Advanced Analytical Chemistry Chapter 0: introduction Chapter 1: X-rays Chapter 2: Element Mass spectrometry Chapter 3: Molecular Mass spectrometry Chapter 4: Surfaces Chapter 5: Radiochemistry Chapter 6: Seminar I: X ray diffraction Chapter 7: Particle Size Determination 1 Chapter 0: Introduction Course Content Lectures ◦ 1/10 ◦ 15/10 ◦ 29/10 ◦ 12/11 ◦ 26/11 ◦ 10/12 ◦ Slides: Toledo (Information in English) 6 times 3 h, instead of 12 times 1.5h, English Exam (70%) Contact: Bram.Verbinnen@uantwerpen.be Seminars (30%): 12h ◦ Exercises ◦ Guest lectures ◦ …. ◦ Dates: TBD, see schedule Toledo Obligatory presence Contact: Antoinette Deschuytere Part1: Spectrometry Atomic X-Ray Spectrometry (XRF) Molecular X-Ray Spectrometry (XRD) Atomic Mass Spectrometry (ICP-MS) Molecular Mass Spectrometry (…-MS) (Raman Spectrometry) (NMR-Spectrometry) Spectrometry: you will always receive a spectrum. X axis will always be a spectrum. Part 2: The study of Surfaces: Electron Spectroscopy (AES) Ion Spectroscopy (SIMS) Surface Photon Spectroscopy (SPS) Electron-stimulated micro-analysis methods (SEM/TEM) Scanning Probe Microscopy (STM/ATM) Part 3: Particle Size Determination Voltammetry … 2 OVERVIEW OF THE TECHNIQUES Belangrijkste figuur van het vak! Classified according to the detection limit, resolution and the y-axis. Detection limit: the lowest concentration that can be distinguished from the background. Resolution: the smallest distance needed between two objects to see them apart from each other. If you can distinguish them apart from each other: the resolution is lower than the distance between the two objects. Example: ICP-MS, if you want to measure As: 75 m/q, but there is also ArCl: 75 m/q. With a bad detector (high resolution) we cannot see them apart, if we have a very low resolution (good) we can see As: 74,9 and ArCl 75,4. has nothing to do with the detection limit because both signals can be far above the detection limit. For every technique that we are going to study: have an idea what the detection limit and spatial resolution are, be able to compare two different techniques (relative, not absolute). 3 Chapter 1: X-rays General: ◦ ◦ ◦ Measuring emission, absorption, scattering, fluorescence, diffraction of X-Rays Determining elements: identification and quantification Elements heavier than Na X-rays: part of EM spectrum. What can we do with x-rays? Different processes. Fluorescence: incoming em wave, it is absorbed, secondary em wave emitted that has a lower energy content than the incoming em wave, higher wavelength. Scattering: verstrooiing, happens in all directions. All of these scattered waves can interfere with each other, can show interference (constructive or destructive), most of them will destruct each other. When you have constructive however, we call it diffraction, you can measure it in this case. X-Rays ◦ ◦ ◦ E = h*c/lambda EM-rays between10-6 nm and 10 nm. Used in spectrometry: from 10-2 nm to 2.5 nm Use of continuous spectrum or line spectrum. High energy = low wavelength. They all interact with matter on a different matter. Gamma rays: X rays: what can happen? You will expell an electron from an atom ionization UV-VIS: electron shift to higher orbital IR: vibration Micro: rotation Radio: electron spin changes (not a lot of energy necessary) Producing X-Rays ◦ ◦ ◦ ◦ By impact of a beam of high energy electrons on a metal By using primary X-rays to produce secondary X-rays by fluorescence Radioactive sources Sources of Synchrotron Rays 4 Making X-Rays: The Continuous Spectrum ◦ ◦ ◦ ◦ ◦ Production of high energy electrons by heating a Cathode Accelerating electrons towards a metal anode (target) using a high potential (e.g. 100 kV) Electrons collide with the target and decelerate: transformation of kinetic energy in radiation energy (photons) Many collisions with different energy produce a continuous spectrum. Is called “White radiation” or “Brehmsstralung”. Cathode = negative. You produce high energy electrons with eg a sharp, needle like tip, by applying an electric field between cathode and anode, you will accelerate the electrons towards the positive anode, they collide with the target and they decelerate (lose energy). Incoming electrons will bend under the influence of the nucleus, the outgoing electron will have a lower energy. The excess in energy is released under the form of X-rays while bending around the nucleus. There are several paths that the electrons can make, the different deflection angles produce different X-ray frequencies. If you have a lot of collisions, you get a continuous spectrum. The atom from the nucleus around which the electron bends are specific X-ray source metal atoms. ◦ ◦ ◦ ◦ ◦ Smallest wavelength (highest energy): l0 Complete transformation of electron kinetic energy in photon Independent of target material, dependent on accelerating potential Formula (Duane-Hunt): V. e = h. n0 = h.c/l0 Allows determination of h 5 At zero intensity, you have lambda zero. Higher energy of your outgoing x-rays the more you go to the left, but intensity is zero. All of you incoming energy is absorbed, you had a collision with your nucleus and the incoming electron has completely stopped. Different acceleration potentials: difference between energy before and after can have a wider range. Intensiteit moet je zien als: ‘hoeveel electronen hebben x-rays uitgezonden met die energie’. The difference is the acceleration Potential. 50 kv acceleration potential has bigger curve. Tongsten = wolfram Lambda zero depends only on you acceleration potential!!! One of the ways in which you can determine the planck constant. 6 Making X-Rays: The Line Spectrum: ◦ ◦ ◦ ◦ Lines superimposed on continuous spectrum Typical for elements with Z >23: Very simple spectrum with 2 groups of lines: K and L Remember: 1 A = 0,1 nm Not only continuous spectrum, but also some lines superimposed on the spectrum. These line spectra originate from the knocking out of an electron in the shell, secondary electron is removed from is orbital. There is now a vacant place and can be filled with an electron from a higher orbital, thus with higher energy. There is thus an excess in energy, this is then expressed by releasing an x-ray. Difference between the 2 orbitals (vacant place vs where the replacement electron comes from) is always the same, thus the line spectrum is always the same. You give all the lines a name, and it depends from which orbital the electron is knocked out (K-L-M orbitals). Alfa or beta refers to which electron has filled the gap (one up is alfa, two up is beta). Kalfa is lower in energy because it is placed at higher wavelengths. Not a statistical distribution as a continuous spectrum. The ability to see the line spectra is also dependent on you acceleration potential, if it is very low you may not be able to (because lamda zero shifts to the right). Atom number larger than 23, at least 11 electrons (at least 1 on the M orbital). Only from atomic number 24 you have L lines. Elementen K en Na zijn interessant 7 ◦ ◦ ◦ Lines disappear when accelerating potential (energy) too low. Minimum required energy increases with atomic number (Z). Comp. Molybdenum (from 20kV) and Tungsten (lines only starting from 70kV) Energy required to see such lines increases with atomic number, influence of nucleus is higher, you need more energy to expel an electron from its orbital. ◦ ◦ Lines are the result of electron jumps of inner electrons. K series: Accelerated electrons collide with electrons from shell n = 1 An electron is ejected forming an excited ion. Electrons from higher levels fall down and emit X-Rays. You have a distribution on your energy levels, but it is much smaller than your continuous spectrum, you still call it a line spectrum. There are different levels for each orbital!! 8 ◦ ◦ Increasing nuclear charge (Zeff) has an effect on energy needed to remove electrons AND on difference between energy levels (shells). Spectrum is INDEPENDENT of the physical or chemical state of atom (in compounds) because it concerns only electrons not involved in chemical bonds. You cannot distinguish with x-ray techniques in the chemical state of an atom with this technique!! You will only expel electrons from the inner orbitals, they do not define the chemical state, do not make bonds or ions. You will not touch the outer electrons with this techniques. Don’t touch outer orbitals, only inner orbitals, so you cannot distinguish between chemical states If the energy is high enough it can expell an electron!! To release a electron it has to be exactly the right energy! There will be a vacant place, an electron from a higher level will decent an there will be secondary x-ray. 9 Producing X-Rays ◦ By impact of a beam of high energy electrons on a metal, ◦ By using primary X-rays to produce seconday X-rays by fluorescence (see later), Primary x-rays can also expell electrons. You need a certain, well defined amount of energy of the incoming beam to expel this electron. The incoming x-ray beam has a very high energy, so not much expelling, but when it does, there is a large excess of energy and that is all converted in kinetic energy of the outgoing electron. Again a vacant place, replacement electron, emission of a secondary x-ray. Benefit over the first manner of making x-rays: you will not have a spectrum of x-rays, you will only have a very narrow band of energies of your x-rays. ◦ Radioactive sources: Nuclear reactions produce excited nuclei. Relaxation of the nucleus produces gamma-rays. Nuclei can also capture electrons (from K-shell). Higher level electron fall down to fill the space and produce X-Rays. Half-life of K-capture varies from a few minutes to thousands of years. ◦ Sources of Synchrotron Rays: Electrons at high speed in magnetic field. The deccelaration (deviation) of the electrons produces radiation (complete spectrum) of which X-rays. Link: http://www.esrf.eu/about/synchrotron-science/synchrotron-lightanimation Electrons emitted by an electron gun are first accelerated in a linear accelerator (linac) and then transmitted to a circular accelerator (booster synchrotron) where they are accelerated to reach an energy level of 6 billion electron volts (6 GeV). These high-energy electrons are then injected into a large storage ring where they circulate in a vacuum environment, at a constant energy, for many hours. Each time these electrons pass through an undulator, a device consisting of series of alternating magnets, they emit X-rays, which are directed along beamlines. CERN: collider with principle of synchrotron Synchrotron: you can also use it to produce x-rays You accelerate electrons at high speed in EM field, you can slow them down again and they produce x-rays. Advantage of this method: you can control the magnetic field and the amount of deceleration, you can determine the frequency of x-rays you obtain. You can very good select one wavelength of x-rays. Movie: electron will be accelerated, when it passes the beamline, some x-rays will be emite. You can use those energy for experience, (expl. With x-situ you can measure wich minerals there were. ) 10 X-rays and Matter ◦ Overzicht: ◦ Absorption Fluorescence Diffraction Comparable with other optical methods ABSORPTION: Techniques: Xanes, Exafs X-rays that are entering an atom. Certain amount of energy that is exactly high enough to expel an electron from the orbital. Highest changes of absorption: exactly the right amount of energy needed is given. Even if the energy of the incoming x-rays increases, the changes that the x-ray will be absorbed will be lower. Principle: Transfer of energy from X-ray photon to electron. Probability decreases with increasing energy. Sharp change (K or L) corresponds to excitation of K (or L) electron to the outside of the atom (without kinetic energy). When energy is higher then the electron also acquires kinetic energy. 11 Mass Absorption Coefficient: Law of Lambert-Beer: ln P0/P = µx with P0 en P: intensity incoming and transmitted radiation x : thickness µ : linear absorption coefficient : f(element, number of atoms) Or also: ln P0/P = µMρx with ρ: density µM : mass absorption coefficient (cm2/g): only f(element) If several elements: mM = WAmA + WBmB + WCmC + …. (with W = %mass) Not on the exam, this technique is not further discussed. ◦ FLUORESCENCE: Principle: Absorption of X-rays produces an excited ion For instance with an empty K-shell After a short time: relaxation to ground state by electron shifts. This produces new x-rays. (fluorescence) Wavelength is slightly greater (less energy): Incoming X-rays ionize atom completely Emission is only inside the atom. Ex. Absorption of Ag happens at 0,485 A (K-edge) and emission results in 0,497 A and 0,559 A. The wavelength of the emitted x-ray is slightly greater, you lose an amount of energy. 12 ◦ DIFFRACTION: Principle: X-rays (like other EM-radiation) interact with electrons of atoms resulting in scattering. In an ordered system (crystal) constructive and destructive interference occurs (because distance between scattering centers = wavelength) Result = Diffraction In phase vs out of phase. Diffraction only happens when the scattered waves are constructively interfered. If you have a crystal structure, you have multiple layers of crystals, the distance between the crystal layers depends on the cyrstals. When the difference between the path of the waves is a natural number of the wavelength. The extra pathlength that the wave has to travel has to be exactly a natural number times the wavelength. Difference in pathlength depends on the distance between crystal layers and the angle under which the incoming x-ray comes in. Bragg’s law: If the difference in path length AP + PC = nl the waves are in phase along line OCD (only reflection). But AP = PC = d sin q Therefore constructive interference if nl = 2 d sin q or sin q = nl/2d 13 Instruments ◦ ◦ ◦ Use of absorption, emission, fluorescence, diffraction Comparable with other optical methods Instruments contain same parts: a) Source, b) System for selecting a wavelength(range), c) “Sample holder” d) Radiation detector, transducer e) Signal processor, f) “Readout”. a) Source for X-Rays: X-ray Tube: Target material: W, Cr, Cu, Mo, Rh, Sc Ag, Fe, Co,…. By heating the filament, you will release electrons, they will be accelerated, will be forced from the negative anode to the positive anode. When they hit target material collision, deceleration for continuous spectrum. You have to make sure that not too many of your x-rays are already absorbed beryllium window. Also nearly vacuum inside the tube. 14 Radioisotopes: Secondary fluorescence Selection of wavelength or spectrum: Filters Filter: the Kbeta line will be completely absorbed by the filter. This is quite old and is not used much today. You only need a very thin sheet of zirconium. 15 Monochromator You can select the angle of the incoming x-rays to your crystals. If you change the angle, you can select the wavelength of your x-rays that will be emitted to your sample or transducer. Relationship sin(theta)= n*lambda/2d. You need one device that rotates the crystal, and one that rotates the sample. Sample needs to rotate with twice the speed. Important that you know it is 2teta 16 Sample holder Measuring the radiation: Transducers ◦ Previously: Photographic detection ◦ Today: transformation of radiation (photons) in an electrical signal “Photon counting”: transform a radiation quantum in a charge pulse that can be counted: Overzicht: Gas-Filled Detector Scintillation Counters Semiconductor Transducers Gas-filled detector Gas: Ar, Xe, Kr X-Ray: Ar e- + cation e- : ionizes several additional atoms e- move to anode, cations (slower) to cathode X ray enters gas filled chamber, when the x ray goes trough window, it hits the gas and it will produce an electron and a cation and they are accelerated and will collide with other gas atoms. They will ionize different additional atoms. Electrons will mote to the anode. cations go the the cathode. 3 types of gas-filled detectors: Ionization chambers Proportional counters Geiger tubes 17 Number of electrons per photon depends on the voltage applied. You have different regions. For x rays we use the proportional region, the amount of electrons that is produced is proportional to the amount of x rays coming in. Geiger counter cannot be used, it’s just a plateau. Geiger counter checkt voor radioactieve stralen, das zo een bakske da ze gebruiken! Let op 10 tot de 10de wow hoog getal, dus zelf kleine ray wordt zelfs gemeten! Geiger region is a plateau, geen verband helaas dus niet te meten? V < V1: recombination V1 < V < V2: Ionization: constant number of e-: not used for X-rays V3 < V < V4: Proportional region: rapid increase of e- with voltage V5 < V < V6: Geiger range: enormous amplification, but limited by ‘space charge’ counteracting the electric field, no distinction between type and energy of incoming radiation Proportional: number of electrons formed is directly related to the amount of x-rays coming in. Dead time ◦ Result of positive space charge ◦ Charge needs to be dissipated before new pulse can be detected ◦ Typically 50-200 µs for Geiger tubes: less used for X-rays 1 µs for proportional counters: most frequently used Dead time because there is such a magnification of your signal, is also present in proportional region but much higher in Geiger region. 18 Scintillation counters Scintillation = production of flash of light by passage of a particle (e.g. a photon) Semiconductor Transducer Important for detecting elements, not used for measuring molecules. Elemental spectroscopy. 19 Methods ◦ Fluorescence (XRF) Most used Because measuring emission is not practical (target must be in sample) Two types: WDXRF: Wavelength-dispersive XRF EDXRF: Energy-dispersive XRF Advantages: fast, quantitative determination of elements (except very light ones), non-destructive. You should be able to compare XRF and ICPF! WDXRF: Wavelength-dispersive XRF Wavelength Dispersive X-ray Fluorescence (WDXRF) is one of two general types of X-ray Fluorescence instrumentation used for elemental analysis applications. In WDXRF spectrometers, all of the elements in the sample are excited simultaneously. The different energies of the characteristic radiation emitted from the sample are diffracted into different directions by an analyzing crystal or monochrometer (similar to the action of a prism dispersing different colors of visible light into different directions). By placing the detector at a certain angle, the intensity of X-rays with a certain wavelength can be measured. Sequential spectrometers use a moving detector on a goniometer to move it through an angular range to measure the intensities of many different wavelengths. Simultaneous spectrometers are equipped with a set of fixed detection systems, where each system measures the radiation of a specific element. The principle advantages of WDXRF systems are high resolution (typically 5 – 20 eV) and minimal spectral overlaps. For both: your sample is always bombarded with x-rays, all of the fluorated x-rays come out of the sample at once. You have to distinguish between them, based on wavelength or energy. EDXRF: Energy-dispersive XRF Energy Dispersive X-ray Fluorescence (EDXRF) is one of two general types of X-ray Fluorescence techniques used for elemental analysis applications. In EDXRF spectrometers, all of the elements in the sample are excited simultaneously, and an energy dispersive detector in combination with a multi-channel analyzer is used to simultaneously collect the fluorescence radiation emitted from the sample and then separate the different energies of the characteristic radiation from each of the different sample elements. Resolution of EDXRF systems is dependent upon the detector, and typically ranges from 150 eV – 600 eV. The principal advantages of EDXRF systems are their simplicity, fast operation, lack of moving parts, and high source efficiency. You select based on their energy content, eg bending in a magnetic field. 20 ◦ Fluorescence (XRF) Most used Because measuring emission is not practical (target must be in sample) Two types: WDXRF: Wavelength-dispersive XRF EDXRF: Energy-dispersive XRF Advantages: fast, quantitative determination of elements (except very light ones), non-destructive. ◦ Absorption: not often used (mostly when heavy elements are present in a light matrix) ex. Lead in Fuel ◦ Diffraction 21 ◦ Oxygen calculated assuming all elements present as standard oxides: Drawback: cannot measure very light elements, oxygen e.g. the oxygen that is present in your samples will be measured as oxides. This is not always correct but it is the standardized way of comparing them as their oxides. ◦ What Nasa does on mars: 22 ◦ XRF matrix effects: Absorption of X-rays by the matrix Countered by: External and internal standard calibration Dilution of samples and standards e.g. fusion with a substance that absorbs X-rays only weakly Li-metaborate/tetraborate (9.75 g) for each 0.75g of sample Thin films Mathematical corrections Matrix effects: matrix of the sample that you measure also absorbes x-rays. If you measure gold, it will not always be on top of the surface, can also be slightly below the surface. Secondary x-rays generated at the gold needs to penetrate the above layer, you will always lose a part of the rays due to absorption. You need to counter is, calibration, dilution. ◦ X-ray diffraction Mostly for crystalline structures, Used for metals, polymers and other solids, Also for organic molecules: vitamins, steroids, antibiotics, nucleic acids…. Also qualitative determination of compounds (>< elements) X rays used in another way so you can extract different information from it. ONLY ISO MOSTLY. Constructive interference can only happen in a crystalline structure! Used a lot in geology, to determine cystal structures, XRD can be used for this. One of the seminars will also focus on the XRD. You can use it for quantitative and qualitative determination, how much is present and is it present? This is used to measure compounds, not an elemental composition. You hav an x ray source, focus on monochromator, then you are going shoot it at the crystal that we are going to analyse. You are going to change the angle, and let it circle around the crystal so that everytime one angle is selected. Then you will have constructive or destructive interference, this depends on the type of crystal that is present in your sample. 23 Different XRD peaks. First one SiO2, then quartz (also SiO2), cristobalite is also SiO2. you see that the XRD patterns as a function of 2*theta, is different for the three different minerals, although they have the same chemical formula. The distance between the crystal layers is different for the 3 SiO2’s, this way you can distinguish. You have clear peaks for quartz and cristobalite. Glass however is not a clear peak, glass is an amorphous material, reflected xrays will be scattered in all directions, no interferences, all wil go to the detecter and you will get a very broad peak. For XRD you really need crystalline materials. Destructive interference next to the peaks, no signal. Nieuwe powerpoint op Toledo met extra afbeelding voor die constructieve interferentie. It only depends on your incoming angle combined with the distance between your crystal layers. Very specific for one crystal structure. Rechtsbeneden: x ray source, the xrays will go directly to your sample, you will rotate the device, you select an angle and you detect which angle will give you constructive interference and you will get the graphs. 24 ◦ X RAY Diffraction Pattern Diffraction pattern with indication of the minerals that contribute most to the most important reflections. Zincite (‘Z’) was added as internal standard. The reflections of anhydrite (‘An’), Halite (‘H) and Sylvite (‘S’) are also indicated. You see a peak from zincite, that is used as an internal standard, to distinguish how much amorphous material is present in the material that you want to measure. Zincite is 100% crystalline, depending on the height of the signal that you get from the xrd, you can distinguish how much of the sample is amorphous or cytsalline, you can also calculate back to the whole sample. You can quantify the concentration of the material in the crystalline material. Based on the signal of the zincite, you can see here that almost 60% of you sample is amorphous. The baseline is due to the amorphicity, it’s scattering in all directions, for every directions you will have a baseline. Further refinement: Rietveld refinement Overlapping peaks Least-square approach Quantification possible 25 Developed a method to quantify xrd analysis, they were able to determine how much of the mineral was present. With xrays you can only measure the top layer, you have to make sure that your sample is a very fine powder so that you have a homogeneous standard. Limit of detection is not very high, if it is below 1%, you will not be able to detect it. ◦ In-Situ XRD More applications: seminar (Gilles Mertens – Qmineral) You can do in-situ measurements, scan along a certain 2theta, and you can see in real time how a reaction proceeds. Makes use of the synchotron xrays, you need always a stable xray, always at the same frequency with high energy. Used at cern in Geneva. Zie filmpje 26 EXAMPLE EXAM QUESTION !! Some genius scientists from KU Leuven recently published an (equally genius) article; the two first pages of this article are attached. A) In the boxed section ‘A’ in the ‘materials and methods’ section, the authors explain the use of XRD for their experiments. Explain in detail the following principles, and illustrate with a drawing (max 1 page for each principle): 2Θ= 5° to 75° Monochromatic Cu Kα radiation B) The detector in this XRD instrument is a proportional counter. Explain the principle of this detector, make use of a drawing. (max. 1 page) C) In the boxed section ‘B’, the authors describe that they measured the total amount of Cr (at parts per billion level, with ICP-MS) and the amount of Cr(VI) (using a colorometric method) by dissolving the samples in water and then performing the analysis on the dissolved samples. Another scientist claims that they had better measured the total amount of Cr in the solid material directly, by using XRF. Is this other scientist right? Why (not)? Which other technique could be used to measure Cr(VI) in the solid material directly? What is the disadvantage of this technique? Concerns the whole xray chapter. 2 pages of the article that he wrote. 1) Be able to tell what the 2theta means, with the goniometer etc you use copper as a metal sheet, you use Kalfa and you select the wavelengths with monoch. 2) Proportional counter with drawing 3) Comparison between icp-ms and xrf. Ppb level!! It is not true, detection limit is only in the 0,1% with xrf and ppb is much lower. You need to refer to the first big drawing that relates the different detection limits with the different techniques. Other surface techniques that are used to distinguish the oxidation states? Last lessons. 2 years ago, exam question regarding this chapter XRD What does 2teta mean? What does it mean; Monochromatic CU Kalfa? Proportional counter? 27 Chapter 2: Mass Spectrometry Last year: GC-MS, LC-MS: Compound identification, you do not get elemental information. Now we are going to study elemental MS, we should be able to compare this with xrf. What do you need to perform MS? Spectrum as a function of mass/charge ratio, we need to bring our elements in a charged state. The big difference between elemental and compound : we first need to dissociate our compounds into elements, and then me need to ionize with ICP and then to the MS. What is plasma? Gasious phases containing only ions, highly excited ions in a gaseous phase, they are moving in an EM field, hence the inductively coupled plasma. Exctied ions can collide with the incoming elements that you want to measure, then elements will be ionized. Plasma has high energy, so its very easy to ionize. Other thechniques are not able to ionize so much of the elements. Nu gaan we het gebruiken voor elemental onderzoek, vroeger geleerd als compound onderzoek. Techiques already used: JCMS Liquid and glass chromatography. JCMS, LCMS… Elemental or compounds ? COMPOUNTS, any organic compounds like methanol, ethanol… This relates to the XRF bc this will give ELEMENTAL information. General: ◦ Determination of elements: identification and quantification ◦ Advantages vs. optical methods: • Much lower detection limits & BROAD linear detection range (see next slide) • Simple and clear spectra • Determination of isotope ratios • Quick and multi-element ◦ Disadvantages: • EXPENSIVE apparatus • Instrument drift (up to 5 – 10% per hour) • Interferences possible (see further) What is inside and how much? AAS: has a plateau that you will reach, the signal that you will reach is always the same. Not true of Mass spectrometry, you have a linear line almost forever. Much broader range in which you can measure. With optical techniques you have several lines, much more difficult to distinguish the peaks that you want to measure. It’s quick and multi-element. With aas using another lamp, new calibration liquid for every new element. With ms you can do it all at once, you only need one calibration solution. Instrument drift: the signal that you are measuring will gradually decrease in time. It’s completely normal that the signal decreases and that the concentration in different measured solutions was the same. We will see also how this comes, you can know how much the signal drift is and you can calculate back. It is quite significant. 28 The spectrum is always a function of mass to charge ratio. We will bring the element that we want to measure in a CHARGED STATE. Those elements we will ionize, we will use ICP, inductive coupled plasma. What is plasma ? Gas phase with only ions… 4th phase. This plasma is moving in a electro magnetic field. Those exited ions can collide with the incoming elements. They will charge them, the elements you want to measure. Used for: IDENTIFICATION and QUANTIFICATION. What is inside and how much. After a time, you will get a plateau with ASS. With mass the line is lineair and the range is way broader. Instrument drift ? The signal decreases after time. After one hour for example, the signal can be 10% lower after ten hours. There are tricks to calculate that drift. GENERAL PROCEDURE ◦ ◦ ◦ ◦ Atomise Conversion to ions (mostly +1) Seperation based on m/z ratio Counting the ions (plot relative intensity vs. m/z) They should be charged +1 , when they are charged +2 there will be interference!! Ions should be charged +1, then you can distinguish between the different isotopes. ATOM MASS: ◦ ◦ ◦ ◦ ◦ Concept mass different from other methods because here we can distinguish different isotopes Mass expressed in amu (atomic mass unit) or Dalton (Da). 1 amu = 1/12 mass 12C-isotope 1 amu = 1,6605387 x 10-27 kg Ar = relative atomic mass = how many times heavier than amu. In atomic mass spectrometry: use of exact mass and not mean mass (based on isotope composition). 29 For example: 12 C 6 protons and 6 neutrons, by definition this is 12 amu, ( atomic mass unit ) 6 _________________________________________________________ 13 C 6 protons and 7 neutrons, this is 13,0033 amu But C12 has an 6 abundance of 98,9% and C13 is 1.1%. SO the average is 12.01 amu, as seen in the PSE. Isotopes differ in neutrons but they have an equal amount of protons. 12/6 C: 6 protons and 6 neutrons: 12 amu 13/6 C: 6protons and 7 neutrons. 13,003355 amu. By definition, this has a mass of 12 atomic mass units. Based on the average composition: what is the amu of ‘C’? 12,01 amu because 98,9% C12 and 1,10% C13. It is based on the mass of the isotop itself. We use a diffeerent concept about mass! Isotope = dffer in neutrons but have an equal amount of protons ATOMS: MASS AND CHARGE ◦ ◦ ◦ ◦ Mass-charge ratio of an atomic (or molecular) ion m/z = ratio between Ar (or Mr) and ion charge Example: m/z of 12C1H4+ = 16,0313/1 = 16,0313. Usually charge is 1 and mass-charge ratio is just called “mass” GENERAL PRINCIPLE ◦ ◦ ◦ ◦ A mass spectrometer produces (starting from a sample) ions (usually with charge +1) That will be separated based on the mass-charge ratio (or just mass) The ions are produced by irradiation with electrons, photons, ions or molecules or even by thermal or electrical energy Negative ions can also be formed Working in vacuum bc you don’t want your ions to interact with parts in the air. That’s why there is a big vacuum pump. Icp is very fit to produce ions with a charge of +1. if you charge all of you atoms present in your sample, you need to separate them on their m/z ratio. You will use allways a very high vacuum. ◦ ◦ ◦ Mass analyzer = separation of the ions = spectrum Transducer = converting the ion beam in an electrical signal for treatment, storage (memory) or display (screen) Need for vacuum = no collisions 30 TYPES OF ATOMIC MASS SPECTROMETRY ◦ Quadrupole Mass Analyzers (most common type) You have 4 rods, AC and DC that are imposed on these rods, by combination of those 2 you can select one m/z ratio that will pass trough the quadrupole and then go to the transducer. 4 rods AC and DC, Split those 4 ods in 2 times 2 rods, the are place on a positive DC current + + The light ones will pass in negative (low-pass) Heavy ones will pass in positive region (high pass) combination fives a small window!! negative and positive change together so ther ewill always be a small region!! 31 2 times 2 rods, 2 rods will always be placed on a positive direct current and the 2 horizontal are negative. Positive are regarded as a high pass filter, the negative rods can be regarded as a light pass filter, all the light elements will pass trough the negative rods. If you find the right combination, you can select a very small window of masses that can pass trough all of the rods. You have a positive voltage: all the ions that are passing trough will always stay in the middle, they will not tend to go towards the positive sides. There is always a AC superimposed, the light elements are easily influenced by the AC because they have a small momentum and they will hit the rods because the rods will sometimes attreact them and sometimes not. Heavy elements will not be influenced because they have a very high momentum, they will pass. Therefore, the positive rods are a filter for the heavy elements. Otherwise for the negative rods, ions will be attracted towards the rods. AC will move away the light elements from the rods, heavy elements will just collide with the negative rods, light can pass so a low pass filter. If you combine those 2 you will always have a specific pass zone. The AC and DC will always increase together. Based on the combination of the AC and DC, you can select. 32 ◦ Time-of-Flight Mass Analyzers Shoot your ions trough a tube, if they all start together, they will spread out because of their different mass. Not used a lot anymore. ◦ Double-Focusing Analyzers Distinguish between different masses. First electrostatic analyser that will filter out some masses, then magnetic analyser, very good filter, based on the curve in a magnetic field you will select a mass/charge ratio. 33 DETECTION OF IONS = TRANSDUCER ◦ Overzicht methods: Electron multipliers Faraday cup Array transducers ◦ Electron multiplier: All the dynodes are on different potentials, will release an elctron, is then accelerated towards another dynode. The potential over the next one is higher than the previous one. It gains more energy, every electron that will hit, 2 wil be knocked out. You get an enormous amplification that you can then measure. Disadvantage: Signal decreases after some years. ◦ Faraday cup: 34 ◦ Array Transducers: Made up of large amount of transducers Allows detection of several elements at the same time. Microchannel Plates (voor EOID): Shoots ions endlessly and it goes a lot faster than using one multiplier. 35 INDUCTIVELY COUPLED PLASMA MASS SPECRTROMETRY ( ICPMS ) Most used system since 1980. Low detection limit, high selectivity, good precision and accuracy. Simple spectra with isotopes of the elements. Identification and quantification. Carrier gas will be argon most of the times, will be used to make the plasma. You mix your sample with carrier gas, then spray chamber. Torch is where the plasma is generated. All of these is just under atmospheric pressure, you are working in vacuum. You need an interface between the parts that are under atmospheric pressure and the parts that are under vacuum. The interfaces are cones with very small openings. Ionization: plasma Plasma: ◦ 4th state of matter ◦ Ionized gas ◦ Particles react strongly ◦ Usually very hot (6000 - 10000 K) We use a very hot plasma. All of the individual ions have a high energy content, will be used to ionise the elements that you want to determine. The gas that is used most is argon. Plasma will be made out of argon positive and argon. Plasma will be made out of argon positive and argon ion 36 Plasma in ICP-MS ◦ Mostly Ar: Ar+ and e◦ In torch: concentric tubes ◦ Seed electrons by sparc ◦ e- move in electric field and collide with Ar ◦ Sample introduction; nebulizer You have a second stream of plasma, you will also have a third stream of plasma, you need also a stream on the outer side to make sure that not everything gets heated. Ions will be sent trough the interface and then onto the quadrupole and then to the electron multiplier. Electrons will be accelerated and will collide with the argon, will ionise the argon atoms and will create the plasma. Stable plasma which you can send your analytes trough. You need to heat it up 5mins to make the plasma stable. Plasma ignition (video) You need to heat it up 5mins to make the plasma stable. (zie filmpje in slides!!) Plasma ionization efficiency For most elements It’s very easy to make 1+ ionic ions. For others: the ionization efficiency is not high, you will not make a lot of positively charged ions, you cannot measure a bunch of them. Argon very low ionization efficiency, still we use it for making our plasma. It’s the most difficult to be ionized, needs a lot of temperature. We need a very continuous stream of argon, no plasma without constant plasma. Every element that passes trough the plasma will be certainly ionized (hier worden dan de elementen in het draaggas argon bedoelt). The big power of icp-ms is that we have a very good ionization. 37 Interface ◦ Between plasma and detector (vacuum) ◦ Sample and skimmer cone ◦ Signal drift! (due to salt build-up) You always have 2 cones with a very small … Not all of the ions will be transferred to your quadrupole. Only around 0,1% will continue trough the cones. ◦ Signal drift due to salt build-up (signal suppression) Salt build-up on the cones, openings become narrower and if 10% less ions can pass, you signal will be 10% lower. You can correct for this, but why do we have these high salt concentrations? Acids that you use contain salts and they will all precipitate on the cones. You can dilute your samples and then your average salt concentration is lower. The picture is very extreme, if you want a better analysis with lesser drift, diluting. But if you really want to measure low concentrations, you cannot dilute and you just have to correct for the salt build up. Signal drift correction with internal standard Typical internal standards: Be, Ga, In , Tl Not common in samples to be analyzed Whole range of PSE We use an internal standard to correct for the signal drift. The amount of signal decrease is not only related to the salt build up, you should have internal standard that is always as close as possible to the element that you want to measure. Internal standard should not be present in the solution that you want to measure. These 4 do not occur that often in a normal environment. 38 Introduction of sample and internal standard Peristaltic pump, will be mixed and sent together to the plasma. You already have a stream of argon, this makes that at the tip of your nebulizer, your solution is nebulized, sprayed into torch with additional argon and also cooling argon. Mass spectrometer ◦ Typically quadrupole followed by electron multiplier 39 You also have a peak of argon at 40. if you cannot measure 40 calcium because of the peak of 40 argon, you can measure another isotope, calcium will have an isotope at eg 41. there you will have no interference, easier to distinguish, you can calculate back because you know the abundance. You have double focusing, higher resolution that eg a quadrupole. (zie filmpje in ICPMS in slides!) MASS SPECTROSCOPY and INTERFERENCE ◦ Advantage of ICP-MS methods (vs. spectroscopy) is simplicity of spectrum: only a few lines vs. several emission lines. You see a lot of peaks, it is difficult to attribute which peak is from which compound. Optical spectrum of a solution of 100 ppm Cerium Mass spectrum of a solution 10 ppm Cerium Sent trough an icp-ms, a lot clearer, but interferences are possible. Everything except from the Ce+’s are interferences. 40 ◦ BUT: there still are interferences in atomic MS: Non-spectroscopic matrix effects (drift due to salt build up) Spectroscopic effects due to the presence of ions with same mass (m/z) as researched ion: Isobaric ions = isotopes with identical mass Examples: 113 + In and 113Cd+ 58 + Ni and 58Fe+ Solution : Measure another isotope (e.g. 56Fe+) and from information of relative composition calculate the interfering isotope. Can often be done with software. 40 argon and 40 calcium, isotopes with identical mass. Polyatomic ions formation of polyatomic ions (such as 16O2+) with same mass as researched ion. Examples: 14 N2+ and 28Si+, or 75As+ and 75ArCl+ Doubly charged ions Example: 138 Ba2+ and 69Ga+ Measure another isotope Oxide and hydroxide ions most important source of interferences. Formation of oxides and hydroxides of the researched ions and/or the matrix. There are still interferences. First of all: non spectroscopic matrix effects: because of the present matrix in the sample, salt build up etc. Spectroscopic effects: effects that you see when you look at your spectrum. 4 different types. Ions that are made out of more than one atom. If you use HCl, Cl- will form ArCl+. Double charged: if you have high barium contect in your sample, it’s a problem, your signal will not always be the same due to the presence of barium or other double charged ions. 41 ◦ List of interferences: Know the elements you want to detect. You see what types of interferences that can occur for these elements . ◦ Dealing with polyatomic interferences: High resolution spectrometers Resolution (for mass spectrometers): R= m/m m en m+m= 2 particles with similar concentration that can just be separated (10% valley-definition) C2H4+; CH2N+ (28,0187); N2+(28,0061); CO+; mass 28 R= 28,0124/(28,0187 – 28,0061)~2200 The height of the valley between the two peaks can max be 10% of the total height of the peak. Spatial resolution (not this), this is a resolution for MS: average mass diviced by the distance (in mass) between the two peaks. 42 Collision cell technology (CCT) Introduction of gases causes ion-molecule (reaction gas) collisions and reactions Polyatomic ions or gas ions are broken up in parts or converted into atoms that do not interfere, mainly by chemical reactions or charge transfer; analyte ions pass almost unaffected NH3 can be used to eliminate many polyatomic interferences Ca: Ar+ + NH3= Ar + NH3+ Ca+: no reaction + + Fe: ArO + NH3 = Ar + O + NH3 Fe+: no reaction + + Cr: ArC + NH3= Ar + C + NH3 Cr+: no reaction Not ideal for all interferences, reactions with many analytes reduces sensitivity; single gas does not give best performance H2 also often used You build in an extra cell before you go to the quadrupole, there will be extra collisions (gas with polyatomic interferences). In the collision cell arO eg will collide with ammonia gas, you will have no more interference at m/z 56. interfering elements: ArO will be splitted. Gases can react with polyatoms or positively charged argon atoms, you split the polyatoms into individual atoms, will not interfere anymore with the elements that you want to measure. Collision cell Mostly placed before the quadrupole Physical chamber where your analytes are sent trough before entering the quadrupole. 43 Interference of ClO on measurement of V and Cr measured concentration (ppb) ◦ 160 140 120 100 80 V51 use of CCT (NH3/He = 6.1%) 60 Cr53 40 20 0 0 200 400 600 800 HCl in solution (ppm) Interference of the polyatom ClO. Solution that only contains HCl, no V or Cr present. It’s due to the polyatomic interferences ClO with m/z 51 and 53. there is only hcl in your solution. With cct you can eliminate those signals, you will measure nothing anymore. What is the influence of using the CCT on the signal? Interference of CaO on measurement of Fe measured concentration (ppb) ◦ 600.0 500.0 400.0 300.0 Fe56 use of CCT (H2/He = 3.5%) 200.0 100.0 0.0 0 50 100 150 200 250 CaO in solution (ppm) 44 ◦ Use: qualitive and semiquantitative applications Semiquantitative is not correct, you can really use it for quantitative applications. You need a calibration, for each element you have a different surface below the peak. It does not mean that if you know the surface below a peak, that you know the concentration, no, you need to calibrate. Each element has another ionization efficiency, depends on that what the surface is below the peak. Also how well is the mass analyser performing etc. You can use it for quantitative analysis. But you need a calibration before this! It is important to calibrate it first and then you can calculate the concentration that depends on the surface below the peak. Quantitative applications: concentration measurements using internal standards and calibration curves. Every element gives another count for 10 mg/ml e.g. you do not reach the plateau, you always have a straight line, you can easily extrapolate to higher concentrations. In AAS, you would say that you can only measure real concentrations that are within this limit, for ICPMS you can safely assume that you can measure concentrations that are a couple times higher than the highest measured concentration. You use internal standards and a calibration curve. Typically you get an outcome like on the right. Always 3 measurements, average, std, %std. beryllium and gallium are used for correcting the drift signal. Std only 1 percent so that is quite good 45 Measurements of isotope ratios. Used in carbon dating Applications: Archeology, geology (dating) Use of isotopes as tracers. Isotope rations interesting for geology/archeology, you cannot measure carbon with ICPMS, but you can use other elements to date some chemical materials. You can get to know the age Isotopes as tracers: important in radiochemistry, you can measure the activity with ICPMS. Liquid samples: Dilution with ultrapure water Conservation ( acidification ) Solid Samples: Dissolution (digestion, acids) and dilution Laser ablation 99% of the times you use liquid samples. To conserve your samples you always use acids (nitric acid). Hydrochlorid acid is not used due to the chlorine. Solids: you first need to dissolve them, ‘digestion’, then you will dilute them. Dilution step is important because if you use a lot of acids, there will be a lot of salts, and those will deposit and cause interference when too little dilution is provided. 46 Laser ablation (LA-ICP-MS) Powerful laserbeam to physically sputter some matter, those are then sent towards the plasma where they can be measured. You always have a vision system, you can look at your sample. Laser source will remove some of the solid material from your sample in very small particles, then send in solid phase to your ICP, then sent to quadrupole and detector. Laser Uv light (eg 213 nm ) Advantages: + Low detection limits + Small sample size ( less than 1 micro gram ) + Small spot size ( 5 – 300 micro grams ) + No sample preparation ( less matrix effects ) Laser uv light, you can do some spot analysis. If you do not know which one of the 2 is coal or pyrite, you can use the spot analysis. Especially helpful with precious compounds. You can direct your laser beam to one specific spot on your material, different from dissolving because you do not know where the material comes from with dissolving. You can really scan a surface and measure on different spots the concentration, you get a profile alongside you surface. 47 INTERMEZZO: ATOM EMISSION SPECTROMETRY Excitation of atoms from ground state Emission of light of specific wavelength Measurement of emitted light Excitation ◦ Flame (FAES) ◦ Plasma (ICP-AES) Analyzing the emission of EM waves in a visible light range. You always excite your atoms from the ground states, when they fall back they emit light of a specific wavelength, they will measure this light. Excitation with eg a flame, you do not need ionization. Flame or plasma can be used. In this case the most powerful way to do this is by using a plasma at much higher temperatures, you will be able to excite a lot more of those atoms. Flame : detection limit will not be very low. If you use plasma, your detection limit will be a lot lower. You don’t need ionisation, only excite them! That’s why we handle this chapter. Plasma is the best to excite them!! Plasma has much higer temperatures. 48 Atom Absorption Spectrometry (AAS) vs. Atom Emission Spectrometry (AES) Excite the calcium atom. AAS: you measure the amount of energy needed for the excitation of the ca-atom. With AES: you measure the emitted light when the excited ca-atom falls back to ground state. IMPORTANT TO REMEMBER! ◦ Flame-AES vs. ICP-AES: AES is the abbreviation of 2 techniques. Augier electron spectroscopy is also possible. If you work at the typical temps of flames, you won’t have a large excited fraction. Big power of using the plasma. You can see at the table, ratio of excited atoms/ atoms in ground state! You can see that when we work at a specific temperature from a flame the ratio isn’t that high. But if you look at if when used a plasma the ratio is higher!! 49 ◦ ICP-AES = ICP-OES You use your plasma to bring your atoms into the ionized state. Detection system is different, here you only need a system that can measure the emitted light. Nebulizer with argon to cool. Almost same setup than ICPMS. You only need a system now that can measure the light. ICP-AES Advantages and disadvantages Detection limits between AAS and ICP-MS Broad linear range (105), better than AAS, equal to ICPMS Capable of measuring Cl, S, P ( ICP-MS) Multi-element analysis More spectral interferences than ICP-MS Less chemical (polyatomic) interferences With ICP-MS, we could not measure Cl, S and P, you needed to bring them to a +1 state, the ionization efficiency is very low and thus difficult to bring them for the +1 state. Here, you only need to excite them, not +1 state. We can measure different elements in one solution at the same time. Much more difficult to interpret the different lines, more difficult to find the elements you are looking for. ICPMS: related to the ionisation! If difficult difficult to measure! For AES you only have to excite them! Lines difficult to seperate! 50 CHAPTER 3: MOLECULAR MASS SPECTROMETRY Already studied last year! Look at it again but we will look tot the different mechanics that can be used to bring the elements into ionised state! Compare different techniques is very important! We will not use a plasma because the temp is to high, you want information about the molecule! Base peak = highest peak! You use it as arefference for the other ones so you set this peak to 100 en set the other peaks into realtion. “Describe the different techniques to bring a molecule to an inionized state. “ Plot with the mass to charge ratio. We will plot the intensity. Only going to look at the different mechanisms that can be used to bring the elements into the ionized state. Some of these techniques can be used to analyse surfaces. Important here to be able to compare the different techniques to bring them into the ionized state. We will not use a plasma here, temp is too high, not wanted that you have dissociation. Plot the relative intensity to the m/z ratio. A lot of segmentation. APPLICATIONS: ◦ Mass Spectrometry (in general) gives information on: Elementary composition, Structure of (in)organic, biological compounds, Composition of mixtures (qualitative and quantitative), Structure and composition of solid surfaces, Isotope ratios. GENERAL PRINCIPLE: ◦ Example: ethylbenzene (C6H5CH2CH3) ◦ Compound (as vapour) is ionized using electron beam: C6H5CH2CH3 + e C6H5CH2CH3.+ + 2e ◦ Formation of a molecular ion (radical) in excited state. ◦ Relaxation of this excited state leads to breaking of the molecule: fragmentation (ex.: C6H5CH2+ by removal of CH3) ◦ Separation of ions based on m/z ratio. Ionized because the release of an electron. The molecule needs to release some of it’s energy. The large molecule will break up into several molecules. All of the fragments that remain will be positively charged. The ones that will be positively charged can be identified in your mass spectrum. 51 Mass Spectrum of ethylbenzene (normalized) You can use several mass analyzers. Base peak is used to compare. You set that one to 100. The other peaks will be compared with that. Just based on the removal of one electron from the compound, the original mass is the same to that of the original molecule. Base peak is the highest peak, what do we use it for? As a reference for the height of the other peaks, you set the abundance of the base peak to 100. Production of ions Important overview: ◦ ◦ ◦ Gas phase: first evaporation, then ionization Desorption: directly from solid or liquid phase. Hard Sources or Soft Sources: Hard: High energy produces ions in high excited states: relaxation leads to many fragments. Information about functional groups (structure). 52 Soft: Less fragmentation. Information about molecular mass. For hard ionization, you use agents that have a lot of energy, the ions that originate from hard energy sources, also have high excited states. If you have a molecule in a high excited state, the relaxation will lead to a lot of fragments, so a lot of peaks in you spectrum. Highly energetic ion can break into many different parts. Useful because some functional groups will always come from the parent ion, a lot of information about functional groups. Soft: less fragmentation, only splitting of one hydrogen atom e.g., this brings us information about MM because the peaks will always be very close to the molecular ion peak. Spectrum will not contain many peaks, concentrated around the molecular ion peak. Know the difference! Hard: use agents that produce or have a lot of energy, high excited states. Molecules highly excited states, the relaxation will release many fragments . Spectrum will have a lot of peaks. many peaks with different….. information of functional groups Soft: information of full molecule ◦ HARD vs SOFT IONIZATION For the hard ionization, you do not see the MM peak anymore. You would never guess that the molecule that you are analysing would be one-decanol. If you have a soft ionization technique, a lot less peaks. 53 ◦ Electron Impact Source Gas Phase Method: sample must be thermally stable and MM not too high: less than 1000 g / mol. Gas phase method, simple to understand. The easiest way to make ions, important you have to bring your analyte into the gas phase, they should be thermally stable. Your MM can not be too high, it can require too much energy then. You have you analyte on the left, then you have filament that is heated, then placing them into an electric field. Electrons will be generated in the filaments and will be accelerated towards the anode. You will have the cloud of analytes and the electrons will collide with analytes and you get excited ions. Then towards mass analyser. This is a hard ionization source. Afterwards the breaking happens before they go to the analyser. The ionization will lead to one molecular ion that then will relax and split into a lot of different smaller molecules. Easiest way to produce ions. In gas phase (stable!), molecular mass should be below 1000g/mol. Before they go to the mass analyser the excited ions falls back. Fragmentation is an advantage but sometimes the molecular peak disappears completely Learns you a lot about the different functional groups, molecular peak disappears sometimes. 2nd graph: molecular ion peaks are gone. 54 ◦ Other method: Chemical Ionization Gaseous molecules are ionized using ions (formed out of a gas using electrons) ex.: methane. Advantage: soft method. Ionization of reagent gas CH4 + e- → CH4+ + 2 e CH4+ + CH4 → CH5+ + CH3 CH3 + CH4+ → CH3+ + CH4 CH3+ + CH4 → C2H5+ + H2 Reaction with analyte CH5+ + MH → MH2+ + CH4 C2H5+ + MH → M+ + C2H6 You use a gaseous molecule that is first ionized. First ionise your methane, then this methane will ionize your sample. All kinds of different types of ionized methane molecules. This step with methane is necessary because if you directly use electrons, you are in the electron impact mode. Here you are going to capture first a part of the energy, the ionized methane molecules have a lot lower energy content, this is thus a soft ionization method. If you want information about MM, you will use a method like this. 55 ◦ Other methods: Field Ionisation Field Ionization = formation of ions in a strong electrical field around a “Carbon Microneedle Emitter” When EI can’t be used Soft ionization Why that shape ? More surface. No direct contact between ions and needles. Important is that the reaction occur at the tip. You need a very fine tip. Why is that ? Specific surface is large. Electrical field: voltage applied over a surface. You will have a very high Ef at the tip. Positive charge at the tip of those needles. It will extract one electron from the analyte molecule. Field ionization, use a strong electric field around a carbon microneedle emitter. The needle is formed like that to have more surface, the interactions occur at the very tips of the needle. Specific surface there is very large, the definition of an electrical field is the voltage applied over a specific surface, there is a large electrical field there. We will bring the molecules in the neighbourhood of those tips, the positive tips will extract one negatively charged electron from the orbital of the atom. The atom becomes positively charged. 56 ◦ Desorption methods For materials that don’t evaporate or thermally unstable. Apply energy in solid or liquid and remove ions directly Results in simple spectra Example: MALDI Matrix-Assisted Laser Desorption Ionization New technique for solids and liquids. We apply energy directly to the solid or liquid. The power of the laser is too strong to send it directly to the compounds, we have to make sure that we keep the integrity of the molecules and do not split them into atoms. Therefore we use a matrix, to absorb a part of the laser energy to make sure that the molecules are still molecules after they are hit. They also prevent that your analytes agglomerate and they protect the destruction of your analytes. MALDI Matrix used to Absorb laser energy Prevent analyte agglomeration Protect analyte destruction by laser beam 57 Electrospray ionization: LC-MS Electrospray, mostly used in LC. You have a voltage over the solution where your analytes are dissolved in, also a counterelectrode. The needletip will spray the solution with your analytes into a vacuum zone, they will form very large droplets, many analytes will be dissolved into a matrix. On the outside there will be the positive charge, the solvent that you use will start to evaporate due to the heat. The surface charge will become higher and higher and at a given time, the surface charge will become too high. The droplet will break into several smaller droplets. Until you only have your analyte left which is now an ion. Field desorption Comparable to field ionization Secondary-ion Mass Spectrometry (SIMS) See later (measurement of surfaces) Plasma desorption Fast atom bombardment Thermospray ionization 58 Applications of Mass Spectrometry Mostly organic molecules, not so important. ◦ ◦ ◦ Identification of pure compounds Determination of MM, Determination of Molecular formulas from mass determination, Determination of Molecular formulas from isotope ratios, Identification using comparison spectra (database), Analysis of mixtures Coupled to gas chromatography Coupled to capilary elektrophoresis Tandem mass spectrometry Quantitative determination in complex mixtures Organic (petroleum, pharma, bio…) Polymers, PMMA 59 CHAPTER 4 : STUDY OF SURFACES 3D-pictures Overview Part1: Spectrometry ◦ Atomic X-Ray Spectrometry (XRF) ◦ Molecular X-Ray Spectrometry (XRD) ◦ Atomic Mass Spectrometry (ICP-MS) ◦ Molecular Mass Spectrometry (…-MS) ◦ (Raman Spectrometry) ◦ (NMR-Spectrometry) Part 2: The study of Surfaces ◦ Electron Spectroscopy (AES, XPS) ◦ Ion Spectroscopy (SIMS) ◦ Surface Photon Spectroscopy ◦ Electron-stimulated micro-analysis methods (SEM/TEM) ◦ Scanning Probe Microscopy (STM/AFM) ◦ Surface area analysis Part 3 ◦ ◦ ◦ ◦ Particle Size Determination Voltammetry Radiochemistry … 60 Part 2: Study of surfaces: Classification of surface analysis techniques ◦ Spectroscopic: detected (secondary) beam: Electron (AES, XPS, EELS) Ion (SIMS, RBS) Photon (ellipsometry, SPR) ◦ Microscopy: primary beam Electron-stimulated methods (SEM, TEM) ◦ 3D techniques Scanning Probe Microscopy (STM, AFM) ◦ Other Surface area analysis (BET) DEFENITION OF SURFACE ◦ Boundary between solid and Vacuum, Gas Liquid ◦ That part of a solid that has a different composition than the average bulk. ◦ Not only the top layer but also underlying transition layers with gradually changing composition. ◦ That part that is observed using a particular technique. Examples: oxidation layer, fouling catalyst: ◦ ◦ Before 1950 Methods to determine the physical properties of a surface: Images, Adsorption isotherms, Roughness Reflectivity … After 1950 Also measurements of chemical compostion using spectroscopic method. METHODS FOR SURFACE MEASUREMENTS ◦ ◦ Qualitative and quantitative CHEMICAL determinations of surface Dimensions: from a few tenths of nanometers to several nanometers 61
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )