according to the applied field strength. The primary and secondary processes, which are observed as a result of this phenomenon, vary according to the field strength, whereby distinct areas are recognizable. U1: U1-U2: U2-U3: >U3: 6 Minimum voltage is required to create an impulse. All produced ion pairs contribute to the impulse without further secondary effects. This is called the “saturation area” because within this area an increase of the field strength does not lead to an increase in the impulse level. Above the field strength corresponding to U2, all created ion pairs contribute to the voltage impulse. A further increase of the field strength above this point causes secondary effects, which enlarge the signal. That means that the ions which are produced in the primary process are accelerated at such a rate that they produce other ions along their path. In this case, the amplification of the signal depends on the type and energy of the radiation. This part of the field strength is called the “proportional area” due to the proportionality between the primary and secondary effects. The acceleration of the electrons is so great that not only the electronic outer shells, but also inner shells of the counter gas are ionized. Upon filling the inner shells, X-rays, which also produce ion pairs, are emitted. In this way, the ionization can encompass the entire counting volume, resulting in a signal that is independent of the strength of the primary ionization. This part of the field strength is referred to as the “Geiger-Müller-Area” or the “release area.” The addition of a suitable quench gas can limit the time-span of these discharges (dead time). In this area of the voltage, it is impossible to distinguish between the types of radiation. An impulse causes the total ionization of the entire counting volume. Through the selection of gases, gas pressure, applied voltage and casing materials of the detector, the response to specific types of radiation can be optimized. Because the specific ionization and the consequent signal produced by an α-particle are considerably greater than those of a β--particle, the voltage for the detection of β--particles must be raised accordingly. 3.1.1. Geiger-Müller Counter (working in the release area) Various designs of GeigerMüller counters exist. For the measurement of solid specimens, an end window counter is used (Fig.11). Its primary components are: The counting chamber: It is usually made of metal (e.g. Cu, Fe, Al, Pb, Mg) or glass with a metal coating. The wall of the counter functions as the cathode. The counting wire: It is the central electrode and functions as the anode. Diameter approx. Fig. 11: Experimental setup of a Geiger-Müller counter 0.05 mm, material: e. g. W, Mo, Fe The end window: This is generally made of mica (a layered silicate), nylon or similar materials with a surface weight between 1.5 and 3 mg/cm2. The fill gas: This is generally Ar or He. By adding methane, the reinforcing factor of the gas by secondary electron formation near the counting wire becomes less dependent on the voltage of the counter. In extreme cases, a type of constant discharge develops, which can be avoided by quenching. With self-quenching counters, alcohols or halogens are used as quench gases. The positively charged gas ions (G+), which are formed by the ionizing radiation, transfer their charge to the quench gas molecules (L): G+ + L → G + L+ The positively charged quench molecules Fig. 12: Ionization effects in a Geiger-Müller counter (L+) dissociate on their way to the negatively charged wall of the counter. At this point, the fragments are no longer able to produce secondary electrons and the discharge process is interrupted. By using a sufficiently thin end window, α-radiation can still be measured, but with a small, poorly reproducible counting yield. The same is true for weak β−radiation. Geiger-Müller counters are most suitable for measuring high-energy β−-radiation (Emax > 1 MeV). γ-Rays are registered with a low counting yield since only approx. 1 % of the γ-quanta which pass through the detector trigger an ionization. The registered signals are due 7 to photo and Compton processes of the γ-quanta at the atoms of the counting chamber. (See Chapter 5 “γ-Absorption”) 3.1.2. Proportional Flow Counter Fig. 13: Typical setup of a proportional counter (methane flow counter) The counter operates in the proportional area with methane or argon/methane employed as the counting gas. The advantage of this detector is that the measurement can be carried out directly without an end window. The sample is brought into the measurement chamber with the help of a rotating disk, such that the radiating particles are brought into direct contact with the counting gas. The voltage used with the proportional counter depends on the type and pressure of the counting gas. This voltage is between Fig. 14: Ionization processes in a methane flow counter 2000 and 4000 V. Proportional flow counters are especially suited for measuring α- and β−-emitters. 8 3.2. Solid Scintillation Detectors Primary components are the single crystal, the photo cathode and a secondary electron multiplier (PSEV). Fig. 15: Experimental setup of a solid scintillation counter Solid scintillation detectors are used mostly for the measurement of γ-radiation. In order to absorb the largest possible amount of the γ-radiation (to make interactions between the radiation and the crystal possible), thick crystals with large densities are used; preferably, sodium iodide crystals doped with thallium (ρ=3.67 g/cm3). During the interaction of the γ-radiation with the thallium ions, electrons are produced, which release scintillations. This results in the emission of bursts of light in the range of 400800 nm. These bursts of light hit the photocathode, from which electrons are consequently emitted. The number of electrons is multiplied over the dynodes in the secondary electron multiplier (from one electron, approximately 109 electrons are produced giving rise to an electric signal). The number of light bursts is dependent on the energy of the γ-rays, and each γ-quant is measured individually. The more light bursts are produced, the higher is the resulting signal. This is because the scintillations are produced so rapidly consecutively that they lead to one impulse. The dead time of such a measurement instrument is very short. By using organic crystals like anthracene, trans-stilbene or p-terphenyl, β−-radiation can be detected. For the detection of β−-radiation, organic scintillators have an advantage due to the low atomic number and, therefore, scarce occurrence of problematic bremsstrahlung (see Chapter 7 “Backscattering of β−-radiation”). The appropriate thickness of the scintillators depends on the measurement task. Often, it is necessary to select a thickness which is similar to the range of the hardest β−-radiation to be detected. Thin organic scintillators have only a small response to γ-quanta and are, thus, suitable for measuring β−-radiation in the presence of otherwise interfering γ-radiation. For the detection of α-radiation, only scintillators without protection layer or with very thin one can be used. For this purpose, ZnS(Ag)-scintillators are often used. For the detection of mixed radiation fields, scintillation detectors coated with ZnS(Ag) and a organic counting material are nowadays commonly used. With such counting tubes, both α- and β−-radiations can be detected with a high sensitivity. 9 4. The relationship between the activity and the counting rate A detector of a given geometry only registers a portion of the radiation emitted from a sample. The net counting rate (RN) is proportional to the activity (A) of the sample and the average number of the emitted quanta per decay (ν): RN = f⋅A⋅ν (1) with f = fG ⋅ fA ⋅ fS ⋅ fR ⋅ fi (2) fG: Geometric factor Absorption factor / extinction coefficient f A: fS: fR: fi: Self absorption factor Backscattering factor Response probability The geometric factor fG takes into account the distance of the sample from the detector and the diameter of the counting chamber window. Assuming a point source of radiation, only a certain angle will be detected. In other words, a cone-shaped volume is measured and converted into a spherical volume, a f G = 0.5 1 − 2 2 (a + r ) (3) where a is the distance of the sample from the detector and 2r is the diameter of the counting chamber window. The absorption factor fA, or rather the extinction coefficient, takes into account the attenuation of the radiation in the layer of air between the sample and the detector. The attenuation of the radiation by the counting chamber window is not accounted for by this factor. f A = e − µ ⋅d = e − µ ⋅d * µ: d: Where * (4) µ*: Mass extinction coefficient Mass per unit area of the absorber d*: d = ρ ⋅ d* and Linear extinction coefficient Thickness of the absorber µ* = µ ⋅ ρ (ρ = density of the absorber) The mass extinction coefficient µ is a function of the energy of the radiation and the atomic number of the absorber material and depends directly on the interaction of the different types of radiation with matter. 10 Because of the small distance between the detector and the sample, the absorption factor can be disregarded with γ- and strong β−-radiation (fA = 1). However, with weak β--radiation (14C!) and α-radiation, the absorption factor must be considered. In the case of α-radiation, the energy changes with the distance, while the particle number remains almost constant up to the maximum range. The self-absorption factor fS describes the absorption of the radiation within the sample. However, by assuming an infinitely thin layer of the sample, this factor can, in approximation, be set to 1. The factor fS is strongly dependent on the preparation of the sample and must be determined empirically (see Chapter 8 “Self-Absorption of β−-radiation”). Since the used samples are very thin, the selfabsorption factor can be disregarded. The backscattering factor fR increases with increasing mass per unit area and increasing atomic number of the material supporting the sample. By using the same material of adequate thickness under the samples, this factor can be assumed to be constant. From equations (1) and (2) we obtain: RN = fG ⋅ fA⋅ fS⋅ fR ⋅ fi ⋅ A⋅ ν (5) If the geometric factor, the absorption factor, the self absorption factor and the backscattering factor are all held constant during the measurement, the net counting rate, given the activity of the emitter, is directly proportional to the response probability fi of the detector. However, this is only the case with one sample measured with different detectors of the same type. For this reason, when comparing different detectors and different nuclides, fG, fA, fS and fR must all be considered in the calculation. If it can be assumed that the backscattering factor fR is identical for all experimental set ups, the product fR.fi, known as the relative response probability frel, can be used to compare detectors. From equation (5), we get for frel: f rel = f R ⋅ f i ⋅ 100 = 11 R N ⋅ 100 % f G ⋅ f A ⋅ f S ⋅ A ⋅ν (6) Experimental 1. Safety regulations There is no danger due to external irradiation resulting from the samples, since they are all of low activity. Never touch the samples with your finger! Always use tweezers! 2. Nuclides to be used 14C Pure β−-emitter with a maximum energy of 0.156 MeV. One particle is emitted per decay (νβ- = 1). Half-life: 5730 years (t1/2 = 5730 a) On 17 Feb., 1986, the sample had an activity (A) of 51,800 Bq. 204Tl β−-emitter (97.9 %) with a maximum energy of 0.765 MeV, which also emits conversion electrons (EC, 2.1 %). One particle is emitted per decay (νβ- = 1). Half-life: 3.78 years (t1/2 = 3.78 a) On 24 Nov., 2021, the sample had an activity (A) of 3 kBq. 60Co γ-emitter which emits two γ-quanta per decay (νγ= 2) with energies of 1.173 MeV (100 %) and 1.332 MeV (100 %), respectively. The γ-radiation is coupled to a preceding β−-decay (0.313 MeV at 100 %; νβ− = 1; (see Chapter 1 Fundamentals, 1.3 γ-Radiation), which is not detected in the experiment due to the use of an Alabsorber. Half-life: 5.26 years (t1/2 = 5.26 a) On 24 Nov., 2021, the sample had an activity (A) of 3 kBq. 241Am α-emitter (100 %) with an energy of 5.4 MeV. One particle is emitted per α−decay (να= 1) Half-life: 432.2 years (t1/2 = 432.2 a) On 17 Feb., 1986, the sample had an activity (A) of 3,700 Bq. Here, only types of radiation which occur at more than 1 % are specified. 1 Bq (Bequerel) = 1 decay/s 12 3. Procedure 1. Familiarize yourself with the operation of the different detectors (interactive experiment)! 2. Determine the reference rates of the different detectors (Geiger-Müller counters with and without cap, proportional flow counter, α-scintillation detector, β-scintillation detector, γ-scintillation detector)! Measurement time: 5 min each! 3. Determine the gross counting rates for the samples (14C, 204Tl, 60Co and 241Am) with each of the different detectors! In the case of Geiger-Müller counter without cap, αscintillation detector, β-scintillation detector, all measurements should be done for two distances between the detector and the sample: 2 cm and 3 cm. In the case of γscintillation detector and Geiger-Müller counter with cap, it is enough to do the measurement at 2 cm distance. While using the proportional flow counter it is not possible to change the distance, although you should measure the gross counting for the whole samples. Please do not forget to write down the unit for each gross counting rate. Measurement time: 1 min each Evaluation 1. Calculate the net counting rates for the individual measurements as well as the standard deviations (2σR = ± 2 ⋅ N0.5 ⋅ t-1) of the net counting rates 2. Calculate the present-day activity of each of the samples and enter your results in Table 4! A = A0⋅e-λ⋅t ; λ = ln 2 ⋅ t1/2-1 3. Calculate the relative proportional response probability frel for all of the detectors and all of the nuclides and enter your results in Table 5! The geometric factors and the absorption factor (for 14C only) are shown in Table 3. The average number of emitted quanta per decay can be found in the “summary of nuclides to be used.” Please note that 60Co can only be treated as a pure γ-emitter when an aluminum absorber is used. (=> ν = 2)! 4. Discuss the response potential of the detectors for the different types of radiation! 13 Detector a [cm] r [cm] fG GM counter 2 2.25 0.17 0.74 GM counter 3 2.25 0.10 0.41 α-scintillation detector 2 2.5 0.19 0.74 α-scintillation detector 3 2.5 0.12 0.41 β--scintillation detector 2 2.5 0.19 0.74 β--scintillation detector 3 2.5 0.12 0.41 2 1.75 0.12 0.74 2 2.0 0.14 0.74 - - 0.5 1 γ-scintillation detector 1·1 Inch γ-scintillation detector 2·2 Inch proportional flow counter Table 3: geometric factor fG; absorption factor fA for 14C only Isotpe Co 241 Am 204 Tl 14 C 60 Activity [Bq] Table 4: Calculated activity today 14 (for fA 14C only) Isotope GM counter 2 cm distance GM counter 3 cm distance GM counter with cap 2 cm distance α-scint. 2 cm distance α-scint. 3 cm distance β-scint. 2 cm distance β-scint. 3 cm distance γ-scint. 1·1 Inch 2 cm distance γ-scint. 2·2 Inch 2 cm distance proport. α Co 60 Am 241 Tl 204 C 14 Isotop proport. α-β Co 60 Am 241 Tl 204 C 14 Table 5: 15 Relative proportional response probability of the various detectors for different types of radiation Data Sheet: Detector Comparison R0: N: t reference rate of the measurement instrument number of measured signals time Detector R0 [Unit] Nuclide 60 GM counter 2 cm distance GM counter 3 cm distance GM counter with cap 2 cm distance α-scint. 2 cm distance α-scint. 3 cm distance β-scint. 2 cm distance β-scint. 3 cm distance 16 Co Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 241 N [Unit] RM = N ∙ t-1 RN = RM – R0 σR = N0.5⋅t-1 t [Unit] RM [Unit] gross counting rate net counting rate measurement rate error RN [Unit] 2 σR [Unit] Detector γ-scint. 1·1 Inch 2 cm distance γ-scint. 2·2 Inch 2 cm distance proportional flow counter α-Plateau proportional flow counter α+β-Plateau 17 R0 [Unit] Nuclide 60 Co Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 60 Co 241 Am 204 Tl 14 C 241 N [Unit] t [Unit] RM [Unit] RN [Unit] 2 σR [Unit] Chapter 2 Decontamination of radioactively contaminated surfaces Task 1. Investigate the possibilities of decontaminating different surfaces. Employ different decontaminants! Determine the percentage of activity remaining on the plates. 2. In a plastic tub, there are contaminations caused by 14C, 60Co and 90Sr. Localize the contaminants with the use of suitable detectors and identify the emitters! Determine the activity present on the surface! Fundamentals When handling unconfined radioactive materials, the danger of causing contamination by spraying radioactive solutions or by the production of radioactive gases, steam or particles of dust always exists. To prevent endangering the health of the individuals working with radioactive materials, regular checks of instruments, workbenches, floors, etc. for possible contamination are mandatory. When confronted with a contamination, the differentiation between two cases is required: ● Adherent contamination: cannot be wiped clean by normal means ● Non-adherent contamination: Possible to be wiped clean Non-adherent contaminations may cause danger due to outer irradiation as well as incorporation. Two basic methods are applicable to check for the presence of contamination: - Direct measurement in the inspection of surfaces Indirect measurement in the inspection of surfaces 1. Direct measurement in the inspection of surfaces To define the contaminated area, the surface is scanned with a suitable radiation detector of sufficient sensitivity. The type of radiation can be determined with absorption measurements. Especially suitable for the quantitative assessment of surface activity are the so-called largearea proportional counters with a maximal counting area of 300 cm2. Calibration factors for individual common nuclides are prepared and available by the manufacturers of such detectors. This is the only possibility to approximate the surface-related activity (as well to convert impulses to the unit Bq/cm2). Advantage: The detection of adherent contamination is possible. Disadvantage: γ-Radiation and strong β--radiation obstruct the detection of further contamination by α- and weak β--emitters, which may also be present. 1 2. Indirect measurement in the inspection of surfaces This procedure is mainly used when the expected contamination is so small that it could be overlooked or its direct measurement could be disturbed by strong β--radiation or γ-radiation when scanned with a radiation detector. With this method, a filter paper is wiped on the investigated surface spiralling from the outside proceeding inward to avoid a spread of contamination. Gloves must be worn when performing these so-called wipe tests. Afterwards, the filter paper is measured with a suitable detector (e.g. methane flow counter for α- and β--contaminations.) Advantage: No interference from nearby emitters is encountered. Small contaminations can be better detected. Disadvantage: Adherent contaminations cannot be ascertained. If contamination is detected by either the direct or indirect method, it is to be removed by proper means. Contamination levels are to be kept as low as possible and, unconditionally, below the limits defined by the German law (Strahlenschutzverordnung § 57). Surface decontamination can be performed with various methods. In each case, the used method depends on the nuclide causing the contamination, the surface and the chemical form of the contamination. When dealing with radioactivity, surfaces which are difficult to decontaminate should be avoided. Commonly used decontamination agents are e.g., water, soap solution, complex forming agents, diluted acids and bases or alcohols. Care must be taken that the employed decontamination agent does not attack the surface itself. Furthermore, the application of organic solvents is not advisable since they commonly diffuse together with the contamination into the surface. As a result, the decontamination becomes firmly fixed into the material and can no longer be removed. 3. Determination of surface-related activity To assess the degree of contamination, the activity related to the surface must be determined. The approach to determining this activity depends on whether direct or indirect methods are used in the measurement. 3.1. Determination with direct measurements In the quantification of surface contamination, a calibrated detector must be used. If the calibration factor of the used nuclides for the employed measuring device is not available, it must be determined with a calibration emitter of known surface-related activity as shown in equation 1. Kn = An (R M − R 0 ) Kn: Calibration factor of nuclide n (1) An: Surface-related activity RM: Gross counting rate of the substance R0: Background of the detector Attention: Such calibration factors depend on the distance between detector and sample. They are valid only at the distances employed in the calibration. 2 3.2. Determination with indirect measurements With the aid of wipe tests, surface-related activity can be assessed. First, a methane-flow detector must be calibrated for this measurement. For this purpose, three samples of known activity are gauged (14C, 60Co, and 90Sr). With the use of equation (2), the β-detection capacity εβ of the detector can be calculated for each nuclide. R − R0 εβ = M A (2) RM: Gross counting rate of the substance R0: Background rate of the detector A: Activity of the substance Between the counting rate RM produced by wipe tests of the samples and the surface-related activity AF exists the following relationship: AF = R M − R0 s ⋅ F ⋅εβ (3) s: Extraction factor F: Area of the inspected surface The surface F inspected for contamination should have an area of 100 - 300 cm2. With a wipe sample, the extraction factor gives the contained amount of activity. This extraction factor depends on the chemical composition of the nuclides responsible for the contamination and the structure of the surface on which the contamination is found. The nuclide can adhere tightly to the surface as a result of a chemical reaction, be bound due to adsorption or be incorporated into the pores of the surface leaving only a small amount that can be removed with the filter paper. Therefore, the extraction factor must be experimentally determined. For example, verification would be possible if the available contamination is assessed before and after the wipe test with a suitable detector. If no such verification of the extraction factor exists, it must be estimated. When in doubt, the factor of the corresponding DIN-Norm (DIN ISO 7503 Part 1) is used. In this case, the extraction factor amounts to 10%. 3 Experimental 1. Precautions There is no danger due to external irradiation resulting from the samples, since all of them are of low activity. All usual precautions are applied in order to avoid contaminations or incorporations. 2. Procedure 2.1. Direct measurement of the contamination of surfaces 1. In a plastic tub, four different plates were prepared; a metal plate, a plastic plate, a linoleum plate and a wooden plate. Each of the materials is contaminated with aqueous 99m Tc solution. With the help of a large-area proportional counter, ascertain the extent of the surface contamination! Which difficulties do arise during this procedure? 2. Examine the decontamination ability of the surfaces with water, ethanol and acetone! Determine the percentage of remaining radioactivity on the surface with equation (4)! R FD = n 2 ⋅ 100 Rn1 (4) FD = percentage of remaining radioactivity Rn1 = Impulse rate before decontamination Rn2 = Impulse rate after decontamination 2.2. Indirect measurement of the contamination of surfaces 1. With the aid of a Geiger-Müller counter, examine the nine fields in the plastic tub for contaminations and note your results. 2. Assign each of the fields in which you have detected a contamination with the responsible emitter (14C, 60Co, 90Sr). Various shielding materials are available at your disposal. 3. A wipe-test of each field was prepared and measured in a methane flow counter. Copy the count number and the measurement duration into the following table. Calculate the results of the wipe tests for the individual fields of the plastic tub along with their surface-related activity! The surface area of the wiped fields is 100 cm². The activities of the standard samples for the calculation of the β-detection capacity εβ are found in data sheet 2. 4 4. Discuss by means of the German law “Strahlenschutzverordnung“ (§ 57) required consequences. For which nuclide is a decontamination required? Which dose limit is useful? How to deal with contamination during daily lab work? 5 Data sheet 1: Decontamination of radioactively contaminated surfaces Direct Measurement Contamination with 99mTc Decontamination agent: Surface Contamination Wood Linoleum Plastic Metal Before Decontamination After 1st Decontamination Percentage remaining radioactivity after the 1st Decontamination After 2nd Decontamination Percentage remaining radioactivity after the 2nd Decontamination Indirect Measurement Contamination with 14C, 60Co and 90Sr Geiger-Müller counter I II III IV V VI VII VIII IX Data sheet 2: Decontamination of radioactively contaminated surfaces Detector: Methane flow counter (α + β-Plateau) 6 High voltage: 2100 V t [min] N RM [min-1] RN [min-1] 2 σR [min-1] Background β--detection capacity of the detector εβ 3700 Bq 14 C 56486 5 3700 Bq 60 Co 250745 5 3700 Bq 90 Sr 921925 5 Results of the wipe tests Field I Field II Field III Field IV Field V Field VI Field VII Field VIII Field IX 7 εβ AF [Bq⋅cm-2] Chapter 3 Calculation of the γ-radiation dose rate Task 1. Determine the dose rate of a γ-emitting sample (60Co) as a function of distance from the source by using the measuring device GR-135. 2. Determine the dose rate of a γ-emitting sample (60Co) as a function of distance from the source by using the measuring device Colibri VLD. 3. Determine the dose rate of a γ-emitting sample (60Co), upon the application of various shielding materials, at distances of 5 cm and 10 cm. 4. Calculate the dose rates corresponding to distances of 5 and 10 cm from a 60Co source! 5. Calculate the dose rate corresponding to a distance of 5 cm from a 60Co source in the presence of shielding with 8 mm thick lead and iron absorbers! 6. Calculate the effective body dose to which a person is exposed when working for a year at a distance of between 5 and 10 cm from a 60Co source! Fundamentals 1. Protection precautions when handling radioactive materials When radiation strikes matter, ionization processes occur. Such occurrences in living organisms can be seriously detrimental. In order to abate damages due to radiation in a given radiation field, principles have been defined by law. To minimize the radiation dose, the mnemonic aid DATS is applicable. 1. D as in Distance The Distance from the radiation source should be as large as possible (Fig. 1). 2. A as in Activity The Activities used in radiochemical experiments should be kept as low as possible. 3. T as in Time The experiment is to be planned in such a way that the Time spent working in the radiation field is minimized. 4. 1 S as in Shielding Always arrange suitable Shielding materials between the radiation source and persons (Fig. 2). Fig. 1: Effect of distance from a radiation source on the dose. Note, that an increase of the distance by a factor 2, decreases the dose by a factor 4! a.) without shielding b.) with aluminium plate c.) with lead plate Fig. 2: Effect of shielding the radiation source on the dose 2. Dose definitions For the description of the dose, which is transferred to a body by ionizing radiation, there are different dose concepts to be used. They are practically very different and each of these concepts has different uses. Table 1 gives an overview of some common dose magnitudes and their relevance. 2 Types Ion Dose I Background Describes the number of charges formed per mass of air. Energy Dose D Indicates how much energy is absorbed per mass. It is a J⋅kg-1 = Gy physical quantity. It does not distinguish between the types of radiation. Considers the relative biological efficacy of the different J⋅kg-1 = Sv types of radiation. H = D · wR Equivalent Dose H Effective Dose Unit C⋅kg-1 for wR: α = 20; β = 1; γ = 1; n = 5 - 20 Is the sum of the tissue-weighted equivalent doses of the J⋅kg-1 = Sv human body. In addition to the biological efficacy of the different types of radiation, it also takes into account the different radiation sensitivities of the organs. It is used as a dose value fixed by the legislator as a maximum to which a person may be exposed (Strahlenschutzverordnung). E = ƩwT·HT Table 1: Overview of relevant dose rates 3. Calculation of the energy dose rate for γ-radiation The equation used to calculate the energy dose rate (absorbed dose rate) for γ-radiation is as follows: A: Activity of the substance [Bq] = A ⋅ I ⋅ r −2 D γ (1) Iγ: γ-Radiation constant [Gy⋅m2⋅h-1⋅Bq-1] r: Distance between the substance and measuring device [m] D : Energy dose rate [Gy⋅h-1] For equation (1) to be valid, the radiation source must be defined as a point source and the radiation source and the measuring device are only separated by air (no further shielding). The interaction of the γ-quanta with air particles can be neglected. The γ-radiation constant Iγ is a function of the energy of the different γ-quanta of the considered nuclide: Iγ = f (ΣEγ) A shield is located between the radiation source and the measuring device, where it exponentially decreases γ-radiation as a function of the thickness and composition of the absorber. The formula used to calculate the dose rate of γ-radiation with shielding is derived from equation (1) with the multiplication of the attenuation term e-μd: = A ⋅ I ⋅ r −2 ⋅ e −µd D γ 3 (2) d : Mass per unit area of the absorber [g⋅cm2 ] µ : Mass extinction coefficient [cm2⋅g-1] d* : Thickness of the absorber [cm] where d = d*⋅ρ ρ : Density of the absorber material [g⋅cm-3] µ* : Linear extinction coefficient [cm-1] where µ = µ*⋅ρ-1 If the radiation source is partially enclosed, a portion of the radiation is reflected by the surrounding material (Fig. 3). This leads to an increase in the measured γ-radiation. In the calculation of the γ-dose rate, the increased portion of γ-radiation is taken into account by the so-called build-up factor B. This factor can also be applied to indicate the degree of γ-quanta dispersion. The build-up factor is both a function of the energy of the γ-quanta as well as the reflecting material. This value must be Fig. 3: Reflection of γ-radiation specifically determined for each experimental assembly with the use of calibrating substances. The formula used for the calculation of the energy dose rate for γ-emitters with shielding and build-up factor considerations is as follows: = A ⋅ I ⋅ r −2 ⋅ e −µd ⋅ B D γ (3) 4. The γ-dose rate measuring devices 4.1 GR-135 (SIAC Exploranium) The dose rate measuring device GR-135 can be used for photons in the energy range of 50 keV to 3 MeV. This device is capable of measuring dose rates in the range of 0.01 μSv to 50 μSv. A NaI-scintillator serves as the radiation detector. Every radiation measuring device undergoes a warm-up phase after switching it on. A temperature difference of 60 ˚C leads to a variation in value of about +/- 15%. Hence, it is important that the measuring device is calibrated prior to actual use. This so-called stabilization of the measuring device is performed by the GR-135 with a 137Cs calibration emitter. During the stabilization, an internal energy calibration of the device is conducted. The gain of the measuring device is checked and adjusted if necessary. If the temperature fluctuates during the measurement, the device corrects the fluctuations automatically. 4.2 Colibri VLD (Canberra) The dose rate measuring device Colibri VLD can be used for photons in the energy range of 59 keV to 1.5 MeV. Compared to the measuring device GR-135, dose rates in the range of 0.01 4 μSv to 1 mSv can be detected. The radiation detector is a CsI (Tl) scintillation crystal. This measuring device does not have to be calibrated, it can immediately be used. 5 1. Precautions No direct danger of contamination exists. Substances are placed into the experimental set-up by assistants. 2. Nuclides to be used For calibration, a 137Cs sample is used (Eγ = 662 keV) A(01.01.2005) = 9.25 kBq, t1/2 = 30.17 a For the measurement, a 60Co sample is used. A(30.12.2006) = 1.5 ⋅ 105 Bq, t1/2 = 5.272 a, Iγ = 3.41⋅10-13 Gy⋅m2⋅h-1⋅Bq-1 (Weighing factor: wγ = 1; Effective dose [Sv] = Energy dose [Gy]) 3. Procedure 1. The two measuring devices are used in the rate-meter mode. The measured values, related to the applied time constant, are continuously adapted to the measured result. To read the measured values properly, please wait until the values are no longer significantly changing. 2. Turn on the measuring device GR-135 with the joystick! 3. Select the menu item “Stabilize” and place the 137Cs source in the calibration opening. Start the stabilization! Remove the 137Cs source after successful calibration! Return to the main menu of the control panel! 4. Select the menu item “Search”! The device will now measure the corresponding dose rate. 5. Measure the dose rate at the distances from the radiation source starting at 5 cm in 5 cm intervals up to r = 100 cm! Make sure that the detector is directed towards the radiation source! 6. Measure the dose rate at a distances of 5 cm and 10 cm from the source with each of the 8 mm thick absorbers of different materials (aluminum, lead, iron, plexiglass, PVC) placed between the radiation source and the detection device! 7. Turn on the measuring device Colibri VLD. 8. Repeat steps 5 and 6 with this measuring device. Evaluation 1.1 Calculate the current activity of the 60Co-containing sample to compare it later on with the experimentally determined activities. A = A0 · e-λ·t ; λ = ln 2 · t1/2-1 1.2 Plot the dose rates measured as a function of 1/r2! The slope m = ∆y/∆x of the resulting line allows the determination of the γ-radiation constant when the activity is known or, inversely, the activity when the γ-radiation constant is known as follows: m = Iγ ⋅ A Iγ = 3.41⋅10-3 µGy⋅cm2⋅h-1⋅Bq-1 Discuss, when a linear course of the order is to be expected! 1.3 Compare the calculated activity with the values obtained from the measurement! 6 2. Calculate the dose rate of the 60Co-containing sample at distances of 5 and 10 cm from the source without shielding! Use the activity, calculated in 1.1, for this task. 3. Calculate the dose rate of the 60Co-containing sample at distances of 5 cm and 10 cm from the source with the use of an 8 mm thick lead absorber as well as an 8 mm thick iron absorber! µPb =5.996⋅10-2 cm2⋅g-1 µFe =5.196⋅10-2 cm2⋅g-1 ρPb =11.34 g⋅cm-3 ρFe =7.86 g⋅cm-3 4. Discuss the effectiveness of the absorber materials in regard to their reduction of the dose rate! Discuss the influence of the shielding and the distance on the dose rate. What is the effective protection against radiation damages? 5. In accordance with StrSchV (radiation protection regulation), a person of the category A may absorb a maximum effective dose of 20 mSv per year. Assuming that the effective body dose is equivalent to the local dose, how much of a dose is absorbed by a person who works at a distance of 5 cm as well as a person who works at 10 cm from the source per year? (The work time consists of 40 hours / week, 50 weeks / year). Is the legal annual dose limit exceeded? Is a person allowed to work under such conditions? Task 1.1 1.2 Magnitude Result Activity theoretically calculated Activity determined by meausurement 5 cm 2 3 5 7 Unit D D with 8 mm lead absorber D with 8 mm iron absorber Dose 10 cm Data Sheet: Dose rate measurement Equivalent dose rate [µSv⋅h-1] Distance [cm] 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 8 Dose rate [µSv⋅h-1] GR-135 Dose rate [µSv⋅h-1] Colibri VLD Distance [cm] Absorber material (8 mm) Aluminium Lead 5 Iron Plexiglass PVC Aluminium Lead 10 Iron Plexiglass PVC 9 Dose rate [µSv⋅h-1] GR-135 Dose rate [µSv⋅h-1] Colibri VLD Chapter 4 Gammaspectrometry; recording of γ-spectra Task 1. Perform an energy calibration of the NaI-scintillation detector by employing isotopes 60 Co and 137Cs! 2. Determine the γ-energies of 22Na! Discuss the γ-spectrum and identify the nuclide with the help of nuclide libraries and the nuclide chart! 3. Determine the γ-energies of the unknown radioisotopes A and B! Discuss the γ-spectra and identify the nuclides with the help of the nuclide libraries and the nuclide chart! Fundamentals γ-Spectrometry allows qualitative and quantitative analysis of unknown nuclides. Scintillation sensors and semi-conductor detectors are employed as detecting devices. 1. γ-Radiation γ-Radiation is emitted from the nucleus of an atom upon its transition from an excited state to a state of lower energy. The energy of γ-radiation lies between 10 keV and 104 MeV, which corresponds to an approximate wavelength range of 10-1 to 10-7 nm. There are three types of interactions between photons and matter: photo effect, Compton effect and pair formation. 2. Interactions of γ-quanta with matter 2.1. The photo effect Fig. 1: Photo effect through the interaction of γ-quanta with matter 1 The photoelectric effect describes the interaction of a low energy photon with an inner-orbital electron of an atom resulting in the complete transfer of the photon’s energy to the electron (Fig. 1). This results in the ejection of the photoelectron, which at this point possesses an energy equal to the energy difference between the initial γ-ray and the binding energy of the electron to the atom. The photo effect occurs predominately at the inner electron shell of the atom, where the binding energy of such electrons is relatively large. Following the dissociation of the photoelectron from the electron shell, a vacancy is available for a higher-energy electron of an outer shell to occupy. The difference in binding energy between the higher and lower energy orbitals causes the emission of X-rays. This resulting radiation is known as characteristic self-radiation or X-ray fluorescence. Due to the strong influence of the atomic number Z on photon absorption (pPhoto ≈ Z5), materials with large Z (e.g. lead) are commonly used in the fabrication of radiation shields. 2.2. The Compton effect The Compton effect occurs when, a photon of higher energy relative to the photo-effect strikes an outer-orbital electron, conveying a portion of its energy. As a result of the encounter, the flight-path of the γ-quant is altered (Compton scattering) and the electron is released from the atom (Fig. 2). The resulting energy of the Compton electron is close to the energy transferred from the photon to this electron, since the binding energy of the electron is small compared to the γ-energy. This so-called Compton process occurs primarily at the outer-shell electrons of the atom. As a result of the collision, the wavelength λ2 is greater than λ1. Depending on the angle of scattering, quanta of different energies are obtained, which form the so-called Compton-continuum of a γ-spectrum. Fig. 2: Compton-effect through the interaction of γ-quanta with matter 2 2.3. Pair Formation In the strong electric field near the atomic nucleus (Coulomb field), γ-quanta with an energy of 1.02 MeV or greater can be converted into an electron and a positron. In this process of “materialization”, the electron-positron pair develops (Fig. 3). The sum of the resting masses of the positron and the electron corresponds to an energy of 1.02 MeV. Therefore, the mass of each of these particles is equivalent to an energy of 0.51 MeV. If the energy of the photon is greater than 1.02 MeV, the excess energy appears in the form of kinetic energy of the positron and negatron (electron). Thereby, the energy surplus is distributed equally to both particles according to the law of momentum conservation. The generated positron has a short lifetime before it combines with an electron in a process reciprocal of materialization, annihilation. As a result of annihilation, the mass of the elementary particles are converted into radiation energy as two photons are generated with an exact energy of 0.51 MeV each. These photons, produced from pair production and the subsequent annihilation can proceed to interact with matter via Compton or photoelectric processes. The probability for the incidence of pair formation at the electrons of an atomic orbital is proportional to the atomic number Z. The probability of pair formation at an atomic nucleus (coulombic field) is proportional to Z2. Fig. 3: Formation of an electron-positron pair through the interaction of γ-quanta with matter 3 3. Design of a NaI-scintillation detector The fundamental components of a NaI-scintillation detector are the scintillation crystal, the photo cathode and the photomultiplier (secondary electron multiplier). Further details are described in Chapter 1 “Comparison of various detectors and nuclides”. 4. Single channel analyzer Analysis of the counts is carried out via frequency discriminators (upper and lower data thresholds are to be set) and an anti-coincidence level. Only such impulses in which the voltage peak exceeds the lower threshold value and is less than the upper threshold value, are forwarded to the counter. The strength of the impulse is proportional to a defined energy. Therefore, this measurement set-up permits the recording of the following spectra. Integral spectrum Differential spectrum Fig. 4: Impulse rate as a function of energy A single-channel analyzer enables the breakdown of the entire energy range into 1000 individual channels. Furthermore, the width of an energy range can be defined by the degree of amplification. For example, a range from 0 to 1000 eV can be measured with 1000 channels (1 channel corresponding to 1 eV) (Fig. 5a), or a range from 0 to 3000 eV can be measured with 1000 channels (1 channel corresponding to 3 eV) (Fig. 5b). Fig. 5: Impulse rate as a function of energy 4 Moreover, it is also possible to scan the content of a certain number of channels simultaneously. At this point, one talks about the “window” which has been set. For this, a lower energy level LL and an upper energy level UL are defined. Therefore, the measured data between two channels are gathered. 5. Multi-channel analyzer The recording of energy-resolved gamma spectra is today carried out practically and exclusively with multi-channel analyzers. In order to simplify, a multi-channel analyzer can be visualized as a large number of single channel analyzers connected in parallel. Whereas only the impulses of certain energy intervals are counted with the single channel spectrometer, the multi-channel spectrometer analyzes and stores almost every occuring impulse in the channel of the corresponding energy. The measurements of these memory channels can be accessed in succession and the content may be printed out or displayed on a screen. An advantage, which multi-channel spectrometers have over single channels spectrometers is that they immediately provide the entire spectrum distributed to a certain number of channels. Calibration with γ-radiation of a known energy allows a comparatively exact reading of the energy values (error approx. 1%). Fig. 6: Scheme of a γ-radiation measuring station Fig. 6 shows a scheme of a γ-radiation measuring station. It is mainly consisting of three components: the detector, the electronic measurement system and the evaluation unit. As detectors, semiconductor and NaI detectors can be used. The electronic consists of high-voltage supply, preamplifier, pulse height amplifier, analogue-digital-converter (ADC) and multichannel analyzer (MCA). The charge pulses, generated in the detector by the photons, are first collected in the preamplifier and then multiplied. The signals are then passed to the pulse height amplifier. There, the detector pulses are amplified and converted. Thereby, the pulse height is proportional to the photon energy. The following multi-channel analyzer (MCA) is the centerpiece of the system. It is composed basically of an analogue-digital-converter (ADC), multichannel memory, control logic and display. It detects the incoming pulses, then measures the voltage value and translates it into digital information. The contents of the MCA memory are then copied into the memory of the PC, by using the software, and displayed as a spectrum on the screen. Modern gamma-spectrometry measuring systems are controlled entirely by the software of the computer. Here the necessary settings such as high voltage, amplification and the evaluation can be carried out. For the nuclide identification various nuclide libraries are 5 available, which are helpful in the identification of the radiators. The activity can be determined by a known activity of a calibration radiator. 6. Set-up of γ-spectra The monoenergetic gamma radiation of a radionuclide leads by the interaction in the detector (photo effect, Compton effect and pair formation) to a complex pulse height spectrum. A characteristic feature of γ-spectra are “photopeaks”. Only with the photoelectric effect the total energy of the γ-quanta is released in a single process. The photopeaks enable the identification of individual nuclides with the use of appropriate tables or computer programs according to the “fingerprint” method. Problematic for analysis, however, are coincidental dispersions in the background and the overlapping of peaks. For each photon emitted during the decay, a photopeak is to be observed in the spectrum. Thereby, two photopeaks are expected in the γ-spectrum of 60Co according to the decay scheme (Fig. 7). When evaluating the spectra, it is important to ensure that all peaks, which are observed in the spectrum, have to be assigned. Otherwise, errors may occur in the interpretation of the results. Fig. 7: The decay scheme of 60 Co The “Compton continuum”, which is observed at lower energies, is based on the Compton effect and is not suitable for the identification of individual nuclides. Its position depends on the maximum energy of the penetrating γ-quanta. In the case of overlapping spectra, the contents of the channels in the range of the Compton continuum are simultaneously summed. The lower the energy of incoming quanta, the higher are the effects of the background radiation. Therefore, an evaluation of the low energy range of the spectrum is difficult or not possible. 6 7. Comparison between semiconductor and NaI detectors Fig. 8: γ-Spectrum of 60Co recorded with NaI and Ge(Li) detectors Fig. 8 clearly shows that considerably smaller linewidths can be obtained with a semiconductor detector. Improved resolution of the sprectra is especially important when the energies of γquanta lie close to each other. In this case, semiconductor detectors are especially well suited for qualitative analysis. In general, a lower number of counts are yielded with a semiconductor detector than with a NaI-scintillation detector. The high response probability of a NaI-scintillation counter is advantagous in quantitative analysis. NaI-scintillation detectors have a better response capacity to radiation of lower energy than higher energy radiation. However, one must keep in mind that in the range of very low energies the linearity pertaining to the energy of the penetrating radiation is not always guaranteed. 8. Energy calibration The width of the pulse height spectrum, and thus also the resolution and the position of the individual energies, depend on different settings in the detector. The high voltage, with which the detector is operated, as well as the setting of the amplification both cause changes of the spectrum. For this reason the assignment of the channels of the corresponding energies have to be first done for every detector, this can be achieved by using gamma emitters of known energy. If the energy calibration is carried out, this is only valid as long as no changes are made to the settings of the high voltage and the coarse and fine gain. 7 Experimental 1. Precautions Because substances of low activities are used, no external danger due to radiation fields are present. The substances are not to be touched with the hands. Tweezers must be used. 2. Procedure Energy calibration of NaI-scintillation detectors 1. 2. 3. 4. 5. 6. 7. 8. 9. Start the computer and open the "Prospect" measurement program, which you will find on the desktop. Select "Technican" as the user. The password is "welcome". Click "Connect to device" and select the detector "Gammaspek1b.10.0.1.4". Click "take over". In the "Detector" submenu you can now select the high voltage of the detector. Set on the value to 700 V. As soon as you click "on", the high voltage is switched on. The gain of the detector can be set in the "MCA" submenu. Set the number of channels to be measured on 1024. With "coarse gain" and "fine gain" you change the amplification. Select x1 for this. In the "Acquisition" submenu you can now set the mode and the time of the measurement. Select PHA (= Pulse Height Analysis) as the mode. Select "live time" and 300 s as the measuring time. Place the calibration radiators 137Cs and 60Co on the detector! Click "Start" to start the measurement. You should now see a spectrum with three signals. Test the influence of the high voltage and the gain of the detector on the position and resolution of the γ-lines. Repeat the measurement with the settings in Table 1. From your results, select the settings, where the photopeaks of 60Co are in the second half of the spectrum. Discuss the settings with the assistant. These settings remain unchanged until the end of your experiment. Spectrum 1 Spectrum 2 Spectrum 3 Spectrum 4 Table 1: Settings 8 High voltage Coarse gain Fine gain High voltage Coarse gain Fine gain High voltage Coarse gain Fine gain High voltage Coarse gain Fine gain 700V x1 x1 700V x8 x5 600V x1 x1 600V x8 x5 10. Measure an impulse height spectrum of the 60Co and 137Cs radiators with the selected settings. 11. Now perform an energy calibration for the selected settings. Select the "calibration" submenu. You can now select the first peak height point with the left mouse button. This should now be shown in green. Select the peak with the right mouse button "use this ROI". Specify the appropriate power 137Cs (γ) - 662 keV. Proceed in the same way for the signals for 60Co (γ1) - 1173 keV and 60Co (γ2) - 1332 keV! 12. Select "close calibration view and return to spectrum view" in the upper right corner. The detector is now calibrated. Determination of the energies of 22Na 1. The calibration emitters are exchanged by 22Na. 2. Click "Clear". 3. Click "Start" to start the new measurement. 4. Now determine the γ-energies of 22Na. 5. Compare your result with the nuclide chart and the interactive nuclide libraries. Click on "Analysis" and select a suitable nuclide library! Determination of the energies of the unknown radioisotope A 1. 22Na is exchanged by the γ-radiator A. 2. Click "Clear". 3. Click "Start" to start the new measurement. 4. Now determine the γ-energies of (A). 5. Compare your result with the interactive nuclide libraries for the identification of γ-radiator A. Determination of the energies of unknown radioisotope B 1. The γ-radiator A is exchanged by the γ-radiator B. 2. Determine the nature of γ-radiator B. Evaluation 1. 2. 3. 4. 5. 6. 7. 9 Discuss why you need to calibrate the detector. Calibrate the detector by taking an impulse height spectrum of 137Cs and 60Co. To do this, test the settings for the high voltage and the gain of the detector and select the most suitable one. Determine the γ-energies of 22Na. Compare your results with the interactive nuclide libraries. Which one is most suitable? What is the influence of the selection of the nuclide library on the evaluation of the γspectrum? Compare your result for 22Na with the nuclide card. Look for the γ-energies for 137Cs and 60 Co. What did you notice? Discuss your observations with your assistant. Determine the γ-energies of the unknown emitter A. Which radiator could it be? Analyze your results critically. What are the difficulties in identifying the radiator? Which nuclide library do you choose? Determine the γ-energies of the unknown emitter B. Analyze your results critically. What are the difficulties in identifying the radiator? Which nuclide library do you choose? Measurement sheet: Gamma spectrometry Coarse gain High voltage [V] Radiator 137Cs 60Co 22Na Fine gain Emitter A Emitter B Energy γ (keV) Energy γ (keV) Energy γ (keV) Energy γ (keV) Energy γ (keV) Energy γ (keV) Energy γ (keV) Energy γ (keV) Radiator 22Na Nuclide library Isotope 10 22 Na Probe A Probe B 1 Chapter 5 Betaspectrometry (Liquid Scintillation Counter) Task 1. Determine the response probability of the liquid scintillation detector for the nuclides 3 H and 14C. 2. Record β–-spectra of the nuclides 3H and 14C and plot the counting rates of the standard substances as a function of the channel. 3. Calculate the contributions of 3H and 14C to the β-radiation of an unknown sample with the help of the net counting rate. Compare the calculated values to an automatic determination. Fundamentals 1. β--Spectrum The energy of the fission of a neutron to a proton (n → p+ + e– + ν ) is divided between the electron and the antineutrino following the linear momentum and energy conservation law. Therefore, when a great number of nuclei undergo β–-fission a continuous energy spectrum of β– particles is obtained. Fig. 1: Schematic distribution of energy of β– particles after the 14C resp. 3H decay The maximum energy of a β– particle is constant for the decay of a specific nuclide. The average energy is about a third of the maximum energy. Often a logarithmic scale of the energy is used when plotting the β– spectra. 1 2 2. Fundamentals of liquid scintillation spectrometry The exact measurement of the β–-radiations of 3H and 14C is of particular interest for the solution of biological and biochemical problems. Both nuclides emit low-energy β– particles, which require special measuring techniques, in particular when the obtained counting rates are low. The most suitable method for this is liquid scintillation counting. The principle of a liquid scintillation counter is based on the interaction of the radiation of a dissolved β–-emitting radioactive nuclide with a scintillation substance in the same liquid phase. For this, a so-called “scintillation cocktail” is used. 2.1. The scintillation cocktail The cocktail is composed of a solvent, a primary scintillator and a secondary scintillator. 2.1.1. Solvent Common solvents, such as water, acetone or alcohols, are unsuitable as solvents for the radioactive substance, since the radioactivity can cause chemical processes with the solvent molecules (radiolytic processes). Therefore, more inert solvents like benzene, toluene, xylene or methoxybenzene are used. They bear characteristic π-electron systems which can readily be excited by β–-particles and show a high quantum yield. Detergents are formulated to provide a good solubitily of polar components in the unpolar solvents. 2.1.2. Primary scintillator Primary scintillators are fluorescent dyes (e.g. PPO: 2,5-Diphenyloxazole, Diphenylbutadiene). These complex aromatic molecules absorb radiation with short wave lengths and emit the energy in form of a radiation of a longer wavelength. They are used in a concentration range of 10–2 to 10–3 mol·l–1. 2.1.3. Secondary scintillator (“Frequency changer”): The secondary scintillators shift the wave lengths of the emitted light (between 340 nm and 480 nm) to slightly longer wave lengths to meet the range of the highest sensitivity of the photocathode (between 400 nm and 500 nm). Examples for such substances are POPOP (1,4-Bis(5phenyloxazol-2-yl)-benzene) or Dimethyl-POPOP. The wavelengths obtained are between 400 and 550 nm. PPO Dimethyl-POPOP Bis-MSB 2.2. The mechanism The mechanisms of the energy absorption and the light developing process in a liquid scintillator system are extremely complex. The following description is strongly simplified and summarizes the steps from the interaction of a charged particle with the solvent in the cocktail up to the light developing process: 2 3 Fig. 2: Interaction of the β--particles with the aromatic solvent and the emission of fluorescent radiation β--particle e– o Solvent molecule in the initial state • Solvent molecule in the excited state 2.2.1. Collision of a β--particle with a solvent molecule The β--particle transfers a part of its energy to the π-electron system of the solvent molecule (L), which changes into an excited state (L*). The β--particle collides with many different solvent molecules, because only a few eV per collision are transferred; the energy of the β--particle is generally several hundred keV. 105 to 106 collisions are expected before the β--particle has released all of its energy and is captured by a molecule. L + β– → L* + β– 2.2.2. Energy transfer between the solvent molecules / phosphorescence The excited solvent molecules transfer their energy to neighboring solvent molecules. L* + L’ → L + L’* However, it is also possible that its excitation energy is emitted as light (e.g. in the range of 260-340 nm; phosphorescence). L* → L + h·ν (260 - 340 nm) Direct measurement of phosphorescence is difficult and not very accurate. It’s better to try to change the wavelength by chemical means. 2.2.3. Absorption of the phosphorescence radiation by the primary scintillator The radiation emitted by the solvent molecules is absorbed by a primary scintillator (P), which changes into an activated state (P*). 3 4 P + h·ν (260 - 340 nm) → P* In contrast to the solvent, the primary scintillator cannot transfer its energy to further molecules. It releases energy exclusively in the form of fluorescence (e.g. within an energy range of 340480 nm). P* → P + h·ν (340 – 480 nm) 2.2.4. Absorption of the fluorescence radiation by a secondary scintillator When the fluorescence spectrum of the primary scintillator does not correspond with the absorption spectrum of the photo cathode, a secondary scintillator (S) is added to the cocktail. This secondary scintillator works in the same way as the primary scintillator. S + h ⋅ν 1 → S * S * → S + h ⋅ν 2 It absorbs radiation within a distinct range (e.g. 340 nm to 480 nm) and emits radiation with higher wavelengths compared to the primary scintillator (e.g. 400 nm and 550 nm). The quantity of light emitted by the scintillator is extremely small. Highly sensitive detectors are needed for the registration. Secondary electron amplifiers are used for these purposes. 2.3. Registration of a β--spectrum The amount of light emitted from the scintillator is very little and therefore, a detector with a very high sensitivity is needed. The photomultiplier transforms a light signal into an amplified amount of electrons that are easier to detect. The photocathode is coated with a light sensitive material that emits electrons when photons of the fluorescence from the scintillators are absorbed. These electrons are accelerated by an electric field (100-200 V) towards a positive charged dynode where each single electron is amplified to 3-5 electrons. Negativly charged focus rings ensure that every electron is amplified. A set of 12 dynodes further increase the amount of electrons by a factor of about 106. The electric impulse is then analyzed by a discriminator. Different settings of the discriminator enable to detect e. g. specific signal intensities (that are proportional to the energy of the β– particle). The discriminator only forwards very specific signal intensities and can be divided into 1000 different intervals (channels). An electric impuls counter is connected to the discriminator to analyze the counts. 2.4. Influence of the energy of the β- particle After a β– decay the β– particle collides with a to its energy proportional amount of solvent molecules. During each collision a similar amount of energy is transferred. The solvent molecules transfer the energy as described above. The overall process takes place in a time interval of about 10–6 seconds. The signal intensity that is measured at the photocathode within this time interval (<10–6 s) is proportional to the energy of the β– particle. The concentration of the sample must be chosen that less then one decay is happening during that time slot. In case that more than one decay appears within 10–6 seconds then the discriminator can not distinguish between the two dacays and counts them as one. 4 5 2.5. Influences on the counting efficiency The counting efficiency is influenced by different side effects. Contributions for the measured background rates beside the dark current are cosmic radiation and naturally occuring radioactivity from the measuring device (e.g. 40K-radiation of the glass walls). Note! Sometimes the radioactivity of the container material is higher than the radioactivity of the sample. For standard substances frequently counting yields of less than 100 % are observed. This is caused by: 2.5.1 Signal loss in the scintillation counter: This contribution is small and mainly appears as loss in the photo cathode. It is usually either caused when photons only have a low energy to not be able to cause the emission of an electron. Secondly, it can happen that the amount of photo electrons from the photocathode is so little, that it is less than the lower limit of the discriminator and, therefore, is not detected. This loss can easily be accounted for by measuring of calibration standards of a known activity and then be corrected. The response probability η (in %) is given by: η= R N ⋅100 A 2.5.2. Losses due to quench effects More important is quenching, which influences the fluorescent dyes and decreases the counting yields significantly. These quench effects shift the whole β– spectrum. Therefore, the maximum energy can be shifted to lower energies as well as the integral counting efficiency can be reduced. Figure 3 shows a non-quenched curce “C” and two quenched curves “B” and “A” where the amount of quenching increases from “B” to “A”. 5 6 Fig. 3: Non-quenched β– spectrum “C” and two quenched spectra “B” and “A” 2.5.3. Loss by primary or chemical quenching Chemical quenching is caused, for example, by strong acids or bases. They chemically destroy the organic molecules of the scintillator and, thus, the concentration of the scintillator substance decreases. Similar conditions must always be applied for preparation of different samples to being able to compare different results. 2.5.4. Loss by secondary or color quenching Parts of the radiation are absorbed during the scintillation process. Thus, they are not registered by the photo cathode. Any substance with absorptions between 200 nm and 500 nm (yellow color!!) causes color quenching. Therefore, the result of colored samples must always be treated with caution. 2.6. Corrections of deleted counts The counting yield and the counting rate of a sample strongly depend on the quantity and the concentration of quenching substances. Internal and external standardization are used to maintain comparable results for solutions containing various concentrations of compounds. Refer to the literature for more information! 2.6.1. Internal standardization First, the sample is analyzed in the scintillation counter. Second, an exactly known amount of the same radionuclide is added to the sample with a volume as low as possible (volume error!) and the sample is measured again. The added amount of radioactivity must be substantially larger than the amount of the sample (20x larger is usually suitable). After measurement of this “standardized” sample the net counts that arize from the added activity are easily calculated by the difference (before and after addition of the activity). As the added activity is exactly known the amount of quenching can be calculated. The exact activity of the sample then is calculated from the response probability and the quenching rate. Even though, this method is very exact it has some disadvantages: very small amounts of solvents have to be pipetted every sample must be processed and measured twice (time-consuming) - radioactive standard solutions are very expensive: There are additional radioactive wastes whose disposal is also expensive. 6 7 2.6.2. External standardization An external standard is a long-life γ-emitter that is placed in close proximity of the sample. The γ-radiation generates Compton electrons in the scintillation solution (compare to “gamma-spectrometry”) that are approximately quenched similarly to the electrons of the β–emmiter of interest. A calibration curve is prepared for the correction of the amount of quenching. Herefore, samples with different concentrations of the quenching substance in the scintillation cocktail are recorded and the logarithm of the counting rate is plotted towards to different concentrations. Usually slightly bent calibrations curves are obtained from which the amount of quenching can be extracted when the concentration of the quenching substance is known. Usually 137Cs or 133Ba are used as external standard that are supplied by the manufacturer of modern liquid scintillation counters. 2.6.3. Channel proportion method Herein, the change of the β– spectrum when being quenched is used (compare to figure 3). Two different channel intervals are set that should have similar counting rates in the not quenched state, which are ideally right and left of the maximum of the β– spectrum (most abundant energy). For an increasing quenching rate both, the maximum energy and the most abundant energy, move to lower energies. The quotient of both channel intervals is formed: 𝑅𝑅𝑐𝑐ℎ𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎 𝐵𝐵 = 𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞𝑞 𝑅𝑅𝑐𝑐ℎ𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎𝑎 𝐴𝐴 For similar impulse rates (non-quenched sample) this quotient is around one. Whereas for increased quenching the β– spectrum is shifted to lower energies, the impulse rate for channel A (lower energy) increases and the impulse rate for channel B (higher energy) decreases strongly. Therefore, the quotient decreases for quenched samples. A plot of the quotient towards the concentration of the quenching substance gives a similar calibration curve as achieved with the external standardization. This method has the disadvantage that for samples with strong quenching the counting rate of channel B (higher energy) is very low and often requires either a very long time to be measured exactly or a high activity is needed. The channel intervals A and B must be selected that this only occurs for higher quenching rates. 2.6.4. Channel proportion method with external standard This is the combination of the two methods as it has been described before. The channel proportional method is used for the determination of the quenching rate in the Compton spectrum. Compton electrons are emitted when a long-life γ-emitter is placed in close proximity to the scintillation cocktail. Different concentrations of quenching substances show the same shift as a β– spectrum. Therefore, the plot of the quotient towards the concentration of quenching substance results in a similar calibration curve as achieved by the external standardization or the channel proportion method. 2.7. Registration of two β--emitters in one cocktail An overlap of different β--spectra is observed when mixtures containing more than one β-emitter are studied. This causes considerable problems in quantitative measurements. For the determination of the individual contributions of 3H and 14C in an isotope mixture, energy windows are selected in such a way that the registered impulses are exclusively (or at least predominately) from one of the nuclides of interest. The experimental net counting rate is recorded for 7 8 each isotope standard (3H or 14C) of known activity in the defined windows. Herewith, the concentration of the corresponding nuclides can be calculated. Fig. 4: β–-spectra of 3H and 14C with two possible windows 8 9 Experimental 1. Precautions There is no danger due to radiation fields due to the small activities. The glass bottles with the cocktails should not be exposed to light in order to prevent undesired fluctuations of the measuring rates. 2. Nuclides to be used 3 H: C: 14 A(11.04.1974) = 2450 Bq A(11.04.1974) = 855 Bq t1/2 = 12.3 a t1/2 = 5730 a Emax = 18 keV Emax = 158 keV 3. Procedure Part I: Determination of the response probability of the liquid scintillation detector for the nuclides 3H and 14C. 1. Load rack No. 1 with the samples (background, 3H standard, 14C standard, unknown sample) in the positions mentioned in the table on data sheet 1. 2. Place the rack in the measuring device on the right hand side and close the device. 3. Wait 5 min and familiarize yourself with the use of the software at the computer during that time (see sheet with software explanation). The software must be set to ready-torecieve before starting the measurement. 4. Start the first measurement after the 5 min as soon as the software is ready to receive data. Click on “Count Single Rack” and “Select User Program” and choose the measureing program (see data sheet 1). Start with the program “4 SPEKTRUM”. 5. Confirm the program with “SELECT” and start the measurement with the “START” button. 6. Each sample is now measured for 1 min and the spectra saved on the computer. The overall counting rates are also saved to a different file. Measurements are performed in the “WIDE” mode, meaning that it is measured over the whole channel region (1-1000). 7. Calculate the response probability for both isotopes. Part II: Plotting of the β–-spectra in Excel and determination of the windows 1. Import the spectrum data of the 3H and 14C standards to one MS Excel sheet. Use the corresponding files “SNMMDDNN” as described in the section “At the computer”. Example: use the files SN011502 and SN011503 for the 2nd and 3rd measurement on January 15 (0115). 2. Normalize the spectra on the maximum of each spectrum. 3. Plot the curve as a line without points in a xy-type graph for the normalized 3H and 14C standards. 4. Determine intervals for the windows A and B for the measurement of the unknown sample. Choose window A as large as possible but take care that only 14C counts are measured in this window and none of 3H. Choose the interval for window B so that it contains as many 9 10 5. counts of 3H as possible with having as little of 14C as possible at the same time. The maximum of the 14C spectrum must be in window A and the maximum of 3H in window B. The windows may not overlap. Discuss the choosen windows with your lab assistant. Part III: 1. 2. 3. 4. 5. 6. For measureing in the range of the windows choose “Review and Edit User Program”, choose the first program “1 PRAKTIKUM” and confirm with “SELECT”. Change the window range settings at “Isotope 1” and “Isotope 2” according to your selection. To set the window range change the values for “lower level, upper level” (e.g. “0,500” for choosing the channel range between channel 0 and channel 500 for window A). Save your settings each time with “SELECT”. After changing the settings save the program by pressing “Previous Menu” until “User Program Is Being Stored …” is displayed for a short period of time. Click on “Main Menu” and start the measurement as described before by going on “Count Single Rack” and choosing the program “1 PRAKTIKUM”. The counting rates are now measured in the windows. “WIND1” is corresponding to window B and “WIND2” to window A. Perform part III of the Analysis. Part IV: 1. 2. 3. 4. 10 Measurement in the choosen windows Automatic measurement of the nuclides 3H and 14C Start a measurement for the automatic determination of the activities of 3H and 14C. Click on “Count Single Rack” and choose the user program “3 AUTOISOTOP”. You obtain automatically determined activities for both nuclides of all samples. Compare these values with the values you calculated yourself. Do they differ? If yes, what could be the reason? 11 At the computer: Software to receive data from the liquid scintillation counter 1. 2. 3. Start the software “LS Data Capture”. Start the capture mode under “Capture” and “Activate”. The program should now show “Ready to receive data …” as well as “WAITING FOR SESSION” or “WAITING FOR NEXT SESSION” (see screenshot). The program must now remain untouched until all measurements are finished. The data are automatically transferred after each measurement and saved as a BSF file (.BSF, including all details) and as a text file (.ASC, only counting rates). Please take care to not use any other software as an internet explorer because they often cause the failure of the data receiving. Figure 5. Receiving data at the computer. Screenshot during activated receiving mode. The program must run in this mode during all measurements. Data are transferred after each measurement automatically. 4. The files are saved in a folder on the desktop. All spectra are saved in the subfolder “USER00”. The file names are given according to the following scheme: 5. A collection of all overall counting rates of each rack are saved to the subfolder “SNGLRACK” after the whole rack is completely measured. Important: Please note the sample number “NN” for each sample that you measure (“NN” used for saving each spectrum in the subfolder USER00) as well as each rack number (different (!) “NN” used in the subfolder SNGLRACK) to be able to know which file corresponds to which measurement. 6. 11 12 Evaluation Part I: 1. 2. Calculate the activities of the standard cocktails of 3H and 14C for today. Calculate the percentage of the response probability η of the liquid scintillation counter for both nuclides 3H and 14C (η = RN ⋅ 100 ⋅ A–1). Part II: 3. 4. 6. 7. 8. Plotting of the β–-spectra in Excel Plot the normalized spectra of 3H and 14C in one Excel plot and choose suitable window settings. Discuss your choice with your assistant. Part III: 5. Response probabilities Determination of the molar account of 3H and 14C of an unkonwn sample Calculate the percentage of the net counting rate (subtracted background) for 3H and 14C in the windows A and B. Determine starting from the percentage of the net counting rate the theoretical counting rate and, herewith, the activity of 14C in the unknown sample. Calculate the overall counting rate of 14C and 3H in the unknown sample (see chapter 2.7). Calculate the molar amount of 3H and 14C in the unknown sample. Part IV: Validation of the results and comparison with the automatic measurement 9. Compare the manually calculated results with the automatic measurement of 3H and 14C. 10. Add the normalized spectrum of the unknown sample to your Excel plot of 3H and 14C. 11. Are there any differences? If yes, how could they possibly be explained? Can you think of a method to improve your results? 12 13 Data Sheet 1: Sample Overview 3 AUTOISOTOP 1 PRAKTIKUM 4 SPEKTRUM channel _____ – _____ window B channel _____ – _____ sample window LL – UL t N –1 [min ] [min] 1 background 1 0 – 1000 1 2 3 H standard 0 – 1000 1 3 14 C standard 0 – 1000 1 4 unkn. sample 0 – 1000 1 A 1 1 background 1 B 1 3 A 1 2 B 1 A 1 B 1 A 1 B 1 spectrum ID* window A position rack ID* program Channel interval for the windows: 2 σR [min–1] H standard 3 14 4 unkn. sample C standard 1 background 1 2 empty 3 3 4 5 RN [min–1] auto auto 1 H standard 3 auto 1 14 14 C auto 1 3 H auto 1 C standard unkn. sample H 14 C auto 1 *The first file of the day will be numbered 01 and then each file with increasing number. This is important for the identification of the correct file at the computer. Please note that your first measurement can have a different number than 01 if you are not the first group of the day. 13 14 Data Sheet 2: Tables for Analysis Part I: Response probabilities 3 14 H C calculated activity today response probability Part II: Window settings window A window B % of 3H signals RN in the window window interval % of 14C signals – – Calculation of the activities and the molar amount 100% counting rate [min–1] activity [min–1] molar amount [mol] 14 unknown sample unknown sample automatically 3 3 H 14 C H 14 C Chapter 6 Backscattering of β--radiation Task 1. Determine the energy of the β-emitter 204Tl by measuring the net counting rate RN as a function of the weight per unit area of an aluminum absorber by determining the maximum range of the radiation and by using the graph: “β-energy as a function of the maximal range [mg⋅cm-2]”! 2. Determine the "atomic numbers" of plexiglass and brass from the dependence of the net counting rate RN on the atomic number of different backscattering materials for 204Tl! Fundamentals Interactions of β-particles with matter are mainly dominated by two types of processes: 1. Ionisation and activation processes 2. Scattering processes at atomic nuclei and orbital electrons 1. Ionisationen For the relationship of the loss of energy of β-radiation through scattering process to the loss of energy through ionization, the following equation roughly applies: ∆E bremsstrahlung ∆E ionization = E ⋅ Z / 800 (1) (E = energy in MeV; Z = atomic number) 2. Scattering processes 2.1. Scattering processes at electrons Inelastic scattering is observed at the orbital electrons. The probability that an electron is scattered with a distinct angle δ is proportional to the number of electrons per volume of the scattering material and, therefore, proportional to the atomic number Z. The scattering of β--particles at orbital electrons is very rare and of less importance compared with the scattering at atomic nuclei. 2.2. Scattering processes at atomic nuclei Coulomb scattering is observed at the atomic nucleus. It occurs most frequently with highenergy β-radiation. In the electrical field of the atomic nucleus the high-energy electrons lose energy and send out a continuous spectrum of X-rays (bremsstrahlung). The loss of energy is proportional to the atomic number of the atomic nucleus and to the energy of the β-radiation. The probability that an electron is scattered with a distinct angle δ is directly proportional to the thickness of the absorber surface, to the number of scattering centers, as well as directly 1 proportional to the square of the atomic number of the absorber substance, but indirectly proportional to the square of the energy of the β-radiation. This probability is described by the following formula: n ( δ) F ⋅ N ⋅ d e4 (Z − A F )2 = ⋅ ⋅ p ( δ) = n0 4 ⋅ r 2 (m e v e2 ) 2 sin 4 ( δ2 ) n(δ) n0 N F d F⋅d 4⋅r2 e meve2 Z AF sin4(δ/2) (2) number of electrons in a distance „r“. The angle „δ“ is defined as the angle between the directions of the arriving and the scattered electrons number of scattered electrons at the surface number of atomic nuclei per volume of the scattering substance (in practice corresponding to the density of the material). area of the scattering surface "thickness" of the scattering surface scattering volume; the product „F⋅N⋅d“ represents the total number of scattering centers distance and the geometry of the detector to the scattering surface elementary charge kinetic energy of the β-particle atomic number of the scattering material atomic form factor, describing the geometry of the atomic nucleus: when a βparticle hits a nucleus, the nucleus deviates from its spherical form describes the angle dependence of the scattering of the β-particles The values which determine the scattering angle δ of β-particles favor small angles. A change of the direction of a β-particle by about 180° due to unique coulomb scattering at an atomic nucleus is very rare. It is almost exclusively caused by multiple scatterings with smaller scattering angles δ, as is shown in figure 1. Fig. 1: (a) max. range of an electron in a substance; (b) backscattering of an electron due to repeated Rutherford scattering (Scattering at atomic nuclei) Backscattering is observed when a β-particle leaves the substance into which it has penetrated with an overall scattering angle of about 180°. The backscattering increases the measuring rate as a function of the material causing the scattering, that is, the material behind the sample. 2 The backscattering factor ϕR is defined as: ϕR = R ( R) Intensity of the backscattering = N Primary intensity of the radiation RN (3) R (NR ) = Counting rate with backscattering; RN = Counting rate without backscattering Fig. 2: Dependence of the backscattering factor ϕR on the area weight and the atomic number of the absorbers and on the energy of the β-radiation Figure 2 shows that the backscattering factor increases at first with increasing thickness of the scattering substance and reaches a maximum value ϕSR when a certain saturation thickness has been reached. It can be assumed that the β-particles, after they have entered the material, interact with the atomic nuclei on their way out of the material. The further the β-particles have penetrated the material, the smaller the chance that they leave it again. This saturation thickness (dS [mg⋅cm-2]) depends solely on the energy of the β-radiation. This thickness could be determined theoretically; practically it corresponds to 1/5 of the maximum range rmax of the β-particles, which penetrated into the lower layers of the material. 3 3. Application of backscattering / Absorption of radiation Attenuation of β-radiation: N = N0 · e–µ·d N = number of not-scattered electrons behind the scattering shield N0 = number of electrons before the scattering shield µ = linear attenuation coefficient [cm–1] d = thickness of the scattering shield The absorption and scattering of radiation of radionuclides is often used in industry for thickness measurements or material testing (see Figure 3). This method can also be used for determination of the thickness of metal coatings or of liquid filling levels. Fig. 3: Thickness measurements by use of radioactive radiation Experimental 1. Precautions Work carefully and pay particular attention to avoid incorporation. Avoid contamination of the the measuring instruments and chamber. 2. Experimental set-up Insert a plexiglass slider into the lowest position of the counting chamber (to avoid backscattering from the chamber material). Insert a second slider with holders for the Al backscatterers, and finally prepare a third slider to carry the radioactive substance. In this setup only the β-particles that are backscattered through the hole in the aluminum carrier are measured (Figure 4). Fig. 4: Experimental setup 4 3. Procedure 1. 2. 3. 4. 5. 6. Determine the background radiation in the chamber with all sliders but without any substances (for checking later the chamber for contamination). Gate time: 5 min. Use tweezers while transferring the substance from the desiccator and insert it into the hole slider with the radioactive side down. The radioactive nuclide 204Tl is located in the notch of the aluminum carrier. Put the aluminum plug into the hole and determine the background again. Here, the bremsstrahlung and the γ-radiation are considered. Gate time: 5 min. Remove the aluminum plug. Measure the aluminum absorbers of different weights per unit area. The appropriate weight per unit area specification can be found inside the cover of the box. Gate time: 60 sec for each measurement. Note: Keep the distance from the absorber surface to the substance and to the window of the Geiger-Mueller counter constant (that means, the surfaces must "grow downwards"). For this reason the numbers of the aluminum absorbers must be put upside down into the hole slider. Ask the assistant how many aluminum absorbers should be used for the measurements (up to a maximum weight per unit area of 14000 g⋅cm-2). Keep the hole slider which contains the 204Tl material in the highest position. a) Install the hole slider with the backscattering disks in the second position from the top. b) Leave the plexiglass slider in the lowest position in the counting chamber. Gate time: 60 sec for each measurement. The thickness of the samples of each material is larger than the saturation thickness for the β-radiation in the given material. The following materials are used: Material: Z: 7. C 6 Al 13 Fe 26 Cu 29 Ag 47 Sn 50 Pb 82 Plexiglas ? Brass ? Remove the 204Tl containing substance from the counting chamber. Determine the background in accordance with 1. Evaluation 1. Use the Program Excel for your evaluation, draw the net counting rates (experimental counting rates minus background with aluminum plug in accordance with 3) as a function of the weight per unit area of the backscatter (with errors). Determine the thickness of the saturation from the curves. Fig. 5: Plot of the counting rates as a function of the weight per unit area of one absorber material (d: Weight per unit area; RN: Net counting rate; (*): Maximum backscattering density, saturation region) 5 Select two scalings for the axis: 0 – 700 g⋅m-2 and 0 – 7000 g⋅m-2 (see above). Determine the β-energy of 204Tl with the help of the determined saturation thickness (consider the experimental error) and figure 6. Fig. 6: β-energy as a function of the maximum range [mg·cm–2] Note the different units used in the two diagrams and the relation: rmax = 5⋅dmax 2. Use the Program Excel for your evaluation, draw the net counting rates (experimental counting rates minus background with aluminum plug in accordance with 3) as a function of the atomic number of the used backscatterer with errors. Determine the “atomic number“ of plexiglass and brass with the help of the curve. Fig. 7: Dependency of the backscattering as a function of the atomic number. 6 Data sheet 1: Backscattering of β–-radiation Geiger-Müller Counter Background (including bremsstrahlung) Nuclide: 204 Tl Weight per unit area of the absorber [g⋅m–2] 7 N R01 = R02 = t [min] RM [min–1] [min–1] [min–1] RN [min–1] 2 σR [min–1] Data sheet 2: Backscattering of β–-radiation Geiger-Müller Counter Background (including bremsstrahlung) Nuclide: 204 Tl 8 Backscattering material Atomic number C Al Fe Cu Ag Sn Pb Plexiglass Brass 6 13 26 29 47 50 82 ? ? N R01 = R02 = t [min] RM [min–1] [min–1] [min–1] RN [min–1] 2 σR [min–1] Chapter 7 Self Absorption of β-Radiation Task 1. Prepare eight CaCO3 samples with different thicknesses. 2. Determine the saturation layer thickness of CaCO3 for the β-radiation of 45Ca. Do this by determining the net counting rate as a function of the area weight of the precipitate. Fundamentals Ionizing radiation can be partly absorbed by the radioactive substance itself. This part of the absorption is called self-absorption. Self-absorption depends on the thickness of the substance and on the type and the energy of the radiation. In the case of γ-radiation self-absorption can be neglected because of the strong penetrating power of this radiation. The extent of self-absorption becomes important with α-emitters and with low-energy β-emitters. One way to eliminate error due to self absorption is to work with samples that are so thick that the radiation from the bottom layer of the sample is completely absorbed and never leaves the sample. This saturation layer thickness must be determined experimentally. The second possibility is to correct the measurement rates to a thickness of zero. Fig. 1: Saturation layer thickness Experimental 1. Precautions There is no danger due to radiation fields because of the low energy of the β-radiation of 45Ca (Eβ = 0.3 MeV; t½ = 165 d). But be careful with incorporation: Danger exists when measuring the dry substances (formation of radioactive dust)! The samples may only be transported in the desiccator. 1 2. Equipment 1 Hahn’s suction filter with witt’s suction bottle, glass cylinder, specialized filter paper, 1 beaker 400 ml, 1 beaker 800 ml, 8 beakers 50 ml, 1 glass rod, 2 tweezers, 8 aluminum holders for the filter cakes, 2 piston pipettes (2-5 ml) 1x active and 1x inactive, 1 waste bottle for liquid 45 Ca waste, 1 waste bucket for solid 45Ca waste, 1 Geiger Mueller counter in a measurement chamber, 1 computer 3. Chemicals CaCl2 solution I (0.25 mol/l; A = 33 300 Bq⋅ml–1) CaCl2 solution II (0.025 mol/l; A = 3 330 Bq⋅ml–1) Aqueous ammonia solution (6 mol/l) Na2CO3 solution (1 mol/l) 4. Experimental procedure Preparation of the precipitates 1. Install the suction bottle in the working area for radioactive materials! 2. Prepare the CaCl2 solution II: 5 ml of the solution I is given into a 50 ml measuring flasks and is filled up to the calibration line with distilled water! 3. Prepare the CaCO3 precipitates! a) The non-radioactive solutions are added to the beakers no. 1 to no. 8 according to the amounts in table. No. Beaker [ml] CaCl2 solution I [ml] CaCl2 solution II [ml] Distilled water [ml] NH3 [ml] Na2CO3 [ml] Expected amount of CaCO3 [mg] 1 2 3 4 5 6 7 8 50 50 50 50 50 50 50 50 1 1 2 3 4 1 4 6 5 - 5 5 10 10 5 10 10 25 1 1 1 1 2 2 2 2 1 1 1 1 2 2 2 2 2.5 10 15 25 37.5 50 75 100 b) A membrane filter is put on the Hahn suction filter and the equipment is checked for tightness! For this the glass cylinder is filled up to ⅔ of its overall height with distilled water. If no liquid runs between the glass cylinder and the frit, the water is sucked off in a way that a small quantity of liquid remains on the filter paper. 4. Filtrate and wash the precipitates: Suction filtration is performed with the precipitates in beakers no. 1 to no. 8. Be careful: Whirl up the precipitates with a glass rod first and then transfer the solids 2 into the glass cylinder of the suction filter with the use of a glass rod to avoid splashing of the radioactive mixture. Please take care that the glass cylinder is never sucked empty (regulation of the vacuum in the Witt’s pot). The solids from the beakers must be transfered quantitatively. The precipitates are washed twice with water and dried with a stream of air. Please avoid a cracking of the solid layers. The sucking device is checked each time for tightness! Additionally, make sure that the 400 ml beaker in the Witt’s pot does not overflow. Empty the liquid into the 800 ml waste container. 5. Prepare the samples! The glass cylinder is carefully removed with a slight vacuum, so that the filter paper remains sticking to the frit. It is then carefully removed with tweezers, whereby the filter cake should not be damaged. Some drops of a glue-acetone mixture are distributed into a small aluminum plate, which carries the number of the filter cake on the reverse side. The filter paper is put into the center of the small plate. Subsequently, the substance is put into a desiccator for drying over night. Decontamination All contaminated devices should be carefully decontaminated subsequently with half concentrated HCl and distilled water. The radioactive waste solutions are collected in the 800 ml beaker. Special attention should be paid to the frit. The glass containers must be checked with a detector after the decontamination by a wipe test, since the weak β-radiation of the 45Ca does not penetrate the glass walls. After successful decontamination, the contents of the 800 ml beaker are emptied into the bottle with the radioactive 45Ca waste. Liquids for the decontamination should be used as sparingly as possible – however, use as much as necessary. Operate economically! The treatment of radioactive waste solutions is very expensive. Measurement Determine the background: Gate time 5 min Each substance is measured for 60 sec. in a slider with a high ring in position 4 (from the top) in the measurement chamber. Subsequently, the background is determined again (5 min); check the measurement chamber, the slider as well as the ring for contamination. After measurement, put the samples in the active waste bottle for solid 45Ca. Evaluation The inner diameter of the glass cylinder is 1.6 cm, the surface is 2.0 cm2 according to the formula: Area = π⋅r2. Using the amounts of the CaCO3 precipates, summarized in the table, the weights per unit area as well as the thickness of the solid layers are calculated for each sample. ρ CaCO3 = 2.93 g ⋅ cm −3 Use the program Excel for the evaluation, the net counting rates as a function of the weight per unit area. Then determine the thickness of the saturation layer for 45CaCO3 precipitates. 3 Data sheet: Self-absorption of β-radiation Geiger Müller Counter Nuclide: 45Ca No. N t [min] RM [min–1] 1 2 3 4 5 6 7 8 Saturation layer thickness [cm]: 4 RN [min–1] R0 = [min–1] 2σR [min–1] Thickness of the layer [mg⋅cm–2] [cm] Weight per unit area saturation layer Chapter 8 The “Thorium Generator” – Determination of the half-lives of different members of the natural thorium decay series Task 1. Separate the daughter nuclides 212Pb, 212Bi, 208Tl and 208Pb from the naturally occurring isotope mixture of the 232Th decay series. 2. Perform an energy calibration of the NaI scintillation counter by using 137Cs 3. Separate the nuclides 212Pb, 212Bi and 208Tl by chemical methods. 4. Determine the half-life of 212Bi indirectly by means of the decay curve of 208Tl. Fundamentals 1. The thorium decay series In nature, the element thorium is found almost exclusively as isotope 232Th. This radioactive isotope is the first member of the 4n-decay series: 232 10 − − a) ,7 a ) ,13 h) →228 Ra β(5 →228 Ac β( 6 →228 Th Th α(10 In geological periods, the short-lived daughter 228Th is formed from the first member, 232Th, of the thorium decay series as shown in the scheme above. 228 (1,9 a) ( 3, 64 d ) Th α →224 Ra α →220 Rn As 228Th continues to decay, 220Rn is formed following two α-transformations. This radioactive noble gas is constantly released from soils containing thorium. Radon, in turn, decays to 212Bi as shown in the following scheme: 220 − h) ( 0.15 s ) ( 55.6 s ) Rn α .6 →212 Pb β(10 →212 Bi →216 Po α Decay of 212Bi proceeds in two different paths. The first possibility (36.2 %) occurs by α-transformation to produce 208Tl; the second (63.8 %) results in 212Po following a β--transformation. Ultimately, both of the paths result in the formation of 208Pb. 1 − 212 Bi α(60.6 min) → 208 Tl β(3.1min) → 208 Pb 212 min) μs) Bi β(9 → 212 Po α(0.3 → 208 Pb − 2. The radiochemical equilbrium The so-called “radiochemical equilibrium” should not be compared to thermodynamic or kinetic equilibriums. The “radiochemical equilibrium” is not reversible i.e. it cannot be attained from both side of the equation. Furthermore, it can never be considered as a stationary state. A “radiochemical equilibrium” is reached when the speeds of decay of the mother and daughter nuclides are matching thus leading to: NM ⋅ λM = ND ⋅ λD Which translates to: AD = AM (1) (2) The activity of the mother and daughter nuclides equalize when the “radiochemical equilibrium” is reached. Three cases can be described: The secular equilibrium: T1/2 (Mother) >> T1/2(Daughter): The activity of the daughter nuclide approaches asymptotically the activity of the mother nuclide. The transient equilibrium T1/2 (Mother) > T1/2(Daughter): The activity of the daughter nuclide decreases along the activity of the mother nuclide. The absent equilibrium T1/2 (Mother) ≤ T1/2(Daugther): The daughter´s activity continuously increases. An equilibrium cannot be attained. Die Zeit tGGW, die für die Einstellung des „Radiochemischen Gleichgewichts“ benötigt wird, kann nach Gleichung 3 ermittelt werden. Beim säkularen Gleichgewicht kann λMutter vernachlässigt werden. The time tEq required for the establishment of the “radiochemical equilibrium” can be determined using equation (3). λMother can be ignored in the case of a secular equilibrium. tEq = (𝜆𝜆 ln 0,01 𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷ℎ𝑒𝑒𝑒𝑒 −𝜆𝜆𝑀𝑀𝑀𝑀𝑀𝑀ℎ𝑒𝑒𝑒𝑒 ) (3) Systems with long-lived mother nuclides and comparatively short-lived daughter nuclides are often used as radionuclide generators for medical or technical applications. The mother nuclide is brought to chemical form suited for a separation from the daughter nuclide. The latter can then be separated or “milked”. Such setups are also referred as “Cow”. Those systems allow the production of radioisotopes without a fission reactor. The half-life of the mother isotope determines the useful life of such a generator. 2 3. The thorium cow Since 220Rn is a gas and is easily separated from its “parents”, its decay products can be collected in a carrier-free form. 216Po daughter nuclei, which develop from the α-decay of 220Rn, as well as 212Pb daughter nuclei, which are subsequently formed from 216Po, are affected by a strong recoil and are positively ionised. Therefore, these daughter nuclei can be deposited on a negatively charged tantalum metal sheet (cathode). A device which makes such separation possible is known as a “thorium generator” or “thorium cow.” This device consists of a sealed cylindrical container with the approximate height of 10 cm, in which a 228Th-containing substance is found. Also, within the device is a tantalum metal plate with a negative potential gradient relative to the chamber wall. The 228Th activity of the substance used in this trial amounts to approximately 37 kBq. Once the thorium generator is closed, a radiochemical equilibrium will be established between 228Th and 220Rn and their decay products. 228 Th: t½ = 1,9 a or. t½ = 5,99 ⋅ 107 s, λ1 = 1,16 ⋅ 10-8 s-1 220 Rn: t½ = 55,6 s, λ2 = 1,25 ⋅ 10-2 s-1 The separation of the daughter nuclides onto the tantalum cathode occurs when a voltage is applied for 24 hours to the device. To remove the plate, the voltage is turned off and the reaction vessel is shortly opened. 4. Separation of 212Bi from a mixture of 212Bi, 212Pb, 208Tl and 208Pb. The daughter nuclides are removed from the tantalum plates using concentrated nitric acid thus oxidizing the metallic deposits according to equation (4) to cationic species: 3 Pb + 8 HNO3 → 3 Pb(NO3 )2 + 4 H2 O + 2NO (4a) 3 Tl + 4 HNO3 → 3 TlNO3 + 2 H2 O + NO (4b) Bi + 4 HNO3 → Bi(NO3 )3 + 2 H2 O + NO (4c) The separation of the dissolved nuclides is possible thanks to the different basicity of the metal hydroxides. The hydroxides of the daughter nuclides are formed are at different pH values and precipitate from the aqueous solution (pH: Bi(OH)3<Pb(OH)2<Tl(OH)). EDTA is used to form a soluble lead complex to a avoid a co-precipitation with the other elements. Equation (5) shows the formation of this complex. 3 O 2O O N Pb2+ + EDTA4- O Pb O N O O O (5) Because lead is masked as EDTA-Lead complex and remains in solution and thallium only precipitates as a hydroxide not before very high pH values are reached, it is possible to selectively isolate bismuth at a pH of 7 according to equation (6). Bi(NO3 )3 + 3 NaOH → Bi(OH)3 ∙ + 3 NaNO3 (6) The formed precipitate is filtered and dissolved in hydrochloric acid leading to chlorido complex of bismuth(III) e.g. [BiCl4]-, [Bi2Cl10]2- or [BiCl5]2-. 5. Indirect determination of the half-life of 212Bi with interval separation from 208Tl 208 Tl is rapidly formed by the decay of 212Bi in the solution obtained above. The half-life of 212Bi is determined indirectly by separating its daughter nuclide 208Tl at known intervals. The bismuth solution is therefore loaded on a strongly basic anion exchange column. The negatively charged bismuth chlorido complexes are retained by the resin while the cationic thallium can be eluted with diluted hydrochloric acid (0.5 M). The decay curve of the obtained 208Tl can be measured precisely and then back extrapolated to the time of the elution. A radiochemical equilibrium was established at the moment of the elution meaning that the decay constant of bismuth and thallium were equal. By repeating the separation at given time intervals- separated at least 30 min. or 10 half-lives of 208Tl to re-establish a radiochemical equilibrium- a series of extrapolated activity at elution is obtained. The decay curve of 212Bi is obtained by linking those points. 6. Carrier free conditions Carrier free conditions are met when a radionuclide is isotopically pure meaning that is not diluted by stable isotope of the same element. The newly formed isotopes are separated under carrier free conditions during the first part of the “Thorium-cow” experiment. 220Rn leaves the 228Th sample and its daughter isotopes are deposited on the cathode. The quantities of daughter nuclides obtained are extremely limited. As such it would be pointless to attempt to precipitate those elements as their solubility products could never be reached. The addition of a non-radioactive “carrier isotope” is thus required for the subsequent work-up. The added compounds should ideally exist in the same chemical form as the radionuclides. Lead acetate is used as a “lead carrier” and bismuth nitrate as a “bismuth carrier”. The solubility product of bismuth hydroxide can only be reached by adding bismuth 4 nitrate thus allowing the isolation of the radioactive bismuth. On the other hand, the addition of non-radioactive lead allows the radioactive lead to remain in solution. The radioactive leadwhich is extremely diluted might co-precipitate with bismuth without such a “retention carrier”. Experimental part 1. Precaution measures The tantalum metal sheets are located in a glass beaker covered with cellulose and are brought to the workspace by an assistant. All isotopes of the decay series prior to 212Pb are almost completely decayed prior to the start of the experiment. Please follow the usual precautions while working with radioactive materials to avoid incorporation and contamination. 2. Chemicals The following solutions are needed for this experiment: HNO3 conc. 0.1 M EDTA 6.0 M NaOH 2.0 M NaOH 2.0 M HCl 0.5 M HCl 0.59 M Bi(NO3)3 0.59 M Pb(CH3COO)2 Thymolphthalein indicator Dissolution Titriplex-III for the complexation of lead Hydroxide precipitation Washing solution Dissolution of the hydroxide Eluant Bismuth carrier Lead carrier Acidic: colorless, basic: blue These solutions have to be prepared: Glass I and Glass II for the dissolution of activity 1.5 ml HNO3 conc. 1.5 ml Bi(NO3)3 solution 1 ml Pb(CH3COO)2 solution Glass A: Glass B: Glass C: Glass D: Glass E: Glass F: 5 5 ml Titriplex for the complexation of lead Thymolphthalein indicator (3 drops) 8 ml 6.0 M NaOH for the hydroxide precipitation 8 ml Titriplex + 2 ml 2.0 M NaOH as washing solution 5 ml 2.0 M HCl for the dissolution of the hydroxide 5 ml distilled water to wash the filter 10 ml 0.5 M HCl as eluant (3x, since there are three elutions) 3. Procedure 3.1 Separation of the isotopes of the Th decay chain the Th cow. A thin 228Th sample is deposited on the bottom of cylindrical pot which is tightly sealed to prevent the diffusion of 220Rn. Several tantalum plates are attached on a metal hook. A cathodic current then flows for 24h though the plates. A “radiochemical equilibrium” establishes itself between the decay products of the Th decay chain. Radon diffuses continuously from the Th probe. Its daughter elements deposit on the cathode as they are ionized by a strong recoil of their alpha decay. The nuclides listed below accumulate on the electrode 212Pb t½ = 10.6 h β-: 0.35 MeV γ: 238.9 keV (47%) 212Bi t½ = 60.6 min α: 6.05 MeV β-: 2.27 MeV γ: 727 keV (7.1%) 212Po t½ = 0.3 µs α: 8.795 MeV γ: 2610 keV (2.6%) 208Tl t½ = 3.1 min β-: 1.79 MeV γ1: 583 keV (86%) γ2: 2615 keV (100%) 208Pb stable 3.2. Setting of the measurement device The measurement is performed using a sodium iodide scintillator. 208Tl has two gamma lines at 583 keV and 2615 keV. A multi-channel analyzer is used for the purpose of this experiment. An energy calibration of the measuring device muss be performed prior use. The calibration is made as such to allow the measurement of the gamma line at 583 keV. The decay lines are measured without resolving the energy. The measurement of the decay curves using a finite time window is thus possible. Please prepare the detector for the measurement. Open the program ProSpect at the computer. Connect the multichannel analyzer and the computer, please choose the menu point “Connect to Device” and choose the detector “Thoriumkuh”. Optimize the high voltage (maximum 700 V) and the gain of the detector in a way that the Cs-Photopeak (E = 662 kV) at around 1/2 of the energy scale appears. Perform an energy calibration, choose the acquisition mode PHA (also see experiment gamma spectrometry). 137 Change the acquisition mode to MCS and use the following settings: Acquisition Mode: MCS MCS Conv Gain: 256 6 Dwell: measuring time Disc Mode: ROI Start Channel: 400 End Channel: 800 Measure the background for 5 min. The measurements last 1min each. For the evaluation, you need to stop the measurement and save the data. You can save the data with de Butten “EXPORT DATA”. You receive a file with the ending *.csv where the data are separated by comma. This file can be opened by a program such as Excel. Use the option “characters like commas separating fields” while importing the data. 3.3. Preparation of the chemical separation Working with radioactive compounds, each step should be carefully planned and required items should be prepared sensibly. All inactive substances required for the procedure are to be prepared beforehand. When working with short-living nuclides like 208Tl, all needed items must be readily available. Make sure that the separation columns are ready to use (ask an assistant). Turn on the water bath and prepare the suction device (tightness test). 3.4. Production of the radioactive parent solution The tantalum plates are brought by the assistant to experimental workspace. To separate the formed radioactive thorium, the metal sheet is placed in nitric acid (Glass I) and warmed for 5 minutes in a water bath. Beaker number I can then be removed from the water bath and placed in the plastic working chamber. Next, the metal sheet is transferred with tweezers to beaker number II and heated for additional 5 minutes in the water bath. Subsequently, the metal sheet is removed from solution II, rinsed with a few drops of water and placed back in the transport container. The two solutions are then pooled into beaker I. 3.5. Separation of bismuth Bismuth is first separated from the radioactive solution, which contains 212Pb, 212Bi, 208Tl and Pb. This separation is successful due to the different basicities of the hydroxides of these metals, which decreases in the order Tl > Pb > Bi. Bismuth hydroxide precipitates at pH 7, whereas thallium remains in solution. In order to completely avoid the co-precipitation of lead, EDTA is used as a lead-complexing agent. 208 For this separation, solution A is added to beaker I, in order to mask the lead which is contained in the solution. Subsequently, a small amount of indicator is added to beaker I and the prepared NaOH (solution B) is then also added until the solution is getting blue. A white precipitate of Bi(OH)3 begins to form and the mixture is again placed in the water bath for 3 minutes. After cooling, the precipitate is filtered with a membrane filter and washed several times with solution C. After the solution from the last washing step is removed by sucking air through the filter, the Bi(OH)3 precipitate is dissolved in HCl. In doing so, the filter paper is carefully transferred using tweezers into beaker D, which is then placed in the warm water bath for a short period of time. Subsequently, this solution is then diluted with 5 ml distilled water (beaker E) by rinsing 7 off the filter paper held above Glass D. The filter paper is then wrapped in cellulose and placed in the radioactive thorium generator waste. 3.6 Separation of bismuth and thallium As a result of radioactive decay, 208Tl is quickly reformed in the freshly produced 212Bi solution. In order to separate 208Tl from 212Bi, Cl- loaded anion exchange resin (Dowex) is employed. Using a glass rod, the HCl-bismuth solution is carefully loaded to the column. This solution contains bismuth as tetrachloro complexes [BiCl4]-, which bind in place of the chloride ions on the anion exchange resin. The solution is allowed to run slowly through the column until the meniscus of the liquid reaches the surface of the exchanger material. In order to avoid separation failure, the column should not be too long. The work is performed carrier-free, therefore, only traces of thallium are present. Once the column is loaded with bismuth, time is required to establish the “radiochemical equilibrium”. 30 minutes after the column is loaded, it can be assumed that the “radiochemical equilibrium” between 212Bi and 208Tl has been established and the separation of 208Tl can begin. This waiting period can be used to prepare the counting device. For the separation of thallium, 10 ml of 0.5 M HCl (beaker F) are added to the column. This solution is allowed to run quickly though the column until its meniscus reaches the surface of the resin. The stopwatch is started once the stopcock is turned and runs throughout the experiment (≈ 1.5 h). The eluate is collected in a suitable container and measured immediately. Work should be performed as efficiently as possible. 3.7. Measurement of the decay curve of 208Tl Choose a measurement time of 1 min before you start the first separation. After the separation put the elution beaker with the solution on the detector, start the measurement at a full or half minute. You need the exact measurement time for the evaluation. Please write down the starting time. Stop the measurement 27 min after the start of the elution, save the data-file. The last value is used as background for the measurement just done. Clean the elution beaker for the second elution that should start exactly 30 min after the first one. All in all, three elution’s are carried out with 10 ml of 0.5 ml of HCl. Evaluation The evaluation is performed using a spreadsheet software e.g. MS Excel. The net count rates of Tl are logarithmically plotted relative to the time elapsed after each chromatographic separation from 212Bi. The logarithmic values can be calculated prior plotting, or an automatic scaling of the Y axis can be selected in the spreadsheet software. Three linear regressions of the decay of 208Tl (in the form y=mx+n)in total can be obtained. The half-life can be obtained from each slope m. The theoretical values at the time of elution of the count rates are then obtained by extrapolating the slopes. The three obtained linear equations are solved according to their respective time of elution (x value of the elution 30/60/90 min). The obtained y values a.k.a. as head points are proportional to both the activity of 208Tl and 212Bi. 208 The three obtained pairs of values (count rates at a specific elution time) are plotted logarithmically-as done with thallium previously. A linear regression of those three points leads to the 8 linear decay slope of 212Bi. As the activity of the bismuth declined throughout the experiment, it can be calculated from the obtained decay slope as done previously for the thallium. Measurement results: Half-life 208Tl from the 1st separation: Half-life 208Tl from the 2st separation: Half-life 208Tl from the 3st separation: Average half-life 208Tl: Half-life 212Bi: 9
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )